Zimmer Biomet, Inc.: 875 FDA recall reports
Every medical device recall the FDA has published for Zimmer Biomet, Inc.. The fault, and how the agency graded it.
875Reports
3Class I
0Ongoing
2020-12-30Most recent
3 Class I on this pageClass I means the FDA judged there was a reasonable probability of serious harm or death. Check the lot numbers against anything you hold.
What the classes mean
The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.
Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.
By class
By year reported
Checking whether it affects you
The lot or code number is what matters. Not the product name. A recall covers
specific lots. Yours may not be one of them.
Check the code on your package against the report. The FDA keeps the current
list at fda.gov/safety/recalls.
The reports
The text below is quoted from the FDA record word for word.
Spinal Rod Cutter, Item number 00392500200. The product is a nonsterile pin cutter - Product Usage: The intend
Z-0696-2021 · Class II · reported 2020-12-30 · Terminated
- Why it was recalled
- Device has the potential for fracture during use.
- Product
- Spinal Rod Cutter, Item number 00392500200. The product is a nonsterile pin cutter - Product Usage: The intended use of this instrument is to be used for cutting orthopedic rods/pins during orthopedic surgeries.
- Where it went
- Domestic distribution nationwide. Foreign distribution to Australia, China, Hong Kong, India, Japan, Korea, New Zealand, Taiwan, Singapore, Thailand, Canada, Netherlands, Argentina, Brazil, Mexico, Venezuela, Panama, and Chile.
- How much
- 4713
Polaris 5.5 Spinal System and Polaris 5.5/Cypher Spinal System, Thoracolumbosacral pedicle screw system
Z-0570-2021 · Class II · reported 2020-12-23 · Terminated
- Why it was recalled
- Zimmer Biomet is conducting a medical device correction to update the Instructions for Use (IFU) and the Surgical Techniques for the Polaris Spinal System Translation Screw and Cypher MIS Screw System. Users should be aware that changes were made to the contraindications section indicating that the Translation Screws should not be used with a direct current stimulation, as a high probability of construct failure exists when using Translation Screws containing CoCr with direct current simulation system.
- Product
- Polaris 5.5 Spinal System and Polaris 5.5/Cypher Spinal System, Thoracolumbosacral pedicle screw system
- Where it went
- Domestic distribution nationwide. As direct stimulation devices are only available in the US, this contraindication does not apply to international customers.
- How much
- 57,715 units
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Neutral, O.D. Cup with Spacers, 5
Z-0344-2021 · Class II · reported 2020-11-11 · Terminated
- Why it was recalled
- Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
- Product
- ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Neutral, O.D. Cup with Spacers, 55 mm Item Number: 00-8065-552-32
- Where it went
- US Nationwide
- How much
- 412 units total
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, 10 degree inclined face, O.D. Cup
Z-0368-2021 · Class II · reported 2020-11-11 · Terminated
- Why it was recalled
- Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
- Product
- ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, 10 degree inclined face, O.D. Cup with Spacers, 51 mm Number: 00-8065-648-26
- Where it went
- US Nationwide
- How much
- 412 units total
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Snap-In O.D. Cup with Spacers, 55
Z-0394-2021 · Class II · reported 2020-11-11 · Terminated
- Why it was recalled
- Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
- Product
- ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Snap-In O.D. Cup with Spacers, 55 mm Number: 00-8065-952-32
- Where it went
- US Nationwide
- How much
- 412 units total
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, 10 degree inclined face, O.D. Cup
Z-0363-2021 · Class II · reported 2020-11-11 · Terminated
- Why it was recalled
- Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
- Product
- ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, 10 degree inclined face, O.D. Cup with Spacers, 49 mm Number: 00-8065-646-22
- Where it went
- US Nationwide
- How much
- 412 units total
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, Neutral, O.D. Cup with Spacers, 47
Z-0328-2021 · Class II · reported 2020-11-11 · Terminated
- Why it was recalled
- Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
- Product
- ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, Neutral, O.D. Cup with Spacers, 47 mm Item Number: 00-8065-544-28
- Where it went
- US Nationwide
- How much
- 412 units total
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, 10 degree inclined face, O.D. Cup
Z-0370-2021 · Class II · reported 2020-11-11 · Terminated
- Why it was recalled
- Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
- Product
- ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, 10 degree inclined face, O.D. Cup with Spacers, 51 mm Number: 00-8065-648-32
- Where it went
- US Nationwide
- How much
- N/A
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, 10 degree inclined face, O.D. Cup
Z-0386-2021 · Class II · reported 2020-11-11 · Terminated
- Why it was recalled
- Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
- Product
- ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, 10 degree inclined face, O.D. Cup with Spacers, 59 mm Number: 00-8065-656-32
- Where it went
- US Nationwide
- How much
- 412 units total
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Neutral, O.D. Cup with Spacers, 5
Z-0348-2021 · Class II · reported 2020-11-11 · Terminated
- Why it was recalled
- Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
- Product
- ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Neutral, O.D. Cup with Spacers, 57 mm Number: 00-8065-554-32
- Where it went
- US Nationwide
- How much
- 412 units total
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, Neutral, O.D. Cup with Spacers, 45
Z-0325-2021 · Class II · reported 2020-11-11 · Terminated
- Why it was recalled
- Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
- Product
- ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, Neutral, O.D. Cup with Spacers, 45mm Item Number: 00-8065-542-26 Hip prosthesis component
- Where it went
- US Nationwide
- How much
- 412 units total
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, 10 degree inclined face, O.D. Cup
Z-0374-2021 · Class II · reported 2020-11-11 · Terminated
- Why it was recalled
- Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
- Product
- ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, 10 degree inclined face, O.D. Cup with Spacers, 53 mm Number: 00-8065-650-32
- Where it went
- US Nationwide
- How much
- 412 units total
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Snap-In O.D. Cup with Spacers, 59
Z-0396-2021 · Class II · reported 2020-11-11 · Terminated
- Why it was recalled
- Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
- Product
- ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Snap-In O.D. Cup with Spacers, 59 mm Number: 00-8065-956-32
- Where it went
- US Nationwide
- How much
- 412 units total
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, 10 degree inclined face, O.D. Cup
Z-0357-2021 · Class II · reported 2020-11-11 · Terminated
- Why it was recalled
- Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
- Product
- ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, 10 degree inclined face, O.D. Cup with Spacers, 43 mm Number: 00-8065-640-22
- Where it went
- US Nationwide
- How much
- 412 units total
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, 10 degree inclined face, O.D. Cup
Z-0371-2021 · Class II · reported 2020-11-11 · Terminated
- Why it was recalled
- Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
- Product
- ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, 10 degree inclined face, O.D. Cup with Spacers, 53 mm Number: 00-8065-650-22
- Where it went
- US Nationwide
- How much
- 412 units total
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, Neutral, O.D. Cup with Spacers, 47
Z-0327-2021 · Class II · reported 2020-11-11 · Terminated
- Why it was recalled
- Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
- Product
- ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, Neutral, O.D. Cup with Spacers, 47 mm Item Number: 00-8065-544-26
- Where it went
- US Nationwide
- How much
- 412 units total
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, 10 degree inclined face, O.D. Cup
Z-0381-2021 · Class II · reported 2020-11-11 · Terminated
- Why it was recalled
- Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
- Product
- ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, 10 degree inclined face, O.D. Cup with Spacers, 57 mm Number: 00-8065-654-28
- Where it went
- US Nationwide
- How much
- 412 units total
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, Neutral, O.D. Cup with Spacers, 5
Z-0351-2021 · Class II · reported 2020-11-11 · Terminated
- Why it was recalled
- Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
- Product
- ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, Neutral, O.D. Cup with Spacers, 59 mm Number: 00-8065-556-28
- Where it went
- US Nationwide
- How much
- 412 units total
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22mm, Neutral, O.D. Cup with Spacers, 51
Z-0333-2021 · Class II · reported 2020-11-11 · Terminated
- Why it was recalled
- Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
- Product
- ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22mm, Neutral, O.D. Cup with Spacers, 51 mm Item Number: 00-8065-548-22
- Where it went
- US Nationwide
- How much
- 412 unit total
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, Neutral, O.D. Cup with Spacers, 57
Z-0346-2021 · Class II · reported 2020-11-11 · Terminated
- Why it was recalled
- Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
- Product
- ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, Neutral, O.D. Cup with Spacers, 57 mm Number: 00-8065-554-26
- Where it went
- US Nationwide
- How much
- 412 units total
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, 10 degree inclined face, O.D. Cup
Z-0364-2021 · Class II · reported 2020-11-11 · Terminated
- Why it was recalled
- Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
- Product
- ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, 10 degree inclined face, O.D. Cup with Spacers, 49 mm Number: 00-8065-646-26
- Where it went
- US Nationwide
- How much
- 412 units total
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, 10 degree inclined face, O.D. Cup
Z-0388-2021 · Class II · reported 2020-11-11 · Terminated
- Why it was recalled
- Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
- Product
- ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, 10 degree inclined face, O.D. Cup with Spacers, 61 mm Number: 00-8065-658-26
- Where it went
- US Nationwide
- How much
- 412 units total
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, 10 degree inclined face, O.D. Cup
Z-0362-2021 · Class II · reported 2020-11-11 · Terminated
- Why it was recalled
- Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
- Product
- ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, 10 degree inclined face, O.D. Cup with Spacers, 47 mm Number: 00-8065-644-28
- Where it went
- US Nationwide
- How much
- 412 units total
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Neutral, O.D. Cup with Spacers, 5
Z-0352-2021 · Class II · reported 2020-11-11 · Terminated
- Why it was recalled
- Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
- Product
- ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Neutral, O.D. Cup with Spacers, 59 mm Number: 00-8065-556-32
- Where it went
- US Nationwide
- How much
- 412 units total
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, Neutral, O.D. Cup with Spacers, 43
Z-0323-2021 · Class II · reported 2020-11-11 · Terminated
- Why it was recalled
- Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
- Product
- ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, Neutral, O.D. Cup with Spacers, 43mm Item Number: 00-8065-540-22 Hip prosthesis component
- Where it went
- US Nationwide
- How much
- 412 units total
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, 10 degree inclined face, O.D. Cup
Z-0365-2021 · Class II · reported 2020-11-11 · Terminated
- Why it was recalled
- Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
- Product
- ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, 10 degree inclined face, O.D. Cup with Spacers, 49 mm Number: 00-8065-646-28
- Where it went
- US Nationwide
- How much
- 412 units total
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, 10 degree inclined face, O.D. Cup
Z-0387-2021 · Class II · reported 2020-11-11 · Terminated
- Why it was recalled
- Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
- Product
- ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, 10 degree inclined face, O.D. Cup with Spacers, 61 mm Number: 00-8065-658-22
- Where it went
- US Nationwide
- How much
- 412 units total
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, Neutral, O.D. Cup with Spacers, 61
Z-0355-2021 · Class II · reported 2020-11-11 · Terminated
- Why it was recalled
- Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
- Product
- ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, Neutral, O.D. Cup with Spacers, 61 mm Number: 00-8065-558-28
- Where it went
- US Nationwide
- How much
- 412 units total
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Snap-In O.D. Cup with Spacers, 57
Z-0395-2021 · Class II · reported 2020-11-11 · Terminated
- Why it was recalled
- Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
- Product
- ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Snap-In O.D. Cup with Spacers, 57 mm Number: 00-8065-954-32
- Where it went
- US Nationwide
- How much
- 412 units total
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Snap-In O.D. Cup with Spacers, 51m
Z-0392-2021 · Class II · reported 2020-11-11 · Terminated
- Why it was recalled
- Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
- Product
- ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Snap-In O.D. Cup with Spacers, 51mm Number: 00-8065-948-32
- Where it went
- US Nationwide
- How much
- 412 units total
Showing the 30 most recent of 875. The rest are searchable at the FDA enforcement report database.
Read these counts with three things in mind
A firm with more recalls usually makes more products. A contract packer runs
hundreds of lines. It will out-recall a company with one product whatever either
does. This is not a safety ranking.
Most recalls here are voluntary. A firm found its own fault and pulled the stock.
That is the system working. It lands in this database looking the same as a recall
the FDA had to push for.
Terminated means the FDA considers the recall finished. It is not a verdict.
It does not mean anyone was harmed.
Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.
← All device recalls