Synthes, Inc.

Synthes, Inc.: 43 FDA recall reports

Every medical device recall the FDA has published for Synthes, Inc.. The fault, and how the agency graded it.

43Reports
1Class I
0Ongoing
2020-08-05Most recent
1 Class I on this page

Class I means the FDA judged there was a reasonable probability of serious harm or death. Check the lot numbers against anything you hold.

What the classes mean

The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.

ClassWhat the FDA is saying
Class IThere is a reasonable probability that using this will cause serious harm or death.
Class IIIt may cause temporary or reversible harm. Serious harm is unlikely.
Class IIIIt breaks a rule but is unlikely to hurt anyone. Often a labelling fault.

Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.

By class

ClassReports
Class II42
Class I1

By year reported

YearReports
201419
201515
20183
20206
Checking whether it affects you

The lot or code number is what matters. Not the product name. A recall covers specific lots. Yours may not be one of them.

Check the code on your package against the report. The FDA keeps the current list at fda.gov/safety/recalls.

The reports

The text below is quoted from the FDA record word for word.

MatrixNEURO Screws - Product Usage: The intended use for this device is in fixation of cranial bones and facia

Z-2698-2020  ·  Class II  ·  reported 2020-08-05  ·  Terminated

Why it was recalled
One lot of MatrixNEURO 5mm screws was packaged in a body clip etched "4", which indicated a 4mm screw. The outer package labeling correctly identifies a 5mm screw, but the body clop incorrectly identifies a shorter length for the screw.
Product
MatrixNEURO Screws - Product Usage: The intended use for this device is in fixation of cranial bones and facial skeleton in procedures such as craniotomies, reconstructive procedures, fracture repair, osteotomies, and selective orthognathic surgery of the maxilla and chin.
Where it went
International distribution including in the countries of Australia, Belgium, Chile, France, Slovakia, South Korea.
How much
104 units

2.4MM TI VA LOCKING SCREW STARDRIVE 18MM STERILE, Part Number 04.210.118TS - Product Usage:Sterile Tube Packag

Z-0882-2020  ·  Class II  ·  reported 2020-02-05  ·  Terminated

Why it was recalled
There is a possibility that the inner cap could open when the outer cap is removed or the entire inner tube remains in the outer tube. In both instances this could prevent the screw from being removed and used causing a surgical delay. The inner tube should remain closed and retained in the outer cap when presented to the sterile field.
Product
2.4MM TI VA LOCKING SCREW STARDRIVE 18MM STERILE, Part Number 04.210.118TS - Product Usage:Sterile Tube Packaging is a packaging system to deliver single, sterile screws to the sterile field.
Where it went
US Nationwide distribution in the states of AL and ID.
How much
25 total

2.4MM TI VA LOCKING SCREW STARDRIVE 10MM STERILE, Part Number 04.210.110TS - Product Usage:Sterile Tube Packag

Z-0879-2020  ·  Class II  ·  reported 2020-02-05  ·  Terminated

Why it was recalled
There is a possibility that the inner cap could open when the outer cap is removed or the entire inner tube remains in the outer tube. In both instances this could prevent the screw from being removed and used causing a surgical delay. The inner tube should remain closed and retained in the outer cap when presented to the sterile field.
Product
2.4MM TI VA LOCKING SCREW STARDRIVE 10MM STERILE, Part Number 04.210.110TS - Product Usage:Sterile Tube Packaging is a packaging system to deliver single, sterile screws to the sterile field.
Where it went
US Nationwide distribution in the states of AL and ID.
How much
25 total

2.4MM TI VA LOCKING SCREW STARDRIVE 24MM STERILE, Part Number 04.210.124TS - Product Usage:Sterile Tube Packag

Z-0883-2020  ·  Class II  ·  reported 2020-02-05  ·  Terminated

Why it was recalled
There is a possibility that the inner cap could open when the outer cap is removed or the entire inner tube remains in the outer tube. In both instances this could prevent the screw from being removed and used causing a surgical delay. The inner tube should remain closed and retained in the outer cap when presented to the sterile field.
Product
2.4MM TI VA LOCKING SCREW STARDRIVE 24MM STERILE, Part Number 04.210.124TS - Product Usage:Sterile Tube Packaging is a packaging system to deliver single, sterile screws to the sterile field.
Where it went
US Nationwide distribution in the states of AL and ID.
How much
25 total

2.4MM TI VA LOCKING SCREW STARDRIVE 16MM STERILE, Part Number 04.210.116TS - Product Usage:Sterile Tube Packag

Z-0881-2020  ·  Class II  ·  reported 2020-02-05  ·  Terminated

Why it was recalled
There is a possibility that the inner cap could open when the outer cap is removed or the entire inner tube remains in the outer tube. In both instances this could prevent the screw from being removed and used causing a surgical delay. The inner tube should remain closed and retained in the outer cap when presented to the sterile field.
Product
2.4MM TI VA LOCKING SCREW STARDRIVE 16MM STERILE, Part Number 04.210.116TS - Product Usage:Sterile Tube Packaging is a packaging system to deliver single, sterile screws to the sterile field.
Where it went
US Nationwide distribution in the states of AL and ID.
How much
25 total

2.4MM TI VA LOCKING SCREW STARDRIVE 12MM STERILE, Part Number 04.210.112TS - Product Usage:Sterile Tube Packag

Z-0880-2020  ·  Class II  ·  reported 2020-02-05  ·  Terminated

Why it was recalled
There is a possibility that the inner cap could open when the outer cap is removed or the entire inner tube remains in the outer tube. In both instances this could prevent the screw from being removed and used causing a surgical delay. The inner tube should remain closed and retained in the outer cap when presented to the sterile field.
Product
2.4MM TI VA LOCKING SCREW STARDRIVE 12MM STERILE, Part Number 04.210.112TS - Product Usage:Sterile Tube Packaging is a packaging system to deliver single, sterile screws to the sterile field.
Where it went
US Nationwide distribution in the states of AL and ID.
How much
25 total

Flex Arm Product Usage: The Synthes Flex Arm is a component of the Synthes Minimally Invasive Support Syst

Z-1511-2018  ·  Class II  ·  reported 2018-05-02  ·  Terminated

Why it was recalled
Quick connect feature of instruments may be unable to connect to the male features of Flex Arm Adaptors, Insight Retractors, or other mating parts.
Product
Flex Arm Product Usage: The Synthes Flex Arm is a component of the Synthes Minimally Invasive Support System, which provides secure table mounting for minimally invasive spine surgery. The Flex Arm is designed to hold Synthes Minimally Invasive Retractors, which are connected to the Flex Arm by adapters.
Where it went
Worldwide distribution - US Nationwide and the countries of Canada, Switzerland and Japan
How much
84

Stem Extractor f/Guide Bar, Part number 03.401.072. Instrumentation for extraction and revision of Epoca shou

Z-1514-2018  ·  Class II  ·  reported 2018-05-02  ·  Terminated

Why it was recalled
There is a potential for the connection screw of the stem extractor to break.
Product
Stem Extractor f/Guide Bar, Part number 03.401.072. Instrumentation for extraction and revision of Epoca shoulder implants.
Where it went
Worldwide Distribution: US (nationwide) and to countries of: Canada and Switzerland.
How much
31

LCP Drill Sleeve 1.5 Drill Bit 1.1mm Part Number 03.114.001 Used in LCP Modular Mini Fragment System, for

Z-0669-2018  ·  Class II  ·  reported 2018-02-28  ·  Terminated

Why it was recalled
May have difficulty attaching the drill sleeve to the plate. If the drill sleeve will not connect to the plate intraoperatively, surgical delay could occur. If the pilot hole is drilled without the drill sleeve, the pilot hole could be off-center which could result in screw and/or plate loosening.
Product
LCP Drill Sleeve 1.5 Drill Bit 1.1mm Part Number 03.114.001 Used in LCP Modular Mini Fragment System, for fixation of small bones and small bone fragments
Where it went
Distribution to US, Canada, and Germany.
How much
567 units

Reamer / Irrigator / Aspirator (RIA): RIA Drive Shafts, RIA Tube Assemblies, sterile, RIA Reamer Head, steril

Z-2092-2015  ·  Class II  ·  reported 2015-07-22  ·  Terminated

Why it was recalled
It was reported that the RIA Drive Shaft, Tube Assembly, and Reamer Head have the potential to break when incorrectly assembled or used improperly. Precautionary statements are being added to the Reamer / Irrigator / Aspirator (RIA) Surgical Technique Guide.
Product
Reamer / Irrigator / Aspirator (RIA): RIA Drive Shafts, RIA Tube Assemblies, sterile, RIA Reamer Head, sterile (12.0mm-19.0mm); intended to clear the medullary canal of bone marrow and debris and to effectively size the medullary canal for the acceptance of an intramedullary implant or prosthesis
Where it went
Nationwide Distribution and Internationally to Canada and Switzerland.
How much
132564

TITANIUM END CAP WITH T40 STARDRIVE RECESS, 0MM EXTENSION FOR TITANIUM TIBIAL NAILS-EX; Synthes Tibial Nail Sy

Z-2051-2015  ·  Class II  ·  reported 2015-07-15  ·  Terminated

Why it was recalled
A specific part and lot number was packaged and shipped prior to the completion of a required internal inspection.
Product
TITANIUM END CAP WITH T40 STARDRIVE RECESS, 0MM EXTENSION FOR TITANIUM TIBIAL NAILS-EX; Synthes Tibial Nail System EX is intended to stabilize fractures of the proximal and distal tibia and the tibial shaft; open and closed tibial shaft fractures; certain pre-and post-isthmic fractures; and tibial malunions and non-unions.
Where it went
US Distribution to states of: MO and SC.
How much
3

DePuy Synthes 6.5 mm Titanium Recon Screws w/T25 Stardrive (70mm, 75mm, 80mm, 85mm, 90mm, 95mm, 105mm); part

Z-1915-2015  ·  Class II  ·  reported 2015-07-08  ·  Terminated

Why it was recalled
certain lots of the Titanium Recon Screws (6.5mm ) were found to contain the incorrect insert GP0760-A (Synthes Titanium Buttress Locking Plate System), instead of the correct insert GP0706-E (Synthes Titanium Intramedullary Nails and Components).
Product
DePuy Synthes 6.5 mm Titanium Recon Screws w/T25 Stardrive (70mm, 75mm, 80mm, 85mm, 90mm, 95mm, 105mm); part of the Expert Lateral Femur Nail System and indicated to stabilize Subtrochanteric fractures, Ipsilateral neck/shaft fractures, Femoral shaft fractures, Impending pathologic fractures, and malunions/Non-unions.
Where it went
Nationwide Distribution.
How much
276

3.7mm Cannulated Locking Screws and 3.7mm Cannulated Conical Screws; . the Cannulated Screws are indicated fo

Z-1800-2015  ·  Class II  ·  reported 2015-06-24  ·  Terminated

Why it was recalled
Certain lots of the 3.7mm Cannulated Locking Screws and 3.7mm Cannulated Conical Screws were found to have been labeled incorrectly.
Product
3.7mm Cannulated Locking Screws and 3.7mm Cannulated Conical Screws; . the Cannulated Screws are indicated for fixation of fractures, osteotomies, and non-unions of the clavicle, scapula, olecranon, humerus, radius, ulna, pelvis, tibia, and fibula, particularly in osteopenic bone.
Where it went
Nationwide Distribution
How much
54

Pull Reduction Device For 4.3mm Percutaneous Drill Guide; The pull reduction device with quick coupling is pla

Z-1799-2015  ·  Class II  ·  reported 2015-06-24  ·  Terminated

Why it was recalled
Certain affected parts and lots of the Pull Reduction Devices may have been manufactured to an incorrect hardness specification which could result in intraoperative breakage.
Product
Pull Reduction Device For 4.3mm Percutaneous Drill Guide; The pull reduction device with quick coupling is placed through the guide sleeve and plate holes to pull or push bone fragments relative to the plate. This instrument can be used for: Minor Varus-valgus adjustments (approximately 2¿ - 4¿), Translational adjustments, Stabilization of plate bone orientation during insertion of the first screws, alignment of segmental fragments and pre-drilling dense or thick cortical bone before placing a 5.0 mm locking screw.
Where it went
Worldwide Distribution-US (nationwide) and internationlly to Canada
How much
3747 US, 19 Foreign

3.2 mm Cannulated Drill Bit/ QC 170 mm 2.7 Cannulated Drill Bit/ QC 160 mm 2.0 mm Cannulated Drill Bit / QC

Z-1711-2015  ·  Class II  ·  reported 2015-06-10  ·  Terminated

Why it was recalled
affected parts and lots of the Cannulated Drill Bits have the potential to break during use
Product
3.2 mm Cannulated Drill Bit/ QC 170 mm 2.7 Cannulated Drill Bit/ QC 160 mm 2.0 mm Cannulated Drill Bit / QC 150 mm there is no indication for Synthes Cannulated Drill Bits which is used with the Synthes Cannulated Screw. The Synthes Cannulated Screw, used with the threaded washer, is generally intended for intra-articular fixation of small bones, such as the hand, wrist, and forefoot.
Where it went
Worldwide Distribution - US (nationwide) and Canada.
How much
1495 US (37 OUS)

TFN-ADVANCED Proximal Femoral Nailing System (TFNA); for treatment of fractures in orthopedics and trauma.

Z-1600-2015  ·  Class II  ·  reported 2015-05-20  ·  Terminated

Why it was recalled
TFNA nails from certain lots were assembled with a locking mechanism too close to the top of the nail. Potential patient impact: surgical delay.
Product
TFN-ADVANCED Proximal Femoral Nailing System (TFNA); for treatment of fractures in orthopedics and trauma.
Where it went
Distribution US Nationwide and Switzerland.
How much
8175

Synthes 130 Degree Aiming Arm and 125 Degree Aiming Arm; The 130 Degree Aiming Arm and 125 Degree Aiming Arm a

Z-1352-2015  ·  Class II  ·  reported 2015-04-08  ·  Terminated

Why it was recalled
The guide sleeve has the potential to jam in the Aiming Arms, which may delay disassembly and removal of instruments.
Product
Synthes 130 Degree Aiming Arm and 125 Degree Aiming Arm; The 130 Degree Aiming Arm and 125 Degree Aiming Arm are a part of the TFN-ADVANCED (TFNA) System. The TFN-ADVANCED Proximal Femoral Nailing System is intended for treatment of fractures in adults and adolescents (12-21) in which the growth plates have fused.
Where it went
US Nationwide Distribution - NY, TX, MD, OR, MS, ND, CA, AZ, MI, IL, PA, TN, AR, SD, SC, FL, VA, OK, and GA
How much
131

Synthes Electric Pen Drive (EPD) System, The Synthes Electric Pen Drive is indicated for screw insertion, pin

Z-1313-2015  ·  Class II  ·  reported 2015-04-01  ·  Terminated

Why it was recalled
In May 2013, the original recall was initiated due to complaints the Hand Switch for the Electric Pen Drive was not labeled correctly, and the on and lock labels were interchanged.
Product
Synthes Electric Pen Drive (EPD) System, The Synthes Electric Pen Drive is indicated for screw insertion, pin and wire replacement, cutting of bone and metal, drilling, decorticating, shaping, and smoothing of bones and teeth in a wide variety of surgical procedures, including, but not limited to general orthopaedic trauma, foot, hand, maxillofacial, neurosurgical, oral, otolaryngolical, reconstructive, and spine surgery.
Where it went
Nationwide Distribution
How much
249

Thumb Screw for the Synthes Recon Locking Aiming Arm for Lateral Entry Femoral Nails-EX (part number 03.010.04

Z-1281-2015  ·  Class II  ·  reported 2015-03-25  ·  Terminated

Why it was recalled
For certain lots, the incorrect raw material of annealed 17-4PH Stainless Steel was used instead of the correct material of heat-treated 304 Stainless Steel. This may lead to stress-corrosion cracking, which can lead to surgical delay or unintended debris entering the surgical site.
Product
Thumb Screw for the Synthes Recon Locking Aiming Arm for Lateral Entry Femoral Nails-EX (part number 03.010.048) and the Thumb Screw for the Aiming Arm for Titanium Cannulated Tibial Nails-EX (part number 03.010.052). The Recon Locking Aiming Arm for Lateral Entry Femoral Recon Nails- EX is used when locking the Femoral Nail-EX. The Aiming Arm for Titanium (TI) Cannulated Tibial Nails-EX is used when locking the Tibial Nail-EX.
Where it went
Nationwide Distribution
How much
183

Synthes Inserter for Titanium Elastic Nails; The Inserter for Titanium Elastic Nails (TEN) is intended for fix

Z-1260-2015  ·  Class II  ·  reported 2015-03-18  ·  Terminated

Why it was recalled
The Inserter has the potential for mechanical failures such as breakage of the main shaft of the Inserter for the TEN, malfunction of the chuck of the Inserter for the TEN in the form of the chuck getting stuck/jammed, breakage of the cross bar of the Inserter for the TEN.
Product
Synthes Inserter for Titanium Elastic Nails; The Inserter for Titanium Elastic Nails (TEN) is intended for fixation of diaphyseal fractures where the canal is narrow or flexibility of the implant is paramount. This includes: lower extremity fractures in pediatric and small-statured patients and upper extremity fractures in patients. This system is also intended to treat metaphyseal and episphyseal fractures, such as radial neck fractures, and is intended for fixation of small long bones, such as carpal and tarsal bones. In pediatric applications, the flexibility of the elastic nail allows it to be inserted at a point which avoids disruption of the bone growth plate.
Where it went
Nationwide Distribution
How much
1550

Hemostatic Bone Putty, for use as a water-soluble implant material and for use in the control of bleeding fro

Z-1094-2015  ·  Class II  ·  reported 2015-02-18  ·  Terminated

Why it was recalled
Revision A of the Hemostatic Bone Putty Brochure (J10847A) was sent to sales consultants instead of Revision B (J10847B), which corrected an error in the indication statement.
Product
Hemostatic Bone Putty, for use as a water-soluble implant material and for use in the control of bleeding from bone surfaces.
Where it went
Distributed in the state of WI
How much
38

The Synthes Mandible External Fixator; all lots of part nos.: 04.305.003 04.305.004 04.305.005 04.305.0

Z-1009-2015  ·  Class II  ·  reported 2015-02-04  ·  Terminated

Why it was recalled
Under new testing protocols for MR Environment safety and compatibility, metal devices are no longer designated MR Safe. Some components in the CMF Mandible External Fixator Systems were labeled and/or etched with information indicating they are MR Safe. They can no longer be labeled MR Safe.
Product
The Synthes Mandible External Fixator; all lots of part nos.: 04.305.003 04.305.004 04.305.005 04.305.010 04.305.011 04.305.012 03.305.006 04.305.100 04.305.101 04.305.102 04.305.103 04.305.128 04.305.130 04.305.134 04.305.138 04.305.140 04.305.142 03.305.105 03.305.500 03.305.500S Intended to stabilize and provide treatment for fractures of the maxillofacial area, including severe open mandibular fractures, highly comminuted closed fractures, nonunions and delayed unions (especially associated with infection), fractures associated with infections, tumor resections, facial deformity corrections, gunshot wounds, pan facial fractures, burn maintenance, and bone grafting defects.
Where it went
Worldwide Distribution.
How much
43159

2.0 Mandible Locking Plate (MLP) Bender / Cutter (part number 329.143) Intended to be used for bending and/

Z-1021-2015  ·  Class II  ·  reported 2015-02-04  ·  Terminated

Why it was recalled
Certain lots have the potential for tip breakage while bending 2.0 mm thick plates. If device is used and breaks while in the Operating Room, surgical delay may occur while alternate bending or cutting instruments are located.
Product
2.0 Mandible Locking Plate (MLP) Bender / Cutter (part number 329.143) Intended to be used for bending and/or cutting Mandible Locking Plates.
Where it went
Worldwide Distribution -- US and Canada.
How much
973

Malleable C- Retractor, Part Number 397.232, used to provide retraction of the soft tissues during use of troc

Z-0973-2015  ·  Class II  ·  reported 2015-01-21  ·  Terminated

Why it was recalled
Potential failure and/or corrosion of the internal spring of the retractor, which may have been manufactured from an incorrect material. Surgical delay may occur. Spring particles could transfer from the instrument to the patient, and may lead to an adverse tissue reaction and/or infection.
Product
Malleable C- Retractor, Part Number 397.232, used to provide retraction of the soft tissues during use of trocar instrumentation. Used in the following systems: the Basic Trocar System, Universal Trocar System, Craniofacial Modular Fixation System, Craniofacial (CMF) Distraction System, MatrixORTHOGNATHIC Plating System, and the Matrix MANDIBLE Plating System.
Where it went
Worldwide Distribution: US (nationwide) and internationally to Canada.
How much
1762

10 mm/135 degree Titanium Cannulated Trochanteric Fixation Nail 170 MM - Sterile: intended to treat stable and

Z-0219-2015  ·  Class II  ·  reported 2014-11-26  ·  Terminated

Why it was recalled
The 10 mm/135 degree Titanium Cannulated Trochanteric Fixation Nail 170 MM (Sterile) was assembled incorrectly. If the non conformance goes undetected, it may impact the locking function post-operatively and compromise the bone reduction and construct stability which may lead to non-union/malunion.
Product
10 mm/135 degree Titanium Cannulated Trochanteric Fixation Nail 170 MM - Sterile: intended to treat stable and unstable fractures of the proximal femur
Where it went
US Distribution in the states of: AK, FL, OH, OK, and TX.
How much
6

DePuy Synthes Radiolucent Retractor, part number 387.580, lot number 3620715

Z-0092-2015  ·  Class II  ·  reported 2014-10-29  ·  Terminated

Why it was recalled
Particle residue on the instrument from adhesive tape which was used to bind the device during transport. There may be a risk of tissue reaction if a patient is exposed to latex particles from the adhesive tape, or a systemic reaction such as anaphylaxis in patients with latex allergy.
Product
DePuy Synthes Radiolucent Retractor, part number 387.580, lot number 3620715
Where it went
US Distribution including the states of PA, NC, TN, NY, WI, WA and MI.
How much
11

DePuy Synthes TI Vectra Plates: TI Vectra-One" Plate 2 Level / 34 mm part 04.613.184 3337672 TI Vect

Z-2674-2014  ·  Class II  ·  reported 2014-10-01  ·  Terminated

Why it was recalled
DePuy Synthes is initiating a voluntary medical device recall of certain lots of the TI Vectra Plates, which is a part of the DePuy Synthes Vectra Anterior Cervical Plate Family of Systems. The Vectra Systems are intended for anterior screw fixation to the cervical spine (C2-C7). It was discovered in certain lots, that the clips are potentially missing from the TI Vectra Plates resulting in the i
Product
DePuy Synthes TI Vectra Plates: TI Vectra-One" Plate 2 Level / 34 mm part 04.613.184 3337672 TI Vectra" Plate 3 Level / 57 mm part 04.613.257 lot 3693878 TI Vectra" Plate 4 Level / 64 mm part 04.613.364 lot 8143905
Where it went
US Distribution including the states of IL, SC, VA, TX, MA, NC, LA, CA, AL, CO, NE, OR and KS., and Internationally to Canada.
How much
16
Class I

DePuy Synthes Craniomaxillofacial Distraction System (AB Distractor Bodies and BC Distractor Bodies)

Z-2148-2014  ·  Class I  ·  reported 2014-08-27  ·  Terminated

Why it was recalled
DePuy Synthes is initiating a recall of certain lots of the Craniomaxillofacial Distraction System (AB Distractor Bodies and BC Distractor Bodies) because they may reverse post-operatively.
Product
DePuy Synthes Craniomaxillofacial Distraction System (AB Distractor Bodies and BC Distractor Bodies)
Where it went
US Nationwide Distribution
How much
1499

Synthes 3.0 MM TI Cannulated Screw Instrument and Implant Set

Z-2265-2014  ·  Class II  ·  reported 2014-08-27  ·  Terminated

Why it was recalled
A screw set was returned from the field for destruction after being exposed to flood water, which was erroneously reprocessed and returned to the field. Initial investigative testing indicated the components within the set may be cytotoxic.
Product
Synthes 3.0 MM TI Cannulated Screw Instrument and Implant Set
Where it went
Distributed in the states of WI and MI.
How much
4

Synthes Guide Blocks for the 2 Column Plate 6 Hole Head (Volar Distal Radius Plate System). Orthopedic surgic

Z-2087-2014  ·  Class II  ·  reported 2014-07-30  ·  Terminated

Why it was recalled
The Guide Blocks for the 2 Column Plate 6 Hole Head (Volar Distal Radius Plate System) were found to be labeled incorrectly. Part no. 03.111.600 (Right) was found in the package labeled part no. 03.111.601 (Left). The laser etching denoting the orientation on the part is correct.
Product
Synthes Guide Blocks for the 2 Column Plate 6 Hole Head (Volar Distal Radius Plate System). Orthopedic surgical instrument.
Where it went
Distribution US Nationwide and Bermuda.
How much
180

Showing the 30 most recent of 43. The rest are searchable at the FDA enforcement report database.

Read these counts with three things in mind

A firm with more recalls usually makes more products. A contract packer runs hundreds of lines. It will out-recall a company with one product whatever either does. This is not a safety ranking.

Most recalls here are voluntary. A firm found its own fault and pulled the stock. That is the system working. It lands in this database looking the same as a recall the FDA had to push for.

Terminated means the FDA considers the recall finished. It is not a verdict. It does not mean anyone was harmed.

Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.

← All device recalls