Synthes (USA) Products LLC

Synthes (USA) Products LLC: 164 FDA recall reports

Every medical device recall the FDA has published for Synthes (USA) Products LLC. The fault, and how the agency graded it.

164Reports
1Class I
24Ongoing
2026-04-08Most recent
1 Class I on this page

Class I means the FDA judged there was a reasonable probability of serious harm or death. Check the lot numbers against anything you hold.

What the classes mean

The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.

ClassWhat the FDA is saying
Class IThere is a reasonable probability that using this will cause serious harm or death.
Class IIIt may cause temporary or reversible harm. Serious harm is unlikely.
Class IIIIt breaks a rule but is unlikely to hurt anyone. Often a labelling fault.

Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.

By class

ClassReports
Class II148
Class III15
Class I1

By year reported

YearReports
20121
20133
20141
201523
201660
20172
201815
20192
202033
20227
20233
202410
20252
20262
Checking whether it affects you

The lot or code number is what matters. Not the product name. A recall covers specific lots. Yours may not be one of them.

Check the code on your package against the report. The FDA keeps the current list at fda.gov/safety/recalls.

The reports

The text below is quoted from the FDA record word for word.

2.4 VOLT TAP LOCKING SCREW TAP TO 90MM. Part Number: 03.424.124-US.

Z-1702-2026  ·  Class II  ·  reported 2026-04-08  ·  Ongoing

Why it was recalled
A full lot swap between VOLT Locking Screw Tap 03.424.124 lot 68900P9 and Cortical Screw Tap 03.424.126 lot 68982P8, resulting in the bones taps having an incorrect thread.
Product
2.4 VOLT TAP LOCKING SCREW TAP TO 90MM. Part Number: 03.424.124-US.
Where it went
US Nationwide distribution in the states of AR, AZ, CA, CO, FL, IN, LA, MD, MN, NC, NJ, OH, OK, OR, PA, TN, TX, UT, VA, WA, WI.
How much
77 units

2.4 VOLT(TM) TAP CORTEX SCREW TO 90MM. Part Number: 03.424.126-US.

Z-1703-2026  ·  Class II  ·  reported 2026-04-08  ·  Ongoing

Why it was recalled
A full lot swap between VOLT Locking Screw Tap 03.424.124 lot 68900P9 and Cortical Screw Tap 03.424.126 lot 68982P8, resulting in the bones taps having an incorrect thread.
Product
2.4 VOLT(TM) TAP CORTEX SCREW TO 90MM. Part Number: 03.424.126-US.
Where it went
US Nationwide distribution in the states of AR, AZ, CA, CO, FL, IN, LA, MD, MN, NC, NJ, OH, OK, OR, PA, TN, TX, UT, VA, WA, WI.
How much
33 units

LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 46MM / XL25/ STER. Product Code: 04.045.346TS. Used in RFN-ADVA

Z-0228-2026  ·  Class II  ·  reported 2025-10-29  ·  Ongoing

Why it was recalled
Although the etched length on the screw is correct, the length marked on the external packaging is incorrect. Specifically, 42mm screws were packaged and labeled as 46mm, while 46mm screws were packaged and labeled as 42mm.
Product
LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 46MM / XL25/ STER. Product Code: 04.045.346TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail Advanced System.
Where it went
Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MN, MO, MS, NJ, NM, NY, OH, TN, TX and the country of Canada.
How much
120 units

LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 42MM / XL25/ STER. Product Code: 04.045.342TS. Used in RFN-ADVA

Z-0229-2026  ·  Class II  ·  reported 2025-10-29  ·  Ongoing

Why it was recalled
Although the etched length on the screw is correct, the length marked on the external packaging is incorrect. Specifically, 42mm screws were packaged and labeled as 46mm, while 46mm screws were packaged and labeled as 42mm.
Product
LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 42MM / XL25/ STER. Product Code: 04.045.342TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail Advanced System.
Where it went
Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MN, MO, MS, NJ, NM, NY, OH, TN, TX and the country of Canada.
How much
122 units

STARDRIVE SCRWDRVR SHFT/T4 50MM/SELF-RETAINING/HXC-Intended for tightening, loosening, or removing of screw im

Z-1916-2024  ·  Class II  ·  reported 2024-06-05  ·  Ongoing

Why it was recalled
Stardrive Screwdrivers manufactured without a 5-degree relief cut around the tip and will not engage with their respective 1.3mm and 1.5mm screws, may result in surgical delay
Product
STARDRIVE SCRWDRVR SHFT/T4 50MM/SELF-RETAINING/HXC-Intended for tightening, loosening, or removing of screw implants during orthopaedic surgery. Part Number: 03.130.010
Where it went
Worldwide distribution - US Nationwide and the country of Canada.
How much
261 units

Opal, Intervertebral Cage, 10x32mm, h 11mm, Revolvable-Indicated for degenerative spine disease Part Number:

Z-1242-2024  ·  Class II  ·  reported 2024-03-13  ·  Ongoing

Why it was recalled
Products not sterilized, sterility cannot be confirmed
Product
Opal, Intervertebral Cage, 10x32mm, h 11mm, Revolvable-Indicated for degenerative spine disease Part Number: 08.803.231S
Where it went
Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Australia, Belgium, Canada, Czech Republic, Germany, Spain, Finland, France, United Kingdom, Ireland, South Korea, Kuwait, Mexico, Netherlands, Slovakia, Taiwan, South Africa.
How much
30 units

RAPIDSORB IPS Battery Pack, Sterile-Intended for use in non-load bearing fracture repair and reconstructive pr

Z-1238-2024  ·  Class II  ·  reported 2024-03-13  ·  Ongoing

Why it was recalled
Products not sterilized, sterility cannot be confirmed
Product
RAPIDSORB IPS Battery Pack, Sterile-Intended for use in non-load bearing fracture repair and reconstructive procedures of the craniofacial skeleton (excluding the upper and lower jaw) in pediatric and adult populations. Part Number: 530.553.01S
Where it went
Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Australia, Belgium, Canada, Czech Republic, Germany, Spain, Finland, France, United Kingdom, Ireland, South Korea, Kuwait, Mexico, Netherlands, Slovakia, Taiwan, South Africa.
How much
10 units

13 mm Titanium Hindfoot Arthrodesis Cannulated Nails EX, sterile -Intended to facilitate tibiotalocalcaneal ar

Z-1234-2024  ·  Class II  ·  reported 2024-03-13  ·  Ongoing

Why it was recalled
Products not sterilized, sterility cannot be confirmed
Product
13 mm Titanium Hindfoot Arthrodesis Cannulated Nails EX, sterile -Intended to facilitate tibiotalocalcaneal arthrodesis to treat severe foot/ankle deformity, arthritis, instability, and skeletal defects after tumor resection. Part Number: 04.008.378S
Where it went
Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Australia, Belgium, Canada, Czech Republic, Germany, Spain, Finland, France, United Kingdom, Ireland, South Korea, Kuwait, Mexico, Netherlands, Slovakia, Taiwan, South Africa.
How much
5 units

TFNA Femoral Nail 9mm, right, 130¿, L 235mm-Intended for temporary fixation and stabilization of proximal fem

Z-1239-2024  ·  Class II  ·  reported 2024-03-13  ·  Ongoing

Why it was recalled
Products not sterilized, sterility cannot be confirmed
Product
TFNA Femoral Nail 9mm, right, 130¿, L 235mm-Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.944S
Where it went
Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Australia, Belgium, Canada, Czech Republic, Germany, Spain, Finland, France, United Kingdom, Ireland, South Korea, Kuwait, Mexico, Netherlands, Slovakia, Taiwan, South Africa.
How much
11 units

1.5MM Drill Bit/QC/85MM Sterile-surgical, orthopedic, pneumatic powered & accessory/attachment Part Number: 3

Z-1235-2024  ·  Class II  ·  reported 2024-03-13  ·  Ongoing

Why it was recalled
Products not sterilized, sterility cannot be confirmed
Product
1.5MM Drill Bit/QC/85MM Sterile-surgical, orthopedic, pneumatic powered & accessory/attachment Part Number: 310.15S
Where it went
Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Australia, Belgium, Canada, Czech Republic, Germany, Spain, Finland, France, United Kingdom, Ireland, South Korea, Kuwait, Mexico, Netherlands, Slovakia, Taiwan, South Africa.
How much
17 units

VA-LCP Clavicle Plate 2.7, shaft, CS1, right-Intended for fixation of clavicle bone fragments. Part Number: 0

Z-1241-2024  ·  Class II  ·  reported 2024-03-13  ·  Ongoing

Why it was recalled
Products not sterilized, sterility cannot be confirmed
Product
VA-LCP Clavicle Plate 2.7, shaft, CS1, right-Intended for fixation of clavicle bone fragments. Part Number: 02.112.621S
Where it went
Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Australia, Belgium, Canada, Czech Republic, Germany, Spain, Finland, France, United Kingdom, Ireland, South Korea, Kuwait, Mexico, Netherlands, Slovakia, Taiwan, South Africa.
How much
47 units

2.5MM DRILL BIT/QC/GOLD/110MM STERILE-surgical, orthopedic, pneumatic powered & accessory/attachment Product

Z-1236-2024  ·  Class II  ·  reported 2024-03-13  ·  Ongoing

Why it was recalled
Products not sterilized, sterility cannot be confirmed
Product
2.5MM DRILL BIT/QC/GOLD/110MM STERILE-surgical, orthopedic, pneumatic powered & accessory/attachment Product Code: 310.25S
Where it went
Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Australia, Belgium, Canada, Czech Republic, Germany, Spain, Finland, France, United Kingdom, Ireland, South Korea, Kuwait, Mexico, Netherlands, Slovakia, Taiwan, South Africa.
How much
30 units

RIA Tube Assembly, minimum 520 mm length, sterile- intended for use in adults and adolescents (12-21 years) t

Z-1237-2024  ·  Class II  ·  reported 2024-03-13  ·  Ongoing

Why it was recalled
Products not sterilized, sterility cannot be confirmed
Product
RIA Tube Assembly, minimum 520 mm length, sterile- intended for use in adults and adolescents (12-21 years) to clear the medullary canal of bone marrow and debris and to effectively size the medullary canal for the acceptance of an intramedullary implant or prosthesis, to harvest finely morselized autogenous bone and bone marrow for any surgical procedure requiring bone graft to facilitate fusion and/or fill bone defects, and to remove infected and necrotic bone and tissue from the medullary canal in the treatment of osteomyelitis. Part Number: 314.746S
Where it went
Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Australia, Belgium, Canada, Czech Republic, Germany, Spain, Finland, France, United Kingdom, Ireland, South Korea, Kuwait, Mexico, Netherlands, Slovakia, Taiwan, South Africa.
How much
144 units

TFNA Femoral Nail 11mm, right, 130¿, L 235mm-Intended for temporary fixation and stabilization of proximal fe

Z-1240-2024  ·  Class II  ·  reported 2024-03-13  ·  Ongoing

Why it was recalled
Products not sterilized, sterility cannot be confirmed
Product
TFNA Femoral Nail 11mm, right, 130¿, L 235mm-Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.144S
Where it went
Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Australia, Belgium, Canada, Czech Republic, Germany, Spain, Finland, France, United Kingdom, Ireland, South Korea, Kuwait, Mexico, Netherlands, Slovakia, Taiwan, South Africa.
How much
11 units

TFN-Advanced Femoral Nail 12MM/125 DEG TI CANN TFNA 200MM STERILE. Intended for temporary fixation and stab

Z-1246-2023  ·  Class II  ·  reported 2023-03-22  ·  Ongoing

Why it was recalled
TFNA Femoral Nails labeled and etched as 200mm may measure 170mm in length, and TFNA Femoral Nails labeled and etched as 170mm may measure 200mm in length
Product
TFN-Advanced Femoral Nail 12MM/125 DEG TI CANN TFNA 200MM STERILE. Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.213S
Where it went
Worldwide - US Nationwide distribution in the states of MS, OR, TX and the countries of Canada, Japan.
How much
6 units

TFN-Advanced Femoral Nail 12MM/125 DEG TI CANN TFNA 170MM - STERILE. Intended for temporary fixation and stab

Z-1247-2023  ·  Class II  ·  reported 2023-03-22  ·  Ongoing

Why it was recalled
TFNA Femoral Nails labeled and etched as 200mm may measure 170mm in length, and TFNA Femoral Nails labeled and etched as 170mm may measure 200mm in length
Product
TFN-Advanced Femoral Nail 12MM/125 DEG TI CANN TFNA 170MM - STERILE. Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.212S
Where it went
Worldwide - US Nationwide distribution in the states of MS, OR, TX and the countries of Canada, Japan.
How much
5 units

TFN-Advanced Femoral Nail 12MM/130 DEG TI CANN TFNA 170MM STERILE. Intended for temporary fixation and stab

Z-1248-2023  ·  Class II  ·  reported 2023-03-22  ·  Ongoing

Why it was recalled
TFNA Femoral Nails labeled and etched as 200mm may measure 170mm in length, and TFNA Femoral Nails labeled and etched as 170mm may measure 200mm in length
Product
TFN-Advanced Femoral Nail 12MM/130 DEG TI CANN TFNA 170MM STERILE. Intended for temporary fixation and stabilization of proximal femur fractures Part Number: 04.037.242S
Where it went
Worldwide - US Nationwide distribution in the states of MS, OR, TX and the countries of Canada, Japan.
How much
6 units

SPEED 9X10X10mm Continuous Compression Implant, Product No. SE-0910

Z-0100-2023  ·  Class II  ·  reported 2022-10-26  ·  Ongoing

Why it was recalled
The subject products are being recalled because they contain the incorrect size drill guide and will not prepare the bone correctly for the size of the implant included in the sterile kit. If the incorrect drill guide is used, there is potential for bone damage due to incorrectly placed drill holes for the implant included in the kit.
Product
SPEED 9X10X10mm Continuous Compression Implant, Product No. SE-0910
Where it went
US Nationwide distribution.
How much
96

DePuy Synthes Hammertoe Continuous Compression Implants, Product No. 46.239.001

Z-0101-2023  ·  Class II  ·  reported 2022-10-26  ·  Ongoing

Why it was recalled
The subject products are being recalled because they contain the incorrect size drill guide and will not prepare the bone correctly for the size of the implant included in the sterile kit. If the incorrect drill guide is used, there is potential for bone damage due to incorrectly placed drill holes for the implant included in the kit.
Product
DePuy Synthes Hammertoe Continuous Compression Implants, Product No. 46.239.001
Where it went
US Nationwide distribution.
How much
196

SPEEDSHIFT20x20x20 Offset 10mm Implant Drill Kit, Product No. SE-2020-10. Used with DePuy Synthes SPEEDSHIFT

Z-0098-2023  ·  Class II  ·  reported 2022-10-26  ·  Ongoing

Why it was recalled
The subject products are being recalled because they contain the incorrect size drill guide and will not prepare the bone correctly for the size of the implant included in the sterile kit. If the incorrect drill guide is used, there is potential for bone damage due to incorrectly placed drill holes for the implant included in the kit.
Product
SPEEDSHIFT20x20x20 Offset 10mm Implant Drill Kit, Product No. SE-2020-10. Used with DePuy Synthes SPEEDSHIFT Continuous Compression Staple Implants for bone fixation and management of fracture and reconstructive surgery.
Where it went
US Nationwide distribution.
How much
81

SPEED 15X12X12mm Continuous Compression Implant, Product No. SE-1512

Z-0099-2023  ·  Class II  ·  reported 2022-10-26  ·  Ongoing

Why it was recalled
The subject products are being recalled because they contain the incorrect size drill guide and will not prepare the bone correctly for the size of the implant included in the sterile kit. If the incorrect drill guide is used, there is potential for bone damage due to incorrectly placed drill holes for the implant included in the kit.
Product
SPEED 15X12X12mm Continuous Compression Implant, Product No. SE-1512
Where it went
US Nationwide distribution.
How much
94 US; 30 OUS

Hammertoe Continuous Compression Implant (CCI) Kit Part Number: 46.239.001

Z-0008-2023  ·  Class II  ·  reported 2022-10-12  ·  Ongoing

Why it was recalled
Drill Pin (drill bit) is not compatible with the Drill Template included in the kit due to a packaging error, result in a surgical delay and may cause soft tissue damage if a larger incision is required to accommodate an alternative form of fixation. Additionally, there is a potential for adverse tissue reaction if the surgeon attempts to use the guide regardless of tight fit and plastic debris is created and ends up in the wound
Product
Hammertoe Continuous Compression Implant (CCI) Kit Part Number: 46.239.001
Where it went
US Nationwide distribution.
How much
191 units

Radial Head Replacement System Implant is a one-piece, stemmed radial head replacement. The radial head is ava

Z-1072-2022  ·  Class II  ·  reported 2022-05-18  ·  Ongoing

Why it was recalled
DePuy Synthes Radial Head Replacement System Contraindications added to Electronic Instructions for Use (e-IFU): Absolute contraindications include: Infection, sepsis, osteomyelitis.
Product
Radial Head Replacement System Implant is a one-piece, stemmed radial head replacement. The radial head is available in diameters of 19, 22 and 25mm. The radial stems range from 5.5 to 8.5mm in diameter and from 21 24mm in length.
Where it went
US Nationwide distribution.
How much
N/A

Universal Chuck-The Universal Chuck is an Instrument Handle provided as an alternative instrument available fo

Z-0634-2022  ·  Class II  ·  reported 2022-02-16  ·  Ongoing

Why it was recalled
Top cap may loosen and detach if the Universal Chuck becomes jammed and the user attempts to manually free the device, internal ball bearings may fall out onto the surgical field and cause surgical delay or, and if not recovered, adverse tissue reaction and infection.
Product
Universal Chuck-The Universal Chuck is an Instrument Handle provided as an alternative instrument available for use in the TN-ADVANCED" Tibial Nailing System (TNA), TFN-ADVANCED" Proximal Femoral Nailing System (TFNA), RFN-ADVANCED" Retrograde Femoral Nailing System (RFNA), FRN-ADVANCED" Femoral Recon Nailing System (FRNA), and Flexible Monobloc Reamers. Instrument Handles are intended for holding instruments for manipulation during orthopedic surgery. Product No.: 03.043.001
Where it went
US Nationwide distribution.
How much
600 units

DePuy Synthes 12.0MM REAMER HEAD FOR RIA 2 STERILE -Orthopedic manual surgical instrument Part Code: 03.404.0

Z-0575-2021  ·  Class II  ·  reported 2020-12-23  ·  Terminated

Why it was recalled
RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed
Product
DePuy Synthes 12.0MM REAMER HEAD FOR RIA 2 STERILE -Orthopedic manual surgical instrument Part Code: 03.404.020S
Where it went
Nationwide Foreign:countries: United Arab Emirates, Austria, Australia, Belgium, Switzerland, China, Czech Republic, Germany, Spain, United Kingdom, Ireland, Israel, Italy, Singapore,
How much
1566 US ; 732 OUS

DePuy Synthes 11.5MM REAMER HEAD FOR RIA 2 STERILE-Orthopedic manual surgical instrument Part Code: 03.404.01

Z-0574-2021  ·  Class II  ·  reported 2020-12-23  ·  Terminated

Why it was recalled
RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed
Product
DePuy Synthes 11.5MM REAMER HEAD FOR RIA 2 STERILE-Orthopedic manual surgical instrument Part Code: 03.404.019S
Where it went
Nationwide Foreign:countries: United Arab Emirates, Austria, Australia, Belgium, Switzerland, China, Czech Republic, Germany, Spain, United Kingdom, Ireland, Israel, Italy, Singapore,
How much
1231US; 422 OUS

DePuy Synthes 15.0MM REAMER HEAD FOR RIA 2 STERILE -Orthopedic manual surgical instrument Part Code: 03.404.

Z-0581-2021  ·  Class II  ·  reported 2020-12-23  ·  Terminated

Why it was recalled
RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed
Product
DePuy Synthes 15.0MM REAMER HEAD FOR RIA 2 STERILE -Orthopedic manual surgical instrument Part Code: 03.404.026S
Where it went
Nationwide Foreign:countries: United Arab Emirates, Austria, Australia, Belgium, Switzerland, China, Czech Republic, Germany, Spain, United Kingdom, Ireland, Israel, Italy, Singapore,
How much
748 US; 381 OUS

DePuy Synthes 13.0MM REAMER HEAD FOR RIA 2 STERILE-Orthopedic manual surgical instrument Part Code: 03.404.0

Z-0577-2021  ·  Class II  ·  reported 2020-12-23  ·  Terminated

Why it was recalled
RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed
Product
DePuy Synthes 13.0MM REAMER HEAD FOR RIA 2 STERILE-Orthopedic manual surgical instrument Part Code: 03.404.022S
Where it went
Nationwide Foreign:countries: United Arab Emirates, Austria, Australia, Belgium, Switzerland, China, Czech Republic, Germany, Spain, United Kingdom, Ireland, Israel, Italy, Singapore,
How much
1447 US; 554 OUS

DePuy Synthes 16.0MM REAMER HEAD FOR RIA 2 STERILE--Orthopedic manual surgical instrument Part Code: 03.404.0

Z-0583-2021  ·  Class II  ·  reported 2020-12-23  ·  Terminated

Why it was recalled
RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed
Product
DePuy Synthes 16.0MM REAMER HEAD FOR RIA 2 STERILE--Orthopedic manual surgical instrument Part Code: 03.404.028S
Where it went
Nationwide Foreign:countries: United Arab Emirates, Austria, Australia, Belgium, Switzerland, China, Czech Republic, Germany, Spain, United Kingdom, Ireland, Israel, Italy, Singapore,
How much
651US : 326 OUS

DePuy Synthes 16.5MM REAMER HEAD FOR RIA 2 STERILE-Orthopedic manual surgical instrument Part Code: 03.404.02

Z-0584-2021  ·  Class II  ·  reported 2020-12-23  ·  Terminated

Why it was recalled
RIA 2 reamer head breakage exceeds the expected occurrence rate resulting in adverse tissue reaction, soft tissue damage, and/or bone damage, when the recommended surgical techniques are not followed
Product
DePuy Synthes 16.5MM REAMER HEAD FOR RIA 2 STERILE-Orthopedic manual surgical instrument Part Code: 03.404.029S
Where it went
Nationwide Foreign:countries: United Arab Emirates, Austria, Australia, Belgium, Switzerland, China, Czech Republic, Germany, Spain, United Kingdom, Ireland, Israel, Italy, Singapore,
How much
531 US; 317 OUS

Showing the 30 most recent of 164. The rest are searchable at the FDA enforcement report database.

Read these counts with three things in mind

A firm with more recalls usually makes more products. A contract packer runs hundreds of lines. It will out-recall a company with one product whatever either does. This is not a safety ranking.

Most recalls here are voluntary. A firm found its own fault and pulled the stock. That is the system working. It lands in this database looking the same as a recall the FDA had to push for.

Terminated means the FDA considers the recall finished. It is not a verdict. It does not mean anyone was harmed.

Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.

← All device recalls