Pentax Medical Company

Pentax Medical Company: 26 FDA recall reports

Every medical device recall the FDA has published for Pentax Medical Company. The fault, and how the agency graded it.

26Reports
1Class I
0Ongoing
2016-08-03Most recent
1 Class I on this page

Class I means the FDA judged there was a reasonable probability of serious harm or death. Check the lot numbers against anything you hold.

What the classes mean

The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.

ClassWhat the FDA is saying
Class IThere is a reasonable probability that using this will cause serious harm or death.
Class IIIt may cause temporary or reversible harm. Serious harm is unlikely.
Class IIIIt breaks a rule but is unlikely to hurt anyone. Often a labelling fault.

Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.

By class

ClassReports
Class II25
Class I1

By year reported

YearReports
20131
20142
201623
Checking whether it affects you

The lot or code number is what matters. Not the product name. A recall covers specific lots. Yours may not be one of them.

Check the code on your package against the report. The FDA keeps the current list at fda.gov/safety/recalls.

The reports

The text below is quoted from the FDA record word for word.

Signmoidoscope Product Usage: Intended to provide optical visualization of, therapeutic access to the Lowe

Z-2280-2016  ·  Class II  ·  reported 2016-08-03  ·  Terminated

Why it was recalled
Pentax Medical Video Processor EPK-i5010 has been shipped with a reusable Pentax Medical water bottle assembly (OS-H4) which has not been cleared for use with this devices in the United States.
Product
Signmoidoscope Product Usage: Intended to provide optical visualization of, therapeutic access to the Lower Gastrointestinal Tract. This anatomy includes, but is not restricted to, the organs; tissues; and subsystems: Large Bowel to the Sigmoid Colon. The instrument is introduced via the rectum when indications consistent with the need for procedure are observed in adult and pediatric patient populations.
Where it went
US Nationwide Distribution
How much
N/A

Colonoscope Product Usage: Intended to provide optical visualization of and therapeutic access to the Lowe

Z-2275-2016  ·  Class II  ·  reported 2016-08-03  ·  Terminated

Why it was recalled
Pentax Medical Video Processor EPK-i5010 has been shipped with a reusable Pentax Medical water bottle assembly (OS-H4) which has not been cleared for use with this devices in the United States.
Product
Colonoscope Product Usage: Intended to provide optical visualization of and therapeutic access to the Lower Gastrointestinal Tract. This anatomy includes, but is not restricted to, the organs; tissues; and subsystems: Large Bowel to the Cecum. The instrument is introduced via the rectum when indications consistent with the need for procedure are observed in adult and pediatric patient populations.
Where it went
US Nationwide Distribution
How much
12,412 units

Ultrasound Gastroscope Product Usage: Intended to provide optical visualization of, ultrasonic Visualizati

Z-2279-2016  ·  Class II  ·  reported 2016-08-03  ·  Terminated

Why it was recalled
Pentax Medical Video Processor EPK-i5010 has been shipped with a reusable Pentax Medical water bottle assembly (OS-H4) which has not been cleared for use with this devices in the United States.
Product
Ultrasound Gastroscope Product Usage: Intended to provide optical visualization of, ultrasonic Visualization of, and therapeutic access to the Upper Gastrointestinal Tract including but not restricted to, the organs; tissues; and subsystems: Esophagus, Stomach, Duodenum, Small Bowel and underlying areas. The instrument is introduced via per orally when indications consistent with the requirement for procedure are observed in adult and pediatric patient populations.
Where it went
US Nationwide Distribution
How much
819 units

Gastroscope Product Usage: Intended to provide optical visualization of and therapeutic access to the Uppe

Z-2278-2016  ·  Class II  ·  reported 2016-08-03  ·  Terminated

Why it was recalled
Pentax Medical Video Processor EPK-i5010 has been shipped with a reusable Pentax Medical water bottle assembly (OS-H4) which has not been cleared for use with this devices in the United States.
Product
Gastroscope Product Usage: Intended to provide optical visualization of and therapeutic access to the Upper Gastrointestinal Tract. This anatomy includes, but is not restricted to, the organs; tissues; and subsystems: Esophagus, Stomach and Duodenum. The instrument is introduced via the mouth when indications consistent with the need for procedure are observed in adult and pediatric patient populations.
Where it went
US Nationwide Distribution
How much
10,074 units

Duodenoscope Product Usage: Intended to provide optical visualization of and therapeutic access to the Gas

Z-2277-2016  ·  Class II  ·  reported 2016-08-03  ·  Terminated

Why it was recalled
Pentax Medical Video Processor EPK-i5010 has been shipped with a reusable Pentax Medical water bottle assembly (OS-H4) which has not been cleared for use with this devices in the United States.
Product
Duodenoscope Product Usage: Intended to provide optical visualization of and therapeutic access to the Gastrointestinal Tract. This anatomy includes, but is not restricted to, the organs; tissues; and subsystems: Esophagus, Stomach, Duodenum, and Large Bowel to the Cecum. The instrument is introduced via the mouth or rectum as decided by the the anatomy being accessed when indications consistent with the need for procedure are observed in adult and pediatric patient populations.
Where it went
US Nationwide Distribution
How much
433 units

Confocal GI Scope Product Usage: Intended to provide optical visualization of and therapeutic access to th

Z-2276-2016  ·  Class II  ·  reported 2016-08-03  ·  Terminated

Why it was recalled
Pentax Medical Video Processor EPK-i5010 has been shipped with a reusable Pentax Medical water bottle assembly (OS-H4) which has not been cleared for use with this devices in the United States.
Product
Confocal GI Scope Product Usage: Intended to provide optical visualization of and therapeutic access to the Gastrointestinal Tract. This anatomy includes, but is not restricted to, the organs; tissues; and subsystems: Esophagus, Stomach, Duodenum, and Large Bowel to the Cecum. The instrument is introduced via the mouth or rectum as decided by the the anatomy being accessed when indications consistent with the need for procedure are observed in adult and pediatric patient populations.
Where it went
US Nationwide Distribution
How much
17 units

Bronchoscope Intended to provide optical visualization of and therapeutic access to the Airway and Bronchial

Z-2065-2016  ·  Class II  ·  reported 2016-07-06  ·  Terminated

Why it was recalled
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
Product
Bronchoscope Intended to provide optical visualization of and therapeutic access to the Airway and Bronchial Tree. This anatomy includes, but is not restricted to, the organs, tissues and subsystems: Nasal Passage, Trachea and Bronchial Tree. The instrument is introduced via the mouth or the nose, as decided by the physician, when indications consistent with the need for procedure are observed in adult and pediatric patient populations.
Where it went
Worldwide Distribution - US (Nation Wide) including Puerto Rico and Internationally to Denmark, Guam
How much
4,195 units

Colonoscope Intended to provide optical visualization of and therapeutic access to the Lower Gastrointestinal

Z-2068-2016  ·  Class II  ·  reported 2016-07-06  ·  Terminated

Why it was recalled
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
Product
Colonoscope Intended to provide optical visualization of and therapeutic access to the Lower Gastrointestinal Tract. This anatomy includes, but is not restricted to, the organs, tissues and subsystems: Large Bowel to the Cecum. The instrument is introduced via the rectum when indications consistent with the need for procedure are observed in adult and pediatric patient populations.
Where it went
Worldwide Distribution - US (Nation Wide) including Puerto Rico and Internationally to Denmark, Guam
How much
12,412 units

Intubation Scope Intended to provide optical visualization of, therapeutic access to, and use to facilitate r

Z-2072-2016  ·  Class II  ·  reported 2016-07-06  ·  Terminated

Why it was recalled
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
Product
Intubation Scope Intended to provide optical visualization of, therapeutic access to, and use to facilitate routine or difficult tracheal intubations of the Upper Airway. This anatomy includes, but is not restricted to, the organs, tissues and subsystems: Larynx and Trachea. The instrument is introduced via the mouth or the nose, as decided by the physician, when indications consistent with the need for procedure are observed in adult and pediatric patient populations.
Where it went
Worldwide Distribution - US (Nation Wide) including Puerto Rico and Internationally to Denmark, Guam
How much
1,220 units

Small Bowel Scope Gastrointestinal Tract. This anatomy includes, but is not restricted to, the organs, tissue

Z-2080-2016  ·  Class II  ·  reported 2016-07-06  ·  Terminated

Why it was recalled
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
Product
Small Bowel Scope Gastrointestinal Tract. This anatomy includes, but is not restricted to, the organs, tissues and subsystems: Esophagus, Stomach and Duodenum. The instrument is introduced via the mouth when indications consistent with the need for procedure are observed in adult and pediatric patient populations.
Where it went
Worldwide Distribution - US (Nation Wide) including Puerto Rico and Internationally to Denmark, Guam
How much
96 units

Laryngo Stroboscope Product Usage Intended to provide optical visualization of, therapeutic access to, the Up

Z-2073-2016  ·  Class II  ·  reported 2016-07-06  ·  Terminated

Why it was recalled
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
Product
Laryngo Stroboscope Product Usage Intended to provide optical visualization of, therapeutic access to, the Upper Airway and Vocal Folds. This anatomy includes, but is not restricted to, the organs, tissues and subsystems: Larynx, Pharynx and Vocal Cords. The instrument is introduced via the mouth or the nose, as decided by the physician, when indications consistent with the need for procedure are observed in adult and pediatric patient populations.
Where it went
Worldwide Distribution - US (Nation Wide) including Puerto Rico and Internationally to Denmark, Guam
How much
206 units

Signmoidoscope Intended to provide optical visualization of, therapeutic access to, the Lower Gastrointestina

Z-2075-2016  ·  Class II  ·  reported 2016-07-06  ·  Terminated

Why it was recalled
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
Product
Signmoidoscope Intended to provide optical visualization of, therapeutic access to, the Lower Gastrointestinal Tract. This anatomy includes, but is not restricted to, the organs, tissues and subsystems: Large Bowel to the Sigmoid Colon. The instrument is introduced via the rectum when indications consistent with the need for procedure are observed in adult and pediatric patient populations
Where it went
Worldwide Distribution - US (Nation Wide) including Puerto Rico and Internationally to Denmark, Guam
How much
1,596 units

Ultrasound Bronchoscope Intended to provide optical visualization of, ultrasonic visualization of and therape

Z-2076-2016  ·  Class II  ·  reported 2016-07-06  ·  Terminated

Why it was recalled
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
Product
Ultrasound Bronchoscope Intended to provide optical visualization of, ultrasonic visualization of and therapeutic access to, the Pulmonary Tract including but not restricted to organs, tissues and subsystems: Nasal Passage, Pharynx, Larynx, Trachea, Bronchial Tree (including access beyond the stem), and underlying areas. The instrument is introduced orally when indications consistent with the requirement for procedure are observed in adult and pediatric patient populations
Where it went
Worldwide Distribution - US (Nation Wide) including Puerto Rico and Internationally to Denmark, Guam
How much
115 units

Ultrasound Gastroscope Intended to provide optical visualization of, ultrasonic visualization of and therapeu

Z-2077-2016  ·  Class II  ·  reported 2016-07-06  ·  Terminated

Why it was recalled
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
Product
Ultrasound Gastroscope Intended to provide optical visualization of, ultrasonic visualization of and therapeutic access to, the Upper Gastrointestinal Track including but not restricted to organs, tissues and subsystems: Esophagus, Stomach, Duodenum, Small Bowel and underlying areas. The instrument is introduced orally when indications consistent with the requirement for procedure are observed in adult and pediatric
Where it went
Worldwide Distribution - US (Nation Wide) including Puerto Rico and Internationally to Denmark, Guam
How much
819 units

Esophagoscope Gastrointestinal Tract. This anatomy includes, but is not restricted to, the organs, tissues an

Z-2079-2016  ·  Class II  ·  reported 2016-07-06  ·  Terminated

Why it was recalled
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
Product
Esophagoscope Gastrointestinal Tract. This anatomy includes, but is not restricted to, the organs, tissues and subsystems: Esophagus, Stomach and Duodenum. The instrument is introduced via the mouth when indications consistent with the need for procedure are observed in adult and pediatric patient populations.
Where it went
Worldwide Distribution - US (Nation Wide) including Puerto Rico and Internationally to Denmark, Guam
How much
405 units

Ureteroscope Intended to provide optical visualization of and therapeutic access to the Urinary Track. This a

Z-2078-2016  ·  Class II  ·  reported 2016-07-06  ·  Terminated

Why it was recalled
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
Product
Ureteroscope Intended to provide optical visualization of and therapeutic access to the Urinary Track. This anatomy includes but is not restricted to the organs, tissues and subsystems: Urethra, Bladder and Renal Pelvis. The instrument is introduced via the urethra when indications consistent with the requirement for procedure are observed in adult and pediatric patient populations.
Where it went
Worldwide Distribution - US (Nation Wide) including Puerto Rico and Internationally to Denmark, Guam
How much
32 units

Cystoscope Intended to provide optical visualization of and therapeutic access to the Urinary Tract. This ana

Z-2070-2016  ·  Class II  ·  reported 2016-07-06  ·  Terminated

Why it was recalled
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
Product
Cystoscope Intended to provide optical visualization of and therapeutic access to the Urinary Tract. This anatomy includes, but is not restricted to, the organs, tissues and subsystems: Urethra to the Bladder. The instrument is introduced via the urethra when indications consistent with the need for procedure are observed in adult and pediatric patient populations.
Where it went
Worldwide Distribution - US (Nation Wide) including Puerto Rico and Internationally to Denmark, Guam
How much
120 units

Choledocho-Nephroscope Intended to provide optical visualization of and therapeutic access to the Abdominal C

Z-2066-2016  ·  Class II  ·  reported 2016-07-06  ·  Terminated

Why it was recalled
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
Product
Choledocho-Nephroscope Intended to provide optical visualization of and therapeutic access to the Abdominal Cavity. This anatomy includes, but is not restricted to, the organs, tissues and subsystems: Biliary tree with access to the Common bile Duct, Hepatic Duct, Cystic Duct, Kidney, Renal Cavity and Renal Calyces. The instrument is introduced via percutaneous cut when indications consistent with the need for procedure are observed in adult and pediatric patient populations.
Where it went
Worldwide Distribution - US (Nation Wide) including Puerto Rico and Internationally to Denmark, Guam
How much
126 units

Confocal GI Scope Intended to provide optical visualization of an therapeutic access to the Gastrointestinal

Z-2069-2016  ·  Class II  ·  reported 2016-07-06  ·  Terminated

Why it was recalled
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
Product
Confocal GI Scope Intended to provide optical visualization of an therapeutic access to the Gastrointestinal Tract. This anatomy includes, but is not restricted to, the organs, tissues and subsystems: Esophagus, Stomach and Duodenum, and Large Bowel to the Cecum. The instrument is introduced via the mouth or rectum as decided by the anatomy being accessed when indications consistent with the need for procedure are observed in adult and pediatric patient populations.
Where it went
Worldwide Distribution - US (Nation Wide) including Puerto Rico and Internationally to Denmark, Guam
How much
17 units

Choledochoscope Intended to provide optical visualization of and therapeutic access to the Biliary Tree. This

Z-2067-2016  ·  Class II  ·  reported 2016-07-06  ·  Terminated

Why it was recalled
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
Product
Choledochoscope Intended to provide optical visualization of and therapeutic access to the Biliary Tree. This anatomy includes, but is not restricted to, the organs, tissues and subsystems: Common Bile, Hepatic, Cystic and Pancreatic Ducts. The instrument is introduced via the working channel of a Duodenoscope positioned for access to the Common bile Duct when indications consistent with the need for procedure are observed in adult and pediatric patient populations.
Where it went
Worldwide Distribution - US (Nation Wide) including Puerto Rico and Internationally to Denmark, Guam
How much
50 units

Gastroscope Intended to provide optical visualization of and therapeutic access to the Upper Gastrointestinal

Z-2071-2016  ·  Class II  ·  reported 2016-07-06  ·  Terminated

Why it was recalled
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
Product
Gastroscope Intended to provide optical visualization of and therapeutic access to the Upper Gastrointestinal Tract. This anatomy includes, but is not restricted to, the organs, tissues and subsystems: Esophagus, Stomach and Duodenum. The instrument is introduced via the mouth when indications consistent with the need for procedure are observed in adult and pediatric patient populations.
Where it went
Worldwide Distribution - US (Nation Wide) including Puerto Rico and Internationally to Denmark, Guam
How much
9,669 units

Naso-Pharyngo-Laryngoscope Intended to provide optical visualization of, therapeutic access to, the Upper Air

Z-2074-2016  ·  Class II  ·  reported 2016-07-06  ·  Terminated

Why it was recalled
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
Product
Naso-Pharyngo-Laryngoscope Intended to provide optical visualization of, therapeutic access to, the Upper Airway. This anatomy includes, but is not restricted to, the organs, tissues and subsystems: Ear, Nasal Passage and Trachea. The instrument is introduced via the mouth or the nose, as decided by the physician, when indications consistent with the need for procedure are observed in adult and pediatric patient populations
Where it went
Worldwide Distribution - US (Nation Wide) including Puerto Rico and Internationally to Denmark, Guam
How much
215 units

Video Duodenoscope

Z-1835-2016  ·  Class II  ·  reported 2016-06-08  ·  Terminated

Why it was recalled
Pentax Medical learned of reports of carbapenem-resistant Enterobacteriaceae (CRE) infection at a medical facility.
Product
Video Duodenoscope
Where it went
Nationwide
How much
423 units

PENTAX Upper/Lower G.I. Endoscope OF-B130 and OF-B194 Gas/Water Feeding Valve

Z-2290-2014  ·  Class II  ·  reported 2014-08-27  ·  Terminated

Why it was recalled
The Instructions for Use for the Pentax OF-B130 and OF-B194 Gas/Water Valve IFU if not followed carefully may cause a potential risk to health.
Product
PENTAX Upper/Lower G.I. Endoscope OF-B130 and OF-B194 Gas/Water Feeding Valve
Where it went
Worldwide Distribution - US Distribution including the states of CA, CO, IN, KS and ID., and the countries of Asia, Canada, Europe and Japan.
How much
98 units US; 1108 units OUS

EG-3630U Ultrasound Video Gastroscope EG-3830UT Ultrasound Video Gastroscope EG-3870UTK Ultrasound Video Gas

Z-2116-2014  ·  Class II  ·  reported 2014-07-30  ·  Terminated

Why it was recalled
Aspiration needles used in combination with Ultrasound Gastroscopes can derail from the accessory elevator at the distal end of the endoscope. This may result in patient injury if the Instruction for Use is not followed carefully by the user.
Product
EG-3630U Ultrasound Video Gastroscope EG-3830UT Ultrasound Video Gastroscope EG-3870UTK Ultrasound Video Gastroscope FG-36UX Ultrasound Fiber Gastroscope Used to provide optical and sonographic visualization of and therapeutic access to the Upper Gastrointestinal Tract.
Where it went
US Nationwide Distribution
How much
531 units
Class I

PENTAX Endoscopic Accessories. OF-B194. Made in Japan. PENTAX Upper/Lower G.I. Endoscopes are provided wi

Z-0084-2014  ·  Class I  ·  reported 2013-11-06  ·  Terminated

Why it was recalled
PENTAX Medical has become aware of a product issue associated with the OF-B194 Gas/Water Feeding Valve. A manufacturing defect may cause improper stoppage of CO2 gas during an endoscopic procedure with the potential for serious hazard to and/or user.
Product
PENTAX Endoscopic Accessories. OF-B194. Made in Japan. PENTAX Upper/Lower G.I. Endoscopes are provided with Air/Water Valve. As an alternative, the optionally available Gas/Water Valve may be used in place of the standard Air/Water Valve to deliver a non-explosive gas for insufflation.
Where it went
Nationwide Distribution including Arizona, California, Florida, Idaho, Indiana, Kansas, Maryland, Missouri, Missippi, North Carolina, New York, Pennsylvania, Virginia, Wisconsin and Wyoming.
How much
76 units
Read these counts with three things in mind

A firm with more recalls usually makes more products. A contract packer runs hundreds of lines. It will out-recall a company with one product whatever either does. This is not a safety ranking.

Most recalls here are voluntary. A firm found its own fault and pulled the stock. That is the system working. It lands in this database looking the same as a recall the FDA had to push for.

Terminated means the FDA considers the recall finished. It is not a verdict. It does not mean anyone was harmed.

Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.