Maquet Cardiovascular, LLC

Maquet Cardiovascular, LLC: 252 FDA recall reports

Every medical device recall the FDA has published for Maquet Cardiovascular, LLC. The fault, and how the agency graded it.

252Reports
5Class I
35Ongoing
2026-07-22Most recent
5 Class I on this page

Class I means the FDA judged there was a reasonable probability of serious harm or death. Check the lot numbers against anything you hold.

What the classes mean

The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.

ClassWhat the FDA is saying
Class IThere is a reasonable probability that using this will cause serious harm or death.
Class IIIt may cause temporary or reversible harm. Serious harm is unlikely.
Class IIIIt breaks a rule but is unlikely to hurt anyone. Often a labelling fault.

Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.

By class

ClassReports
Class II247
Class I5

By year reported

YearReports
20134
20141
20171
2019184
202024
20211
20223
202313
202412
20258
20261
Checking whether it affects you

The lot or code number is what matters. Not the product name. A recall covers specific lots. Yours may not be one of them.

Check the code on your package against the report. The FDA keeps the current list at fda.gov/safety/recalls.

The reports

The text below is quoted from the FDA record word for word.

The Vasoview Hemopro 2 Endoscopic Vessel Harvesting System (Model VH-4000; UDI 00607567700406) is a sterile,

Z-2741-2026  ·  Class II  ·  reported 2026-07-22  ·  Ongoing

Why it was recalled
Getinge identified that Vasoview Hemopro 2 devices from Lot 3000547797 were exposed to an out-of-specification sterilization dose during a gamma sterilization run (Run ID 1183-28782A) performed by STERIS at its Whippany, NJ facility between March 31 and April 1 of 2026.
Product
The Vasoview Hemopro 2 Endoscopic Vessel Harvesting System (Model VH-4000; UDI 00607567700406) is a sterile, single-use Class II device that was cleared for marketing by FDA under 510(k) number K101274.
Where it went
US Nationwide distribution in the states of NY, CA, SC, FL, TX, GA, OR.
How much
205 units

Vasoview Hemopro 3 Endoscopic Vessel Harvesting System Model No. C-VH-6000. The VASOVIEW HEMOPRO 3 Endoscopi

Z-2636-2025  ·  Class II  ·  reported 2025-10-01  ·  Ongoing

Why it was recalled
Reports of out-of-box failures discovered during the limited launch phase. Failure modes included angled deformity at the attachment site of the Harvesting Tool Jaws and the distal end of the Harvesting Tool Shaft, jaws not fully closed out of box, and heater wire lifted/flexed away from jaws.
Product
Vasoview Hemopro 3 Endoscopic Vessel Harvesting System Model No. C-VH-6000. The VASOVIEW HEMOPRO 3 Endoscopic Vessel Harvesting (EVH) System (VH-6000, VH-6001, and VH-6010) is a vessel harvesting kit designed for use with a 7 mm endoscope to cut and cauterize vessel branches. The HP3 Harvesting Tool features two curved, insulated jaws with heating elements that operate through heat and pressure rather than radiofrequency energy.
Where it went
US Nationwide distribution in the states of FL, IL, MD & TX.
How much
98 units

Heartstring III Proximal Seal System. Intravascular anastomosis occluder.

Z-2586-2025  ·  Class II  ·  reported 2025-09-24  ·  Ongoing

Why it was recalled
Three failure modes have been identified: failure of the Heartstring Seal to load, failure of the Heartstring Seal to deploy into the aortotomy, and failure of the deployed Heartstring Seal to provide adequate hemostasis.
Product
Heartstring III Proximal Seal System. Intravascular anastomosis occluder.
Where it went
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Canada, China, Colombia, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Israel, Italy, Japan, Luxembourg, Malaysia, Netherlands, New Zealand, Norway, Poland, Portugal, Saudi Arabia, Singapore, Slovakia, Slovenia, South Korea, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom.
How much
126 units (91 US, 35 OUS)

Heartstring III Proximal Seal System, 3.8 MM. Intravascular anastomosis occluder.

Z-2587-2025  ·  Class II  ·  reported 2025-09-24  ·  Ongoing

Why it was recalled
Three failure modes have been identified: failure of the Heartstring Seal to load, failure of the Heartstring Seal to deploy into the aortotomy, and failure of the deployed Heartstring Seal to provide adequate hemostasis.
Product
Heartstring III Proximal Seal System, 3.8 MM. Intravascular anastomosis occluder.
Where it went
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Canada, China, Colombia, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Israel, Italy, Japan, Luxembourg, Malaysia, Netherlands, New Zealand, Norway, Poland, Portugal, Saudi Arabia, Singapore, Slovakia, Slovenia, South Korea, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom.
How much
28,331 units (16,615 US, 11,716 OUS)

Heartstring III Proximal Seal System, 4.3 MM. Intravascular anastomosis occluder.

Z-2588-2025  ·  Class II  ·  reported 2025-09-24  ·  Ongoing

Why it was recalled
Three failure modes have been identified: failure of the Heartstring Seal to load, failure of the Heartstring Seal to deploy into the aortotomy, and failure of the deployed Heartstring Seal to provide adequate hemostasis.
Product
Heartstring III Proximal Seal System, 4.3 MM. Intravascular anastomosis occluder.
Where it went
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Canada, China, Colombia, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Israel, Italy, Japan, Luxembourg, Malaysia, Netherlands, New Zealand, Norway, Poland, Portugal, Saudi Arabia, Singapore, Slovakia, Slovenia, South Korea, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom.
How much
12,471 units (6,565 US, 5,906 OUS)

The VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting System is designed for use in conjunction with the 7 mm ex

Z-1622-2025  ·  Class II  ·  reported 2025-04-30  ·  Ongoing

Why it was recalled
There have been at least 12 complaints from users as of March 3, 2025, reporting that they identified the ceramic C-Ring to be broken in half during use of the device. The ceramic C-Ring has a potential to break while in use. Maquet/Getinge determined that the issue was limited to the units of VH-4000 and VH-4001 manufactured with ceramic C-Ring, and the ceramic C-Ring was implemented in the Hemopro 2 devices from November 5, 2024, through February 14, 2025.
Product
The VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting System is designed for use in conjunction with the 7 mm extended length endoscope. The Harvesting Cannula has four lumens which house the endoscope, C-Ring, distal lens washer tube, and Harvesting Tool for cutting and cauterizing vessel branches. The Harvesting Tool has two curved jaws. One of the jaws contains the heating elements for branch cutting and cauterization and spot cautery. An area near the tip of the convex side of the jaw is used for spot cautery. The activation toggle is used to control the jaws to activate the heating elements. Positioning the device, cutting, and cauterization are performed under endoscopic visualization. The device is intended for specific use with the VASOVIEW HEMOPRO Power Supply, VASOVIEW HEMOPRO 2 extension cable, and an adapter cable.
Where it went
Worldwide distribution - US Nationwide and the countries of Mexico, United Arab Emirates, Singapore, India, Australia, Canada, Brazil, Hong Kong, Japan, South Korea, New Zealand.
How much
22,244 units

Maquet VH-3010 Power Supply. Used to deliver power to the harvesting tool of the Vasoview Hemopro EVH System (

Z-1577-2025  ·  Class II  ·  reported 2025-04-23  ·  Ongoing

Why it was recalled
Incorrect resistor utilized in the VH-3010 Power Supply, which may cause the jaws of the HemoPro 2 to not adequately heat up.
Product
Maquet VH-3010 Power Supply. Used to deliver power to the harvesting tool of the Vasoview Hemopro EVH System (HemoPro 2).
Where it went
Worldwide - US Nationwide distribution in the states of AL, CA, CO, DC, FL, GA, IA, ID, IL, IN, KY, MD, MI, MN, MO, NC, NH, NJ, NY, OH, PA, PR, TX, UT, VA, WI, WV and the countries of China, Japan, Netherlands, New Zealand, Singapore, United Kingdom, United States.
How much
110 units
Class I

VasoView HemoPro 2 Endoscopic Vessel Harvesting System. Model Numbers: VH-4000. Indicated for use in minimall

Z-0711-2025  ·  Class I  ·  reported 2025-01-15  ·  Ongoing

Why it was recalled
Potential for two issues: 1. Bent or detached heater wire; 2. Silicone peeling or detaching from the Jaws of the Harvesting Tool
Product
VasoView HemoPro 2 Endoscopic Vessel Harvesting System. Model Numbers: VH-4000. Indicated for use in minimally invasive surgery allowing access for vessel harvesting.
Where it went
Nationwide distribution. International distribution to Australia, Austria, Bahrain, Bangladesh, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Guatemala, Hong Kong, India, Iran, Iraq, Ireland, Italy, Japan, Jordan, Kenya, Kuwait, Malaysia, Mexico, Netherlands, Netherlands, New Zealand, Oman, Panama, Puerto Rico, Philippines, Poland, Portugal, Qatar, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Turkey, United Arab Emirates.
How much
418,076 units (326,608 US, 91,468 OUS)
Class I

Vasoview Hemopro 2 (w/Vasoshield) Endoscopic Vessel Harvesting System. Model Numbers: VH-4001. Indicated for

Z-0712-2025  ·  Class I  ·  reported 2025-01-15  ·  Ongoing

Why it was recalled
Potential for two issues: 1. Bent or detached heater wire; 2. Silicone peeling or detaching from the Jaws of the Harvesting Tool
Product
Vasoview Hemopro 2 (w/Vasoshield) Endoscopic Vessel Harvesting System. Model Numbers: VH-4001. Indicated for use in minimally invasive surgery allowing access for vessel harvesting.
Where it went
Nationwide distribution. International distribution to Australia, Austria, Bahrain, Bangladesh, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Guatemala, Hong Kong, India, Iran, Iraq, Ireland, Italy, Japan, Jordan, Kenya, Kuwait, Malaysia, Mexico, Netherlands, Netherlands, New Zealand, Oman, Panama, Puerto Rico, Philippines, Poland, Portugal, Qatar, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Turkey, United Arab Emirates.
How much
11,101 units US

Acrobat-i Vacuum Stabilizer System. Model Number C-OM-10000Z

Z-0486-2025  ·  Class II  ·  reported 2024-11-27  ·  Ongoing

Why it was recalled
Deviations in the manufacturing process created a compromise in the sterile barrier, in which sterility cannot be assured during the duration of the product's 2-year shelf life.
Product
Acrobat-i Vacuum Stabilizer System. Model Number C-OM-10000Z
Where it went
Nationwide distribution. International distribution to Albania, Algeria, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brunei Canada, Croatia, Czech Republic, Denmark, Finland, France and Belarus, Germany, Greece, Hong Kong, India, Indonesia, Italy, Japan, Lebanon, Luxembourg, Malaysia, Netherlands, Oman, Pakistan, Poland, Portugal, Slovenia, South Africa, Spain, Switzerland, the Republic of Korea (South Korea), Turkey, United Arab Emirates, United Kingdom.
How much
33,823 units

Acrobat SUV Vacuum Stabilizer System, ST. Model Number C-OM-9000Z

Z-0485-2025  ·  Class II  ·  reported 2024-11-27  ·  Ongoing

Why it was recalled
Deviations in the manufacturing process created a compromise in the sterile barrier, in which sterility cannot be assured during the duration of the product's 2-year shelf life.
Product
Acrobat SUV Vacuum Stabilizer System, ST. Model Number C-OM-9000Z
Where it went
Nationwide distribution. International distribution to Albania, Algeria, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brunei Canada, Croatia, Czech Republic, Denmark, Finland, France and Belarus, Germany, Greece, Hong Kong, India, Indonesia, Italy, Japan, Lebanon, Luxembourg, Malaysia, Netherlands, Oman, Pakistan, Poland, Portugal, Slovenia, South Africa, Spain, Switzerland, the Republic of Korea (South Korea), Turkey, United Arab Emirates, United Kingdom.
How much
10,778 units

Acrobat-i Positioner. Model Number C-XP-5000Z

Z-0484-2025  ·  Class II  ·  reported 2024-11-27  ·  Ongoing

Why it was recalled
Deviations in the manufacturing process created a compromise in the sterile barrier, in which sterility cannot be assured during the duration of the product's 2-year shelf life.
Product
Acrobat-i Positioner. Model Number C-XP-5000Z
Where it went
Nationwide distribution. International distribution to Albania, Algeria, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brunei Canada, Croatia, Czech Republic, Denmark, Finland, France and Belarus, Germany, Greece, Hong Kong, India, Indonesia, Italy, Japan, Lebanon, Luxembourg, Malaysia, Netherlands, Oman, Pakistan, Poland, Portugal, Slovenia, South Africa, Spain, Switzerland, the Republic of Korea (South Korea), Turkey, United Arab Emirates, United Kingdom.
How much
14,338 units
Class I

VasoView HemoPro Endoscopic Vessel Harvesting System, Product Codes VH-3000-W and VH-3500

Z-0092-2025  ·  Class I  ·  reported 2024-11-06  ·  Ongoing

Why it was recalled
There were reports of the silicone detaching from the Jaws of the Harvesting Tool during use.
Product
VasoView HemoPro Endoscopic Vessel Harvesting System, Product Codes VH-3000-W and VH-3500
Where it went
Worldwide distribution. US Nationwide. Brazil, China, and Hong Kong.
How much
28,809 units

The VASOVIEW HEMOPRO Endoscopic Vessel Harvesting Systems is designed for use in conjunction with the 7 mm Ext

Z-2337-2024  ·  Class II  ·  reported 2024-07-17  ·  Ongoing

Why it was recalled
There were 27 complaints between March 22, 2024, and April 30, 2024, related to VH-3500 and VH-4000, reporting that the C-Ring wire was straightened (due to insufficient bend radius) and that the distance between C-Ring and Harvesting Tool was closer than normal. No adverse events have been reported to date in direct association with this issue.
Product
The VASOVIEW HEMOPRO Endoscopic Vessel Harvesting Systems is designed for use in conjunction with the 7 mm Extended Length Endoscope. The Harvesting Cannula has four lumens which house the Endoscope, C-Ring, distal lens washer tube, and harvesting tool for cutting and cauterizing vessel branches. The harvesting tool has two curved jaws. One of the jaws contains the heating elements for branch cutting and cauterization and spot cautery. An area near the tip of the convex side of the jaw is used for spot cautery. The activation toggle is used to control the jaws to activate the heating elements. Positioning of the device, cutting, and cauterization are performed under endoscopic visualization.
Where it went
NationwideForeign: AT DE DK ES GB HK IT NL ZA
How much
600 units

The VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting System is designed for use in conjunction with the 7 mm Ex

Z-2338-2024  ·  Class II  ·  reported 2024-07-17  ·  Ongoing

Why it was recalled
There were 27 complaints between March 22, 2024, and April 30, 2024, related to VH-3500 and VH-4000, reporting that the C-Ring wire was straightened (due to insufficient bend radius) and that the distance between C-Ring and Harvesting Tool was closer than normal. No adverse events have been reported to date in direct association with this issue.
Product
The VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting System is designed for use in conjunction with the 7 mm Extended Length Endoscope. The Harvesting Cannula has four lumens which house the Endoscope, C-Ring, distal lens washer tube, and harvesting tool for cutting and cauterizing vessel branches. The harvesting tool has two curved jaws. One of the jaws contains the heating elements for branch cutting and cauterization and spot cautery. An area near the tip of the convex side of the jaw is used for spot cautery. The activation toggle is used to control the jaws to activate the heating elements. Positioning of the device, cutting, and cauterization are performed under endoscopic visualization.
Where it went
NationwideForeign: AT DE DK ES GB HK IT NL ZA
How much
2463 units units US /330 units OUS

The VASOVIEW HemoPro 2 Endoscopic Vessel Harvesting System, labeled as the following: 1. VasoView Hemopro 2,

Z-2175-2024  ·  Class II  ·  reported 2024-07-03  ·  Ongoing

Why it was recalled
Potential for there to be a fluid ingress into the handle of the harvesting tool, which may lead to the device not providing cautery or providing cautery when not intended.
Product
The VASOVIEW HemoPro 2 Endoscopic Vessel Harvesting System, labeled as the following: 1. VasoView Hemopro 2, Model Number: VH-4000. 2. VasoView Hemopro 2 with Vasoshield. Model Number: VH-4001. The VASOVIEW HemoPro 2 Endoscopic Vessel Harvesting Systems are designed for use in conjunction with the 7 mm Extended Length Endoscope. The Harvesting Cannula has four lumens which house the Endoscope, C-Ring, distal lens washer tube, and harvesting tool for cutting and cauterizing vessel branches. The harvesting tool has two curved jaws. One of the jaws contains the heating elements for branch cutting and cauterization and spot cautery. An area near the tip of the convex side of the jaw is used for spot cautery. The activation toggle is used to control the jaws to activate the heating elements. Positioning of the device, cutting, and cauterization are performed under endoscopic visualization.
Where it went
Worldwide distribution - US Nationwide and the countries of Australia Brazil, Canada, China, EEA, India, Korea, New Zealand, Southern Africa, Taiwan, Thailand, Turkey, & United Kingdom.
How much
430,037 units

The VASOVIEW HemoPro Endoscopic Vessel Harvesting Systems are designed for use in conjunction with the 7 mm Ex

Z-2176-2024  ·  Class II  ·  reported 2024-07-03  ·  Ongoing

Why it was recalled
Potential for there to be a fluid ingress into the handle of the harvesting tool, which may lead to the device not providing cautery or providing cautery when not intended.
Product
The VASOVIEW HemoPro Endoscopic Vessel Harvesting Systems are designed for use in conjunction with the 7 mm Extended Length Endoscope. The Harvesting Cannula has four lumens which house the Endoscope, C-Ring, distal lens washer tube, and harvesting tool for cutting and cauterizing vessel branches. The harvesting tool has two curved jaws. One of the jaws contains the heating elements for branch cutting and cauterization and spot cautery. An area near the tip of the convex side of the jaw is used for spot cautery. The activation toggle is used to control the jaws to activate the heating elements. Positioning of the device, cutting, and cauterization are performed under endoscopic visualization.
Where it went
Worldwide distribution - US Nationwide and the countries of Australia Brazil, Canada, China, EEA, India, Korea, New Zealand, Southern Africa, Taiwan, Thailand, Turkey, & United Kingdom.
How much
37,186 units

Vasoview Hemopro Endoscopic Vessel Harvesting System, model VH-3000-W .

Z-0723-2024  ·  Class II  ·  reported 2024-01-24  ·  Ongoing

Why it was recalled
Risk of polyimide particulate matter being present on the endoscopic harvesting vessel system. This issue may result in polyimide particulate matter being introduced into the arterial conduit for patients undergoing a coronary artery bypass graft (CABG) or peripheral bypass graft surgery.
Product
Vasoview Hemopro Endoscopic Vessel Harvesting System, model VH-3000-W .
Where it went
US: OUS: China, Brazil, Hong Kong.
How much
554 units (OUS)
Class I

Product REF 4050-100N, DRAIN, EXPRESS BRU W/AC, Blood Recovery Unit

Z-0710-2024  ·  Class I  ·  reported 2024-01-24  ·  Ongoing

Why it was recalled
Express chest drains are pre-packaged with sterile water syringes which have bene recalled. If the device is not kept in an upright position, the patient may be exposed to infectious material. Subsequent hazards include but are not limited to infection, abscess, sepsis, and death.
Product
Product REF 4050-100N, DRAIN, EXPRESS BRU W/AC, Blood Recovery Unit
Where it went
Worldwide distribution: US (nationwide) and OUS (Foreign) to countries of: Australia, Brazil, Canada, Luxembourg, Netherlands, Norway, Spain and UK
How much
(7,501 cases/45,006 devices US); no OUS
Class I

Product REF 4000-100N, DRAIN, EXPRESS SINGLE W/AC, Single Collection

Z-0709-2024  ·  Class I  ·  reported 2024-01-24  ·  Ongoing

Why it was recalled
Express chest drains are pre-packaged with sterile water syringes which have bene recalled. If the device is not kept in an upright position, the patient may be exposed to infectious material. Subsequent hazards include but are not limited to infection, abscess, sepsis, and death.
Product
Product REF 4000-100N, DRAIN, EXPRESS SINGLE W/AC, Single Collection
Where it went
Worldwide distribution: US (nationwide) and OUS (Foreign) to countries of: Australia, Brazil, Canada, Luxembourg, Netherlands, Norway, Spain and UK
How much
(34,356 cases/206,136 devices US); (3,902 cases/23,412 devices OUS)

Vasoview Hemopro Endoscopic Vessel Harvesting System, model VH-3500.

Z-0724-2024  ·  Class II  ·  reported 2024-01-24  ·  Ongoing

Why it was recalled
Risk of polyimide particulate matter being present on the endoscopic harvesting vessel system. This issue may result in polyimide particulate matter being introduced into the arterial conduit for patients undergoing a coronary artery bypass graft (CABG) or peripheral bypass graft surgery.
Product
Vasoview Hemopro Endoscopic Vessel Harvesting System, model VH-3500.
Where it went
US: OUS: China, Brazil, Hong Kong.
How much
38,094 units (US)

Acrobat-i Vacuum Positioner System, Model Number XP-5000Z

Z-0589-2024  ·  Class II  ·  reported 2023-12-27  ·  Ongoing

Why it was recalled
The Positioner Arm may not tighten or lock during normal use.
Product
Acrobat-i Vacuum Positioner System, Model Number XP-5000Z
Where it went
US Nationwide. Global Distribution.
How much
6472 units

Vasoview 6 Pro (VV6) - Endoscopic Vessel Harvesting System, Model VH-2400 printed on outer label, inner label

Z-0210-2024  ·  Class II  ·  reported 2023-11-08  ·  Ongoing

Why it was recalled
Device packaged inside an incorrect outer box; displaying artwork & illustrations for a different device.
Product
Vasoview 6 Pro (VV6) - Endoscopic Vessel Harvesting System, Model VH-2400 printed on outer label, inner label and IFU, outer box artwork for Vasoview Hemopro 2 (HP2).
Where it went
US (AL, AR, CA, FL, GA, IN, KY, MI, NJ, OK, PA, TN, TX) and Canada.
How much
N/A

Maquet Acrobat-i Vacuum Stabilizer System Intended for use during performance of cardiac surgical procedures t

Z-0137-2024  ·  Class II  ·  reported 2023-11-01  ·  Ongoing

Why it was recalled
Housing Mount Jaw of these devices breaking during normal use and may result in injury to the epicardial tissue or vasculature, or a coronary artery or previously placed bypass graft, or could result in procedural delay and/or conversion.
Product
Maquet Acrobat-i Vacuum Stabilizer System Intended for use during performance of cardiac surgical procedures through a sternotomy incision approach. Product Code: OM-10000Z
Where it went
Nationwide Foreign: AT BE DE ES FR GB IT JP KR LU NL PL PT TR
How much
2951 units

ACROBAT SUV Vacuum Off-Pump System, Part Number C-OM-9000S. Component of a chest stabilization system for car

Z-2272-2023  ·  Class II  ·  reported 2023-08-09  ·  Ongoing

Why it was recalled
The Standard Blade may not securely latch onto the Activator Drive. This may result in the inability to stabilize and position or may result in the loss of stabilization and positioning of heart, release of device component into the patient, and/or a procedural delay and/or conversion.
Product
ACROBAT SUV Vacuum Off-Pump System, Part Number C-OM-9000S. Component of a chest stabilization system for cardiac and chest surgery.
Where it went
Worldwide distribution - US Nationwide and the countries of Bangladesh, Brazil, Chile, Colombia, Iran, Jordan, Malaysia, Mexico, Netherlands, Singapore, Sudan, Taiwan, Thailand, and United Arab Emirates.
How much
401 units

ACCESSRAIL Platform (Standard Blade) Part Number SB-1000, and (Deep Blades) Part Number C-DB-1000. Component

Z-2271-2023  ·  Class II  ·  reported 2023-08-09  ·  Ongoing

Why it was recalled
The Standard Blade may not securely latch onto the Activator Drive. This may result in the inability to stabilize and position or may result in the loss of stabilization and positioning of heart, release of device component into the patient, and/or a procedural delay and/or conversion.
Product
ACCESSRAIL Platform (Standard Blade) Part Number SB-1000, and (Deep Blades) Part Number C-DB-1000. Component of a chest stabilization system for cardiac and chest surgery.
Where it went
Worldwide distribution - US Nationwide and the countries of Bangladesh, Brazil, Chile, Colombia, Iran, Jordan, Malaysia, Mexico, Netherlands, Singapore, Sudan, Taiwan, Thailand, and United Arab Emirates.
How much
785 units

ACROBAT-i Vacuum Stabilizer System, Part Number C-OM-10000. Component of a chest stabilization system for card

Z-2274-2023  ·  Class II  ·  reported 2023-08-09  ·  Ongoing

Why it was recalled
The Standard Blade may not securely latch onto the Activator Drive. This may result in the inability to stabilize and position or may result in the loss of stabilization and positioning of heart, release of device component into the patient, and/or a procedural delay and/or conversion.
Product
ACROBAT-i Vacuum Stabilizer System, Part Number C-OM-10000. Component of a chest stabilization system for cardiac and chest surgery.
Where it went
Worldwide distribution - US Nationwide and the countries of Bangladesh, Brazil, Chile, Colombia, Iran, Jordan, Malaysia, Mexico, Netherlands, Singapore, Sudan, Taiwan, Thailand, and United Arab Emirates.
How much
334 units

ACROBAT V Vacuum Off-Pump System, Part Number C-OM-9100S. Component of a chest stabilization system for cardia

Z-2273-2023  ·  Class II  ·  reported 2023-08-09  ·  Ongoing

Why it was recalled
The Standard Blade may not securely latch onto the Activator Drive. This may result in the inability to stabilize and position or may result in the loss of stabilization and positioning of heart, release of device component into the patient, and/or a procedural delay and/or conversion.
Product
ACROBAT V Vacuum Off-Pump System, Part Number C-OM-9100S. Component of a chest stabilization system for cardiac and chest surgery.
Where it went
Worldwide distribution - US Nationwide and the countries of Bangladesh, Brazil, Chile, Colombia, Iran, Jordan, Malaysia, Mexico, Netherlands, Singapore, Sudan, Taiwan, Thailand, and United Arab Emirates.
How much
594 units

VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting System, Model Number C-VH-4000

Z-1336-2023  ·  Class II  ·  reported 2023-04-12  ·  Ongoing

Why it was recalled
Some batches of product were not sterilized to their minimum sterilization specification.
Product
VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting System, Model Number C-VH-4000
Where it went
Worldwide distribution - US Nationwide and the countries of Brazil, China.
How much
2300 units

VASOVIEW HEMOPRO, OUS, Endoscopic Vessel Harvesting System, Model Number C-VH-3000-W

Z-1337-2023  ·  Class II  ·  reported 2023-04-12  ·  Ongoing

Why it was recalled
Some batches of product were not sterilized to their minimum sterilization specification.
Product
VASOVIEW HEMOPRO, OUS, Endoscopic Vessel Harvesting System, Model Number C-VH-3000-W
Where it went
Worldwide distribution - US Nationwide and the countries of Brazil, China.
How much
235 units

Showing the 30 most recent of 252. The rest are searchable at the FDA enforcement report database.

Read these counts with three things in mind

A firm with more recalls usually makes more products. A contract packer runs hundreds of lines. It will out-recall a company with one product whatever either does. This is not a safety ranking.

Most recalls here are voluntary. A firm found its own fault and pulled the stock. That is the system working. It lands in this database looking the same as a recall the FDA had to push for.

Terminated means the FDA considers the recall finished. It is not a verdict. It does not mean anyone was harmed.

Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.