Hill-Rom, Inc.

Hill-Rom, Inc.: 37 FDA recall reports

Every medical device recall the FDA has published for Hill-Rom, Inc.. The fault, and how the agency graded it.

37Reports
1Class I
7Ongoing
2022-03-23Most recent
1 Class I on this page

Class I means the FDA judged there was a reasonable probability of serious harm or death. Check the lot numbers against anything you hold.

What the classes mean

The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.

ClassWhat the FDA is saying
Class IThere is a reasonable probability that using this will cause serious harm or death.
Class IIIt may cause temporary or reversible harm. Serious harm is unlikely.
Class IIIIt breaks a rule but is unlikely to hurt anyone. Often a labelling fault.

Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.

By class

ClassReports
Class II36
Class I1

By year reported

YearReports
20122
20139
20141
20153
20165
20171
20181
20204
20218
20223
Checking whether it affects you

The lot or code number is what matters. Not the product name. A recall covers specific lots. Yours may not be one of them.

Check the code on your package against the report. The FDA keeps the current list at fda.gov/safety/recalls.

The reports

The text below is quoted from the FDA record word for word.

(1)Traverse Rail Carrier E-System, wide 31017XX component to Liko overhead lift, (2)Traverse Rail Carrier 31

Z-0783-2022  ·  Class II  ·  reported 2022-03-23  ·  Terminated

Why it was recalled
Traverse rail carriage delivered with non-conforming screws are too short to ensure appropriate installation. There is potential that the screws are unable to secure the traverse rail appropriately and could loosen over time causing the rail system to fall, potentially causing injury to caregivers and/or patients
Product
(1)Traverse Rail Carrier E-System, wide 31017XX component to Liko overhead lift, (2)Traverse Rail Carrier 31025XX component to Liko overhead lift. Intended for to make safe lifting of a patient possible. Affected Devices: 3101704 TRAV RAIL CARR E-SYST STD 3102514 TR CARRIER SLIM/RAISED, LR 3101705 TRAV RAIL CARR ESYST WIDE 3102517 TR CARR. NORMAL/LOWER. LR 3102506 TR CARRIER WIDE/RAISED ML 3102519 TR CARR. LOW.LR 30MM/PAIR 3102511 TR CARRIER NORMAL, LR 3102531 TRAVERSE RL CARRIER STD/RAISED 3102512 TR CARRIER WIDE, LR 3102532 TRAVERSE RL CARRIER WD/RAISED 3102513 TR CARRIER SLIM, LR 3102562 TR CARRIER, PARALLEL/WIDE
Where it went
Worldwide distribution - US Nationwide and the countries of AT, CA, CZ, DE, DK, ES, FI, FR, GB, HK, IE, IL, IT, JP, NL, NO, NZ, RU, SE, SG, SK.
How much
Traverse Rail Carrier E-System, wide 31017XX 5 Traverse Rail Carrier 31025XX 1798

Liko Mobile Lifts: Primarily intended for use in nursing homes LikoM230 Model: 2050015

Z-0640-2022  ·  Class II  ·  reported 2022-02-23  ·  Ongoing

Why it was recalled
Missing bushing on Liko M220 and M230 causing wear and potential for patient falling
Product
Liko Mobile Lifts: Primarily intended for use in nursing homes LikoM230 Model: 2050015
Where it went
Worldwide distribution - US Nationwide and the countries of Austria, Austria, Bahrain, Canada, Switzerland, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, Italy, Norway, Saudi Arabia, Sweden, Slovenia, Taiwan (Province of China), South Africa, Oman.
How much
1215 units

Liko Mobile Lifts: Primarily intended for use in nursing homes Liko M220, Model 2050010

Z-0639-2022  ·  Class II  ·  reported 2022-02-23  ·  Ongoing

Why it was recalled
Missing bushing on Liko M220 and M230 causing wear and potential for patient falling
Product
Liko Mobile Lifts: Primarily intended for use in nursing homes Liko M220, Model 2050010
Where it went
Worldwide distribution - US Nationwide and the countries of Austria, Austria, Bahrain, Canada, Switzerland, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, Italy, Norway, Saudi Arabia, Sweden, Slovenia, Taiwan (Province of China), South Africa, Oman.
How much
5470 units

These are communication systems, labeled as: a. NaviCare Nurse Call, Part Numbers: 3.6, 3.7, 3.9 b. Voalte

Z-0402-2022  ·  Class II  ·  reported 2021-12-29  ·  Terminated

Why it was recalled
An unrecoverable error may occur where a component will fail to complete the reboot cycle, become non-functional and must be replaced.
Product
These are communication systems, labeled as: a. NaviCare Nurse Call, Part Numbers: 3.6, 3.7, 3.9 b. Voalte Nurse Call, Part Number 4.0 *Product was rebranded from NaviCare to Voalte in 2020
Where it went
Worldwide distribution - US Nationwide.
How much
1,050 units

LikoStretch Mod 600 IC Model Number: 3156065 LikoStretch Mod 600 IC, Wide Model Number: 3156065B (obsolete si

Z-2601-2021  ·  Class II  ·  reported 2021-10-06  ·  Ongoing

Why it was recalled
Stretch Leveller has the potential to lose the function of maintaining a patient in the horizontal position when lifting and in motion. When this function is lost the patient can slide backwards and potentially may fall out of the Oct lift sheet back onto the surface it is being lifted from.
Product
LikoStretch Mod 600 IC Model Number: 3156065 LikoStretch Mod 600 IC, Wide Model Number: 3156065B (obsolete since 12JUN2020) - The combination of this products allows the user to easily adjust/control the center of gravity during lift of patient.
Where it went
Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, DE, IL, ME, MI, MO, MS, NJ, OH, PA, VA, WA and the countries of Argentina, Australia, Austria, Bolivia, Canada, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Greece, Hong Kong, India, Ireland, Italy, Japan, Jordan, Kuwait, Latvia, Lithuania, Mexico, Netherlands, Norway, Poland, Portugal, Serbia, Slovakia, Spain, Sweden, Switzerland, Taiwan, Province of China, Thailand, Turkey, United Arab Emirates.
How much
Undetermined

LikoStretch 1900 Model Number: 3156051 - The combination of this products allows the user to easily adjust/con

Z-2600-2021  ·  Class II  ·  reported 2021-10-06  ·  Ongoing

Why it was recalled
Stretch Leveller has the potential to lose the function of maintaining a patient in the horizontal position when lifting and in motion. When this function is lost the patient can slide backwards and potentially may fall out of the Oct lift sheet back onto the surface it is being lifted from.
Product
LikoStretch 1900 Model Number: 3156051 - The combination of this products allows the user to easily adjust/control the center of gravity during lift of patient.
Where it went
Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, DE, IL, ME, MI, MO, MS, NJ, OH, PA, VA, WA and the countries of Argentina, Australia, Austria, Bolivia, Canada, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Greece, Hong Kong, India, Ireland, Italy, Japan, Jordan, Kuwait, Latvia, Lithuania, Mexico, Netherlands, Norway, Poland, Portugal, Serbia, Slovakia, Spain, Sweden, Switzerland, Taiwan, Province of China, Thailand, Turkey, United Arab Emirates.
How much
9 units

Stretch Leveller Model Number: 3156200 - The combination of this products allows the user to easily adjust/con

Z-2599-2021  ·  Class II  ·  reported 2021-10-06  ·  Ongoing

Why it was recalled
Stretch Leveller has the potential to lose the function of maintaining a patient in the horizontal position when lifting and in motion. When this function is lost the patient can slide backwards and potentially may fall out of the Oct lift sheet back onto the surface it is being lifted from.
Product
Stretch Leveller Model Number: 3156200 - The combination of this products allows the user to easily adjust/control the center of gravity during lift of patient.
Where it went
Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, DE, IL, ME, MI, MO, MS, NJ, OH, PA, VA, WA and the countries of Argentina, Australia, Austria, Bolivia, Canada, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Greece, Hong Kong, India, Ireland, Italy, Japan, Jordan, Kuwait, Latvia, Lithuania, Mexico, Netherlands, Norway, Poland, Portugal, Serbia, Slovakia, Spain, Sweden, Switzerland, Taiwan, Province of China, Thailand, Turkey, United Arab Emirates.
How much
122 units

FlexoStretch Model Number: 3156057 - The combination of this products allows the user to easily adjust/control

Z-2598-2021  ·  Class II  ·  reported 2021-10-06  ·  Ongoing

Why it was recalled
Stretch Leveller has the potential to lose the function of maintaining a patient in the horizontal position when lifting and in motion. When this function is lost the patient can slide backwards and potentially may fall out of the Oct lift sheet back onto the surface it is being lifted from.
Product
FlexoStretch Model Number: 3156057 - The combination of this products allows the user to easily adjust/control the center of gravity during lift of patient.
Where it went
Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, DE, IL, ME, MI, MO, MS, NJ, OH, PA, VA, WA and the countries of Argentina, Australia, Austria, Bolivia, Canada, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Greece, Hong Kong, India, Ireland, Italy, Japan, Jordan, Kuwait, Latvia, Lithuania, Mexico, Netherlands, Norway, Poland, Portugal, Serbia, Slovakia, Spain, Sweden, Switzerland, Taiwan, Province of China, Thailand, Turkey, United Arab Emirates.
How much
129 units

OctoStretch with Stretch Leveller - The combination of this products allows the user to easily adjust/control

Z-2597-2021  ·  Class II  ·  reported 2021-10-06  ·  Ongoing

Why it was recalled
Stretch Leveller has the potential to lose the function of maintaining a patient in the horizontal position when lifting and in motion. When this function is lost the patient can slide backwards and potentially may fall out of the Oct lift sheet back onto the surface it is being lifted from.
Product
OctoStretch with Stretch Leveller - The combination of this products allows the user to easily adjust/control the center of gravity during lift of patient. Model Number: 3156056
Where it went
Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, DE, IL, ME, MI, MO, MS, NJ, OH, PA, VA, WA and the countries of Argentina, Australia, Austria, Bolivia, Canada, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Greece, Hong Kong, India, Ireland, Italy, Japan, Jordan, Kuwait, Latvia, Lithuania, Mexico, Netherlands, Norway, Poland, Portugal, Serbia, Slovakia, Spain, Sweden, Switzerland, Taiwan, Province of China, Thailand, Turkey, United Arab Emirates.
How much
440 units

Affinity Four Birthing Bed-used as a birthing bed for women of child bearing age in an LDR (Labor, Delivery, R

Z-2156-2021  ·  Class II  ·  reported 2021-08-04  ·  Terminated

Why it was recalled
Latch mechanism on the Lift-Off foot section of the Affinity Four Birthing Bed is damaged, it could potentially cause the installed foot section to be improperly engaged onto the bed, could result in injury to the user due to a fall.
Product
Affinity Four Birthing Bed-used as a birthing bed for women of child bearing age in an LDR (Labor, Delivery, Recovery) or LDRP (Labor, Delivery, Recovery, Postpartum) setting within the acute care Labor and Delivery market Model Numbers: P3700B, P3700C, P3700D, & P3700E
Where it went
Worldwide distribution - US Nationwide distribution in the states of AL, CA, IA, MS, NC, NY, OH, PA, TX, VA and the countries of Australia, Austria, Azerbaijan, Belgium, Chile, China, Denmark, Finland, France, Germany, Ireland, Israel, Italy, Jordan, Kuwait, Netherlands, New Zealand, Norway, Portugal, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Taiwan, United Arab Emirates, United Kingdom.
How much
1,096 Beds
Class I

Liko Multirall 200 overhead lift- overhead lift is a general-purpose lift with the intended use in healthcare

Z-0925-2021  ·  Class I  ·  reported 2021-02-10  ·  Terminated

Why it was recalled
Multirall Q-link strap lock has not been fully engaged in the S65 rail carriage hook by user, leading to the Multirall lift becoming detached from the rail system, potentially resulting in the fall of both the lift motor and the patient
Product
Liko Multirall 200 overhead lift- overhead lift is a general-purpose lift with the intended use in healthcare, intensive care and rehabilitation. Affected devices: Liko Multirall 200 (product number 3130001) Universal SlingBar 450 R2R (product number 3156095) Universal SlingBar 350 R2R (product number 3156094) Carriage D45 with Double Hook (product number 3136100) Extension belt 300-400 mm (product number 3136226) Extension belt 400-600 mm (product number 3136227) Extension belt 600-1000 mm (product number 3136228) Extension belt 1000-1400mm (product number 3136229)
Where it went
Worldwide distribution: US (Nationwide) and countries of::AT, AU, BR, CA, CH, CZ, DE, DK, ES, FI, FR, GB, GR, IE, IL, IS, IT, JP, LV, NL, NO, NZ, SE & SK.
How much
10470 units Expanded Recall: 154 units

Progressa Bed, Catalog No. P7500 - Product Usage: intended to be used to treat or prevent pulmonary or other c

Z-2737-2020  ·  Class II  ·  reported 2020-08-05  ·  Terminated

Why it was recalled
Affected beds may have screws installed that do not meet material tensile strength requirements. Failure of this screw could result in serious patient or user injury.
Product
Progressa Bed, Catalog No. P7500 - Product Usage: intended to be used to treat or prevent pulmonary or other complications associated with immobility; to treat or prevent pressure ulcers; or for any other use where medical benefits may be derived from either Continuous Lateral Rotation Therapy or Percussion/Vibration Therapy.
Where it went
Worldwide distribution - US nationwide distribution including in the states of AR, AZ, CA, DE, IN, KY, MD, ME, MI, MO, NC, NM, NY, OH, OK, PA, SC, TN, TX, VA, WA, WI, WV and the country of Canada.
How much
257

Centrella Smart+ Bed, Catalog No. P7900 - Product Usage: intended to be used to treat or prevent pulmonary or

Z-2738-2020  ·  Class II  ·  reported 2020-08-05  ·  Terminated

Why it was recalled
Affected beds may have screws installed that do not meet material tensile strength requirements. Failure of this screw could result in serious patient or user injury.
Product
Centrella Smart+ Bed, Catalog No. P7900 - Product Usage: intended to be used to treat or prevent pulmonary or other complications associated with immobility; to treat or prevent pressure ulcers; or for any other use where medical benefits may be derived from either Continuous Lateral Rotation Therapy or Percussion/Vibration Therapy.
Where it went
Worldwide distribution - US nationwide distribution including in the states of AR, AZ, CA, DE, IN, KY, MD, ME, MI, MO, NC, NM, NY, OH, OK, PA, SC, TN, TX, VA, WA, WI, WV and the country of Canada.
How much
1033

Centrella Smart+ Bed, Catalog Number P7900

Z-2246-2020  ·  Class II  ·  reported 2020-06-10  ·  Terminated

Why it was recalled
The Bed Exit System may fail to send a remote alert through the nurse call system if a remote alert was previously sent and cancelled at the in-room nurse call wall unit.
Product
Centrella Smart+ Bed, Catalog Number P7900
Where it went
Domestic distribution nationwide. Foreign distribution to Canada and Australia.
How much
4381 beds

NaviCare Patient Safety Software Versions 3.5.400-3.9.300. The NaviCare Patient Safety (Patient Safety) Sof

Z-1554-2020  ·  Class II  ·  reported 2020-04-01  ·  Terminated

Why it was recalled
There is a software issue which may result in failure to monitor or control patient's bed exit mechanism.
Product
NaviCare Patient Safety Software Versions 3.5.400-3.9.300. The NaviCare Patient Safety (Patient Safety) Software is compatible with the following beds: 1. VersaCare Revision A-J with upgrade kit Revision K with UCB 1.18 2. Centrella (all versions) 3. Progressa (firmware version 1.15.0.0)
Where it went
Colorado, Georgia, Indiana, Louisiana, North Carolina, Nevada, Oregon, South Carolina, Tennessee, Virginia, Washington.
How much
35 Software Systems

Progressa Bed - AC-powered adjustable hospital bed

Z-1319-2018  ·  Class II  ·  reported 2018-04-18  ·  Terminated

Why it was recalled
A component used within the braking system may break resulting in one or more brake pedals on the bed staying in brake or not being engaged which could result in unwanted bed movement and possible patient or user injury.
Product
Progressa Bed - AC-powered adjustable hospital bed
Where it went
Worldwide distribution. US nationwide, Canada, Chile, Colombia, Costa Rica, France, Hong Kong, Israel, Jamaica, Japan, Jordan, Malaysia, Mexico, Oman, Panama, Russian Federation, Saudi Arabia, Singapore, Spain, Taiwan, UAE, and UK.
How much
2117

TotalCare Bed Systems, Model Numbers P1900, PR1900, and U1900 with 3-Level Patient Position Monitoring and Sc

Z-1816-2017  ·  Class II  ·  reported 2017-04-19  ·  Terminated

Why it was recalled
The bed exit alarm may not function properly if the alarm has been armed for a period longer than seventy-two (72) hours without the alarm being triggered by patient movement or removal of patient weight within that time period.
Product
TotalCare Bed Systems, Model Numbers P1900, PR1900, and U1900 with 3-Level Patient Position Monitoring and Scale Upgrade Kits, Model Numbers P1947A and P1948A).
Where it went
US Nationwide. AUSTRALIA BANGLADESH BRAZIL CANADA CHINA GERMANY INDIA ITALY KUWAIT PAKISTAN SAUDI ARABIA THAILAND UNITED ARAB EMIRATES CYPRUS GREECE NETHERLANDS
How much
3037

Product Number 3102071 Ceiling Bracket, Arch 71 White with use on ceiling installed rail systems: Multirall 2

Z-0326-2017  ·  Class II  ·  reported 2016-11-09  ·  Terminated

Why it was recalled
Ceiling Bracket 71 was shipped with a larger center hole. A potentially hazardous situation exists that if the Ceiling Bracket 71 with a larger center hole is used for a concrete ceiling Ultra System installation with the M8 screw and the safe working load is exceeded, the system could fall.
Product
Product Number 3102071 Ceiling Bracket, Arch 71 White with use on ceiling installed rail systems: Multirall 200, Likorall 200, 242, 243, 250, and LikoGuard L and XL.
Where it went
WA NC OH PA MI Australia Sweden
How much
188

Affinity Four Birthing Bed Model P3700 The primary packaging is pictured below, and utilizes a metal pack

Z-2750-2016  ·  Class II  ·  reported 2016-09-14  ·  Terminated

Product
Affinity Four Birthing Bed Model P3700 The primary packaging is pictured below, and utilizes a metal packing brace, standard corrugated cardboard and plastic banding. When possible these beds are shipped as a quantity of two (2) together as shown with the nylon strapping. These are shipped in standard freight containers. There is no secondary packaging.
Where it went
US Distribution to the states of : CA, FL, IN, KS, LA, MI, MO, NC, NY, OK, OR, PA, TN, TX, WA, WY Foreign:BULGARIA, INDIA, UNITED ARAB EMIRATES
How much
94 beds

Volker 2080, 2082, S 280 and S 282 series beds with lateral release button. A/C Powered Adjustable Hospital B

Z-2376-2016  ·  Class II  ·  reported 2016-08-17  ·  Terminated

Why it was recalled
The current construction of the side rail latch may not be able to support the full patient weight and may give way allowing the side rail to abruptly lower and the patient to fall, potentially resulting in harm. To date there has been one report of a related injury.
Product
Volker 2080, 2082, S 280 and S 282 series beds with lateral release button. A/C Powered Adjustable Hospital Bed Product Usage: Volker model 2080/3010/3080 nursing beds are intended for the care of residents in rooms in nursing homes, in residential homes for the elderly and in suitable rooms in private residences. They may not, however, be used for transportation purposes. Any deviation from this use is expressly excluded from possible liability claims.
Where it went
Worldwide Distribution - US Nationwide and the countries of: Canada, AUSTRALIA, CHINA, HONG KONG, THAILAND, POLAND, RUSSIAN FEDERATION, SLOVENIA, UNITED ARAB EMIRATES, AUSTRIA, BELGIUM, DENMARK, FINLAND, FRANCE, GERMANY, ICELAND, IRELAND, ISRAEL, ITALY, LUXEMBOURG, NETHERLANDS, NORWAY, SPAIN, SWITZERLAND, and UNITED KINGDOM.
How much
24262

Liko Universal SlingBars Universal SlingBar" 350, Universal SlingBar" 350 QRH, Universal SlingBar" 350

Z-1916-2016  ·  Class II  ·  reported 2016-06-15  ·  Terminated

Why it was recalled
The center bolt of the sling bar, which connects the bar to the patient lift, has failed during use. If the sling bar is not used as intended the bolt may be weakened. If the bolt is weakened there is a potential risk for breakage with the result of a free fall of the patient. This hazard could cause potentially minor to catastrophic injuries to a patient.
Product
Liko Universal SlingBars Universal SlingBar" 350, Universal SlingBar" 350 QRH, Universal SlingBar" 350 R2R, Universal SlingBar" 450, Universal SlingBar" 450 QRH, Universal SlingBar" 450 R2R, Universal SlingBar" 600 Universal SlingBar" 600 QRH Component of Liko Lifts Product Usage: Universal SlingBar 350 It is commonly used for lifting children. Universal SlingBar 350 is available with three assembly options - fixed assembly, Quick-release Hook (QRH) or Room 2 Room (R2R). Universal SlingBar 450 Is appropriate in most cases and with most slings. Is installed on most lifts on delivery (does not apply to Viking XL). Universal SlingBar 450 is available with three assembly options - fixed assembly, Quick-release Hook (QRH) or Room 2 Room (R2R). Universal SlingBar 600 It is intended for use with the ComfortSling Plus mod 300/310 and 350/360, or with slings the intention of which is to overlap the leg supports, Universal SlingBar 600 is available with two assembly options - fixed assembly and Quick-release Hook (QRH).
Where it went
Worldwide and US Nationwide Distribution AE, AO, AR , AT , AU , AZ , BD , BE , BG , BH , BM , BN , BR , CA , CH , CL , CN , CO , CS , CZ , DE , DK , EG , ES , FI , FR , GB , GR , HK , HR , HU , ID , IE , IL , IN , IR , IS , IT , JO , JP , KR , KW , LB , LI , LK , LT , LU , LV , LY , MA , MT , MX , MY , NL , NO , NZ , OM , PH , PK , PL , PT , QA , RO , RS , RU, SA , SE , SG , SI , SK , SY , TH , TN , TR , TW , UA , US , UY and ZA.
How much
163,561

Hill-Rom Bariatric Power Recliner, Product Model # P9096 Hill-Roms 660 lb Powered Bariatric Recliner was cre

Z-0860-2016  ·  Class II  ·  reported 2016-03-02  ·  Terminated

Why it was recalled
The recliners were shipped without the foot rest ottoman linkage side shields, which could result in consumer injury.
Product
Hill-Rom Bariatric Power Recliner, Product Model # P9096 Hill-Roms 660 lb Powered Bariatric Recliner was created to provide an aesthetically pleasing design that will enhance the overall experience for the bariatric patient. Using spring box construction and an exceptionally smooth mechanism, the patient is able to move into a reclined position with little effort and infinite positions up to 38 degrees. An independent footrest enables patients to be in a fully seated position with their feet up for added comfort.
Where it went
Worldwide Distribution - US including AL, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, VT, WA, WI, WV, WY, VA Medical Center and Internationally to Canada and Mexico. **Consumer Recommended Depth - Consumers/User**
How much
216

Hill-Rom 100 Low Bed (GPAC) AC Powered adjustable hospital bed

Z-0274-2016  ·  Class II  ·  reported 2015-11-25  ·  Terminated

Why it was recalled
Hill-Rom is voluntarily recalling the Hill-Rom 100 Low Bed due to complaints of patient injury when exiting the beds set at the lowest position setting.
Product
Hill-Rom 100 Low Bed (GPAC) AC Powered adjustable hospital bed
Where it went
Worldwide Distribution - USA (nationwide) to the states of : AL, AR, AZ. CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MS, MT, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WI and WV., and to the countries of : Canada and France.
How much
12251 units

Steriseal-Cannula Ophthalmic; Bi-Manual Irrigating/Aspirating Handpiece.

Z-1820-2015  ·  Class II  ·  reported 2015-07-01  ·  Terminated

Why it was recalled
The metal cannula attached to the handpiece can rotate and become dislodged from the plastic handpiece.
Product
Steriseal-Cannula Ophthalmic; Bi-Manual Irrigating/Aspirating Handpiece.
Where it went
Worldwide Distribution-US (nationwide) including the states of MI & PA, and the countries of Belgium, England, Italy, Czech Republic, Turkey, Greece, Ireland, Scotland, Spain, and Lebanon.
How much
66691 units

Viking M, L, and XL Mobile Lifts. Non-AC-powered patient lift.

Z-1818-2015  ·  Class II  ·  reported 2015-06-24  ·  Terminated

Why it was recalled
Complaints (including one reported death) allegedly of the lift arm drifting down suddenly. If the lift arm assembly is manually lifted, the actuator can become damaged and get stuck in the highest position. If a patient is lifted into the sling while the actuator is stuck, there is a potential risk of a free fall of the patient, resulting in minor or potentially catastrophic injuries.
Product
Viking M, L, and XL Mobile Lifts. Non-AC-powered patient lift.
Where it went
Worldwide Distribution -- AE, AO, AR, AT, AU, AZ, BD, BE, BG, BH, BR, CA, CH, CL, CN, CO, CS, CZ, DE, DK, EG, ES, FI, FR, GR, HK, HR, HU, ID, IE, IL, IN, IQ, IR, IS, IT, JO, JP, KR, KW, LB, LI, LT, LU, LV, MA, MT, MX, MY, NL, NO, NZ, OM, PH, PL, PT, QA, RO, SA, SE, SG, SI, SK, SY, TH, TN, TR, TW, UK, US, VE, and ZA.
How much
24, 798

Affinity¿ Four Birthing Bed, Product Usage: Product is intended to be used as birthing beds for women of

Z-2562-2014  ·  Class II  ·  reported 2014-09-10  ·  Terminated

Why it was recalled
The attachment latch mechanism can become bent depending on user handling. Latch mechanism bending could lead to improper engagement of the foot section and this could result in injury to the user due to a fall.
Product
Affinity¿ Four Birthing Bed, Product Usage: Product is intended to be used as birthing beds for women of child bearing age in an LDR (Labor, Delivery, Recovery) or LDRP (Labor, Delivery, Recovery, Postpartum) setting within the acute care Labor and Delivery market.
Where it went
Worldwide Distribution - US Nationwide and the following countries: Argentina , Aruba, Australia, Austria, Bahamas, Bahrain, Belgium, Belize, Bermuda, Brazil, Brunei, Bulgaria, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, Equatorial Guinea, Finland, France, Germany, Greece, Guam, Guyana, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Lebanon, Libya, Malaysia, Mauritius, Mexico, Morocco, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turks and Caicos Islands, United Arab Emirates, United Kingdom, US Virgin Islands, Venezuela and Vietnam.
How much
13,000 units

TotalCare Bed used with AccuMax Quantum VPC mattress AccuMax Quantum Complete mattress, and AccuMax Quantum C

Z-1794-2013  ·  Class II  ·  reported 2013-07-31  ·  Terminated

Why it was recalled
When the bed is put into the chair egress position, the magnets on the mattress may not hold the bottom of the mattress flat to the bed frame, and the mattress foot section will extend away from the bed at an angle. A patient attempting to get back into the bed may miss the seat and sit on the foot section instead. The patient can slide down to the floor, or fall, with possible injury.
Product
TotalCare Bed used with AccuMax Quantum VPC mattress AccuMax Quantum Complete mattress, and AccuMax Quantum Convertible mattress The TotalCare Bed System is intended to provide a patient support ideally suited to be used in health care environments. The TotalCare Bed System may be used in a variety of settings including, but not limited to, acute care, including critical care, step down/progressive care, medical/surgical, high acuity sub-acute care, post anesthesia care unit (PACU), and sections of the emergency department (ED). The TotalCare Bed System is capable of being used with a broad patient population as determined appropriate by the caregiver or institution.
Where it went
Worldwide distribution: USA (nationwide) and countries of: Canada, Colombia, Nigeria, Kuwait and Saudi Arabia.
How much
4,121 Units

SlingBar 350. Designed to meet the needs for lifting humans.

Z-1473-2013  ·  Class II  ·  reported 2013-06-12  ·  Terminated

Why it was recalled
Retrospective review found 4 sling bars that may present a potential for impalement if a patient, caregiver or bystander were to fall onto the hooks based on a simulated test method. In effort to eliminate any risk of injury, Hill-Rom is requesting products be removed from field use and replaced with an updated design.
Product
SlingBar 350. Designed to meet the needs for lifting humans.
Where it went
Worldwide Distribution-USA (nationwide) including the states of AZ, CA, CO, FL, IL, IN, KS, LA, MA, MI, MN, MO, NC, NH, NJ, NV, OH, OK, PA, SC, TN, UT, WA, and WI, and the countries of Canada, Australia, Austria, Belgium, Switzerland, Germany Denmark, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, Italy, Japan, Netherlands, Norway, New Zealand, Portugal, Sweden, and Turkey.
How much
3556

SlingBar Standard. Designed to meet the needs for lifting humans.

Z-1476-2013  ·  Class II  ·  reported 2013-06-12  ·  Terminated

Why it was recalled
Retrospective review found 4 sling bars that may present a potential for impalement if a patient, caregiver or bystander were to fall onto the hooks based on a simulated test method. In effort to eliminate any risk of injury, Hill-Rom is requesting products be removed from field use and replaced with an updated design.
Product
SlingBar Standard. Designed to meet the needs for lifting humans.
Where it went
Worldwide Distribution-USA (nationwide) including the states of AZ, CA, CO, FL, IL, IN, KS, LA, MA, MI, MN, MO, NC, NH, NJ, NV, OH, OK, PA, SC, TN, UT, WA, and WI, and the countries of Canada, Australia, Austria, Belgium, Switzerland, Germany Denmark, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, Italy, Japan, Netherlands, Norway, New Zealand, Portugal, Sweden, and Turkey.
How much
726

Hill-Rom 1000 Bed, Product No. P1160. The Hill-Rom 1000 Bed is intended for low to moderate acuity patient i

Z-1466-2013  ·  Class II  ·  reported 2013-06-12  ·  Terminated

Why it was recalled
Complaints have been reported for brakes disengaging when the bed is subjected to a hard jolt or force to the frame.
Product
Hill-Rom 1000 Bed, Product No. P1160. The Hill-Rom 1000 Bed is intended for low to moderate acuity patient in the medical surgical area of the hospital.
Where it went
Worldwide distribution: US (Nationwide) and countries of: Argentina , Bahamas, Bahrain, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, Egypt, Ethiopia, France, Ghana, Guam, Hong Kong, India, Indonesia, Japan, Jordan, Kuwait, Lebanon, Libya, Malaysia, Mexico, N. Mariana Islands, Pakistan, panama, Paraguay, Puerto Rico, Qatar, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, United Arab Emirates, United States, US Virgin Islands, Venezuela and Vietnam.
How much
8,737 devices

Showing the 30 most recent of 37. The rest are searchable at the FDA enforcement report database.

Read these counts with three things in mind

A firm with more recalls usually makes more products. A contract packer runs hundreds of lines. It will out-recall a company with one product whatever either does. This is not a safety ranking.

Most recalls here are voluntary. A firm found its own fault and pulled the stock. That is the system working. It lands in this database looking the same as a recall the FDA had to push for.

Terminated means the FDA considers the recall finished. It is not a verdict. It does not mean anyone was harmed.

Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.

← All device recalls