Roche Diagnostics Corporation

Roche Diagnostics Corporation: 39 FDA recall reports

Every medical device recall the FDA has published for Roche Diagnostics Corporation. The fault, and how the agency graded it.

39Reports
1Class I
0Ongoing
2019-07-24Most recent
1 Class I on this page

Class I means the FDA judged there was a reasonable probability of serious harm or death. Check the lot numbers against anything you hold.

What the classes mean

The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.

ClassWhat the FDA is saying
Class IThere is a reasonable probability that using this will cause serious harm or death.
Class IIIt may cause temporary or reversible harm. Serious harm is unlikely.
Class IIIIt breaks a rule but is unlikely to hurt anyone. Often a labelling fault.

Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.

By class

ClassReports
Class II36
Class III2
Class I1

By year reported

YearReports
20176
201821
201912
Checking whether it affects you

The lot or code number is what matters. Not the product name. A recall covers specific lots. Yours may not be one of them.

Check the code on your package against the report. The FDA keeps the current list at fda.gov/safety/recalls.

The reports

The text below is quoted from the FDA record word for word.

The Tina-quant lgA Gen.2 is an immunoturbidimetric assay. Anti-lgA antibodies react with antigen in the sample

Z-1979-2019  ·  Class II  ·  reported 2019-07-24  ·  Terminated

Why it was recalled
Calibration failures with Std.E, Sens.E, or Dup.E error due to atypical reaction kinetics and quality control failures have occurred due to contamination of R1 reagent with R3 reagent.
Product
The Tina-quant lgA Gen.2 is an immunoturbidimetric assay. Anti-lgA antibodies react with antigen in the sample to form an antigen/antibody complex which is measured turbidimetrically.
Where it went
The products were distributed to the following US states: AL, IN, and KY.
How much
8 units

cobas p 501 post-analytical units, Part Number 05158494001. For diagnostic sample storage and retrieval.

Z-1927-2019  ·  Class III  ·  reported 2019-07-10  ·  Terminated

Why it was recalled
There is a potential for the tray input flap on the post analytical units to become loose, potentially detaching from the instrument.
Product
cobas p 501 post-analytical units, Part Number 05158494001. For diagnostic sample storage and retrieval.
Where it went
US Nationwide Distribution in the states of AL, CA, FL, IA, IL, KY, MD, OR & PA
How much
4 units

cobas p 701 post-analytical units, Part Number 05349966001. For diagnostic sample storage and retrieval.

Z-1928-2019  ·  Class III  ·  reported 2019-07-10  ·  Terminated

Why it was recalled
There is a potential for the tray input flap on the post analytical units to become loose, potentially detaching from the instrument.
Product
cobas p 701 post-analytical units, Part Number 05349966001. For diagnostic sample storage and retrieval.
Where it went
US Nationwide Distribution in the states of AL, CA, FL, IA, IL, KY, MD, OR & PA
How much
9 units

Cobas¿ infinity central lab IT solution Material Number: 07154003001 Calculator/Data Processing Module,

Z-1368-2019  ·  Class II  ·  reported 2019-05-29  ·  Terminated

Why it was recalled
Using the following versions of cobas infinity software (2.0 thorough 2.5), there is a potential where the results from the first sample subsequent to the bottle change, (when using the Standby bottle function), does not display a quality control (QC) alarm when one should be present. If the QC is out of range on the stand-by bottle and it is not removed from the system, there is a potential that when the status of the Standby bottle changes from the Standby bottle to the Current bottle status, QC alarms will not be displayed next to the sample result.
Product
Cobas¿ infinity central lab IT solution Material Number: 07154003001 Calculator/Data Processing Module, For Clinical Use
Where it went
US Distribution to states of: AL, AZ, CA, GA, IA, MO, NJ, NY, SC, TN and TX.
How much
15 units

cobas e 801 immunoassay analyzer

Z-1249-2019  ·  Class II  ·  reported 2019-05-08  ·  Terminated

Why it was recalled
The firm has confirmed complaints regarding signal drops on the cobas e 801 module after a ProCell II M bottle changeover, which may lead to incorrect medical decisions with respect to diagnostics and patient treatment.
Product
cobas e 801 immunoassay analyzer
Where it went
Distributed to accounts in AL AZ CA CT FL HI IA IL IN KY LA MA MD ME MI MN MO NC NH NJ NY OH OR PA PR SC TN TX VA WA WI
How much
149

Lactate Gen.2, Catalog Nos. 03183700190, 05171881190, 11822837190, ***added as of 5/10/19*** 03183700190

Z-1146-2019  ·  Class II  ·  reported 2019-04-24  ·  Terminated

Why it was recalled
New endogenous interference claims have been established for the following assays used on the cobas c 311 analyzer and the cobas c 501, 502, 701, 702, and Modular Analytics P/D modules which may have potential medical risk. ***Update: Cobas c111 and COBAS INTEGRA 400 plus analyzers are also impacted for BILT3 and LACT2 assays***
Product
Lactate Gen.2, Catalog Nos. 03183700190, 05171881190, 11822837190, ***added as of 5/10/19*** 03183700190 Product Usage: An In vitro diagnostic reagent system intended for use on cobas Integra for the quantitative determination of the lactate concentration in plasma and cerebrospinal fluid.
Where it went
US Nationwide distribution.
How much
5292 analytical units

Albumin BCP Product Usage: ALBP is an in vitro test used for the quantitative determination of albumin in

Z-1194-2019  ·  Class II  ·  reported 2019-04-24  ·  Terminated

Why it was recalled
The reagent lot is exhibiting calibration failures, QC imprecision, and elevated QC recovery due to elevated pH level causing higher absorbance values.
Product
Albumin BCP Product Usage: ALBP is an in vitro test used for the quantitative determination of albumin in human serum and plasma on Roche/Hitachi cobas c systems.
Where it went
US Nationwide distribution in the states of: GA, IL, KY, MI, and TX.
How much
774

Bilirubin Total Gen.3, Cat. Nos. 05795397190, 05795419190, 05795320160, 05795338160, 05795346160 ***added a

Z-1145-2019  ·  Class II  ·  reported 2019-04-24  ·  Terminated

Why it was recalled
New endogenous interference claims have been established for the following assays used on the cobas c 311 analyzer and the cobas c 501, 502, 701, 702, and Modular Analytics P/D modules which may have potential medical risk. ***Update: Cobas c111 and COBAS INTEGRA 400 plus analyzers are also impacted for BILT3 and LACT2 assays***
Product
Bilirubin Total Gen.3, Cat. Nos. 05795397190, 05795419190, 05795320160, 05795338160, 05795346160 ***added as of 5/10/19***05795397190, 05795648190 Product Usage: In vitro test for the quantitative determination of total bilirubin in serum and plasma of adults and neonates on Roche/Hitachi cobas c systems.
Where it went
US Nationwide distribution.
How much
5292 analytical units

Tina-quant IgM Gen.2, Catalog Nos. 03507190190, 05220726190, 03507041190, 03507149190 Product Usage: Immun

Z-1147-2019  ·  Class II  ·  reported 2019-04-24  ·  Terminated

Why it was recalled
New endogenous interference claims have been established for the following assays used on the cobas c 311 analyzer and the cobas c 501, 502, 701, 702, and Modular Analytics P/D modules which may have potential medical risk. ***Update: Cobas c111 and COBAS INTEGRA 400 plus analyzers are also impacted for BILT3 and LACT2 assays***
Product
Tina-quant IgM Gen.2, Catalog Nos. 03507190190, 05220726190, 03507041190, 03507149190 Product Usage: Immunoturbidimetric assay for the quantitative in vitro determination of IgM in human serum and plasma on automated clinical chemistry analyzers analyzers.
Where it went
US Nationwide distribution.
How much
5292 analytical units

COBAS INTEGRA 400 plus analyzer used in the following assays: a)Tina-quant Apolipoprotein A-1 ver.2 Catalog

Z-1124-2019  ·  Class II  ·  reported 2019-04-17  ·  Terminated

Why it was recalled
COBAS INTEGRA¿ 400 plus and cobas c 111 analyzers Additional Extra Wash Cycle Configurations and/or Batch Mode Needed for Specific Tests Due to Carry-over
Product
COBAS INTEGRA 400 plus analyzer used in the following assays: a)Tina-quant Apolipoprotein A-1 ver.2 Catalog Number: 03032566122 b)Tina-quant Albumin Gen.2 Catalog Number 04469658190 c)Creatinine plus ver.2 Catalog Number 03263991190 d) riglycerides Catalog Number :20767107322 - Product Usage: Tina-quant Apolipoprotein A-1 ver.2 A lipoprotein test system is a device intended to measure lipoprotein in serum and plasma. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders and atherosclerosis Tina-quant Albumin Gen.2 Immunoturbidimetric assay for the quantitative, in vitro determination of albumin in human urine, serum, plasma, and CSF on Roche/Hitachi cobas c systems
Where it went
US Nationwide distribution.
How much
656 units

cobas c 111 with ISE analyzers analyzer used in the following assays: a)Creatinine plus ver.2 Catalog Numbe

Z-1125-2019  ·  Class II  ·  reported 2019-04-17  ·  Terminated

Why it was recalled
COBAS INTEGRA¿ 400 plus and cobas c 111 analyzers Additional Extra Wash Cycle Configurations and/or Batch Mode Needed for Specific Tests Due to Carry-over
Product
cobas c 111 with ISE analyzers analyzer used in the following assays: a)Creatinine plus ver.2 Catalog Number: 05401470190; b)Triglycerides Catalog Number: 04657594190 - Product Usage: Tina-quant Apolipoprotein A-1 ver.2 A lipoprotein test system is a device intended to measure lipoprotein in serum and plasma. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders and atherosclerosis Tina-quant Albumin Gen.2 Immunoturbidimetric assay for the quantitative, in vitro determination of albumin in human urine, serum, plasma, and CSF on Roche/Hitachi cobas c systems
Where it went
US Nationwide distribution.
How much
137 units

Anti-HAV IgM test System, Material No. 11820591160 Product Usage: The Roche Elecsys Anti-HAV IgM immunoass

Z-0784-2019  ·  Class II  ·  reported 2019-02-06  ·  Terminated

Why it was recalled
Potential for erroneous result messages for the Anti-HAV IgM assay when run on the cobas e 602 module. Discrepant result reporting may result.
Product
Anti-HAV IgM test System, Material No. 11820591160 Product Usage: The Roche Elecsys Anti-HAV IgM immunoassay is used for the in vitro qualitative detection of IgM antibodies to hepatitis A virus (anti-HAV IgM) in human serum and plasma (potassium EDTA, lithium or sodium heparin, sodium citrate). The assay is intended for use as an aid in the laboratory diagnosis of an acute or recently acquired hepatitis A virus infection. Assay results, in conjunction with other laboratory results and clinical information, may be used to provide presumptive evidence of infection with hepatitis A virus in persons with signs and symptoms of hepatitis and in persons at risk for hepatitis A infection. The electrochemiluminescence immunoassay ECLIA is intended for use on Elecsys and cobas e immunoassay analyzers.
Where it went
US Nationwide Distribution
How much
70,639 units
Class I

CoaguChek¿ XS System Prothrombin time test: CoaguChek XS PT Test 2X24 Strips, Catalog Number: 04625315160; C

Z-0360-2019  ·  Class I  ·  reported 2018-11-14  ·  Terminated

Why it was recalled
Abnormally high INR test results with the affected CoaguChek test strips
Product
CoaguChek¿ XS System Prothrombin time test: CoaguChek XS PT Test 2X24 Strips, Catalog Number: 04625315160; CoaguChek XS PT Test 6 Strips, Catalog Number: 04625374160; CoaguChek XS PT Test 24 Tests, Catalog Number: 07797826160;
Where it went
Nationwide
How much
1,163,952 boxes

Roche Diagnostics URISYS 1100 Urine Analyzer Part Number: 03617556001 Product Usage: The URISYS 1100 Urin

Z-0301-2019  ·  Class II  ·  reported 2018-11-07  ·  Terminated

Why it was recalled
Limits of Detection (LoD) for protein, nitrite, leukocytes, and erythrocytes on the Urisys 1100 urine analyzer with Chemstrip 5 OB, Chemstrip 7, Chemstrip 10 MD, and Chemstrip 10 UA test strips were higher than what is listed in the test strip method sheets may lead to false negative results for the affected parameters
Product
Roche Diagnostics URISYS 1100 Urine Analyzer Part Number: 03617556001 Product Usage: The URISYS 1100 Urine Analyzer is a reflectance photometer designed to automatically read and evaluate the results of Chemstrip 5 OB, Chemstrip 7, Chemstrip 10 MD, and Chemstrip 10 UA urine test strips for various urine analytes.: specific gravity, pH, leukocytes, nitrite, protein, glucose, ketones, urobilinogen, bilirubin and blood
Where it went
US Nationwide Distribution
How much
14,398 units

Roche cobas p 612 pre-analytical system (63x) Model: 07563116001 The cobas p 612 pre-analytical system is

Z-2944-2018  ·  Class II  ·  reported 2018-09-05  ·  Terminated

Why it was recalled
Sample material may potentially come in contact with the pipetting nozzle during operation of cobas p 612 pre-analytical system (63x) (LCP1) with non-filter tips. The possible presence of biological material on the nozzle may cause contamination resulting in potential false positive results, depending on the sensitivity of the analytical technology.
Product
Roche cobas p 612 pre-analytical system (63x) Model: 07563116001 The cobas p 612 pre-analytical system is a computer controlled fully automatic system for sorting of open and closed barcoded and centrifuged and non-centrifuged sample tubes. The cobas p 612 is intended for use with analyzers that perform tests in the area of clinical chemistry, immuno chemistry, coagulation, hematology, urinalysis, nucleic acid testing.
Where it went
US Distribution to states of: CA, NY,TX, WA, and WI.
How much
12 units

Elecsys Vitamin D total II, Vitamin D Test System, Material number 07464215160 Product Usage: This assay i

Z-2390-2018  ·  Class II  ·  reported 2018-07-18  ·  Terminated

Why it was recalled
The device may give a falsely elevated result that is non-reproducible. If vitamin D level is monitored by laboratory tests and adapted based on the result, a dosage reduction of vitamin D supplementation cannot be excluded, and as a consequence, vitamin D deficiency may develop over time.
Product
Elecsys Vitamin D total II, Vitamin D Test System, Material number 07464215160 Product Usage: This assay is intended for the quantitative determination of total 25-hydroxyvitamin D in human serum and plasma. This assay is to be used as an aid in the assessment of vitamin D sufficiency in adults. The electrochemiluminescence binding assay is intended for use on the immunoassay analyzer.
Where it went
US Nationwide distribution.
How much
3591 units

Elecsys Vitamin D total II, Vitamin D Test System, Material number 07028148190 Product Usage: This assay i

Z-2389-2018  ·  Class II  ·  reported 2018-07-18  ·  Terminated

Why it was recalled
The device may give a falsely elevated result that is non-reproducible. If vitamin D level is monitored by laboratory tests and adapted based on the result, a dosage reduction of vitamin D supplementation cannot be excluded, and as a consequence, vitamin D deficiency may develop over time.
Product
Elecsys Vitamin D total II, Vitamin D Test System, Material number 07028148190 Product Usage: This assay is intended for the quantitative determination of total 25-hydroxyvitamin D in human serum and plasma. This assay is to be used as an aid in the assessment of vitamin D sufficiency in adults. The electrochemiluminescence binding assay is intended for use on the immunoassay analyzer.
Where it went
US Nationwide distribution.
How much
347 units

cobas e 801 immunoassay analyzer, cobas 8000 Modular Series The device is a fully automated, software contr

Z-2136-2018  ·  Class II  ·  reported 2018-06-20  ·  Terminated

Why it was recalled
The syringe plunger on the device may be mounted in a tilted position, which applies mechanical stress to the plunger. In the worst case scenario, the plunger may crack leading to impaired syringe function.
Product
cobas e 801 immunoassay analyzer, cobas 8000 Modular Series The device is a fully automated, software controlled analyzer system for in vitro determination of analytes in human body fluids. It is part of the cobas 8000 modular analyzer series. It uses electrochemiluminescent technology for signal generation and measurement.
Where it went
US Distribution to the states of : CA, CT, IA, IN, KY, ME, MI, NC, NY, OH, SC, TX, and VA.
How much
35

cobas b 221 AutoQC Module Complete; Catalog numbers: 3313522001 and 3313522692 The Roche Diagnostics Omni S

Z-1613-2018  ·  Class II  ·  reported 2018-05-09  ·  Terminated

Why it was recalled
The software responsible for starting scheduled AutoQC measurements (scheduler) will not activate.
Product
cobas b 221 AutoQC Module Complete; Catalog numbers: 3313522001 and 3313522692 The Roche Diagnostics Omni S Analyzer is a fully automated critical care analyzer intended to be used for the measurement of pH, PCO2, PO2, sodium, potassium, ionized calcium, chloride, hematocrit, glucose, lactate, urea/BUN, total hemoglobin, oxygen saturation, oxyhemoglobin, deoxyhemoglobin, carboxyhemoglobin, and methemoglobin in samples of whole blood, serum, plasma, and aqueous solutions as appropriate.
Where it went
USA ( nationwide)
How much
1026 total products

cobas b 221<2>Roche OMNI S2 system catalog numbers: 3337111001 and 3337111692 The Roche Diagnostics Omni S

Z-1610-2018  ·  Class II  ·  reported 2018-05-09  ·  Terminated

Why it was recalled
The software responsible for starting scheduled AutoQC measurements (scheduler) will not activate.
Product
cobas b 221<2>Roche OMNI S2 system catalog numbers: 3337111001 and 3337111692 The Roche Diagnostics Omni S Analyzer is a fully automated critical care analyzer intended to be used for the measurement of pH, PCO2, PO2, sodium, potassium, ionized calcium, chloride, hematocrit, glucose, lactate, urea/BUN, total hemoglobin, oxygen saturation, oxyhemoglobin, deoxyhemoglobin, carboxyhemoglobin, and methemoglobin in samples of whole blood, serum, plasma, and aqueous solutions as appropriate.
Where it went
USA ( nationwide)
How much
1026 total products

cobas b 221<6>Roche OMNI S6 system; Catalog numbers: 3337154001 and 3337154692 The Roche Diagnostics Omni

Z-1612-2018  ·  Class II  ·  reported 2018-05-09  ·  Terminated

Why it was recalled
The software responsible for starting scheduled AutoQC measurements (scheduler) will not activate.
Product
cobas b 221<6>Roche OMNI S6 system; Catalog numbers: 3337154001 and 3337154692 The Roche Diagnostics Omni S Analyzer is a fully automated critical care analyzer intended to be used for the measurement of pH, PCO2, PO2, sodium, potassium, ionized calcium, chloride, hematocrit, glucose, lactate, urea/BUN, total hemoglobin, oxygen saturation, oxyhemoglobin, deoxyhemoglobin, carboxyhemoglobin, and methemoglobin in samples of whole blood, serum, plasma, and aqueous solutions as appropriate.
Where it went
USA ( nationwide)
How much
1026 total products

Cobas b 221<4>Roche OMNI S4 system; Catalog numbers: 3337138001 and 3337138692 The Roche Diagnostics Omni

Z-1611-2018  ·  Class II  ·  reported 2018-05-09  ·  Terminated

Why it was recalled
The software responsible for starting scheduled AutoQC measurements (scheduler) will not activate.
Product
Cobas b 221<4>Roche OMNI S4 system; Catalog numbers: 3337138001 and 3337138692 The Roche Diagnostics Omni S Analyzer is a fully automated critical care analyzer intended to be used for the measurement of pH, PCO2, PO2, sodium, potassium, ionized calcium, chloride, hematocrit, glucose, lactate, urea/BUN, total hemoglobin, oxygen saturation, oxyhemoglobin, deoxyhemoglobin, carboxyhemoglobin, and methemoglobin in samples of whole blood, serum, plasma, and aqueous solutions as appropriate.
Where it went
USA ( nationwide)
How much
1026 total products

Roche Acetaminophen assay 03255379160-HAcetaminophen (P) 20767174160- cobas c Integra AAcetaminophen150 test

Z-1519-2018  ·  Class II  ·  reported 2018-05-02  ·  Terminated

Why it was recalled
The previous version of method sheets for the Acetaminophen assay on the COBAS INTEGRA 400 plus/800/800 CTS analyzers and the MODULAR ANALYTICS P module are missing the following statement in the Limitations-interference section: "In very rare cases, gammopathy, in particular type IgM (Waldenstr¿m s macroglobulinemia), may cause unreliable results."
Product
Roche Acetaminophen assay 03255379160-HAcetaminophen (P) 20767174160- cobas c Integra AAcetaminophen150 test
Where it went
Nationwide Distribution
How much
35,111

Elecsys BRAHMS PCT Immunoassay for the in vitro quantitative determination of PCT (procalcitonin) in human

Z-1311-2018  ·  Class II  ·  reported 2018-04-11  ·  Terminated

Why it was recalled
Product exhibits a decreased recovery of patient samples on the cobas e 801 module compared to the MODULAR ANALYTICS E 170, cobas e 411, 601, and 602 analyzers.
Product
Elecsys BRAHMS PCT Immunoassay for the in vitro quantitative determination of PCT (procalcitonin) in human serum and plasma (K2 and K3 EDTA, Li-Heparin).
Where it went
US Distribution in states of: CA, FL and IN.
How much
47

c6000, Chemistry (Photometric, Discrete), for clinical use. - cobas e 601 module: Immunoassay Analyzer, Catalo

Z-1268-2018  ·  Class II  ·  reported 2018-04-04  ·  Terminated

Why it was recalled
Complaints have been received concerning questionable results on analyzers due to customers utilizing 13 sample tubes without tube adapters.
Product
c6000, Chemistry (Photometric, Discrete), for clinical use. - cobas e 601 module: Immunoassay Analyzer, Catalog Numbers: 04745922001, 05036348001, 05860652001, 04745922692, 05036348001 & 05036348692
Where it went
US Nationwide

cobas e 411 Immunoassay Analyzer (Discrete Photometric Analyzer for Clinical use), Catalog Numbers: 047752010

Z-1270-2018  ·  Class II  ·  reported 2018-04-04  ·  Terminated

Why it was recalled
Complaints have been received concerning questionable results on analyzers due to customers utilizing 13 mm sample tubes without tube adapters.
Product
cobas e 411 Immunoassay Analyzer (Discrete Photometric Analyzer for Clinical use), Catalog Numbers: 04775201001 (rack system), 04775279001 (disk system), 04775201973 (rack system) & 04775279973 (disk system).
Where it went
US Nationwide

Modular E 170 Analyzer: Immunoassay Chemistry Analyzer, Chemistry (Photometric, Discrete) for clinical use, C

Z-1269-2018  ·  Class II  ·  reported 2018-04-04  ·  Terminated

Why it was recalled
Complaints have been received concerning questionable results on analyzers due to customers utilizing sample 13 mm tubes without tube adapters.
Product
Modular E 170 Analyzer: Immunoassay Chemistry Analyzer, Chemistry (Photometric, Discrete) for clinical use, Catalog Numbers: 03739040692, 03023109001, 05023599001, 05023572001, 03739040001 & 03023109973.
Where it went
US Nationwide

cobas 6000 Modular Series system; Part Numbers: 1. 04745914001 cobas c 501 module; 2. 05036453001 cobas 60

Z-0909-2018  ·  Class II  ·  reported 2018-03-14  ·  Terminated

Why it was recalled
A warning is being added to the cobas¿ 6000 MODULAR Series System - c501 and the cobas¿ 8000 MODULAR Analyzer Series - c502 Operator Manuals to prevent operator injury during maintenance.
Product
cobas 6000 Modular Series system; Part Numbers: 1. 04745914001 cobas c 501 module; 2. 05036453001 cobas 6000 UL c 501 + Core Fix configuration; 3. 05036453692 - cobas 6000 UL + CORE CU; and 4. 05860636001 cobas 6000 c 501 (UL) V
Where it went
Distributed in all 50 states: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MN, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, the District of Columbia, and Puerto Rico.
How much
3433

cobas 8000 Modular Analyzer Series; Part Numbers: 1. 05964067001 cobas 8000 c 502 module

Z-0910-2018  ·  Class II  ·  reported 2018-03-14  ·  Terminated

Why it was recalled
A warning is being added to the cobas¿ 6000 analyzer and the cobas¿ 8000 modular analyzer series Operators Manual to prevent operator injury during maintenance.
Product
cobas 8000 Modular Analyzer Series; Part Numbers: 1. 05964067001 cobas 8000 c 502 module
Where it went
Distributed in all 50 states: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MN, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, the District of Columbia, and Puerto Rico.
How much
245

Cobas 8000 Modular Series system; Software Version 05-02 and 06-03; UDI: 05641446001

Z-0675-2018  ·  Class II  ·  reported 2018-02-28  ·  Terminated

Why it was recalled
A software failure may incorrectly set the system settings to "default" settings, creating a risk of incorrect results.
Product
Cobas 8000 Modular Series system; Software Version 05-02 and 06-03; UDI: 05641446001
Where it went
Distributed in 36 states: AL, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, NC, NE, NH, NJ, NY, OH, OK, OR, PA, RI, TN, TX, UT, VA, WA, WI and Puerto Rico.
How much
399

Showing the 30 most recent of 39. The rest are searchable at the FDA enforcement report database.

Read these counts with three things in mind

A firm with more recalls usually makes more products. A contract packer runs hundreds of lines. It will out-recall a company with one product whatever either does. This is not a safety ranking.

Most recalls here are voluntary. A firm found its own fault and pulled the stock. That is the system working. It lands in this database looking the same as a recall the FDA had to push for.

Terminated means the FDA considers the recall finished. It is not a verdict. It does not mean anyone was harmed.

Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.

← All device recalls