GE OEC Medical Systems, Inc

GE OEC Medical Systems, Inc: 48 FDA recall reports

Every medical device recall the FDA has published for GE OEC Medical Systems, Inc. The fault, and how the agency graded it.

48Reports
6Class I
9Ongoing
2024-12-25Most recent
6 Class I on this page

Class I means the FDA judged there was a reasonable probability of serious harm or death. Check the lot numbers against anything you hold.

What the classes mean

The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.

ClassWhat the FDA is saying
Class IThere is a reasonable probability that using this will cause serious harm or death.
Class IIIt may cause temporary or reversible harm. Serious harm is unlikely.
Class IIIIt breaks a rule but is unlikely to hurt anyone. Often a labelling fault.

Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.

By class

ClassReports
Class II42
Class I6

By year reported

YearReports
201219
20131
20145
20153
20163
20176
20212
20235
20244
Checking whether it affects you

The lot or code number is what matters. Not the product name. A recall covers specific lots. Yours may not be one of them.

Check the code on your package against the report. The FDA keeps the current list at fda.gov/safety/recalls.

The reports

The text below is quoted from the FDA record word for word.

OEC 3D REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC 3D is a mobile fluoroscopic C-Arm

Z-0703-2025  ·  Class II  ·  reported 2024-12-25  ·  Ongoing

Why it was recalled
Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.
Product
OEC 3D REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC 3D is a mobile fluoroscopic C-Arm imaging system used to assist trained surgeons and other qualified physicians. The system is used to provide fluoroscopic Xray images and volumetric reconstructions during diagnostic, interventional, and surgical procedures.
Where it went
U.S.: AK, AZ, CA, CO, CT, DC, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. O.U.S.: Australia, Brazil, Canada, Chile, Costa Rica, India, Italy, Philippines, Poland, Portugal, Sweden, Switzerland, and United Kingdom
How much
84 systems

OEC Elite REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC Elite mobile C-arm is intended

Z-0704-2025  ·  Class II  ·  reported 2024-12-25  ·  Ongoing

Why it was recalled
Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.
Product
OEC Elite REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC Elite mobile C-arm is intended to provide fluoroscopic and digital spot images of the patient anatomy, interventional tools/devices, and contrast agents during diagnostic, interventional, and surgical procedures.
Where it went
U.S.: AK, AZ, CA, CO, CT, DC, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. O.U.S.: Australia, Brazil, Canada, Chile, Costa Rica, India, Italy, Philippines, Poland, Portugal, Sweden, Switzerland, and United Kingdom
How much
100 systems

OEC 9900 REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC 9900 Elite mobile fluoroscopy sy

Z-0705-2025  ·  Class II  ·  reported 2024-12-25  ·  Ongoing

Why it was recalled
Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.
Product
OEC 9900 REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC 9900 Elite mobile fluoroscopy system is designed to provide fluoroscopic and spot-film images of the patient during diagnostic surgical and interventional procedures. Examples of clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures.
Where it went
U.S.: AK, AZ, CA, CO, CT, DC, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. O.U.S.: Australia, Brazil, Canada, Chile, Costa Rica, India, Italy, Philippines, Poland, Portugal, Sweden, Switzerland, and United Kingdom
How much
78 systems

OEC 9800 REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The Series OEC 9800 is designed to provi

Z-0706-2025  ·  Class II  ·  reported 2024-12-25  ·  Ongoing

Why it was recalled
Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.
Product
OEC 9800 REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The Series OEC 9800 is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. Clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures.
Where it went
U.S.: AK, AZ, CA, CO, CT, DC, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. O.U.S.: Australia, Brazil, Canada, Chile, Costa Rica, India, Italy, Philippines, Poland, Portugal, Sweden, Switzerland, and United Kingdom
How much
25 systems

OEC Flexiview 8800 Systems with 9-inch Image Intensifier

Z-2275-2023  ·  Class II  ·  reported 2023-08-09  ·  Ongoing

Why it was recalled
Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which could lead to injuries such as fractures, hematoma, contusion, bruise, pain.
Product
OEC Flexiview 8800 Systems with 9-inch Image Intensifier
Where it went
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Algeria, Argentina, Australia, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Burkina Faso, Canada, Chile, Colombia, Czechia, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, France, Georgia, Germany, Hong Kong, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Kazakhstan, KOREA, Malaysia, Mauritius, Mexico, Mongolia, Morocco, Nepal, New Caledonia, New Zealand, Nigeria, Paraguay, Peru, Poland, Portugal, Romania, Russia, Singapore, Spain, Switzerland, Taiwan, Tanzania, Thailand, Trinidad and Tobago, United Arab Emirates, United Kingdom, Uruguay, Vietnam.
How much
2

Field Upgrade Kits: P/N 5451208 RIU, 9 INCH, TOSHIBA; P/N 5485304 SERVICE PART, IMAGE INTENSIFIER, 9 INCH, TOS

Z-2279-2023  ·  Class II  ·  reported 2023-08-09  ·  Ongoing

Why it was recalled
Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which could lead to injuries such as fractures, hematoma, contusion, bruise, pain.
Product
Field Upgrade Kits: P/N 5451208 RIU, 9 INCH, TOSHIBA; P/N 5485304 SERVICE PART, IMAGE INTENSIFIER, 9 INCH, TOSHIBA; P/N 5761613 SERVICE PART, IMAGE INTENSIFIER, 9 INCH - used with OEC Flexiview 8800, OEC 9800, and OEC 9900 Systems
Where it went
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Algeria, Argentina, Australia, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Burkina Faso, Canada, Chile, Colombia, Czechia, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, France, Georgia, Germany, Hong Kong, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Kazakhstan, KOREA, Malaysia, Mauritius, Mexico, Mongolia, Morocco, Nepal, New Caledonia, New Zealand, Nigeria, Paraguay, Peru, Poland, Portugal, Romania, Russia, Singapore, Spain, Switzerland, Taiwan, Tanzania, Thailand, Trinidad and Tobago, United Arab Emirates, United Kingdom, Uruguay, Vietnam.
How much
16

OEC 9900 Systems with 9-inch Image Intensifier

Z-2277-2023  ·  Class II  ·  reported 2023-08-09  ·  Ongoing

Why it was recalled
Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which could lead to injuries such as fractures, hematoma, contusion, bruise, pain.
Product
OEC 9900 Systems with 9-inch Image Intensifier
Where it went
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Algeria, Argentina, Australia, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Burkina Faso, Canada, Chile, Colombia, Czechia, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, France, Georgia, Germany, Hong Kong, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Kazakhstan, KOREA, Malaysia, Mauritius, Mexico, Mongolia, Morocco, Nepal, New Caledonia, New Zealand, Nigeria, Paraguay, Peru, Poland, Portugal, Romania, Russia, Singapore, Spain, Switzerland, Taiwan, Tanzania, Thailand, Trinidad and Tobago, United Arab Emirates, United Kingdom, Uruguay, Vietnam.
How much
908

OEC 9800 Systems with 9-inch Image Intensifier

Z-2276-2023  ·  Class II  ·  reported 2023-08-09  ·  Ongoing

Why it was recalled
Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which could lead to injuries such as fractures, hematoma, contusion, bruise, pain.
Product
OEC 9800 Systems with 9-inch Image Intensifier
Where it went
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Algeria, Argentina, Australia, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Burkina Faso, Canada, Chile, Colombia, Czechia, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, France, Georgia, Germany, Hong Kong, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Kazakhstan, KOREA, Malaysia, Mauritius, Mexico, Mongolia, Morocco, Nepal, New Caledonia, New Zealand, Nigeria, Paraguay, Peru, Poland, Portugal, Romania, Russia, Singapore, Spain, Switzerland, Taiwan, Tanzania, Thailand, Trinidad and Tobago, United Arab Emirates, United Kingdom, Uruguay, Vietnam.
How much
58

OEC Elite Systems with 9-inch Image Intensifier

Z-2278-2023  ·  Class II  ·  reported 2023-08-09  ·  Ongoing

Why it was recalled
Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which could lead to injuries such as fractures, hematoma, contusion, bruise, pain.
Product
OEC Elite Systems with 9-inch Image Intensifier
Where it went
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Algeria, Argentina, Australia, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Burkina Faso, Canada, Chile, Colombia, Czechia, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, France, Georgia, Germany, Hong Kong, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Kazakhstan, KOREA, Malaysia, Mauritius, Mexico, Mongolia, Morocco, Nepal, New Caledonia, New Zealand, Nigeria, Paraguay, Peru, Poland, Portugal, Romania, Russia, Singapore, Spain, Switzerland, Taiwan, Tanzania, Thailand, Trinidad and Tobago, United Arab Emirates, United Kingdom, Uruguay, Vietnam.
How much
812

OEC 3D Imaging, Image-Intensified Fluoroscopic X-Ray System, Mobile - Product Usage: designed to provide fluor

Z-2160-2021  ·  Class II  ·  reported 2021-08-04  ·  Terminated

Why it was recalled
There is a potential that the coin cell battery used to monitor X-Ray tube temperature may deplete early causing the system to become inoperable.
Product
OEC 3D Imaging, Image-Intensified Fluoroscopic X-Ray System, Mobile - Product Usage: designed to provide fluoroscopic and digital spot images of adult and pediatric populations during diagnostic, interventional, and surgical procedures
Where it went
Worldwide distribution - US Nationwide distribution including in the states of CA, PA, WA, OH, LA, MS, AL, FL, MO, OK, AZ, NJ, CO, MI, NV, MN, KS, TN, NC, IL, MD, SC,GA, UT, WI, MT, AR, WV, ID,CT, IA, WY, VT, KY, DC, SD,ME, Virgin Islands and the countries of Colombia, Egypt, Peru, Chile, Spain, France, Germany, Hong Kong, Canada, Japan, Brazil, Turkey, United Arab Emirates, Singapore, Australia, Thailand, United Kingdom, Italy, Korea, Russia, CHINA, Romania, Nigeria, Czechia, Argentina, Taiwan, Bangladeh, Malaysia, Bosnia and Herzegovina, Poland, Belgium, Indonesia, Dominican Republic, Switzerland, Hungary, New Zealand, Vietnam, Mexico, India, Norway, Philippines, Ecuador, Lithuania, HONDURAS, Uruguay, Sweden, Croatia, Jamaica, Kuwait, Ethiopia, Saudi Arabia, Finland, Bahamas, Trinidad and Tobago, Pakistan, Slovakia, Costa Rica, Israel, Panama, Greece, South Africa and DENMARK.
How much
3 units

OEC Elite mobile C-arm - Product Usage: intended to provide fluoroscopic and digital spot images of the patien

Z-2161-2021  ·  Class II  ·  reported 2021-08-04  ·  Terminated

Why it was recalled
There is a potential that the coin cell battery used to monitor X-Ray tube temperature may deplete early causing the system to become inoperable.
Product
OEC Elite mobile C-arm - Product Usage: intended to provide fluoroscopic and digital spot images of the patient anatomy, interventional tools/devices, and contrast agents during diagnostic, interventional, and surgical procedures.
Where it went
Worldwide distribution - US Nationwide distribution including in the states of CA, PA, WA, OH, LA, MS, AL, FL, MO, OK, AZ, NJ, CO, MI, NV, MN, KS, TN, NC, IL, MD, SC,GA, UT, WI, MT, AR, WV, ID,CT, IA, WY, VT, KY, DC, SD,ME, Virgin Islands and the countries of Colombia, Egypt, Peru, Chile, Spain, France, Germany, Hong Kong, Canada, Japan, Brazil, Turkey, United Arab Emirates, Singapore, Australia, Thailand, United Kingdom, Italy, Korea, Russia, CHINA, Romania, Nigeria, Czechia, Argentina, Taiwan, Bangladeh, Malaysia, Bosnia and Herzegovina, Poland, Belgium, Indonesia, Dominican Republic, Switzerland, Hungary, New Zealand, Vietnam, Mexico, India, Norway, Philippines, Ecuador, Lithuania, HONDURAS, Uruguay, Sweden, Croatia, Jamaica, Kuwait, Ethiopia, Saudi Arabia, Finland, Bahamas, Trinidad and Tobago, Pakistan, Slovakia, Costa Rica, Israel, Panama, Greece, South Africa and DENMARK.
How much
3570 units

OEC FlexiView 8800, Digital Mobile Imaging System, GE Medical Systems

Z-1469-2017  ·  Class II  ·  reported 2017-03-22  ·  Terminated

Why it was recalled
OEC FlexiView 8800 workstation power cable assembly issue which can result in a loss of imaging functionality due to intermittent power cord failures.
Product
OEC FlexiView 8800, Digital Mobile Imaging System, GE Medical Systems
Where it went
One consignee in CA.
How much
1

OEC 9900 Elite. MDL Numbers: D148942, D155043

Z-1060-2017  ·  Class II  ·  reported 2017-02-01  ·  Terminated

Why it was recalled
GE Healthcare Surgery announces a voluntary field action for the OEC 9800 and OEC 9900 Elite due to additional supplementation to the user documentation to include areas for enhanced planned maintenance and component replacement frequencies.
Product
OEC 9900 Elite. MDL Numbers: D148942, D155043
Where it went
Worldwide including govt/VA/military and foreign consignees.
How much
26,256 total

OEC 9800. MDL Numbers: D222250, D141598

Z-1059-2017  ·  Class II  ·  reported 2017-02-01  ·  Terminated

Why it was recalled
GE Healthcare Surgery announces a voluntary field action for the OEC 9800 and OEC 9900 Elite due to additional supplementation to the user documentation to include areas for enhanced planned maintenance and component replacement frequencies.
Product
OEC 9800. MDL Numbers: D222250, D141598
Where it went
Worldwide including govt/VA/military and foreign consignees.
How much
26,256 total

OEC 9800. MDL Numbers: D222250, D141598 The OEC 9800 is designed to provide fluoroscopic and spot-film imagin

Z-0973-2017  ·  Class II  ·  reported 2017-01-18  ·  Terminated

Why it was recalled
GE Healthcare Surgery announces a voluntary field action for the OEC MiniView 6800, OEC 9900 Elite, and OEC 9800 because the workstation power cable assembly can result in intermittent power cord failures.
Product
OEC 9800. MDL Numbers: D222250, D141598 The OEC 9800 is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. Clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures.
Where it went
Worldwide Distribution - US including U.S. States: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PUERTO RICO, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and Internationally to Canada, Mexico, and Costa Rica

OEC MiniView 6800. MDL Number: D002533 The OEC MiniView 6800 Digital Mobile C-Arm is designed to provide the

Z-0974-2017  ·  Class II  ·  reported 2017-01-18  ·  Terminated

Why it was recalled
GE Healthcare Surgery announces a voluntary field action for the OEC MiniView 6800, OEC 9900 Elite, and OEC 9800 because the workstation power cable assembly can result in intermittent power cord failures.
Product
OEC MiniView 6800. MDL Number: D002533 The OEC MiniView 6800 Digital Mobile C-Arm is designed to provide the physician with general fluoroscopic visualization of the patient including but not limited to surgical orthopedic and extremity imaging. The device is not intended for whole-body pediatric imaging
Where it went
Worldwide Distribution - US including U.S. States: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PUERTO RICO, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and Internationally to Canada, Mexico, and Costa Rica

The OEC 9900 Elite mobile fluoroscopy system is designed to provide fluoroscopic and spot-film images of the p

Z-0975-2017  ·  Class II  ·  reported 2017-01-18  ·  Terminated

Why it was recalled
GE Healthcare Surgery announces a voluntary field action for the OEC MiniView 6800, OEC 9900 Elite, and OEC 9800 because the workstation power cable assembly can result in intermittent power cord failures.
Product
The OEC 9900 Elite mobile fluoroscopy system is designed to provide fluoroscopic and spot-film images of the patient during diagnostic surgical and interventional procedure. Examples of clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedure. The system may be used for other imaging applications at the physician's discretionOEC 9900 Elite. MDL Numbers: D148942, D155043
Where it went
Worldwide Distribution - US including U.S. States: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PUERTO RICO, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and Internationally to Canada, Mexico, and Costa Rica

OEC Brivo 715 Prime, OEC Brivo 785 Essential, OEC Brivo 865 Advance Product Usage: The OEC Brivo Mobile C-

Z-0071-2017  ·  Class II  ·  reported 2016-10-19  ·  Terminated

Why it was recalled
GE Healthcare Surgery announces a voluntary field action for the OEC Brivo 715 Prime, OEC Brivo 685 Essential, and OEC Brivo 865 Advance because the circuit board that controls vertical motion of the C-arm may fail, resulting in the C-Arm moving up or down without command.
Product
OEC Brivo 715 Prime, OEC Brivo 785 Essential, OEC Brivo 865 Advance Product Usage: The OEC Brivo Mobile C-Arm X-Ray Products are designed to provide digital spot film imaging and fluoroscopic image guidance for all adult and pediatric populations for orientations between patient anatomy and surgical instruments. The product is used for general surgical applications and musculoskeletal procedures to visualize, for example, implant localizations or needle positions for aspirations, injections or biopsy. The OEC Brivo is not indicated for interventional use.
Where it went
Worldwide Distribution - US Nationwide in the states of AL, AZ, CA, FL, KS, MS, NJ, NV, NY, PA, TX, UT, and WV. No Candaian or Va/govt/military consignees. There are 133 other foreign consignees.
How much
148 total

OEC UroView 2800, fluoroscopic x-ray system

Z-2628-2016  ·  Class II  ·  reported 2016-08-31  ·  Terminated

Why it was recalled
A firmware issue in the supplied LCD workstation monitor(s). Under some circumstances, the monitor(s) may appear dark with no video displayed.
Product
OEC UroView 2800, fluoroscopic x-ray system
Where it went
Distribution US nationwide and Korea.
How much
47 total

OEC 9800, fluoroscopic x-ray system

Z-2629-2016  ·  Class II  ·  reported 2016-08-31  ·  Terminated

Why it was recalled
A firmware issue in the supplied LCD workstation monitor(s). Under some circumstances, the monitor(s) may appear dark with no video displayed.
Product
OEC 9800, fluoroscopic x-ray system
Where it went
Distribution US nationwide and Korea.
How much
47 total

GE OEC Brivo 865 plus, Mobile X-ray system laser aimer assemblies (part number 520002DW)

Z-1638-2015  ·  Class II  ·  reported 2015-06-17  ·  Terminated

Why it was recalled
Laser aimer assemblies were shipped without certification of the overall assembly or submitting a laser product report and therefore do not comply with the US regulations for laser products.
Product
GE OEC Brivo 865 plus, Mobile X-ray system laser aimer assemblies (part number 520002DW)
Where it went
US Distribution to the states of: MD, NC, MT, NJ, UT and NY
How much
9

GE OEC Brivo 865 Plus Mobile C-Arm X-Ray Product Used for general surgical applications and musculoskeletal

Z-0841-2015  ·  Class II  ·  reported 2015-01-21  ·  Terminated

Why it was recalled
mA Accuracy may exceed the design and labeling specification of+ 10%. Per 1020.32(f): Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as stated by the manufacturer in accordance with 1020.30(h)(3).
Product
GE OEC Brivo 865 Plus Mobile C-Arm X-Ray Product Used for general surgical applications and musculoskeletal procedures to visualize.
Where it went
Worldwide Distribution
How much
61 units installed in US

GE OEC Brivo 715 Prime Mobile C-Arm X-Ray Product Used for general surgical applications and musculoskeleta

Z-0840-2015  ·  Class II  ·  reported 2015-01-21  ·  Terminated

Why it was recalled
mA Accuracy may exceed the design and labeling specification of+ 10%. Per 1020.32(f): Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as stated by the manufacturer in accordance with 1020.30(h)(3).
Product
GE OEC Brivo 715 Prime Mobile C-Arm X-Ray Product Used for general surgical applications and musculoskeletal procedures to visualize.
Where it went
Worldwide Distribution
How much
61 units installed in US

FlexiView 8800 Digital Mobile Imaging System. The FlexiView 8800 Digital Mobile Imaging System is designed

Z-0194-2015  ·  Class II  ·  reported 2014-11-19  ·  Terminated

Why it was recalled
GE Healthcare initiated a field correction because prior revision of components were potentially installed on certain OEC 9800, FlexiView 8800, OEC MiniView 6800 and OEC UroView 2800 systems.
Product
FlexiView 8800 Digital Mobile Imaging System. The FlexiView 8800 Digital Mobile Imaging System is designed to provide fluoroscopic and spot film, imaging of the patient during diagnostic, surgical and interventional procedures.
Where it went
Worldwide Distribution.
How much
5

Series OEC 9800. The Series OEC 9800 is designed to provide fluoroscopic and spot-film imaging of the patie

Z-0195-2015  ·  Class II  ·  reported 2014-11-19  ·  Terminated

Why it was recalled
GE Healthcare initiated a field correction because prior revision of components were potentially installed on certain OEC 9800, FlexiView 8800, OEC MiniView 6800 and OEC UroView 2800 systems.
Product
Series OEC 9800. The Series OEC 9800 is designed to provide fluoroscopic and spot-film imaging of the patient,during diagnostic, surgical and interventional procedures.
Where it went
Worldwide Distribution.
How much
311

OEC¿ UroView 2800. The UroView 2800 is designed to provide fluoroscopic and spot-film imaging of the patien

Z-0192-2015  ·  Class II  ·  reported 2014-11-19  ·  Terminated

Why it was recalled
GE Healthcare initiated a field correction because prior revision of components were potentially installed on certain OEC 9800, FlexiView 8800, OEC MiniView 6800 and OEC UroView 2800 systems.
Product
OEC¿ UroView 2800. The UroView 2800 is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical. and interventional procedures
Where it went
Worldwide Distribution.
How much
2

OEC¿ MiniView 6800. The Mini 6800 Digital Mobile C-Arm is designed to provide the physician with general fl

Z-0193-2015  ·  Class II  ·  reported 2014-11-19  ·  Terminated

Why it was recalled
GE Healthcare initiated a field correction because prior revision of components were potentially installed on certain OEC 9800, FlexiView 8800, OEC MiniView 6800 and OEC UroView 2800 systems.
Product
OEC¿ MiniView 6800. The Mini 6800 Digital Mobile C-Arm is designed to provide the physician with general fluoroscopic visualization of the patient including but not limited to surgical orthopedic and extremity imaging. The device is not intended for whole-body pediatric imaging.
Where it went
Worldwide Distribution.
How much
26

The OEC 9900 Elite mobile fluoroscopy system (image intensified fluoroscopic x-ray) is designed to provide flu

Z-1462-2014  ·  Class II  ·  reported 2014-04-23  ·  Terminated

Why it was recalled
GE Healthcare has identified a potential safety issue related to the screws that attach the workstation monitor spring arm to the extension arm on certain OEC 9900 systems, causing the spring arm and monitor to fall forward.
Product
The OEC 9900 Elite mobile fluoroscopy system (image intensified fluoroscopic x-ray) is designed to provide fluoroscopic and spot-film images of the patient during diagnostic surgical and interventional procedure. Examples of clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedure. The system may be used for other imaging applications at the physician's discretion.
Where it went
Worldwide Distribution - USA, and the countries of Australia, Argentina, Canada, China, Cambodia, Columbia, Japan, Ecuador, Japan, Korea, Hong Kong, India, Israel, Libya, Mexico, Nigeria, Peru, Russia, Thailand, Taiwan, Saudi Arabia and Singapore.
How much
453 units

InstaTrak¿ 3500 Plus, ENTrakTM Plus, InstaTrak¿ 3500, ENTrakTM Computer-Assisted Navigation Systems (USA),

Z-0524-2014  ·  Class II  ·  reported 2013-12-25  ·  Terminated

Why it was recalled
GE OEC Medical Systems Inc is recalling InstaTrak¿ 3500 Plus, ENTrakTM Plus, InstaTrak¿ 3500, ENTrakTM Computer-Assisted Navigation Systems (USA) due to several potential safety issues related to the use of these products.
Product
InstaTrak¿ 3500 Plus, ENTrakTM Plus, InstaTrak¿ 3500, ENTrakTM Computer-Assisted Navigation Systems (USA),
Where it went
Nationwide Distribution
How much
654
Class I

OEC IT3000; IT2500 (EnTrak); IT2500 Plus (EnTrak Plus); IT3500 (InstaTrak); IT3500 Plus (InstaTrak Plus); Conn

Z-2251-2012  ·  Class I  ·  reported 2012-09-05  ·  Terminated

Why it was recalled
GE Healthcare Surgery had discovered that using the Inverted Headset Placement is not a validated configuration with the IT 3000, 2500, 2500 plus, 3500, 3500 plus, ConneCTstat, and ConneCTstat Plus Surgical Navigation equipment.
Product
OEC IT3000; IT2500 (EnTrak); IT2500 Plus (EnTrak Plus); IT3500 (InstaTrak); IT3500 Plus (InstaTrak Plus); ConneCTstat; and ConneCTstat Plus picture archiving and communications systems. The systems are an aid to locate anatomical structures during open or percutaneous surgical procedures. It is intended for use in medical conditions that may benefit from the use of stereotactic surgical technique. The system provides a reference to rigid anatomical structures such as sinus, skull, long bone, or vertebra, which are visible on medical images such as CT, MRI, or X-ray.
Where it went
Worldwide Distribution - USA (nationwide) and the countries of India, Japan, Canada, Switzerland, Italy, United Kingdom, Germany, Austria, Hong Kong, France, and Denmark.
How much
768 units

Showing the 30 most recent of 48. The rest are searchable at the FDA enforcement report database.

Read these counts with three things in mind

A firm with more recalls usually makes more products. A contract packer runs hundreds of lines. It will out-recall a company with one product whatever either does. This is not a safety ranking.

Most recalls here are voluntary. A firm found its own fault and pulled the stock. That is the system working. It lands in this database looking the same as a recall the FDA had to push for.

Terminated means the FDA considers the recall finished. It is not a verdict. It does not mean anyone was harmed.