GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL

GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING: 41 FDA recall reports

Every medical device recall the FDA has published for GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING. The fault, and how the agency graded it.

41Reports
6Class I
41Ongoing
2026-01-28Most recent
6 Class I on this page

Class I means the FDA judged there was a reasonable probability of serious harm or death. Check the lot numbers against anything you hold.

What the classes mean

The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.

ClassWhat the FDA is saying
Class IThere is a reasonable probability that using this will cause serious harm or death.
Class IIIt may cause temporary or reversible harm. Serious harm is unlikely.
Class IIIIt breaks a rule but is unlikely to hurt anyone. Often a labelling fault.

Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.

By class

ClassReports
Class II35
Class I6

By year reported

YearReports
20236
20241
202529
20265
Checking whether it affects you

The lot or code number is what matters. Not the product name. A recall covers specific lots. Yours may not be one of them.

Check the code on your package against the report. The FDA keeps the current list at fda.gov/safety/recalls.

The reports

The text below is quoted from the FDA record word for word.

GE Healthcare Omni Legend, sold under the following names: OMNI Legend 16, Omni Legend 21, OMNI Legend 32,

Z-1112-2026  ·  Class II  ·  reported 2026-01-28  ·  Ongoing

Why it was recalled
There is a potential intermittent issue on certain Omni Legend systems that can result in a streaking artifact in the PET clinical scan images. This streaking artifact is most easily identified in transaxial slices on the acquisition console (both corrected and non-corrected for attenuation).
Product
GE Healthcare Omni Legend, sold under the following names: OMNI Legend 16, Omni Legend 21, OMNI Legend 32, Omni Legend PET Gantry 16cm - MEA, Omni Legend PET Gantry 16cm Haifa LM, Omni Legend PET Gantry 16cm Mobile, Omni Legend PET Gantry 21cm - BJ for Global, Omni Legend PET Gantry 21cm - Global Haifa LM, Omni Legend PET Gantry 21cm - Mobile, Omni Legend PET Gantry 21cm WSO LM, Omni Legend PET Gantry 32cm - MEA, Omni Legend PET Gantry 32cm Haifa LM, Omni Legend PET Gantry 32cm Mobile, OMNI PET Gantry 15cm, Omni PET Gantry 15cm - WSO, OMNI PET Gantry 15cm Cardiac- WSO, OMNI PET Gantry 15cm ES - WSO, Omni PET Gantry 15cm Mobile WSO, Omni PET Gantry 21cm - WSO, Omni PET Gantry 21cm Mobile - WSO, OMNI PET Gantry 30cm, Omni PET Gantry 30cm - WSO; System, Tomography, Computed, Emission
Where it went
Worldwide distribution. US Nationwide, Argentina, Australia, Belgium, Brazil, Canada, Chile, Colombia, El Salvador, Estonia, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Israel, Italy, Japan, Korea (Republic Of), Lithuania, Luxembourg, Mexico, Netherlands, New Zealand, Poland, Romania, Russia, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkiye, United Kingdom, and Uzbekistan.
How much
364 units

SMV DSI, System, Tomography, Computed, Emission

Z-1045-2026  ·  Class II  ·  reported 2026-01-21  ·  Ongoing

Why it was recalled
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.
Product
SMV DSI, System, Tomography, Computed, Emission
Where it went
US, Brazil, Egypt, Canada, Mexico, Turkey, Italy, Sweden, Switzerland, France, Spain Chile, Germany, Iran, Greece
How much
3 units

SMV BodyTrack, System, Tomography, Computed, Emission

Z-1046-2026  ·  Class II  ·  reported 2026-01-21  ·  Ongoing

Why it was recalled
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.
Product
SMV BodyTrack, System, Tomography, Computed, Emission
Where it went
US, Brazil, Egypt, Canada, Mexico, Turkey, Italy, Sweden, Switzerland, France, Spain Chile, Germany, Iran, Greece
How much
12 units

SMV DSX New Line, System, Tomography, Computed, Emission

Z-1044-2026  ·  Class II  ·  reported 2026-01-21  ·  Ongoing

Why it was recalled
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.
Product
SMV DSX New Line, System, Tomography, Computed, Emission
Where it went
US, Brazil, Egypt, Canada, Mexico, Turkey, Italy, Sweden, Switzerland, France, Spain Chile, Germany, Iran, Greece
How much
7 units

MAXXUS, System, Tomography, Computed, Emission

Z-1047-2026  ·  Class II  ·  reported 2026-01-21  ·  Ongoing

Why it was recalled
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.
Product
MAXXUS, System, Tomography, Computed, Emission
Where it went
US, Brazil, Egypt, Canada, Mexico, Turkey, Italy, Sweden, Switzerland, France, Spain Chile, Germany, Iran, Greece
How much
38 units

Elscint SPX6; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging

Z-0293-2026  ·  Class II  ·  reported 2025-10-29  ·  Ongoing

Why it was recalled
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
Product
Elscint SPX6; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Where it went
Worldwide
How much
28 units

Millenium MPR; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging

Z-0299-2026  ·  Class II  ·  reported 2025-10-29  ·  Ongoing

Why it was recalled
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
Product
Millenium MPR; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Where it went
Worldwide
How much
292 units

Elscint Model 4XX; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging

Z-0291-2026  ·  Class II  ·  reported 2025-10-29  ·  Ongoing

Why it was recalled
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
Product
Elscint Model 4XX; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Where it went
Worldwide
How much
64 units

Millenium MG/MC; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging

Z-0297-2026  ·  Class II  ·  reported 2025-10-29  ·  Ongoing

Why it was recalled
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
Product
Millenium MG/MC; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Where it went
Worldwide
How much
1386 units

Millenium Myosight; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging

Z-0298-2026  ·  Class II  ·  reported 2025-10-29  ·  Ongoing

Why it was recalled
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
Product
Millenium Myosight; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Where it went
Worldwide
How much
82 units

Starcam/Camstar; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging

Z-0303-2026  ·  Class II  ·  reported 2025-10-29  ·  Ongoing

Why it was recalled
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
Product
Starcam/Camstar; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Where it went
Worldwide
How much
102 units

Elscint Helix; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging

Z-0295-2026  ·  Class II  ·  reported 2025-10-29  ·  Ongoing

Why it was recalled
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
Product
Elscint Helix; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Where it went
Worldwide
How much
3 units

Sopha Medical Vision (SMV) DSXi/DST-XLi,/ DST-XL/ DSTi; dual-head nuclear medicine gamma camera designed speci

Z-0302-2026  ·  Class II  ·  reported 2025-10-29  ·  Ongoing

Why it was recalled
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
Product
Sopha Medical Vision (SMV) DSXi/DST-XLi,/ DST-XL/ DSTi; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Where it went
Worldwide
How much
205 units

Elscint SPX4; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging

Z-0292-2026  ·  Class II  ·  reported 2025-10-29  ·  Ongoing

Why it was recalled
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
Product
Elscint SPX4; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Where it went
Worldwide
How much
45 units

Elscint Varicam; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging

Z-0294-2026  ·  Class II  ·  reported 2025-10-29  ·  Ongoing

Why it was recalled
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
Product
Elscint Varicam; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Where it went
Worldwide
How much
15 units

Millenium MPS; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging

Z-0300-2026  ·  Class II  ·  reported 2025-10-29  ·  Ongoing

Why it was recalled
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
Product
Millenium MPS; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Where it went
Worldwide
How much
54 units

Elscint Cardial; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging

Z-0290-2026  ·  Class II  ·  reported 2025-10-29  ·  Ongoing

Why it was recalled
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
Product
Elscint Cardial; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Where it went
Worldwide
How much
11 units

Maxicam; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging

Z-0296-2026  ·  Class II  ·  reported 2025-10-29  ·  Ongoing

Why it was recalled
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
Product
Maxicam; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Where it went
Worldwide
How much
37 units

Optima NX; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging

Z-0301-2026  ·  Class II  ·  reported 2025-10-29  ·  Ongoing

Why it was recalled
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
Product
Optima NX; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Where it went
Worldwide
How much
1 unit

Infinia, Infinia Hawkeye, Infinia II, Infinia II Hawkeye 4 Nuclear Medicine system

Z-2097-2025  ·  Class II  ·  reported 2025-07-16  ·  Ongoing

Why it was recalled
Unintended radial detector motion may occur during patient setup or during patient scan if system does not have correct version of gantry software installed. Unintended detector motion may result in life-threatening injury.
Product
Infinia, Infinia Hawkeye, Infinia II, Infinia II Hawkeye 4 Nuclear Medicine system
Where it went
Worldwide - US Nationwide distribution.
How much
1822

Varicam, Millennium VG, Millennium VG Hawkeye, Discovery VH Nuclear Medicine System

Z-2098-2025  ·  Class II  ·  reported 2025-07-16  ·  Ongoing

Why it was recalled
Unintended radial detector motion may occur during patient setup or during patient scan if system does not have correct version of gantry software installed. Unintended detector motion may result in life-threatening injury.
Product
Varicam, Millennium VG, Millennium VG Hawkeye, Discovery VH Nuclear Medicine System
Where it went
Worldwide - US Nationwide distribution.
How much
21

GE Healthcare Brivo NM 615, Model/Catalog Numbers: 1) 5376204-15, 2) H2401MA, 3) H3100TA; System, Tomog

Z-0889-2025  ·  Class II  ·  reported 2025-01-22  ·  Ongoing

Why it was recalled
GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
Product
GE Healthcare Brivo NM 615, Model/Catalog Numbers: 1) 5376204-15, 2) H2401MA, 3) H3100TA; System, Tomography, Computed, Emission
Where it went
Worldwide
How much
116 units

GE Healthcare Infinia I, Model/Catalog Numbers: 1) 5178728, 2) H3000WE, 3) H3000WM, 4) H3000YG; Syste

Z-0888-2025  ·  Class II  ·  reported 2025-01-22  ·  Ongoing

Why it was recalled
GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
Product
GE Healthcare Infinia I, Model/Catalog Numbers: 1) 5178728, 2) H3000WE, 3) H3000WM, 4) H3000YG; System, Tomography, Computed, Emission
Where it went
Worldwide
How much
36 units

GE Healthcare Discovery NM/CT 670 CZT, Model/Catalog Numbers: 1) H3906AB; System, Tomography, Computed, Emi

Z-0895-2025  ·  Class II  ·  reported 2025-01-22  ·  Ongoing

Why it was recalled
GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
Product
GE Healthcare Discovery NM/CT 670 CZT, Model/Catalog Numbers: 1) H3906AB; System, Tomography, Computed, Emission
Where it went
Worldwide
How much
8 units

GE Healthcare Discovery NM/CT 670 DR, Model/Catalog Numbers: 1) 5376204-70-56-2; 2) H2401AW; 3) H3100AN

Z-0892-2025  ·  Class II  ·  reported 2025-01-22  ·  Ongoing

Why it was recalled
GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
Product
GE Healthcare Discovery NM/CT 670 DR, Model/Catalog Numbers: 1) 5376204-70-56-2; 2) H2401AW; 3) H3100AN; 4) H3100BT; 5) H3100PX; System, Tomography, Computed, Emission
Where it went
Worldwide
How much
76 units

GE Healthcare Discovery NM/CT 670 ES, Model/Catalog Numbers: 1) H3100XF ; System, Tomography, Computed, Emi

Z-0894-2025  ·  Class II  ·  reported 2025-01-22  ·  Ongoing

Why it was recalled
GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
Product
GE Healthcare Discovery NM/CT 670 ES, Model/Catalog Numbers: 1) H3100XF ; System, Tomography, Computed, Emission
Where it went
Worldwide
How much
17 units

GE Healthcare Discovery NM/CT 670, Model/Catalog Numbers: 1) 5376204-70-54-2, 2) H2401BC, 3) H2401CE,

Z-0891-2025  ·  Class II  ·  reported 2025-01-22  ·  Ongoing

Why it was recalled
GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
Product
GE Healthcare Discovery NM/CT 670, Model/Catalog Numbers: 1) 5376204-70-54-2, 2) H2401BC, 3) H2401CE, 4) H2401CG, 5) H3100AG, 6) H3100JW, 7) H3100NC; System, Tomography, Computed, Emission
Where it went
Worldwide
How much
859 units

GE Healthcare Infinia I Hawkeye 1, Model/Catalog Numbers: 1) H3000WN; 2) H3000YM; 3) H3000YS; System, T

Z-0898-2025  ·  Class II  ·  reported 2025-01-22  ·  Ongoing

Why it was recalled
GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
Product
GE Healthcare Infinia I Hawkeye 1, Model/Catalog Numbers: 1) H3000WN; 2) H3000YM; 3) H3000YS; System, Tomography, Computed, Emission
Where it went
Worldwide
How much
93 units

GE Healthcare Infinia II, Model/Catalog Numbers: 1) H2799ZP; 2) H3000WT; System, Tomography, Computed, Em

Z-0897-2025  ·  Class II  ·  reported 2025-01-22  ·  Ongoing

Why it was recalled
GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
Product
GE Healthcare Infinia II, Model/Catalog Numbers: 1) H2799ZP; 2) H3000WT; System, Tomography, Computed, Emission
Where it went
Worldwide
How much
877 units

GE Healthcare Optima NM/CT 640, Model/Catalog Numbers: 1) H3100JZ; System, Tomography, Computed, Emission

Z-0896-2025  ·  Class II  ·  reported 2025-01-22  ·  Ongoing

Why it was recalled
GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
Product
GE Healthcare Optima NM/CT 640, Model/Catalog Numbers: 1) H3100JZ; System, Tomography, Computed, Emission
Where it went
Worldwide
How much
430 units

Showing the 30 most recent of 41. The rest are searchable at the FDA enforcement report database.

Read these counts with three things in mind

A firm with more recalls usually makes more products. A contract packer runs hundreds of lines. It will out-recall a company with one product whatever either does. This is not a safety ranking.

Most recalls here are voluntary. A firm found its own fault and pulled the stock. That is the system working. It lands in this database looking the same as a recall the FDA had to push for.

Terminated means the FDA considers the recall finished. It is not a verdict. It does not mean anyone was harmed.

Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.