Stryker Spine

Stryker Spine: 29 FDA recall reports

Every medical device recall the FDA has published for Stryker Spine. The fault, and how the agency graded it.

29Reports
1Class I
6Ongoing
2025-04-09Most recent
1 Class I on this page

Class I means the FDA judged there was a reasonable probability of serious harm or death. Check the lot numbers against anything you hold.

What the classes mean

The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.

ClassWhat the FDA is saying
Class IThere is a reasonable probability that using this will cause serious harm or death.
Class IIIt may cause temporary or reversible harm. Serious harm is unlikely.
Class IIIIt breaks a rule but is unlikely to hurt anyone. Often a labelling fault.

Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.

By class

ClassReports
Class II28
Class I1

By year reported

YearReports
20128
20133
20144
20152
20162
20181
20213
20231
20241
20254
Checking whether it affects you

The lot or code number is what matters. Not the product name. A recall covers specific lots. Yours may not be one of them.

Check the code on your package against the report. The FDA keeps the current list at fda.gov/safety/recalls.

The reports

The text below is quoted from the FDA record word for word.

Monterey AL Implant Inserter; 14/16mm; Catalog 48019120.

Z-1489-2025  ·  Class II  ·  reported 2025-04-09  ·  Ongoing

Why it was recalled
Potential for the gold unlock button to separate from the inserter.
Product
Monterey AL Implant Inserter; 14/16mm; Catalog 48019120.
Where it went
Domestic: CA, CO, FL, GA, IL, MA, NE, NH, NY, OK, PA, TX, UT. International: Canada.
How much
56 units (53 US, 3 OUS)

Monterey AL Implant Inserter; 10/12mm; Catalog 48019100.

Z-1488-2025  ·  Class II  ·  reported 2025-04-09  ·  Ongoing

Why it was recalled
Potential for the gold unlock button to separate from the inserter.
Product
Monterey AL Implant Inserter; 10/12mm; Catalog 48019100.
Where it went
Domestic: CA, CO, FL, GA, IL, MA, NE, NH, NY, OK, PA, TX, UT. International: Canada.
How much
64 units (60 US, 4 OUS)

Monterey AL Implant Inserter; 18/20mm; Catalog 48019130.

Z-1490-2025  ·  Class II  ·  reported 2025-04-09  ·  Ongoing

Why it was recalled
Potential for the gold unlock button to separate from the inserter.
Product
Monterey AL Implant Inserter; 18/20mm; Catalog 48019130.
Where it went
Domestic: CA, CO, FL, GA, IL, MA, NE, NH, NY, OK, PA, TX, UT. International: Canada.
How much
46 units (44 US, 2 OUS)

Monterey AL Implant Inserter; 22mm; Catalog 48019140.

Z-1491-2025  ·  Class II  ·  reported 2025-04-09  ·  Ongoing

Why it was recalled
Potential for the gold unlock button to separate from the inserter.
Product
Monterey AL Implant Inserter; 22mm; Catalog 48019140.
Where it went
Domestic: CA, CO, FL, GA, IL, MA, NE, NH, NY, OK, PA, TX, UT. International: Canada.
How much
44 units (41 US, 3 OUS)

AVS Anchor-C Cervical Cage System Instructions For Use (IFU)- eIFU (revision 5)-Indicated for anterior cervica

Z-3210-2024  ·  Class II  ·  reported 2024-10-02  ·  Ongoing

Why it was recalled
Labeling Correction: Instructions For Use (IFU)-Updated electronic Rev 5, specific language for indications for use for the United States and Canada was inadvertently removed.
Product
AVS Anchor-C Cervical Cage System Instructions For Use (IFU)- eIFU (revision 5)-Indicated for anterior cervical interbody fusion procedures in skeletally mature patients with cervical disc disease at one level from the C2-C3 disc to the C7-T1 disc.
Where it went
Worldwide distribution - US Nationwide and the country of Canada.
How much
8,589 units

LITe Decompression Snake Arm, REF 48080230

Z-0610-2024  ·  Class II  ·  reported 2023-12-27  ·  Ongoing

Why it was recalled
Product arm was manufactured with the incorrect clamp component. As a result, the affected product arm cannot attach to the arm post shaft.
Product
LITe Decompression Snake Arm, REF 48080230
Where it went
US: MD, MI, NM, TX & OUS: Canada & France.
How much
9

Cortoss Bone Augmentation Material 5cc Cartridge, Catalog Number 2101-0005, GTIN 00808232000962

Z-2100-2021  ·  Class II  ·  reported 2021-07-28  ·  Terminated

Why it was recalled
The products experienced a 12-hour excursion from the required refrigerated temperature conditions at the distribution center prior to distribution.
Product
Cortoss Bone Augmentation Material 5cc Cartridge, Catalog Number 2101-0005, GTIN 00808232000962
Where it went
US Nationwide Distribution
How much
559 total

Cortoss Bone Augmentation Material 10cc Cartridge, Catalog Number 2101-0010, GTIN 00808232000979

Z-2101-2021  ·  Class II  ·  reported 2021-07-28  ·  Terminated

Why it was recalled
The products experienced a 12-hour excursion from the required refrigerated temperature conditions at the distribution center prior to distribution.
Product
Cortoss Bone Augmentation Material 10cc Cartridge, Catalog Number 2101-0010, GTIN 00808232000979
Where it went
US Nationwide Distribution
How much
559 total

Tritanium X TL Instruments - intervertebral fusion device with bone graft, lumbar. Catalogs:TL Inserter (58022

Z-1437-2021  ·  Class II  ·  reported 2021-04-28  ·  Terminated

Why it was recalled
Stryker identified non-conforming instruments that are components of the Tritanium X TL sets. The firm found non-conformances with the TL Inserter, TL Draw Rod, Tube Set/Syringe Adapter Assembly, and Syringe Plunger that could lead to device failure during a procedure.
Product
Tritanium X TL Instruments - intervertebral fusion device with bone graft, lumbar. Catalogs:TL Inserter (580220100), TL Draw Rod (580220101), Tube Set/Syringe Adapter Assembly (680220100), Syringe Plunger (680220102).
Where it went
US Nationwide distribution in the states of AZ, CA, CO, GA, KY, MA, MT, NC, NE, NJ, NY, OH, PA, WY.
How much
169 sets

AVS Aria Cage, Catalog Number 48753212 Product Usage: The Stryker Spine AVS¿ ARIA 1M PEEK Spacers are inte

Z-0092-2019  ·  Class II  ·  reported 2018-10-17  ·  Terminated

Why it was recalled
The anterior and posterior markers on the cage were on the incorrect/opposite sides of the cage.
Product
AVS Aria Cage, Catalog Number 48753212 Product Usage: The Stryker Spine AVS¿ ARIA 1M PEEK Spacers are intervertebral body fusion devices indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one level or two contiguous levels from L2 to S1 . The AVS¿ ARIA TM PEEK Spacers are intended to be used with supplemental fixation systems that have been cleared for use in the lumbosacral spine.
Where it went
US Nationwide in the states: CA, FL, GA, MI, MN, NJ, NM, and OR
How much
5

Stryker Aero-AL Impaction handle Catalog Number 48921007. Stryker Spine Manufactured by : Stryker Spine SAS F

Z-0828-2016  ·  Class II  ·  reported 2016-03-02  ·  Terminated

Why it was recalled
Reported jamming of the impaction handle when assembled to the Inserter guide at the threads.
Product
Stryker Aero-AL Impaction handle Catalog Number 48921007. Stryker Spine Manufactured by : Stryker Spine SAS France Used to facilitate insertion of the Aero-AL cage into the disc space.
Where it went
Nationwide Distribution
How much
31 units

AccuLIF PL Tubing Set and AccuLIF TL Tubing Set AccuLIF is an interbody device system that uses hydraulic p

Z-0852-2016  ·  Class II  ·  reported 2016-03-02  ·  Terminated

Why it was recalled
There were two potential interference conditions identified with the way the tubing set attaches to the inserter.
Product
AccuLIF PL Tubing Set and AccuLIF TL Tubing Set AccuLIF is an interbody device system that uses hydraulic pressure to expand the implant in-situ. The purpose of the AccuLIF PL Tubing Set (cat#900301) and AccuLIF TL Tubing Set (cat#905103) is to deliver pressurized saline from a syringe to the AccuLIF implant for in-situ expansion. The Tubing Set attaches to the AccuLIF Insertion handle at two district locations: the "junction block" and the "manifold." The PL and TL Tubing Sets are Class 2 instruments not intended to sustain or support life.
Where it went
US Nationwide Distribution
How much
413 units

Stryker Radius 5.5 to 3.5 Rod to Red Connector, REF 48664530. Intended to provide additional support during

Z-0270-2016  ·  Class II  ·  reported 2015-11-18  ·  Terminated

Why it was recalled
Stryker Spine is recalling Radius Rod to Rod Connector 5.5 to 3.5 mm (Catalog #486614530) because it does not have an associated 510(k) submisstion and subsequent clearance, which is required as this is a class 2 device.
Product
Stryker Radius 5.5 to 3.5 Rod to Red Connector, REF 48664530. Intended to provide additional support during fusion using autograft or allograft in skeletally mature patients.
Where it went
Nationwide Distribution including the states of AK, AL, AR, AZ, CA, CO, FL, GA, IA, IL, IN, KS, KY, MA, MD, MI, MN, MO, MT, NC, NM, NY, OH, OK, OR, PA, TN, TX, UT, WA and WI.
How much
42 units

Stryker UniVise Spinous Process Fixation Plate System Inserter, Stryker Spine, Non-Sterile. Reference Number

Z-0222-2016  ·  Class II  ·  reported 2015-11-11  ·  Terminated

Why it was recalled
The Inserter Inner Shaft would not fit through the Inserter. The inner diameter of the cannulated shaft of the inserter was undersized.
Product
Stryker UniVise Spinous Process Fixation Plate System Inserter, Stryker Spine, Non-Sterile. Reference Number 48590100. Manual Surgical Instrument.
Where it went
Nationwide Distribution.
How much
41 units

Specialty Audible Torque Wrench Mfg by: Stryker Spine, Catalog number IS3002XLP, Non-sterile. Used to final

Z-0803-2015  ·  Class II  ·  reported 2014-12-24  ·  Terminated

Why it was recalled
Stryker has received a complaint from customers relating to a fracture during the use of the Specialty Audible Torque Wrench in surgery.
Product
Specialty Audible Torque Wrench Mfg by: Stryker Spine, Catalog number IS3002XLP, Non-sterile. Used to final tighten to 12 Nm the blocker of the XIA 3, Hooks, Monoaxial Screws, Rod to Rod Connectors and Off-set connectors.
Where it went
Distributed to GA, IN, MI, NJ, OH, PA, TX & VA.
How much
15 units

Stryker Aria Implant Inserter, REF 48758500, This instrument is part of the Aria product brand. It is use

Z-2626-2014  ·  Class II  ·  reported 2014-09-17  ·  Terminated

Why it was recalled
Stryker has initiated the recall of its Aria Standard and Specialty Implant Inserters due to complaints of product fracturing.
Product
Stryker Aria Implant Inserter, REF 48758500, This instrument is part of the Aria product brand. It is used for Implant insertion into the disc space. The Aria Inserter PEEK Spacer System offers implants that are interbody fusion devices intended for use as an aid in spinal fixation. These hollow implants are offered in a variety of lengths, heights, and lordotic angles designed to adapt to different patient anatomies. They have serrations on the superior and inferior surfaces designed for multidirectional fixation, ergonomically shaped anterior edges, and flat posterior edges. This device does not sustain or support life.
Where it went
Worldwide Distribution - USA including OR, AZ, TN, MA, MD, GA, FL, NY, IA, IN, TX, MI, PA, UT, CA, MN, MO, OH, KS, OK, MT, FL, AL, CO, TN, and WI and Internationally to Australia.
How much
64 units

Specialty Aria Inserter, REF IS3021ARI, Non-Sterile, This instrument is part of the Aria product brand. I

Z-2627-2014  ·  Class II  ·  reported 2014-09-17  ·  Terminated

Why it was recalled
Stryker has initiated the recall of its Aria Standard and Specialty Implant Inserters due to complaints of product fracturing.
Product
Specialty Aria Inserter, REF IS3021ARI, Non-Sterile, This instrument is part of the Aria product brand. It is used for Implant insertion into the disc space. The Aria Inserter PEEK Spacer System offers implants that are interbody fusion devices intended for use as an aid in spinal fixation. These hollow implants are offered in a variety of lengths, heights, and lordotic angles designed to adapt to different patient anatomies. They have serrations on the superior and inferior surfaces designed for multidirectional fixation, ergonomically shaped anterior edges, and flat posterior edges. This device does not sustain or support life.
Where it went
Worldwide Distribution - USA including OR, AZ, TN, MA, MD, GA, FL, NY, IA, IN, TX, MI, PA, UT, CA, MN, MO, OH, KS, OK, MT, FL, AL, CO, TN, and WI and Internationally to Australia.
How much
57 units

Specialty Straight Static Inserter Assembly Product Usage: This instrument is a part of the Navigator p

Z-1086-2014  ·  Class II  ·  reported 2014-03-05  ·  Terminated

Why it was recalled
Stryker has received complaints from customers relating to fracture along the weld that binds the handle to the shaft of the static inserter.
Product
Specialty Straight Static Inserter Assembly Product Usage: This instrument is a part of the Navigator product brand . It is used for implant insertion in to the disc space. The AVS navigator Peek Spacer System offers implants that are interbody fusion devices intended for use as an aid in spinal fixation. These hollow implants are offered in a variety of lengths, heights and lordotic angles designed to adapt to different patient anatomies. They have serrations on the superior and inferior surfaces designed for multidirectional fixation and ergonomically shaped anterior edges and flat posterior edges.
Where it went
US Nationwide Distribution
How much
21 units

Stryker 1) XIA Torque Wrench 2) XIA 3 Torque Wrench 3) XIA Elegance Short Torque Wrench Manufactured by

Z-0375-2014  ·  Class II  ·  reported 2013-11-27  ·  Terminated

Why it was recalled
Multiple complaints have reported that the hex tip of the torque wrench fracturing during final tightening.
Product
Stryker 1) XIA Torque Wrench 2) XIA 3 Torque Wrench 3) XIA Elegance Short Torque Wrench Manufactured by: Stryker Spine SAS Z.I. Marticot -33610 -CESTAS -FRANCE +33 (0) 5.57.97.06.30 http://www.stryker.com Distributed in the USA by: Stryker Spine 2 Pearl Court Allendale, NJ 07401-1677 USA +1-201-760-8000 XIA 3 Spinal System- Intended for use in the noncervical spine. When used as an anterior/anterolateral and posterior, non cervical pedicle and on-pedicled fixation system, the XIA 3 Spinal system is intended to provide additional support during fusion autograft or allograft in skeletally mature patients in the treatment of the following acute and chronic instabilities or deformities; degenerate disc disease (DDD) ; spondylolisthesis; trauma spinal stenosis; curvatures; tumor; pseudoarthrosis and failed previous fusion. XIA - Intended for anterior/anterolateral and posterior, non cervical pedicle and non-pedicle fixation of the spine to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion for the following indications: degenerative disc disease; spondylolisthesis, trauma, spinal stenosis, curvatures, tumor,...
Where it went
Worldwide Distribution - USA (nationwide) and Internationally to Argentina, Australia, Canada, Chile, France, Canada, Germany, Hong Kong, Ireland, Italy, Japan, Korea, Netherlands, New Zealand, Spain, Iraq, Egypt, and Turkey.
How much
220 US units; 431 OUS units
Class I

Stryker Oasys Midline Occiput Plate-1) Small 2) Medium 3) Large 4) Large Long 5) Mini. Allows for position

Z-1976-2013  ·  Class I  ·  reported 2013-08-28  ·  Terminated

Why it was recalled
Stryker has received reports from customers indicating post-operative fracture of the pin that connects the tulip head to the plate body.
Product
Stryker Oasys Midline Occiput Plate-1) Small 2) Medium 3) Large 4) Large Long 5) Mini. Allows for positioning directly on the midline keel of the occiput.
Where it went
Nationwide Distribution.
How much
1536 units (US) 880 (Foreign)

Stryker XIA 3 Iliac Screwdriver, 3-PCS. The XIA 3 Iliac Screwdriver is used to drive XIA 3 closed head Mo

Z-0893-2013  ·  Class II  ·  reported 2013-03-06  ·  Terminated

Why it was recalled
Surgeons have experienced unthreading of the screwdrivers outer shafts during use.
Product
Stryker XIA 3 Iliac Screwdriver, 3-PCS. The XIA 3 Iliac Screwdriver is used to drive XIA 3 closed head Monoaxial and closed head Polyaxial screws into the Ilium. Orthopedic manual surgical instrument.
Where it went
Nationwide Distribution including the states of AL, AZ, CA, CO, FL, GA, IA, IL,IN, KS, LA, MA, MD, MI, MO, MN, NC, NY. OH, OK, OR, PA, TN, TX, UT and WI.
How much
46 units US

SPECIALTY AUDIBLE TORQUE WRENCH, XIA IS2217XLP MFG BY: STRYKER SPINE 2 PEARL COURT ALLENDALE NJ 0740

Z-0382-2013  ·  Class II  ·  reported 2012-11-28  ·  Terminated

Why it was recalled
Multiple complaints have reported the hex tip of the torque wrench fracturing during final tightening.
Product
SPECIALTY AUDIBLE TORQUE WRENCH, XIA IS2217XLP MFG BY: STRYKER SPINE 2 PEARL COURT ALLENDALE NJ 07401-1677 The Xia 3/Mantis Redux/Specialty Audible Torque Wrenches are used to final tighten the blocker of the Xia 3/ Mantis Redux Polyaxial Screws, Hooks, Monoaxial Screws, Rod to Rod Connectors and Off-set connectors. The Torque Wrench is a t-handle deflection beam torque wrench with a 5 mm hex tip. The instrument allows the controlled application of 12 Nm torque via a visual indicator on the bell shaped cylinder just below the handle. The Specialty Audible Torque Wrench has an extra feature where it creates and audible "click" when the desired final tightening torque is reached.
Where it went
Worldwide Distribution - USA (nationwide) and internationally to Australia, Canada, France, Germany, Netherlands, Japan, Spain, Denmark, Switzerland, and the United Kingdom.
How much
28 units

Stryker MANTIS TORQUE WRENCH REF 48287028 Manufactured by: Stryker pine SAS Z.I MARTICOT-33610-CESTAS-

Z-0381-2013  ·  Class II  ·  reported 2012-11-28  ·  Terminated

Why it was recalled
Multiple complaints have reported the hex tip of the torque wrench fracturing during final tightening.
Product
Stryker MANTIS TORQUE WRENCH REF 48287028 Manufactured by: Stryker pine SAS Z.I MARTICOT-33610-CESTAS-FRANCE +33 (0)5.57.97.06.30 http://www.stryker.com Distributed in the USA by: Stryker Spine, 2 Pearl Ct. Allendale NJ 07401-1677 USA +1-201-780-8000 The Xia 3/Mantis Redux/Specialty Audible Torque Wrenches are used to final tighten the blocker of the Xia 3/ Mantis Redux Polyaxial Screws, Hooks, Monoaxial Screws, Rod to Rod Connectors and Off-set connectors. The Torque Wrench is a t-handle deflection beam torque wrench with a 5 mm hex tip. The instrument allows the controlled application of 12 Nm torque via a visual indicator on the bell shaped cylinder just below the handle. The Specialty Audible Torque Wrench has an extra feature where it creates and audible "click" when the desired final tightening torque is reached.
Where it went
Worldwide Distribution - USA (nationwide) and internationally to Australia, Canada, France, Germany, Netherlands, Japan, Spain, Denmark, Switzerland, and the United Kingdom.
How much
19 units

Stryker XIA 3 TORQUE WRENCH REF 48237028 Manufactured by: Stryker pine SAS Z.I MARTICOT-33610-CESTAS-F

Z-0380-2013  ·  Class II  ·  reported 2012-11-28  ·  Terminated

Why it was recalled
Multiple complaints have reported the hex tip of the torque wrench fracturing during final tightening.
Product
Stryker XIA 3 TORQUE WRENCH REF 48237028 Manufactured by: Stryker pine SAS Z.I MARTICOT-33610-CESTAS-FRANCE +33 (0)5.57.97.06.30 http://www.stryker.com Distributed in the USA by: Stryker Spine, 2 Pearl Ct. Allendale NJ 07401-1677 USA +1-201-780-8000 The Xia 3/Mantis Redux/Specialty Audible Torque Wrenches are used to final tighten the blocker of the Xia 3/ Mantis Redux Polyaxial Screws, Hooks, Monoaxial Screws, Rod to Rod Connectors and Off-set connectors. The Torque Wrench is a t-handle deflection beam torque wrench with a 5 mm hex tip. The instrument allows the controlled application of 12 Nm torque via a visual indicator on the bell shaped cylinder just below the handle. The Specialty Audible Torque Wrench has an extra feature where it creates and audible "click" when the desired final tightening torque is reached.
Where it went
Worldwide Distribution - USA (nationwide) and internationally to Australia, Canada, France, Germany, Netherlands, Japan, Spain, Denmark, Switzerland, and the United Kingdom.
How much
54

Aviator Drill Bit (10 mm, 12 mm, 14 mm, 16 mm, Manufactured by: Stryker Spine SAS Z.1 Marticot -33610- Cesta

Z-0163-2013  ·  Class II  ·  reported 2012-11-07  ·  Terminated

Why it was recalled
Aviator drill bits are 10 mm longer than the Reflex Hybrid drill bits and are not interchangeable which could lead to over-drilling by as much as 10 mm.
Product
Aviator Drill Bit (10 mm, 12 mm, 14 mm, 16 mm, Manufactured by: Stryker Spine SAS Z.1 Marticot -33610- Cestas France +33 (0) 5.57.97.06.30 http://www.stryker.com Distributed in the USA by Stryker Spine 2 Pearl Ct., Allendale NJ 07401-1677 USA +1-201-750-8000 Drill bits are reusable instruments, sterilized prior to use, used in conjunction with an Aviator Drill Guide to provide a pathway for screw insertion. The aviator Anterior Cervical Plating (ACP) System is intended for use as an aid in cervical spinal fusion and is intended for unilateral fixation.
Where it went
Nationwide distribution.
How much
6154 units

The Instructions For Use for the following: Stryker TRIO+ 4.5 x 45 mm STANDARD POST SCREW Stryker TRIO+ 4.5

Z-2235-2012  ·  Class II  ·  reported 2012-08-29  ·  Terminated

Why it was recalled
The front cover of Instructions For Use (IFU) #NOLI135B03Rev02 was improperly marked with a "Sterile" logo in the lower right corner. The products these IFUs were packed with are sold as "Non-Sterile" and require sterilization before use in surgery.
Product
The Instructions For Use for the following: Stryker TRIO+ 4.5 x 45 mm STANDARD POST SCREW Stryker TRIO+ 4.5 x 50 mm STANDARD POST SCREW Stryker TRIO+ 4.5 x 55 mm STANDARD POST SCREW Stryker TRIO+ 5.5 x 45 mm STANDARD POST SCREW Stryker TRIO+ 5.5 x 60 mm STANDARD POST SCREW Manufactured by: STRYKER Spine SAS, Z.I. Marticot - 33610 - Cestas-France +33 (0)5.57.97.06.30 http://www.stryker.com Distributed in the USA by: Stryker Spine, 2 Pearl Ct., Allendale, NJ 07401-1677 USA +1-201-750-8000. The system is intended for posterior, non-cervical pedicle and non-pedicle pedicle fixation of the spine. It is indicated for: degenerative disc disease (DDD), spondylolisthesis, fracture, dislocation, spinal stenosis, scoliosis kyphosis, lordosis, spinal tumor, pseudarthrosis, and failed previous fusion.
Where it went
Distributed in Atlanta, Georgia.
How much
40 units.

LUXOR RETRACTOR BASE; Catalog Number: 48250030; Manufactured by: STRYKER Spine SAS; Z.I MARTICOT-33610-CESTAS-

Z-2155-2012  ·  Class II  ·  reported 2012-08-15  ·  Terminated

Why it was recalled
Stryker received three reports involving blades of the Luxor Retractor becoming disassembled from the base causing the retractor to collapse insitu.
Product
LUXOR RETRACTOR BASE; Catalog Number: 48250030; Manufactured by: STRYKER Spine SAS; Z.I MARTICOT-33610-CESTAS-FRANCE; Distributed in the USA by: Stryker Spine. 2 Pearl Ct. Allendale, NJ 07401. Provides access to the thoracic and lumbar spine from a posterior approach via a small incision.
Where it went
Worldwide Distribution -- US, Switzerland, Netherlands, Brazil, Hong Kong, Germany, and Taiwan.
How much
30 devices

Stryker Reliance AL Implant Blocker (Various sizes) The SLIDE Implant Blocker is a reusable instrument, st

Z-2176-2012  ·  Class II  ·  reported 2012-08-15  ·  Terminated

Why it was recalled
Stryker Spine has received six reports regarding the breakage of one or more prongs at the extremity of the instrument.
Product
Stryker Reliance AL Implant Blocker (Various sizes) The SLIDE Implant Blocker is a reusable instrument, sterilized prior to use. It allows the surgeon to distract the disc space in the lumbar region of the spine using an anterior or anterolateral approach while also allowing them to insert the implant into final position.
Where it went
Nationwide Distribution including NC, AR, TN, MA, IL, GA, NY, LA, MN, MO, OH, KS, NM, TX, KS,and IN.
How much
109 Lots

Stryker AVS Navigator Trials 6mm-15mm Manufactured by: Stryker Spine SAS Z.1 Marticot-33610-Cestas -France

Z-2033-2012  ·  Class II  ·  reported 2012-07-25  ·  Terminated

Why it was recalled
Beginning in February 2011, Stryker Spine began receiving reports related to the shaft of the AVS Navigator trial failing at the junction of the trial head.
Product
Stryker AVS Navigator Trials 6mm-15mm Manufactured by: Stryker Spine SAS Z.1 Marticot-33610-Cestas -France +33 (0) 5.57.97.06.30 http;//www.stryker.com Distributed in the USA by Stryker Spine 2 Pearl Court. Allendale NJ 07401-1677 USA +1-201-780-8000 Cestas France 33610. Intended Use: The trials are designed and intended to measure the height of the disc space intra-operatively.
Where it went
Nationwide.
How much
273 units
Read these counts with three things in mind

A firm with more recalls usually makes more products. A contract packer runs hundreds of lines. It will out-recall a company with one product whatever either does. This is not a safety ranking.

Most recalls here are voluntary. A firm found its own fault and pulled the stock. That is the system working. It lands in this database looking the same as a recall the FDA had to push for.

Terminated means the FDA considers the recall finished. It is not a verdict. It does not mean anyone was harmed.

Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.

← All device recalls