Siemens Healthcare Diagnostics, Inc.

Siemens Healthcare Diagnostics, Inc.: 662 FDA recall reports

Every medical device recall the FDA has published for Siemens Healthcare Diagnostics, Inc.. The fault, and how the agency graded it.

662Reports
13Class I
167Ongoing
2026-07-15Most recent
13 Class I on this page

Class I means the FDA judged there was a reasonable probability of serious harm or death. Check the lot numbers against anything you hold.

What the classes mean

The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.

ClassWhat the FDA is saying
Class IThere is a reasonable probability that using this will cause serious harm or death.
Class IIIt may cause temporary or reversible harm. Serious harm is unlikely.
Class IIIIt breaks a rule but is unlikely to hurt anyone. Often a labelling fault.

Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.

By class

ClassReports
Class II611
Class III38
Class I13

By year reported

YearReports
201225
201368
201462
201546
201686
201736
201891
201926
202051
202144
202232
202337
202425
202527
20266
Checking whether it affects you

The lot or code number is what matters. Not the product name. A recall covers specific lots. Yours may not be one of them.

Check the code on your package against the report. The FDA keeps the current list at fda.gov/safety/recalls.

The reports

The text below is quoted from the FDA record word for word.

Atellica CH Enzymatic Hemoglobin A1c (A1c_E). Material Number: 11097536.

Z-2703-2026  ·  Class II  ·  reported 2026-07-15  ·  Ongoing

Why it was recalled
The potential for intermittent well-to-well bias on the A1c_E assay on the Atellica CH and Atellica CI analyzer which could cause falsely depressed or falsely elevated HbA1c patient results.
Product
Atellica CH Enzymatic Hemoglobin A1c (A1c_E). Material Number: 11097536.
Where it went
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahamas, Bangladesh, Bolivia, Brazil, Bulgaria, Burkina¿Faso, Canada, Chile, China, Colombia, Croatia, Cura¿ao,¿St¿Eus, Czech¿Republic, Democratic Republic Congo, Denmark, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Georgia, Germany, Greece, Hong Kong, India, Indonesia, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Lebanon, Lithuania, Macedonia, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, Pakistan, Paraguay, Poland, Portugal, Qatar, Republic of Korea, Romania, Saudi¿Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Switzerland, Thailand, Turkey, Turkmenistan, U.A.E., Uganda, United¿Kingdom, Uruguay, Vatican City, Vietnam.
How much
24,077 units

Siemens Dimension Tacrolimus Flex Reagent Cartridge (TAC), REF DF207, SMN 10700795; enzyme immunoassay

Z-2398-2026  ·  Class II  ·  reported 2026-06-17  ·  Ongoing

Why it was recalled
A potential positive bias in quality control (QC) and patient sample results in three lots of the Dimension Tacrolimus (TAC) Flex reagent cartridge.
Product
Siemens Dimension Tacrolimus Flex Reagent Cartridge (TAC), REF DF207, SMN 10700795; enzyme immunoassay
Where it went
Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Brazil, Croatia, Germany, Hungary, India, Italy, Japan, Korea, Lithuania, Pakistan, Poland, Saudi Arabia, Spain, South Africa, Sri Lanka, Turkey, United Arab Emirates.
How much
2878 units

Dimension Creatinine Flex reagent cartridge. Material Number: 10872079.

Z-1980-2026  ·  Class II  ·  reported 2026-05-06  ·  Ongoing

Why it was recalled
Siemens Healthcare Diagnostics is recalling certain lots of Dimension Creatinine Flex reagent cartridge (Dimension CRE2) because a potential imprecision in the Dimension CRE2 quality control (QC) and patient sample results using lot numbers GA6307 and BA7005 on the Dimension System. This imprecision may lead to erroneously decreased or increased creatinine patient results
Product
Dimension Creatinine Flex reagent cartridge. Material Number: 10872079.
Where it went
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahamas, Brazil, Cambodia, Canada, Chile, China, Columbia, Costa Rica, Ecuador, Egypt, Germany, Honduras, India, Indonesia, Korea, Kuwait, Lebanon, Malaysia, Mexico, Nicaragua, Pakistan, Paraguay, Philippines, Qatar, Saudi Arabia, Serbia, Slovenia, South Africa, Sri Lanka, Taiwan, Thailand, U.A.E., Ukraine, Uruguay, Vietnam, Yemen.
How much
18,080 units

epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;

Z-2005-2026  ·  Class II  ·  reported 2026-05-06  ·  Ongoing

Why it was recalled
Use of the affected product can result in discrepant low pH and measured Total Carbon Dioxide (mTCO2) results and corresponding calculated values in arterial, venous, and capillary patient samples. An erroneously low pH or mTCO2 reading can lead to inappropriate treatment of nonexistent acidosis (respiratory of metabolic) or missed diagnosis of alkalosis (a true high pH: respiratory or metabolic alkalosis), potentially causing harm through unnecessary interventions with the former or failure to deliver appropriate treatment with the latter.
Product
epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;
Where it went
Worldwide distribution - US Nationwide and the countries of Italy, Spain, Austria, Portugal, Ireland, France, Romania, Canada, India, Philippines, Mexico.
How much
71725 (38025 US, 33700 OUS)

Atellica CH Enzymatic Hemoglobin A1c (A1c_E); Siemens Material Number (SMN): 11097536;

Z-1657-2026  ·  Class II  ·  reported 2026-04-08  ·  Ongoing

Why it was recalled
A potential for depressed results for Atellica CH A1c_E when processing Atellica CH RCRP on the same analyzer. Customers processing A1c_E and RCRP on separate analyzers are not impacted by the recall. The frequency and magnitude of the negative bias and imprecision increase over time and are not consistent with the assay Instructions For Use (IFU). When this issue is present, Quality Control (QC) results demonstrate a negative trend and increased imprecision. The clinical impact of a negative (A1c_E) bias may lead to a delay in diagnosis for diabetes or lead to less-than-optimal treatment for existing diabetes.
Product
Atellica CH Enzymatic Hemoglobin A1c (A1c_E); Siemens Material Number (SMN): 11097536;
Where it went
Domestic: Nationwide Distribution International: Argentina, Australia, Bangladesh, Brazil, Bulgaria , Canada, Chile, China, Colombia, Croatia , Cura¿ao, St Eus, Czech Republic , Denmark , Ecuador, Egypt , Estonia , Finland , France , Georgia , Germany , Greece , Hong Kong, India, Iraq , Italy , Jordan , Kuwait , Latvia , Lithuania , Macedonia , Malaysia, Maldives, Mexico, Netherlands , New Zealand, Norway , Oman , Pakistan , Paraguay, Qatar , Romania , Saudi Arabia , Serbia , Singapore, Slovakia , Slovenia , South Africa, South Korea, Spain , Turkey , Turkmenistan , U.A.E. , Uganda , United Kingdom , Uruguay , Vatikancity;
How much
21,291 units (9750 US, 11541 OUS)

Atellica CH Urine Albumin (UAlb). Material Number: 11537225

Z-1484-2026  ·  Class II  ·  reported 2026-03-11  ·  Ongoing

Why it was recalled
Falsely depressed UAlb patient results may occur. Affected samples with any urine albumin concentration above the UAlb measuring interval >40.0 mg/dL may report values between 6.2 mg/dL and 40.0 mg/dL.
Product
Atellica CH Urine Albumin (UAlb). Material Number: 11537225
Where it went
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Canada, Chile, Curacao,¿St¿Eus, Czech¿Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Ireland, Israel, Italy, Korea, Latvia, Malaysia, Netherlands, New Zealand, Norway, Portugal, Romania, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, United¿Kingdom.
How much
4,885 units

Food Panel 3 Specific Allergen; Siemens Material Number (SMN): 10385891; Catalog Number: FP3L4;

Z-0493-2026  ·  Class II  ·  reported 2025-11-19  ·  Ongoing

Why it was recalled
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
Product
Food Panel 3 Specific Allergen; Siemens Material Number (SMN): 10385891; Catalog Number: FP3L4;
Where it went
Worldwide - US Nationwide distribution in the state of CA and the countries of Argentina, Australia, Austria, Belgium, Bosnia Herzeg. , Brazil, Canary Islands, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Estonia, France, Germany, Greece, Guatemala, Hong Kong, Hungary, Israel, Italy, Latvia, Macedonia, Mexico, Netherlands, Norway, Pakistan, Panama, Poland, Portugal, Romania, Russian Fed., Serbia, Slovakia, Spain, Sweden, Turkey, U.A.E., Ukraine, United Kingdom, Uruguay.
How much
28 units (OUS only)

3gAllergy Specific IgE Universal Kit; Kit Component: 3gAllergy Specific IgE Adjustor Antibody (L2UNS1), 3gAll

Z-0492-2026  ·  Class II  ·  reported 2025-11-19  ·  Ongoing

Why it was recalled
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
Product
3gAllergy Specific IgE Universal Kit; Kit Component: 3gAllergy Specific IgE Adjustor Antibody (L2UNS1), 3gAllergy Specific IgE Control Antibody (L2UNS2); Siemens Material Number (SMN): 10380875; Catalog Number: L2KUN6;
Where it went
Worldwide - US Nationwide distribution in the state of CA and the countries of Argentina, Australia, Austria, Belgium, Bosnia Herzeg. , Brazil, Canary Islands, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Estonia, France, Germany, Greece, Guatemala, Hong Kong, Hungary, Israel, Italy, Latvia, Macedonia, Mexico, Netherlands, Norway, Pakistan, Panama, Poland, Portugal, Romania, Russian Fed., Serbia, Slovakia, Spain, Sweden, Turkey, U.A.E., Ukraine, United Kingdom, Uruguay.
How much
2802 units (OUS only)

IMMULITE 2000 Oak Mix Specific Allergen; Siemens Material Number (SMN): 10386105; Catalog Number: L2KUN6;

Z-0491-2026  ·  Class II  ·  reported 2025-11-19  ·  Ongoing

Why it was recalled
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
Product
IMMULITE 2000 Oak Mix Specific Allergen; Siemens Material Number (SMN): 10386105; Catalog Number: L2KUN6;
Where it went
Worldwide - US Nationwide distribution in the state of CA and the countries of Argentina, Australia, Austria, Belgium, Bosnia Herzeg. , Brazil, Canary Islands, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Estonia, France, Germany, Greece, Guatemala, Hong Kong, Hungary, Israel, Italy, Latvia, Macedonia, Mexico, Netherlands, Norway, Pakistan, Panama, Poland, Portugal, Romania, Russian Fed., Serbia, Slovakia, Spain, Sweden, Turkey, U.A.E., Ukraine, United Kingdom, Uruguay.
How much
34 units (1 US, 33 OUS)

Assay: IMMULITE 2000 Intact PTH; Test Code: iPT; Siemens Material Number (SMN): (1) 10381441, (2) 10381442;

Z-0366-2026  ·  Class II  ·  reported 2025-11-05  ·  Ongoing

Why it was recalled
The potential for falsely depressed Intact PTH patient results at the low end of the assay range, less than or equal to 50 pg/mL (less than or equal to 5.3 pmol/L), when using specific lots identified by the firm on the IMMULITE 2000/IMMULITE 2000 XPi systems. The bias is observed in both serum and plasma samples.
Product
Assay: IMMULITE 2000 Intact PTH; Test Code: iPT; Siemens Material Number (SMN): (1) 10381441, (2) 10381442; Catalog Number: (1) L2KPP2, (2) L2KPP6;
Where it went
Worldwide - US Nationwide distribution in the states of AL, CA, DC, FL, ID, IN, LA, MD, NC, NJ, NY, PA, SC and the countries of Argentina, Australia, Bolivia, Bosnia Herzeg, Brazil, Bulgaria, Canada, Chile, Colombia, Cyprus, Czech Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Honduras, Hungary, India, Iran, Israel, Italy, Kosovo, Kuwait, Latvia, Lebanon, Macedonia, Mexico, Nepal, Netherlands, P.R. China, Pakistan, Palestinian Ter, Paraguay, Peru, Poland, Portugal, Qatar, Romania, Russian Fed., Serbia, Slovakia, Slovenia, South Africa, Spain, Taiwan, Tunisia, Turkey, Turkmenistan, U.A.E., Ukraine, Uruguay, Uzbekistan.
How much
3318 units (169 US, 3149 OUS)

Atellica CH Urinary/Cerebrospinal Fluid Protein (UCFP). Material Number: 11097543.

Z-2542-2025  ·  Class II  ·  reported 2025-09-17  ·  Ongoing

Why it was recalled
Potential for falsely depressed patient, quality control (QC), and/or calibration results or delayed results.
Product
Atellica CH Urinary/Cerebrospinal Fluid Protein (UCFP). Material Number: 11097543.
Where it went
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cura¿ao,¿St¿Eus, Czech¿Republic, Denmark, Egypt, Estonia, Ethiopia, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Macedonia, Malaysia, Maldives, Mexico, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Turkmenistan, U.A.E., United¿Kingdom, Uruguay, Vatikancity, Vietnam, Zaire.
How much
11,811 units

epoc BGEM BUN Test Card [25pk]. Material Number: 10736515.

Z-2272-2025  ·  Class II  ·  reported 2025-08-13  ·  Ongoing

Why it was recalled
Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in some epoc test card lots. All other analytes are performing as intended. The observed average bias for sodium was -4.4 mmol/L. The maximum bias observed was -14 mmol/L, which occurred at a higher sodium concentration around 150 mmol/L. No positive bias has been observed. Quality Control (QC) fluid may or may not detect the issue depending on the bias magnitude and the frequency of QC testing.
Product
epoc BGEM BUN Test Card [25pk]. Material Number: 10736515.
Where it went
Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Bahrain, Bolivia, Brazil, Canada, Canary Islands, Chile, Croatia, Denmark, Ecuador, Estonia, France, Fren. Polynesia, Germany, Greece, Guatemala, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Jordan, Kazakhstan, Lithuania, Malaysia, Martinique, Mexico, Morocco, Netherlands, New Caledonia, Paraguay, Philippines, Poland, Portugal, Qatar, Rep. Korea (S), Romania, Russian Fed., San Marino, Slovakia, Slovenia, South Africa, Spain, Srilanka, Sweden, Switzerland, Taiwan, Thailand, Turkey, Uganda, Ukraine, Unit.Arab Emir., United Kingdom, Uruguay, Japan.
How much
37,554 units

RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.

Z-2230-2025  ·  Class II  ·  reported 2025-08-06  ·  Ongoing

Why it was recalled
Siemens Healthcare Diagnostics is recalling the RAPIDPoint 500 Systems Wash/Waste Cartridge Lot WW/08925 due to a manufacturing error causing positive and negative bias in Ionized Calcium, Potassium, pH, Chloride, pCO2, Glucose and Lactate analytes and their derivatives. Four complaints were filed related to this event that led to Siemens Healthcare Diagnostics identifying an error in the manufacturing process and initiated corrective actions. Use of the affected device may result in errors during QC which could lead to a delay in patient treatment. If bias generated from this issue is not caught during QC erroneous results may cause temporary or medically reversible adverse health consequences to patients.
Product
RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.
Where it went
worldwide distribution - US Nationwide and the countries of Canada, Chile, Japan, Paraguay.
How much
1,477 units

ADVIA¿ 120/2120/2120i 3-in-1 TESTpoint Control (ABN2 Control)

Z-2089-2025  ·  Class II  ·  reported 2025-07-09  ·  Ongoing

Why it was recalled
Products have the incorrect platelet (PLT) value assignments for the ADVIA 120/2120/2120i Hematology Systems.
Product
ADVIA¿ 120/2120/2120i 3-in-1 TESTpoint Control (ABN2 Control)
Where it went
Worldwide distribution - US Nationwide and the countries of AR, AU, BR, CA, CL, CN, CO, CR, HK, ID, IN, JP, KR, MX, MY, PE, PY, SA, SG, TH, TW, UY, VN, & ZA.
How much
3,266 units

ADVIA¿ 120/2120/2120i 3-in-1 TESTpoint Control (Norm Control)

Z-2088-2025  ·  Class II  ·  reported 2025-07-09  ·  Ongoing

Why it was recalled
Products have the incorrect platelet (PLT) value assignments for the ADVIA 120/2120/2120i Hematology Systems.
Product
ADVIA¿ 120/2120/2120i 3-in-1 TESTpoint Control (Norm Control)
Where it went
Worldwide distribution - US Nationwide and the countries of AR, AU, BR, CA, CL, CN, CO, CR, HK, ID, IN, JP, KR, MX, MY, PE, PY, SA, SG, TH, TW, UY, VN, & ZA.
How much
3,435 units

ADVIA¿ 120/2120/2120i 3-in-1 TESTpoint Control (ABN1 Control)

Z-2087-2025  ·  Class II  ·  reported 2025-07-09  ·  Ongoing

Why it was recalled
Products have the incorrect platelet (PLT) value assignments for the ADVIA 120/2120/2120i Hematology Systems.
Product
ADVIA¿ 120/2120/2120i 3-in-1 TESTpoint Control (ABN1 Control)
Where it went
Worldwide distribution - US Nationwide and the countries of AR, AU, BR, CA, CL, CN, CO, CR, HK, ID, IN, JP, KR, MX, MY, PE, PY, SA, SG, TH, TW, UY, VN, & ZA.
How much
4,214 units

ADVIA¿ 120/2120/2120i SETpoint Calibrator

Z-2086-2025  ·  Class II  ·  reported 2025-07-09  ·  Ongoing

Why it was recalled
Products have the incorrect platelet (PLT) value assignments for the ADVIA 120/2120/2120i Hematology Systems.
Product
ADVIA¿ 120/2120/2120i SETpoint Calibrator
Where it went
Worldwide distribution - US Nationwide and the countries of AR, AU, BR, CA, CL, CN, CO, CR, HK, ID, IN, JP, KR, MX, MY, PE, PY, SA, SG, TH, TW, UY, VN, & ZA.
How much
2,020 units

Product: Atellica CH Diluent - CONS 2 x 1.5L; Siemens Material Number (SMN): 11099300;

Z-2027-2025  ·  Class II  ·  reported 2025-07-02  ·  Ongoing

Why it was recalled
Contamination of the diluent by sodium hypochlorite (NaOCl) which may impact calibrator, quality control (QC), and patient results for all assays except for sodium, potassium, and chloride.
Product
Product: Atellica CH Diluent - CONS 2 x 1.5L; Siemens Material Number (SMN): 11099300;
Where it went
Worldwide distribution - US Nationwide and the countries of Canada, Bahamas.
How much
1,625 units (1450 US, 175 OUS)

Atellica IM Homocysteine Calibrator (2 Pack). Siemens Material Number: 10995498

Z-1821-2025  ·  Class II  ·  reported 2025-06-04  ·  Ongoing

Why it was recalled
Certain lots of Atellica IM Homocysteine Calibrator and ADVIA Centaur HCY Calibrator do not meet the assay Instructions for Use (IFU) reference interval due to calibrator instability resulting in increased positive bias.
Product
Atellica IM Homocysteine Calibrator (2 Pack). Siemens Material Number: 10995498
Where it went
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KY, MA, MD, ME, MI, MO, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TX, UT, VA, VT, WI, WY and the countries of Austria, Belgium, Croatia, Curacao,¿St¿Eus, Czech¿Republic, Denmark, Egypt, Finland, France, Georgia, Germany, Greece, Hungary, Iran, Ireland, Israel, Italy, Jordan, Kuwait, Latvia, Lithuania, Nepal, Netherlands, Norway, Pakistan, Poland, Portugal, Romania, Serbia, Slovakia, Spain, Sweden, Switzerland, Turkey, Turkmenistan, U.A.E., United¿Kingdom, Vatikan City, Zaire.
How much
2,212 packs (4,424 units) - 448 packs US, 1,764 packs OUS

ADVIA Centaur HCY Calibrator (2 Pack). Siemens Material Number: 10310376.

Z-1822-2025  ·  Class II  ·  reported 2025-06-04  ·  Ongoing

Why it was recalled
Certain lots of Atellica IM Homocysteine Calibrator and ADVIA Centaur HCY Calibrator do not meet the assay Instructions for Use (IFU) reference interval due to calibrator instability resulting in increased positive bias.
Product
ADVIA Centaur HCY Calibrator (2 Pack). Siemens Material Number: 10310376.
Where it went
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KY, MA, MD, ME, MI, MO, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TX, UT, VA, VT, WI, WY and the countries of Austria, Belgium, Croatia, Curacao,¿St¿Eus, Czech¿Republic, Denmark, Egypt, Finland, France, Georgia, Germany, Greece, Hungary, Iran, Ireland, Israel, Italy, Jordan, Kuwait, Latvia, Lithuania, Nepal, Netherlands, Norway, Pakistan, Poland, Portugal, Romania, Serbia, Slovakia, Spain, Sweden, Switzerland, Turkey, Turkmenistan, U.A.E., United¿Kingdom, Vatikan City, Zaire.
How much
1,072 packs (2,144 units) - 156 packs US, 916 packs OUS

epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515; e

Z-1684-2025  ·  Class II  ·  reported 2025-05-07  ·  Ongoing

Why it was recalled
The occurrence of discrepant high pH results in samples introduced with higher injection volumes with epoc sensor configuration 45.n.
Product
epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515; epoc Software Version & Sensor Configuration: epoc Host SW v3.41.2, epoc NXS SW v4.14.9 & 4.14.11, Sensor Configuration 45.n.
Where it went
Domestic: Nationwide Distribution; Foreign: Albania, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia, Brazil, Cambodia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Ecuador, Estonia, France, Germany, Ghana, Greece, Guadeloupe, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Lithuania, Malaysia, Mali, Martinique, Mexico, Morocco, Netherlands, New Caledonia, New Zealand, Nigeria, Oman, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, San Marino, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Vietnam.
How much
264,252 units

Stratus CS Acute Care cTNI TestPak.IVD test for cardiac troponin I in heparinized plasma.

Z-1673-2025  ·  Class II  ·  reported 2025-05-07  ·  Ongoing

Why it was recalled
Siemens Healthcare Diagnostics Inc., is recalling their Stratus CS Acute Care cTNI TestPak, an IVD test for cardiac troponin I in heparinized plasma by removal. The reason for the recall is an increased occurrence of random non-repeatable false positive cardiac Troponin l(cTnI) results at any point during the TestPaks shelf life when using the Stratus CS cTnI Acute Care Testpak.
Product
Stratus CS Acute Care cTNI TestPak.IVD test for cardiac troponin I in heparinized plasma.
Where it went
Worldwide - US Nationwide distribution in the states of CA, IL, KS, MI, MO, NJ, NY, OH, OK, PA, SC & WI. The countries of Austria, Canada, Chile, France, Germany, Iraq, Italy, Malaysia, North Macedonia, Saudi Arabia, Serbia, Slovenia, Spain, Sweden, Switzerland & Vietnam.
How much
367 units

epoc BGEM Crea Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736382;

Z-1685-2025  ·  Class II  ·  reported 2025-05-07  ·  Ongoing

Why it was recalled
The occurrence of discrepant high pH results in samples introduced with higher injection volumes with epoc sensor configuration 45.n.
Product
epoc BGEM Crea Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736382; epoc Software Version & Sensor Configuration: epoc Host SW v3.41.2, epoc NXS SW v4.14.9 & 4.14.11, Sensor Configuration 45.n.
Where it went
Domestic: Nationwide Distribution; Foreign: Albania, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia, Brazil, Cambodia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Ecuador, Estonia, France, Germany, Ghana, Greece, Guadeloupe, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Lithuania, Malaysia, Mali, Martinique, Mexico, Morocco, Netherlands, New Caledonia, New Zealand, Nigeria, Oman, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, San Marino, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Vietnam.
How much
5,207 units

Atellica CH Revised C Reactive Protein (RCRP); Siemens Material Number (SMN): 11537223.

Z-1551-2025  ·  Class II  ·  reported 2025-04-16  ·  Ongoing

Why it was recalled
Incorrect software flagging may occur for the assay that may potentially lead to an erroneous result.
Product
Atellica CH Revised C Reactive Protein (RCRP); Siemens Material Number (SMN): 11537223.
Where it went
Domestic: Nationwide Distribution. International: Australia, Austria, Bahamas, Bahrain, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Cura¿ao,¿St¿Eus, Czech¿Republic, Denmark,¿Estonia, Finland, France, Germany, Greece,¿Hong Kong, Hungary, India, Ireland, Israel, Italy, Korea, Kuwait, Latvia, Lithuania, Malaysia, Maldives, Netherlands, New Zealand, Norway, Oman,¿Pakistan, Poland, Portugal, Qatar, Romania, Saudi¿Arabia, Serbia, Singapore, Slovakia, Slovenia, South¿Africa, Spain, Sweden, Switzerland, Taiwan, U.A.E., Uganda, United¿Kingdom, Uzbekistan, Vatikancity.¿ ¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿ ¿¿¿¿¿¿¿ ¿ ¿¿¿¿ ¿¿¿¿¿¿ ¿¿¿¿¿¿¿¿¿¿ ¿¿¿¿¿¿¿¿ ¿¿¿¿¿ ¿ ¿¿¿¿¿¿
How much
N/A

Magnesium Flex reagent cartridge, 360 tests, Catalog Number: K3057; Siemens Material Number (SMN): 10445158;

Z-0767-2025  ·  Class II  ·  reported 2025-01-01  ·  Ongoing

Why it was recalled
Siemens Healthcare Diagnostics is recalling two lots of their Dimension Vista¿ Total Magnesium (MG) Flex Reagent Cartridges (lots 24057BB and 24064BC) due to the potential for falsely low measurements of magnesium in plasma. Because the error is intermittent, existing quality control procedures may not recognize that a patients results are incorrect, and the results may be reported out
Product
Magnesium Flex reagent cartridge, 360 tests, Catalog Number: K3057; Siemens Material Number (SMN): 10445158;
Where it went
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Canada, Denmark, France, Germany, Italy, Slovenia, Spain.
How much
13,115 units (11,520 US, 1,595 ROW)

epoc NXS Host; Siemens Material Number (SMN): 11413498(JP) ; Software Version Number: 4.14.9. Intended for

Z-0736-2025  ·  Class II  ·  reported 2025-01-01  ·  Ongoing

Why it was recalled
Siemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement by the device if the user of the device does not ensure verification of the correct patient name. The software malfunction has the potential to impact patient safety via misdiagnosis and/or mistreatment by a clinician if the wrong patient information is saved with the test result from the subject device.
Product
epoc NXS Host; Siemens Material Number (SMN): 11413498(JP) ; Software Version Number: 4.14.9. Intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care.
Where it went
Nationwide Foreign: Albania Angola Argentina Australia Austria Bahrain Bangladesh Belarus Belgium Bolivia Bosnia and Herzegovina Brazil Cambodia Canada Chile China Colombia Croatia Czech Republic Denmark Ecuador Egypt Estonia Finland France Germany Ghana Greece Guadeloupe Hong Kong Hungary Iceland India Indonesia Iraq Ireland Israel Italy Japan Jordan Kazakhstan Kuwait Latvia Lithuania Luxembourg Macau Madagascar Malaysia Mexico Monaco Mongolia Morocco Myanmar Netherlands New Caledonia New Zealand Norway Paraguay Peru Philippines Poland Portugal Republic Korea Romania Russian Federation Saudi Arabia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Swaziland Sweden Switzerland Thailand Tunisia Turkey Ukraine United Arab Emirates United Kingdom Great Britain Uruguay Vietnam
How much
670 units

epoc NXS Host; Siemens Material Number (SMN): 11413879(IN) ; Software Version Number: 4.14.9. Intended for

Z-0741-2025  ·  Class II  ·  reported 2025-01-01  ·  Ongoing

Why it was recalled
Siemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement by the device if the user of the device does not ensure verification of the correct patient name. The software malfunction has the potential to impact patient safety via misdiagnosis and/or mistreatment by a clinician if the wrong patient information is saved with the test result from the subject device.
Product
epoc NXS Host; Siemens Material Number (SMN): 11413879(IN) ; Software Version Number: 4.14.9. Intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care.
Where it went
Nationwide Foreign: Albania Angola Argentina Australia Austria Bahrain Bangladesh Belarus Belgium Bolivia Bosnia and Herzegovina Brazil Cambodia Canada Chile China Colombia Croatia Czech Republic Denmark Ecuador Egypt Estonia Finland France Germany Ghana Greece Guadeloupe Hong Kong Hungary Iceland India Indonesia Iraq Ireland Israel Italy Japan Jordan Kazakhstan Kuwait Latvia Lithuania Luxembourg Macau Madagascar Malaysia Mexico Monaco Mongolia Morocco Myanmar Netherlands New Caledonia New Zealand Norway Paraguay Peru Philippines Poland Portugal Republic Korea Romania Russian Federation Saudi Arabia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Swaziland Sweden Switzerland Thailand Tunisia Turkey Ukraine United Arab Emirates United Kingdom Great Britain Uruguay Vietnam
How much
276 units

epoc NXS Host; Siemens Material Number (SMN): 11413518 (ROW) ; Software Version Number: 4.14.9. Intended fo

Z-0739-2025  ·  Class II  ·  reported 2025-01-01  ·  Ongoing

Why it was recalled
Siemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement by the device if the user of the device does not ensure verification of the correct patient name. The software malfunction has the potential to impact patient safety via misdiagnosis and/or mistreatment by a clinician if the wrong patient information is saved with the test result from the subject device.
Product
epoc NXS Host; Siemens Material Number (SMN): 11413518 (ROW) ; Software Version Number: 4.14.9. Intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care.
Where it went
Nationwide Foreign: Albania Angola Argentina Australia Austria Bahrain Bangladesh Belarus Belgium Bolivia Bosnia and Herzegovina Brazil Cambodia Canada Chile China Colombia Croatia Czech Republic Denmark Ecuador Egypt Estonia Finland France Germany Ghana Greece Guadeloupe Hong Kong Hungary Iceland India Indonesia Iraq Ireland Israel Italy Japan Jordan Kazakhstan Kuwait Latvia Lithuania Luxembourg Macau Madagascar Malaysia Mexico Monaco Mongolia Morocco Myanmar Netherlands New Caledonia New Zealand Norway Paraguay Peru Philippines Poland Portugal Republic Korea Romania Russian Federation Saudi Arabia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Swaziland Sweden Switzerland Thailand Tunisia Turkey Ukraine United Arab Emirates United Kingdom Great Britain Uruguay Vietnam
How much
2240 units

epoc NXS Host; Siemens Material Number (SMN): 11413475 (US); Software Version Number: 4.14.9. Intended for us

Z-0734-2025  ·  Class II  ·  reported 2025-01-01  ·  Ongoing

Why it was recalled
Siemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement by the device if the user of the device does not ensure verification of the correct patient name. The software malfunction has the potential to impact patient safety via misdiagnosis and/or mistreatment by a clinician if the wrong patient information is saved with the test result from the subject device.
Product
epoc NXS Host; Siemens Material Number (SMN): 11413475 (US); Software Version Number: 4.14.9. Intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care.
Where it went
Nationwide Foreign: Albania Angola Argentina Australia Austria Bahrain Bangladesh Belarus Belgium Bolivia Bosnia and Herzegovina Brazil Cambodia Canada Chile China Colombia Croatia Czech Republic Denmark Ecuador Egypt Estonia Finland France Germany Ghana Greece Guadeloupe Hong Kong Hungary Iceland India Indonesia Iraq Ireland Israel Italy Japan Jordan Kazakhstan Kuwait Latvia Lithuania Luxembourg Macau Madagascar Malaysia Mexico Monaco Mongolia Morocco Myanmar Netherlands New Caledonia New Zealand Norway Paraguay Peru Philippines Poland Portugal Republic Korea Romania Russian Federation Saudi Arabia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Swaziland Sweden Switzerland Thailand Tunisia Turkey Ukraine United Arab Emirates United Kingdom Great Britain Uruguay Vietnam
How much
6,738 units

epoc NXS Host; Siemens Material Number (SMN): 11413506 (CA); Software Version Number: 4.14.9. Intended for

Z-0737-2025  ·  Class II  ·  reported 2025-01-01  ·  Ongoing

Why it was recalled
Siemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement by the device if the user of the device does not ensure verification of the correct patient name. The software malfunction has the potential to impact patient safety via misdiagnosis and/or mistreatment by a clinician if the wrong patient information is saved with the test result from the subject device.
Product
epoc NXS Host; Siemens Material Number (SMN): 11413506 (CA); Software Version Number: 4.14.9. Intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care.
Where it went
Nationwide Foreign: Albania Angola Argentina Australia Austria Bahrain Bangladesh Belarus Belgium Bolivia Bosnia and Herzegovina Brazil Cambodia Canada Chile China Colombia Croatia Czech Republic Denmark Ecuador Egypt Estonia Finland France Germany Ghana Greece Guadeloupe Hong Kong Hungary Iceland India Indonesia Iraq Ireland Israel Italy Japan Jordan Kazakhstan Kuwait Latvia Lithuania Luxembourg Macau Madagascar Malaysia Mexico Monaco Mongolia Morocco Myanmar Netherlands New Caledonia New Zealand Norway Paraguay Peru Philippines Poland Portugal Republic Korea Romania Russian Federation Saudi Arabia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Swaziland Sweden Switzerland Thailand Tunisia Turkey Ukraine United Arab Emirates United Kingdom Great Britain Uruguay Vietnam
How much
320 units

Showing the 30 most recent of 662. The rest are searchable at the FDA enforcement report database.

Read these counts with three things in mind

A firm with more recalls usually makes more products. A contract packer runs hundreds of lines. It will out-recall a company with one product whatever either does. This is not a safety ranking.

Most recalls here are voluntary. A firm found its own fault and pulled the stock. That is the system working. It lands in this database looking the same as a recall the FDA had to push for.

Terminated means the FDA considers the recall finished. It is not a verdict. It does not mean anyone was harmed.

Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.