thejoe
Haemonetics Corporation
Haemonetics Corporation: 31 FDA recall reports
Every medical device recall the FDA has published for Haemonetics Corporation. The fault, and how the agency graded it.
31Reports
0Class I
0Ongoing
2022-07-06Most recent
What the classes mean
The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.
Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.
By class
By year reported
Checking whether it affects you
The lot or code number is what matters. Not the product name. A recall covers
specific lots. Yours may not be one of them.
Check the code on your package against the report. The FDA keeps the current
list at fda.gov/safety/recalls.
The reports
The text below is quoted from the FDA record word for word.
TEG 5000 Functional Fibrinogen Reagent, Item No. 07-034
Z-1310-2022 · Class II · reported 2022-07-06 · Terminated
- Why it was recalled
- Due to a shift in the citrated blood reference range for a normal population, the ranges stated in the IFU are changing. If the clinician relies solely on affected results in comparison to the normal donor reference range, it could lead to misdiagnosis and incorrect treatment or failure to treat.
- Product
- TEG 5000 Functional Fibrinogen Reagent, Item No. 07-034
- Where it went
- Domestic distribution US Nationwide.
- How much
- 3933 kits (US); 2429 kits (OUS)
Product Name: TEG5000 Analyzer (07-022, 07-033) with TEG Analytical Software (07-030, 07-031) and TEG Platelet
Z-1298-2022 · Class II · reported 2022-07-06 · Terminated
- Why it was recalled
- When the TEG 5000 Analyzer including TEG Analytical Software is used with PlateletMapping (ADP or AA) Assays and TEG Mangement Software, incorrect values of the PlateletMapping ADP or AA % Inhibition and % Aggretagion may be displayed on TEG Manager for the affected assays.
- Product
- Product Name: TEG5000 Analyzer (07-022, 07-033) with TEG Analytical Software (07-030, 07-031) and TEG Platelet Mapping Assay - PlateletMapping ADP & AA (07-014), PlateletMapping ADP (07-015), PlateletMapping AA (07-016). Issue occurs when Platelet Mapping Assay results from a TEG 5000 analyzer are viewed using TEG Manager Software (versions 1.1-4.3.1)
- Where it went
- US Nationwide.
- How much
- 111
TEG Manager software versions used in conjunction with TEG 5000 are 1.1.0, 3.0.0, 4.0.0, 4.1.0 and 4.1.1. In
Z-1617-2020 · Class II · reported 2020-04-08 · Terminated
- Why it was recalled
- Software defect in TEG Manager impacts the displayed alert for out of range test results. Due to this defect, reference range values received from TEG 5000 and displayed on the TEG Manager test result screen are rounded to the nearest whole number and lead to TEG Manager displaying an out of range alert when the test result is actually in range, or vice versa.
- Product
- TEG Manager software versions used in conjunction with TEG 5000 are 1.1.0, 3.0.0, 4.0.0, 4.1.0 and 4.1.1. Intended: For in vitro coagulation studies
- Where it went
- US Nationwide distribution.
- How much
- 88 units
JAPAN CELL SAVER ELITE SET - 225ML - Product Usage: its related accessory components are intended for use to r
Z-0904-2020 · Class II · reported 2020-02-05 · Terminated
- Why it was recalled
- Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.
- Product
- JAPAN CELL SAVER ELITE SET - 225ML - Product Usage: its related accessory components are intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, and pumping this processed red cell product to either a bag for gravity reinfusion into the patient or to the arterial line of an extracorporeal circuit for reinfusion into the patient.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.
- How much
- 12370
CS5/5+ Fastpacks, 225 mL, 20¿ Res Autotransfusion Device - Product Usage: its related accessory components are
Z-0890-2020 · Class II · reported 2020-02-05 · Terminated
- Why it was recalled
- Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.
- Product
- CS5/5+ Fastpacks, 225 mL, 20¿ Res Autotransfusion Device - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.
- How much
- 8641
JAPAN DISP SET, 261J, 205J, HAR - Product Usage: its related accessory components are intended to be used for
Z-0900-2020 · Class II · reported 2020-02-05 · Terminated
- Why it was recalled
- Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.
- Product
- JAPAN DISP SET, 261J, 205J, HAR - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma. The recovered blood is processed by a centrifugation and washing procedure before pumping the red cell product to either a bag for gravity reinfusion into the patient or an extracorporeal circuit.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.
- How much
- 739
CS Elite Fastpack, 225 mL, 150 Res Autotransfusion Device - Product Usage: its related accessory components ar
Z-0895-2020 · Class II · reported 2020-02-05 · Terminated
- Why it was recalled
- Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.
- Product
- CS Elite Fastpack, 225 mL, 150 Res Autotransfusion Device - Product Usage: its related accessory components are intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, and pumping this processed red cell product to either a bag for gravity reinfusion into the patient or to the arterial line of an extracorporeal circuit for reinfusion into the patient.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.
- How much
- 42739
Japan DISP SET, 263J, 205J, HAR - Product Usage: its related accessory components are intended to be used for
Z-0906-2020 · Class II · reported 2020-02-05 · Terminated
- Why it was recalled
- Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.
- Product
- Japan DISP SET, 263J, 205J, HAR - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.
- How much
- 1584
CS5/5+ Fastpacks, 125 mL, 20¿ Res Autotransfusion Device- Product Usage: its related accessory components are
Z-0889-2020 · Class II · reported 2020-02-05 · Terminated
- Why it was recalled
- Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.
- Product
- CS5/5+ Fastpacks, 125 mL, 20¿ Res Autotransfusion Device- Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.
- How much
- 1340
CS Elite Fastpack, 125 mL, 150 Res Autotransfusion Device - Product Usage: its related accessory components ar
Z-0896-2020 · Class II · reported 2020-02-05 · Terminated
- Why it was recalled
- Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.
- Product
- CS Elite Fastpack, 125 mL, 150 Res Autotransfusion Device - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.
- How much
- 14698
Cell Saver¿5/5+ Bowl Kit (225 ml) Autotransfusion Device - Product Usage: its related accessory components are
Z-0891-2020 · Class II · reported 2020-02-05 · Terminated
- Why it was recalled
- Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.
- Product
- Cell Saver¿5/5+ Bowl Kit (225 ml) Autotransfusion Device - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.
- How much
- 690043
JAPAN DISP SET, 261J, 205J, 208 - Product Usage: its related accessory components are intended to be used for
Z-0899-2020 · Class II · reported 2020-02-05 · Terminated
- Why it was recalled
- Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.
- Product
- JAPAN DISP SET, 261J, 205J, 208 - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma. The recovered blood is processed by a centrifugation and washing procedure before pumping the red cell product to either a bag for gravity reinfusion into the patient or an extracorporeal circuit.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.
- How much
- 25529
Japan CS Elite FASTPACK DISP. SET - 125ML - Product Usage: its related accessory components are intended for u
Z-0901-2020 · Class II · reported 2020-02-05 · Terminated
- Why it was recalled
- Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.
- Product
- Japan CS Elite FASTPACK DISP. SET - 125ML - Product Usage: its related accessory components are intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, and pumping this processed red cell product to either a bag for gravity reinfusion into the patient or to the arterial line of an extracorporeal circuit for reinfusion into the patient.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.
- How much
- 14278
CS Elite Processing Set (225 ml) Autotransfusion Device - Product Usage:its related accessory components are i
Z-0892-2020 · Class II · reported 2020-02-05 · Terminated
- Why it was recalled
- Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.
- Product
- CS Elite Processing Set (225 ml) Autotransfusion Device - Product Usage:its related accessory components are intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, and pumping this processed red cell product to either a bag for gravity reinfusion into the patient or to the arterial line of an extracorporeal circuit for reinfusion into the patient.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.
- How much
- 257345
CS Elite Fastpack, 225 mL, 20¿ Res Autotransfusion Device - Product Usage: its related accessory components ar
Z-0894-2020 · Class II · reported 2020-02-05 · Terminated
- Why it was recalled
- Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.
- Product
- CS Elite Fastpack, 225 mL, 20¿ Res Autotransfusion Device - Product Usage: its related accessory components are intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, and pumping this processed red cell product to either a bag for gravity reinfusion into the patient or to the arterial line of an extracorporeal circuit for reinfusion into the patient.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.
- How much
- 1092
CS5/5+ Bowl Kit-125 mL Autotransfusion Device - Product Usage: its related accessory components are intended t
Z-0886-2020 · Class II · reported 2020-02-05 · Terminated
- Why it was recalled
- Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.
- Product
- CS5/5+ Bowl Kit-125 mL Autotransfusion Device - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.
- How much
- 333443
CS5/5+ Fastpacks, 225 mL, 150¿ Res Autotransfusion Device - Product Usage: its related accessory components ar
Z-0885-2020 · Class II · reported 2020-02-05 · Terminated
- Why it was recalled
- Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.
- Product
- CS5/5+ Fastpacks, 225 mL, 150¿ Res Autotransfusion Device - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.
- How much
- 44821
JAPAN CS ELITE FASTPACK, 225ML - Product Usage: its related accessory components are intended for use to recov
Z-0902-2020 · Class II · reported 2020-02-05 · Terminated
- Why it was recalled
- Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.
- Product
- JAPAN CS ELITE FASTPACK, 225ML - Product Usage: its related accessory components are intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, and pumping this processed red cell product to either a bag for gravity reinfusion into the patient or to the arterial line of an extracorporeal circuit for reinfusion into the patient.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.
- How much
- 8114
Japan DISP SET, 263J, 205J, 208 - Product Usage: its related accessory components are intended to be used for
Z-0905-2020 · Class II · reported 2020-02-05 · Terminated
- Why it was recalled
- Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.
- Product
- Japan DISP SET, 263J, 205J, 208 - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.
- How much
- 14271
CS5/5+ Fastpacks, 125 mL, 150¿ Res Autotransfusion Device - Product Usage: its related accessory components ar
Z-0888-2020 · Class II · reported 2020-02-05 · Terminated
- Why it was recalled
- Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.
- Product
- CS5/5+ Fastpacks, 125 mL, 150¿ Res Autotransfusion Device - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.
- How much
- 14095
CS Elite Processing Set (125 ml) Autotransfusion Device - Product Usage:its related accessory components are i
Z-0893-2020 · Class II · reported 2020-02-05 · Terminated
- Why it was recalled
- Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.
- Product
- CS Elite Processing Set (125 ml) Autotransfusion Device - Product Usage:its related accessory components are intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, and pumping this processed red cell product to either a bag for gravity reinfusion into the patient or to the arterial line of an extracorporeal circuit for reinfusion into the patient.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.
- How much
- 176742
High Speed Cell Saver¿5/5+ Bowl Set Autotransfusion Device - Product Usage: its related accessory components a
Z-0887-2020 · Class II · reported 2020-02-05 · Terminated
- Why it was recalled
- Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.
- Product
- High Speed Cell Saver¿5/5+ Bowl Set Autotransfusion Device - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.
- How much
- 3563
JAPAN CELL SAVER ELITE SET - 125ML - Product Usage: its related accessory components are intended for use to r
Z-0903-2020 · Class II · reported 2020-02-05 · Terminated
- Why it was recalled
- Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.
- Product
- JAPAN CELL SAVER ELITE SET - 125ML - Product Usage: its related accessory components are intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, and pumping this processed red cell product to either a bag for gravity reinfusion into the patient or to the arterial line of an extracorporeal circuit for reinfusion into the patient.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.
- How much
- 21113
JAPAN BASIC HI SPEED CS5 SET,LN 263J - Product Usage: its related accessory components are intended to be used
Z-0898-2020 · Class II · reported 2020-02-05 · Terminated
- Why it was recalled
- Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.
- Product
- JAPAN BASIC HI SPEED CS5 SET,LN 263J - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma. The recovered blood is processed by a centrifugation and washing procedure before pumping the red cell product to either a bag for gravity reinfusion into the patient or an extracorporeal circuit.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.
- How much
- 28462
JAPAN LOW VOL(125ML) CS5 SET,LN 261J - Product Usage: its related accessory components are intended to be used
Z-0897-2020 · Class II · reported 2020-02-05 · Terminated
- Why it was recalled
- Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.
- Product
- JAPAN LOW VOL(125ML) CS5 SET,LN 261J - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma. The recovered blood is processed by a centrifugation and washing procedure before pumping the red cell product to either a bag for gravity reinfusion into the patient or an extracorporeal circuit.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.
- How much
- 38699
Cell Saver Elite Processing Kit, 70 ml, Product ID: CSE-P-70 Indications for Use: The Haemonetics Cell Save
Z-0779-2019 · Class II · reported 2019-02-06 · Terminated
- Why it was recalled
- Haemonetics has received isolated reports that in some cases the 70ml centrifuge bowl may leak during cell salvage processing.
- Product
- Cell Saver Elite Processing Kit, 70 ml, Product ID: CSE-P-70 Indications for Use: The Haemonetics Cell Saver 5/5+ System and its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma.
- Where it went
- US Nationwide Distribution to states of: AL, AZ, CA, CO, CT, FL, GA, ID, IL, IN, LA, MA, MD, ME, MN, MO, NC, NE, NH, NJ, NY, OH, OR, PA, SC, TN, TX, VA, WA, and WI.
- How much
- 4576 units
Cell Saver 5/5+ Bowl Set, 70ml, Product ID: 0291E-00, 0291A-00 Indications for Use: The Haemonetics Cell Sa
Z-0778-2019 · Class II · reported 2019-02-06 · Terminated
- Why it was recalled
- Haemonetics has received isolated reports that in some cases the 70ml centrifuge bowl may leak during cell salvage processing.
- Product
- Cell Saver 5/5+ Bowl Set, 70ml, Product ID: 0291E-00, 0291A-00 Indications for Use: The Haemonetics Cell Saver 5/5+ System and its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma.
- Where it went
- US Nationwide Distribution to states of: AL, AZ, CA, CO, CT, FL, GA, ID, IL, IN, LA, MA, MD, ME, MN, MO, NC, NE, NH, NJ, NY, OH, OR, PA, SC, TN, TX, VA, WA, and WI.
- How much
- 11,208 units
Haemonetics OrthoPAT System-Integrated Set. Product Code: 1150H-00 OrthoPAT orthopedic perioperative auto
Z-2950-2018 · Class II · reported 2018-09-05 · Terminated
- Why it was recalled
- OrthoPAT¿ reservoirs are missing the cannula inside the reservoirs. The absence of this cannula prevents the emptying of blood from the reservoirs. The inability to empty the reservoir means that the collected blood cannot be processed.
- Product
- Haemonetics OrthoPAT System-Integrated Set. Product Code: 1150H-00 OrthoPAT orthopedic perioperative autotransfusion system indicated for use to salvage RBCs from blood lost intraoperatively and postoperatively during surgical procedures
- Where it went
- US Nationwide Distribution
- How much
- 432 units
Haemonetics¿ OrthoPAT¿ System-OPAT QuickConnect Reservoirs. Product Code: OPT-R-1000 The OrthoPAT orthoped
Z-2951-2018 · Class II · reported 2018-09-05 · Terminated
- Why it was recalled
- OrthoPAT¿ reservoirs are missing the cannula inside the reservoirs. The absence of this cannula prevents the emptying of blood from the reservoirs. The inability to empty the reservoir means that the collected blood cannot be processed.
- Product
- Haemonetics¿ OrthoPAT¿ System-OPAT QuickConnect Reservoirs. Product Code: OPT-R-1000 The OrthoPAT orthopedic perioperative autotransfusion system indicated for use to salvage RBCs from blood lost intraoperatively and postoperatively during surgical procedures
- Where it went
- US Nationwide Distribution
- How much
- 348 units
Haemonetics TEG Hemostasis System Level II Control, Catalog number: 8002. Level II Control is a biological
Z-1130-2015 · Class II · reported 2015-02-25 · Terminated
- Why it was recalled
- Specific lots of Level II Control Kits produce high Maximum Amplitude (MA) results, which are outside of the manufacturer's expected Quality Control range for MA parameter.
- Product
- Haemonetics TEG Hemostasis System Level II Control, Catalog number: 8002. Level II Control is a biological control for use with the Thrombelastograph (TEG) Hemostasis System.
- Where it went
- Worldwide Distribution -- US, Canada, Czech Republic, Denmark, Great Britain, and Sweden.
- How much
- 4,143 kits
Showing the 30 most recent of 31. The rest are searchable at the FDA enforcement report database.
Read these counts with three things in mind
A firm with more recalls usually makes more products. A contract packer runs
hundreds of lines. It will out-recall a company with one product whatever either
does. This is not a safety ranking.
Most recalls here are voluntary. A firm found its own fault and pulled the stock.
That is the system working. It lands in this database looking the same as a recall
the FDA had to push for.
Terminated means the FDA considers the recall finished. It is not a verdict.
It does not mean anyone was harmed.
Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.
← All device recalls