Ventana Medical Systems Inc
Ventana Medical Systems Inc: 41 FDA recall reports
Every medical device recall the FDA has published for Ventana Medical Systems Inc. The fault, and how the agency graded it.
41Reports
33Class I
4Ongoing
2024-08-28Most recent
33 Class I on this pageClass I means the FDA judged there was a reasonable probability of serious harm or death. Check the lot numbers against anything you hold.
What the classes mean
The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.
Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.
By class
By year reported
Checking whether it affects you
The lot or code number is what matters. Not the product name. A recall covers
specific lots. Yours may not be one of them.
Check the code on your package against the report. The FDA keeps the current
list at fda.gov/safety/recalls.
The reports
The text below is quoted from the FDA record word for word.
Ventana anti-CD10 (SP67) Rabbit Monoclonal Primary Antibody, REF: 790-4506, IVD
Z-2644-2024 · Class II · reported 2024-08-28 · Ongoing
- Why it was recalled
- Their is a potential for high background and off-target staining when using the monoclonal primary antibody which can result in false positive results.
- Product
- Ventana anti-CD10 (SP67) Rabbit Monoclonal Primary Antibody, REF: 790-4506, IVD
- Where it went
- Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and the countries of Austria, Belgium, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Ireland, Italy, Latvia, Lithuania, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Albania, Argentina, Armenia, Australia, Azerbaijan, Bahrain, Belarus, Brazil, Cameroon, Chile, China, Colombia, Costa Rica, Cuba, Dominican Republic, Ecuador, Egypt, El Salvador, Georgia, Ghana, Guatemala, Honduras, Hong Kong, India, Iran, Iraq, Israel, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Malaysia, Mali, Mauritania, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, New Zealand, Nicaragua, Nigeria, North Macedonia, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Russian Fed., Rwanda, Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Sri Lanka, Taiwan, Tanzania, Thailand, T¿rkiye, Uganda, United Arab Emirates, United Kingdom,...
- How much
- 22,839 kits
VENTANA PD-L1 (SP142) VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 07709374001 VENTANA PD-L1 (SP142) As
Z-0907-2023 · Class II · reported 2023-01-18 · Ongoing
- Why it was recalled
- Internal comparison studies revealed unacceptable, light staining. Light staining affects the borderline of positive versus negative test results. It can cause a reduction in both staining intensity and the percent positivity, which could potentially lead to a false-negative PD-L1 result.
- Product
- VENTANA PD-L1 (SP142) VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 07709374001 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 08008540001 VENTANA PD-L1 (SP142) IHC Assay (CE IVD Analytical) 07011571001
- Where it went
- Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AZ, CA, CO,CT, DC, DE, FL, GA, IL, IN, KS, KY, MA, MD, MI, MN, NC, ND, NE, NJ, NM, NV, NY, OH, OR, PA, Puerto Rico, SC, TN, TX, UT, VA, VT, WA, and WI. The countries of Canada, Austria, Belgium, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Ireland, Italy, Latvia, Lithuania, Luxembourg, Netherlands, Norway, Poland, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland and United Kingdom, Albania, Armenia, Bangladesh, Belarus, Costa Rica, Cote d'Ivoire, Cuba, Dominican Republic, Egypt, El Salvador, Ghana, Guatemala, Iran, Israel, Jordan, Kenya, Kosovo, Kuwait, Lebanon, Mauritius, Morocco, Myanmar, Nigeria, North Macedonia, Oman, Panama, Philippines, Russian Fed., Saudi Arabia, Serbia, South Africa, Tanzania, Turkey, Turkmenistan and United Arab Emirates. Update (04-Jan-2023): Argentina, Australia, Brazil, Chile, China, Colombia, Ecuador, Hong Kong, Honduras, Indonesia, India, Japan, Korea, Mexico, Malaysia, New Zealand, Peru, Pakistan, Paraguay, Singapore, Thailand, Taiwan, Uruguay and Vietnam.
- How much
- 8,927 kits
Ventana HE 600 System, automated slide preparer, for use in laboratories.
Z-1374-2022 · Class II · reported 2022-07-20 · Ongoing
- Why it was recalled
- There is a potential for fluid leak inside the Ventana HE 600 instrument which could result in electrical short and fire
- Product
- Ventana HE 600 System, automated slide preparer, for use in laboratories.
- Where it went
- US: AL AR AZ CA CO CT DC FL GA HI IA IL IN KS KY MA MD MI MN MO MS MT NC NE NJ NY OH OK OR PA PR SC SD TN TX UT VA WA WI WV OUS: Canada, Austria, Belgium, Croatia, Denmark, Finland, France, Germany, Greece, Ireland, Italy, Luxembourg, Netherlands, Norway, Poland, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey, UK, Argentina, Australia, Brazil, Chile, China, Colombia, Costa Rica, Ecuador, Egypt, El Salvador, Guatemala, Hong Kong, Israel, Japan, Kuwait, Lebanon, Malaysia, Saudi Arabia, Singapore, South Korea, Taiwan, Thailand, United Arab Emirates (UAE), Vietnam
- How much
- 763 instruments
BenchMark ULTRA and DISCOVERY ULTRA Instruments
Z-0894-2022 · Class II · reported 2022-04-13 · Ongoing
- Why it was recalled
- Potential for Fluid Leak inside a staining system that could cause an electrical short circuit of the General Purpose Input/ Output Board's J1 Connector that could result in smoke and fire.
- Product
- BenchMark ULTRA and DISCOVERY ULTRA Instruments
- Where it went
- US: CA, IL, PA, FL, AZ, NY, OH, CO, AR, NC, MD, AL, AK, LA, IA, MN, MA, ND, IN, TX, TN, CT, SC, GA, UT, WI, VA, SD, KY, MS, MO, DE, MI, MT, NJ, WV, OK, WA, OR, PR, DC, NM, ID, NH, NE, ME, KS, HI, NV, RI, WY, VT OUS: Australia, Austria, Belgium, Canada, China, Croatia, Denmark, Finland, France, Germany, Hong Kong, Hungary, India, Israel, Italy, Japan, Netherlands, Norway, Poland, Portugal, Russian Fed., Singapore, South Korea, Spain, Sweden, Switzerland, United Kingdom, Algeria, Argentina, Aruba, Bosnia-Herzegovina., Brazil, Bulgaria, Chile, Colombia, Costa Rica, Cuba, Cyprus, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Greece, Indonesia, Ireland, Kazakhstan, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Macedonia, Malaysia, Martinique, Mexico, Monaco, Morocco, Namibia, New Zealand, Oman, Pakistan, Panama, Peru, Philippines, Reunion, Romania, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, Vietnam, White Russia, Zimbabwe
- How much
- 9214
Model Number: 06917259001 UDI: 04015630976010 - Product Usage: The System s intended use environment inclu
Z-2262-2020 · Class II · reported 2020-06-17 · Completed
- Why it was recalled
- The firm became aware of a cleaning solution leaking issue from the middle staining module of the Ventana HE 600 instrument, used for histologic section samples in anatomic pathology laboratory settings, which caused fluid to travel down to the electrical wiring connector of the Coverslipper solvent shield circuit board. This can cause an electrical short and result in burned components.
- Product
- Model Number: 06917259001 UDI: 04015630976010 - Product Usage: The System s intended use environment includes anatomic pathology (AP) laboratories and histology facilities within hospital, clinic, reference, or private lab settings. The System is to be operated by qualified users in
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, MD, MI, MA, MN, MO, MS, MT, NC, NE, NJ, NY, OH, OK, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI and the countries of Australia, Austria, Belgium, Brazil, Canada, Chile, China, Columbia, Costa Rica, Croatia, Denmark, Ecuador, El Salvador, Finland, France, Germany, Hong Kong, Ireland, Italy, Japan, Kuwait, Lebanon, Malaysia, Norway, Peru, Poland, Saudi Arabia, Slovakia, Slovenia, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, Vietnam.
- How much
- 461
Class I
INFORM HPV III Fam 16 Probe - US Export, Catalog Number 05278856001, model 800-4295 Immunohistochemistry (I
Z-3014-2018 · Class I · reported 2018-09-26 · Terminated
- Why it was recalled
- Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.
- Product
- INFORM HPV III Fam 16 Probe - US Export, Catalog Number 05278856001, model 800-4295 Immunohistochemistry (IHC) for in vitro diagnostic use.
- Where it went
- worldwide
- How much
- 195 units
Class I
VENTANA anti-Helicobacter Pylori (SP48) PAB 250, Catalog Number 06425623001, model 790-1015 Immunohistochem
Z-3010-2018 · Class I · reported 2018-09-26 · Terminated
- Why it was recalled
- Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.
- Product
- VENTANA anti-Helicobacter Pylori (SP48) PAB 250, Catalog Number 06425623001, model 790-1015 Immunohistochemistry (IHC) for in vitro diagnostic use.
- Where it went
- worldwide
- How much
- 2753 units
Class I
iView DAB Detection Kit, Catalog Number 05266157001, model 760-091 Immunohistochemistry (IHC) for in vitro di
Z-2998-2018 · Class I · reported 2018-09-26 · Terminated
- Why it was recalled
- Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.
- Product
- iView DAB Detection Kit, Catalog Number 05266157001, model 760-091 Immunohistochemistry (IHC) for in vitro diagnostic use.
- Where it went
- worldwide
- How much
- 3068 units
Class I
ANTI-PAN Keratin Primary Antibody, 25 mL, Catalog Number 05266840001, model 760-2135 Immunohistochemistry
Z-3003-2018 · Class I · reported 2018-09-26 · Terminated
- Why it was recalled
- Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.
- Product
- ANTI-PAN Keratin Primary Antibody, 25 mL, Catalog Number 05266840001, model 760-2135 Immunohistochemistry (IHC) for in vitro diagnostic use.
- Where it went
- worldwide
- How much
- 354 units
Class I
VENTANA ISH iView Blue Detection Kit, Catalog Number 05278511001, model 800-092 Immunohistochemistry (IHC)
Z-3008-2018 · Class I · reported 2018-09-26 · Terminated
- Why it was recalled
- Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers
- Product
- VENTANA ISH iView Blue Detection Kit, Catalog Number 05278511001, model 800-092 Immunohistochemistry (IHC) for in vitro diagnostic use.
- Where it went
- worldwide
- How much
- 559 units
Class I
iView DAB Detection Kit w/o Secondary Antibody, Catalog Number 05266173001, Model 760-093. For laboratory
Z-3001-2018 · Class I · reported 2018-09-26 · Terminated
- Why it was recalled
- Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.
- Product
- iView DAB Detection Kit w/o Secondary Antibody, Catalog Number 05266173001, Model 760-093. For laboratory use.
- Where it went
- worldwide
- How much
- 10 units
Class I
CINtec Histology (250), Catalog Number 06680011001, model 725-4793 Immunohistochemistry (IHC) for in vitro d
Z-3011-2018 · Class I · reported 2018-09-26 · Terminated
- Why it was recalled
- Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.
- Product
- CINtec Histology (250), Catalog Number 06680011001, model 725-4793 Immunohistochemistry (IHC) for in vitro diagnostic use.
- Where it went
- worldwide
- How much
- 120 units
Class I
ISH Protease 3, Catalog Number 05273331001, model 780-4149 Immunohistochemistry (IHC) for in vitro diagnos
Z-3005-2018 · Class I · reported 2018-09-26 · Terminated
- Why it was recalled
- Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.
- Product
- ISH Protease 3, Catalog Number 05273331001, model 780-4149 Immunohistochemistry (IHC) for in vitro diagnostic use.
- Where it went
- worldwide
- How much
- 2231 units
Class I
NEXES VEN IVIEW DAB DET KT JPN-US EXPORT, Catalog Number 05266084001 Model 760-041 Immunohistochemistry (IH
Z-3015-2018 · Class I · reported 2018-09-26 · Terminated
- Why it was recalled
- Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.
- Product
- NEXES VEN IVIEW DAB DET KT JPN-US EXPORT, Catalog Number 05266084001 Model 760-041 Immunohistochemistry (IHC) for in vitro diagnostic use.
- Where it went
- worldwide
- How much
- 2172 units
Class I
CONFIRM anti-Progesterone Receptor (PR) (1E2) Rabbit Monoclonal Primary Antibody, Catalog Number 05278392001,
Z-3006-2018 · Class I · reported 2018-09-26 · Terminated
- Why it was recalled
- Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.
- Product
- CONFIRM anti-Progesterone Receptor (PR) (1E2) Rabbit Monoclonal Primary Antibody, Catalog Number 05278392001, model 790-4296 Immunohistochemistry (IHC) for in vitro diagnostic use.
- Where it went
- worldwide
- How much
- 1655 units
Class I
ultraView SISH DNP Detection Kit US, Catalog Number 05572037001, model 760-098 Immunohistochemistry (IHC) f
Z-2999-2018 · Class I · reported 2018-09-26 · Terminated
- Why it was recalled
- Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.
- Product
- ultraView SISH DNP Detection Kit US, Catalog Number 05572037001, model 760-098 Immunohistochemistry (IHC) for in vitro diagnostic use.
- Where it went
- worldwide
- How much
- 96 units
Class I
OptiView Amplification Kit (250 test), Catalog Number 06718663001, model 860-099 Immunohistochemistry (IHC
Z-3000-2018 · Class I · reported 2018-09-26 · Terminated
- Why it was recalled
- Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.
- Product
- OptiView Amplification Kit (250 test), Catalog Number 06718663001, model 860-099 Immunohistochemistry (IHC) for in vitro diagnostic use.
- Where it went
- worldwide
- How much
- 461 units
Class I
ultraView Universal Alkaline Phosphatase Red Detection Kit, Catalog Number 05269814001, model 760-501 Immu
Z-3004-2018 · Class I · reported 2018-09-26 · Terminated
- Why it was recalled
- Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.
- Product
- ultraView Universal Alkaline Phosphatase Red Detection Kit, Catalog Number 05269814001, model 760-501 Immunohistochemistry (IHC) for in vitro diagnostic use.
- Where it went
- worldwide
- How much
- 8211 units
Class I
VENTANA Basal Cell Cktl, (34BE12+p63)250, Catalog Number 06419445001, Model 790-1010 Immunohistochemistry
Z-3009-2018 · Class I · reported 2018-09-26 · Terminated
- Why it was recalled
- Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.
- Product
- VENTANA Basal Cell Cktl, (34BE12+p63)250, Catalog Number 06419445001, Model 790-1010 Immunohistochemistry (IHC) for in vitro diagnostic use.
- Where it went
- worldwide
- How much
- 75 units
Class I
CINtec p16 Histology (250), Catalog Number 06695230001, Model 725-4713 Immunohistochemistry (IHC) for in vitr
Z-3012-2018 · Class I · reported 2018-09-26 · Terminated
- Why it was recalled
- Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.
- Product
- CINtec p16 Histology (250), Catalog Number 06695230001, Model 725-4713 Immunohistochemistry (IHC) for in vitro diagnostic use.
- Where it went
- worldwide
- How much
- 468 units
Class I
ultraView Universal DAB Detection Kit, Catalog Number 05269806001, model 760-500 Immunohistochemistry (IHC
Z-2997-2018 · Class I · reported 2018-09-26 · Terminated
- Why it was recalled
- Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.
- Product
- ultraView Universal DAB Detection Kit, Catalog Number 05269806001, model 760-500 Immunohistochemistry (IHC) for in vitro diagnostic use.
- Where it went
- worldwide
- How much
- 14,655 units
Class I
ultraView SISH DNP Detection Kit, Catalog Number 05907136001, model 800-098 Immunohistochemistry (IHC) for in
Z-3016-2018 · Class I · reported 2018-09-26 · Terminated
- Why it was recalled
- Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.
- Product
- ultraView SISH DNP Detection Kit, Catalog Number 05907136001, model 800-098 Immunohistochemistry (IHC) for in vitro diagnostic use.
- Where it went
- worldwide
- How much
- 342 units
Class I
CONFIRM anti-Estrogen Receptor (ER) (SP1) Rabbit Monoclonal Primary Antibody, Catalog Number 05278414001, Mod
Z-3007-2018 · Class I · reported 2018-09-26 · Terminated
- Why it was recalled
- Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.
- Product
- CONFIRM anti-Estrogen Receptor (ER) (SP1) Rabbit Monoclonal Primary Antibody, Catalog Number 05278414001, Model 790-4325 Immunohistochemistry (IHC) for in vitro diagnostic use.
- Where it went
- worldwide
- How much
- 1719 units
Class I
ISH iVIEW Blue Plus Detection Kit, Catalog Number 05266181001, model 760-097 Immunohistochemistry (IHC) fo
Z-3002-2018 · Class I · reported 2018-09-26 · Terminated
- Why it was recalled
- Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.
- Product
- ISH iVIEW Blue Plus Detection Kit, Catalog Number 05266181001, model 760-097 Immunohistochemistry (IHC) for in vitro diagnostic use.
- Where it went
- worldwide
- How much
- 2136 units
Class I
CINtec p16 Histology (250) CE, Catalog Number 06695256001, model 825-4713 Immunohistochemistry (IHC) for i
Z-3013-2018 · Class I · reported 2018-09-26 · Terminated
- Why it was recalled
- Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.
- Product
- CINtec p16 Histology (250) CE, Catalog Number 06695256001, model 825-4713 Immunohistochemistry (IHC) for in vitro diagnostic use.
- Where it went
- worldwide
- How much
- 240 units
Class I
OptiView DAB IHC Detection Kit, Catalog Number 06396500001, model 760-700 Immunohistochemistry (IHC) for in
Z-2996-2018 · Class I · reported 2018-09-26 · Terminated
- Why it was recalled
- Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.
- Product
- OptiView DAB IHC Detection Kit, Catalog Number 06396500001, model 760-700 Immunohistochemistry (IHC) for in vitro diagnostic use.
- Where it went
- worldwide
- How much
- 11,712 units
Class I
ultraView Universal DAB Detection Kit, Ventana part Number 760-500, Roche GMMI 05269806001 in vitro diagnosti
Z-0730-2018 · Class I · reported 2018-03-07 · Terminated
- Why it was recalled
- Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
- Product
- ultraView Universal DAB Detection Kit, Ventana part Number 760-500, Roche GMMI 05269806001 in vitro diagnostic
- Where it went
- Worldwide distribution, including US nationwide, Argentina, Australia, Hong Kong, Indonesia, Philippines, Brazil, Chile, Thailand, Korea, Malaysia, New Zealand, Ecuador, Mexico, Pakistan, Vietnam, Germany, Japan, India, Colombia, Singapore, Uruguay, Peru, Canada, Taiwan, China,
- How much
- 72730 units
Class I
Hematoxylin II, Ventana Part Number 790-22087, Roche GMMI 05277965001 in vitro diagnostic
Z-0738-2018 · Class I · reported 2018-03-07 · Terminated
- Why it was recalled
- Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
- Product
- Hematoxylin II, Ventana Part Number 790-22087, Roche GMMI 05277965001 in vitro diagnostic
- Where it went
- Worldwide distribution, including US nationwide, Argentina, Australia, Hong Kong, Indonesia, Philippines, Brazil, Chile, Thailand, Korea, Malaysia, New Zealand, Ecuador, Mexico, Pakistan, Vietnam, Germany, Japan, India, Colombia, Singapore, Uruguay, Peru, Canada, Taiwan, China,
- How much
- 71640 units
Class I
iView DAB IHC Detection Kit, Ventana part Number 760-091, Roche GMMI 05266157001 in vitro diagnostic
Z-0731-2018 · Class I · reported 2018-03-07 · Terminated
- Why it was recalled
- Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
- Product
- iView DAB IHC Detection Kit, Ventana part Number 760-091, Roche GMMI 05266157001 in vitro diagnostic
- Where it went
- Worldwide distribution, including US nationwide, Argentina, Australia, Hong Kong, Indonesia, Philippines, Brazil, Chile, Thailand, Korea, Malaysia, New Zealand, Ecuador, Mexico, Pakistan, Vietnam, Germany, Japan, India, Colombia, Singapore, Uruguay, Peru, Canada, Taiwan, China,
- How much
- 1980 units
Class I
NEXES VEN IVIEW DAB DET KT JPN-US EXPORT, Ventana Part Number 760-041, Roche GMMI 05266084001 in vitro diagnos
Z-0740-2018 · Class I · reported 2018-03-07 · Terminated
- Why it was recalled
- Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
- Product
- NEXES VEN IVIEW DAB DET KT JPN-US EXPORT, Ventana Part Number 760-041, Roche GMMI 05266084001 in vitro diagnostic
- Where it went
- Worldwide distribution, including US nationwide, Argentina, Australia, Hong Kong, Indonesia, Philippines, Brazil, Chile, Thailand, Korea, Malaysia, New Zealand, Ecuador, Mexico, Pakistan, Vietnam, Germany, Japan, India, Colombia, Singapore, Uruguay, Peru, Canada, Taiwan, China,
- How much
- 3122 units
Showing the 30 most recent of 41. The rest are searchable at the FDA enforcement report database.
Read these counts with three things in mind
A firm with more recalls usually makes more products. A contract packer runs
hundreds of lines. It will out-recall a company with one product whatever either
does. This is not a safety ranking.
Most recalls here are voluntary. A firm found its own fault and pulled the stock.
That is the system working. It lands in this database looking the same as a recall
the FDA had to push for.
Terminated means the FDA considers the recall finished. It is not a verdict.
It does not mean anyone was harmed.
Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.