Navilyst Medical, Inc: 73 FDA recall reports
Every medical device recall the FDA has published for Navilyst Medical, Inc. The fault, and how the agency graded it.
73Reports
0Class I
0Ongoing
2015-03-04Most recent
What the classes mean
The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.
Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.
By class
By year reported
Checking whether it affects you
The lot or code number is what matters. Not the product name. A recall covers
specific lots. Yours may not be one of them.
Check the code on your package against the report. The FDA keeps the current
list at fda.gov/safety/recalls.
The reports
The text below is quoted from the FDA record word for word.
NAMIC Convenience Kit, 3 VALVE KIT, UPN H749600709971, REF/Catalog No. 60070997, STERILE, Rx ONLY --- For sin
Z-1164-2015 · Class II · reported 2015-03-04 · Terminated
- Why it was recalled
- The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
- Product
- NAMIC Convenience Kit, 3 VALVE KIT, UPN H749600709971, REF/Catalog No. 60070997, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
- Where it went
- Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
- How much
- 602 units
NAMIC Convenience Kit, RADIOLOGY KIT, UPN H7496002072111, REF/Catalog No. 600207211, STERILE, Rx ONLY --- For
Z-1207-2015 · Class II · reported 2015-03-04 · Terminated
- Why it was recalled
- The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
- Product
- NAMIC Convenience Kit, RADIOLOGY KIT, UPN H7496002072111, REF/Catalog No. 600207211, STERILE, Rx ONLY --- For single use only. Device Listing Number D136683. NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
- Where it went
- Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
- How much
- 133 units
NAMIC Convenience Kit, RADIOLOGY KIT, UPN H749601506921, REF/Catalog No. 60150692, STERILE, Rx ONLY --- For si
Z-1176-2015 · Class II · reported 2015-03-04 · Terminated
- Why it was recalled
- The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
- Product
- NAMIC Convenience Kit, RADIOLOGY KIT, UPN H749601506921, REF/Catalog No. 60150692, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
- Where it went
- Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
- How much
- 22 units
NAMIC Convenience Kit, UPN H749601904231, REF/Catalog No. 60190423, STERILE, Rx ONLY --- For single use only.
Z-1184-2015 · Class II · reported 2015-03-04 · Terminated
- Why it was recalled
- The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
- Product
- NAMIC Convenience Kit, UPN H749601904231, REF/Catalog No. 60190423, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
- Where it went
- Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
- How much
- 1 unit
NAMIC Convenience Kit, UPN H965601931651, REF/Catalog No. 60193165, STERILE, Rx ONLY --- For single use only.
Z-1202-2015 · Class II · reported 2015-03-04 · Terminated
- Why it was recalled
- The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
- Product
- NAMIC Convenience Kit, UPN H965601931651, REF/Catalog No. 60193165, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
- Where it went
- Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
- How much
- 74 units
NAMIC Convenience Kit, LEFT HEART KIT, UPN H749600407061, REF/Catalog No. 60040706, STERILE, Rx ONLY --- For
Z-1155-2015 · Class II · reported 2015-03-04 · Terminated
- Why it was recalled
- The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
- Product
- NAMIC Convenience Kit, LEFT HEART KIT, UPN H749600407061, REF/Catalog No. 60040706, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
- Where it went
- Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
- How much
- 489 units
NAMIC Convenience Kit, RADIOLOGY KIT, UPN H749600617521, REF/Catalog No. 60061752, STERILE, Rx ONLY ---For sin
Z-1161-2015 · Class II · reported 2015-03-04 · Terminated
- Why it was recalled
- The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
- Product
- NAMIC Convenience Kit, RADIOLOGY KIT, UPN H749600617521, REF/Catalog No. 60061752, STERILE, Rx ONLY ---For single use only. --- Device Listing Number D136683. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
- Where it went
- Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
- How much
- 30 units
NAMIC Convenience Kit, LEFT HEART KIT, UPN H749602207461, REF/Catalog No. 60220746, STERILE, Rx ONLY ---For si
Z-1188-2015 · Class II · reported 2015-03-04 · Terminated
- Why it was recalled
- The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
- Product
- NAMIC Convenience Kit, LEFT HEART KIT, UPN H749602207461, REF/Catalog No. 60220746, STERILE, Rx ONLY ---For single use only. --- Device Listing Number D136588. NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
- Where it went
- Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
- How much
- 17 units
NAMIC Convenience Kit, UPN H749602301481, REF/Catalog No. 60230148, STERILE, Rx ONLY --- For single use only.
Z-1189-2015 · Class II · reported 2015-03-04 · Terminated
- Why it was recalled
- The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
- Product
- NAMIC Convenience Kit, UPN H749602301481, REF/Catalog No. 60230148, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
- Where it went
- Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
- How much
- 2502 units
NAMIC Convenience Kit, 3 VALVE KIT CARDIOVASCULAR PROCEDURE KIT, UPN H749600403131, REF/Catalog No. 60040313,
Z-1154-2015 · Class II · reported 2015-03-04 · Terminated
- Why it was recalled
- The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
- Product
- NAMIC Convenience Kit, 3 VALVE KIT CARDIOVASCULAR PROCEDURE KIT, UPN H749600403131, REF/Catalog No. 60040313, STERILE, Rx ONLY --- --- For single use only. --- Device Listing Number D136588. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
- Where it went
- Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
- How much
- 438 units
NAMIC Convenience Kit, 3V WMMII KIT, UPN H965601852521, REF/Catalog No. 60185252, STERILE, Rx ONLY --- For sin
Z-1200-2015 · Class II · reported 2015-03-04 · Terminated
- Why it was recalled
- The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
- Product
- NAMIC Convenience Kit, 3V WMMII KIT, UPN H965601852521, REF/Catalog No. 60185252, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
- Where it went
- Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
- How much
- 7 units
NAMIC Convenience Kit, UPN H749601308341, REF/Catalog No. 60130834, STERILE, Rx ONLY --- For single use only.
Z-1169-2015 · Class II · reported 2015-03-04 · Terminated
- Why it was recalled
- The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
- Product
- NAMIC Convenience Kit, UPN H749601308341, REF/Catalog No. 60130834, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
- Where it went
- Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
- How much
- 390 units
NAMIC Convenience Kit, UPN H749602001451, REF/Catalog No. 60200145, STERILE, Rx ONLY ---For single use only. -
Z-1187-2015 · Class II · reported 2015-03-04 · Terminated
- Why it was recalled
- The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
- Product
- NAMIC Convenience Kit, UPN H749602001451, REF/Catalog No. 60200145, STERILE, Rx ONLY ---For single use only. --- Device Listing Number D136588. NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
- Where it went
- Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
- How much
- 238 units
NAMIC Convenience Kit, UPN H7496013148131, REF/Catalog No. 601314813, STERILE, Rx ONLY --- For single use only
Z-1171-2015 · Class II · reported 2015-03-04 · Terminated
- Why it was recalled
- The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
- Product
- NAMIC Convenience Kit, UPN H7496013148131, REF/Catalog No. 601314813, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
- Where it went
- Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
- How much
- 124 units
NAMIC Convenience Kit, LEFT HEART KIT, UPN H7496007057101, REF/Catalog No. 600705710, STERILE, Rx ONLY --- For
Z-1163-2015 · Class II · reported 2015-03-04 · Terminated
- Why it was recalled
- The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
- Product
- NAMIC Convenience Kit, LEFT HEART KIT, UPN H7496007057101, REF/Catalog No. 600705710, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
- Where it went
- Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
- How much
- 5 units
NAMIC Convenience Kit, MOBILE CATH LAB KIT, UPN H749600419211, REF/Catalog No. 60041921, STERILE, Rx ONLY ---
Z-1157-2015 · Class II · reported 2015-03-04 · Terminated
- Why it was recalled
- The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
- Product
- NAMIC Convenience Kit, MOBILE CATH LAB KIT, UPN H749600419211, REF/Catalog No. 60041921, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
- Where it went
- Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
- How much
- 16 units
NAMIC Convenience Kit, 5 VALVE MANIFOLD KIT, UPN H749600105121, REF/Catalog No. 60010512, STERILE, Rx ONLY ---
Z-1149-2015 · Class II · reported 2015-03-04 · Terminated
- Why it was recalled
- The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
- Product
- NAMIC Convenience Kit, 5 VALVE MANIFOLD KIT, UPN H749600105121, REF/Catalog No. 60010512, STERILE, Rx ONLY --- For single use only. Device Listing Number D136588. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
- Where it went
- Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
- How much
- 19 units
NAMIC Convenience Kit, UPN H7496016046121, REF/Catalog No. 601604612, STERILE, Rx ONLY --- For single use only
Z-1177-2015 · Class II · reported 2015-03-04 · Terminated
- Why it was recalled
- The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
- Product
- NAMIC Convenience Kit, UPN H7496016046121, REF/Catalog No. 601604612, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
- Where it went
- Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
- How much
- 1008 units
NAMIC Convenience Kit, UPN H749600203741, REF/Catalog No. 60020374, STERILE, Rx ONLY --- For single use only.
Z-1151-2015 · Class II · reported 2015-03-04 · Terminated
- Why it was recalled
- The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
- Product
- NAMIC Convenience Kit, UPN H749600203741, REF/Catalog No. 60020374, STERILE, Rx ONLY --- For single use only. Device Listing Number D136588. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
- Where it went
- Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
- How much
- 4 units
NAMIC Convenience Kit, ANGIO DOUBLE SPIKE KIT, UPN H9656003028111, REF/Catalog No. 600302811, STERILE, Rx ONLY
Z-1194-2015 · Class II · reported 2015-03-04 · Terminated
- Why it was recalled
- The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
- Product
- NAMIC Convenience Kit, ANGIO DOUBLE SPIKE KIT, UPN H9656003028111, REF/Catalog No. 600302811, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136683. NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
- Where it went
- Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
- How much
- 2 units
NAMIC Convenience Kit, LEFT HEART KIT, UPN H749602322341, REF/Catalog No. 60232234, STERILE, Rx ONLY --- For s
Z-1190-2015 · Class II · reported 2015-03-04 · Terminated
- Why it was recalled
- The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
- Product
- NAMIC Convenience Kit, LEFT HEART KIT, UPN H749602322341, REF/Catalog No. 60232234, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
- Where it went
- Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
- How much
- 61 units
NAMIC Convenience Kit, UPN H749601827221, REF/Catalog No. 60182722, STERILE, Rx ONLY --- For single use only.
Z-1181-2015 · Class II · reported 2015-03-04 · Terminated
- Why it was recalled
- The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
- Product
- NAMIC Convenience Kit, UPN H749601827221, REF/Catalog No. 60182722, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
- Where it went
- Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
- How much
- 1 unit
NAMIC Convenience Kit, UPN H965607011741, REF/Catalog No. 60701174, STERILE, Rx ONLY --- For single use only.
Z-1204-2015 · Class II · reported 2015-03-04 · Terminated
- Why it was recalled
- The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
- Product
- NAMIC Convenience Kit, UPN H965607011741, REF/Catalog No. 60701174, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
- Where it went
- Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
- How much
- 541 units
NAMIC Convenience Kit, LEFT HEART KIT, UPN H749601617931, REF/Catalog No. 60161793, STERILE, Rx ONLY ---For si
Z-1179-2015 · Class II · reported 2015-03-04 · Terminated
- Why it was recalled
- The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
- Product
- NAMIC Convenience Kit, LEFT HEART KIT, UPN H749601617931, REF/Catalog No. 60161793, STERILE, Rx ONLY ---For single use only. --- Device Listing Number D136588. NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
- Where it went
- Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
- How much
- 69 units
NAMIC Convenience Kit, UPN H749606204751, REF/Catalog No. 60620475, STERILE, Rx ONLY --- For single use only.
Z-1191-2015 · Class II · reported 2015-03-04 · Terminated
- Why it was recalled
- The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
- Product
- NAMIC Convenience Kit, UPN H749606204751, REF/Catalog No. 60620475, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
- Where it went
- Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
- How much
- 13 units
NAMIC Convenience Kit, UPN H965601955411, REF/Catalog No. 60195541, STERILE, Rx ONLY --- For single use only.
Z-1203-2015 · Class II · reported 2015-03-04 · Terminated
- Why it was recalled
- The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
- Product
- NAMIC Convenience Kit, UPN H965601955411, REF/Catalog No. 60195541, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
- Where it went
- Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
- How much
- 20 units
NAMIC Convenience Kit, LEFT HEART, UPN H7496006044121, REF/Catalog No. 600604412, STERILE, Rx ONLY ---For sing
Z-1160-2015 · Class II · reported 2015-03-04 · Terminated
- Why it was recalled
- The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
- Product
- NAMIC Convenience Kit, LEFT HEART, UPN H7496006044121, REF/Catalog No. 600604412, STERILE, Rx ONLY ---For single use only. --- Device Listing Number D136588. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
- Where it went
- Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
- How much
- 2409 units
NAMIC Convenience Kit, PTCA KIT, UPN H9656019110161, REF/Catalog No. 601911016, STERILE, Rx ONLY --- For singl
Z-1201-2015 · Class II · reported 2015-03-04 · Terminated
- Why it was recalled
- The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
- Product
- NAMIC Convenience Kit, PTCA KIT, UPN H9656019110161, REF/Catalog No. 601911016, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
- Where it went
- Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
- How much
- 70 units
NAMIC Convenience Kit, LEFT/RIGHT HEART KIT, UPN H7496008010201, REF/Catalog No. 600801020, STERILE, Rx ONLY -
Z-1165-2015 · Class II · reported 2015-03-04 · Terminated
- Why it was recalled
- The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
- Product
- NAMIC Convenience Kit, LEFT/RIGHT HEART KIT, UPN H7496008010201, REF/Catalog No. 600801020, STERILE, Rx ONLY ---For single use only. ---Device Listing Number D136588. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
- Where it went
- Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
- How much
- 105 units
NAMIC Convenience Kit, UPN H7496005216131, REF/Catalog No. 600521613, STERILE, Rx ONLY --- For single use onl
Z-1159-2015 · Class II · reported 2015-03-04 · Terminated
- Why it was recalled
- The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
- Product
- NAMIC Convenience Kit, UPN H7496005216131, REF/Catalog No. 600521613, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
- Where it went
- Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
- How much
- 161 units
Showing the 30 most recent of 73. The rest are searchable at the FDA enforcement report database.
Read these counts with three things in mind
A firm with more recalls usually makes more products. A contract packer runs
hundreds of lines. It will out-recall a company with one product whatever either
does. This is not a safety ranking.
Most recalls here are voluntary. A firm found its own fault and pulled the stock.
That is the system working. It lands in this database looking the same as a recall
the FDA had to push for.
Terminated means the FDA considers the recall finished. It is not a verdict.
It does not mean anyone was harmed.