Skeletal Kinetics, Llc: 49 FDA recall reports
Every medical device recall the FDA has published for Skeletal Kinetics, Llc. The fault, and how the agency graded it.
49Reports
0Class I
0Ongoing
2021-01-27Most recent
What the classes mean
The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.
Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.
By class
By year reported
Checking whether it affects you
The lot or code number is what matters. Not the product name. A recall covers
specific lots. Yours may not be one of them.
Check the code on your package against the report. The FDA keeps the current
list at fda.gov/safety/recalls.
The reports
The text below is quoted from the FDA record word for word.
OsteoVation, Inject Bone Filler Kits, 5cc (part number 390-2005, manufacturing lot number 1129957) - Product U
Z-0866-2021 · Class II · reported 2021-01-27 · Terminated
- Why it was recalled
- A discrepancy was found between the Mixing Instruction Card and the Instructions for Use for the 5cc and 10 cc bone void filler kits. No customer complaints or reports of injury have been received.
- Product
- OsteoVation, Inject Bone Filler Kits, 5cc (part number 390-2005, manufacturing lot number 1129957) - Product Usage: The product provides a bone void filler that is replaced by bone during the healing process.
- Where it went
- U.S. Nationwide distribution including in the states of TX, NY, FL and AZ.
- How much
- 2 units
OsteoVation, Inject, Bone Filler Kits 10cc (part number 390-2010, manufacturing lot 1130303) - Product Usage:
Z-0867-2021 · Class II · reported 2021-01-27 · Terminated
- Why it was recalled
- A discrepancy was found between the Mixing Instruction Card and the Instructions for Use for the 5cc and 10 cc bone void filler kits. No customer complaints or reports of injury have been received.
- Product
- OsteoVation, Inject, Bone Filler Kits 10cc (part number 390-2010, manufacturing lot 1130303) - Product Usage: The product provides a bone void filler that is replaced by bone during the healing process.
- Where it went
- U.S. Nationwide distribution including in the states of TX, NY, FL and AZ.
- How much
- 49 units
OSTEOVATION EX, 5CC, IMPACT FORMULA. Product Number: 390-0205, UDI: 813845020047 - Product Usage: Callos is i
Z-2613-2020 · Class II · reported 2020-07-22 · Terminated
- Why it was recalled
- Bone void filler kits may not maintain its setting characteristics for the labeled shelf-life duration. Use may increase the required surgical and anesthesia time for the patient or possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.
- Product
- OSTEOVATION EX, 5CC, IMPACT FORMULA. Product Number: 390-0205, UDI: 813845020047 - Product Usage: Callos is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of AL, CA, CO, FL, IL, MO, MS, OH, OK, OR, TN, TX, WI and the countries of Australia, Canada, South Africa, Switzerland.
- How much
- 88
EXACTECH OSSILIX MP FIL, 10CC. Product Number: 662-02-10, UDI: 813845021051 - Product Usage: Callos is indica
Z-2636-2020 · Class II · reported 2020-07-22 · Terminated
- Why it was recalled
- Bone void filler kits may not maintain its setting characteristics for the labeled shelf-life duration. Use may increase the required surgical and anesthesia time for the patient or possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.
- Product
- EXACTECH OSSILIX MP FIL, 10CC. Product Number: 662-02-10, UDI: 813845021051 - Product Usage: Callos is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of AL, CA, CO, FL, IL, MO, MS, OH, OK, OR, TN, TX, WI and the countries of Australia, Canada, South Africa, Switzerland.
- How much
- 60
CRANIOSCULPT FLOW, OUS, 10CC. Product Number: C-FLOW10CC-OUS, UDI: 813845020382 - Product Usage: Callos is in
Z-2603-2020 · Class II · reported 2020-07-22 · Terminated
- Why it was recalled
- Bone void filler kits may not maintain its setting characteristics for the labeled shelf-life duration. Use may increase the required surgical and anesthesia time for the patient or possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.
- Product
- CRANIOSCULPT FLOW, OUS, 10CC. Product Number: C-FLOW10CC-OUS, UDI: 813845020382 - Product Usage: Callos is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of AL, CA, CO, FL, IL, MO, MS, OH, OK, OR, TN, TX, WI and the countries of Australia, Canada, South Africa, Switzerland.
- How much
- 31
CRANIOSCULPT FLOW, CAN, 5CC. Product Number: C-FLOW5CC-CAN, UDI: 813845020344 - Product Usage: Callos is indi
Z-2602-2020 · Class II · reported 2020-07-22 · Terminated
- Why it was recalled
- Bone void filler kits may not maintain its setting characteristics for the labeled shelf-life duration. Use may increase the required surgical and anesthesia time for the patient or possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.
- Product
- CRANIOSCULPT FLOW, CAN, 5CC. Product Number: C-FLOW5CC-CAN, UDI: 813845020344 - Product Usage: Callos is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.-
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of AL, CA, CO, FL, IL, MO, MS, OH, OK, OR, TN, TX, WI and the countries of Australia, Canada, South Africa, Switzerland.
- How much
- 35
OSTEOVATIONEX 5CC, INJECT FORMULA. Product Number: 390-0005, UDI: 813845020023 - Product Usage: Callos is ind
Z-2622-2020 · Class II · reported 2020-07-22 · Terminated
- Why it was recalled
- Bone void filler kits may not maintain its setting characteristics for the labeled shelf-life duration. Use may increase the required surgical and anesthesia time for the patient or possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.
- Product
- OSTEOVATIONEX 5CC, INJECT FORMULA. Product Number: 390-0005, UDI: 813845020023 - Product Usage: Callos is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of AL, CA, CO, FL, IL, MO, MS, OH, OK, OR, TN, TX, WI and the countries of Australia, Canada, South Africa, Switzerland.
- How much
- 375
OSTEOVATION, 5CC, IMPACT FORMULA. Product Number: 390-2105, UDI: 813845020085 - Product Usage: Callos is indi
Z-2619-2020 · Class II · reported 2020-07-22 · Terminated
- Why it was recalled
- Bone void filler kits may not maintain its setting characteristics for the labeled shelf-life duration. Use may increase the required surgical and anesthesia time for the patient or possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.
- Product
- OSTEOVATION, 5CC, IMPACT FORMULA. Product Number: 390-2105, UDI: 813845020085 - Product Usage: Callos is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of AL, CA, CO, FL, IL, MO, MS, OH, OK, OR, TN, TX, WI and the countries of Australia, Canada, South Africa, Switzerland.
- How much
- 1393
OSTEOVATION RMX 10CC. Product Number: 390-6002, UDI: 813845021099 - Product Usage: Callos is indicated to fil
Z-2615-2020 · Class II · reported 2020-07-22 · Terminated
- Why it was recalled
- Bone void filler kits may not maintain its setting characteristics for the labeled shelf-life duration. Use may increase the required surgical and anesthesia time for the patient or possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.
- Product
- OSTEOVATION RMX 10CC. Product Number: 390-6002, UDI: 813845021099 - Product Usage: Callos is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of AL, CA, CO, FL, IL, MO, MS, OH, OK, OR, TN, TX, WI and the countries of Australia, Canada, South Africa, Switzerland.
- How much
- 379
SKAFFOLD CMF IMPRESS (MP), 10CC. Product Number: CMMP-IM10, UDI: 813845020696 - Product Usage: Callos is indi
Z-2625-2020 · Class II · reported 2020-07-22 · Terminated
- Why it was recalled
- Bone void filler kits may not maintain its setting characteristics for the labeled shelf-life duration. Use may increase the required surgical and anesthesia time for the patient or possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.
- Product
- SKAFFOLD CMF IMPRESS (MP), 10CC. Product Number: CMMP-IM10, UDI: 813845020696 - Product Usage: Callos is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of AL, CA, CO, FL, IL, MO, MS, OH, OK, OR, TN, TX, WI and the countries of Australia, Canada, South Africa, Switzerland.
- How much
- 12
MEDTRONIC BONE VOID FILLER, LARGE. Product Number: C07B, UDI: 813845020511 - Product Usage: Callos is indicat
Z-2610-2020 · Class II · reported 2020-07-22 · Terminated
- Why it was recalled
- Bone void filler kits may not maintain its setting characteristics for the labeled shelf-life duration. Use may increase the required surgical and anesthesia time for the patient or possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.
- Product
- MEDTRONIC BONE VOID FILLER, LARGE. Product Number: C07B, UDI: 813845020511 - Product Usage: Callos is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of AL, CA, CO, FL, IL, MO, MS, OH, OK, OR, TN, TX, WI and the countries of Australia, Canada, South Africa, Switzerland.
- How much
- 1053
EXACTECH OSSILIX FIL (MX), 10CC. Product Number: 660-02-10, UDI: 813845021013 - - Product Usage: Callos is in
Z-2606-2020 · Class II · reported 2020-07-22 · Terminated
- Why it was recalled
- Bone void filler kits may not maintain its setting characteristics for the labeled shelf-life duration. Use may increase the required surgical and anesthesia time for the patient or possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.
- Product
- EXACTECH OSSILIX FIL (MX), 10CC. Product Number: 660-02-10, UDI: 813845021013 - - Product Usage: Callos is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of AL, CA, CO, FL, IL, MO, MS, OH, OK, OR, TN, TX, WI and the countries of Australia, Canada, South Africa, Switzerland.
- How much
- 121
ACUMED CALLOS INJECT 5CC. Product Number: 65-0005-S, UDI: 813845020238 - Product Usage: Callos is indicated t
Z-2594-2020 · Class II · reported 2020-07-22 · Terminated
- Why it was recalled
- Bone void filler kits may not maintain its setting characteristics for the labeled shelf-life duration. Use may increase the required surgical and anesthesia time for the patient or possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.
- Product
- ACUMED CALLOS INJECT 5CC. Product Number: 65-0005-S, UDI: 813845020238 - Product Usage: Callos is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of AL, CA, CO, FL, IL, MO, MS, OH, OK, OR, TN, TX, WI and the countries of Australia, Canada, South Africa, Switzerland.
- How much
- 442
EXACTECH OSSILIX FIL (MX), 5CC. Product Number: 660-02-05, UDI: 813845021006 - Product Usage: Callos is indic
Z-2607-2020 · Class II · reported 2020-07-22 · Terminated
- Why it was recalled
- Bone void filler kits may not maintain its setting characteristics for the labeled shelf-life duration. Use may increase the required surgical and anesthesia time for the patient or possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.
- Product
- EXACTECH OSSILIX FIL (MX), 5CC. Product Number: 660-02-05, UDI: 813845021006 - Product Usage: Callos is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of AL, CA, CO, FL, IL, MO, MS, OH, OK, OR, TN, TX, WI and the countries of Australia, Canada, South Africa, Switzerland.
- How much
- 217
CRANIOSCULPT, CMF IMPACT, 3CC Product Number: C-CMF3CC, UDI: 813845020931 - Product Usage: Callos is indicate
Z-2633-2020 · Class II · reported 2020-07-22 · Terminated
- Why it was recalled
- Bone void filler kits may not maintain its setting characteristics for the labeled shelf-life duration. Use may increase the required surgical and anesthesia time for the patient or possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.
- Product
- CRANIOSCULPT, CMF IMPACT, 3CC Product Number: C-CMF3CC, UDI: 813845020931 - Product Usage: Callos is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of AL, CA, CO, FL, IL, MO, MS, OH, OK, OR, TN, TX, WI and the countries of Australia, Canada, South Africa, Switzerland.
- How much
- 27
SKAFFOLD FLOW, 10CC. Product Number: SKMX-FL10, UDI: 813845020900 - Product Usage: Callos is indicated to fil
Z-2629-2020 · Class II · reported 2020-07-22 · Terminated
- Why it was recalled
- Bone void filler kits may not maintain its setting characteristics for the labeled shelf-life duration. Use may increase the required surgical and anesthesia time for the patient or possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.
- Product
- SKAFFOLD FLOW, 10CC. Product Number: SKMX-FL10, UDI: 813845020900 - Product Usage: Callos is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of AL, CA, CO, FL, IL, MO, MS, OH, OK, OR, TN, TX, WI and the countries of Australia, Canada, South Africa, Switzerland.
- How much
- 54
CRANIOSCULPT, CMF IMPACT, 10CC, Product Number: C-CMF10CC, UDI: 813845020955 - Product Usage: Callos is indic
Z-2635-2020 · Class II · reported 2020-07-22 · Terminated
- Why it was recalled
- Bone void filler kits may not maintain its setting characteristics for the labeled shelf-life duration. Use may increase the required surgical and anesthesia time for the patient or possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.
- Product
- CRANIOSCULPT, CMF IMPACT, 10CC, Product Number: C-CMF10CC, UDI: 813845020955 - Product Usage: Callos is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of AL, CA, CO, FL, IL, MO, MS, OH, OK, OR, TN, TX, WI and the countries of Australia, Canada, South Africa, Switzerland.
- How much
- 51
CRANIOSCULPT FLOW, 3CC. Product Number: C-FLOW3CC, UDI: 813845020320 - - Product Usage: Callos is indicated t
Z-2600-2020 · Class II · reported 2020-07-22 · Terminated
- Why it was recalled
- Bone void filler kits may not maintain its setting characteristics for the labeled shelf-life duration. Use may increase the required surgical and anesthesia time for the patient or possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.
- Product
- CRANIOSCULPT FLOW, 3CC. Product Number: C-FLOW3CC, UDI: 813845020320 - - Product Usage: Callos is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of AL, CA, CO, FL, IL, MO, MS, OH, OK, OR, TN, TX, WI and the countries of Australia, Canada, South Africa, Switzerland.
- How much
- 70
SKAFFOLD FLOW (MX) 5CC. Product Number: SKMX-FL05, UDI: 813845020894 - Product Usage: Callos is indicated to
Z-2628-2020 · Class II · reported 2020-07-22 · Terminated
- Why it was recalled
- Bone void filler kits may not maintain its setting characteristics for the labeled shelf-life duration. Use may increase the required surgical and anesthesia time for the patient or possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.
- Product
- SKAFFOLD FLOW (MX) 5CC. Product Number: SKMX-FL05, UDI: 813845020894 - Product Usage: Callos is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of AL, CA, CO, FL, IL, MO, MS, OH, OK, OR, TN, TX, WI and the countries of Australia, Canada, South Africa, Switzerland.
- How much
- 41
CRANIOSCULPT FLOW, CAN, 10CC, Product Number: C-FLOW10CC-CAN, UDI: 813845020375 - Product Usage: Callos is in
Z-2632-2020 · Class II · reported 2020-07-22 · Terminated
- Why it was recalled
- Bone void filler kits may not maintain its setting characteristics for the labeled shelf-life duration. Use may increase the required surgical and anesthesia time for the patient or possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.
- Product
- CRANIOSCULPT FLOW, CAN, 10CC, Product Number: C-FLOW10CC-CAN, UDI: 813845020375 - Product Usage: Callos is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of AL, CA, CO, FL, IL, MO, MS, OH, OK, OR, TN, TX, WI and the countries of Australia, Canada, South Africa, Switzerland.
- How much
- 10
OSTEOVATION IMPACT 10CC and OSTEOVATION, 10CC, IMPACT FORMULA, Product Number: 390-2110, UDI: 813845020092 -
Z-2614-2020 · Class II · reported 2020-07-22 · Terminated
- Why it was recalled
- Bone void filler kits may not maintain its setting characteristics for the labeled shelf-life duration. Use may increase the required surgical and anesthesia time for the patient or possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.
- Product
- OSTEOVATION IMPACT 10CC and OSTEOVATION, 10CC, IMPACT FORMULA, Product Number: 390-2110, UDI: 813845020092 - Product Usage: Callos is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of AL, CA, CO, FL, IL, MO, MS, OH, OK, OR, TN, TX, WI and the countries of Australia, Canada, South Africa, Switzerland.
- How much
- 661
OSTEOVATION RMX 5CC. Product Number: 390-6001, UDI: 813845021082 - Product Usage: Callos is indicated to fill
Z-2616-2020 · Class II · reported 2020-07-22 · Terminated
- Why it was recalled
- Bone void filler kits may not maintain its setting characteristics for the labeled shelf-life duration. Use may increase the required surgical and anesthesia time for the patient or possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.
- Product
- OSTEOVATION RMX 5CC. Product Number: 390-6001, UDI: 813845021082 - Product Usage: Callos is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of AL, CA, CO, FL, IL, MO, MS, OH, OK, OR, TN, TX, WI and the countries of Australia, Canada, South Africa, Switzerland.
- How much
- 111
ACUMED CALLOS INJECT 10CC. Product Number: 65-0010-S, UDI: 813845020245 - Product Usage: Callos is indicated
Z-2592-2020 · Class II · reported 2020-07-22 · Terminated
- Why it was recalled
- Bone void filler kits may not maintain its setting characteristics for the labeled shelf-life duration. Use may increase the required surgical and anesthesia time for the patient or possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.
- Product
- ACUMED CALLOS INJECT 10CC. Product Number: 65-0010-S, UDI: 813845020245 - Product Usage: Callos is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of AL, CA, CO, FL, IL, MO, MS, OH, OK, OR, TN, TX, WI and the countries of Australia, Canada, South Africa, Switzerland.
- How much
- 379
ACUMED CALLOS INJECT 3CC. Product Number: 65-0003-S, UDI: 813845020221 - Product Usage: Callos is indicated
Z-2593-2020 · Class II · reported 2020-07-22 · Terminated
- Why it was recalled
- Bone void filler kits may not maintain its setting characteristics for the labeled shelf-life duration. Use may increase the required surgical and anesthesia time for the patient or possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.
- Product
- ACUMED CALLOS INJECT 3CC. Product Number: 65-0003-S, UDI: 813845020221 - Product Usage: Callos is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of AL, CA, CO, FL, IL, MO, MS, OH, OK, OR, TN, TX, WI and the countries of Australia, Canada, South Africa, Switzerland.
- How much
- 184
SKAFFOLD CMF IMPRESS (MP), 5CC. Product Number: CMMP-IM05, UDI: 813845020689 - Product Usage: Callos is indic
Z-2626-2020 · Class II · reported 2020-07-22 · Terminated
- Why it was recalled
- Bone void filler kits may not maintain its setting characteristics for the labeled shelf-life duration. Use may increase the required surgical and anesthesia time for the patient or possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.
- Product
- SKAFFOLD CMF IMPRESS (MP), 5CC. Product Number: CMMP-IM05, UDI: 813845020689 - Product Usage: Callos is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of AL, CA, CO, FL, IL, MO, MS, OH, OK, OR, TN, TX, WI and the countries of Australia, Canada, South Africa, Switzerland.
- How much
- 4
SKAFFOLD FLOW, 3CC and SKAFFOLD FLOW (MX) 3CC. Product Number: SKMX-FL03, UDI: 813845020887 - Product Usage:
Z-2627-2020 · Class II · reported 2020-07-22 · Terminated
- Why it was recalled
- Bone void filler kits may not maintain its setting characteristics for the labeled shelf-life duration. Use may increase the required surgical and anesthesia time for the patient or possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.
- Product
- SKAFFOLD FLOW, 3CC and SKAFFOLD FLOW (MX) 3CC. Product Number: SKMX-FL03, UDI: 813845020887 - Product Usage: Callos is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of AL, CA, CO, FL, IL, MO, MS, OH, OK, OR, TN, TX, WI and the countries of Australia, Canada, South Africa, Switzerland.
- How much
- 29
OSTEOVATION, 10CC, INJECT FORMULA. Product Number: 390-2010, UDI: 813845020061 - Product Usage: Callos is ind
Z-2617-2020 · Class II · reported 2020-07-22 · Terminated
- Why it was recalled
- Bone void filler kits may not maintain its setting characteristics for the labeled shelf-life duration. Use may increase the required surgical and anesthesia time for the patient or possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.
- Product
- OSTEOVATION, 10CC, INJECT FORMULA. Product Number: 390-2010, UDI: 813845020061 - Product Usage: Callos is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of AL, CA, CO, FL, IL, MO, MS, OH, OK, OR, TN, TX, WI and the countries of Australia, Canada, South Africa, Switzerland.
- How much
- 116
CRANIOSCULPT C, 5CC. Product Number: C-C5CC, UDI: 813845021112 - - Product Usage: Callos is indicated to fill
Z-2599-2020 · Class II · reported 2020-07-22 · Terminated
- Why it was recalled
- Bone void filler kits may not maintain its setting characteristics for the labeled shelf-life duration. Use may increase the required surgical and anesthesia time for the patient or possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.
- Product
- CRANIOSCULPT C, 5CC. Product Number: C-C5CC, UDI: 813845021112 - - Product Usage: Callos is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of AL, CA, CO, FL, IL, MO, MS, OH, OK, OR, TN, TX, WI and the countries of Australia, Canada, South Africa, Switzerland.
- How much
- 263
OSTEOVATIONEX 2CC INJECT FORMULA. Product Number: 390-0002, UDI: 813845020009 - Product Usage: Callos is indi
Z-2621-2020 · Class II · reported 2020-07-22 · Terminated
- Why it was recalled
- Bone void filler kits may not maintain its setting characteristics for the labeled shelf-life duration. Use may increase the required surgical and anesthesia time for the patient or possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.
- Product
- OSTEOVATIONEX 2CC INJECT FORMULA. Product Number: 390-0002, UDI: 813845020009 - Product Usage: Callos is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of AL, CA, CO, FL, IL, MO, MS, OH, OK, OR, TN, TX, WI and the countries of Australia, Canada, South Africa, Switzerland.
- How much
- 35
CRANIOSCULPT C, 3CC. Product Number: C-C3CC, UDI: 813845021105 - Product Usage: Callos is indicated to fill b
Z-2598-2020 · Class II · reported 2020-07-22 · Terminated
- Why it was recalled
- Bone void filler kits may not maintain its setting characteristics for the labeled shelf-life duration. Use may increase the required surgical and anesthesia time for the patient or possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.
- Product
- CRANIOSCULPT C, 3CC. Product Number: C-C3CC, UDI: 813845021105 - Product Usage: Callos is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of AL, CA, CO, FL, IL, MO, MS, OH, OK, OR, TN, TX, WI and the countries of Australia, Canada, South Africa, Switzerland.
- How much
- 48
Showing the 30 most recent of 49. The rest are searchable at the FDA enforcement report database.
Read these counts with three things in mind
A firm with more recalls usually makes more products. A contract packer runs
hundreds of lines. It will out-recall a company with one product whatever either
does. This is not a safety ranking.
Most recalls here are voluntary. A firm found its own fault and pulled the stock.
That is the system working. It lands in this database looking the same as a recall
the FDA had to push for.
Terminated means the FDA considers the recall finished. It is not a verdict.
It does not mean anyone was harmed.
Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.