Microvention, Inc.

Microvention, Inc.: 37 FDA recall reports

Every medical device recall the FDA has published for Microvention, Inc.. The fault, and how the agency graded it.

37Reports
0Class I
10Ongoing
2026-05-20Most recent

What the classes mean

The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.

ClassWhat the FDA is saying
Class IThere is a reasonable probability that using this will cause serious harm or death.
Class IIIt may cause temporary or reversible harm. Serious harm is unlikely.
Class IIIIt breaks a rule but is unlikely to hurt anyone. Often a labelling fault.

Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.

By class

ClassReports
Class II34
Class III3

By year reported

YearReports
20171
202022
20214
20221
20232
20242
20254
20261
Checking whether it affects you

The lot or code number is what matters. Not the product name. A recall covers specific lots. Yours may not be one of them.

Check the code on your package against the report. The FDA keeps the current list at fda.gov/safety/recalls.

The reports

The text below is quoted from the FDA record word for word.

LVIS Intraluminal Support Device, REF/Outside Diameter(OD)xTotal Length(TL)xWorking Length(WL): 214022-CAS/4.0

Z-2200-2026  ·  Class II  ·  reported 2026-05-20  ·  Ongoing

Why it was recalled
Stent-assisted coiling device may due to commingling during manufacturing have dimensional nonconformance and/or incorrect device configuration. If device is oversized may lead to increased vessel wall stress, neointimal hyperplasia, vessel injury/rupture; if undersized may lead to incomplete expansion/malposition, acute stent thrombosis, distal embolization, ischemic neurological events
Product
LVIS Intraluminal Support Device, REF/Outside Diameter(OD)xTotal Length(TL)xWorking Length(WL): 214022-CAS/4.0 mm x 22 mm x 18 mm, 214518-CAS/4.5 mm x 18 mm x 14 mm, 213517-CAS/3.5 mm x 17 mm x 13 mm
Where it went
International distribution to the country of China.
How much
38

Flow Re-Direction Endoluminal Devices: FRED 27, REF: FRED5019, FRED5029, FRED4528, FRED4026, FRED5526, FRED450

Z-0108-2026  ·  Class II  ·  reported 2025-10-22  ·  Ongoing

Why it was recalled
Flow diverter may have a manufacturing issue related to the tantalum length and/or tantalum attachment pattern, which may lead the device to be incompletely open or not properly apposed to the vessel wall, which may necessitate device recapture and withdrawal of the delivery system and microcatheter together. Insufficient apposition could lead to long-term health consequences.
Product
Flow Re-Direction Endoluminal Devices: FRED 27, REF: FRED5019, FRED5029, FRED4528, FRED4026, FRED5526, FRED4508, FRED5514, FRED3536, FRED3511, FRED5014, FRED3516, MV-F451827; FRED X 27, REF: XFRED4017-MVE, XFRED4528-MVE, MV-F501427X, XFRED4518-MVE, XFRED5526-MVE, XFRED4539-MVE, FREDX5522-PMA, FREDX3522-PMA, FREDX4525-PMA, FREDX5015-PMA, FREDX5532-PMA, FREDX4013-PMA, FREDX3513-PMA, FREDX4515-PMA, FREDX3517-PMA, MV-F352427X
Where it went
Worldwide - US Nationwide distribution in the states of SD, CA, SC, TX, FL, CO, NE, NY and the countries of DE, ES, PL, GB, IT, IN, AU, KZ, JP, AT, SI, RS, GR, TR, AR, CH, SG, KR, SK, MX, DK, IR, SA, BY, PA, VN, ZA, BR, TW, CN, BE, FR, VE, MY.
How much
467

LIFEPEARL Drug Elutable Microspheres: 200 +/- 50 micrometers, REF: 8LP2S200; 400 +/- 50 micrometers, REF: 8LP2

Z-2188-2025  ·  Class II  ·  reported 2025-08-06  ·  Ongoing

Why it was recalled
Drug Elutable Microspheres have a smaller actual average diameter that is not within specification, which may lead to inability to reach the desired treatment location (non-targeted embolization), additional procedure/treatment required, blockage other than target vessel, increased procedure time, incomplete embolization, and/or inability to treat the patient.
Product
LIFEPEARL Drug Elutable Microspheres: 200 +/- 50 micrometers, REF: 8LP2S200; 400 +/- 50 micrometers, REF: 8LP2S400
Where it went
OUS: Italy, Spain, Portugal, Poland, Belgium, Germany, Denmark, France, Kazakhstan, Slovakia, Norway, Greece, Serbia, Turkey, Czech Republic, United Kingdom, Netherlands, Russian Federation, Kuwait, Republic of North Macedonia, Switzerland, Saudi Arabia, Georgia, Lithuania, Romania, Croatia, Chile, Hong Kong , India, Mexico, Taiwan, Vietnam
How much
1451

MicoVention Terumo, LVIS Intraluminal Support Device, REF: 213522-CAS, od, 3.5 mm, TL 22 mm, WL 18 mm, Sterile

Z-1121-2025  ·  Class II  ·  reported 2025-02-19  ·  Ongoing

Why it was recalled
Potential for Polytetrafluoethylene (PTFE) material encasing the stent loop may prevent the stent from being pushed out of the introducer.
Product
MicoVention Terumo, LVIS Intraluminal Support Device, REF: 213522-CAS, od, 3.5 mm, TL 22 mm, WL 18 mm, Sterile R, CE0297. The intended purpose of the LVIS Device is to serve as a stent-assisted coiling (SAC) device providing support to the embolic coils used in the treatment of wide-necked aneurysms.
Where it went
US: None OUS: China
How much
11 units

Microvention, Headway" 17 Advanced Straight Microcatheter with Hydrophilic Coating, REF: MC172150SX-CN, Steril

Z-0990-2025  ·  Class II  ·  reported 2025-01-29  ·  Ongoing

Why it was recalled
Sterility assurance; microcatheter inner packaging may contain incomplete seal.
Product
Microvention, Headway" 17 Advanced Straight Microcatheter with Hydrophilic Coating, REF: MC172150SX-CN, Sterile R, Rx Only, Lot# 0000629846
Where it went
US: None OUS: China
How much
107 units

ERIC Retrieval Device REF: ER173020 ER174030 ER176044 The ERIC Retrieval Device is a mechanical thrombe

Z-3157-2024  ·  Class II  ·  reported 2024-09-25  ·  Ongoing

Why it was recalled
Due to Retrieval Device product labels stating the incorrect expiration dates that (5 years instead of 3 years) which extends beyond their true expiration date.
Product
ERIC Retrieval Device REF: ER173020 ER174030 ER176044 The ERIC Retrieval Device is a mechanical thrombectomy device designed to restore blood flow by effectively removing clots from occlusive vasculature in patients suffering from acute ischemic stroke.
Where it went
U.S. Nationwide distribution in the states of AR, CA, FL, IN, LA, MA, MN, MO, NC, NJ, NY, OH, PA, SC, TX and VA.
How much
1,370 devices

TERUMO HydroPearl Compressible Microspheres for Embolisation REF 8HP2S600

Z-0868-2024  ·  Class II  ·  reported 2024-02-07  ·  Ongoing

Why it was recalled
Due to, during the manufacturing process, the prescribed manufacturing and quality processes were not followed and the product was inadvertently distributed.
Product
TERUMO HydroPearl Compressible Microspheres for Embolisation REF 8HP2S600
Where it went
U.S. Nationwide distribution in the states of AZ, IL, LA, NC, TN, VA, and WI.
How much
22 units

AZUR CX 35 Peripheral Coil System Detachable 13mm x 24 CM, REF: 45-751324

Z-0187-2024  ·  Class II  ·  reported 2023-11-08  ·  Ongoing

Why it was recalled
Peripheral coil system detachable has a potential of unsealed pouch packaging.
Product
AZUR CX 35 Peripheral Coil System Detachable 13mm x 24 CM, REF: 45-751324
Where it went
US Nationwide distribution in the states of TX, NY, VA, NV, MI, CA, AZ, WA.
How much
14

FRED Flow Re-Direction Endoluminal Device, REF: FRED3507-PMA, FRED3511, FRED3511-PMA, FRED3516-CA, FRED3516-PM

Z-2238-2023  ·  Class II  ·  reported 2023-08-02  ·  Ongoing

Why it was recalled
Intracranial aneurysm flow diverter device may be manufactured with an incorrect inner stent length and may not perform as intended.
Product
FRED Flow Re-Direction Endoluminal Device, REF: FRED3507-PMA, FRED3511, FRED3511-PMA, FRED3516-CA, FRED3516-PMA, FRED3524, FRED3524-CA, FRED3524-PMA, FRED3536-CA, FRED3536-PMA, FRED4007-PMA, FRED4012-CA, FRED4012-PMA, FRED4017-PMA, FRED4026, FRED4026-PMA, FRED4038-PMA, FRED4508, FRED4508-PMA, FRED4513-PMA, FRED4518-PMA, FRED4528, FRED4528-PMA, FRED4539, FRED4539-CA, FRED4539-PMA, FRED5009, FRED5009-PMA, FRED5014-PMA, FRED5019-PMA, FRED5029, FRED5029-CA, FRED5029-PMA, FRED5514-CA, FRED5514-PMA, FRED5526-CA, FRED5526-PMA, MV-F351127, MV-F352427, MV-F401227, MV-F401727, MV-F451327, MV-F500927, MV-F501427, MV-F501927, MV-F552627
Where it went
Worldwide - US Nationwide distribution including in the states of AZ, NY, FL, CA, PA, SC, TX, IL, MA, NJ, VA, GA, OH, IN, MI, KY, AL, MO, SD, LA, WA, HI, KS, MD, OK, DE, NC, MS and the countries of Taiwan, Chile, India, South Korea, Malaysia, Canada, France, Morocco, Belgium, Spain, Turkey, Poland, Russian Federation, Czech Republic, Slovenia, Germany, United Kingdom, El Salvador, United Arab Emirates, Bulgaria, Portugal, Brazil, Iran, Malta, Vietnam, South Africa, Jordan, Japan, Uruguay.
How much
474

WEB Detachment Controller, REF: WDC-2, WDC, MV-WDC2. Component of WEB Aneurysm Embolization System.

Z-0170-2023  ·  Class II  ·  reported 2022-11-09  ·  Ongoing

Why it was recalled
Detachment controller, of an aneurysm embolization system, has an out of specification tunnel opening dimension, which may prevent the implant device pusher from being fully inserted into the controller, which may result in the inability of the controller to properly detach the implant device, which could cause delays and excessive manipulation that may cause the implant to move and damage vessels
Product
WEB Detachment Controller, REF: WDC-2, WDC, MV-WDC2. Component of WEB Aneurysm Embolization System.
Where it went
Worldwide - US Nationwide distribution in the states of MN, PA, NY, MI, FL, VA, TX, OH, CA, NE, MN, NC, LA, IL, TN, IN, UT, KS and the countries of Japan, China, Argentina, Australia, South Korea, Italy, UK.
How much
2018

Aspiration Syringe Kit, REF: MVSK60

Z-0202-2022  ·  Class II  ·  reported 2021-11-10  ·  Terminated

Why it was recalled
Syringe Kits labeled for distribution OUS were inadvertently shipped to US customers.
Product
Aspiration Syringe Kit, REF: MVSK60
Where it went
US Nationwide distribution in the states of AL, NY, NV.
How much
33

VIA Microcatheter VIA 21 REF VIA-21-154-01 - Product Usage: is a single lumen catheter designed to be introduc

Z-1829-2021  ·  Class III  ·  reported 2021-06-16  ·  Terminated

Why it was recalled
Due to inadvertently shipping to U.S. customers with international labeling which has some differences with US-cleared labeling.
Product
VIA Microcatheter VIA 21 REF VIA-21-154-01 - Product Usage: is a single lumen catheter designed to be introduced over a steerable guidewire into the vasculature.
Where it went
U.S. Nationwide distribution in the states of FL, IA, MI, MN, TN and TX.
How much
N/A

VIA Microcatheter VIA 17 REF VIA-17-154-01 - Product Usage: is a single lumen catheter designed to be introduc

Z-1828-2021  ·  Class III  ·  reported 2021-06-16  ·  Terminated

Why it was recalled
Due to inadvertently shipping to U.S. customers with international labeling which has some differences with US-cleared labeling.
Product
VIA Microcatheter VIA 17 REF VIA-17-154-01 - Product Usage: is a single lumen catheter designed to be introduced over a steerable guidewire into the vasculature.
Where it went
U.S. Nationwide distribution in the states of FL, IA, MI, MN, TN and TX.
How much
N/A

VIA Microcatheter VIA 27 REF VIA-27-154-01 - Product Usage: is a single lumen catheter designed to be introduc

Z-1830-2021  ·  Class III  ·  reported 2021-06-16  ·  Terminated

Why it was recalled
Due to inadvertently shipping to U.S. customers with international labeling which has some differences with US-cleared labeling.
Product
VIA Microcatheter VIA 27 REF VIA-27-154-01 - Product Usage: is a single lumen catheter designed to be introduced over a steerable guidewire into the vasculature.
Where it went
U.S. Nationwide distribution in the states of FL, IA, MI, MN, TN and TX.
How much
N/A

MicroVention Terumo HydroSoft 10, HydroCoil Embolic System, Endovascular Embolization Coil, Helical, Sterile,

Z-1392-2020  ·  Class II  ·  reported 2020-03-04  ·  Terminated

Why it was recalled
The devices may be missing the implant coil.
Product
MicroVention Terumo HydroSoft 10, HydroCoil Embolic System, Endovascular Embolization Coil, Helical, Sterile, Rx, REF numbers 100101H2HS-V; 100102H2HS-V, 100103H2HS-V, 100104H2HS-V, 100151H2HS-V; 100152H2HS-V; 100153H2HS-V; 100154H2HS-V; 100201H2HS-V; 100202H2HS-V; 100203H2HS-V; 100204H2HS-V; 100206H2HS-V, 100208H2HS-V, 100254H2HS-V, 100256H2HS-V, 100304H2HS-V; 100306H2HS-V; 100308H2HS-V; 100310H2HS-V; 100404H2HS-V; 100406H2HS-V, 100408H2HS-V, 100410H2HS-V, 100506H2HS-V, 100508H2HS-V; 100510H2HS-V, 100515H2HS-V, 100520H2HS-V, 100620H2HS-V, and MV-00208HHSA - Product Usage: The HES is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. The firm name on the label is MicroVention, Inc., Tustin, CA.
Where it went
All distribution was OUS. The countries receiving affected product are: Albania, Austria, Belarus, Belgium, Brazil, Bulgaria, China, Cyprus, Czech Republic, Denmark, Egypt, Finalnd, France, Georgia, Germany, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Malta, Morocco, Netherlands, Norway, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Slovak Republic, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Turkey, Ukraine, United Arab Emirates, and United Kingdom.
How much
2,762 units

MicroVention Terumo Microplex 18, Platinum Coil System, Endovascular Embolization Coil, Complex 18, Sterile, R

Z-1401-2020  ·  Class II  ·  reported 2020-03-04  ·  Terminated

Why it was recalled
The devices may be missing the implant coil.
Product
MicroVention Terumo Microplex 18, Platinum Coil System, Endovascular Embolization Coil, Complex 18, Sterile, Rx, REF numbers 180512CC-V, 180820CC-V, 181026CC-V, 181434CC-V, and 182050CC-V. Product Usage: The MCS is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. The firm name on the label is MicroVention, Inc., Tustin, CA.
Where it went
All distribution was OUS. The countries receiving affected product are: Albania, Austria, Belarus, Belgium, Brazil, Bulgaria, China, Cyprus, Czech Republic, Denmark, Egypt, Finalnd, France, Georgia, Germany, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Malta, Morocco, Netherlands, Norway, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Slovak Republic, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Turkey, Ukraine, United Arab Emirates, and United Kingdom.
How much
19 units

MicroVention Terumo Microplex 10, Platinum Coil System, Endovascular Embolization Coil, Compass 10, Sterile, R

Z-1404-2020  ·  Class II  ·  reported 2020-03-04  ·  Terminated

Why it was recalled
The devices may be missing the implant coil.
Product
MicroVention Terumo Microplex 10, Platinum Coil System, Endovascular Embolization Coil, Compass 10, Sterile, Rx, REF numbers 100203CMSR-V, 100408CMSR-V, 100412CMSR-V, 100510CMSR-V, 100516CMSR-V, 100612CMSR-V, 100618CMSR-V, and 100721CMSR-V. Product Usage: The MCS is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. The firm name on the label is MicroVention, Inc., Tustin, CA.
Where it went
All distribution was OUS. The countries receiving affected product are: Albania, Austria, Belarus, Belgium, Brazil, Bulgaria, China, Cyprus, Czech Republic, Denmark, Egypt, Finalnd, France, Georgia, Germany, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Malta, Morocco, Netherlands, Norway, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Slovak Republic, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Turkey, Ukraine, United Arab Emirates, and United Kingdom.
How much
85 units

MicroVention Terumo Microplex 18, Platinum Coil System, Endovascular Embolization Coil, Helical 18 Soft, Steri

Z-1397-2020  ·  Class II  ·  reported 2020-03-04  ·  Terminated

Why it was recalled
The devices may be missing the implant coil.
Product
MicroVention Terumo Microplex 18, Platinum Coil System, Endovascular Embolization Coil, Helical 18 Soft, Sterile, Rx, REF numbers 180204HC-S-V, 180208HC-S-V, 180308HC-S-V, 180410HC-S-V, and 180512HC-S-V. Product Usage: The MCS is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.
Where it went
All distribution was OUS. The countries receiving affected product are: Albania, Austria, Belarus, Belgium, Brazil, Bulgaria, China, Cyprus, Czech Republic, Denmark, Egypt, Finalnd, France, Georgia, Germany, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Malta, Morocco, Netherlands, Norway, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Slovak Republic, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Turkey, Ukraine, United Arab Emirates, and United Kingdom.
How much
169 units

MicroVention Terumo HydroFrame 18, HydroCoil Embolic System, Endovascular Embolization Coil, Sterile, Rx, REF

Z-1403-2020  ·  Class II  ·  reported 2020-03-04  ·  Terminated

Why it was recalled
The devices may be missing the implant coil.
Product
MicroVention Terumo HydroFrame 18, HydroCoil Embolic System, Endovascular Embolization Coil, Sterile, Rx, REF numbers 180723HFRM-V, 180931HFRM-V, 181644HFRM-V, MV-80619HHFA, MV80827HHFA, MV-80931HHFA, and MV-81036HHFA. Product Usage: The HES is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. The firm name on the label is MicroVention, Inc., Tustin, CA.
Where it went
All distribution was OUS. The countries receiving affected product are: Albania, Austria, Belarus, Belgium, Brazil, Bulgaria, China, Cyprus, Czech Republic, Denmark, Egypt, Finalnd, France, Georgia, Germany, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Malta, Morocco, Netherlands, Norway, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Slovak Republic, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Turkey, Ukraine, United Arab Emirates, and United Kingdom.
How much
32 units

MicroVention Terumo Microplex 10, Platinum Coil System, Endovascular Embolization Coil, HyperSoft Helical, Ste

Z-1407-2020  ·  Class II  ·  reported 2020-03-04  ·  Terminated

Why it was recalled
The devices may be missing the implant coil.
Product
MicroVention Terumo Microplex 10, Platinum Coil System, Endovascular Embolization Coil, HyperSoft Helical, Sterile, Rx, REF numbers 100101HS-V, 100102HS-V, 100103HS-V, 100104HS-V, 100151HS-V, 100152HS-V, 100153HS-V, 100154HS-V, 100201HS-V, 100202HS-V, 100203HS-V, 100204HS-V, 100206HS-V, 100254HS-V, 100256HS-V, 100304HS-V, 100306HS-V, 100308HS-V, 100406HS-V, 100408HS-V, and 100508HS-V. Product Usage: The MCS is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. The firm name on the label is MicroVention, Inc., Tustin, CA.
Where it went
All distribution was OUS. The countries receiving affected product are: Albania, Austria, Belarus, Belgium, Brazil, Bulgaria, China, Cyprus, Czech Republic, Denmark, Egypt, Finalnd, France, Georgia, Germany, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Malta, Morocco, Netherlands, Norway, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Slovak Republic, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Turkey, Ukraine, United Arab Emirates, and United Kingdom.
How much
5,777 units

MicroVention Terumo Microplex 10, Platinum Coil System, Endovascular Embolization Coil, Helical 10 Soft, Steri

Z-1395-2020  ·  Class II  ·  reported 2020-03-04  ·  Terminated

Why it was recalled
The devices may be missing the implant coil.
Product
MicroVention Terumo Microplex 10, Platinum Coil System, Endovascular Embolization Coil, Helical 10 Soft, Sterile, Rx, REF numbers 100206HCSR-S-V, 100208HC-S-V, 100208HCSR-S-V, 100306HCSR-S-V, 100308HCSR-S-V, 100310HC-S-V, 100310HCSR-S-V, 100410HC-S-V, 100410HCSR-S-V, Product Usage: The MCS is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. The firm name on the label is MicroVention, Inc., Tustin, CA.
Where it went
All distribution was OUS. The countries receiving affected product are: Albania, Austria, Belarus, Belgium, Brazil, Bulgaria, China, Cyprus, Czech Republic, Denmark, Egypt, Finalnd, France, Georgia, Germany, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Malta, Morocco, Netherlands, Norway, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Slovak Republic, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Turkey, Ukraine, United Arab Emirates, and United Kingdom.
How much
780 units

MicroVention Terumo Microplex 10, Platinum Coil System, Endovascular Embolization Coil, Complex 10, Sterile, R

Z-1400-2020  ·  Class II  ·  reported 2020-03-04  ·  Terminated

Why it was recalled
The devices may be missing the implant coil.
Product
MicroVention Terumo Microplex 10, Platinum Coil System, Endovascular Embolization Coil, Complex 10, Sterile, Rx, REF numbers 100204CC-V, 100307CC-V, 100410CC-V, 100512CC-V, 100615CC-V, 100718CC-V, 100820CC-V, 100924CC-V, and 101026CC-V. Product Usage: The MCS is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. The firm name on the label is MicroVention, Inc., Tustin, CA.
Where it went
All distribution was OUS. The countries receiving affected product are: Albania, Austria, Belarus, Belgium, Brazil, Bulgaria, China, Cyprus, Czech Republic, Denmark, Egypt, Finalnd, France, Georgia, Germany, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Malta, Morocco, Netherlands, Norway, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Slovak Republic, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Turkey, Ukraine, United Arab Emirates, and United Kingdom.
How much
409 units

MicroVention Terumo Microplex, Platinum Coil System, Endovascular Embolization Coil, VFC, Sterile, Rx, REF num

Z-1406-2020  ·  Class II  ·  reported 2020-03-04  ·  Terminated

Why it was recalled
The devices may be missing the implant coil.
Product
MicroVention Terumo Microplex, Platinum Coil System, Endovascular Embolization Coil, VFC, Sterile, Rx, REF numbers MV-00306VFC, MV-00315VFC, MV-00310VFC, MV-00630VFC, VFC010303-V, VFC030606-V, VFC030610-V, VFC030615-V, VFC061020-V, VFC061030-V, VFC101530-V, VFC101540-V, VFC152040-V, and VFC152060-V. Product Usage: . The MCS is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. The firm name on the label is MicroVention, Inc., Tustin, CA.
Where it went
All distribution was OUS. The countries receiving affected product are: Albania, Austria, Belarus, Belgium, Brazil, Bulgaria, China, Cyprus, Czech Republic, Denmark, Egypt, Finalnd, France, Georgia, Germany, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Malta, Morocco, Netherlands, Norway, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Slovak Republic, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Turkey, Ukraine, United Arab Emirates, and United Kingdom.
How much
666 units

MicroVention Terumo Microplex 10, Platinum Coil System, Endovascular Embolization Coil, Helical 10 Regular, St

Z-1394-2020  ·  Class II  ·  reported 2020-03-04  ·  Terminated

Why it was recalled
The devices may be missing the implant coil.
Product
MicroVention Terumo Microplex 10, Platinum Coil System, Endovascular Embolization Coil, Helical 10 Regular, Sterile, Rx, REF numbers 100515HC-R-V, 100515HCSR-R-V, 100520HC-R-V, 100520HCSR-R-V, 100620HC-R-V, 100620HCSR-R-V, 100730HC-R-V, 100730HCSR-R-V, 100830HC-R-V, 100930HC-R-V, 100930HCSR-R-V, 101030HC-R-V, and 101030HCSR-R-V. Product Usage: The MCS is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. The firm name on the label is MicroVention, Inc., Tustin, CA.
Where it went
All distribution was OUS. The countries receiving affected product are: Albania, Austria, Belarus, Belgium, Brazil, Bulgaria, China, Cyprus, Czech Republic, Denmark, Egypt, Finalnd, France, Georgia, Germany, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Malta, Morocco, Netherlands, Norway, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Slovak Republic, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Turkey, Ukraine, United Arab Emirates, and United Kingdom.
How much
486 units

MicroVention Terumo Microplex 10, Platinum Coil System, Endovascular Embolization Coil, Cosmos 10, Sterile, Rx

Z-1398-2020  ·  Class II  ·  reported 2020-03-04  ·  Terminated

Why it was recalled
The devices may be missing the implant coil.
Product
MicroVention Terumo Microplex 10, Platinum Coil System, Endovascular Embolization Coil, Cosmos 10, Sterile, Rx, REF numbers 100202CSSR-V, 100254CSSR-V, 100306CSSR-V, 100408CSSR-V, 100412CSSR-V, 100515CSSR-V, 100522CSSR-V, 100618CSSR-V, 100626CSSR-V, 100722CSSR-V, 100731CSSR-V, 100825CSSR-V, 100837CSSR-V, 100933CSSR-V, and 101036CSSR-V. Product Usage: The MCS is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. The firm name on the label is MicroVention, Inc., Tustin, CA.
Where it went
All distribution was OUS. The countries receiving affected product are: Albania, Austria, Belarus, Belgium, Brazil, Bulgaria, China, Cyprus, Czech Republic, Denmark, Egypt, Finalnd, France, Georgia, Germany, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Malta, Morocco, Netherlands, Norway, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Slovak Republic, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Turkey, Ukraine, United Arab Emirates, and United Kingdom.
How much
3,442 units

MicroVention Terumo HydroFrame 10, HydroCoil Embolic System, Endovascular Embolization Coil, Sterile, Rx, REF

Z-1402-2020  ·  Class II  ·  reported 2020-03-04  ·  Terminated

Why it was recalled
The devices may be missing the implant coil.
Product
MicroVention Terumo HydroFrame 10, HydroCoil Embolic System, Endovascular Embolization Coil, Sterile, Rx, REF numbers 100306HFRM-V, 100405HFRM-V, 100408HFRM-V, 100410HFRM-V, 100412HFRM-V, 100415HFRM-V, 100510HFRM-V, 100515HFRM-V, 100519HFRM-V, 100612HFRM-V, 100619HFRM-V, 100623HFRM-V, 100715HFRM-V, 100728HFRM-V, 100817HFRM-V, 100833HFRM-V, 100931HFRM-V; 101036HFRM-V, MV-00408HHFA, MV-00510HHFA, MV-00515HHFA, and MV-00619HHFA. Product Usage: The HES is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.
Where it went
All distribution was OUS. The countries receiving affected product are: Albania, Austria, Belarus, Belgium, Brazil, Bulgaria, China, Cyprus, Czech Republic, Denmark, Egypt, Finalnd, France, Georgia, Germany, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Malta, Morocco, Netherlands, Norway, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Slovak Republic, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Turkey, Ukraine, United Arab Emirates, and United Kingdom.
How much
991 units

MicroVention Terumo Microplex 18, Platinum Embolization Coil, Cosmos 18, Sterile, Rx, REF numbers 180619CS-V,

Z-1399-2020  ·  Class II  ·  reported 2020-03-04  ·  Terminated

Why it was recalled
The devices may be missing the implant coil.
Product
MicroVention Terumo Microplex 18, Platinum Embolization Coil, Cosmos 18, Sterile, Rx, REF numbers 180619CS-V, 180723CS-V, 180827CS-V, 180931CS-V, 181036CS-V, 181139CS-V, 181243CS-V, 181347CS-V, 181451CS-V, 181652CS-V, 181859CS-V, 182065CS-V, 182263CS-V, and 182468CS-V; Product Usage: The MCS is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. The firm name on the label is MicroVention, Inc., Tustin, CA.
Where it went
All distribution was OUS. The countries receiving affected product are: Albania, Austria, Belarus, Belgium, Brazil, Bulgaria, China, Cyprus, Czech Republic, Denmark, Egypt, Finalnd, France, Georgia, Germany, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Malta, Morocco, Netherlands, Norway, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Slovak Republic, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Turkey, Ukraine, United Arab Emirates, and United Kingdom.
How much
403 units

MicroVention Terumo HydroSoft 10, HydroCoil Embolic System, Endovascular Embolization Coil, HydroSoft 3D, Ster

Z-1405-2020  ·  Class II  ·  reported 2020-03-04  ·  Terminated

Why it was recalled
The devices may be missing the implant coil.
Product
MicroVention Terumo HydroSoft 10, HydroCoil Embolic System, Endovascular Embolization Coil, HydroSoft 3D, Sterile, Rx, REF numbers 100102HFRM-V, 100152HFRM-V, 100153HFRM-V, 100154HFRM-V, 100202HFRM-V, 100206HFRM-V, 100208HFRM-V, 100254HFRM-V, 100256HFRM-V, 100308HFRM-V, MV-00202HHTA, MV-00203HHTA, MV-00204HHTA, MV-00206HHTA, MV-00208HHTA, MV-00304HHTA, MV-00306HHTA, MV-00308HHTA, MV-00310HHTA, MV-00408HHTA, MV-00412HHTA, MV-00510HHTA, MV-00515HHTA, MV-00612HHTA, MV-00619HHTA, MV-00715HHTA, MV-00728HHTA, MV-00817HHTA, MV-00833HHTA, MV-01502HHTA, MV-01503HHTA, MV-01504HHTA, MV-02504HHTA, MV-02506HHTA, and MV-02508HHTA. Product Usage: The HES is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. The firm name on the label is MicroVention, Inc., Tustin, CA.
Where it went
All distribution was OUS. The countries receiving affected product are: Albania, Austria, Belarus, Belgium, Brazil, Bulgaria, China, Cyprus, Czech Republic, Denmark, Egypt, Finalnd, France, Georgia, Germany, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Malta, Morocco, Netherlands, Norway, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Slovak Republic, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Turkey, Ukraine, United Arab Emirates, and United Kingdom.
How much
847 units

MicroVention Terumo Microplex 18, Platinum Coil System, Endovascular Embolization Coil, Helical 18 Regular, St

Z-1396-2020  ·  Class II  ·  reported 2020-03-04  ·  Terminated

Why it was recalled
The devices may be missing the implant coil.
Product
MicroVention Terumo Microplex 18, Platinum Coil System, Endovascular Embolization Coil, Helical 18 Regular, Sterile, Rx, REF numbers 180520HC-R-V and 181230HC-R-V. Product Usage: The MCS is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. The firm name on the label is MicroVention, Inc., Tustin, CA.
Where it went
All distribution was OUS. The countries receiving affected product are: Albania, Austria, Belarus, Belgium, Brazil, Bulgaria, China, Cyprus, Czech Republic, Denmark, Egypt, Finalnd, France, Georgia, Germany, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Malta, Morocco, Netherlands, Norway, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Slovak Republic, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Turkey, Ukraine, United Arab Emirates, and United Kingdom.
How much
9 units

MicroVention Terumo Microplex 10, Platinum Coil System, Endovascular Embolization Coil, HyperSoft 3D, Sterile,

Z-1393-2020  ·  Class II  ·  reported 2020-03-04  ·  Terminated

Why it was recalled
The devices may be missing the implant coil.
Product
MicroVention Terumo Microplex 10, Platinum Coil System, Endovascular Embolization Coil, HyperSoft 3D, Sterile, Rx, REF numbers 100102HS3D-V, 100103HS3D-V, 100152HS3D-V, 100153HS3D-V, 100154HS3D-V, 100202HS3D-V, 100203HS3D-V, 100204HS3D-V, 100206HS3D-V, 100208HS3D-V, 100254HS3D-V, 100256HS3D-V, 100304HS3D-V, 100306HS3D-V, 100310HS3D-V, 100355HS3D-V, 100358HS3D-V, 100406HS3D-V, 100408HS3D-V, 100412HS3D-V, 100415HS3D-V, 100510HS3D-V, 100515HS3D-V; MV-00102HTDA, MV-00103HTDA, MV-00204HTDA,MV-00304HTDA, MV-00306HTDA, MV-00406HTDA, MV-00408HTDA, MV-00510HTDA, MV-01502HTDA, MV-01503HTDA, MV-01504HTDA; MV-02504HTDA, MV-02506HTDA, MV-03505HTDA, and MV-03508HTDA. Product Usage: The MCS is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. The firm name on the label is MicroVention, Inc., Tustin, CA.
Where it went
All distribution was OUS. The countries receiving affected product are: Albania, Austria, Belarus, Belgium, Brazil, Bulgaria, China, Cyprus, Czech Republic, Denmark, Egypt, Finalnd, France, Georgia, Germany, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Malta, Morocco, Netherlands, Norway, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Slovak Republic, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Turkey, Ukraine, United Arab Emirates, and United Kingdom.
How much
1,725 units

Showing the 30 most recent of 37. The rest are searchable at the FDA enforcement report database.

Read these counts with three things in mind

A firm with more recalls usually makes more products. A contract packer runs hundreds of lines. It will out-recall a company with one product whatever either does. This is not a safety ranking.

Most recalls here are voluntary. A firm found its own fault and pulled the stock. That is the system working. It lands in this database looking the same as a recall the FDA had to push for.

Terminated means the FDA considers the recall finished. It is not a verdict. It does not mean anyone was harmed.

Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.

← All device recalls