K2M, Inc: 39 FDA recall reports
Every medical device recall the FDA has published for K2M, Inc. The fault, and how the agency graded it.
39Reports
0Class I
3Ongoing
2022-07-06Most recent
What the classes mean
The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.
Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.
By class
By year reported
Checking whether it affects you
The lot or code number is what matters. Not the product name. A recall covers
specific lots. Yours may not be one of them.
Check the code on your package against the report. The FDA keeps the current
list at fda.gov/safety/recalls.
The reports
The text below is quoted from the FDA record word for word.
Yukon Straight Rod 4.0 x120mm; Cat# 7601-540120. Used to provide immobilization and stabilization of spinal
Z-1307-2022 · Class II · reported 2022-07-06 · Ongoing
- Why it was recalled
- Some of the rods in this lot were observed to have discoloration, later confirmed to be indication of trace amounts of inorganic phosphorous compound. This may produce a mild irritating or inflammatory response; adverse local tissue reactions may occur related to cytotoxicity with the nonconforming rod and surrounding tissue.
- Product
- Yukon Straight Rod 4.0 x120mm; Cat# 7601-540120. Used to provide immobilization and stabilization of spinal segments.
- Where it went
- Domestic distribution to FL GA NE NV NY OH OK SC Foreign distribution to Canada, EU, and Hong Kong.
- How much
- 53 devices (19 US; 34 OUS)
Everest MI XT Inner Dilator, Catalog Number 5101-90167
Z-0705-2022 · Class II · reported 2022-03-09 · Ongoing
- Why it was recalled
- Stryker received two (2) complaints for units from specific lots of Inner and Outer Dilators not fitting together properly during a surgical procedure. Some dilators from these lots and from other lots were subsequently determined to be affected by a manufacturing nonconformance.
- Product
- Everest MI XT Inner Dilator, Catalog Number 5101-90167
- Where it went
- Worldwide distribution - US Nationwide.
- How much
- 686 Total Devices
Everest MI XT Outer Dilator, Catalog Number 5101-90168
Z-0706-2022 · Class II · reported 2022-03-09 · Ongoing
- Why it was recalled
- Stryker received two (2) complaints for units from specific lots of Inner and Outer Dilators not fitting together properly during a surgical procedure. Some dilators from these lots and from other lots were subsequently determined to be affected by a manufacturing nonconformance.
- Product
- Everest MI XT Outer Dilator, Catalog Number 5101-90168
- Where it went
- Worldwide distribution - US Nationwide.
- How much
- 686 Total Devices
Chesapeake Anterior Lumbar (AL) Removal Tool; Cat. no. 2008-90068
Z-0104-2022 · Class II · reported 2021-10-20 · Terminated
- Why it was recalled
- Units may arrive with missing pins and/or in a state in disassembly. Pins dislodging from the instrument during a procedure may enter the surgical cavity, requiring intraoperative removal or revision surgery.
- Product
- Chesapeake Anterior Lumbar (AL) Removal Tool; Cat. no. 2008-90068
- Where it went
- Domestic distribution to CO, IL, LA, MA, MS, OK, PA, TN, and TX. Foreign distribution to Columbia, UK, Netherlands, and Canada.
- How much
- 23
Cascadia AN Interbody Convex 10x28x14mm Catalog Number: 6101-2102814NC-G2 Lumbar implants are interverte
Z-1697-2021 · Class II · reported 2021-06-09 · Terminated
- Why it was recalled
- Mislabeled product labeling references incorrect length or height, or length and height for specific Lots, Sizes 10x22x14mm, 10x28x14mm, 10x22x15mm
- Product
- Cascadia AN Interbody Convex 10x28x14mm Catalog Number: 6101-2102814NC-G2 Lumbar implants are intervertebral body fusion devices indicated for use with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft when used as anadjunct to fusion in patients with degenerative disc disease.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of MA, MI, NJ, PA, UT and the countries of Italy, Netherlands.
- How much
- 3 units
Cascadia AN Interbody Convex 10x22x15mm Catalog Number: 6101-2102215NC-G2 Lumbar implants are interverte
Z-1696-2021 · Class II · reported 2021-06-09 · Terminated
- Why it was recalled
- Mislabeled product labeling references incorrect length or height, or length and height for specific Lots, Sizes 10x22x14mm, 10x28x14mm, 10x22x15mm
- Product
- Cascadia AN Interbody Convex 10x22x15mm Catalog Number: 6101-2102215NC-G2 Lumbar implants are intervertebral body fusion devices indicated for use with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft when used as anadjunct to fusion in patients with degenerative disc disease.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of MA, MI, NJ, PA, UT and the countries of Italy, Netherlands.
- How much
- 7 units
Cascadia AN Interbody Convex 10x22x14mm Catalog Number: 6101-2102214NC-G2 Lumbar implants are intervertebr
Z-1695-2021 · Class II · reported 2021-06-09 · Terminated
- Why it was recalled
- Mislabeled product labeling references incorrect length or height, or length and height for specific Lots, Sizes 10x22x14mm, 10x28x14mm, 10x22x15mm
- Product
- Cascadia AN Interbody Convex 10x22x14mm Catalog Number: 6101-2102214NC-G2 Lumbar implants are intervertebral body fusion devices indicated for use with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft when used as anadjunct to fusion in patients with degenerative disc disease.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of MA, MI, NJ, PA, UT and the countries of Italy, Netherlands.
- How much
- 3 units
Chesapeake Cervical-Ti 10mm Drill Guide Inserter- a multi-screw construct providing stability to the anterior
Z-1410-2021 · Class II · reported 2021-04-21 · Terminated
- Why it was recalled
- Interference condition between the 10mm Drill Guide Inserter and the inner shaft that mates with the inserters. As a result, inner shafts do not mate flush with the collet inserters, and splaying of the collet. If implant falls into the spinal cord may cause a spinal cord injury or Incomplete seating of the inserter to the implant could result in misalignment of the screws with the inserter screw guide, potentially causing malposition of the screws into the adjacent vertebral bodies
- Product
- Chesapeake Cervical-Ti 10mm Drill Guide Inserter- a multi-screw construct providing stability to the anterior column (cervical spine) Catalog number: 3608-90033
- Where it went
- PA
- How much
- 1 unit
Yukon Polyaxial Screw Size 4.0x26mm (catalog number 7601-04026), component of YUKON Spinal System
Z-0931-2021 · Class II · reported 2021-02-03 · Terminated
- Why it was recalled
- The Yukon polyaxial screw used in the Yukon Spinal System was incorrectly laser marked with size 5.0 mm (diameter) & 24 mm (length).
- Product
- Yukon Polyaxial Screw Size 4.0x26mm (catalog number 7601-04026), component of YUKON Spinal System
- Where it went
- Domestic: Nationwide Foreign: Australia, Canada, Netherlands, Singapore
- How much
- 248 units/1 each
Everest MI XT Tab Removal Tool (i.e. All-In-One Tab Removal Tool)-used to break the tab off the Everest XT scr
Z-0542-2021 · Class II · reported 2020-12-16 · Terminated
- Why it was recalled
- All-in-One Tab Removal Tool may cause cracking and/or separating at the threaded interface between tool s metal shaft component and handle, resulting in leakage of Loctite from handle cracks
- Product
- Everest MI XT Tab Removal Tool (i.e. All-In-One Tab Removal Tool)-used to break the tab off the Everest XT screw by rocking the instrument back and forth until the entire tab is fully detached from the screw head. Catalog number: 5101-90195
- Where it went
- Nationwide Foreign: Netherlands, Spain, UK, Belgium, Germany, Italy, Norway and Sweden
- How much
- 197 units
Yukon OCT Spinal System Yukon Anti-Torque Alignment Tube - Product Usage: The Anti-Torque Alignment Tube (the
Z-0520-2021 · Class II · reported 2020-12-09 · Terminated
- Why it was recalled
- Shafts do not pass through the inner cannula of the Anti-Torque Alignment Tube from Lot KTGG due to a manufacturing nonconformance. This could result in a delay of surgery or result in surgical intervention later on.
- Product
- Yukon OCT Spinal System Yukon Anti-Torque Alignment Tube - Product Usage: The Anti-Torque Alignment Tube (the Alignment Tube ) provides counter rotation to reduce the torque applied to the anatomy when the size 15 Torque limiting shaft (the shaft is used to tighten the Yukon Set Screw to the specified torque.
- Where it went
- US Nationwide distribution.
- How much
- 4 devices
K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog
Z-2009-2020 · Class II · reported 2020-05-27 · Terminated
- Why it was recalled
- The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Product
- K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog # 5203-90109, UDI # 10888857108226, Size 24x6mm, 6 Product Usage: Used to simulate the height of the cage and expand the spine disc space to determine the optimal implant size base on patient anatomy.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of Colorado, Michigan, Texas, California, New York, Florida, Maryland, Oregon, Kansas, Montana, Missouri and the country of Europe.
- How much
- N/A
K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog
Z-2014-2020 · Class II · reported 2020-05-27 · Terminated
- Why it was recalled
- The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Product
- K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog # 5203-90123, UDI # 10888857108363, Size 24x6mm, 12 Product Usage: Used to simulate the height of the cage and expand the spine disc space to determine the optimal implant size base on patient anatomy.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of Colorado, Michigan, Texas, California, New York, Florida, Maryland, Oregon, Kansas, Montana, Missouri and the country of Europe.
- How much
- N/A
K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog
Z-2031-2020 · Class II · reported 2020-05-27 · Terminated
- Why it was recalled
- The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Product
- K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog # 5203-90181, UDI # 10888857108943, Size 28x8mm, 18 Product Usage: Used to simulate the height of the cage and expand the spine disc space to determine the optimal implant size base on patient anatomy.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of Colorado, Michigan, Texas, California, New York, Florida, Maryland, Oregon, Kansas, Montana, Missouri and the country of Europe.
- How much
- N/A
K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog
Z-2021-2020 · Class II · reported 2020-05-27 · Terminated
- Why it was recalled
- The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Product
- K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog # 5203-90151, UDI # 10888857108646, Size 28x6mm, 6 Product Usage: Used to simulate the height of the cage and expand the spine disc space to determine the optimal implant size base on patient anatomy.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of Colorado, Michigan, Texas, California, New York, Florida, Maryland, Oregon, Kansas, Montana, Missouri and the country of Europe.
- How much
- N/A
K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog
Z-2023-2020 · Class II · reported 2020-05-27 · Terminated
- Why it was recalled
- The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Product
- K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog # 5203-90155, UDI # 10888857108684, Size 28x10mm, 6 Product Usage: Used to simulate the height of the cage and expand the spine disc space to determine the optimal implant size base on patient anatomy.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of Colorado, Michigan, Texas, California, New York, Florida, Maryland, Oregon, Kansas, Montana, Missouri and the country of Europe.
- How much
- N/A
K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog
Z-2015-2020 · Class II · reported 2020-05-27 · Terminated
- Why it was recalled
- The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Product
- K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog # 5203-90125, UDI # 10888857108387, Size 24x8mm, 12 Product Usage: Used to simulate the height of the cage and expand the spine disc space to determine the optimal implant size base on patient anatomy.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of Colorado, Michigan, Texas, California, New York, Florida, Maryland, Oregon, Kansas, Montana, Missouri and the country of Europe.
- How much
- N/A
K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog
Z-2022-2020 · Class II · reported 2020-05-27 · Terminated
- Why it was recalled
- The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Product
- K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog # 5203-90153, UDI # 10888857108660, Size 28x8mm, 6 Product Usage: Used to simulate the height of the cage and expand the spine disc space to determine the optimal implant size base on patient anatomy.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of Colorado, Michigan, Texas, California, New York, Florida, Maryland, Oregon, Kansas, Montana, Missouri and the country of Europe.
- How much
- N/A
K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog
Z-2029-2020 · Class II · reported 2020-05-27 · Terminated
- Why it was recalled
- The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Product
- K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog # 5203-90177, UDI # 10888857108905, Size 28x4mm, 18 Product Usage: Used to simulate the height of the cage and expand the spine disc space to determine the optimal implant size base on patient anatomy.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of Colorado, Michigan, Texas, California, New York, Florida, Maryland, Oregon, Kansas, Montana, Missouri and the country of Europe.
- How much
- N/A
K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog
Z-2020-2020 · Class II · reported 2020-05-27 · Terminated
- Why it was recalled
- The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Product
- K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog # 5203-90149, UDI # 10888857108622, Size 28x4mm, 6 Product Usage: Used to simulate the height of the cage and expand the spine disc space to determine the optimal implant size base on patient anatomy.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of Colorado, Michigan, Texas, California, New York, Florida, Maryland, Oregon, Kansas, Montana, Missouri and the country of Europe.
- How much
- N/A
K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog
Z-2026-2020 · Class II · reported 2020-05-27 · Terminated
- Why it was recalled
- The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Product
- K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog # 5203-90165, UDI # 10888857108783, Size 28x6mm, 12 Product Usage: Used to simulate the height of the cage and expand the spine disc space to determine the optimal implant size base on patient anatomy.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of Colorado, Michigan, Texas, California, New York, Florida, Maryland, Oregon, Kansas, Montana, Missouri and the country of Europe.
- How much
- N/A
K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog
Z-2008-2020 · Class II · reported 2020-05-27 · Terminated
- Why it was recalled
- The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Product
- K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog # 5203-90107, UDI # 10888857108202, Size 24x4mm, 6 Product Usage: Used to simulate the height of the cage and expand the spine disc space to determine the optimal implant size base on patient anatomy.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of Colorado, Michigan, Texas, California, New York, Florida, Maryland, Oregon, Kansas, Montana, Missouri and the country of Europe.
- How much
- 211
K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog
Z-2027-2020 · Class II · reported 2020-05-27 · Terminated
- Why it was recalled
- The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Product
- K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog # 5203-90167, UDI # 10888857108806, Size 28x8mm, 12 Product Usage: Used to simulate the height of the cage and expand the spine disc space to determine the optimal implant size base on patient anatomy.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of Colorado, Michigan, Texas, California, New York, Florida, Maryland, Oregon, Kansas, Montana, Missouri and the country of Europe.
- How much
- N/A
K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog
Z-2016-2020 · Class II · reported 2020-05-27 · Terminated
- Why it was recalled
- The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Product
- K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog # 5203-90127, UDI # 10888857108400, Size 24x10mm, 12 Product Usage: Used to simulate the height of the cage and expand the spine disc space to determine the optimal implant size base on patient anatomy.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of Colorado, Michigan, Texas, California, New York, Florida, Maryland, Oregon, Kansas, Montana, Missouri and the country of Europe.
- How much
- N/A
K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog
Z-2028-2020 · Class II · reported 2020-05-27 · Terminated
- Why it was recalled
- The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Product
- K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog # 5203-90169, UDI # 10888857108820, Size 28x10mm, 12 Product Usage: Used to simulate the height of the cage and expand the spine disc space to determine the optimal implant size base on patient anatomy.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of Colorado, Michigan, Texas, California, New York, Florida, Maryland, Oregon, Kansas, Montana, Missouri and the country of Europe.
- How much
- N/A
K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog
Z-2025-2020 · Class II · reported 2020-05-27 · Terminated
- Why it was recalled
- The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Product
- K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog # 5203-90163, UDI # 10888857108769, Size 28x4mm, 12 Product Usage: Used to simulate the height of the cage and expand the spine disc space to determine the optimal implant size base on patient anatomy.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of Colorado, Michigan, Texas, California, New York, Florida, Maryland, Oregon, Kansas, Montana, Missouri and the country of Europe.
- How much
- N/A
K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog
Z-2017-2020 · Class II · reported 2020-05-27 · Terminated
- Why it was recalled
- The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Product
- K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog # 5203-90135, UDI # 10888857108486, Size 24x4mm, 18 Product Usage: Used to simulate the height of the cage and expand the spine disc space to determine the optimal implant size base on patient anatomy.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of Colorado, Michigan, Texas, California, New York, Florida, Maryland, Oregon, Kansas, Montana, Missouri and the country of Europe.
- How much
- N/A
K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog
Z-2024-2020 · Class II · reported 2020-05-27 · Terminated
- Why it was recalled
- The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Product
- K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog # 5203-90157, UDI # 10888857108707, Size 28x12mm, 6 Product Usage: Used to simulate the height of the cage and expand the spine disc space to determine the optimal implant size base on patient anatomy.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of Colorado, Michigan, Texas, California, New York, Florida, Maryland, Oregon, Kansas, Montana, Missouri and the country of Europe.
- How much
- N/A
K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog
Z-2013-2020 · Class II · reported 2020-05-27 · Terminated
- Why it was recalled
- The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Product
- K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog # 5203-90121, UDI # 10888857108349, Size 24x4mm, 12 Product Usage: Used to simulate the height of the cage and expand the spine disc space to determine the optimal implant size base on patient anatomy.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of Colorado, Michigan, Texas, California, New York, Florida, Maryland, Oregon, Kansas, Montana, Missouri and the country of Europe.
- How much
- N/A
K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog
Z-2030-2020 · Class II · reported 2020-05-27 · Terminated
- Why it was recalled
- The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Product
- K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog # 5203-90179, UDI # 10888857108929, Size 28x6mm, 18 Product Usage: Used to simulate the height of the cage and expand the spine disc space to determine the optimal implant size base on patient anatomy.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of Colorado, Michigan, Texas, California, New York, Florida, Maryland, Oregon, Kansas, Montana, Missouri and the country of Europe.
- How much
- N/A
Showing the 30 most recent of 39. The rest are searchable at the FDA enforcement report database.
Read these counts with three things in mind
A firm with more recalls usually makes more products. A contract packer runs
hundreds of lines. It will out-recall a company with one product whatever either
does. This is not a safety ranking.
Most recalls here are voluntary. A firm found its own fault and pulled the stock.
That is the system working. It lands in this database looking the same as a recall
the FDA had to push for.
Terminated means the FDA considers the recall finished. It is not a verdict.
It does not mean anyone was harmed.
Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.