Medical Components, Inc dba MedComp: 38 FDA recall reports
Every medical device recall the FDA has published for Medical Components, Inc dba MedComp. The fault, and how the agency graded it.
38Reports
0Class I
1Ongoing
2021-03-17Most recent
What the classes mean
The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.
Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.
By class
By year reported
Checking whether it affects you
The lot or code number is what matters. Not the product name. A recall covers
specific lots. Yours may not be one of them.
Check the code on your package against the report. The FDA keeps the current
list at fda.gov/safety/recalls.
The reports
The text below is quoted from the FDA record word for word.
Power Injectable CT Port insertion kits - Product Usage: is an implantable access device designed to provide r
Z-1184-2021 · Class II · reported 2021-03-17 · Ongoing
- Why it was recalled
- The 5F Dignity CT Port kits were packaged with the incorrect size port. The kits include a 6.6F port instead of the correct 5F port. The inability to connect the catheter lumen to the port stem because of the incorrect size may not be noted until well into the procedure, after the catheter lumen is placed in the vessel. An alternate port will need to be obtained prolonging the procedure. A prolonged insertion procedure increases any risk associated with a surgical procedure.
- Product
- Power Injectable CT Port insertion kits - Product Usage: is an implantable access device designed to provide repeated access to the vascular system.
- Where it went
- US Nationwide distribution in the states of CA, DC, GA, IL, IN, MA, MI, NJ, NM, NY, OH, PA, TX. Updated 3/11/2021
- How much
- 96 pieces total (updated 3/11/2021 additional 72 pieces)
Medcomp 15.SF X 24CM Titan HD CATHETER (CUFF 19CM FROM TIP) Ref: THD155024
Z-2940-2020 · Class II · reported 2020-09-16 · Terminated
- Why it was recalled
- Kits were packaged with the incorrect size valve peelable introducer. The kits include a 15F instead of the correct 16F valve peelable introducer
- Product
- Medcomp 15.SF X 24CM Titan HD CATHETER (CUFF 19CM FROM TIP) Ref: THD155024
- Where it went
- AZ, CA. FL. GA, NC, ND, NJ, NY, OH, PA, VA, WA
- How much
- 132 eaches
C3 Wave App, v. 2.0.5
Z-0222-2020 · Class II · reported 2019-11-06 · Terminated
- Why it was recalled
- When the iPad is updated with Apple iOS software version 12, the C3 application malfunctions. Upon launching a new procedure, the ECG waves do not appear on the screen, in surface or internal mode. If the C3 application is not operating properly, the PICC procedure must be completed without using alternate methods to confirm PICC tip placement.
- Product
- C3 Wave App, v. 2.0.5
- Where it went
- Distributed to accounts in AZ, CA, CO, FL, IL, LA, MA, ME, NC, OH, TN, VA, and WV. Foreign distribution to Canada, Croatia, Czech Republic, Great Britain, Greece, Italy, Netherlands, Slovenia, and Sweden.
- How much
- 24
14F x 24CM SLX Double Lumen Full Tray Hemodialysis catheter insertion kits, Item No. MC061424 Product Usa
Z-2216-2019 · Class II · reported 2019-08-14 · Terminated
- Why it was recalled
- Tray label indicates the kits are packaged with a 15F "valved" peelable introducer, but the kits contain a 15F "non-valved" peelable introducer.
- Product
- 14F x 24CM SLX Double Lumen Full Tray Hemodialysis catheter insertion kits, Item No. MC061424 Product Usage: An aid for introducing a hemodialysis catheter
- Where it went
- US Nationwide Distribution CA FL GA IN KY LA MA MI MO MS NJ NY OH OK PA SC TN TX VA WV.
- How much
- 6830
14F x 20CM SLX Double Lumen Full Tray Hemodialysis catheter insertion kits, Item No. MC061423 Product Usa
Z-2215-2019 · Class II · reported 2019-08-14 · Terminated
- Why it was recalled
- Tray label indicates the kits are packaged with a 15F "valved" peelable introducer, but the kits contain a 15F "non-valved" peelable introducer.
- Product
- 14F x 20CM SLX Double Lumen Full Tray Hemodialysis catheter insertion kits, Item No. MC061423 Product Usage: An aid for introducing a hemodialysis catheter
- Where it went
- US Nationwide Distribution CA FL GA IN KY LA MA MI MO MS NJ NY OH OK PA SC TN TX VA WV.
- How much
- 6830
14F x 30CM SLX Double Lumen Full Tray Hemodialysis catheter insertion kits, Item No. MC061425 Product Usa
Z-2217-2019 · Class II · reported 2019-08-14 · Terminated
- Why it was recalled
- Tray label indicates the kits are packaged with a 15F "valved" peelable introducer, but the kits contain a 15F "non-valved" peelable introducer.
- Product
- 14F x 30CM SLX Double Lumen Full Tray Hemodialysis catheter insertion kits, Item No. MC061425 Product Usage: An aid for introducing a hemodialysis catheter
- Where it went
- US Nationwide Distribution CA FL GA IN KY LA MA MI MO MS NJ NY OH OK PA SC TN TX VA WV.
- How much
- 6830
14F SLX Double Lumen Full Tray Hemodialysis catheter insertion kits, Item No. MC061422
Z-1149-2019 · Class II · reported 2019-04-24 · Terminated
- Why it was recalled
- Tray label indicates the kits are packaged with a 15F "valved" peelable introducer, but the kits contain a 15F "non-valved" peelable introducer.
- Product
- 14F SLX Double Lumen Full Tray Hemodialysis catheter insertion kits, Item No. MC061422
- Where it went
- Distributed to accounts in CA FL GA IN KY LA MA MI MO MS NJ NY OH OK PA SC TN TX VA WV.
- How much
- 6830
SOFT-LINE DUO-FLOW DOUBLE LUMEN CATHETER/CATHETER SET; T94C T94M T96M T98M
Z-2940-2018 · Class II · reported 2018-09-05 · Terminated
- Why it was recalled
- The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.
- Product
- SOFT-LINE DUO-FLOW DOUBLE LUMEN CATHETER/CATHETER SET; T94C T94M T96M T98M
- Where it went
- Nationwide distribution; worldwide distribution.
- How much
- N/A
Duo-Flow XTP Straight Full Set; Hemodialysis catheter insertion kits, AAC02106 AAC02107 AAC02108
Z-2932-2018 · Class II · reported 2018-09-05 · Terminated
- Why it was recalled
- The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.
- Product
- Duo-Flow XTP Straight Full Set; Hemodialysis catheter insertion kits, AAC02106 AAC02107 AAC02108
- Where it went
- Nationwide distribution; worldwide distribution.
- How much
- N/A
Duo-Flow Catheter Full Tray; MCDLT3500 MCDLT4000 MCDLT4400 MCDLT4424# Hemodialysis catheter insertion kits
Z-2936-2018 · Class II · reported 2018-09-05 · Terminated
- Why it was recalled
- The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.
- Product
- Duo-Flow Catheter Full Tray; MCDLT3500 MCDLT4000 MCDLT4400 MCDLT4424# Hemodialysis catheter insertion kits
- Where it went
- Nationwide distribution; worldwide distribution.
- How much
- N/A
Duo-Flow Double Lumen Catheter Only/Catheter Set; T114C T114M T116C T116M T116ME T118C T118M T118ME T119M
Z-2938-2018 · Class II · reported 2018-09-05 · Terminated
- Why it was recalled
- The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.
- Product
- Duo-Flow Double Lumen Catheter Only/Catheter Set; T114C T114M T116C T116M T116ME T118C T118M T118ME T119M
- Where it went
- Nationwide distribution; worldwide distribution.
- How much
- N/A
SOFT-LINE DUO-FLOW DOUBLE LUMEN CATHETER TRAY; T3500 T4000 T4400 T4424
Z-2939-2018 · Class II · reported 2018-09-05 · Terminated
- Why it was recalled
- The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.
- Product
- SOFT-LINE DUO-FLOW DOUBLE LUMEN CATHETER TRAY; T3500 T4000 T4400 T4424
- Where it went
- Nationwide distribution; worldwide distribution.
- How much
- N/A
Double Lumen CRRT/ABP Catheter Set; Hemodialysis catheter insertion kits, DL 11/15 DL 11/20 DL 9/15
Z-2933-2018 · Class II · reported 2018-09-05 · Terminated
- Why it was recalled
- The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.
- Product
- Double Lumen CRRT/ABP Catheter Set; Hemodialysis catheter insertion kits, DL 11/15 DL 11/20 DL 9/15
- Where it went
- Nationwide distribution; worldwide distribution.
- How much
- N/A
Duo-Flow IJ Full Tray; MCDLT114IJ MCDLT116IJ MCDLT118IJ
Z-2935-2018 · Class II · reported 2018-09-05 · Terminated
- Why it was recalled
- The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.
- Product
- Duo-Flow IJ Full Tray; MCDLT114IJ MCDLT116IJ MCDLT118IJ
- Where it went
- Nationwide distribution; worldwide distribution.
- How much
- N/A
DUO FLOW DOUBLE LUMEN CATHETER, XTP114CT, XTP114MT, ; RAULERSON DUO FLOW IJ DOUBLE LUMEN CATHETER/CATHETER SE
Z-2942-2018 · Class II · reported 2018-09-05 · Terminated
- Why it was recalled
- The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.
- Product
- DUO FLOW DOUBLE LUMEN CATHETER, XTP114CT, XTP114MT, ; RAULERSON DUO FLOW IJ DOUBLE LUMEN CATHETER/CATHETER SET, XTP114IJC, XTP114IJS, XTP114IJSE, XTP114MT-C, XTP116IJC, XTP116IJS, XTP116IJSE RAULERSON PRE-CURVED DUO-FLOW IJ DOUBLE LUMEN CATHETER W/SOFT TIP SET, W/MINI SCALPEL, SYRINGE, AND NEEDLE, XTP118IJS-ID; DOUBLE LUMEN CATHETER WITH CURVED EXTENSIONS (CATHETER ONLY), XTP114IJC=, XTP116IJC=, DOUBLE LUMEN IJ CATHETER SET, XTP114IJS=, XTP115IJS=, XTP116IJS=, XTP116IJS=JO, DOUBLE LUMEN CATHETER SET, XTP114MT= DUO-FLOW DOUBLE LUMEN CATHETER, XTP116CT, INTERNAL JUGULAR CATHETER SET, XTP96IJS= (additional product codes/catalog numbers in attached spreadsheet)
- Where it went
- Nationwide distribution; worldwide distribution.
- How much
- N/A
Duo-Flow XTP Straight Full Set, RMS02105 RMS02108
Z-2937-2018 · Class II · reported 2018-09-05 · Terminated
- Why it was recalled
- The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.
- Product
- Duo-Flow XTP Straight Full Set, RMS02105 RMS02108
- Where it went
- Nationwide distribution; worldwide distribution.
- How much
- N/A
DUO FLOW CUSTOM TRAYS; TRAY #414 TRAY #415 TRAY #67
Z-2941-2018 · Class II · reported 2018-09-05 · Terminated
- Why it was recalled
- The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.
- Product
- DUO FLOW CUSTOM TRAYS; TRAY #414 TRAY #415 TRAY #67
- Where it went
- Nationwide distribution; worldwide distribution.
- How much
- N/A
Duo-Flow Double Lumen Soft Tip CRRT/ABP Catheter Set; DL11/24 . Hemodialysis catheter insertion kits
Z-2934-2018 · Class II · reported 2018-09-05 · Terminated
- Why it was recalled
- The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.
- Product
- Duo-Flow Double Lumen Soft Tip CRRT/ABP Catheter Set; DL11/24 . Hemodialysis catheter insertion kits
- Where it went
- Nationwide distribution; worldwide distribution.
- How much
- N/A
8F Plastic Dignity¿ Low Profile CT Port W/Pre-Attached ChronoFlex¿ Polyurethane Catheter
Z-1271-2018 · Class II · reported 2018-04-04 · Terminated
- Why it was recalled
- The kit was packaged with the incorrect guidewire. The label indicates the kit contains a marked guidewire. The kits were packaged with an unmarked guidewire.
- Product
- 8F Plastic Dignity¿ Low Profile CT Port W/Pre-Attached ChronoFlex¿ Polyurethane Catheter
- Where it went
- CA, AR. AL
- How much
- 24 Eaches
12F Tesio CATHETER SET (CUFF 32CM FROM TIP), REF MCTC1247SD, UDI 884908027266 Product Usage: The Medcomp R
Z-1124-2018 · Class II · reported 2018-03-28 · Terminated
- Why it was recalled
- There is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli.
- Product
- 12F Tesio CATHETER SET (CUFF 32CM FROM TIP), REF MCTC1247SD, UDI 884908027266 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female her connectors, clamps, or repairing extensions where there is a minimum of 4.5cm viable extension tubing.
- Where it went
- Worldwide Distribution - US Nationwide in the states of AL, AR, CA, CO, CT, DE, FL, GA, IL, IN, KS, LA, MA, MD, MI, MN, MS, NC, NH, NJ, NM, NV, NY, OH, OK, PA, RI, TN, TX, UT, VA, VT, WA & WI. International: Australia, Austria, Belgium, Canada, Ecuador, France, Germany, Great Britain Greece, Ireland, Israel, Italy, Netherlands, Netherlands, South Africa, Spain Sweden, Turkey & United Arab Emirates.
- How much
- N/A
12F X 40CM TESIO CATHETER SET WITH "Y" ADAPTOR, REF CMD755, UDI 884908005530 Product Usage: The Medcomp Re
Z-1120-2018 · Class II · reported 2018-03-28 · Terminated
- Why it was recalled
- There is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli.
- Product
- 12F X 40CM TESIO CATHETER SET WITH "Y" ADAPTOR, REF CMD755, UDI 884908005530 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female her connectors, clamps, or repairing extensions where there is a minimum of 4.5cm viable extension tubing.
- Where it went
- Worldwide Distribution - US Nationwide in the states of AL, AR, CA, CO, CT, DE, FL, GA, IL, IN, KS, LA, MA, MD, MI, MN, MS, NC, NH, NJ, NM, NV, NY, OH, OK, PA, RI, TN, TX, UT, VA, VT, WA & WI. International: Australia, Austria, Belgium, Canada, Ecuador, France, Germany, Great Britain Greece, Ireland, Israel, Italy, Netherlands, Netherlands, South Africa, Spain Sweden, Turkey & United Arab Emirates.
- How much
- 712 kits
12F X 60CM TESIO CATHETER SET WITH "Y" ADAPTOR, REF CMD757, UDI 884908005554 Product Usage: The Medcomp Re
Z-1121-2018 · Class II · reported 2018-03-28 · Terminated
- Why it was recalled
- There is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli.
- Product
- 12F X 60CM TESIO CATHETER SET WITH "Y" ADAPTOR, REF CMD757, UDI 884908005554 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female her connectors, clamps, or repairing extensions where there is a minimum of 4.5cm viable extension tubing.
- Where it went
- Worldwide Distribution - US Nationwide in the states of AL, AR, CA, CO, CT, DE, FL, GA, IL, IN, KS, LA, MA, MD, MI, MN, MS, NC, NH, NJ, NM, NV, NY, OH, OK, PA, RI, TN, TX, UT, VA, VT, WA & WI. International: Australia, Austria, Belgium, Canada, Ecuador, France, Germany, Great Britain Greece, Ireland, Israel, Italy, Netherlands, Netherlands, South Africa, Spain Sweden, Turkey & United Arab Emirates.
- How much
- 55 kits
12F Tesio CATHETER SET (CUFF 22CM FROM TIP), REF MCTC1235SD, UDI 884908027242 Product Usage: The Medcomp R
Z-1122-2018 · Class II · reported 2018-03-28 · Terminated
- Why it was recalled
- There is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli.
- Product
- 12F Tesio CATHETER SET (CUFF 22CM FROM TIP), REF MCTC1235SD, UDI 884908027242 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female her connectors, clamps, or repairing extensions where there is a minimum of 4.5cm viable extension tubing.
- Where it went
- Worldwide Distribution - US Nationwide in the states of AL, AR, CA, CO, CT, DE, FL, GA, IL, IN, KS, LA, MA, MD, MI, MN, MS, NC, NH, NJ, NM, NV, NY, OH, OK, PA, RI, TN, TX, UT, VA, VT, WA & WI. International: Australia, Austria, Belgium, Canada, Ecuador, France, Germany, Great Britain Greece, Ireland, Israel, Italy, Netherlands, Netherlands, South Africa, Spain Sweden, Turkey & United Arab Emirates.
- How much
- 385 kits
REPAIR KIT, REF RPK-01, UDI 884908039856 Product Usage: The Medcomp Repair Kit is indicated for use in rep
Z-1127-2018 · Class II · reported 2018-03-28 · Terminated
- Why it was recalled
- There is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli.
- Product
- REPAIR KIT, REF RPK-01, UDI 884908039856 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female her connectors, clamps, or repairing extensions where there is a minimum of 4.5cm viable extension tubing.
- Where it went
- Worldwide Distribution - US Nationwide in the states of AL, AR, CA, CO, CT, DE, FL, GA, IL, IN, KS, LA, MA, MD, MI, MN, MS, NC, NH, NJ, NM, NV, NY, OH, OK, PA, RI, TN, TX, UT, VA, VT, WA & WI. International: Australia, Austria, Belgium, Canada, Ecuador, France, Germany, Great Britain Greece, Ireland, Israel, Italy, Netherlands, Netherlands, South Africa, Spain Sweden, Turkey & United Arab Emirates.
- How much
- 1575 kits
CATHETER REPAIR KIT, REF RMS52001, UDI 884908039658 Product Usage: The Medcomp Repair Kit is indicated for
Z-1126-2018 · Class II · reported 2018-03-28 · Terminated
- Why it was recalled
- There is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli.
- Product
- CATHETER REPAIR KIT, REF RMS52001, UDI 884908039658 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female her connectors, clamps, or repairing extensions where there is a minimum of 4.5cm viable extension tubing.
- Where it went
- Worldwide Distribution - US Nationwide in the states of AL, AR, CA, CO, CT, DE, FL, GA, IL, IN, KS, LA, MA, MD, MI, MN, MS, NC, NH, NJ, NM, NV, NY, OH, OK, PA, RI, TN, TX, UT, VA, VT, WA & WI. International: Australia, Austria, Belgium, Canada, Ecuador, France, Germany, Great Britain Greece, Ireland, Israel, Italy, Netherlands, Netherlands, South Africa, Spain Sweden, Turkey & United Arab Emirates.
- How much
- 315 kits
TWO PART ADAPTER, REF MCTLA2000, UDI 884908027310 Product Usage: The Medcomp Repair Kit is indicated for u
Z-1125-2018 · Class II · reported 2018-03-28 · Terminated
- Why it was recalled
- There is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli.
- Product
- TWO PART ADAPTER, REF MCTLA2000, UDI 884908027310 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female her connectors, clamps, or repairing extensions where there is a minimum of 4.5cm viable extension tubing.
- Where it went
- Worldwide Distribution - US Nationwide in the states of AL, AR, CA, CO, CT, DE, FL, GA, IL, IN, KS, LA, MA, MD, MI, MN, MS, NC, NH, NJ, NM, NV, NY, OH, OK, PA, RI, TN, TX, UT, VA, VT, WA & WI. International: Australia, Austria, Belgium, Canada, Ecuador, France, Germany, Great Britain Greece, Ireland, Israel, Italy, Netherlands, Netherlands, South Africa, Spain Sweden, Turkey & United Arab Emirates.
- How much
- 35 kits
CATHETER REPAIR KIT, REF AAC52001, UDI 884908001891 Product Usage: The Medcomp Repair Kit is indicated fo
Z-1118-2018 · Class II · reported 2018-03-28 · Terminated
- Why it was recalled
- There is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli.
- Product
- CATHETER REPAIR KIT, REF AAC52001, UDI 884908001891 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female her connectors, clamps, or repairing extensions where there is a minimum of 4.5cm viable extension tubing.
- Where it went
- Worldwide Distribution - US Nationwide in the states of AL, AR, CA, CO, CT, DE, FL, GA, IL, IN, KS, LA, MA, MD, MI, MN, MS, NC, NH, NJ, NM, NV, NY, OH, OK, PA, RI, TN, TX, UT, VA, VT, WA & WI. International: Australia, Austria, Belgium, Canada, Ecuador, France, Germany, Great Britain Greece, Ireland, Israel, Italy, Netherlands, Netherlands, South Africa, Spain Sweden, Turkey & United Arab Emirates.
- How much
- 93 kits
REPAIR KIT, REF ARA101, UDI 884908083637 Product Usage: The Medcomp Repair Kit is indicated for use in repla
Z-1119-2018 · Class II · reported 2018-03-28 · Terminated
- Why it was recalled
- There is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli.
- Product
- REPAIR KIT, REF ARA101, UDI 884908083637 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female her connectors, clamps, or repairing extensions where there is a minimum of 4.5cm viable extension tubing.
- Where it went
- Worldwide Distribution - US Nationwide in the states of AL, AR, CA, CO, CT, DE, FL, GA, IL, IN, KS, LA, MA, MD, MI, MN, MS, NC, NH, NJ, NM, NV, NY, OH, OK, PA, RI, TN, TX, UT, VA, VT, WA & WI. International: Australia, Austria, Belgium, Canada, Ecuador, France, Germany, Great Britain Greece, Ireland, Israel, Italy, Netherlands, Netherlands, South Africa, Spain Sweden, Turkey & United Arab Emirates.
- How much
- 5 kits
12F MODIFIED Tesio CATHETER SET, REF MCTC1235SM, UDI 884908027259 Product Usage: The Medcomp Repair Kit is
Z-1123-2018 · Class II · reported 2018-03-28 · Terminated
- Why it was recalled
- There is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli.
- Product
- 12F MODIFIED Tesio CATHETER SET, REF MCTC1235SM, UDI 884908027259 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female her connectors, clamps, or repairing extensions where there is a minimum of 4.5cm viable extension tubing.
- Where it went
- Worldwide Distribution - US Nationwide in the states of AL, AR, CA, CO, CT, DE, FL, GA, IL, IN, KS, LA, MA, MD, MI, MN, MS, NC, NH, NJ, NM, NV, NY, OH, OK, PA, RI, TN, TX, UT, VA, VT, WA & WI. International: Australia, Austria, Belgium, Canada, Ecuador, France, Germany, Great Britain Greece, Ireland, Israel, Italy, Netherlands, Netherlands, South Africa, Spain Sweden, Turkey & United Arab Emirates.
- How much
- 40 kits
6.6F Plastic Dignity¿ Low Profile CT Port W/Attachable ChronoFlex¿ Polyurethane Catheter, Catalog # MRCTI66041
Z-0533-2018 · Class II · reported 2018-02-14 · Terminated
- Why it was recalled
- The kit was packaged with the incorrect introducer needle size. The label indicates the kit contains an 18G Needle w/Echo Tip. The kits were packaged with a 21G Needle w/Echo Tip.
- Product
- 6.6F Plastic Dignity¿ Low Profile CT Port W/Attachable ChronoFlex¿ Polyurethane Catheter, Catalog # MRCTI66041, UDI#- 884908031157 Product Usage: Indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples. When used with a power injectable needle, the Power Injectable Implantable Infusion Port device is indicated for power injection of contrast media.
- Where it went
- Arabia, Australia, Israel, Brunei & Jamaica
- How much
- 65 units
Showing the 30 most recent of 38. The rest are searchable at the FDA enforcement report database.
Read these counts with three things in mind
A firm with more recalls usually makes more products. A contract packer runs
hundreds of lines. It will out-recall a company with one product whatever either
does. This is not a safety ranking.
Most recalls here are voluntary. A firm found its own fault and pulled the stock.
That is the system working. It lands in this database looking the same as a recall
the FDA had to push for.
Terminated means the FDA considers the recall finished. It is not a verdict.
It does not mean anyone was harmed.
Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.