Medical Components, Inc dba MedComp

Medical Components, Inc dba MedComp: 38 FDA recall reports

Every medical device recall the FDA has published for Medical Components, Inc dba MedComp. The fault, and how the agency graded it.

38Reports
0Class I
1Ongoing
2021-03-17Most recent

What the classes mean

The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.

ClassWhat the FDA is saying
Class IThere is a reasonable probability that using this will cause serious harm or death.
Class IIIt may cause temporary or reversible harm. Serious harm is unlikely.
Class IIIIt breaks a rule but is unlikely to hurt anyone. Often a labelling fault.

Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.

By class

ClassReports
Class II38

By year reported

YearReports
20121
20131
20141
20153
20161
20171
201823
20195
20201
20211
Checking whether it affects you

The lot or code number is what matters. Not the product name. A recall covers specific lots. Yours may not be one of them.

Check the code on your package against the report. The FDA keeps the current list at fda.gov/safety/recalls.

The reports

The text below is quoted from the FDA record word for word.

Power Injectable CT Port insertion kits - Product Usage: is an implantable access device designed to provide r

Z-1184-2021  ·  Class II  ·  reported 2021-03-17  ·  Ongoing

Why it was recalled
The 5F Dignity CT Port kits were packaged with the incorrect size port. The kits include a 6.6F port instead of the correct 5F port. The inability to connect the catheter lumen to the port stem because of the incorrect size may not be noted until well into the procedure, after the catheter lumen is placed in the vessel. An alternate port will need to be obtained prolonging the procedure. A prolonged insertion procedure increases any risk associated with a surgical procedure.
Product
Power Injectable CT Port insertion kits - Product Usage: is an implantable access device designed to provide repeated access to the vascular system.
Where it went
US Nationwide distribution in the states of CA, DC, GA, IL, IN, MA, MI, NJ, NM, NY, OH, PA, TX. Updated 3/11/2021
How much
96 pieces total (updated 3/11/2021 additional 72 pieces)

Medcomp 15.SF X 24CM Titan HD CATHETER (CUFF 19CM FROM TIP) Ref: THD155024

Z-2940-2020  ·  Class II  ·  reported 2020-09-16  ·  Terminated

Why it was recalled
Kits were packaged with the incorrect size valve peelable introducer. The kits include a 15F instead of the correct 16F valve peelable introducer
Product
Medcomp 15.SF X 24CM Titan HD CATHETER (CUFF 19CM FROM TIP) Ref: THD155024
Where it went
AZ, CA. FL. GA, NC, ND, NJ, NY, OH, PA, VA, WA
How much
132 eaches

C3 Wave App, v. 2.0.5

Z-0222-2020  ·  Class II  ·  reported 2019-11-06  ·  Terminated

Why it was recalled
When the iPad is updated with Apple iOS software version 12, the C3 application malfunctions. Upon launching a new procedure, the ECG waves do not appear on the screen, in surface or internal mode. If the C3 application is not operating properly, the PICC procedure must be completed without using alternate methods to confirm PICC tip placement.
Product
C3 Wave App, v. 2.0.5
Where it went
Distributed to accounts in AZ, CA, CO, FL, IL, LA, MA, ME, NC, OH, TN, VA, and WV. Foreign distribution to Canada, Croatia, Czech Republic, Great Britain, Greece, Italy, Netherlands, Slovenia, and Sweden.
How much
24

14F x 24CM SLX Double Lumen Full Tray Hemodialysis catheter insertion kits, Item No. MC061424 Product Usa

Z-2216-2019  ·  Class II  ·  reported 2019-08-14  ·  Terminated

Why it was recalled
Tray label indicates the kits are packaged with a 15F "valved" peelable introducer, but the kits contain a 15F "non-valved" peelable introducer.
Product
14F x 24CM SLX Double Lumen Full Tray Hemodialysis catheter insertion kits, Item No. MC061424 Product Usage: An aid for introducing a hemodialysis catheter
Where it went
US Nationwide Distribution CA FL GA IN KY LA MA MI MO MS NJ NY OH OK PA SC TN TX VA WV.
How much
6830

14F x 20CM SLX Double Lumen Full Tray Hemodialysis catheter insertion kits, Item No. MC061423 Product Usa

Z-2215-2019  ·  Class II  ·  reported 2019-08-14  ·  Terminated

Why it was recalled
Tray label indicates the kits are packaged with a 15F "valved" peelable introducer, but the kits contain a 15F "non-valved" peelable introducer.
Product
14F x 20CM SLX Double Lumen Full Tray Hemodialysis catheter insertion kits, Item No. MC061423 Product Usage: An aid for introducing a hemodialysis catheter
Where it went
US Nationwide Distribution CA FL GA IN KY LA MA MI MO MS NJ NY OH OK PA SC TN TX VA WV.
How much
6830

14F x 30CM SLX Double Lumen Full Tray Hemodialysis catheter insertion kits, Item No. MC061425 Product Usa

Z-2217-2019  ·  Class II  ·  reported 2019-08-14  ·  Terminated

Why it was recalled
Tray label indicates the kits are packaged with a 15F "valved" peelable introducer, but the kits contain a 15F "non-valved" peelable introducer.
Product
14F x 30CM SLX Double Lumen Full Tray Hemodialysis catheter insertion kits, Item No. MC061425 Product Usage: An aid for introducing a hemodialysis catheter
Where it went
US Nationwide Distribution CA FL GA IN KY LA MA MI MO MS NJ NY OH OK PA SC TN TX VA WV.
How much
6830

14F SLX Double Lumen Full Tray Hemodialysis catheter insertion kits, Item No. MC061422

Z-1149-2019  ·  Class II  ·  reported 2019-04-24  ·  Terminated

Why it was recalled
Tray label indicates the kits are packaged with a 15F "valved" peelable introducer, but the kits contain a 15F "non-valved" peelable introducer.
Product
14F SLX Double Lumen Full Tray Hemodialysis catheter insertion kits, Item No. MC061422
Where it went
Distributed to accounts in CA FL GA IN KY LA MA MI MO MS NJ NY OH OK PA SC TN TX VA WV.
How much
6830

SOFT-LINE DUO-FLOW DOUBLE LUMEN CATHETER/CATHETER SET; T94C T94M T96M T98M

Z-2940-2018  ·  Class II  ·  reported 2018-09-05  ·  Terminated

Why it was recalled
The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.
Product
SOFT-LINE DUO-FLOW DOUBLE LUMEN CATHETER/CATHETER SET; T94C T94M T96M T98M
Where it went
Nationwide distribution; worldwide distribution.
How much
N/A

Duo-Flow XTP Straight Full Set; Hemodialysis catheter insertion kits, AAC02106 AAC02107 AAC02108

Z-2932-2018  ·  Class II  ·  reported 2018-09-05  ·  Terminated

Why it was recalled
The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.
Product
Duo-Flow XTP Straight Full Set; Hemodialysis catheter insertion kits, AAC02106 AAC02107 AAC02108
Where it went
Nationwide distribution; worldwide distribution.
How much
N/A

Duo-Flow Catheter Full Tray; MCDLT3500 MCDLT4000 MCDLT4400 MCDLT4424# Hemodialysis catheter insertion kits

Z-2936-2018  ·  Class II  ·  reported 2018-09-05  ·  Terminated

Why it was recalled
The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.
Product
Duo-Flow Catheter Full Tray; MCDLT3500 MCDLT4000 MCDLT4400 MCDLT4424# Hemodialysis catheter insertion kits
Where it went
Nationwide distribution; worldwide distribution.
How much
N/A

Duo-Flow Double Lumen Catheter Only/Catheter Set; T114C T114M T116C T116M T116ME T118C T118M T118ME T119M

Z-2938-2018  ·  Class II  ·  reported 2018-09-05  ·  Terminated

Why it was recalled
The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.
Product
Duo-Flow Double Lumen Catheter Only/Catheter Set; T114C T114M T116C T116M T116ME T118C T118M T118ME T119M
Where it went
Nationwide distribution; worldwide distribution.
How much
N/A

SOFT-LINE DUO-FLOW DOUBLE LUMEN CATHETER TRAY; T3500 T4000 T4400 T4424

Z-2939-2018  ·  Class II  ·  reported 2018-09-05  ·  Terminated

Why it was recalled
The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.
Product
SOFT-LINE DUO-FLOW DOUBLE LUMEN CATHETER TRAY; T3500 T4000 T4400 T4424
Where it went
Nationwide distribution; worldwide distribution.
How much
N/A

Double Lumen CRRT/ABP Catheter Set; Hemodialysis catheter insertion kits, DL 11/15 DL 11/20 DL 9/15

Z-2933-2018  ·  Class II  ·  reported 2018-09-05  ·  Terminated

Why it was recalled
The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.
Product
Double Lumen CRRT/ABP Catheter Set; Hemodialysis catheter insertion kits, DL 11/15 DL 11/20 DL 9/15
Where it went
Nationwide distribution; worldwide distribution.
How much
N/A

Duo-Flow IJ Full Tray; MCDLT114IJ MCDLT116IJ MCDLT118IJ

Z-2935-2018  ·  Class II  ·  reported 2018-09-05  ·  Terminated

Why it was recalled
The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.
Product
Duo-Flow IJ Full Tray; MCDLT114IJ MCDLT116IJ MCDLT118IJ
Where it went
Nationwide distribution; worldwide distribution.
How much
N/A

DUO FLOW DOUBLE LUMEN CATHETER, XTP114CT, XTP114MT, ; RAULERSON DUO FLOW IJ DOUBLE LUMEN CATHETER/CATHETER SE

Z-2942-2018  ·  Class II  ·  reported 2018-09-05  ·  Terminated

Why it was recalled
The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.
Product
DUO FLOW DOUBLE LUMEN CATHETER, XTP114CT, XTP114MT, ; RAULERSON DUO FLOW IJ DOUBLE LUMEN CATHETER/CATHETER SET, XTP114IJC, XTP114IJS, XTP114IJSE, XTP114MT-C, XTP116IJC, XTP116IJS, XTP116IJSE RAULERSON PRE-CURVED DUO-FLOW IJ DOUBLE LUMEN CATHETER W/SOFT TIP SET, W/MINI SCALPEL, SYRINGE, AND NEEDLE, XTP118IJS-ID; DOUBLE LUMEN CATHETER WITH CURVED EXTENSIONS (CATHETER ONLY), XTP114IJC=, XTP116IJC=, DOUBLE LUMEN IJ CATHETER SET, XTP114IJS=, XTP115IJS=, XTP116IJS=, XTP116IJS=JO, DOUBLE LUMEN CATHETER SET, XTP114MT= DUO-FLOW DOUBLE LUMEN CATHETER, XTP116CT, INTERNAL JUGULAR CATHETER SET, XTP96IJS= (additional product codes/catalog numbers in attached spreadsheet)
Where it went
Nationwide distribution; worldwide distribution.
How much
N/A

Duo-Flow XTP Straight Full Set, RMS02105 RMS02108

Z-2937-2018  ·  Class II  ·  reported 2018-09-05  ·  Terminated

Why it was recalled
The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.
Product
Duo-Flow XTP Straight Full Set, RMS02105 RMS02108
Where it went
Nationwide distribution; worldwide distribution.
How much
N/A

DUO FLOW CUSTOM TRAYS; TRAY #414 TRAY #415 TRAY #67

Z-2941-2018  ·  Class II  ·  reported 2018-09-05  ·  Terminated

Why it was recalled
The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.
Product
DUO FLOW CUSTOM TRAYS; TRAY #414 TRAY #415 TRAY #67
Where it went
Nationwide distribution; worldwide distribution.
How much
N/A

Duo-Flow Double Lumen Soft Tip CRRT/ABP Catheter Set; DL11/24 . Hemodialysis catheter insertion kits

Z-2934-2018  ·  Class II  ·  reported 2018-09-05  ·  Terminated

Why it was recalled
The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.
Product
Duo-Flow Double Lumen Soft Tip CRRT/ABP Catheter Set; DL11/24 . Hemodialysis catheter insertion kits
Where it went
Nationwide distribution; worldwide distribution.
How much
N/A

8F Plastic Dignity¿ Low Profile CT Port W/Pre-Attached ChronoFlex¿ Polyurethane Catheter

Z-1271-2018  ·  Class II  ·  reported 2018-04-04  ·  Terminated

Why it was recalled
The kit was packaged with the incorrect guidewire. The label indicates the kit contains a marked guidewire. The kits were packaged with an unmarked guidewire.
Product
8F Plastic Dignity¿ Low Profile CT Port W/Pre-Attached ChronoFlex¿ Polyurethane Catheter
Where it went
CA, AR. AL
How much
24 Eaches

12F Tesio CATHETER SET (CUFF 32CM FROM TIP), REF MCTC1247SD, UDI 884908027266 Product Usage: The Medcomp R

Z-1124-2018  ·  Class II  ·  reported 2018-03-28  ·  Terminated

Why it was recalled
There is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli.
Product
12F Tesio CATHETER SET (CUFF 32CM FROM TIP), REF MCTC1247SD, UDI 884908027266 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female her connectors, clamps, or repairing extensions where there is a minimum of 4.5cm viable extension tubing.
Where it went
Worldwide Distribution - US Nationwide in the states of AL, AR, CA, CO, CT, DE, FL, GA, IL, IN, KS, LA, MA, MD, MI, MN, MS, NC, NH, NJ, NM, NV, NY, OH, OK, PA, RI, TN, TX, UT, VA, VT, WA & WI. International: Australia, Austria, Belgium, Canada, Ecuador, France, Germany, Great Britain Greece, Ireland, Israel, Italy, Netherlands, Netherlands, South Africa, Spain Sweden, Turkey & United Arab Emirates.
How much
N/A

12F X 40CM TESIO CATHETER SET WITH "Y" ADAPTOR, REF CMD755, UDI 884908005530 Product Usage: The Medcomp Re

Z-1120-2018  ·  Class II  ·  reported 2018-03-28  ·  Terminated

Why it was recalled
There is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli.
Product
12F X 40CM TESIO CATHETER SET WITH "Y" ADAPTOR, REF CMD755, UDI 884908005530 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female her connectors, clamps, or repairing extensions where there is a minimum of 4.5cm viable extension tubing.
Where it went
Worldwide Distribution - US Nationwide in the states of AL, AR, CA, CO, CT, DE, FL, GA, IL, IN, KS, LA, MA, MD, MI, MN, MS, NC, NH, NJ, NM, NV, NY, OH, OK, PA, RI, TN, TX, UT, VA, VT, WA & WI. International: Australia, Austria, Belgium, Canada, Ecuador, France, Germany, Great Britain Greece, Ireland, Israel, Italy, Netherlands, Netherlands, South Africa, Spain Sweden, Turkey & United Arab Emirates.
How much
712 kits

12F X 60CM TESIO CATHETER SET WITH "Y" ADAPTOR, REF CMD757, UDI 884908005554 Product Usage: The Medcomp Re

Z-1121-2018  ·  Class II  ·  reported 2018-03-28  ·  Terminated

Why it was recalled
There is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli.
Product
12F X 60CM TESIO CATHETER SET WITH "Y" ADAPTOR, REF CMD757, UDI 884908005554 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female her connectors, clamps, or repairing extensions where there is a minimum of 4.5cm viable extension tubing.
Where it went
Worldwide Distribution - US Nationwide in the states of AL, AR, CA, CO, CT, DE, FL, GA, IL, IN, KS, LA, MA, MD, MI, MN, MS, NC, NH, NJ, NM, NV, NY, OH, OK, PA, RI, TN, TX, UT, VA, VT, WA & WI. International: Australia, Austria, Belgium, Canada, Ecuador, France, Germany, Great Britain Greece, Ireland, Israel, Italy, Netherlands, Netherlands, South Africa, Spain Sweden, Turkey & United Arab Emirates.
How much
55 kits

12F Tesio CATHETER SET (CUFF 22CM FROM TIP), REF MCTC1235SD, UDI 884908027242 Product Usage: The Medcomp R

Z-1122-2018  ·  Class II  ·  reported 2018-03-28  ·  Terminated

Why it was recalled
There is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli.
Product
12F Tesio CATHETER SET (CUFF 22CM FROM TIP), REF MCTC1235SD, UDI 884908027242 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female her connectors, clamps, or repairing extensions where there is a minimum of 4.5cm viable extension tubing.
Where it went
Worldwide Distribution - US Nationwide in the states of AL, AR, CA, CO, CT, DE, FL, GA, IL, IN, KS, LA, MA, MD, MI, MN, MS, NC, NH, NJ, NM, NV, NY, OH, OK, PA, RI, TN, TX, UT, VA, VT, WA & WI. International: Australia, Austria, Belgium, Canada, Ecuador, France, Germany, Great Britain Greece, Ireland, Israel, Italy, Netherlands, Netherlands, South Africa, Spain Sweden, Turkey & United Arab Emirates.
How much
385 kits

REPAIR KIT, REF RPK-01, UDI 884908039856 Product Usage: The Medcomp Repair Kit is indicated for use in rep

Z-1127-2018  ·  Class II  ·  reported 2018-03-28  ·  Terminated

Why it was recalled
There is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli.
Product
REPAIR KIT, REF RPK-01, UDI 884908039856 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female her connectors, clamps, or repairing extensions where there is a minimum of 4.5cm viable extension tubing.
Where it went
Worldwide Distribution - US Nationwide in the states of AL, AR, CA, CO, CT, DE, FL, GA, IL, IN, KS, LA, MA, MD, MI, MN, MS, NC, NH, NJ, NM, NV, NY, OH, OK, PA, RI, TN, TX, UT, VA, VT, WA & WI. International: Australia, Austria, Belgium, Canada, Ecuador, France, Germany, Great Britain Greece, Ireland, Israel, Italy, Netherlands, Netherlands, South Africa, Spain Sweden, Turkey & United Arab Emirates.
How much
1575 kits

CATHETER REPAIR KIT, REF RMS52001, UDI 884908039658 Product Usage: The Medcomp Repair Kit is indicated for

Z-1126-2018  ·  Class II  ·  reported 2018-03-28  ·  Terminated

Why it was recalled
There is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli.
Product
CATHETER REPAIR KIT, REF RMS52001, UDI 884908039658 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female her connectors, clamps, or repairing extensions where there is a minimum of 4.5cm viable extension tubing.
Where it went
Worldwide Distribution - US Nationwide in the states of AL, AR, CA, CO, CT, DE, FL, GA, IL, IN, KS, LA, MA, MD, MI, MN, MS, NC, NH, NJ, NM, NV, NY, OH, OK, PA, RI, TN, TX, UT, VA, VT, WA & WI. International: Australia, Austria, Belgium, Canada, Ecuador, France, Germany, Great Britain Greece, Ireland, Israel, Italy, Netherlands, Netherlands, South Africa, Spain Sweden, Turkey & United Arab Emirates.
How much
315 kits

TWO PART ADAPTER, REF MCTLA2000, UDI 884908027310 Product Usage: The Medcomp Repair Kit is indicated for u

Z-1125-2018  ·  Class II  ·  reported 2018-03-28  ·  Terminated

Why it was recalled
There is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli.
Product
TWO PART ADAPTER, REF MCTLA2000, UDI 884908027310 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female her connectors, clamps, or repairing extensions where there is a minimum of 4.5cm viable extension tubing.
Where it went
Worldwide Distribution - US Nationwide in the states of AL, AR, CA, CO, CT, DE, FL, GA, IL, IN, KS, LA, MA, MD, MI, MN, MS, NC, NH, NJ, NM, NV, NY, OH, OK, PA, RI, TN, TX, UT, VA, VT, WA & WI. International: Australia, Austria, Belgium, Canada, Ecuador, France, Germany, Great Britain Greece, Ireland, Israel, Italy, Netherlands, Netherlands, South Africa, Spain Sweden, Turkey & United Arab Emirates.
How much
35 kits

CATHETER REPAIR KIT, REF AAC52001, UDI 884908001891 Product Usage: The Medcomp Repair Kit is indicated fo

Z-1118-2018  ·  Class II  ·  reported 2018-03-28  ·  Terminated

Why it was recalled
There is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli.
Product
CATHETER REPAIR KIT, REF AAC52001, UDI 884908001891 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female her connectors, clamps, or repairing extensions where there is a minimum of 4.5cm viable extension tubing.
Where it went
Worldwide Distribution - US Nationwide in the states of AL, AR, CA, CO, CT, DE, FL, GA, IL, IN, KS, LA, MA, MD, MI, MN, MS, NC, NH, NJ, NM, NV, NY, OH, OK, PA, RI, TN, TX, UT, VA, VT, WA & WI. International: Australia, Austria, Belgium, Canada, Ecuador, France, Germany, Great Britain Greece, Ireland, Israel, Italy, Netherlands, Netherlands, South Africa, Spain Sweden, Turkey & United Arab Emirates.
How much
93 kits

REPAIR KIT, REF ARA101, UDI 884908083637 Product Usage: The Medcomp Repair Kit is indicated for use in repla

Z-1119-2018  ·  Class II  ·  reported 2018-03-28  ·  Terminated

Why it was recalled
There is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli.
Product
REPAIR KIT, REF ARA101, UDI 884908083637 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female her connectors, clamps, or repairing extensions where there is a minimum of 4.5cm viable extension tubing.
Where it went
Worldwide Distribution - US Nationwide in the states of AL, AR, CA, CO, CT, DE, FL, GA, IL, IN, KS, LA, MA, MD, MI, MN, MS, NC, NH, NJ, NM, NV, NY, OH, OK, PA, RI, TN, TX, UT, VA, VT, WA & WI. International: Australia, Austria, Belgium, Canada, Ecuador, France, Germany, Great Britain Greece, Ireland, Israel, Italy, Netherlands, Netherlands, South Africa, Spain Sweden, Turkey & United Arab Emirates.
How much
5 kits

12F MODIFIED Tesio CATHETER SET, REF MCTC1235SM, UDI 884908027259 Product Usage: The Medcomp Repair Kit is

Z-1123-2018  ·  Class II  ·  reported 2018-03-28  ·  Terminated

Why it was recalled
There is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli.
Product
12F MODIFIED Tesio CATHETER SET, REF MCTC1235SM, UDI 884908027259 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female her connectors, clamps, or repairing extensions where there is a minimum of 4.5cm viable extension tubing.
Where it went
Worldwide Distribution - US Nationwide in the states of AL, AR, CA, CO, CT, DE, FL, GA, IL, IN, KS, LA, MA, MD, MI, MN, MS, NC, NH, NJ, NM, NV, NY, OH, OK, PA, RI, TN, TX, UT, VA, VT, WA & WI. International: Australia, Austria, Belgium, Canada, Ecuador, France, Germany, Great Britain Greece, Ireland, Israel, Italy, Netherlands, Netherlands, South Africa, Spain Sweden, Turkey & United Arab Emirates.
How much
40 kits

6.6F Plastic Dignity¿ Low Profile CT Port W/Attachable ChronoFlex¿ Polyurethane Catheter, Catalog # MRCTI66041

Z-0533-2018  ·  Class II  ·  reported 2018-02-14  ·  Terminated

Why it was recalled
The kit was packaged with the incorrect introducer needle size. The label indicates the kit contains an 18G Needle w/Echo Tip. The kits were packaged with a 21G Needle w/Echo Tip.
Product
6.6F Plastic Dignity¿ Low Profile CT Port W/Attachable ChronoFlex¿ Polyurethane Catheter, Catalog # MRCTI66041, UDI#- 884908031157 Product Usage: Indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples. When used with a power injectable needle, the Power Injectable Implantable Infusion Port device is indicated for power injection of contrast media.
Where it went
Arabia, Australia, Israel, Brunei & Jamaica
How much
65 units

Showing the 30 most recent of 38. The rest are searchable at the FDA enforcement report database.

Read these counts with three things in mind

A firm with more recalls usually makes more products. A contract packer runs hundreds of lines. It will out-recall a company with one product whatever either does. This is not a safety ranking.

Most recalls here are voluntary. A firm found its own fault and pulled the stock. That is the system working. It lands in this database looking the same as a recall the FDA had to push for.

Terminated means the FDA considers the recall finished. It is not a verdict. It does not mean anyone was harmed.

Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.