Roche Diabetes Care, Inc.
Roche Diabetes Care, Inc.: 29 FDA recall reports
Every medical device recall the FDA has published for Roche Diabetes Care, Inc.. The fault, and how the agency graded it.
29Reports
0Class I
5Ongoing
2024-07-03Most recent
What the classes mean
The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.
Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.
By class
By year reported
Checking whether it affects you
The lot or code number is what matters. Not the product name. A recall covers
specific lots. Yours may not be one of them.
Check the code on your package against the report. The FDA keeps the current
list at fda.gov/safety/recalls.
The reports
The text below is quoted from the FDA record word for word.
Accu-Chek Guide (SC) Kit -Intended to quantitatively measure glucose in fresh capillary whole blood from the f
Z-2174-2024 · Class II · reported 2024-07-03 · Ongoing
- Why it was recalled
- The reason for recall is the meters may show an incorrect measurement unit (mmol/L rather than mg/dL) and may result in the wrong unit of measure. Which may subsequently cause an error in interpreting a result as approximately 18 times lower that the actual blood glucose. Potential immediate health consequences of the glucose meter providing a reading in the wrong units of measurement (UOM) may result in inappropriate rescue therapy for presumed hypoglycemia including urgent and repeat administration of carbohydrates. This in turn could lead to frequent episodes of hyperglycemia which is unrecognized and undertreated.Long range consequences of continued use of the affected meters could include mismanagement of the patient s diabetes for longer periods of time potentially resulting in a major health event (in general persistent and/or severe hyperglycemia) that may lead them to seek medical attention, and depending on the duration of hyperglycemia, could contribute to the development of microvascular (i.e., retinopathy, neuropathy, nephropathy) and/or macrovascular (i.e., myocardial infarction, ischemic stroke) complications of diabetes and could lead to fetal damage (abnormal...
- Product
- Accu-Chek Guide (SC) Kit -Intended to quantitatively measure glucose in fresh capillary whole blood from the fingertip, palm, and upper arm as an aid in monitoring the effectiveness of glucose control. Product REF Number: 08453071001
- Where it went
- Nationwide and Worldwide
- How much
- 48 units
RocheDiabetes Care Platform Software version 2.5.2 material number 08888353001. This is a web application soft
Z-1008-2022 · Class II · reported 2022-05-04 · Terminated
- Why it was recalled
- Potential for patient data mismatch when using browser "back" button to navigate between patients when using the diabetes management software.
- Product
- RocheDiabetes Care Platform Software version 2.5.2 material number 08888353001. This is a web application software hosted on a server. There is no physical product.
- Where it went
- US Nationwide Distribution: AK, AZ, CA, CO, DC, FL, GA, IL, IN, MN, NC, NJ, NY, PA, RI, SC, TX, WI
- How much
- 81 registered accounts
Accu-Chek Performa Test Strip, Blood Glucose Monitoring System, Model number 07299702001
Z-2352-2021 · Class II · reported 2021-09-01 · Ongoing
- Why it was recalled
- Test strip vials may open while inside sealed cartons during shipment. An open vial might expose the test strips to humidity, which might damage the strips and result in inaccurate results, potentially leading to inappropriate therapy decisions and adverse health consequences.
- Product
- Accu-Chek Performa Test Strip, Blood Glucose Monitoring System, Model number 07299702001
- Where it went
- Worldwide distribution - US Nationwide distribution.
- How much
- 146,854 (US); 59,815,555 (OUS)
Accu-Chek Aviva Plus Test Strip, Blood Glucose Monitoring System, Model numbers: 06908152001, 06908217001, 069
Z-2351-2021 · Class II · reported 2021-09-01 · Ongoing
- Why it was recalled
- Test strip vials may open while inside sealed cartons during shipment. An open vial might expose the test strips to humidity, which might damage the strips and result in inaccurate results, potentially leading to inappropriate therapy decisions and adverse health consequences.
- Product
- Accu-Chek Aviva Plus Test Strip, Blood Glucose Monitoring System, Model numbers: 06908152001, 06908217001, 06908268001, 06908349001, 06908373001
- Where it went
- Worldwide distribution - US Nationwide distribution.
- How much
- 10,546,420 (US), 19,120,668 (OUS)
Accu-Chek SmartView Test Strip, Blood Glucose Monitoring System, Model numbers: 06337538001, 06337546001, 063
Z-2353-2021 · Class II · reported 2021-09-01 · Ongoing
- Why it was recalled
- Test strip vials may open while inside sealed cartons during shipment. An open vial might expose the test strips to humidity, which might damage the strips and result in inaccurate results, potentially leading to inappropriate therapy decisions and adverse health consequences.
- Product
- Accu-Chek SmartView Test Strip, Blood Glucose Monitoring System, Model numbers: 06337538001, 06337546001, 06337562001
- Where it went
- Worldwide distribution - US Nationwide distribution.
- How much
- 1,150,755 (US)
Accu-Chek Inform II Test Strip, Whole Blood Glucose Test System, Model number 05942891001
Z-2354-2021 · Class II · reported 2021-09-01 · Ongoing
- Why it was recalled
- Test strip vials may open while inside sealed cartons during shipment. An open vial might expose the test strips to humidity, which might damage the strips and result in inaccurate results, potentially leading to inappropriate therapy decisions and adverse health consequences.
- Product
- Accu-Chek Inform II Test Strip, Whole Blood Glucose Test System, Model number 05942891001
- Where it went
- Worldwide distribution - US Nationwide distribution.
- How much
- 9,555,555 (US); 5,128,758 (OUS)
Accu-Chek Aviva Plus Health Network Strips- intended for Blood Glucose Monitoring System Catalog Number: 069
Z-0997-2020 · Class II · reported 2020-02-12 · Terminated
- Why it was recalled
- Inability to dose the test strips, may result in a temporary inability to get a result that could lead to a delay in treatment and/or therapy decisions
- Product
- Accu-Chek Aviva Plus Health Network Strips- intended for Blood Glucose Monitoring System Catalog Number: 06908349001
- Where it went
- Nationwide: Foreign: Austalia, India
- How much
- 82,332 vials
Accu-Chek Aviva Plus Retail Strips 50 ct - intended for Blood Glucose Monitoring System Catalog Number: 06908
Z-0998-2020 · Class II · reported 2020-02-12 · Terminated
- Why it was recalled
- Inability to dose the test strips, may result in a temporary inability to get a result that could lead to a delay in treatment and/or therapy decisions
- Product
- Accu-Chek Aviva Plus Retail Strips 50 ct - intended for Blood Glucose Monitoring System Catalog Number: 06908217001
- Where it went
- Nationwide: Foreign: Austalia, India
- How much
- 92,836 vials
ACCU-CHEK AVIVA ASPAC STRIP 50CT- intended for Blood Glucose Monitoring System Catalog Number: 06453970020
Z-0999-2020 · Class II · reported 2020-02-12 · Terminated
- Why it was recalled
- Inability to dose the test strips, may result in a temporary inability to get a result that could lead to a delay in treatment and/or therapy decisions
- Product
- ACCU-CHEK AVIVA ASPAC STRIP 50CT- intended for Blood Glucose Monitoring System Catalog Number: 06453970020
- Where it went
- Nationwide: Foreign: Austalia, India
- How much
- 61,329 units (OUS)
Blood Glucose Monitoring System, Model Number 08453063001 ACCU-CHEK GUIDE SAMPLE KIT FCN MG/DL US - Product U
Z-0751-2020 · Class II · reported 2020-01-08 · Terminated
- Why it was recalled
- Potential power issues with the Accu-Chek¿ Guide Blood Glucose Monitoring System.
- Product
- Blood Glucose Monitoring System, Model Number 08453063001 ACCU-CHEK GUIDE SAMPLE KIT FCN MG/DL US - Product Usage: The Accu-Chek Guide Blood Glucose Monitoring System is intended to quantitatively measure glucose in fresh capillary whole blood from the fingertip, palm, and upper arm as an aid in monitoring the effectiveness of glucose control. The Accu-Chek Guide blood Glucose Monitoring System is intended for in vitro diagnostic single-patient use by people with diabetes. The Accu-Chek Guide Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.
- Where it went
- US Nationwide distribution and countries of AR, AT, AU, BE, BR, CA, CH, CL, DE, DK, EC, ES, FI, FR, GB, GR, HK, HR, ID, IN, IT, JP, KR, LB, LT, LU, MM, MY, NL, NO, PH, PK, PT, SE, SG, TH, TW, ZA.
- How much
- 355,400
Blood Glucose Monitoring System, Model Number 08313202001 ACCU CHEK GUIDE HCP SAMPLE LAUNCH BOX - Product Usa
Z-0752-2020 · Class II · reported 2020-01-08 · Terminated
- Why it was recalled
- Potential power issues with the Accu-Chek¿ Guide Blood Glucose Monitoring System.
- Product
- Blood Glucose Monitoring System, Model Number 08313202001 ACCU CHEK GUIDE HCP SAMPLE LAUNCH BOX - Product Usage: The Accu-Chek Guide Blood Glucose Monitoring System is intended to quantitatively measure glucose in fresh capillary whole blood from the fingertip, palm, and upper arm as an aid in monitoring the effectiveness of glucose control. The Accu-Chek Guide blood Glucose Monitoring System is intended for in vitro diagnostic single-patient use by people with diabetes. The Accu-Chek Guide Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.
- Where it went
- US Nationwide distribution and countries of AR, AT, AU, BE, BR, CA, CH, CL, DE, DK, EC, ES, FI, FR, GB, GR, HK, HR, ID, IN, IT, JP, KR, LB, LT, LU, MM, MY, NL, NO, PH, PK, PT, SE, SG, TH, TW, ZA.
- How much
- 452,392
Blood Glucose Monitoring System, Model Number 07400926001 ACCU-CHEK GUIDE RETAIL KIT MG/DL - Product Usage: T
Z-0748-2020 · Class II · reported 2020-01-08 · Terminated
- Why it was recalled
- Potential power issues with the Accu-Chek¿ Guide Blood Glucose Monitoring System.
- Product
- Blood Glucose Monitoring System, Model Number 07400926001 ACCU-CHEK GUIDE RETAIL KIT MG/DL - Product Usage: The Accu-Chek Guide Blood Glucose Monitoring System is intended to quantitatively measure glucose in fresh capillary whole blood from the fingertip, palm, and upper arm as an aid in monitoring the effectiveness of glucose control. The Accu-Chek Guide blood Glucose Monitoring System is intended for in vitro diagnostic single-patient use by people with diabetes. The Accu-Chek Guide Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.
- Where it went
- US Nationwide distribution and countries of AR, AT, AU, BE, BR, CA, CH, CL, DE, DK, EC, ES, FI, FR, GB, GR, HK, HR, ID, IN, IT, JP, KR, LB, LT, LU, MM, MY, NL, NO, PH, PK, PT, SE, SG, TH, TW, ZA.
- How much
- 215,259
Blood Glucose Monitoring System, Model Number 07400870001 ACCU-CHEK GUIDE SAMPLE KIT MG/DL - Product Usage: T
Z-0747-2020 · Class II · reported 2020-01-08 · Terminated
- Why it was recalled
- Potential power issues with the Accu-Chek¿ Guide Blood Glucose Monitoring System.
- Product
- Blood Glucose Monitoring System, Model Number 07400870001 ACCU-CHEK GUIDE SAMPLE KIT MG/DL - Product Usage: The Accu-Chek Guide Blood Glucose Monitoring System is intended to quantitatively measure glucose in fresh capillary whole blood from the fingertip, palm, and upper arm as an aid in monitoring the effectiveness of glucose control. The Accu-Chek Guide blood Glucose Monitoring System is intended for in vitro diagnostic single-patient use by people with diabetes. The Accu-Chek Guide Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.
- Where it went
- US Nationwide distribution and countries of AR, AT, AU, BE, BR, CA, CH, CL, DE, DK, EC, ES, FI, FR, GB, GR, HK, HR, ID, IN, IT, JP, KR, LB, LT, LU, MM, MY, NL, NO, PH, PK, PT, SE, SG, TH, TW, ZA.
- How much
- 318,678
Blood Glucose Monitoring System, Model Number 07400942001 ACCU-CHEK GUIDE METER ONLY MG/DL US - Product Usage
Z-0749-2020 · Class II · reported 2020-01-08 · Terminated
- Why it was recalled
- Potential power issues with the Accu-Chek¿ Guide Blood Glucose Monitoring System.
- Product
- Blood Glucose Monitoring System, Model Number 07400942001 ACCU-CHEK GUIDE METER ONLY MG/DL US - Product Usage: The Accu-Chek Guide Blood Glucose Monitoring System is intended to quantitatively measure glucose in fresh capillary whole blood from the fingertip, palm, and upper arm as an aid in monitoring the effectiveness of glucose control. The Accu-Chek Guide blood Glucose Monitoring System is intended for in vitro diagnostic single-patient use by people with diabetes. The Accu-Chek Guide Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.
- Where it went
- US Nationwide distribution and countries of AR, AT, AU, BE, BR, CA, CH, CL, DE, DK, EC, ES, FI, FR, GB, GR, HK, HR, ID, IN, IT, JP, KR, LB, LT, LU, MM, MY, NL, NO, PH, PK, PT, SE, SG, TH, TW, ZA.
- How much
- 21,094
Blood Glucose Monitoring System, Model Number 08453071001 ACCU-CHEK GUIDE RETAIL KIT FCN MG/DL US - Product U
Z-0750-2020 · Class II · reported 2020-01-08 · Terminated
- Why it was recalled
- Potential power issues with the Accu-Chek¿ Guide Blood Glucose Monitoring System.
- Product
- Blood Glucose Monitoring System, Model Number 08453071001 ACCU-CHEK GUIDE RETAIL KIT FCN MG/DL US - Product Usage: The Accu-Chek Guide Blood Glucose Monitoring System is intended to quantitatively measure glucose in fresh capillary whole blood from the fingertip, palm, and upper arm as an aid in monitoring the effectiveness of glucose control. The Accu-Chek Guide blood Glucose Monitoring System is intended for in vitro diagnostic single-patient use by people with diabetes. The Accu-Chek Guide Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.
- Where it went
- US Nationwide distribution and countries of AR, AT, AU, BE, BR, CA, CH, CL, DE, DK, EC, ES, FI, FR, GB, GR, HK, HR, ID, IN, IT, JP, KR, LB, LT, LU, MM, MY, NL, NO, PH, PK, PT, SE, SG, TH, TW, ZA.
- How much
- 3,996
Accu-Chek Aviva II Meter - Product Usage: a. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is inten
Z-2765-2019 · Class II · reported 2019-10-09 · Terminated
- Why it was recalled
- Meters may contain a damaged C29 capacitor resulting in issues with power allegation (short battery life or meter not turning on).
- Product
- Accu-Chek Aviva II Meter - Product Usage: a. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. b. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Accu-Chek Aviva Plus Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady state times (when glucose is not changing rapidly). c. The Accu-Chek Aviva Plus Test Strips are for use with the Accu-Chek Aviva Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm.
- Where it went
- Worldwide distribution - US Nationwide distribution.
- How much
- 26,600
Accu-Chek Performa II Meter - Product Usage: a. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is in
Z-2766-2019 · Class II · reported 2019-10-09 · Terminated
- Why it was recalled
- Meters may contain a damaged C29 capacitor resulting in issues with power allegation (short battery life or meter not turning on).
- Product
- Accu-Chek Performa II Meter - Product Usage: a. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. b. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Accu-Chek Aviva Plus Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady state times (when glucose is not changing rapidly). c. The Accu-Chek Aviva Plus Test Strips are for use with the Accu-Chek Aviva Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm.
- Where it went
- Worldwide distribution - US Nationwide distribution.
- How much
- N/A
Accu-Chek Connect Diabetes Management App
Z-2753-2019 · Class II · reported 2019-10-02 · Terminated
- Why it was recalled
- Users with Android OS 8.0 and above may be unable to transfer values obtained on their meter to the App.
- Product
- Accu-Chek Connect Diabetes Management App
- Where it went
- Nationwide and worldwide usage of the Connect App.
Accu-Chek Performa Test Strip, Blood glucose monitoring system, Catalog no. 07299702001 Product Usage: The
Z-2186-2018 · Class II · reported 2018-06-20 · Terminated
- Why it was recalled
- The recalling firm has identified two lots of test strips that are out of specification. Affected test strips show an increased potential for: strip errors prior to applying a blood sample; strips not detected once inserted; or a biased result may occur, which may not be detectable. In the case of a biased result, incorrect action could be taken by the user in response to the displayed blood glucose result as compared to the actual blood glucose value.
- Product
- Accu-Chek Performa Test Strip, Blood glucose monitoring system, Catalog no. 07299702001 Product Usage: The Accu-Chek Performa test strips are for use with the Accu-Chek Performa meter to quantitatively measure glucose (sugar) in venous whole blood or fresh capillary whole blood samples from the fingertips.
- Where it went
- Worldwide Distribution - US Nationwide distribution. Foreign distribution to Argentina Armenia Australia Austria Azerbaijan Bahamas Bahrain Barbados Belgium Brazil Bulgaria Canada Chile China Colombia Costa Rica Denmark Dominican Republic El Salvador Finland Georgia Germany Ghana Greece Guatemala Honduras Hong Kong (SAR China) Hungary India Indonesia Israel Italy Jamaica Japan Jordan Ireland Kenya Korea Kuwait Lebanon Lithuania Luxembourg Malaysia Malta Mexico Myanmar Netherlands New Zealand Norway Oman Pakistan Philippines Poland Russian Federation New Zealand Saudi Arabia Serbia Singapore Slovenia South Africa Spain Sweden Switzerland Taiwan (Republic of China), Thailand, United Arab Emirates, United Kingdom, Vietnam, Portugal"
- How much
- 295,248
Accu-Chek Aviva Plus Test Strip, Blood glucose monitoring system, Catalog nos. 06908217001, 06908268001, 06908
Z-2185-2018 · Class II · reported 2018-06-20 · Terminated
- Why it was recalled
- The recalling firm identified four US lots of test strips out of specification. Affected strips show increased potential for strip errors prior to applying blood, strips not detected once inserted, or biased results, which may not be detectable. In the case of a biased result, incorrect action could be taken by the user in response to the displayed blood glucose result.
- Product
- Accu-Chek Aviva Plus Test Strip, Blood glucose monitoring system, Catalog nos. 06908217001, 06908268001, 06908349001,06908217001 Product Usage: The Accu-Chek Aviva Plus test strips are for use with the Accu-Chek Aviva Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm.
- Where it went
- Worldwide Distribution - US Nationwide distribution. Foreign distribution to Argentina Armenia Australia Austria Azerbaijan Bahamas Bahrain Barbados Belgium Brazil Bulgaria Canada Chile China Colombia Costa Rica Denmark Dominican Republic El Salvador Finland Georgia Germany Ghana Greece Guatemala Honduras Hong Kong (SAR China) Hungary India Indonesia Israel Italy Jamaica Japan Jordan Ireland Kenya Korea Kuwait Lebanon Lithuania Luxembourg Malaysia Malta Mexico Myanmar Netherlands New Zealand Norway Oman Pakistan Philippines Poland Russian Federation New Zealand Saudi Arabia Serbia Singapore Slovenia South Africa Spain Sweden Switzerland Taiwan (Republic of China), Thailand, United Arab Emirates, United Kingdom, Vietnam, Portugal"
- How much
- 295, 248
Accu-Chek Inform II Test Strip, Whole blood glucose test system, Catalog no. 05942861001 Product Usage: Th
Z-2187-2018 · Class II · reported 2018-06-20 · Terminated
- Why it was recalled
- The recalling firm has identified two lots of test strips that are out of specification. Affected test strips show an increased potential for: strip errors prior to applying a blood sample; strips not detected once inserted; or a biased result may occur, which may not be detectable. In the case of a biased result, incorrect action could be taken by the user in response to the displayed blood glucose result as compared to the actual blood glucose value.
- Product
- Accu-Chek Inform II Test Strip, Whole blood glucose test system, Catalog no. 05942861001 Product Usage: The Accu-Chek Inform II test strips are for use with the Accu-Chek Inform II meter to quantitatively measure glucose (sugar) in venous whole blood, arterial whole blood, neonatal heel stick, or fresh capillary whole blood samples drawn from the fingertips as an aid in monitoring the effectiveness of glucose control. The system is not for use in diagnosis or screening of diabetes mellitus, nor for testing neonate cord blood samples.
- Where it went
- Worldwide Distribution - US Nationwide distribution. Foreign distribution to Argentina Armenia Australia Austria Azerbaijan Bahamas Bahrain Barbados Belgium Brazil Bulgaria Canada Chile China Colombia Costa Rica Denmark Dominican Republic El Salvador Finland Georgia Germany Ghana Greece Guatemala Honduras Hong Kong (SAR China) Hungary India Indonesia Israel Italy Jamaica Japan Jordan Ireland Kenya Korea Kuwait Lebanon Lithuania Luxembourg Malaysia Malta Mexico Myanmar Netherlands New Zealand Norway Oman Pakistan Philippines Poland Russian Federation New Zealand Saudi Arabia Serbia Singapore Slovenia South Africa Spain Sweden Switzerland Taiwan (Republic of China), Thailand, United Arab Emirates, United Kingdom, Vietnam, Portugal"
- How much
- 295,248
Accu-Chek Connect Diabetes Management App, Catalog number 07562462001 / GTIN number 00365702700000 & Catalog n
Z-0625-2018 · Class II · reported 2018-02-21 · Terminated
- Why it was recalled
- Certain iOS and Android App versions contain a program error (bug) in the Bolus Advisor feature. Due to a software bug, when the OS region of the phone setting is changed, the unit of measure within the app may unexpectedly change. This creates a risk the app might not transfer the blood glucose result or the user might not correctly input numerical values for carbohydrate used for bolus advice.
- Product
- Accu-Chek Connect Diabetes Management App, Catalog number 07562462001 / GTIN number 00365702700000 & Catalog number 07250452001 / GTIN number 00365702700017
- Where it went
- US Nationwide
- How much
- 67,040 unique users of version 2.1.0
Accu-Chek Connect diabetes management software app, Catalog number 07562462001 / GTIN number 00365702700000, C
Z-0197-2018 · Class II · reported 2017-12-13 · Terminated
- Why it was recalled
- Certain iOS and Android App versions contain a program error (bug) in the Bolus Advisor feature. After pairing a meter with the app for the first time, a customer may encounter the rare condition in which the countdown timer is not displayed and correction bolus advice is not available for the most recent, valid glucose reading. This same blood glucose value may become available for bolus advice calculation at a later time (countdown timer is displayed).
- Product
- Accu-Chek Connect diabetes management software app, Catalog number 07562462001 / GTIN number 00365702700000, Catalog number 07250452001 / GTIN number 00365702700017 The Accu-Chek Connect Diabetes Management App is indicated as an aid in the treatment of diabetes. The software provides for electronic download of blood glucose meters, manual data entry, storage, display, transfer, and self-managing of blood glucose and other related health indicators which can be shown in report and graphical format. The Accu-Chek Bolus Advisor, as a component of the Accu-Chek Connect Diabetes Management App, is indicated for the management of diabetes by calculating an insulin dose or carbohydrate intake based on user-entered data
- Where it went
- Worldwide Distribution - US (Nationwide) and Canada
- How much
- 1,134 bolus advisor features used
Accu-Chek Connect Diabetes Management App versions 2.0.0, 2.0.1 and 2.1.0 for iOS and Android Product Usage
Z-1899-2017 · Class II · reported 2017-05-10 · Terminated
- Why it was recalled
- A program error (bug) in the Bolus Advisor feature, which could result in incorrect bolus advice and a potential insulin over-delivery.
- Product
- Accu-Chek Connect Diabetes Management App versions 2.0.0, 2.0.1 and 2.1.0 for iOS and Android Product Usage: This is a digital product available for download direct to customers from the Apple App Store and the Google Play Store. The total number of downloads is not available. The number of unique users for each affected version from the date of initial distribution through 03/23/2017 is noted in 7a. However, some users will have used all three affected versions as they upgraded from version to version when the versions became available and many others will have tried the app and discontinued use. In the last 30 days, 02/22/2017 03/23/2017, there have been 16,781 unique users in the US with 264 of those seeking bolus advice and 19,268 unique users in the rest of world (excluding Canada) with 415 of those users seeking bolus advice.
- Where it went
- US Nationwide in the states of: AL, AR, AZ, CA, CO, CT, DE, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV
- How much
- 64,324
Accu-Chek Connect Diabetes Management App
Z-1099-2017 · Class II · reported 2017-02-01 · Terminated
- Why it was recalled
- iOS and Android: Under certain conditions the affected app versions may disregard historical bolus data potentially leading to an incorrect bolus insulin recommendation being provided to the user. iOS only: Pairing and using multiple meters with the Accu-Chek Connect app can under rare circumstances cause the bolus advisor to fail to offer a correction bolus recommendation within the eligible time window following a blood glucose measurement (10 15 minutes). Depending on the individual metabolic situation potentially incorrect bolus advice could lead to serious health consequences such as hypoglycemia. Both software issues may also cause the amount of active insulin displayed during the bolus calculation process to be incorrect and should not be used to manually calculate a bolus.
- Product
- Accu-Chek Connect Diabetes Management App
- Where it went
- Worldwide Distribution - Nationwide Distribution The Accu-Chek Connect App for iOS is distributed directly to end users via the Apple App Store. The Accu-Chek Connect App for Android is distributed directly to end users via the Google Play App Store.
- How much
- 27243
Version 1.2.0 of Accu-Chek Connect Diabetes Management App (iOS) released on July 11, 2016
Z-0586-2017 · Class II · reported 2016-11-23 · Terminated
- Why it was recalled
- Roche Diabetes Care, Inc. initiated a voluntary recall for the version 1.2.0 of Accu-Chek Connect Diabetes Management App (iOS), released on July 11, 2016, due to the software containing a programming error (bug) in the Bolus Advisor feature, which could lead to an incorrect insulin bolus recommendation.
- Product
- Version 1.2.0 of Accu-Chek Connect Diabetes Management App (iOS) released on July 11, 2016
- Where it went
- US Distribution to the states of : Alabama, Arizona, Arkansas, California, Colorado, Connecticut, DC, Delaware, Florida, Georgia, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Virginia, Washington, West Virginia, Wisconsin Foreign: None provided
- How much
- 7,909 downloads of iOS version 1.2.0 of the Accu-Chek Connect Diabetes Management App in the US and 8,775 downloads Internationally
AccuChek Spirit Combo insulin infusion pump sold as part of Accu-Chek Combo System.
Z-1733-2015 · Class II · reported 2015-06-17 · Terminated
- Why it was recalled
- Roche has become aware of an increase of mechanical errors experienced by customers using the Accu-Chek Spirit Combo insulin infusion pump and Accu-Chek Spirit insulin infusion pump when the insulin pumps display E6 and El0 error messages. If the user does not act upon the E6 and E10 error messages appropriately, insulin delivery will be stopped and if unnoticed may lead to severe hyperglycemia u
- Product
- AccuChek Spirit Combo insulin infusion pump sold as part of Accu-Chek Combo System.
- Where it went
- Nationwide Distribution
- How much
- 35,795
AccuChek Spirit insulin infusion pump.
Z-1734-2015 · Class II · reported 2015-06-17 · Terminated
- Why it was recalled
- Roche has become aware of an increase of mechanical errors experienced by customers using the Accu-Chek Spirit Combo insulin infusion pump and Accu-Chek Spirit insulin infusion pump when the insulin pumps display E6 and El0 error messages. If the user does not act upon the E6 and E10 error messages appropriately, insulin delivery will be stopped and if unnoticed may lead to severe hyperglycemia u
- Product
- AccuChek Spirit insulin infusion pump.
- Where it went
- Nationwide Distribution
- How much
- 35,795
ACCU-CHEK Connect Diabetes Management App; Instruction Manual Designed to transfer data for diabetes manage
Z-1369-2015 · Class II · reported 2015-04-08 · Terminated
- Why it was recalled
- Roche Diabetes Care has become aware of an issue with the Accu-Chek Connect diabetes management app that could potentially lead to inaccurate bolus advice being provided to the user. A thorough investigation of the situation revealed that this issue may occur if the user changes the screen orientation of the phone from portrait to landscape or vise versa while looking at the Bolus Advisor or Carb
- Product
- ACCU-CHEK Connect Diabetes Management App; Instruction Manual Designed to transfer data for diabetes management.
- Where it went
- There are no US distributors as the affected device is not marketed in the US. The affected device was distributed in Germany, Italy and South Africa. The application has been downloaded total of 644 times (Italy-219, South Africa-24, Germany-401) and there are 113 bolus advice activations.
- How much
- The application has been downloaded total of 644 times (Italy-219, South Africa-24, Germany-401) and there are 113 bolus advice activations.
Read these counts with three things in mind
A firm with more recalls usually makes more products. A contract packer runs
hundreds of lines. It will out-recall a company with one product whatever either
does. This is not a safety ranking.
Most recalls here are voluntary. A firm found its own fault and pulled the stock.
That is the system working. It lands in this database looking the same as a recall
the FDA had to push for.
Terminated means the FDA considers the recall finished. It is not a verdict.
It does not mean anyone was harmed.
Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.
← All device recalls