TELEFLEX LLC: 140 FDA recall reports
Every medical device recall the FDA has published for TELEFLEX LLC. The fault, and how the agency graded it.
140Reports
129Class I
139Ongoing
2026-07-08Most recent
129 Class I on this pageClass I means the FDA judged there was a reasonable probability of serious harm or death. Check the lot numbers against anything you hold.
What the classes mean
The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.
Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.
By class
By year reported
Checking whether it affects you
The lot or code number is what matters. Not the product name. A recall covers
specific lots. Yours may not be one of them.
Check the code on your package against the report. The FDA keeps the current
list at fda.gov/safety/recalls.
The reports
The text below is quoted from the FDA record word for word.
Brand Name: Pilling Wecksorb Product Name: Pilling Wecksorb Neurosponges (Cotton) Pilling¿ Wecksorb Neurospo
Z-2639-2026 · Class II · reported 2026-07-08 · Ongoing
- Why it was recalled
- Due to out-of-specification, or potential out-of-specification, levels of endotoxins
- Product
- Brand Name: Pilling Wecksorb Product Name: Pilling Wecksorb Neurosponges (Cotton) Pilling¿ Wecksorb Neurosponge Strips (Cotton) Model/Catalog Number: 200100 200101 200102 200103 200104 200105 200106 200108 200110 200112 200120 200121 200123 200131 Software Version: N/A Product Description: Wecksorb Neurosponges are made of an absorbent cotton material. Certain Neurosponge varieties are available with an x-ray detectable thread to assist with location within the surgical site. Some Neurosponges (patties) have a string for accountability during the post-surgical verification process. Wecksorb Neurosponges are for use in the protection of tissue during surgical procedures; tissues include the brain and other central nervous system tissues. Neuro sponges are used during cranial and spine surgery to moisten the dura or to cushion against bone edges. They are also used to protect brain tissue under the dura or elsewhere during incision. Component: N/A
- Where it went
- Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The country of Panama.
- How much
- 1488700
Pilling LOWSLEY PROSTATIC TRACTOR CVD, REF 243200; retractor
Z-2096-2023 · Class II · reported 2023-07-12 · Ongoing
- Why it was recalled
- There are incomplete pre-cleaning instructions within the Instructions for Use (IFU).
- Product
- Pilling LOWSLEY PROSTATIC TRACTOR CVD, REF 243200; retractor
- Where it went
- Nationwide and Australia, Canada, S. Korea, Singapore
- How much
- 170 units
Pilling O'REILLY ESOPHAGEAL RETRACTOR, REF 381801A; ENT retractor
Z-2095-2023 · Class II · reported 2023-07-12 · Ongoing
- Why it was recalled
- There are incomplete pre-cleaning instructions within the Instructions for Use (IFU).
- Product
- Pilling O'REILLY ESOPHAGEAL RETRACTOR, REF 381801A; ENT retractor
- Where it went
- Nationwide and Australia, Canada, S. Korea, Singapore
- How much
- 365 units
Pilling LOWSLEY PROSTATIC TRACTOR STR, REF 243201; retractor
Z-2097-2023 · Class II · reported 2023-07-12 · Ongoing
- Why it was recalled
- There are incomplete pre-cleaning instructions within the Instructions for Use (IFU).
- Product
- Pilling LOWSLEY PROSTATIC TRACTOR STR, REF 243201; retractor
- Where it went
- Nationwide and Australia, Canada, S. Korea, Singapore
- How much
- 32 units
Class I
Slick Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170085
Z-1834-2023 · Class I · reported 2023-07-05 · Ongoing
- Why it was recalled
- Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
- Product
- Slick Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170085
- Where it went
- US Nationwide distribution including Puerto Rico.
- How much
- 4370 units
Class I
Endotracheal Tube oral/nasal uncuffed/plain - Murphy, REF 100382035
Z-1875-2023 · Class I · reported 2023-07-05 · Ongoing
- Why it was recalled
- Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
- Product
- Endotracheal Tube oral/nasal uncuffed/plain - Murphy, REF 100382035
- Where it went
- US Nationwide distribution including Puerto Rico.
- How much
- 25435 units
Class I
Endotracheal Tube oral/nasal uncuffed/plain - Murphy, REF 100382055
Z-1937-2023 · Class I · reported 2023-07-05 · Ongoing
- Why it was recalled
- Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
- Product
- Endotracheal Tube oral/nasal uncuffed/plain - Murphy, REF 100382055
- Where it went
- US Nationwide distribution including Puerto Rico.
- How much
- 7800 units
Class I
Endotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff, REF 112082075
Z-1889-2023 · Class I · reported 2023-07-05 · Ongoing
- Why it was recalled
- Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
- Product
- Endotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff, REF 112082075
- Where it went
- US Nationwide distribution including Puerto Rico.
- How much
- 300613 units
Class I
Slick Set Uncuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 150055
Z-1905-2023 · Class I · reported 2023-07-05 · Ongoing
- Why it was recalled
- Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
- Product
- Slick Set Uncuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 150055
- Where it went
- US Nationwide distribution including Puerto Rico.
- How much
- 70 units
Class I
Preformed AGT Oral Endotracheal Tube uncuffed/plain - Murphy, REF 100181055
Z-1922-2023 · Class I · reported 2023-07-05 · Ongoing
- Why it was recalled
- Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
- Product
- Preformed AGT Oral Endotracheal Tube uncuffed/plain - Murphy, REF 100181055
- Where it went
- US Nationwide distribution including Puerto Rico.
- How much
- 24752 units
Class I
Endotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff, REF 112082090
Z-1892-2023 · Class I · reported 2023-07-05 · Ongoing
- Why it was recalled
- Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
- Product
- Endotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff, REF 112082090
- Where it went
- US Nationwide distribution including Puerto Rico.
- How much
- 30423 units
Class I
Endotracheal Tube oral/nasal uncuffed/plain - Magill, REF 100380065
Z-1933-2023 · Class I · reported 2023-07-05 · Ongoing
- Why it was recalled
- Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
- Product
- Endotracheal Tube oral/nasal uncuffed/plain - Magill, REF 100380065
- Where it went
- US Nationwide distribution including Puerto Rico.
- How much
- 190 units
Class I
RUSCHELIT Safety Clear Tracheal Tube, oral/nasal, Magill, REF 112480030
Z-1896-2023 · Class I · reported 2023-07-05 · Ongoing
- Why it was recalled
- Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
- Product
- RUSCHELIT Safety Clear Tracheal Tube, oral/nasal, Magill, REF 112480030
- Where it went
- US Nationwide distribution including Puerto Rico.
- How much
- 21284 units
Class I
Endotracheal Tube oral/nasal Magill, High Volume, Low Pressure Cuff, REF 112080060
Z-1959-2023 · Class I · reported 2023-07-05 · Ongoing
- Why it was recalled
- Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
- Product
- Endotracheal Tube oral/nasal Magill, High Volume, Low Pressure Cuff, REF 112080060
- Where it went
- US Nationwide distribution including Puerto Rico.
- How much
- 4160 units
Class I
Preformed AGT Oral Endotracheal Tube uncuffed/plain - Murphy, REF 100180070
Z-1869-2023 · Class I · reported 2023-07-05 · Ongoing
- Why it was recalled
- Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
- Product
- Preformed AGT Oral Endotracheal Tube uncuffed/plain - Murphy, REF 100180070
- Where it went
- US Nationwide distribution including Puerto Rico.
- How much
- 970 units
Class I
Endotracheal Tube oral/nasal uncuffed/plain - Magill, REF 100380050
Z-1930-2023 · Class I · reported 2023-07-05 · Ongoing
- Why it was recalled
- Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
- Product
- Endotracheal Tube oral/nasal uncuffed/plain - Magill, REF 100380050
- Where it went
- US Nationwide distribution including Puerto Rico.
- How much
- 4110 units
Class I
Preformed AGT Oral Endotracheal Tube Murphy Eye, High Volume, Low Pressure Cuff, REF 111780085
Z-1949-2023 · Class I · reported 2023-07-05 · Ongoing
- Why it was recalled
- Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
- Product
- Preformed AGT Oral Endotracheal Tube Murphy Eye, High Volume, Low Pressure Cuff, REF 111780085
- Where it went
- US Nationwide distribution including Puerto Rico.
- How much
- 670 units
Class I
Endotracheal Tube oral/nasal uncuffed/plain - Magill, REF 100380030
Z-1926-2023 · Class I · reported 2023-07-05 · Ongoing
- Why it was recalled
- Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
- Product
- Endotracheal Tube oral/nasal uncuffed/plain - Magill, REF 100380030
- Where it went
- US Nationwide distribution including Puerto Rico.
- How much
- 5990 units
Class I
Slick Set Uncuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170140
Z-1841-2023 · Class I · reported 2023-07-05 · Ongoing
- Why it was recalled
- Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
- Product
- Slick Set Uncuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170140
- Where it went
- US Nationwide distribution including Puerto Rico.
- How much
- 10650 units
Class I
Preformed AGT Oral Endotracheal Tube Murphy Eye, High Volume, Low Pressure Cuff, REF 111780060
Z-1945-2023 · Class I · reported 2023-07-05 · Ongoing
- Why it was recalled
- Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
- Product
- Preformed AGT Oral Endotracheal Tube Murphy Eye, High Volume, Low Pressure Cuff, REF 111780060
- Where it went
- US Nationwide distribution including Puerto Rico.
- How much
- 16570 units
Class I
Slick Set Uncuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 150025
Z-1901-2023 · Class I · reported 2023-07-05 · Ongoing
- Why it was recalled
- Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
- Product
- Slick Set Uncuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 150025
- Where it went
- US Nationwide distribution including Puerto Rico.
- How much
- 2382 units
Class I
Preformed AGT Oral Endotracheal Tube uncuffed/plain - Murphy, REF 100181045
Z-1872-2023 · Class I · reported 2023-07-05 · Ongoing
- Why it was recalled
- Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
- Product
- Preformed AGT Oral Endotracheal Tube uncuffed/plain - Murphy, REF 100181045
- Where it went
- US Nationwide distribution including Puerto Rico.
- How much
- 53216 units
Class I
Endotracheal Tube oral/nasal Magill, High Volume, Low Pressure Cuff, REF 112080100
Z-1883-2023 · Class I · reported 2023-07-05 · Ongoing
- Why it was recalled
- Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
- Product
- Endotracheal Tube oral/nasal Magill, High Volume, Low Pressure Cuff, REF 112080100
- Where it went
- US Nationwide distribution including Puerto Rico.
- How much
- 690 units
Class I
Slick Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170050
Z-1827-2023 · Class I · reported 2023-07-05 · Ongoing
- Why it was recalled
- Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
- Product
- Slick Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170050
- Where it went
- US Nationwide distribution including Puerto Rico.
- How much
- 11560 units
Class I
Slick Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170080
Z-1833-2023 · Class I · reported 2023-07-05 · Ongoing
- Why it was recalled
- Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
- Product
- Slick Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170080
- Where it went
- US Nationwide distribution including Puerto Rico.
- How much
- 18150 units
Class I
Slick Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170090
Z-1835-2023 · Class I · reported 2023-07-05 · Ongoing
- Why it was recalled
- Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
- Product
- Slick Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170090
- Where it went
- US Nationwide distribution including Puerto Rico.
- How much
- 1460 units
Class I
Preformed AGT Oral Endotracheal Tube Murphy Eye, High Volume, Low Pressure Cuff, REF 111780075
Z-1877-2023 · Class I · reported 2023-07-05 · Ongoing
- Why it was recalled
- Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
- Product
- Preformed AGT Oral Endotracheal Tube Murphy Eye, High Volume, Low Pressure Cuff, REF 111780075
- Where it went
- US Nationwide distribution including Puerto Rico.
- How much
- 15810 units
Class I
Slick Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170065
Z-1830-2023 · Class I · reported 2023-07-05 · Ongoing
- Why it was recalled
- Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
- Product
- Slick Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170065
- Where it went
- US Nationwide distribution including Puerto Rico.
- How much
- 15020
Class I
Endotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff, REF 112082070
Z-1888-2023 · Class I · reported 2023-07-05 · Ongoing
- Why it was recalled
- Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
- Product
- Endotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff, REF 112082070
- Where it went
- US Nationwide distribution including Puerto Rico.
- How much
- 860472 units
Class I
Preformed AGT Oral Endotracheal Tube uncuffed/plain - Murphy, REF 100180050
Z-1865-2023 · Class I · reported 2023-07-05 · Ongoing
- Why it was recalled
- Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
- Product
- Preformed AGT Oral Endotracheal Tube uncuffed/plain - Murphy, REF 100180050
- Where it went
- US Nationwide distribution including Puerto Rico.
- How much
- 4146 units
Showing the 30 most recent of 140. The rest are searchable at the FDA enforcement report database.
Read these counts with three things in mind
A firm with more recalls usually makes more products. A contract packer runs
hundreds of lines. It will out-recall a company with one product whatever either
does. This is not a safety ranking.
Most recalls here are voluntary. A firm found its own fault and pulled the stock.
That is the system working. It lands in this database looking the same as a recall
the FDA had to push for.
Terminated means the FDA considers the recall finished. It is not a verdict.
It does not mean anyone was harmed.
Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.
← All device recalls