thejoe
Maquet Cardiovascular, LLC
Maquet Cardiovascular, LLC: 252 FDA recall reports
Every medical device recall the FDA has published for Maquet Cardiovascular, LLC. The fault, and how the agency graded it.
252Reports
5Class I
35Ongoing
2026-07-22Most recent
5 Class I on this pageClass I means the FDA judged there was a reasonable probability of serious harm or death. Check the lot numbers against anything you hold.
What the classes mean
The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.
Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.
By class
By year reported
Checking whether it affects you
The lot or code number is what matters. Not the product name. A recall covers
specific lots. Yours may not be one of them.
Check the code on your package against the report. The FDA keeps the current
list at fda.gov/safety/recalls.
The reports
The text below is quoted from the FDA record word for word.
The Vasoview Hemopro 2 Endoscopic Vessel Harvesting System (Model VH-4000; UDI 00607567700406) is a sterile,
Z-2741-2026 · Class II · reported 2026-07-22 · Ongoing
- Why it was recalled
- Getinge identified that Vasoview Hemopro 2 devices from Lot 3000547797 were exposed to an out-of-specification sterilization dose during a gamma sterilization run (Run ID 1183-28782A) performed by STERIS at its Whippany, NJ facility between March 31 and April 1 of 2026.
- Product
- The Vasoview Hemopro 2 Endoscopic Vessel Harvesting System (Model VH-4000; UDI 00607567700406) is a sterile, single-use Class II device that was cleared for marketing by FDA under 510(k) number K101274.
- Where it went
- US Nationwide distribution in the states of NY, CA, SC, FL, TX, GA, OR.
- How much
- 205 units
Vasoview Hemopro 3 Endoscopic Vessel Harvesting System Model No. C-VH-6000. The VASOVIEW HEMOPRO 3 Endoscopi
Z-2636-2025 · Class II · reported 2025-10-01 · Ongoing
- Why it was recalled
- Reports of out-of-box failures discovered during the limited launch phase. Failure modes included angled deformity at the attachment site of the Harvesting Tool Jaws and the distal end of the Harvesting Tool Shaft, jaws not fully closed out of box, and heater wire lifted/flexed away from jaws.
- Product
- Vasoview Hemopro 3 Endoscopic Vessel Harvesting System Model No. C-VH-6000. The VASOVIEW HEMOPRO 3 Endoscopic Vessel Harvesting (EVH) System (VH-6000, VH-6001, and VH-6010) is a vessel harvesting kit designed for use with a 7 mm endoscope to cut and cauterize vessel branches. The HP3 Harvesting Tool features two curved, insulated jaws with heating elements that operate through heat and pressure rather than radiofrequency energy.
- Where it went
- US Nationwide distribution in the states of FL, IL, MD & TX.
- How much
- 98 units
Heartstring III Proximal Seal System. Intravascular anastomosis occluder.
Z-2586-2025 · Class II · reported 2025-09-24 · Ongoing
- Why it was recalled
- Three failure modes have been identified: failure of the Heartstring Seal to load, failure of the Heartstring Seal to deploy into the aortotomy, and failure of the deployed Heartstring Seal to provide adequate hemostasis.
- Product
- Heartstring III Proximal Seal System. Intravascular anastomosis occluder.
- Where it went
- Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Canada, China, Colombia, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Israel, Italy, Japan, Luxembourg, Malaysia, Netherlands, New Zealand, Norway, Poland, Portugal, Saudi Arabia, Singapore, Slovakia, Slovenia, South Korea, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom.
- How much
- 126 units (91 US, 35 OUS)
Heartstring III Proximal Seal System, 3.8 MM. Intravascular anastomosis occluder.
Z-2587-2025 · Class II · reported 2025-09-24 · Ongoing
- Why it was recalled
- Three failure modes have been identified: failure of the Heartstring Seal to load, failure of the Heartstring Seal to deploy into the aortotomy, and failure of the deployed Heartstring Seal to provide adequate hemostasis.
- Product
- Heartstring III Proximal Seal System, 3.8 MM. Intravascular anastomosis occluder.
- Where it went
- Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Canada, China, Colombia, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Israel, Italy, Japan, Luxembourg, Malaysia, Netherlands, New Zealand, Norway, Poland, Portugal, Saudi Arabia, Singapore, Slovakia, Slovenia, South Korea, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom.
- How much
- 28,331 units (16,615 US, 11,716 OUS)
Heartstring III Proximal Seal System, 4.3 MM. Intravascular anastomosis occluder.
Z-2588-2025 · Class II · reported 2025-09-24 · Ongoing
- Why it was recalled
- Three failure modes have been identified: failure of the Heartstring Seal to load, failure of the Heartstring Seal to deploy into the aortotomy, and failure of the deployed Heartstring Seal to provide adequate hemostasis.
- Product
- Heartstring III Proximal Seal System, 4.3 MM. Intravascular anastomosis occluder.
- Where it went
- Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Canada, China, Colombia, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Israel, Italy, Japan, Luxembourg, Malaysia, Netherlands, New Zealand, Norway, Poland, Portugal, Saudi Arabia, Singapore, Slovakia, Slovenia, South Korea, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom.
- How much
- 12,471 units (6,565 US, 5,906 OUS)
The VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting System is designed for use in conjunction with the 7 mm ex
Z-1622-2025 · Class II · reported 2025-04-30 · Ongoing
- Why it was recalled
- There have been at least 12 complaints from users as of March 3, 2025, reporting that they identified the ceramic C-Ring to be broken in half during use of the device. The ceramic C-Ring has a potential to break while in use. Maquet/Getinge determined that the issue was limited to the units of VH-4000 and VH-4001 manufactured with ceramic C-Ring, and the ceramic C-Ring was implemented in the Hemopro 2 devices from November 5, 2024, through February 14, 2025.
- Product
- The VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting System is designed for use in conjunction with the 7 mm extended length endoscope. The Harvesting Cannula has four lumens which house the endoscope, C-Ring, distal lens washer tube, and Harvesting Tool for cutting and cauterizing vessel branches. The Harvesting Tool has two curved jaws. One of the jaws contains the heating elements for branch cutting and cauterization and spot cautery. An area near the tip of the convex side of the jaw is used for spot cautery. The activation toggle is used to control the jaws to activate the heating elements. Positioning the device, cutting, and cauterization are performed under endoscopic visualization. The device is intended for specific use with the VASOVIEW HEMOPRO Power Supply, VASOVIEW HEMOPRO 2 extension cable, and an adapter cable.
- Where it went
- Worldwide distribution - US Nationwide and the countries of Mexico, United Arab Emirates, Singapore, India, Australia, Canada, Brazil, Hong Kong, Japan, South Korea, New Zealand.
- How much
- 22,244 units
Maquet VH-3010 Power Supply. Used to deliver power to the harvesting tool of the Vasoview Hemopro EVH System (
Z-1577-2025 · Class II · reported 2025-04-23 · Ongoing
- Why it was recalled
- Incorrect resistor utilized in the VH-3010 Power Supply, which may cause the jaws of the HemoPro 2 to not adequately heat up.
- Product
- Maquet VH-3010 Power Supply. Used to deliver power to the harvesting tool of the Vasoview Hemopro EVH System (HemoPro 2).
- Where it went
- Worldwide - US Nationwide distribution in the states of AL, CA, CO, DC, FL, GA, IA, ID, IL, IN, KY, MD, MI, MN, MO, NC, NH, NJ, NY, OH, PA, PR, TX, UT, VA, WI, WV and the countries of China, Japan, Netherlands, New Zealand, Singapore, United Kingdom, United States.
- How much
- 110 units
Class I
VasoView HemoPro 2 Endoscopic Vessel Harvesting System. Model Numbers: VH-4000. Indicated for use in minimall
Z-0711-2025 · Class I · reported 2025-01-15 · Ongoing
- Why it was recalled
- Potential for two issues: 1. Bent or detached heater wire; 2. Silicone peeling or detaching from the Jaws of the Harvesting Tool
- Product
- VasoView HemoPro 2 Endoscopic Vessel Harvesting System. Model Numbers: VH-4000. Indicated for use in minimally invasive surgery allowing access for vessel harvesting.
- Where it went
- Nationwide distribution. International distribution to Australia, Austria, Bahrain, Bangladesh, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Guatemala, Hong Kong, India, Iran, Iraq, Ireland, Italy, Japan, Jordan, Kenya, Kuwait, Malaysia, Mexico, Netherlands, Netherlands, New Zealand, Oman, Panama, Puerto Rico, Philippines, Poland, Portugal, Qatar, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Turkey, United Arab Emirates.
- How much
- 418,076 units (326,608 US, 91,468 OUS)
Class I
Vasoview Hemopro 2 (w/Vasoshield) Endoscopic Vessel Harvesting System. Model Numbers: VH-4001. Indicated for
Z-0712-2025 · Class I · reported 2025-01-15 · Ongoing
- Why it was recalled
- Potential for two issues: 1. Bent or detached heater wire; 2. Silicone peeling or detaching from the Jaws of the Harvesting Tool
- Product
- Vasoview Hemopro 2 (w/Vasoshield) Endoscopic Vessel Harvesting System. Model Numbers: VH-4001. Indicated for use in minimally invasive surgery allowing access for vessel harvesting.
- Where it went
- Nationwide distribution. International distribution to Australia, Austria, Bahrain, Bangladesh, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Guatemala, Hong Kong, India, Iran, Iraq, Ireland, Italy, Japan, Jordan, Kenya, Kuwait, Malaysia, Mexico, Netherlands, Netherlands, New Zealand, Oman, Panama, Puerto Rico, Philippines, Poland, Portugal, Qatar, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Turkey, United Arab Emirates.
- How much
- 11,101 units US
Acrobat-i Vacuum Stabilizer System. Model Number C-OM-10000Z
Z-0486-2025 · Class II · reported 2024-11-27 · Ongoing
- Why it was recalled
- Deviations in the manufacturing process created a compromise in the sterile barrier, in which sterility cannot be assured during the duration of the product's 2-year shelf life.
- Product
- Acrobat-i Vacuum Stabilizer System. Model Number C-OM-10000Z
- Where it went
- Nationwide distribution. International distribution to Albania, Algeria, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brunei Canada, Croatia, Czech Republic, Denmark, Finland, France and Belarus, Germany, Greece, Hong Kong, India, Indonesia, Italy, Japan, Lebanon, Luxembourg, Malaysia, Netherlands, Oman, Pakistan, Poland, Portugal, Slovenia, South Africa, Spain, Switzerland, the Republic of Korea (South Korea), Turkey, United Arab Emirates, United Kingdom.
- How much
- 33,823 units
Acrobat SUV Vacuum Stabilizer System, ST. Model Number C-OM-9000Z
Z-0485-2025 · Class II · reported 2024-11-27 · Ongoing
- Why it was recalled
- Deviations in the manufacturing process created a compromise in the sterile barrier, in which sterility cannot be assured during the duration of the product's 2-year shelf life.
- Product
- Acrobat SUV Vacuum Stabilizer System, ST. Model Number C-OM-9000Z
- Where it went
- Nationwide distribution. International distribution to Albania, Algeria, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brunei Canada, Croatia, Czech Republic, Denmark, Finland, France and Belarus, Germany, Greece, Hong Kong, India, Indonesia, Italy, Japan, Lebanon, Luxembourg, Malaysia, Netherlands, Oman, Pakistan, Poland, Portugal, Slovenia, South Africa, Spain, Switzerland, the Republic of Korea (South Korea), Turkey, United Arab Emirates, United Kingdom.
- How much
- 10,778 units
Acrobat-i Positioner. Model Number C-XP-5000Z
Z-0484-2025 · Class II · reported 2024-11-27 · Ongoing
- Why it was recalled
- Deviations in the manufacturing process created a compromise in the sterile barrier, in which sterility cannot be assured during the duration of the product's 2-year shelf life.
- Product
- Acrobat-i Positioner. Model Number C-XP-5000Z
- Where it went
- Nationwide distribution. International distribution to Albania, Algeria, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brunei Canada, Croatia, Czech Republic, Denmark, Finland, France and Belarus, Germany, Greece, Hong Kong, India, Indonesia, Italy, Japan, Lebanon, Luxembourg, Malaysia, Netherlands, Oman, Pakistan, Poland, Portugal, Slovenia, South Africa, Spain, Switzerland, the Republic of Korea (South Korea), Turkey, United Arab Emirates, United Kingdom.
- How much
- 14,338 units
Class I
VasoView HemoPro Endoscopic Vessel Harvesting System, Product Codes VH-3000-W and VH-3500
Z-0092-2025 · Class I · reported 2024-11-06 · Ongoing
- Why it was recalled
- There were reports of the silicone detaching from the Jaws of the Harvesting Tool during use.
- Product
- VasoView HemoPro Endoscopic Vessel Harvesting System, Product Codes VH-3000-W and VH-3500
- Where it went
- Worldwide distribution. US Nationwide. Brazil, China, and Hong Kong.
- How much
- 28,809 units
The VASOVIEW HEMOPRO Endoscopic Vessel Harvesting Systems is designed for use in conjunction with the 7 mm Ext
Z-2337-2024 · Class II · reported 2024-07-17 · Ongoing
- Why it was recalled
- There were 27 complaints between March 22, 2024, and April 30, 2024, related to VH-3500 and VH-4000, reporting that the C-Ring wire was straightened (due to insufficient bend radius) and that the distance between C-Ring and Harvesting Tool was closer than normal. No adverse events have been reported to date in direct association with this issue.
- Product
- The VASOVIEW HEMOPRO Endoscopic Vessel Harvesting Systems is designed for use in conjunction with the 7 mm Extended Length Endoscope. The Harvesting Cannula has four lumens which house the Endoscope, C-Ring, distal lens washer tube, and harvesting tool for cutting and cauterizing vessel branches. The harvesting tool has two curved jaws. One of the jaws contains the heating elements for branch cutting and cauterization and spot cautery. An area near the tip of the convex side of the jaw is used for spot cautery. The activation toggle is used to control the jaws to activate the heating elements. Positioning of the device, cutting, and cauterization are performed under endoscopic visualization.
- Where it went
- NationwideForeign: AT DE DK ES GB HK IT NL ZA
- How much
- 600 units
The VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting System is designed for use in conjunction with the 7 mm Ex
Z-2338-2024 · Class II · reported 2024-07-17 · Ongoing
- Why it was recalled
- There were 27 complaints between March 22, 2024, and April 30, 2024, related to VH-3500 and VH-4000, reporting that the C-Ring wire was straightened (due to insufficient bend radius) and that the distance between C-Ring and Harvesting Tool was closer than normal. No adverse events have been reported to date in direct association with this issue.
- Product
- The VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting System is designed for use in conjunction with the 7 mm Extended Length Endoscope. The Harvesting Cannula has four lumens which house the Endoscope, C-Ring, distal lens washer tube, and harvesting tool for cutting and cauterizing vessel branches. The harvesting tool has two curved jaws. One of the jaws contains the heating elements for branch cutting and cauterization and spot cautery. An area near the tip of the convex side of the jaw is used for spot cautery. The activation toggle is used to control the jaws to activate the heating elements. Positioning of the device, cutting, and cauterization are performed under endoscopic visualization.
- Where it went
- NationwideForeign: AT DE DK ES GB HK IT NL ZA
- How much
- 2463 units units US /330 units OUS
The VASOVIEW HemoPro 2 Endoscopic Vessel Harvesting System, labeled as the following: 1. VasoView Hemopro 2,
Z-2175-2024 · Class II · reported 2024-07-03 · Ongoing
- Why it was recalled
- Potential for there to be a fluid ingress into the handle of the harvesting tool, which may lead to the device not providing cautery or providing cautery when not intended.
- Product
- The VASOVIEW HemoPro 2 Endoscopic Vessel Harvesting System, labeled as the following: 1. VasoView Hemopro 2, Model Number: VH-4000. 2. VasoView Hemopro 2 with Vasoshield. Model Number: VH-4001. The VASOVIEW HemoPro 2 Endoscopic Vessel Harvesting Systems are designed for use in conjunction with the 7 mm Extended Length Endoscope. The Harvesting Cannula has four lumens which house the Endoscope, C-Ring, distal lens washer tube, and harvesting tool for cutting and cauterizing vessel branches. The harvesting tool has two curved jaws. One of the jaws contains the heating elements for branch cutting and cauterization and spot cautery. An area near the tip of the convex side of the jaw is used for spot cautery. The activation toggle is used to control the jaws to activate the heating elements. Positioning of the device, cutting, and cauterization are performed under endoscopic visualization.
- Where it went
- Worldwide distribution - US Nationwide and the countries of Australia Brazil, Canada, China, EEA, India, Korea, New Zealand, Southern Africa, Taiwan, Thailand, Turkey, & United Kingdom.
- How much
- 430,037 units
The VASOVIEW HemoPro Endoscopic Vessel Harvesting Systems are designed for use in conjunction with the 7 mm Ex
Z-2176-2024 · Class II · reported 2024-07-03 · Ongoing
- Why it was recalled
- Potential for there to be a fluid ingress into the handle of the harvesting tool, which may lead to the device not providing cautery or providing cautery when not intended.
- Product
- The VASOVIEW HemoPro Endoscopic Vessel Harvesting Systems are designed for use in conjunction with the 7 mm Extended Length Endoscope. The Harvesting Cannula has four lumens which house the Endoscope, C-Ring, distal lens washer tube, and harvesting tool for cutting and cauterizing vessel branches. The harvesting tool has two curved jaws. One of the jaws contains the heating elements for branch cutting and cauterization and spot cautery. An area near the tip of the convex side of the jaw is used for spot cautery. The activation toggle is used to control the jaws to activate the heating elements. Positioning of the device, cutting, and cauterization are performed under endoscopic visualization.
- Where it went
- Worldwide distribution - US Nationwide and the countries of Australia Brazil, Canada, China, EEA, India, Korea, New Zealand, Southern Africa, Taiwan, Thailand, Turkey, & United Kingdom.
- How much
- 37,186 units
Vasoview Hemopro Endoscopic Vessel Harvesting System, model VH-3000-W .
Z-0723-2024 · Class II · reported 2024-01-24 · Ongoing
- Why it was recalled
- Risk of polyimide particulate matter being present on the endoscopic harvesting vessel system. This issue may result in polyimide particulate matter being introduced into the arterial conduit for patients undergoing a coronary artery bypass graft (CABG) or peripheral bypass graft surgery.
- Product
- Vasoview Hemopro Endoscopic Vessel Harvesting System, model VH-3000-W .
- Where it went
- US: OUS: China, Brazil, Hong Kong.
- How much
- 554 units (OUS)
Class I
Product REF 4050-100N, DRAIN, EXPRESS BRU W/AC, Blood Recovery Unit
Z-0710-2024 · Class I · reported 2024-01-24 · Ongoing
- Why it was recalled
- Express chest drains are pre-packaged with sterile water syringes which have bene recalled. If the device is not kept in an upright position, the patient may be exposed to infectious material. Subsequent hazards include but are not limited to infection, abscess, sepsis, and death.
- Product
- Product REF 4050-100N, DRAIN, EXPRESS BRU W/AC, Blood Recovery Unit
- Where it went
- Worldwide distribution: US (nationwide) and OUS (Foreign) to countries of: Australia, Brazil, Canada, Luxembourg, Netherlands, Norway, Spain and UK
- How much
- (7,501 cases/45,006 devices US); no OUS
Class I
Product REF 4000-100N, DRAIN, EXPRESS SINGLE W/AC, Single Collection
Z-0709-2024 · Class I · reported 2024-01-24 · Ongoing
- Why it was recalled
- Express chest drains are pre-packaged with sterile water syringes which have bene recalled. If the device is not kept in an upright position, the patient may be exposed to infectious material. Subsequent hazards include but are not limited to infection, abscess, sepsis, and death.
- Product
- Product REF 4000-100N, DRAIN, EXPRESS SINGLE W/AC, Single Collection
- Where it went
- Worldwide distribution: US (nationwide) and OUS (Foreign) to countries of: Australia, Brazil, Canada, Luxembourg, Netherlands, Norway, Spain and UK
- How much
- (34,356 cases/206,136 devices US); (3,902 cases/23,412 devices OUS)
Vasoview Hemopro Endoscopic Vessel Harvesting System, model VH-3500.
Z-0724-2024 · Class II · reported 2024-01-24 · Ongoing
- Why it was recalled
- Risk of polyimide particulate matter being present on the endoscopic harvesting vessel system. This issue may result in polyimide particulate matter being introduced into the arterial conduit for patients undergoing a coronary artery bypass graft (CABG) or peripheral bypass graft surgery.
- Product
- Vasoview Hemopro Endoscopic Vessel Harvesting System, model VH-3500.
- Where it went
- US: OUS: China, Brazil, Hong Kong.
- How much
- 38,094 units (US)
Acrobat-i Vacuum Positioner System, Model Number XP-5000Z
Z-0589-2024 · Class II · reported 2023-12-27 · Ongoing
- Why it was recalled
- The Positioner Arm may not tighten or lock during normal use.
- Product
- Acrobat-i Vacuum Positioner System, Model Number XP-5000Z
- Where it went
- US Nationwide. Global Distribution.
- How much
- 6472 units
Vasoview 6 Pro (VV6) - Endoscopic Vessel Harvesting System, Model VH-2400 printed on outer label, inner label
Z-0210-2024 · Class II · reported 2023-11-08 · Ongoing
- Why it was recalled
- Device packaged inside an incorrect outer box; displaying artwork & illustrations for a different device.
- Product
- Vasoview 6 Pro (VV6) - Endoscopic Vessel Harvesting System, Model VH-2400 printed on outer label, inner label and IFU, outer box artwork for Vasoview Hemopro 2 (HP2).
- Where it went
- US (AL, AR, CA, FL, GA, IN, KY, MI, NJ, OK, PA, TN, TX) and Canada.
- How much
- N/A
Maquet Acrobat-i Vacuum Stabilizer System Intended for use during performance of cardiac surgical procedures t
Z-0137-2024 · Class II · reported 2023-11-01 · Ongoing
- Why it was recalled
- Housing Mount Jaw of these devices breaking during normal use and may result in injury to the epicardial tissue or vasculature, or a coronary artery or previously placed bypass graft, or could result in procedural delay and/or conversion.
- Product
- Maquet Acrobat-i Vacuum Stabilizer System Intended for use during performance of cardiac surgical procedures through a sternotomy incision approach. Product Code: OM-10000Z
- Where it went
- Nationwide Foreign: AT BE DE ES FR GB IT JP KR LU NL PL PT TR
- How much
- 2951 units
ACROBAT SUV Vacuum Off-Pump System, Part Number C-OM-9000S. Component of a chest stabilization system for car
Z-2272-2023 · Class II · reported 2023-08-09 · Ongoing
- Why it was recalled
- The Standard Blade may not securely latch onto the Activator Drive. This may result in the inability to stabilize and position or may result in the loss of stabilization and positioning of heart, release of device component into the patient, and/or a procedural delay and/or conversion.
- Product
- ACROBAT SUV Vacuum Off-Pump System, Part Number C-OM-9000S. Component of a chest stabilization system for cardiac and chest surgery.
- Where it went
- Worldwide distribution - US Nationwide and the countries of Bangladesh, Brazil, Chile, Colombia, Iran, Jordan, Malaysia, Mexico, Netherlands, Singapore, Sudan, Taiwan, Thailand, and United Arab Emirates.
- How much
- 401 units
ACCESSRAIL Platform (Standard Blade) Part Number SB-1000, and (Deep Blades) Part Number C-DB-1000. Component
Z-2271-2023 · Class II · reported 2023-08-09 · Ongoing
- Why it was recalled
- The Standard Blade may not securely latch onto the Activator Drive. This may result in the inability to stabilize and position or may result in the loss of stabilization and positioning of heart, release of device component into the patient, and/or a procedural delay and/or conversion.
- Product
- ACCESSRAIL Platform (Standard Blade) Part Number SB-1000, and (Deep Blades) Part Number C-DB-1000. Component of a chest stabilization system for cardiac and chest surgery.
- Where it went
- Worldwide distribution - US Nationwide and the countries of Bangladesh, Brazil, Chile, Colombia, Iran, Jordan, Malaysia, Mexico, Netherlands, Singapore, Sudan, Taiwan, Thailand, and United Arab Emirates.
- How much
- 785 units
ACROBAT-i Vacuum Stabilizer System, Part Number C-OM-10000. Component of a chest stabilization system for card
Z-2274-2023 · Class II · reported 2023-08-09 · Ongoing
- Why it was recalled
- The Standard Blade may not securely latch onto the Activator Drive. This may result in the inability to stabilize and position or may result in the loss of stabilization and positioning of heart, release of device component into the patient, and/or a procedural delay and/or conversion.
- Product
- ACROBAT-i Vacuum Stabilizer System, Part Number C-OM-10000. Component of a chest stabilization system for cardiac and chest surgery.
- Where it went
- Worldwide distribution - US Nationwide and the countries of Bangladesh, Brazil, Chile, Colombia, Iran, Jordan, Malaysia, Mexico, Netherlands, Singapore, Sudan, Taiwan, Thailand, and United Arab Emirates.
- How much
- 334 units
ACROBAT V Vacuum Off-Pump System, Part Number C-OM-9100S. Component of a chest stabilization system for cardia
Z-2273-2023 · Class II · reported 2023-08-09 · Ongoing
- Why it was recalled
- The Standard Blade may not securely latch onto the Activator Drive. This may result in the inability to stabilize and position or may result in the loss of stabilization and positioning of heart, release of device component into the patient, and/or a procedural delay and/or conversion.
- Product
- ACROBAT V Vacuum Off-Pump System, Part Number C-OM-9100S. Component of a chest stabilization system for cardiac and chest surgery.
- Where it went
- Worldwide distribution - US Nationwide and the countries of Bangladesh, Brazil, Chile, Colombia, Iran, Jordan, Malaysia, Mexico, Netherlands, Singapore, Sudan, Taiwan, Thailand, and United Arab Emirates.
- How much
- 594 units
VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting System, Model Number C-VH-4000
Z-1336-2023 · Class II · reported 2023-04-12 · Ongoing
- Why it was recalled
- Some batches of product were not sterilized to their minimum sterilization specification.
- Product
- VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting System, Model Number C-VH-4000
- Where it went
- Worldwide distribution - US Nationwide and the countries of Brazil, China.
- How much
- 2300 units
VASOVIEW HEMOPRO, OUS, Endoscopic Vessel Harvesting System, Model Number C-VH-3000-W
Z-1337-2023 · Class II · reported 2023-04-12 · Ongoing
- Why it was recalled
- Some batches of product were not sterilized to their minimum sterilization specification.
- Product
- VASOVIEW HEMOPRO, OUS, Endoscopic Vessel Harvesting System, Model Number C-VH-3000-W
- Where it went
- Worldwide distribution - US Nationwide and the countries of Brazil, China.
- How much
- 235 units
Showing the 30 most recent of 252. The rest are searchable at the FDA enforcement report database.
Read these counts with three things in mind
A firm with more recalls usually makes more products. A contract packer runs
hundreds of lines. It will out-recall a company with one product whatever either
does. This is not a safety ranking.
Most recalls here are voluntary. A firm found its own fault and pulled the stock.
That is the system working. It lands in this database looking the same as a recall
the FDA had to push for.
Terminated means the FDA considers the recall finished. It is not a verdict.
It does not mean anyone was harmed.