Teleflex Medical: 204 FDA recall reports
Every medical device recall the FDA has published for Teleflex Medical. The fault, and how the agency graded it.
204Reports
39Class I
6Ongoing
2020-01-22Most recent
39 Class I on this pageClass I means the FDA judged there was a reasonable probability of serious harm or death. Check the lot numbers against anything you hold.
What the classes mean
The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.
Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.
By class
By year reported
Checking whether it affects you
The lot or code number is what matters. Not the product name. A recall covers
specific lots. Yours may not be one of them.
Check the code on your package against the report. The FDA keeps the current
list at fda.gov/safety/recalls.
The reports
The text below is quoted from the FDA record word for word.
Teleflex MEDICAL RuSCH One piece Male External Regular Catheter W/O Tape, REF A1500
Z-0796-2020 · Class III · reported 2020-01-22 · Terminated
- Why it was recalled
- The device label incorrectly states that the device is sterile. This product is an externally applied male condom-type catheter designed for the management of male incontinence as an alternative to incontinence pads. In normal use, the device is neither expected nor required to be sterile.
- Product
- Teleflex MEDICAL RuSCH One piece Male External Regular Catheter W/O Tape, REF A1500
- Where it went
- ME,NC, WA, MO, NJ, FL, CAL, TX, OH, IL, MA, GA, FL, GA, Canada
- How much
- 39456 devices
Teleflex MEDICAL RuSCH One piece Male External Medium Catheter W/O Tape, REF A1500M
Z-0797-2020 · Class III · reported 2020-01-22 · Terminated
- Why it was recalled
- The device label incorrectly states that the device is sterile. This product is an externally applied male condom-type catheter designed for the management of male incontinence as an alternative to incontinence pads. In normal use, the device is neither expected nor required to be sterile.
- Product
- Teleflex MEDICAL RuSCH One piece Male External Medium Catheter W/O Tape, REF A1500M
- Where it went
- ME,NC, WA, MO, NJ, FL, CAL, TX, OH, IL, MA, GA, FL, GA, Canada
- How much
- 13104 devices
Arrow EZ-IO Intraosseous Vascular Access System 25mm Needle+Stabilizer Kit, Product Code 9001P-VC-005 - Produc
Z-0689-2020 · Class II · reported 2019-12-25 · Ongoing
- Why it was recalled
- The safety cap attached to needles within the needle sets may become dislodged exposing the needle and potentially causing the needle to protrude through the packaging. If this issue is not detected, the immediate risk of exposure to the affected devices is needle stick injury to the clinician or health care professional. In addition, a puncture of the packaging may compromise the sterility of the needle.
- Product
- Arrow EZ-IO Intraosseous Vascular Access System 25mm Needle+Stabilizer Kit, Product Code 9001P-VC-005 - Product Usage: The Arrow EZ-IO Intraosseous Vascular Access System provides intraosseous access in the proximal humerus, proximal tibia, and distal tibia of adult and pediatric patients, and the distal femur in pediatric patients when vascular access is difficult to obtain in emergent, urgent, or medically necessary cases for up to 24 hours.
- Where it went
- US nationwide distribution.
- How much
- 526,000 total
Arrow EZ-IO Intraosseous Vascular Access System 45mm Needle+Stabilizer Kit, Product Code 9079P-VC-005 - Produc
Z-0690-2020 · Class II · reported 2019-12-25 · Ongoing
- Why it was recalled
- The safety cap attached to needles within the needle sets may become dislodged exposing the needle and potentially causing the needle to protrude through the packaging. If this issue is not detected, the immediate risk of exposure to the affected devices is needle stick injury to the clinician or health care professional. In addition, a puncture of the packaging may compromise the sterility of the needle.
- Product
- Arrow EZ-IO Intraosseous Vascular Access System 45mm Needle+Stabilizer Kit, Product Code 9079P-VC-005 - Product Usage: The Arrow EZ-IO Intraosseous Vascular Access System provides intraosseous access in the proximal humerus, proximal tibia, and distal tibia of adult and pediatric patients, and the distal femur in pediatric patients when vascular access is difficult to obtain in emergent, urgent, or medically necessary cases for up to 24 hours.
- Where it went
- US nationwide distribution.
- How much
- 526,000 total
HUDSON RCI SHERIDAN SHERI-BRONCH Endobronchial Tube REF 5-16135 - Product Usage: The Sheridan Endobronchial Tu
Z-0703-2020 · Class II · reported 2019-12-25 · Terminated
- Why it was recalled
- The tracheal and brochial swivel connectors are incorrectly labeled.
- Product
- HUDSON RCI SHERIDAN SHERI-BRONCH Endobronchial Tube REF 5-16135 - Product Usage: The Sheridan Endobronchial Tubes are intended for use in Thoracic surgery, brochospirometry, adminitration of endobronchial anesthesia and other uses commonly requiring endobronchial intubation. The Sheridan Endobronchial tube is indicated for main stem bronchus intubation and allows for selective inflation or deflation of either lung.
- Where it went
- US Nationwide distribution in the states of CT, OK, LA, IL, GA, NY, CA, MT, MI, MA, FL.
- How much
- 53 units
Arrow EZ-IO Intraosseous Vascular Access System 15mm Needle+Stabilizer Kit, Product Code 9018P-VC-005 - Produc
Z-0688-2020 · Class II · reported 2019-12-25 · Ongoing
- Why it was recalled
- The safety cap attached to needles within the needle sets may become dislodged exposing the needle and potentially causing the needle to protrude through the packaging. If this issue is not detected, the immediate risk of exposure to the affected devices is needle stick injury to the clinician or health care professional. In addition, a puncture of the packaging may compromise the sterility of the needle.
- Product
- Arrow EZ-IO Intraosseous Vascular Access System 15mm Needle+Stabilizer Kit, Product Code 9018P-VC-005 - Product Usage: The Arrow EZ-IO Intraosseous Vascular Access System provides intraosseous access in the proximal humerus, proximal tibia, and distal tibia of adult and pediatric patients, and the distal femur in pediatric patients when vascular access is difficult to obtain in emergent, urgent, or medically necessary cases for up to 24 hours.
- Where it went
- US nationwide distribution.
- How much
- 526,000 total
WECK Hem-O-Lok AutoEndo5, 5mm Automatic Endoscopie Hem-o-lok Applier, REF 543965 Automatic Endoscopic Su
Z-0497-2020 · Class II · reported 2019-11-27 · Ongoing
- Why it was recalled
- Certain clip appliers have an increase in misloading and/or jamming related complaints that may be due to a discrepancy in the specification a component. Misloading and/or jamming may lead to clips falling out of the applier, clip breakage, or the jaws becoming locked in a partially closed position (jammed).
- Product
- WECK Hem-O-Lok AutoEndo5, 5mm Automatic Endoscopie Hem-o-lok Applier, REF 543965 Automatic Endoscopic Surgical Ligating Clip Appliers are delivery devices for Hem-o-lok ML non-absorbable polymer ligating clips
- Where it went
- Worldwide distribution, including US nationwide (including Puerto Rico).
- How much
- 4099 devices
WECK AutoEndo5, Automatic Hem-O-Lok Clip Applier, REF AE05ML Automatic Endoscopic Surgical Ligating Clip
Z-0496-2020 · Class II · reported 2019-11-27 · Ongoing
- Why it was recalled
- Certain clip appliers have an increase in misloading and/or jamming related complaints that may be due to a discrepancy in the specification a component. Misloading and/or jamming may lead to clips falling out of the applier, clip breakage, or the jaws becoming locked in a partially closed position (jammed).
- Product
- WECK AutoEndo5, Automatic Hem-O-Lok Clip Applier, REF AE05ML Automatic Endoscopic Surgical Ligating Clip Appliers are delivery devices for Hem-o-lok ML non-absorbable polymer ligating clips
- Where it went
- Worldwide distribution, including US nationwide (including Puerto Rico).
- How much
- 34491 devices
Class I
Hudson RCI Sheridan EZ-ENDO: a) 5.0 mm, REF 5-22510 b) 5.5 mm, REF 5-22511 c) 7.0 mm, REF 5-22514 d) 7.5 m
Z-2313-2019 · Class I · reported 2019-09-18 · Terminated
- Why it was recalled
- Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.
- Product
- Hudson RCI Sheridan EZ-ENDO: a) 5.0 mm, REF 5-22510 b) 5.5 mm, REF 5-22511 c) 7.0 mm, REF 5-22514 d) 7.5 mm, REF 5-22515 e) 8.0 mm, REF 5-22516 f) 8.5 mm, REF 5-22517 g) 9.0 mm, REF 5-22518 h) 6.0 mm, REF 5-22512 i) 6.5 mm, REF 5-22513 Product Usage: Tracheal tube/airway management
- Where it went
- US Nationwide Distribution
- How much
- 57181units
Class I
Hudson RCI Sheridan/HVT: a) 5.0 mm, REF 5-10310 b) 5.5 mm, REF 5-10311 c) 6.0 mm, REF 5-10312 d) 6.5 mm, R
Z-2316-2019 · Class I · reported 2019-09-18 · Terminated
- Why it was recalled
- Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.
- Product
- Hudson RCI Sheridan/HVT: a) 5.0 mm, REF 5-10310 b) 5.5 mm, REF 5-10311 c) 6.0 mm, REF 5-10312 d) 6.5 mm, REF 5-10313 e) 7.0 mm, REF 5-10314 f) 7.5 mm, REF 5-10315 g) 8.0 mm, REF 5-10316 h) 8.5 mm, REF 5-10317 i) 9.0 mm, REF 5-10318 j) 10.0 mm, REF 5-10320 Product Usage: Tracheal tube/airway management
- Where it went
- US Nationwide Distribution
- How much
- 3375808 units
Class I
Hudson RCI Sheridan/CF: a) 3.0 mm, REF 5-10106, REF 5-10206 b) 3.5 mm, REF 5-10107 c) 4.0 mm, REF 5-10108,
Z-2312-2019 · Class I · reported 2019-09-18 · Terminated
- Why it was recalled
- Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.
- Product
- Hudson RCI Sheridan/CF: a) 3.0 mm, REF 5-10106, REF 5-10206 b) 3.5 mm, REF 5-10107 c) 4.0 mm, REF 5-10108, REF 5-10208 d) 4.5 mm, REF 5-10109 e) 5.0 mm, REF 5-10110, REF 5-10210 f) 5.5 mm, REF 5-10111 g) 6.0 mm, REF 5-10112, REF 5-10212 h) 6.5 mm, REF 5-10113, REF 5-10213 I) 7.0 mm, REF 5-10114, REF 5-10214 j) 7.5 mm, REF V5-10115, REF 5-10115, REF 5-10215 k) 8.0 mm, REF 5-10116, REF 5-10216, REF V5-10116 l) 8.5 mm, REF 5-10117, REF 5-10217 m) 9.0 mm, REF 5-10118, REF5-10218 n) 9.5 mm, REF 5-10119 o) 10.0 mm, REF 5-10120, REF 5-10220 Product Usage: Tracheal tube/airway management
- Where it went
- US Nationwide Distribution
- How much
- 1805730 units
Class I
Hudson RCI Sheridan Preformed: a) 3.0 mm, REF 5-22006, REF 5-22106 b) 3.5 mm, REF 5-22007, REF 5-22107 c) 4
Z-2314-2019 · Class I · reported 2019-09-18 · Terminated
- Why it was recalled
- Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.
- Product
- Hudson RCI Sheridan Preformed: a) 3.0 mm, REF 5-22006, REF 5-22106 b) 3.5 mm, REF 5-22007, REF 5-22107 c) 4.0 mm, REF 5-22008, REF 5-22108, REF 5-22208 d) 4.5 mm, REF 5-22009, REF 5-22109, REF 5-22209 e) 5.0 mm, REF 5-22010, REF 5-22110, REF 5-22210 f) 5.5 mm, REF 5-22011, REF 5-22111, REF 5-22211 g) 6.0 mm, REF 5-22212, REF 5-22312, REF 5-22112 h) 6.5 mm, REF 5-22313, REF 5-22213, REF 5-22013, REF 5-22113 i) 7.0 mm, REF 5-22214, REF 5-22314, REF 5-22014, REF 5-22114 j) 7.5 mm, REF 5-22215, REF 5-22315 k) 8.0 mm, REF 5-22216, REF 5-22316 l) 8.5 mm, REF 5-22217 m) 9.0 mm, REF 5-22218 Product Usage: Tracheal tube/airway management
- Where it went
- US Nationwide Distribution
- How much
- 278142 units
Class I
Hudson RCI Sheridan Uncuffed: a) 2.0 mm, REF 5-10404 b) 2.5 mm, REF 5-10405 c) 3.0 mm, REF 5-10406 d) 3.
Z-2315-2019 · Class I · reported 2019-09-18 · Terminated
- Why it was recalled
- Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.
- Product
- Hudson RCI Sheridan Uncuffed: a) 2.0 mm, REF 5-10404 b) 2.5 mm, REF 5-10405 c) 3.0 mm, REF 5-10406 d) 3.5 mm, REF 5-10407 e) 4.0 mm, REF 5-10408 f) 4.5 mm, REF 5-10409 g) 5.0 mm, REF 5-10410 h) 5.5 mm, REF 5-10411 i) 6.0 mm, REF 5-10412 j) 6.5 mm, REF 5-10413 k) 7.0 mm, REF 5-10414 Product Usage: Tracheal tube/airway management
- Where it went
- US Nationwide Distribution
- How much
- 170740 units
Class I
Hudson RCI Sheridan LTS: a) 4.0 mm, REF 5-11108 b) 5.0 mm, REF 5-11110 c) 6.0 mm, REF 5-11112 Product Us
Z-2311-2019 · Class I · reported 2019-09-18 · Terminated
- Why it was recalled
- Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.
- Product
- Hudson RCI Sheridan LTS: a) 4.0 mm, REF 5-11108 b) 5.0 mm, REF 5-11110 c) 6.0 mm, REF 5-11112 Product Usage: Tracheal tube/airway management
- Where it went
- US Nationwide Distribution
- How much
- 7963 units
Class I
Sheridan/HVT Novaplus: a) 7.0 mm, REF V5-10314 b) 7.5 mm, REF V5-10315 c) 8.0 mm, REF V5-10316 d) 8.5 mm,
Z-2318-2019 · Class I · reported 2019-09-18 · Terminated
- Why it was recalled
- Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.
- Product
- Sheridan/HVT Novaplus: a) 7.0 mm, REF V5-10314 b) 7.5 mm, REF V5-10315 c) 8.0 mm, REF V5-10316 d) 8.5 mm, REF V5-10317 Product Usage: Tracheal tube/airway management
- Where it went
- US Nationwide Distribution
- How much
- 1391004 units
Class I
Sheridan/CF Novaplus 7.0 mm, REF V5-10114 Product Usage: Tracheal tube/airway management
Z-2317-2019 · Class I · reported 2019-09-18 · Terminated
- Why it was recalled
- Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.
- Product
- Sheridan/CF Novaplus 7.0 mm, REF V5-10114 Product Usage: Tracheal tube/airway management
- Where it went
- US Nationwide Distribution
- How much
- 77136 units
Rusch EasyCath Kit
Z-2086-2019 · Class II · reported 2019-08-07 · Terminated
- Why it was recalled
- Potential for product not to be sterile.
- Product
- Rusch EasyCath Kit
- Where it went
- Nationwide distribution to CA, FL, IL, NC, NJ, OH, OK, TX.
- How much
- 3,700 devices
RUSCH Urinary Drainage bag, 2000mlm REF 390060 urine collection device
Z-1975-2019 · Class III · reported 2019-07-24 · Terminated
- Why it was recalled
- The device labels are not UDI compliant. The missing UDI compliance information on the labelling would not affect the function/performance of the device itself. Other label information and warnings are unaffected by this labelling.
- Product
- RUSCH Urinary Drainage bag, 2000mlm REF 390060 urine collection device
- Where it went
- Worldwide Distribution: US (nationwide) and countries of: Canada, Bolivia, Guatemala, Jamaica, Peru, Trinidad, and El Salvador.
- How much
- 76725 US eaches
Class I
Teleflex HUDSON RDI NEONATAL ConchaSmart Breathing Circuit with Dual Heated Limb and ConchaSmart Column: a)
Z-1912-2019 · Class I · reported 2019-07-24 · Terminated
- Why it was recalled
- Reported complaints of cracks being observed, prior to use, on swivel wye adaptors of the Neonatal ConchaSmart Breathing Circuit.
- Product
- Teleflex HUDSON RDI NEONATAL ConchaSmart Breathing Circuit with Dual Heated Limb and ConchaSmart Column: a) REF 870-07KIT b) REF 870-09KIT Product Usage: The Neonatal ConchaSmart breathing circuits are intended to deliver humidified breathing gases for administration to an infant/neonate patient. These gases may be delivered by nasal prongs using the CareFusion Infant Flow system and the single limb circuit, or through bypassing upper airways, for example through the use of an endotracheal tube with the dual limb configuration.
- Where it went
- US Nationwide Distribution in the states of AL, TX
- How much
- 20 units
RUSCH CARE Premium Drain bag, 2000mlm REF 390000 urine collection device
Z-1974-2019 · Class III · reported 2019-07-24 · Terminated
- Why it was recalled
- The device labels are not UDI compliant. The missing UDI compliance information on the labelling is not likely to affect the function/performance of the device itself. Other label information and warnings are unaffected by this labelling.
- Product
- RUSCH CARE Premium Drain bag, 2000mlm REF 390000 urine collection device
- Where it went
- Worldwide Distribution: US (nationwide) and countries of: Canada, Bolivia, Guatemala, Jamaica, Peru, Trinidad, and El Salvador.
- How much
- 105426 US eaches
Class I
Hudson RCI Sheridan: 1) EZ-ENDO 6.0 mm, Product Code 5-22512 2) EZ-ENDO 6.5 mm, Product Code 5-22513 3) EZ-
Z-1761-2019 · Class I · reported 2019-06-26 · Terminated
- Why it was recalled
- This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.
- Product
- Hudson RCI Sheridan: 1) EZ-ENDO 6.0 mm, Product Code 5-22512 2) EZ-ENDO 6.5 mm, Product Code 5-22513 3) EZ-ENDO 7.0 mm, Product Code 5-22514 4) EZ-ENDO 7.5 mm, Product Code 5-22515 5) EZ-ENDO 8.0 mm, Product Code 5-22516 6) EZ-ENDO 8.5 mm, Product Code 5-22517 Product Usage: Tracheal tube/airway management
- Where it went
- Worldwide - US Nationwide Distribution
- How much
- 55312 units
Class I
Hudson RCI Sheridan: 1) CF 6.0 mm, Product Codes: 5-10112, 5-10212 2) CF 6.5 mm, Product Codes: 5-10113, 5
Z-1760-2019 · Class I · reported 2019-06-26 · Terminated
- Why it was recalled
- This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.
- Product
- Hudson RCI Sheridan: 1) CF 6.0 mm, Product Codes: 5-10112, 5-10212 2) CF 6.5 mm, Product Codes: 5-10113, 5-10213 3) CF 7.0 mm, Product Codes: 5-10114, 5-10214 4) CF 7.5 mm, Product Codes: V5-10115, 5-10115, 5-10215 5) CF 8.0 mm, Product Codes: 5-10116, 5-10216, V5-10116 6) CF 8.5 mm, Product Codes: 5-10117, 5-10217 Product Usage: Tracheal tube/airway management
- Where it went
- Worldwide - US Nationwide Distribution
- How much
- 1101073 units
Class I
Hudson RCI Sheridan: 1) HVT 6.0 mm, Product Code 5-10312 2) HVT 6.5 mm, Product Code 5-10313 3) HVT 7.0 mm,
Z-1764-2019 · Class I · reported 2019-06-26 · Terminated
- Why it was recalled
- This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.
- Product
- Hudson RCI Sheridan: 1) HVT 6.0 mm, Product Code 5-10312 2) HVT 6.5 mm, Product Code 5-10313 3) HVT 7.0 mm, Product Code 5-10314 4) HVT 7.5 mm, Product Code 5-10315 5) HVT 8.0 mm, Product Code 5-10316 6) HVT 8.5 mm, Product Code 5-10317 Product Usage: Tracheal tube/airway management
- Where it went
- Worldwide - US Nationwide Distribution
- How much
- 3238600 units
Class I
Sheridan/CF Novaplus 7.0 mm, Product Code V5-10114 Product Usage: Tracheal tube/airway management
Z-1765-2019 · Class I · reported 2019-06-26 · Terminated
- Why it was recalled
- This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.
- Product
- Sheridan/CF Novaplus 7.0 mm, Product Code V5-10114 Product Usage: Tracheal tube/airway management
- Where it went
- Worldwide - US Nationwide Distribution
- How much
- 77136 units
Class I
Hudson RCI Sheridan LTS, Product Code 5-11112 Product Usage: Tracheal tube/airway management
Z-1759-2019 · Class I · reported 2019-06-26 · Terminated
- Why it was recalled
- This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.
- Product
- Hudson RCI Sheridan LTS, Product Code 5-11112 Product Usage: Tracheal tube/airway management
- Where it went
- Worldwide - US Nationwide Distribution
- How much
- 3483 units
Class I
Sheridan/HVT: 1) Novaplus 7.0 mm, Product Code V5-10314 2) Novaplus 7.5 mm, Product Code V5-10315 3) Novapl
Z-1766-2019 · Class I · reported 2019-06-26 · Terminated
- Why it was recalled
- This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.
- Product
- Sheridan/HVT: 1) Novaplus 7.0 mm, Product Code V5-10314 2) Novaplus 7.5 mm, Product Code V5-10315 3) Novaplus 8.0 mm, Product Code V5-10316 4) Novaplus 8.5 mm, Product Code V5-10317 Product Usage: Tracheal tube/airway management
- Where it went
- Worldwide - US Nationwide Distribution
- How much
- 1390944 units
Class I
Hudson RCI Sheridan: 1) Preformed 6.0 mm, Product Codes: a) 5-22212, b) 5-22312, c) 5-22112, d) 5-22
Z-1762-2019 · Class I · reported 2019-06-26 · Terminated
- Why it was recalled
- This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.
- Product
- Hudson RCI Sheridan: 1) Preformed 6.0 mm, Product Codes: a) 5-22212, b) 5-22312, c) 5-22112, d) 5-22012, 2) Preformed 6.5 mm, Product Codes: a) 5-22313, b) 5-22213, c) 5-22013, d) 5-22113, 3) Preformed 7.0 mm, Product Codes: a) 5-22214, b) 5-22314, c) 5-22014, d) 5-22114, 4) Preformed 7.5 mm, Product Codes: a) 5-22215, b) 5-22315, 5) Preformed 8.0 mm, Product Codes: a) 5-22216, b) 5-22316, 6) Preformed 8.5 mm, Product Codes: 5-22217 Product Usage: Tracheal tube/airway management
- Where it went
- Worldwide - US Nationwide Distribution
- How much
- 186592 units
Class I
Hudson RCI Sheridan: 1) Uncuffed 6.0 mm, Product Code 5-10412 2) Uncuffed 6.5 mm, Product Code 5-10413 3) U
Z-1763-2019 · Class I · reported 2019-06-26 · Terminated
- Why it was recalled
- This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.
- Product
- Hudson RCI Sheridan: 1) Uncuffed 6.0 mm, Product Code 5-10412 2) Uncuffed 6.5 mm, Product Code 5-10413 3) Uncuffed 7.0 mm, Product Code 5-10414 Product Usage: Tracheal tube/airway management
- Where it went
- Worldwide - US Nationwide Distribution
- How much
- 14357
RUSCH LASERTUBE (Rubber), Order No. : 102004-000050, REF 102004, SIZE I.D. mm 5,0
Z-1683-2019 · Class II · reported 2019-06-05 · Terminated
- Why it was recalled
- The firm received reports indicating that the laser guard foil partially separated and/or slightly detached at the edges.
- Product
- RUSCH LASERTUBE (Rubber), Order No. : 102004-000050, REF 102004, SIZE I.D. mm 5,0
- Where it went
- Nationwide
- How much
- 5995 units
RUSCH LASERTUBE (Rubber), Order No. : 102004-000040, REF 102004, SIZE I.D. mm 4,0
Z-1682-2019 · Class II · reported 2019-06-05 · Terminated
- Why it was recalled
- The firm received reports indicating that the laser guard foil partially separated and/or slightly detached at the edges.
- Product
- RUSCH LASERTUBE (Rubber), Order No. : 102004-000040, REF 102004, SIZE I.D. mm 4,0
- Where it went
- Nationwide
- How much
- 1787 units
Showing the 30 most recent of 204. The rest are searchable at the FDA enforcement report database.
Read these counts with three things in mind
A firm with more recalls usually makes more products. A contract packer runs
hundreds of lines. It will out-recall a company with one product whatever either
does. This is not a safety ranking.
Most recalls here are voluntary. A firm found its own fault and pulled the stock.
That is the system working. It lands in this database looking the same as a recall
the FDA had to push for.
Terminated means the FDA considers the recall finished. It is not a verdict.
It does not mean anyone was harmed.
Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.
← All device recalls