Philips Medical Systems (Cleveland) In

Philips Medical Systems (Cleveland) Inc: 315 FDA recall reports

Every medical device recall the FDA has published for Philips Medical Systems (Cleveland) Inc. The fault, and how the agency graded it.

315Reports
5Class I
4Ongoing
2025-09-24Most recent
5 Class I on this page

Class I means the FDA judged there was a reasonable probability of serious harm or death. Check the lot numbers against anything you hold.

What the classes mean

The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.

ClassWhat the FDA is saying
Class IThere is a reasonable probability that using this will cause serious harm or death.
Class IIIt may cause temporary or reversible harm. Serious harm is unlikely.
Class IIIIt breaks a rule but is unlikely to hurt anyone. Often a labelling fault.

Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.

By class

ClassReports
Class II302
Class III8
Class I5

By year reported

YearReports
201212
201326
201426
201585
201664
201715
201862
201922
20221
20252
Checking whether it affects you

The lot or code number is what matters. Not the product name. A recall covers specific lots. Yours may not be one of them.

Check the code on your package against the report. The FDA keeps the current list at fda.gov/safety/recalls.

The reports

The text below is quoted from the FDA record word for word.

Pinnacle Radiation Therapy Planning System: Multimodality Simulation Workspace 18.0.5 Model 870258 UDI cod

Z-2570-2025  ·  Class II  ·  reported 2025-09-24  ·  Ongoing

Why it was recalled
Due to a software issue, there is a potential image error of the Region of Interest for expansion/contraction for HFP (Head First Prone), FFS (Feet First Supine) and FFP (Feet First Prone) orientations.
Product
Pinnacle Radiation Therapy Planning System: Multimodality Simulation Workspace 18.0.5 Model 870258 UDI code: (01)00884838103566(10)18.0.5.22236 Pinnacle TumorLOC 16.2 Model 870226 UDI code: (01)00884838091122(10)16.2.0.60010 Pinnacle TumorLOC 16.2.1 Model 870227 UDI code: (01)00884838091122(10)16.2.1.60098 Pinnacle TumorLOC 16.0.2 Model 870218 UDI code: (01)00884838085602(10)16.0.2.60004 Pinnacle TumorLOC 14.0 Model 870200 UDI code: N/A Pinnacle3 Radiation Therapy Planning System is a software package intended to provide planning support for the treatment of disease processes. The full Pinnacle3 Radiation Therapy Planning System software package provides planning support for the treatment of disease processes, utilizing photon, electron and brachytherapy techniques.
Where it went
U.S Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV.
How much
810 (350 U.S. and 460 O.U.S.)

Pinnacle 3 with TumorLOC, (870258) Radiation Therapy Planning System Software

Z-2239-2025  ·  Class II  ·  reported 2025-08-13  ·  Ongoing

Why it was recalled
Due to software issue, Radiation Therapy Planning system may provide incorrect dataset calculations when performing the "Stopping Power Ratio" (SPR) ,
Product
Pinnacle 3 with TumorLOC, (870258) Radiation Therapy Planning System Software
Where it went
Worldwide - US Nationwide distribution in the state of WV and the countries of Macao & United Kingdom.
How much
149 Serial Numbers

Pinnacle3 Radiation Therapy Planning System, Model numbers 870231 and 870237.

Z-1744-2022  ·  Class II  ·  reported 2022-09-21  ·  Ongoing

Why it was recalled
When computing a radiation dose in the system, the exported dose information is incorrect when there is more than one beam attached to the prescription and certain options are selected on the Edit Prescription screen.
Product
Pinnacle3 Radiation Therapy Planning System, Model numbers 870231 and 870237.
Where it went
Distribution was made to CA, FL, GA, IA, IL, MA, MD, ME, MI, MN, NJ, NM, NV, NY, OH, OK, RI, SC, TN, TX, and WA. There was government distribution and no military distribution. Foreign distribution was made to Australia, Austria, Belgium, China, France, Germany, Italy, Japan, Jordan, Lebanon, Poland, Portugal, Romania, Russian Federation, Singapore, Spain, Switzerland, Taiwan, Thailand, and United Kingdom.
How much
169 systems
Class I

Forte Gamma Camera System is intended to produce images depicting the anatomical distribution of single-photon

Z-0217-2020  ·  Class I  ·  reported 2019-11-13  ·  Terminated

Why it was recalled
An issue with the Detector for the Forte Family of cameras may result in either detector 1 or detector 2 falling unimpeded vertically to the end stops of its travel limit.
Product
Forte Gamma Camera System is intended to produce images depicting the anatomical distribution of single-photon and positron-emitting radioisotopes within the body for interpretation by medical personnel.
Where it went
Nationwide Distribution.
How much
1167 in total
Class I

Forte Gamma Camera System is intended to produce images depicting the anatomical distribution of single-photon

Z-0216-2020  ·  Class I  ·  reported 2019-11-13  ·  Terminated

Why it was recalled
An issue with the Detector for the Forte Family of cameras may result in either detector 1 or detector 2 falling unimpeded vertically to the end stops of its travel limit.
Product
Forte Gamma Camera System is intended to produce images depicting the anatomical distribution of single-photon and positron-emitting radioisotopes within the body for interpretation by medical personnel.
Where it went
Nationwide Distribution.
How much
1167 in total
Class I

Forte Gamma Camera System is intended to produce images depicting the anatomical distribution of single-photon

Z-0213-2020  ·  Class I  ·  reported 2019-11-13  ·  Terminated

Why it was recalled
An issue with the Detector for the Forte Family of cameras may result in either detector 1 or detector 2 falling unimpeded vertically to the end stops of its travel limit.
Product
Forte Gamma Camera System is intended to produce images depicting the anatomical distribution of single-photon and positron-emitting radioisotopes within the body for interpretation by medical personnel.
Where it went
Nationwide Distribution.
How much
1167 in total
Class I

Forte Gamma Camera System is intended to produce images depicting the anatomical distribution of single-photon

Z-0215-2020  ·  Class I  ·  reported 2019-11-13  ·  Terminated

Why it was recalled
An issue with the Detector for the Forte Family of cameras may result in either detector 1 or detector 2 falling unimpeded vertically to the end stops of its travel limit.
Product
Forte Gamma Camera System is intended to produce images depicting the anatomical distribution of single-photon and positron-emitting radioisotopes within the body for interpretation by medical personnel.
Where it went
Nationwide Distribution.
How much
1167 in total
Class I

Forte Gamma Camera System is intended to produce images depicting the anatomical distribution of single-photon

Z-0214-2020  ·  Class I  ·  reported 2019-11-13  ·  Terminated

Why it was recalled
An issue with the Detector for the Forte Family of cameras may result in either detector 1 or detector 2 falling unimpeded vertically to the end stops of its travel limit.
Product
Forte Gamma Camera System is intended to produce images depicting the anatomical distribution of single-photon and positron-emitting radioisotopes within the body for interpretation by medical personnel.
Where it went
Nationwide Distribution.
How much
1167 in total

Ingenuity Core128 Model # 728323, computed tomography x-ray system

Z-2187-2019  ·  Class II  ·  reported 2019-08-14  ·  Terminated

Why it was recalled
In certain instances when performing a cardiac step and shoot acquisition with phase tolerance selected, images may be labeled as if phase tolerance had been applied when phase tolerance had not been applied
Product
Ingenuity Core128 Model # 728323, computed tomography x-ray system
Where it went
Worldwide distribution. US nationwide including Puerto Rico, Canada, Mexico, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Cote D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Djibouti, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, French Polynesia, Gabon, Georgia, Germany, Ghana, Gibraltar, Greece, Greenland, Guyana, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malawi, Malaysia, Maldives, Mali, Malta, Martinique, Mauritius, Monaco, Mongolia, Morocco, Mozambique, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, State of Palestine, Panama, Peru, Philippines, Poland, Qatar, Reunion, Romania, Russian Federation, Rwanda, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden,...
How much
116

Brilliance 64 Model # 728231, computed tomography x-ray system

Z-2191-2019  ·  Class II  ·  reported 2019-08-14  ·  Terminated

Why it was recalled
In certain instances when performing a cardiac step and shoot acquisition with phase tolerance selected, images may be labeled as if phase tolerance had been applied when phase tolerance had not been applied
Product
Brilliance 64 Model # 728231, computed tomography x-ray system
Where it went
Worldwide distribution. US nationwide including Puerto Rico, Canada, Mexico, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Cote D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Djibouti, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, French Polynesia, Gabon, Georgia, Germany, Ghana, Gibraltar, Greece, Greenland, Guyana, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malawi, Malaysia, Maldives, Mali, Malta, Martinique, Mauritius, Monaco, Mongolia, Morocco, Mozambique, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, State of Palestine, Panama, Peru, Philippines, Poland, Qatar, Reunion, Romania, Russian Federation, Rwanda, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden,...
How much
582

iCT, Model # 728306, computed tomography x-ray system

Z-2189-2019  ·  Class II  ·  reported 2019-08-14  ·  Terminated

Why it was recalled
In certain instances when performing a cardiac step and shoot acquisition with phase tolerance selected, images may be labeled as if phase tolerance had been applied when phase tolerance had not been applied
Product
iCT, Model # 728306, computed tomography x-ray system
Where it went
Worldwide distribution. US nationwide including Puerto Rico, Canada, Mexico, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Cote D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Djibouti, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, French Polynesia, Gabon, Georgia, Germany, Ghana, Gibraltar, Greece, Greenland, Guyana, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malawi, Malaysia, Maldives, Mali, Malta, Martinique, Mauritius, Monaco, Mongolia, Morocco, Mozambique, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, State of Palestine, Panama, Peru, Philippines, Poland, Qatar, Reunion, Romania, Russian Federation, Rwanda, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden,...
How much
128

Ingenuity CT, Model # 728326, computed tomography x-ray system

Z-2186-2019  ·  Class II  ·  reported 2019-08-14  ·  Terminated

Why it was recalled
In certain instances when performing a cardiac step and shoot acquisition with phase tolerance selected, images may be labeled as if phase tolerance had been applied when phase tolerance had not been applied
Product
Ingenuity CT, Model # 728326, computed tomography x-ray system
Where it went
Worldwide distribution. US nationwide including Puerto Rico, Canada, Mexico, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Cote D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Djibouti, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, French Polynesia, Gabon, Georgia, Germany, Ghana, Gibraltar, Greece, Greenland, Guyana, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malawi, Malaysia, Maldives, Mali, Malta, Martinique, Mauritius, Monaco, Mongolia, Morocco, Mozambique, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, State of Palestine, Panama, Peru, Philippines, Poland, Qatar, Reunion, Romania, Russian Federation, Rwanda, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden,...
How much
92

Ingenuity TF PET/CT Model # 882456, computed tomography x-ray system

Z-2193-2019  ·  Class II  ·  reported 2019-08-14  ·  Terminated

Why it was recalled
In certain instances when performing a cardiac step and shoot acquisition with phase tolerance selected, images may be labeled as if phase tolerance had been applied when phase tolerance had not been applied
Product
Ingenuity TF PET/CT Model # 882456, computed tomography x-ray system
Where it went
Worldwide distribution. US nationwide including Puerto Rico, Canada, Mexico, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Cote D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Djibouti, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, French Polynesia, Gabon, Georgia, Germany, Ghana, Gibraltar, Greece, Greenland, Guyana, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malawi, Malaysia, Maldives, Mali, Malta, Martinique, Mauritius, Monaco, Mongolia, Morocco, Mozambique, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, State of Palestine, Panama, Peru, Philippines, Poland, Qatar, Reunion, Romania, Russian Federation, Rwanda, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden,...
How much
12

iCT SP Model # 728311, computed tomography x-ray system

Z-2190-2019  ·  Class II  ·  reported 2019-08-14  ·  Terminated

Why it was recalled
In certain instances when performing a cardiac step and shoot acquisition with phase tolerance selected, images may be labeled as if phase tolerance had been applied when phase tolerance had not been applied
Product
iCT SP Model # 728311, computed tomography x-ray system
Where it went
Worldwide distribution. US nationwide including Puerto Rico, Canada, Mexico, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Cote D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Djibouti, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, French Polynesia, Gabon, Georgia, Germany, Ghana, Gibraltar, Greece, Greenland, Guyana, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malawi, Malaysia, Maldives, Mali, Malta, Martinique, Mauritius, Monaco, Mongolia, Morocco, Mozambique, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, State of Palestine, Panama, Peru, Philippines, Poland, Qatar, Reunion, Romania, Russian Federation, Rwanda, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden,...
How much
27

Ingenuity Core Model # 728321, computed tomography x-ray system

Z-2188-2019  ·  Class II  ·  reported 2019-08-14  ·  Terminated

Why it was recalled
In certain instances when performing a cardiac step and shoot acquisition with phase tolerance selected, images may be labeled as if phase tolerance had been applied when phase tolerance had not been applied
Product
Ingenuity Core Model # 728321, computed tomography x-ray system
Where it went
Worldwide distribution. US nationwide including Puerto Rico, Canada, Mexico, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Cote D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Djibouti, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, French Polynesia, Gabon, Georgia, Germany, Ghana, Gibraltar, Greece, Greenland, Guyana, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malawi, Malaysia, Maldives, Mali, Malta, Martinique, Mauritius, Monaco, Mongolia, Morocco, Mozambique, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, State of Palestine, Panama, Peru, Philippines, Poland, Qatar, Reunion, Romania, Russian Federation, Rwanda, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden,...
How much
202

Vereos PET/CT Model # 882446, computed tomography x-ray system

Z-2194-2019  ·  Class II  ·  reported 2019-08-14  ·  Terminated

Why it was recalled
In certain instances when performing a cardiac step and shoot acquisition with phase tolerance selected, images may be labeled as if phase tolerance had been applied when phase tolerance had not been applied
Product
Vereos PET/CT Model # 882446, computed tomography x-ray system
Where it went
Worldwide distribution. US nationwide including Puerto Rico, Canada, Mexico, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Cote D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Djibouti, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, French Polynesia, Gabon, Georgia, Germany, Ghana, Gibraltar, Greece, Greenland, Guyana, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malawi, Malaysia, Maldives, Mali, Malta, Martinique, Mauritius, Monaco, Mongolia, Morocco, Mozambique, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, State of Palestine, Panama, Peru, Philippines, Poland, Qatar, Reunion, Romania, Russian Federation, Rwanda, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden,...
How much
26

IQon Spectral CT Model # 728332, computed tomography x-ray system

Z-2192-2019  ·  Class II  ·  reported 2019-08-14  ·  Terminated

Why it was recalled
In certain instances when performing a cardiac step and shoot acquisition with phase tolerance selected, images may be labeled as if phase tolerance had been applied when phase tolerance had not been applied
Product
IQon Spectral CT Model # 728332, computed tomography x-ray system
Where it went
Worldwide distribution. US nationwide including Puerto Rico, Canada, Mexico, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Cote D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Djibouti, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, French Polynesia, Gabon, Georgia, Germany, Ghana, Gibraltar, Greece, Greenland, Guyana, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malawi, Malaysia, Maldives, Mali, Malta, Martinique, Mauritius, Monaco, Mongolia, Morocco, Mozambique, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, State of Palestine, Panama, Peru, Philippines, Poland, Qatar, Reunion, Romania, Russian Federation, Rwanda, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden,...
How much
25

BrightView XCT , Model Number 882482

Z-2011-2019  ·  Class II  ·  reported 2019-07-31  ·  Terminated

Why it was recalled
An issue with the handcontroller may result in either spontaneous uncommanded motions or continued motion after the buttons were released.
Product
BrightView XCT , Model Number 882482
Where it went
Worldwide Distribution - US Nationwide
How much
427

BrightView SPECT, Model Number 882480

Z-2009-2019  ·  Class II  ·  reported 2019-07-31  ·  Terminated

Why it was recalled
An issue with the handcontroller may result in either spontaneous uncommanded motions or continued motion after the buttons were released.
Product
BrightView SPECT, Model Number 882480
Where it went
Worldwide Distribution - US Nationwide
How much
696

BrightView X , Model Number 882478

Z-2010-2019  ·  Class II  ·  reported 2019-07-31  ·  Terminated

Why it was recalled
An issue with the handcontroller may result in either spontaneous uncommanded motions or continued motion after the buttons were released.
Product
BrightView X , Model Number 882478
Where it went
Worldwide Distribution - US Nationwide
How much
117

Vereos PET/CT, Model No. 882446 Product Usage: Positron Emission Tomography (PET) and X-ray Computed Tomog

Z-1929-2019  ·  Class II  ·  reported 2019-07-10  ·  Terminated

Why it was recalled
The Vereos PET/CT may have misaligned front and rear covers, which may result in gaps between cover parts and/or protrusion of the cover into the bore diameter.
Product
Vereos PET/CT, Model No. 882446 Product Usage: Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The CT subsystem images anatomical cross-sections by computer reconstruction of X-ray transmission data. The PET subsystem images the distribution of PET anatomy-specific radiopharmaceuticals in the patient. The PET/CT system is used for the purposes of detecting, localizing, diagnosing, staging, re-staging, and follow-up for monitoring therapy response of various diseases in oncology, cardiology, and neurology. The system is intended to image the whole body, heart, brain, lung, gastrointestinal, bone, lymphatic, and other major organs for a wide range of patient types, sizes, and extent of diseases. Both subsystems can also be operated as fully functional, independent diagnostic tools, including application of the CT scanner for diagnosis and for use in radiation therapy planning.
Where it went
Worldwide Distribution - US Nationwide Domestic distribution to CA GA LA MD MI MN MO NE NJ NY OH TN TX VT WI. Foreign distribution to Mexico, Argentina Austria Belgium China Colombia France Germany Israel Italy Japan Latvia Luxembourg Netherlands Portugal Singapore Spain Switzerland
How much
70

Brilliance 16 Air, 728246 Product Usage: Computed Tomography X-ray systems intended to produce cross-secti

Z-1142-2019  ·  Class II  ·  reported 2019-04-24  ·  Terminated

Why it was recalled
Philips has identified that the compensator within the X-ray tube collimator may fail, resulting in ring or smudge artifacts on images. The firm has identified systems that may be at risk for this failure due to the collimator compensator being past its lifetime expected usage.
Product
Brilliance 16 Air, 728246 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories
Where it went
Distributed only to China and UK.
How much
23

Brilliance iCT, 728306 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectiona

Z-1144-2019  ·  Class II  ·  reported 2019-04-24  ·  Terminated

Why it was recalled
Philips has identified that the compensator within the X-ray tube collimator may fail, resulting in ring or smudge artifacts on images. The firm has identified systems that may be at risk for this failure due to the collimator compensator being past its lifetime expected usage.
Product
Brilliance iCT, 728306 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories
Where it went
Distributed only to China and UK.
How much
2

Brilliance 64, 728231 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional

Z-1143-2019  ·  Class II  ·  reported 2019-04-24  ·  Terminated

Why it was recalled
Philips has identified that the compensator within the X-ray tube collimator may fail, resulting in ring or smudge artifacts on images. The firm has identified systems that may be at risk for this failure due to the collimator compensator being past its lifetime expected usage.
Product
Brilliance 64, 728231 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories
Where it went
Distributed only to China and UK.
How much
14

Ingenuity TF PET/CT (non-RoHS), Model Number 882456 This device is a diagnostic imaging system that combine

Z-0823-2019  ·  Class II  ·  reported 2019-02-20  ·  Terminated

Why it was recalled
A software update is being issued to correct multiple issues identified in the previous software version.
Product
Ingenuity TF PET/CT (non-RoHS), Model Number 882456 This device is a diagnostic imaging system that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The CT subsystem produces cross-sectional images of the body by computer reconstruction of x-ray transmission data. The PET subsystem produces images of the distribution of PET radiopharmaceuticals in the patient body (specific radiopharmaceuticals are used for whole body, brain, heart and other organ imaging).
Where it went
Worldwide distribution: US (nationwide) to states of: CA, IL, NY, OH, and PA; and to countries of: Australia, Austria, Columbia, Egypt, France, Germany, India, Italy, Lithuania, Netherlands, and Switzerland.
How much
24

Brilliance 6 Slice (Air), Model 728256 Product Usage: Computed Tomography X-ray systems intended to produc

Z-0548-2019  ·  Class II  ·  reported 2018-12-12  ·  Terminated

Why it was recalled
There is a potential for separation of fiberglass resin which may cause the gantry front cover and the metal support on the cover to fail/fall
Product
Brilliance 6 Slice (Air), Model 728256 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories
Where it went
US Nationwide Distribution
How much
6,343 units in total

Brilliance 40, Model 728235 Product Usage: Computed Tomography X-ray systems intended to produce cross-sec

Z-0544-2019  ·  Class II  ·  reported 2018-12-12  ·  Terminated

Why it was recalled
There is a potential for separation of fiberglass resin which may cause the gantry front cover and the metal support on the cover to fail/fall
Product
Brilliance 40, Model 728235 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories
Where it went
US Nationwide Distribution
How much
6,343 units in total

Brilliance 16 Slice (Power), Model 728240 Product Usage: Computed Tomography X-ray systems intended to pro

Z-0545-2019  ·  Class II  ·  reported 2018-12-12  ·  Terminated

Why it was recalled
There is a potential for separation of fiberglass resin which may cause the gantry front cover and the metal support on the cover to fail/fall
Product
Brilliance 16 Slice (Power), Model 728240 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories
Where it went
US Nationwide Distribution
How much
6,343 units in total

Brilliance 64, Model 728231 Product Usage: Computed Tomography X-ray systems intended to produce cross-se

Z-0543-2019  ·  Class II  ·  reported 2018-12-12  ·  Terminated

Why it was recalled
There is a potential for separation of fiberglass resin which may cause the gantry front cover and the metal support on the cover to fail/fall
Product
Brilliance 64, Model 728231 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories
Where it went
US Nationwide Distribution
How much
6,343 units in total

Brilliance 10 Slice (Air), Model 728251 Product Usage: Computed Tomography X-ray systems intended to produ

Z-0547-2019  ·  Class II  ·  reported 2018-12-12  ·  Terminated

Why it was recalled
There is a potential for separation of fiberglass resin which may cause the gantry front cover and the metal support on the cover to fail/fall
Product
Brilliance 10 Slice (Air), Model 728251 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories
Where it went
US Nationwide Distribution
How much
N/A

Showing the 30 most recent of 315. The rest are searchable at the FDA enforcement report database.

Read these counts with three things in mind

A firm with more recalls usually makes more products. A contract packer runs hundreds of lines. It will out-recall a company with one product whatever either does. This is not a safety ranking.

Most recalls here are voluntary. A firm found its own fault and pulled the stock. That is the system working. It lands in this database looking the same as a recall the FDA had to push for.

Terminated means the FDA considers the recall finished. It is not a verdict. It does not mean anyone was harmed.

Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.

← All device recalls