Getinge Usa Sales Inc

Getinge Usa Sales Inc: 48 FDA recall reports

Every medical device recall the FDA has published for Getinge Usa Sales Inc. The fault, and how the agency graded it.

48Reports
9Class I
46Ongoing
2025-02-19Most recent
9 Class I on this page

Class I means the FDA judged there was a reasonable probability of serious harm or death. Check the lot numbers against anything you hold.

What the classes mean

The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.

ClassWhat the FDA is saying
Class IThere is a reasonable probability that using this will cause serious harm or death.
Class IIIt may cause temporary or reversible harm. Serious harm is unlikely.
Class IIIIt breaks a rule but is unlikely to hurt anyone. Often a labelling fault.

Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.

By class

ClassReports
Class II39
Class I9

By year reported

YearReports
202211
20234
202432
20251
Checking whether it affects you

The lot or code number is what matters. Not the product name. A recall covers specific lots. Yours may not be one of them.

Check the code on your package against the report. The FDA keeps the current list at fda.gov/safety/recalls.

The reports

The text below is quoted from the FDA record word for word.

Class I

Vaporizer Sevoflurane, Quik-Fil. Model Number: 6682285. Device is a component of Flow-i C20, Flow-i C30, Flow

Z-1074-2025  ·  Class I  ·  reported 2025-02-19  ·  Ongoing

Why it was recalled
Sevoflurane used with the vaporizer may degrade to hydrogen fluoride, which could present a risk to patients or health care providers through inhalation and/or skin exposure. Potential harms include irritation of respiratory tract, which in worst case may lead to lung edema and/or severe hypocalcemia, blistering, superficial ulceration, and/or hypomagnesemia.
Product
Vaporizer Sevoflurane, Quik-Fil. Model Number: 6682285. Device is a component of Flow-i C20, Flow-i C30, Flow-i C40, Flow-c, Flow-e. For use during anesthesia.
Where it went
Domestic distribution to Kentucky and Connecticut.
How much
6 units

MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before

Z-0462-2025  ·  Class II  ·  reported 2024-11-20  ·  Ongoing

Why it was recalled
Error code "50037" occurs on the IR-Hand Control when a MEERA-table is controlled, and the table stops the movement. This error may occur sporadically when the MEERA-table is controlled via an IR-Hand Control and may result in procedural delay
Product
MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during and after surgical interventions, as well as for examination and treatment Item Number: 700001B0
Where it went
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bhutan, Brazil, Canada, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Ecuador, Egypt, France, Germany, Greece, India, Israel, Japan, Jordan, Korea, Republic Of Kuwait, "Macedonia, The Former Yugoslav Republic Of", Malaysia, Mauritius, Mexico, Montenegro, Netherlands, New Zealand, Panama, Peru, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Viet Nam.
How much
112 units

MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before

Z-0460-2025  ·  Class II  ·  reported 2024-11-20  ·  Ongoing

Why it was recalled
Error code "50037" occurs on the IR-Hand Control when a MEERA-table is controlled, and the table stops the movement. This error may occur sporadically when the MEERA-table is controlled via an IR-Hand Control and may result in procedural delay
Product
MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during and after surgical interventions, as well as for examination and treatment Item Number: 710001B0
Where it went
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bhutan, Brazil, Canada, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Ecuador, Egypt, France, Germany, Greece, India, Israel, Japan, Jordan, Korea, Republic Of Kuwait, "Macedonia, The Former Yugoslav Republic Of", Malaysia, Mauritius, Mexico, Montenegro, Netherlands, New Zealand, Panama, Peru, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Viet Nam.
How much
112 units

MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before

Z-0458-2025  ·  Class II  ·  reported 2024-11-20  ·  Ongoing

Why it was recalled
Error code "50037" occurs on the IR-Hand Control when a MEERA-table is controlled, and the table stops the movement. This error may occur sporadically when the MEERA-table is controlled via an IR-Hand Control and may result in procedural delay
Product
MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during and after surgical interventions, as well as for examination and treatment Item Number: 720001F0
Where it went
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bhutan, Brazil, Canada, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Ecuador, Egypt, France, Germany, Greece, India, Israel, Japan, Jordan, Korea, Republic Of Kuwait, "Macedonia, The Former Yugoslav Republic Of", Malaysia, Mauritius, Mexico, Montenegro, Netherlands, New Zealand, Panama, Peru, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Viet Nam.
How much
22 units

MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before

Z-0463-2025  ·  Class II  ·  reported 2024-11-20  ·  Ongoing

Why it was recalled
Error code "50037" occurs on the IR-Hand Control when a MEERA-table is controlled, and the table stops the movement. This error may occur sporadically when the MEERA-table is controlled via an IR-Hand Control and may result in procedural delay
Product
MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during and after surgical interventions, as well as for examination and treatment Item Number: 700001F0
Where it went
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bhutan, Brazil, Canada, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Ecuador, Egypt, France, Germany, Greece, India, Israel, Japan, Jordan, Korea, Republic Of Kuwait, "Macedonia, The Former Yugoslav Republic Of", Malaysia, Mauritius, Mexico, Montenegro, Netherlands, New Zealand, Panama, Peru, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Viet Nam.
How much
1 unit

MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before

Z-0456-2025  ·  Class II  ·  reported 2024-11-20  ·  Ongoing

Why it was recalled
Error code "50037" occurs on the IR-Hand Control when a MEERA-table is controlled, and the table stops the movement. This error may occur sporadically when the MEERA-table is controlled via an IR-Hand Control and may result in procedural delay
Product
MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during and after surgical interventions, as well as for examination and treatment, Item Number: 720001B0
Where it went
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bhutan, Brazil, Canada, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Ecuador, Egypt, France, Germany, Greece, India, Israel, Japan, Jordan, Korea, Republic Of Kuwait, "Macedonia, The Former Yugoslav Republic Of", Malaysia, Mauritius, Mexico, Montenegro, Netherlands, New Zealand, Panama, Peru, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Viet Nam.
How much
342 units

MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before

Z-0461-2025  ·  Class II  ·  reported 2024-11-20  ·  Ongoing

Why it was recalled
Error code "50037" occurs on the IR-Hand Control when a MEERA-table is controlled, and the table stops the movement. This error may occur sporadically when the MEERA-table is controlled via an IR-Hand Control and may result in procedural delay
Product
MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during and after surgical interventions, as well as for examination and treatment Item Number: 710001B2
Where it went
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bhutan, Brazil, Canada, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Ecuador, Egypt, France, Germany, Greece, India, Israel, Japan, Jordan, Korea, Republic Of Kuwait, "Macedonia, The Former Yugoslav Republic Of", Malaysia, Mauritius, Mexico, Montenegro, Netherlands, New Zealand, Panama, Peru, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Viet Nam.
How much
3 units

MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before

Z-0457-2025  ·  Class II  ·  reported 2024-11-20  ·  Ongoing

Why it was recalled
Error code "50037" occurs on the IR-Hand Control when a MEERA-table is controlled, and the table stops the movement. This error may occur sporadically when the MEERA-table is controlled via an IR-Hand Control and may result in procedural delay
Product
MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during and after surgical interventions, as well as for examination and treatment Item Number: 720001B2
Where it went
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bhutan, Brazil, Canada, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Ecuador, Egypt, France, Germany, Greece, India, Israel, Japan, Jordan, Korea, Republic Of Kuwait, "Macedonia, The Former Yugoslav Republic Of", Malaysia, Mauritius, Mexico, Montenegro, Netherlands, New Zealand, Panama, Peru, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Viet Nam.
How much
234 units

MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before

Z-0459-2025  ·  Class II  ·  reported 2024-11-20  ·  Ongoing

Why it was recalled
Error code "50037" occurs on the IR-Hand Control when a MEERA-table is controlled, and the table stops the movement. This error may occur sporadically when the MEERA-table is controlled via an IR-Hand Control and may result in procedural delay
Product
MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during and after surgical interventions, as well as for examination and treatment Item Number: 720001F2
Where it went
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bhutan, Brazil, Canada, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Ecuador, Egypt, France, Germany, Greece, India, Israel, Japan, Jordan, Korea, Republic Of Kuwait, "Macedonia, The Former Yugoslav Republic Of", Malaysia, Mauritius, Mexico, Montenegro, Netherlands, New Zealand, Panama, Peru, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Viet Nam.
How much
52 units

ALPHAMAXX Mobile Operating Table, Model Number/Part Number: 113322B4

Z-2271-2024  ·  Class II  ·  reported 2024-07-17  ·  Ongoing

Why it was recalled
The recall is prompted by a defective charger unit (0970.2873) for models 1133.22B4, 1133.22B5, and 1133.22F5 of the ALPHAMAXX mobile operating table. This issue poses a risk of inadequate charging and operational failure.
Product
ALPHAMAXX Mobile Operating Table, Model Number/Part Number: 113322B4
Where it went
Domestic: KY, NC, NY, VA. International: Australia, Bulgaria, Croatia, Czech Republic, Denmark, France, Greece, Hong Kong, Ireland, Israel, Japan, Jordan, Kuwait, Mexico, New Zealand, Norway, Poland, Romania, Saudi Arabia, South Africa, Spain, Switzerland, Thailand, United Arab Emirates, United Kingdom.
How much
73 units

ALPHAMAXX Mobile Operating Table, Model Number/Part Number: 113322F5

Z-2273-2024  ·  Class II  ·  reported 2024-07-17  ·  Ongoing

Why it was recalled
The recall is prompted by a defective charger unit (0970.2873) for models 1133.22B4, 1133.22B5, and 1133.22F5 of the ALPHAMAXX mobile operating table. This issue poses a risk of inadequate charging and operational failure.
Product
ALPHAMAXX Mobile Operating Table, Model Number/Part Number: 113322F5
Where it went
Domestic: KY, NC, NY, VA. International: Australia, Bulgaria, Croatia, Czech Republic, Denmark, France, Greece, Hong Kong, Ireland, Israel, Japan, Jordan, Kuwait, Mexico, New Zealand, Norway, Poland, Romania, Saudi Arabia, South Africa, Spain, Switzerland, Thailand, United Arab Emirates, United Kingdom.
How much
7 units

ALPHAMAXX Mobile Operating Table, Model Number/Part Number: 113322B5

Z-2272-2024  ·  Class II  ·  reported 2024-07-17  ·  Ongoing

Why it was recalled
The recall is prompted by a defective charger unit (0970.2873) for models 1133.22B4, 1133.22B5, and 1133.22F5 of the ALPHAMAXX mobile operating table. This issue poses a risk of inadequate charging and operational failure.
Product
ALPHAMAXX Mobile Operating Table, Model Number/Part Number: 113322B5
Where it went
Domestic: KY, NC, NY, VA. International: Australia, Bulgaria, Croatia, Czech Republic, Denmark, France, Greece, Hong Kong, Ireland, Israel, Japan, Jordan, Kuwait, Mexico, New Zealand, Norway, Poland, Romania, Saudi Arabia, South Africa, Spain, Switzerland, Thailand, United Arab Emirates, United Kingdom.
How much
171 Units
Class I

Vaporizer Sevoflurane Maquet Filling, Product code/REF number (6682282), Component of: Flow-i C20, Flow-i C30,

Z-1877-2024  ·  Class I  ·  reported 2024-06-12  ·  Ongoing

Why it was recalled
Sevoflurane used with the vaporizer may degrade to hydrogen fluoride, which could present a risk to patients or health care providers through inhalation and/or skin exposure. Potential harms include irritation of respiratory tract, which in worst case may lead to lung edema and/or severe hypocalcemia, blistering, superficial ulceration, and/or hypomagnesemia.
Product
Vaporizer Sevoflurane Maquet Filling, Product code/REF number (6682282), Component of: Flow-i C20, Flow-i C30,Flow-i C40, Flow-c, Flow-e.
Where it went
Domestic: CA, CO, IN, MO, OR, VA, WA. International: Aruba, Australia, Austria, Belgium, Bolivia, Bosnia-Herzegovina, Brazil, Bulgaria, Chile, China, Costa Rica, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Macau, Macedonia, Malaysia, Mexico, Moldavia, Morocco, Mozambique, Namibia, Netherlands, Nicaragua, Norway, Oman, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Vietnam Please note that a list of foreign consignees was not provided by the manufacturer and it is not available at this time.
How much
120 units (US)

Flow-c Anesthesia System, model 6887700

Z-0990-2024  ·  Class II  ·  reported 2024-02-14  ·  Ongoing

Why it was recalled
Potential cybersecurity vulnerability that could lead to a Denial of Service (DoS) attack, tampering, or remote code execution via remote login to the Flow Anesthesia System, any of which could lead to a hazardous situation.
Product
Flow-c Anesthesia System, model 6887700
Where it went
Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, DC, FL, GA, IN, MA, MI, MO, NY, OH, OR, PA, TX, VA, WA & WV.
How much
14 (US)

Flow-i C20 Anesthesia System, model 6677200

Z-0987-2024  ·  Class II  ·  reported 2024-02-14  ·  Ongoing

Why it was recalled
Potential cybersecurity vulnerability that could lead to a Denial of Service (DoS) attack, tampering, or remote code execution via remote login to the Flow Anesthesia System, any of which could lead to a hazardous situation.
Product
Flow-i C20 Anesthesia System, model 6677200
Where it went
Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, DC, FL, GA, IN, MA, MI, MO, NY, OH, OR, PA, TX, VA, WA & WV.
How much
128 (US)

Flow-e Anesthesia System, model 6887900

Z-0991-2024  ·  Class II  ·  reported 2024-02-14  ·  Ongoing

Why it was recalled
Potential cybersecurity vulnerability that could lead to a Denial of Service (DoS) attack, tampering, or remote code execution via remote login to the Flow Anesthesia System, any of which could lead to a hazardous situation.
Product
Flow-e Anesthesia System, model 6887900
Where it went
Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, DC, FL, GA, IN, MA, MI, MO, NY, OH, OR, PA, TX, VA, WA & WV.
How much
86 (US)

Flow-i C30 Anesthesia System, model 6677300

Z-0988-2024  ·  Class II  ·  reported 2024-02-14  ·  Ongoing

Why it was recalled
Potential cybersecurity vulnerability that could lead to a Denial of Service (DoS) attack, tampering, or remote code execution via remote login to the Flow Anesthesia System, any of which could lead to a hazardous situation.
Product
Flow-i C30 Anesthesia System, model 6677300
Where it went
Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, DC, FL, GA, IN, MA, MI, MO, NY, OH, OR, PA, TX, VA, WA & WV.
How much
4 (US)

Flow-i C40 Anesthesia System, model 6677400

Z-0989-2024  ·  Class II  ·  reported 2024-02-14  ·  Ongoing

Why it was recalled
Potential cybersecurity vulnerability that could lead to a Denial of Service (DoS) attack, tampering, or remote code execution via remote login to the Flow Anesthesia System, any of which could lead to a hazardous situation.
Product
Flow-i C40 Anesthesia System, model 6677400
Where it went
Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, DC, FL, GA, IN, MA, MI, MO, NY, OH, OR, PA, TX, VA, WA & WV.
How much
16 (US)

Maquet PowerLED/HLED and PowerLED300 OR Light Systems, Model Numbers and Catalog Numbers EPS MB1 - ARDINT05900

Z-0763-2024  ·  Class II  ·  reported 2024-01-24  ·  Ongoing

Why it was recalled
A potential for a light system to fall in the operating room.
Product
Maquet PowerLED/HLED and PowerLED300 OR Light Systems, Model Numbers and Catalog Numbers EPS MB1 - ARDINT059002A, EPS MB1 - ARDINT059002C, EPS MB1 TR - ARDINT059000A, EPS MB1 TR - ARDINT059000C, EPS MB3 - ARDINT059003A, EPS MB3 - ARDINT059003C, EPS MB3 TR - ARDINT059001A, EPS MB3 TR - ARDINT059001C, OPT MBAT - ARD568210010A, OPT MBAT - ARD568210010C, PWD30 DUAL - ARD267800310C, PWD30 DUAL - ARD267800330A, PWD30 DUAL - ARD267800330C, PWD30 DUAL - ARD267800340A, PWD30 DUAL - ARD267800340C, PWD30+DF - ARD268300170A, PWD30+DF - ARD568424110A, PWD30+DF - ARD568424110C, PWD30+DF - ARD568424114A, PWD30+DF - ARD568424114C, PWD30+DF - ARD568424511A, PWD30+DF - ARD568424511C, PWD30+DF - ARD568424512A, PWD30+DF - ARD568424512C, PWD30+DF - ARD568424513A, PWD30+DF - ARD568424513C, PWD30+DF - ARD568484114A, PWD30+DF - ARD568484114C, PWD30+DF - ARD569424110A, PWD30+DF - ARD569424110C, PWD30+DF - ARD569424114A, PWD30+DF - ARD569424114C, PWD30+DF - ARD569424511A, PWD30+DF - ARD569424511C, PWD30+DF - ARD569424512A, PWD30+DF - ARD569424512C, PWD30+DF - ARD569424513A, PWD30+DF - ARD569424513C, PWD30+DF - ARD569484114A, PWD30+DF - ARD569484114C, PWD30+DF - ARDPWD208008A, PWD30+DF - ARDPWD208008C,...
Where it went
US Nationwide. Global Distribution.
How much
236793 units in total

Maquet Blueline Series 30/80 OR Light Systems, Model Numbers and Catalog Numbers BLUE 30 - ARD569015111A, BLUE

Z-0770-2024  ·  Class II  ·  reported 2024-01-24  ·  Ongoing

Why it was recalled
A potential for a light system to fall in the operating room.
Product
Maquet Blueline Series 30/80 OR Light Systems, Model Numbers and Catalog Numbers BLUE 30 - ARD569015111A, BLUE 30 - ARD569015111C, BLUE 30 - ARD569015131C, BLUE 30 - ARD569025111C, BLUE 30 - ARD569025131C, BLUE 30 - ARD569035111C, BLUE 30 - ARD569035131C, BLUE 30 - ARD569045111C, BLUE 30 - ARD569045131C, BLUE 30 - ARD569055111C, BLUE 30 - ARD569055131C, BLUE 30 - ARD569065112C, BLUE 30 - ARD569065132C, BLUE 30 C - ARDBLU209000C, BLUE 3030 - ARD569015411C, BLUE 3030 - ARD569015431C, BLUE 80 - ARD569014111A, BLUE 80 - ARD569014111C, BLUE 80 - ARD569014131C, BLUE 80 - ARD569064112C, BLUE 80 - ARD569064132C, BLUE 8030 - ARD569014411C, BLUE 8030 - ARD569014431C, BLUE 8080 - ARD569014211C, BLUE 8080 - ARD569014231C, BLUE 30 - ARD569065113C, BLUE 30 - ARD569065123C, BLUE 30 - ARD569075113A, BLUE 30 - ARD569075113C, BLUE 30 - ARD569075123C, BLUE 80 - ARD569064113C, BLUE 80 - ARD569064123C, BLUE 80 - ARD569074113A, BLUE 80 - ARD569074113C, BLUE 80 - ARD569074123C, HLX BLUE 3 - ARD569132113A
Where it went
US Nationwide. Global Distribution.
How much
236793 units in total

Maquet Rolite OR Light Systems, Model Numbers and Catalog Numbers ROL PWDII50SF - ARDPWT309001A, ROL PWDII50SF

Z-0760-2024  ·  Class II  ·  reported 2024-01-24  ·  Ongoing

Why it was recalled
A potential for a light system to fall in the operating room.
Product
Maquet Rolite OR Light Systems, Model Numbers and Catalog Numbers ROL PWDII50SF - ARDPWT309001A, ROL PWDII50SF - ARDPWT309002A, ROL PWDII50SF - ARDPWT309009A, ROL PWDII50SF - ARDPWT309010A, ROL PWDII70SF - ARDPWT309007A, ROL PWDII70SF - ARDPWT309008A, ROL PWDII70SF - ARDPWT309015A, ROL PWDII70SF - ARDPWT309016A, ROL VCS40SF - ARDROL309031A, ROL VCS60SF - ARDROL309032A, ROL VST40SF - ARDROL309033A, ROL VST40SF - ARDROL309034A, ROL VST40SF B - ARDROL309037A, ROL VST60SF - ARDROL309035A, ROL VST60SF - ARDROL309036A, ROL VST60SF B - ARDROL309038A, ROL VCSII40SF - ARDROL309039A, ROL VCSII40SF - ARDROL309040A, ROL VCSII60SF - ARDROL309041A, ROL VCSII60SF - ARDROL309042A, ROL PWDII50SF - ARDROL309043A, ROL PWDII50SF - ARDROL309044A, ROL PWDII70SF - ARDROL309045A, ROL PWDII70SF - ARDROL309046A, ROL PWDII50SF - ARDROL309047A, ROL PWDII50SF - ARDROL309048A, ROL PWDII70SF - ARDROL309049A, ROL PWDII70SF - ARDROL309050A, ROL VSTII40SF - ARDROL309051A, ROL VSTII60SF - ARDROL309052A, ROL VSTII40SF - ARDROL309053A, ROL VSTII60SF - ARDROL309054A
Where it went
US Nationwide. Global Distribution.
How much
236793 units in total

Maquet Lucea - Lucea10/40, Lucea50/100 OR Light Systems, Model Numbers and Catalog Numbers LCA 10 - ARD2LCA000

Z-0759-2024  ·  Class II  ·  reported 2024-01-24  ·  Ongoing

Why it was recalled
A potential for a light system to fall in the operating room.
Product
Maquet Lucea - Lucea10/40, Lucea50/100 OR Light Systems, Model Numbers and Catalog Numbers LCA 10 - ARD2LCA00040C, LCA 10 - ARDLCA109000C, LCA 10 - ARDLCA109002C, LCA 10 - ARDLCA109006C, LCA 10 - ARDLCA209008A, LCA 10 - ARDLCA209008C, LCA 10 - ARDLCA209009A, LCA 10 - ARDLCA209009C, LCA 10 - ARDLCA209010A, LCA 10 - ARDLCA209010C, LCA 10 - ARDLCA409000C, LCA 10 - ARDLCA409001C, LCA 10 - ARDLCA409002A, LCA 10 - ARDLCA409002C, LCA 10 - ARDLCA409003C, LCA 10 - ARDLCA409004A, LCA 10 - ARDLCA409004C, LCA 10 - ARDLCA409005C, LCA 10 DES - ARDLCA409001A, LCA 10 MOBILE - ARDLCA309008A, LCA 10 RAIL - ARDLCA409005A, LCA 10 WALL - ARDLCA109006A, LCA 40 - ARDLCA109001C, LCA 40 - ARDLCA208000A, LCA 40 - ARDLCA208000C, LCA 40 - ARDLCA208001A, LCA 40 - ARDLCA208001C, LCA 40 - ARDLCA209000A, LCA 40 - ARDLCA209000C, LCA 40 - ARDLCA209001A, LCA 40 - ARDLCA209001C, LCA 40 - ARDLCA209002A, LCA 40 - ARDLCA209002C, LCA 40 - ARDLCA209003A, LCA 40 - ARDLCA209003C, LCA 40 - ARDLCA209014A, LCA 40 - ARDLCA209014C, LCA 40 - ARDLCA209017A, LCA 40 - ARDLCA209017C, LCA 40 - ARDLCA209018A, LCA 40 - ARDLCA209018C, LCA 40 - ARDLCA209019A, LCA 40 - ARDLCA209019C, LCA 40 - ARDLCA209020A, LCA 40 - ARDLCA209020C, LCA 40...
Where it went
US Nationwide. Global Distribution.
How much
236793 units in total

Maquet PowerLEDII OR Light System, Model Names and Catalog Numbers PWDII77SF - ARDPWT629100A, PWDII555DF - ARD

Z-0757-2024  ·  Class II  ·  reported 2024-01-24  ·  Ongoing

Why it was recalled
A potential for a light system to fall in the operating room.
Product
Maquet PowerLEDII OR Light System, Model Names and Catalog Numbers PWDII77SF - ARDPWT629100A, PWDII555DF - ARDPWT629101A, PWDII555SF - ARDPWT629102A, PWDII755DF - ARDPWT629103A, PWDII755SF - ARDPWT629104A, PWDII775DF - ARDPWT629105A, PWDII775SF - ARDPWT629106A, PWDII50DF - ARDPWT629109A, PWDII50SF - ARDPWT629110A, PWDII50SF - ARDPWT629111A, PWDII55DF - ARDPWT629112A, PWDII55SF - ARDPWT629113A, PWDII70DF - ARDPWT629114A, PWDII70DF - ARDPWT629115A, PWDII70SF - ARDPWT629116A, PWDII70SF - ARDPWT629117A, PWDII75DF - ARDPWT629118A, PWDII75SF - ARDPWT629119A, PWDII50DF - ARDPWT629120A, PWDII50SF - ARDPWT629121A, PWDII50SF - ARDPWT629122A, PWDII55DF - ARDPWT629123A, PWDII55SF - ARDPWT629124A, PWDII70DF - ARDPWT629125A, PWDII70DF - ARDPWT629126A, PWDII70SF - ARDPWT629127A, PWDII70SF - ARDPWT629128A, PWDII75DF - ARDPWT629129A, PWDII75SF - ARDPWT629130A, PWDII50DF - ARDPWT629131A, PWDII50SF - ARDPWT629132A, PWDII50SF - ARDPWT629133A, PWDII55DF - ARDPWT629134A, PWDII55SF - ARDPWT629135A, PWDII70DF - ARDPWT629136A, PWDII70DF - ARDPWT629137A, PWDII70SF - ARDPWT629138A, PWDII70SF - ARDPWT629139A, PWDII75DF - ARDPWT629140A, PWDII75SF - ARDPWT629141A, PWDII50DF - ARDPWT629142A, PWDII50SF -...
Where it went
US Nationwide. Global Distribution.
How much
236793 units in total

Maquet Equipment OR Light Systems, Model Numbers and Catalog Numbers BRAINLAB CAMERA - ARDEQT269000A, BRAINLAB

Z-0761-2024  ·  Class II  ·  reported 2024-01-24  ·  Ongoing

Why it was recalled
A potential for a light system to fall in the operating room.
Product
Maquet Equipment OR Light Systems, Model Numbers and Catalog Numbers BRAINLAB CAMERA - ARDEQT269000A, BRAINLAB CAMERA - ARDEQT269000C, BRAINLAB CAMERA - ARDEQT269001A, BRAINLAB CAMERA - ARDEQT269001C, BRAINLAB CAMERA - ARDEQT269004A, BRAINLAB CAMERA - ARDEQT269004C, BRAINLAB CAMERA - ARDEQT269005A, BRAINLAB CAMERA - ARDEQT269005C, EQTBRB SAT12 - ARDSAT269000A, EQTBRB SAT13 - ARDSAT269001A, EQTFHS010 11 - ARDEQT239014A, EQTFHS010 SAT12 - ARDSAT239005A, EQTFHS010 SAT13 - ARDSAT239006A, EQTFHS010 SAT16 - ARDSAT239023A, EQTMHD231 - ARDEQT239013A, EQTMHD231 11 - ARDEQT239016A, EQTMHD231 16 - ARDEQT239023A, EQTMHD231 SAT12 - ARDSAT239013A, EQTMHD231 SAT13 - ARDSAT239014A, EQTMHS013 08 - ARD2EQT00010A, EQTMHS013 11 - ARDEQT239007A, EQTMHS013 14 - ARDEQT239008A, EQTMHS013 16 - ARDEQT239009A, EQTMHS013 SAT12 - ARDSAT239007A, EQTMHS013 SAT13 - ARDSAT239008A, EQTMHS021 11 - ARDEQT239010A, EQTMHS021 14 - ARDEQT239011A, EQTMHS021 16 - ARDEQT239012A, EQTMHS021 SAT12 - ARDSAT239009A, EQTMHS021 SAT13 - ARDSAT239010A, EQTSC07 SAT12 - ARD515031140A, EQTSC07 SAT13 - ARD515032140A, EQTSC07 SAT16 - ARD2SAT00120A, EQTSC07 SAT16 - ARDSAT229007A, EQTSHD 11 - ARDEQT229001A, EQTSHD SAT - ARD515031990A,...
Where it went
US Nationwide. Global Distribution.
How much
236793 units in total

Maquet Volista, Volista Access, Volista Access II, Volista Standop OR Light System, Model Numbers and Catalog

Z-0758-2024  ·  Class II  ·  reported 2024-01-24  ·  Ongoing

Why it was recalled
A potential for a light system to fall in the operating room.
Product
Maquet Volista, Volista Access, Volista Access II, Volista Standop OR Light System, Model Numbers and Catalog Numbers ROL B FOR VLT400SF STP - ARDROL309001A, ROL B FOR VLT400SF STP - ARDROL309001C, ROL B FOR VLT600SF STP - ARDROL309003A, ROL B FOR VLT600SF STP - ARDROL309003C, ROL FOR VLT400SF STP - ARDROL309000A, ROL FOR VLT400SF STP - ARDROL309000C, ROL FOR VLT600SF STP - ARDROL309002A, ROL FOR VLT600SF STP - ARDROL309002C, VCS40 ROL - ARDROL309024A, VCS40 ROL - ARDROL309024C, VCS40 ROL - ARDROL309028A, VCS40 ROL - ARDROL309028C, VCS60 ROL - ARDROL309025A, VCS60 ROL - ARDROL309025C, VCS60 ROL - ARDROL309029A, VCS60 ROL - ARDROL309029C, VST40 - ARDROL309016A, VST40 - ARDROL309016C, VST40 - ARDROL309017A, VST40 - ARDROL309017C, VST40 - ARDROL309020A, VST40 - ARDROL309020C, VST40 - ARDROL309022A, VST40 - ARDROL309022C, VST60 - ARDROL309018A, VST60 - ARDROL309018C, VST60 - ARDROL309019A, VST60 - ARDROL309019C, VST60 - ARDROL309021A, VST60 - ARDROL309021C, VST60 - ARDROL309023A, VST60 - ARDROL309023C, VCS40DF - ARDVCS209001A, VCS40DF - ARDVCS209001C, VCS40DF - ARDVCS209010A, VCS40DF - ARDVCS209012A, VCS40DF - ARDVCS209021A, VCS40SF - ARDVCS209000A, VCS40SF - ARDVCS209000C, VCS40SF -...
Where it went
US Nationwide. Global Distribution.
How much
236793 units in total

Maquet Hanaulux HLX3000 OR Light Systems, Model Numbers and Catalog Numbers H3 BC - HM567811111C

Z-0765-2024  ·  Class II  ·  reported 2024-01-24  ·  Ongoing

Why it was recalled
A potential for a light system to fall in the operating room.
Product
Maquet Hanaulux HLX3000 OR Light Systems, Model Numbers and Catalog Numbers H3 BC - HM567811111C
Where it went
US Nationwide. Global Distribution.
How much
236793 units in total

Maquet G8 / G8E OR Light Systems, Model Numbers and Catalog Numbers G8 LC E - ARD568131211C, G8G8 DF E - ARD56

Z-0769-2024  ·  Class II  ·  reported 2024-01-24  ·  Ongoing

Why it was recalled
A potential for a light system to fall in the operating room.
Product
Maquet G8 / G8E OR Light Systems, Model Numbers and Catalog Numbers G8 LC E - ARD568131211C, G8G8 DF E - ARD568121111C, G8G8FS DF - ARD568121131C
Where it went
US Nationwide. Global Distribution.
How much
236793 units in total

Maquet Axcel / Axcel + OR Light Systems, Model Numbers and Catalog Numbers AXL 5001 - ARD567501961C, AXL 5001S

Z-0764-2024  ·  Class II  ·  reported 2024-01-24  ·  Ongoing

Why it was recalled
A potential for a light system to fall in the operating room.
Product
Maquet Axcel / Axcel + OR Light Systems, Model Numbers and Catalog Numbers AXL 5001 - ARD567501961C, AXL 5001S - ARD567505961C, AXL 5002 - ARD567501962C, AXL 5002S - ARD567505962C, AXL 5501 - ARD567501960A, AXL 5501 - ARD567501960C, AXL 5501S - ARD567505960C, AXL+5001 DF - ARD567502961C, AXL+5001 SF - ARD567502861C, AXL+5301 DF - ARD567502971C, AXL+5501 DF - ARD567502960C, AXL+5501 SF - ARD567502860C, AXL+5531 DF - ARD567502930C, AXL50 RLC - ARDRLC309000C, AXL50 RLC - ARDRLC309006C, AXL5003 - ARD567505968C, AXL5003 US - ARD567505969C
Where it went
US Nationwide. Global Distribution.
How much
236793 units in total

Maquet Hanaulux HLX2000 OR Light Systems, Model Numbers and Catalog Numbers 3001 - ARD567424001C, 3001 HMS07 -

Z-0768-2024  ·  Class II  ·  reported 2024-01-24  ·  Ongoing

Why it was recalled
A potential for a light system to fall in the operating room.
Product
Maquet Hanaulux HLX2000 OR Light Systems, Model Numbers and Catalog Numbers 3001 - ARD567424001C, 3001 HMS07 - ARD567722001C, 3001 HXS12 - ARD567703001C, 4001 S/DF - ARD567211211C, 4001 SAD/DF - ARD567221211C, 4001 SAD/DF - ARD567221241C, 4001 SAD/SF - ARD567221141C, 4001 SAI/DF - ARD567220111C, 4001K S/DF - ARD567211213C, 4003 - ARD567215111C, 4401 S/DF - ARD567212211C, 4401 SAD/DF - ARD567223211C, 4401 SAD/DF - ARD567223241C, 4401 SAD/SF - ARD567223111C, 4441 SAD/DF - ARD567225241C, 6001K ACS/T - ARD567234113C, 6001K SAL/T - ARD567226113C, 6101 ACS/SU - ARD567235001C, 6301 SAL/T - ARD567423111C, 6401 ACS/DF - ARD567235211C, 6401 ACS/DF - ARD567235231C, 6401 ACS/DF - ARD567235241C, 6401 SAL/DF - ARD567227211C, 6401 SAL/DF - ARD567227241C, 6401K ACS/D - ARD567235243C, 6401K SAL/SF - ARD567227113C, 6411 SAL/DF - ARD567228009C, 6441 ACS/DF - ARD567236241C, 8401 ACS/DF - ARD567231225C, 8401 ACS/DF - ARD567231241C, 8401K ACS/D - ARD567231243C, 8431 ACS/DF - ARD567431221C, 8431 ACS/DF - ARD567431241C, 8441 ACS/DF - ARD567232241C, PRISMALIX - ARD567761211C
Where it went
US Nationwide. Global Distribution.
How much
236793 units in total

Maquet Hanaulux 2006/ 2007, Blue 100, Blue 130/ 90, Blue Series 30/ 80, and Prismatic OR Light Systems. No Mo

Z-0771-2024  ·  Class II  ·  reported 2024-01-24  ·  Ongoing

Why it was recalled
A potential for a light system to fall in the operating room.
Product
Maquet Hanaulux 2006/ 2007, Blue 100, Blue 130/ 90, Blue Series 30/ 80, and Prismatic OR Light Systems. No Model Numbers.
Where it went
US Nationwide. Global Distribution.
How much
236793 units in total

Showing the 30 most recent of 48. The rest are searchable at the FDA enforcement report database.

Read these counts with three things in mind

A firm with more recalls usually makes more products. A contract packer runs hundreds of lines. It will out-recall a company with one product whatever either does. This is not a safety ranking.

Most recalls here are voluntary. A firm found its own fault and pulled the stock. That is the system working. It lands in this database looking the same as a recall the FDA had to push for.

Terminated means the FDA considers the recall finished. It is not a verdict. It does not mean anyone was harmed.

Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.

← All device recalls