Philips Medical Systems, Inc.

Philips Medical Systems, Inc.: 65 FDA recall reports

Every medical device recall the FDA has published for Philips Medical Systems, Inc.. The fault, and how the agency graded it.

65Reports
1Class I
0Ongoing
2016-04-06Most recent
1 Class I on this page

Class I means the FDA judged there was a reasonable probability of serious harm or death. Check the lot numbers against anything you hold.

What the classes mean

The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.

ClassWhat the FDA is saying
Class IThere is a reasonable probability that using this will cause serious harm or death.
Class IIIt may cause temporary or reversible harm. Serious harm is unlikely.
Class IIIIt breaks a rule but is unlikely to hurt anyone. Often a labelling fault.

Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.

By class

ClassReports
Class II64
Class I1

By year reported

YearReports
20131
201435
201527
20162
Checking whether it affects you

The lot or code number is what matters. Not the product name. A recall covers specific lots. Yours may not be one of them.

Check the code on your package against the report. The FDA keeps the current list at fda.gov/safety/recalls.

The reports

The text below is quoted from the FDA record word for word.

Foot Switches used with Philips MultiDiagnost Eleva with Flat Detector; Product Codes: 708037, 708038 This Ph

Z-1279-2016  ·  Class II  ·  reported 2016-04-06  ·  Terminated

Why it was recalled
Loss of key image functionality due to a bent pedal of the Footswitch.
Product
Foot Switches used with Philips MultiDiagnost Eleva with Flat Detector; Product Codes: 708037, 708038 This Philips MultiDiagnost Eleva is intended to be used as a multifunctional/universal system. General RIF, Fluoroscopy, Radiography and Angiography can be performed along with more specialized interventional applications. This includes the following general areas: Digestive system, Skeletal system, Reproductive system, Respiratory system, Circulatory system, Various:Arthrograms, Myelograms, Facet joint injections, Discography, Sialography
Where it went
Worldwide Distribution - US (nationwide) and Internationally to: CZECH REPUBLIC ALBANIA ALGERIA ARGENTINA AUSTRALIA AUSTRIA AZARBAIJAN BAHAMAS BANGLADESH BELGIUM BRAZIL BULGARIA CAYMAN ISLANDS CHILE CHINA COLUMBIA COSTA RICA CROATIA CUBA CZECHOSLOVAKIA DENMARK DOMINICAN REPUB ECUADOR EGYPT ESTONIA FINLAND FRANCE GABON GERMANY GREECE HONG KONG HUNGARY ICELAND INDIA INDONESIA IRAN IRAQ IRELAND ISRAEL ITALY JAPAN JORDAN KAZAKHSTAN KENYA KOREA SOUTH KUWAIT LATVIA LEBANON LIBYA LITHUANIA MACEDONIA MALAYSIA MAURITIUS MEXICO MOLDAVIA MONGOLIA MOROCCO NEPAL NETHERLANDS NETHERLANDS ANT NEW ZEALAND NORWAY OMAN PAKISTAN PALESTINE PANAMA Papua New Guinea PARAGUAY PERU PHILIPPINES POLAND PORTUGAL PUORTO RICO QATAR REUNION ROMANIA RUSSIAN FEDERAT SAUDI ARABIA SERBIA SINGAPORE SLOVENIA SOUTH AFRICA SPAIN SUDAN SWEDEN SWITZERLAND SYRIA TAIWAN TANZANIA THAILAND TUNISIA TURKEY UKRAINE UNITED ARAB EMI United Arab Emirates UNITED KINGDOM URUGUAY USA UZBEKISTAN VENEZUELA VIETNAM YEMEN GUADELOUPE SYRIAN ARAB REPUBLIC
How much
168

Foot Switches used with the following systems: Philips Allura Xper Systems; 722001 AlluraXperFDlO C, 722002 Al

Z-1280-2016  ·  Class II  ·  reported 2016-04-06  ·  Terminated

Why it was recalled
Loss of key image functionality due to a bent pedal of the Footswitch.
Product
Foot Switches used with the following systems: Philips Allura Xper Systems; 722001 AlluraXperFDlO C, 722002 Allura Xper FD1O F, 722003 Allura Xper FD1O, 722005 Allura Xper FD1O/10, 722006 Allura Xper FD2O, 722008 Allura Xper FD2O Biplane, 722010 Allura Xper FD1O, 722011 Allura Xper FD1O/10, 722012 Allura Xper FD2O, 722013 Allura XPER FD2O BIPLANE, 722014 Allura Xper FD1O OR Table, 722015 Allura Xper FD2O OR Table, 722019 Allura Xper FD1O/10 OR Table, 722020 Allura Xper FD2O Biplane OR Table, 722022 Allura Xper FD1O OR Table, 722023 Allura Xper FD2O OR Table, 722024 Allura Xper FD1O/10 OR Table, 722025 Allura Xper FD2O Biplane OR Table, 722026 Allura Xper FD1O, 722027 Allura Xper FD1O/10, 722028 Allura Xper FD2O, 722029 Allura Xper FD2O/10, 722031 Allura CV2O, 722033 Allura Xper FD1O OR Table, 722034 Allura Xper FD1O/10 OR Table, 722035 Allura Xper FD2O OR Table, 722036 Allura Xper FD2O/10 OR Table, 722038 Allura Xper FD2O/20, 722039 Allura Xper FD2O/20 OR Table, 722058 Allura Xper FD20115, 722059 Allura Xper FD2O/15 OR Table, 722123 Field extensions Xper cardio systems, 722124 Field extensions Xper vascular systems, 722126 SmartPath to AlluraClarity cardio, 722127 SmartPath to...
Where it went
Worldwide Distribution - US (nationwide) and Internationally to: CZECH REPUBLIC ALBANIA ALGERIA ARGENTINA AUSTRALIA AUSTRIA AZARBAIJAN BAHAMAS BANGLADESH BELGIUM BRAZIL BULGARIA CAYMAN ISLANDS CHILE CHINA COLUMBIA COSTA RICA CROATIA CUBA CZECHOSLOVAKIA DENMARK DOMINICAN REPUB ECUADOR EGYPT ESTONIA FINLAND FRANCE GABON GERMANY GREECE HONG KONG HUNGARY ICELAND INDIA INDONESIA IRAN IRAQ IRELAND ISRAEL ITALY JAPAN JORDAN KAZAKHSTAN KENYA KOREA SOUTH KUWAIT LATVIA LEBANON LIBYA LITHUANIA MACEDONIA MALAYSIA MAURITIUS MEXICO MOLDAVIA MONGOLIA MOROCCO NEPAL NETHERLANDS NETHERLANDS ANT NEW ZEALAND NORWAY OMAN PAKISTAN PALESTINE PANAMA Papua New Guinea PARAGUAY PERU PHILIPPINES POLAND PORTUGAL PUORTO RICO QATAR REUNION ROMANIA RUSSIAN FEDERAT SAUDI ARABIA SERBIA SINGAPORE SLOVENIA SOUTH AFRICA SPAIN SUDAN SWEDEN SWITZERLAND SYRIA TAIWAN TANZANIA THAILAND TUNISIA TURKEY UKRAINE UNITED ARAB EMI United Arab Emirates UNITED KINGDOM URUGUAY USA UZBEKISTAN VENEZUELA VIETNAM YEMEN GUADELOUPE SYRIAN ARAB REPUBLIC
How much
4180 devices

Philips IntelliVue Information Center iX (release A.00, A.01, and A.02) are impacted by this issue: 866023 In

Z-0374-2016  ·  Class II  ·  reported 2015-12-16  ·  Terminated

Why it was recalled
Sp02 and/or Non Invasive Blood Pressure (NBP) alarms may become disabled without visual notification
Product
Philips IntelliVue Information Center iX (release A.00, A.01, and A.02) are impacted by this issue: 866023 IntelliVue Info Center iX 866024 PIIC iX Upgrade 866117 PIIC Classic Upgrade The intended use of the Information Center Software is to display physiologic waves, parameters, and trends, format data for strip chart recordings and printed reports, and provide the secondary annunciation of alarms from other networked medical devices at a centralized location. An additional intended use of the Information Center Software is to provide primary annunciation of alarms and configuration and control access for networked telemetry monitors.
Where it went
Worldwide Distribution: US (nationwide) and countries: AUSTRALIA, AUSTRIA, BELGIUM, BULGARIA, CANADA, DENMARK, ESTONIA, FINLAND, FRANCE, GERMANY, HUNGARY, ICELAND, INDIA, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, LUXEMBOURG, MOROCCO, NETHERLANDS, NEW ZEALAND, NIGERIA, NORWAY, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIAN FEDERATION, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, UNITED ARAB EMIRATES, and UNITED KINGDOM.
How much
10,013 units

Philips Healthcare INTEGRIS vascular System Model : 722122 Product Usage: The Allura Xper series is int

Z-0349-2016  ·  Class II  ·  reported 2015-12-09  ·  Terminated

Why it was recalled
Monitor Ceiling Suspension system may fall
Product
Philips Healthcare INTEGRIS vascular System Model : 722122 Product Usage: The Allura Xper series is intended for use in cardiovascular and vascular X-ray imaging applications including diagnostics, interventional procedures (such as PTCA, stent placement, embolization and thrombolysis and atherectomics), pacemaker Implantations and electrophysiology (EP). Use of the Allura Xper series system in the OR environment is restricted; the system should only be used for endo-vascular procedures. The Philips Integris Allura 9F is intended for diagnostic cardiovascular, vascular and interventional procedures
Where it went
Worldwide Distribution - US Nationwide and the countries of: Canada,Australia, Austria, Brazil, Chile, China, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, South Africa, Spain, Switzerland and United Kingdom.
How much
3 units

Philips Healthcare Xper vascular systems R7.6 Model : 722134 Product Usage: The Allura Xper series is in

Z-0354-2016  ·  Class II  ·  reported 2015-12-09  ·  Terminated

Why it was recalled
Monitor Ceiling Suspension system may fall
Product
Philips Healthcare Xper vascular systems R7.6 Model : 722134 Product Usage: The Allura Xper series is intended for use in cardiovascular and vascular X-ray imaging applications including diagnostics, interventional procedures (such as PTCA, stent placement, embolization and thrombolysis and atherectomics), pacemaker Implantations and electrophysiology (EP). Use of the Allura Xper series system in the OR environment is restricted; the system should only be used for endo-vascular procedures. The Philips Integris Allura 9F is intended for diagnostic cardiovascular, vascular and interventional procedures
Where it went
Worldwide Distribution - US Nationwide and the countries of: Canada,Australia, Austria, Brazil, Chile, China, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, South Africa, Spain, Switzerland and United Kingdom.
How much
22 units

Philips Healthcare INTEGRIS H5000F/Allura 9F Model : 722017 Product Usage: The Allura Xper series is in

Z-0350-2016  ·  Class II  ·  reported 2015-12-09  ·  Terminated

Why it was recalled
Monitor Ceiling Suspension system may fall
Product
Philips Healthcare INTEGRIS H5000F/Allura 9F Model : 722017 Product Usage: The Allura Xper series is intended for use in cardiovascular and vascular X-ray imaging applications including diagnostics, interventional procedures (such as PTCA, stent placement, embolization and thrombolysis and atherectomics), pacemaker Implantations and electrophysiology (EP). Use of the Allura Xper series system in the OR environment is restricted; the system should only be used for endo-vascular procedures. The Philips Integris Allura 9F is intended for diagnostic cardiovascular, vascular and interventional procedures
Where it went
Worldwide Distribution - US Nationwide and the countries of: Canada,Australia, Austria, Brazil, Chile, China, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, South Africa, Spain, Switzerland and United Kingdom.
How much
1

Philips Healthcare Allura Xper Series: Allura Xper FD10 Allura Xper FD10/10 Allura Xper FD20 Allura Xper F

Z-0346-2016  ·  Class II  ·  reported 2015-12-09  ·  Terminated

Why it was recalled
Monitor Ceiling Suspension system may fall
Product
Philips Healthcare Allura Xper Series: Allura Xper FD10 Allura Xper FD10/10 Allura Xper FD20 Allura Xper FD21 Allura Xper FD22 Allura Xper FD23 Allura Xper FD24 Allura Xper FD25 Product Usage: The Allura Xper series is intended for use in cardiovascular and vascular X-ray imaging applications including diagnostics, interventional procedures (such as PTCA, stent placement, embolization and thrombolysis and atherectomics), pacemaker Implantations and electrophysiology (EP). Use of the Allura Xper series system in the OR environment is restricted; the system should only be used for endo-vascular procedures. The Philips Integris Allura 9F is intended for diagnostic cardiovascular, vascular and interventional procedures
Where it went
Worldwide Distribution - US Nationwide and the countries of: Canada,Australia, Austria, Brazil, Chile, China, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, South Africa, Spain, Switzerland and United Kingdom.
How much
9 units

Philips Healthcare INTEGRIS cardio system Model : 722121 Product Usage: The Allura Xper series is inten

Z-0348-2016  ·  Class II  ·  reported 2015-12-09  ·  Terminated

Why it was recalled
Monitor Ceiling Suspension system may fall
Product
Philips Healthcare INTEGRIS cardio system Model : 722121 Product Usage: The Allura Xper series is intended for use in cardiovascular and vascular X-ray imaging applications including diagnostics, interventional procedures (such as PTCA, stent placement, embolization and thrombolysis and atherectomics), pacemaker Implantations and electrophysiology (EP). Use of the Allura Xper series system in the OR environment is restricted; the system should only be used for endo-vascular procedures. The Philips Integris Allura 9F is intended for diagnostic cardiovascular, vascular and interventional procedures
Where it went
Worldwide Distribution - US Nationwide and the countries of: Canada,Australia, Austria, Brazil, Chile, China, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, South Africa, Spain, Switzerland and United Kingdom.
How much
8 units

Patient Support Material Number: 9890 010 87431 used with: BuckyDiagnost ; Model: 704060/706037

Z-0364-2016  ·  Class II  ·  reported 2015-12-09  ·  Terminated

Why it was recalled
The hook does not securely hold the footplate in vertical position. Therefore the footplate can fall or swing down and may hurt the operator or patient.
Product
Patient Support Material Number: 9890 010 87431 used with: BuckyDiagnost ; Model: 704060/706037
Where it went
Distributed in the states of MN, KY, MA, VA, TX, NC, HI, FL, AZ, CT, and WY and the countries of Argentina, Australia, Austria, Canada, Denmark, Germany, Netherlands, Russia, Saudi Arabia, Singapore, Spain, and Switzerland.
How much
1517 systems, not all units have the patient support accessory

Patient Support Material Number: 9890 010 87431, used with:EasyDiagnost ; Model: 706050

Z-0365-2016  ·  Class II  ·  reported 2015-12-09  ·  Terminated

Why it was recalled
The hook does not securely hold the footplate in vertical position. Therefore the footplate can fall or swing down and may hurt the operator or patient.
Product
Patient Support Material Number: 9890 010 87431, used with:EasyDiagnost ; Model: 706050
Where it went
Distributed in the states of MN, KY, MA, VA, TX, NC, HI, FL, AZ, CT, and WY and the countries of Argentina, Australia, Austria, Canada, Denmark, Germany, Netherlands, Russia, Saudi Arabia, Singapore, Spain, and Switzerland.
How much
260 units, not all units have the patient support accessory

Philips Healthcare Xper vascular system Model : 722124 Product Usage: The Allura Xper series is intend

Z-0352-2016  ·  Class II  ·  reported 2015-12-09  ·  Terminated

Why it was recalled
Monitor Ceiling Suspension system may fall
Product
Philips Healthcare Xper vascular system Model : 722124 Product Usage: The Allura Xper series is intended for use in cardiovascular and vascular X-ray imaging applications including diagnostics, interventional procedures (such as PTCA, stent placement, embolization and thrombolysis and atherectomics), pacemaker Implantations and electrophysiology (EP). Use of the Allura Xper series system in the OR environment is restricted; the system should only be used for endo-vascular procedures. The Philips Integris Allura 9F is intended for diagnostic cardiovascular, vascular and interventional procedures
Where it went
Worldwide Distribution - US Nationwide and the countries of: Canada,Australia, Austria, Brazil, Chile, China, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, South Africa, Spain, Switzerland and United Kingdom.
How much
10 units

Philips Healthcare Xper cardio systems R7.6 Model : 722133 Product Usage: The Allura Xper series is inte

Z-0353-2016  ·  Class II  ·  reported 2015-12-09  ·  Terminated

Why it was recalled
Monitor Ceiling Suspension system may fall
Product
Philips Healthcare Xper cardio systems R7.6 Model : 722133 Product Usage: The Allura Xper series is intended for use in cardiovascular and vascular X-ray imaging applications including diagnostics, interventional procedures (such as PTCA, stent placement, embolization and thrombolysis and atherectomics), pacemaker Implantations and electrophysiology (EP). Use of the Allura Xper series system in the OR environment is restricted; the system should only be used for endo-vascular procedures. The Philips Integris Allura 9F is intended for diagnostic cardiovascular, vascular and interventional procedures
Where it went
Worldwide Distribution - US Nationwide and the countries of: Canada,Australia, Austria, Brazil, Chile, China, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, South Africa, Spain, Switzerland and United Kingdom.
How much
14 units

Philips Healthcare Xper cardio systems Model : 722123 Product Usage: The Allura Xper series is intended

Z-0351-2016  ·  Class II  ·  reported 2015-12-09  ·  Terminated

Why it was recalled
Monitor Ceiling Suspension system may fall
Product
Philips Healthcare Xper cardio systems Model : 722123 Product Usage: The Allura Xper series is intended for use in cardiovascular and vascular X-ray imaging applications including diagnostics, interventional procedures (such as PTCA, stent placement, embolization and thrombolysis and atherectomics), pacemaker Implantations and electrophysiology (EP). Use of the Allura Xper series system in the OR environment is restricted; the system should only be used for endo-vascular procedures. The Philips Integris Allura 9F is intended for diagnostic cardiovascular, vascular and interventional procedures
Where it went
Worldwide Distribution - US Nationwide and the countries of: Canada,Australia, Austria, Brazil, Chile, China, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, South Africa, Spain, Switzerland and United Kingdom.
How much
16 units

Philips Healthcare Allure BiPlane Series: Model 722013 Allura Xper FD20 BIPLANE R7.6 722013 All

Z-0347-2016  ·  Class II  ·  reported 2015-12-09  ·  Terminated

Why it was recalled
Monitor Ceiling Suspension system may fall
Product
Philips Healthcare Allure BiPlane Series: Model 722013 Allura Xper FD20 BIPLANE R7.6 722013 Allura Xper FD20 BIPLANE R7.7 722013 Allura Xper FD20 BIPLANE R7.8 722013 Allura Xper FD20 BIPLANE R7.9 722013 Allura Xper FD20 BIPLANE R7.10 Product Usage: The Allura Xper series is intended for use in cardiovascular and vascular X-ray imaging applications including diagnostics, interventional procedures (such as PTCA, stent placement, embolization and thrombolysis and atherectomics), pacemaker Implantations and electrophysiology (EP). Use of the Allura Xper series system in the OR environment is restricted; the system should only be used for endo-vascular procedures. The Philips Integris Allura 9F is intended for diagnostic cardiovascular, vascular and interventional procedures
Where it went
Worldwide Distribution - US Nationwide and the countries of: Canada,Australia, Austria, Brazil, Chile, China, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, South Africa, Spain, Switzerland and United Kingdom.
How much
5 units

Ingenia, Intera, Achieva and Achieve dStream MR systems on Software versions R5.1.7 and R5.1.8 with the Mobi

Z-0325-2016  ·  Class II  ·  reported 2015-12-02  ·  Terminated

Why it was recalled
When a fused series of a sagittal, coronal or radial multi station scan is generated in MobiView, the resultant image order is reversed. This occurs with software release R5.1.7/R5.1.8.
Product
Ingenia, Intera, Achieva and Achieve dStream MR systems on Software versions R5.1.7 and R5.1.8 with the Mobi View software option. Diagnostic imaging system.
Where it went
Worldwide Distribution - USA (nationwide) and to the countries of: Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, Denmark Estonia, Ethiopia. France, Germany, Hong Kong, India, Indonesia, Iraq, Ireland, Italy, Japan, Jordan, Korea, Republic of, Libya, Malaysia, Mexico, Nepal, Netherlands, New Zealand, Norway, Oman, Palestinian territory, Peru, Poland, Russian Federation, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates and United Kingdom.
How much
307 units

Philips HeartStart MRx Monitor/Defibrillator Models:M3535A, M3536A, M3536M, M3536MC, M3536M4, M3536M5 The He

Z-0262-2016  ·  Class II  ·  reported 2015-11-18  ·  Terminated

Why it was recalled
MRx Defib can be susceptible to one or both issues: 1. The C02 Inlet Port associated with end-tidal carbon dioxide (EtCO2) monitoring on MRx Monitor/Defibrillators can be pushed into the MRx housing, making it inaccessible. 2. The handle can separate from the MRx housing due to breakage of mounts on the rear case.
Product
Philips HeartStart MRx Monitor/Defibrillator Models:M3535A, M3536A, M3536M, M3536MC, M3536M4, M3536M5 The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician
Where it went
Worldwide Distribution - US (nationwide) and Internationally to ARGENTINA, ARUBA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELGIUM, BOLIVIA, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, CROATIA (local name: Hrvatska), CZECH REPUBLIC, DENMARK, EGYPT, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, GUADELOUPE, HONG KONG, ICELAND, INDIA, INDONESIA, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KENYA, KOREA, REPUBLIC OF, KUWAIT, LATVIA, LEBANON, LITHUANIA, MALAYSIA, MEXICO, NETHERLANDS, NETHERLAND ANTILLES, NEW CALEDONIA, NEW ZEALAND, NORWAY, OMAN, PHILIPPINES, POLAND, PORTUGAL, PUERTO RICO, QATAR, REUNION, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SOUTH AFRICA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, TAIWAN, PROVINCE OF CHINA, THAILAND, TRINIDAD AND TOBAGO, TUNISIA, TURKEY, UGANDA, UNITED ARAB EMIRATES, UNITED KINGDOM, VENEZUELA, AND YEMEN.
How much
27,918 units

Philips HeartStart MRx Monitor/Defibrillators Models: M3535A and M3536A

Z-0204-2016  ·  Class II  ·  reported 2015-11-11  ·  Terminated

Why it was recalled
1. Device will perform the weekly automated tests hourly, which could cause the therapy capacitors to degrade sooner than intended and 2. While connected to AC or DC power and with no battery installed or the battery installed has a charge level of less than 10%, the Ready for Use (RFU) indicator will not provide the expected low battery indication
Product
Philips HeartStart MRx Monitor/Defibrillators Models: M3535A and M3536A
Where it went
Worldwide Distribution - USA (nationwide), Australia, Austria, Bahrain, Belgium, Brunei Darussalam, Canada, China, Finland, France, Germany, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Republic of Korea, Libyan Arab Jamahiriya, Lithuania, Mexico, Namibia, Netherlands, Palestine, Philippines, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Taiwan, United Arab Emirates and United Kingdom.
How much
1553 units

HeartStart MRx monitor/defibrillator Models M3535A, M3536A, M3536J, M3536M, M3536MC, M3536M2, M3536M4, M3536M

Z-0219-2016  ·  Class II  ·  reported 2015-11-11  ·  Terminated

Why it was recalled
MRx monitor/defibrillator could reboot at an indeterminate time, potentially causing therapy to be interrupted or delayed
Product
HeartStart MRx monitor/defibrillator Models M3535A, M3536A, M3536J, M3536M, M3536MC, M3536M2, M3536M4, M3536M5, M3536M6 Product Usage: The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician
Where it went
Worldwide Distribution - US Nationwide and the countries of AFGHANISTAN, ALBANIA, ARGENTINA, ARUBA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BANGLADESH, BELGIUM, BOLIVIA, BOSNIA AND HERZEGOWINA, BRAZIL, BRUNEI DARUSSALAM, BULGARIA, CAMBODIA, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA (local name: Hrvatska), CYPRUS, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, EL SALVADOR, ESTONIA, ETHIOPIA, FINLAND, FRANCE, GERMANY, GHANA, GREECE, GUADELOUPE, HONG KONG, ICELAND, INDIA, INDONESIA, IRAN (ISLAMIC REPUBLIC OF), IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, Jordan, KAZAKHSTAN, KENYA, REPUBLIC OF KOREA, KUWAIT, LAOS, LATVIA, LEBANON, LIBYAN ARAB JAMAHIRIYA, LITHUANIA, LUXEMBOURG, MACAU, MACEDONIA, MALAYSIA, MALTA, MEXICO, MONGOLIA, MOROCCO, MOZAMBIQUE, MYANMAR (Burma), NAMIBIA, NEPAL, NETHERLANDS, NETHERLANDS ANTILLES, NEW CALEDONIA, NEW ZEALAND, NICARAGUA, NIGERIA, NORWAY, OMAN, PAKISTAN, PALESTINE, PANAMA, PAPUA NEW GUINEA, PARAGUAY, PERU, PHILIPPINES, POLAND, PORTUGAL, PUERTO RICO, QATAR, REUNION, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, TAIWAN, UNITED REPUBLIC OF TANZANIA, THAILAND, TRINIDAD AND TOBAGO, TUNISIA,...
How much
75,693 units

All Philips Ingenia, Intera, Achieva and Multiva MR systems using R5.1i and R5.1 .2 version of software. Indi

Z-0135-2016  ·  Class II  ·  reported 2015-10-28  ·  Terminated

Why it was recalled
In spine clinical workflows, cross reference lines may be used to determine the position of slices. In cases, where MobiView fused Images are used to show the cross reference lines, the cross reference lines may be positioned incorrectly.
Product
All Philips Ingenia, Intera, Achieva and Multiva MR systems using R5.1i and R5.1 .2 version of software. Indicated for use as a diagnostic device. It can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body, or extremities.
Where it went
Worldwide Distribution-US: AL AR AZ CA CO FL GA IL IN KS KY MA MI NY OH OR PA SC TX VA VT WA OUS: Argentina Australia Austria Bangladesh Belgium Brazil Canada Chile China Colombia Denmark Ecuador Estonia Ethiopia France Germany India Indonesia Iraq Ireland Italy Japan Jordan Korea, South Malaysia Mexico Netherlands New Zealand Norway Oman Palestine Poland Russian Federation South Africa Spain Sweden Switzerland Taiwan Turkey United Arab Emirates United Kingdom.
How much
317

Philips Healthcare DigitalDiagnost System X-Ray

Z-1554-2015  ·  Class II  ·  reported 2015-05-27  ·  Terminated

Why it was recalled
The system is designed to emit a beep upon termination of an exposure. However, if the system has been powered on for over 12 hours, the system will no longer emit this signal. This is a failure to comply with 21CFR 1020.31(j).
Product
Philips Healthcare DigitalDiagnost System X-Ray
Where it went
Worldwide Distribution - Nationwide Distribution and the countries of Australia, Austria, Argentina, Canada, China, Chile, Colombia Czech Republic, Great Britain, Germany, Egypt, Finland, France, India, Ireland, Japan, Malaysia, Macedonia, Mexico, Netherlands, New Zealand, Nepal, Peru, Poland, Russia, Saudi Arabia Singapore, Switzerland, Sweden, Spain, United Arab Emirates and United Kingdom.
How much
54 DigitalDiagnost

Pinnacle3 Software Version 10.0, Model 459800200841. Product Usage: The Pinnacle 3 RTP software allows qu

Z-1643-2015  ·  Class II  ·  reported 2015-05-27  ·  Terminated

Why it was recalled
A problem (the dose computed in planning mode is calculated incorrectly ) has been detected in the Philips Pinnacle3 software version 10.0 that, if it were to re-occur, could pose a risk for patients or users. Specifically, the dose engine is being passed the wrong snout position. The snout position is used to determine the penumbra of the beam which includes the calculation of the source si
Product
Pinnacle3 Software Version 10.0, Model 459800200841. Product Usage: The Pinnacle 3 RTP software allows qualified medical professionals to enter patient data into the system, use that data to construct a plan for radiation therapy and evaluate the plan. Optionally, the qualified medical personnel may output the plan in an electronic or printed form for use by other systems in the delivery of treatment to a patient.
Where it went
US Nationwide Distribution in the states of MO, FL, OH
How much
4

Philips Healthcare DuraDiagnost X- Ray

Z-1555-2015  ·  Class II  ·  reported 2015-05-27  ·  Terminated

Why it was recalled
The system is designed to emit a beep upon termination of an exposure. However, if the system has been powered on for over 12 hours, the system will no longer emit this signal. This is a failure to comply with 21CFR 1020.31(j).
Product
Philips Healthcare DuraDiagnost X- Ray
Where it went
Worldwide Distribution - Nationwide Distribution and the countries of Australia, Austria, Argentina, Canada, China, Chile, Colombia Czech Republic, Great Britain, Germany, Egypt, Finland, France, India, Ireland, Japan, Malaysia, Macedonia, Mexico, Netherlands, New Zealand, Nepal, Peru, Poland, Russia, Saudi Arabia Singapore, Switzerland, Sweden, Spain, United Arab Emirates and United Kingdom.
How much
3 DuraDiagnost

Lithium Polymer Fiber Optic Module Batteries used in the Intera MR System upgraded to CDAS, Part Number: 9065

Z-1307-2015  ·  Class II  ·  reported 2015-04-01  ·  Terminated

Why it was recalled
VCG battery ignited in a VCG unit when recharging.
Product
Lithium Polymer Fiber Optic Module Batteries used in the Intera MR System upgraded to CDAS, Part Number: 9065 and 9067. These rechargeable batteries are used to supply power to the following products: Wireless ECG Module used with the Precess MRI compatible Patient Monitoring System, Wireless SpO2 Module used with the Precess MRI Compatible Patient Monitoring System, Philips Basic Triggering Set, Philips Monitoring Sensor Set, Philips Battery Set
Where it went
Distributed in the following states: AL, CA, FL, GA, IL, MA, MN, NC, ND, NE, NJ, NY, OH, OR, PA, TN, TX, UT, WA, WI; OUS: Australia, Austria, Belgium, China, Ecuador, Finland, France, Germany, Italy, Japan, Mexico, Netherlands, slovenia, South Africa, Sweden, Switzerland, United Kingdom
How much
79

Philips DuraDiagnost X-Ray system, Stationary X-Ray System Radiographic Models: 712210, 712211

Z-1251-2015  ·  Class II  ·  reported 2015-03-18  ·  Terminated

Why it was recalled
Improper installation of Tube arm, and Wall stand may cause it to fall down, Improper installation of table may cause it to float.
Product
Philips DuraDiagnost X-Ray system, Stationary X-Ray System Radiographic Models: 712210, 712211
Where it went
Worldwide Distribution. Argentina, Australia, Burkina-Faso, Chile, China, Colombia, Ecuador, France, Germany, Hungary, Indonesia, Iraq (NL), Kuwait, Libya, Macedonia, Malaysia, Mayotte, Poland, Saudi Arabia, Slovenia, Spain, Switzerland, Turkey, United Kingdom, and Uzbekistan.
How much
71 units

Philips Medical System Allura Xper X-Ray Angiographic

Z-1120-2015  ·  Class II  ·  reported 2015-03-11  ·  Terminated

Why it was recalled
In certain circumstances, a software error can lead to a situation where the five minute fluoroscopy audible signal does not sound.
Product
Philips Medical System Allura Xper X-Ray Angiographic
Where it went
Worldwide Distribution - USA nationwide including Puerto Rico, Austria, Azerbaijan, Australia, Algeria, Argentina, Albania, Brazil, Belgium, Bangladesh, Bahamas, Bulgaria, Cambodia, Chile, Canada, China, Croatia, Czech Republic, Netherlands, New Zealand, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Reunion, Singapore, Slovenia, South Africa, Spain, Sudan, Sweden, Saudi Arabia, Switzerland, Syria, Taiwan, Thailand, France, Dominican Republic, Colombia, Turkey, Germany, Greece, Hong Kong, Denmark, Ecuador, Egypt, Estonia, Finland, Vietnam, Russian Federation, Iran, Iraq, Ireland, Israel, Italy, India, Indonesia, Japan, Jordan, Kenya, South Korea, Kuwait, Latvia, Lebanon, Lithuania, Macedonia, Malaysia, Mauritius, Mexico, Moldavia, Mongolia, Morocco, Nepal, Ukraine, United Arab Emirates, and United Kingdom
How much
7439

Philips Multi Diagnost Eleva II

Z-1044-2015  ·  Class II  ·  reported 2015-03-04  ·  Terminated

Why it was recalled
The possibility of the Reference Air Kerma Rate for low dose mode will not be reduced to 50% of the value for the normal mode, but will be approximately 70-80%, leading to a higher patient dose than expected.
Product
Philips Multi Diagnost Eleva II
Where it went
Worldwide Distribution - USA ( nationwide ) to the states of : AR, AZ, CA, CO, DE, FL, GA, HI, IL, IA, KY, KS, LA, MI, MO, MS, NC, NY, OH, PA, TN, TX, VA, VT, WA, WY and WV., and the countries of Austria, Australia, Colombia, Germany, Switzerland, Turkey, Ireland, Dubia and United Kingdom.
How much
238

Philips Multi Diagnost Eleva with FlatDetector

Z-1045-2015  ·  Class II  ·  reported 2015-03-04  ·  Terminated

Why it was recalled
The possibility of the Reference Air Kerma Rate for low dose mode will not be reduced to 50% of the value for the normal mode, but will be approximately 70-80%, leading to a higher patient dose than expected.
Product
Philips Multi Diagnost Eleva with FlatDetector
Where it went
Worldwide Distribution - USA ( nationwide ) to the states of : AR, AZ, CA, CO, DE, FL, GA, HI, IL, IA, KY, KS, LA, MI, MO, MS, NC, NY, OH, PA, TN, TX, VA, VT, WA, WY and WV., and the countries of Austria, Australia, Colombia, Germany, Switzerland, Turkey, Ireland, Dubia and United Kingdom.
How much
238

Philips IntelliVue TcG10; Monitor, carbon-dioxide, cutaneous. The IntelliVue TcG10 is a device for the meas

Z-0997-2015  ·  Class II  ·  reported 2015-01-28  ·  Terminated

Why it was recalled
Philips has discovered that the Instructions for Use (IFU) for the IntelliVue TcG10 Transcutaneous Gas Measurement Module is missing the contraindication that the device is not to be used on patients under gas anesthesia.
Product
Philips IntelliVue TcG10; Monitor, carbon-dioxide, cutaneous. The IntelliVue TcG10 is a device for the measurement of the transcutaneous O2 and CO2 partial pressure in neonates, pediatrics and adults.
Where it went
Distributed in the states of CA, HI, IL, MA, MI, MO, MT, NC, OH, and PA.
How much
86

HeartStart MRx Monitor/Defibrillator; Model. Numbers: M3535A, M3536A, M3536J, M3536M, M3536MC, M3536M2, M3536M

Z-0963-2015  ·  Class II  ·  reported 2015-01-21  ·  Terminated

Why it was recalled
Device may malfunction, which could cause therapy to be delayed, disabled, or delivered inadvertently.
Product
HeartStart MRx Monitor/Defibrillator; Model. Numbers: M3535A, M3536A, M3536J, M3536M, M3536MC, M3536M2, M3536M4, M3536M5, M3536M6 Product Usage: The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician.
Where it went
Worldwide Distribution - US Nationwide
How much
74,499 devices

Philips Brilliance iCT Computed Tomography X-Ray System Product Usage: The Brilliance iCT/Brilliance iCT S

Z-0236-2015  ·  Class II  ·  reported 2014-12-24  ·  Terminated

Why it was recalled
It was discovered that a software defect may result in the scanner not terminating the CT scan at the intended location.
Product
Philips Brilliance iCT Computed Tomography X-Ray System Product Usage: The Brilliance iCT/Brilliance iCT SP is a Computed Tomography X-Ray System intended to produce crosssectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. (d) A description of the defect in the
Where it went
US Distribution
How much
16 systems

Showing the 30 most recent of 65. The rest are searchable at the FDA enforcement report database.

Read these counts with three things in mind

A firm with more recalls usually makes more products. A contract packer runs hundreds of lines. It will out-recall a company with one product whatever either does. This is not a safety ranking.

Most recalls here are voluntary. A firm found its own fault and pulled the stock. That is the system working. It lands in this database looking the same as a recall the FDA had to push for.

Terminated means the FDA considers the recall finished. It is not a verdict. It does not mean anyone was harmed.