Elekta, Inc.: 116 FDA recall reports
Every medical device recall the FDA has published for Elekta, Inc.. The fault, and how the agency graded it.
116Reports
0Class I
33Ongoing
2026-07-15Most recent
What the classes mean
The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.
Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.
By class
By year reported
Checking whether it affects you
The lot or code number is what matters. Not the product name. A recall covers
specific lots. Yours may not be one of them.
Check the code on your package against the report. The FDA keeps the current
list at fda.gov/safety/recalls.
The reports
The text below is quoted from the FDA record word for word.
Brand Name: Fletcher CT/MR Shielded Applicator Set Product Name: Intracavitary Remote Afterloading Applicator
Z-2710-2026 · Class II · reported 2026-07-15 · Ongoing
- Why it was recalled
- Due to safety concerns regarding the detachment of the tip of the intrauterine tube (IU tube) that may remain within a patient.
- Product
- Brand Name: Fletcher CT/MR Shielded Applicator Set Product Name: Intracavitary Remote Afterloading Applicator Product Description: The Fletcher CT/MR Shielded Applicator Set is intended for gynecological brachytherapy treatment of the cervix and endometrium.
- Where it went
- Worldwide - U.S. Nationwide distribution including in the states of AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TX, UT, VA, WI, and WV. The countries of Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Botswana, Brazil, Burkina Faso, Cambodia, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Georgia, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iran (Islam Republic of), Iraq, Ireland, Italy, Japan, Kazakhstan, Kenya, Korea (Republic of), Kosovo, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Rwanda, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Swaziland, Sweden, Switzerland, Taiwan, Thailand, Togo, Turkey, Ukraine, United Arab Emirates, United Kingdom, Venezuela, Viet Nam, and Zambia.
- How much
- 1
Brand Name: Brachytherapy Applicator Product Name: Advanced Gynecological Applicator Product Description: Th
Z-2711-2026 · Class II · reported 2026-07-15 · Ongoing
- Why it was recalled
- Due to safety concerns regarding the detachment of the tip of the intrauterine tube (IU tube) that may remain within a patient.
- Product
- Brand Name: Brachytherapy Applicator Product Name: Advanced Gynecological Applicator Product Description: The intended use of the Advanced Gynecological Applicator is intracavitary brachytherapy for cancer treatment of the cervix and endometrium. Optional needles can be placed for interstitial brachytherapy.
- Where it went
- Worldwide - U.S. Nationwide distribution including in the states of AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TX, UT, VA, WI, and WV. The countries of Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Botswana, Brazil, Burkina Faso, Cambodia, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Georgia, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iran (Islam Republic of), Iraq, Ireland, Italy, Japan, Kazakhstan, Kenya, Korea (Republic of), Kosovo, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Rwanda, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Swaziland, Sweden, Switzerland, Taiwan, Thailand, Togo, Turkey, Ukraine, United Arab Emirates, United Kingdom, Venezuela, Viet Nam, and Zambia.
- How much
- 586
Brand Name: Geneva Product Name: Brachytherapy Applicator Product Description: The device is intended for in
Z-2712-2026 · Class II · reported 2026-07-15 · Ongoing
- Why it was recalled
- Due to safety concerns regarding the detachment of the tip of the intrauterine tube (IU tube) that may remain within a patient.
- Product
- Brand Name: Geneva Product Name: Brachytherapy Applicator Product Description: The device is intended for intracavitary brachytherapy for cancer treatment of the cervix and the endometrium. Optional needles can be placed for interstitial therapy.
- Where it went
- Worldwide - U.S. Nationwide distribution including in the states of AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TX, UT, VA, WI, and WV. The countries of Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Botswana, Brazil, Burkina Faso, Cambodia, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Georgia, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iran (Islam Republic of), Iraq, Ireland, Italy, Japan, Kazakhstan, Kenya, Korea (Republic of), Kosovo, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Rwanda, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Swaziland, Sweden, Switzerland, Taiwan, Thailand, Togo, Turkey, Ukraine, United Arab Emirates, United Kingdom, Venezuela, Viet Nam, and Zambia.
- How much
- 1317
MOSAIQ Oncology Information System
Z-2406-2026 · Class II · reported 2026-06-17 · Ongoing
- Why it was recalled
- Software may fail to load adapted baseline shift plan during treatment delivery when a communication error occurs.
- Product
- MOSAIQ Oncology Information System
- Where it went
- Worldwide - US Nationwide distribution in the states of WI, FL, LA, NY, NY, NJ, KS, FL, IA, NM, TX and the countries of Bahrain, Belgium, Canada, China, Denmark, France, Germany, Hong Kong, India, Israel, Italy, Japan, Lithuania, Malta, Mexico, Netherlands, Pakistan, Panama, Poland, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom.
- How much
- 12 programs
Brand Name: Leksell Vantage Arc System REF: 1053958
Z-2006-2026 · Class II · reported 2026-05-06 · Ongoing
- Why it was recalled
- Locking mechanisms may not properly secure to the Arc system, resulting in shifting of coordinates during procedures.
- Product
- Brand Name: Leksell Vantage Arc System REF: 1053958
- Where it went
- US: AR AZ CA CO DC FL GA ID IL KS MA MD ME MO MT NC NE NH NJ NM NV NY OH OK PA SC TX UT VA WA WI WV OUS: Algeria Australia Austria Azerbaijan Belgium China Colombia Costa Rica Croatia Czech Republic Egypt Estonia Finland France Germany Greece Hong Kong Hungary India Israel Italy Japan Korea, Republic of Kuwait Lithuania Mexico Morocco Netherlands New Zealand Norway Pakistan Panama Poland Portugal Saudi Arabia Singapore Slovenia South Africa Spain Sweden Switzerland Taiwan Turkey United Kingdom
- How much
- 768 (364Left/364Right)
Brand Name: Leksell Gamma Knife Product Description: Leksell Gamma Knife¿ Perfexion Leksell Gamma Knife¿ Ico
Z-1581-2026 · Class II · reported 2026-03-25 · Ongoing
- Why it was recalled
- Due to adapter not being properly aligned or locked in place.
- Product
- Brand Name: Leksell Gamma Knife Product Description: Leksell Gamma Knife¿ Perfexion Leksell Gamma Knife¿ Icon" Leksell Gamma Knife¿ Elekta Esprit Leksell Gamma Knife¿ (available models Elekta Esprit, Icon", and Perfexion") is a radiosurgery system for use in the stereotactic irradiation of head structures. Surgery is achieved by delivering a prescribed dose as one or more shots of ionizing radiation to the exact site of the target.
- Where it went
- Worldwide - U.S. Nationwide distribution including in the states of AK, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KY, LA, MA, MI, MN, MO, MS, MT, NC, NJ, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The countries of Algeria, Argentina, Australia, Austria, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, France, Georgia, Germany, Greece, Hong Kong, India, Indonesia, Iran, Iraq, Italy, Japan, Jordan, Kazakhstan, Republic of Korea, Kuwait, Libyan Arab Jamahiriya, Lithuania, Malaysia, Mexico, Morocco, Netherlands, Norway, Pakistan, Peru, Philippines, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Taiwan, Thailand, Turkey, United Kingdom, and Vietnam.
- How much
- 352
Brand Name: Leksell GammaPlan Product Name: Radionuclide RT Treatment planning system Software Version: 11,3
Z-1340-2026 · Class II · reported 2026-02-18 · Ongoing
- Why it was recalled
- Due to an error when defining a new stereotactic reference that users fail to first change an already defined obsolete reference to a pre-plan reference. This may result in a treatment plan that potentially irradiates the incorrect patient location if the error is not detected by users.
- Product
- Brand Name: Leksell GammaPlan Product Name: Radionuclide RT Treatment planning system Software Version: 11,3, 11.4 Product Description: Leksell GammaPlan¿ is a computer-based system designed for Leksell Gamma Knife¿ treatment planning.
- Where it went
- Worldwide - U.S. Nationwide distribution including in the states of AK, AZ, CA, CO, ST, FL, GA, HI, IA, IL, LA, MA, MI, MN. MO, NC, NJ, NM, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The countries of Algeria, Argentina, Australia, Belarus, Belgium, Brazil, Bulgaria, Canada, China, Colombia, Croatia, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, France, Georgia, Germany, Hong Kong, India, Iran, Iraq, Italy, Japan, Jordan, Republic of Korea, Kuwait, Libyan Arab Jamahiriya, Lithuania, Malaysia, Mexico, Morocco, Netherlands, Norway, Pakistan, Peru, Poland, Romania, Russian Federation, Saudi Arabia, South Africa, Spain, Sweden, Taiwan, Thailand, Turkey, United Kingdom, and Vietnam.
- How much
- N/A
Brand Name: Leksell GammaPlan Product Name: Radionuclide RT Treatment planning system Software Version: 11.1
Z-1339-2026 · Class II · reported 2026-02-18 · Ongoing
- Why it was recalled
- Due to an error when defining a new stereotactic reference that users fail to first change an already defined obsolete reference to a pre-plan reference. This may result in a treatment plan that potentially irradiates the incorrect patient location if the error is not detected by users.
- Product
- Brand Name: Leksell GammaPlan Product Name: Radionuclide RT Treatment planning system Software Version: 11.1 Product Description: Leksell GammaPlan¿ is a computer-based system designed for Leksell Gamma Knife¿ treatment planning.
- Where it went
- Worldwide - U.S. Nationwide distribution including in the states of AK, AZ, CA, CO, ST, FL, GA, HI, IA, IL, LA, MA, MI, MN. MO, NC, NJ, NM, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The countries of Algeria, Argentina, Australia, Belarus, Belgium, Brazil, Bulgaria, Canada, China, Colombia, Croatia, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, France, Georgia, Germany, Hong Kong, India, Iran, Iraq, Italy, Japan, Jordan, Republic of Korea, Kuwait, Libyan Arab Jamahiriya, Lithuania, Malaysia, Mexico, Morocco, Netherlands, Norway, Pakistan, Peru, Poland, Romania, Russian Federation, Saudi Arabia, South Africa, Spain, Sweden, Taiwan, Thailand, Turkey, United Kingdom, and Vietnam.
- How much
- N/A
Brand Name: Elekta Medical Linear Accelerator Product Name: Elekta Synergy, Elekta Harmony, Elekta Infinity,
Z-1164-2026 · Class II · reported 2026-02-04 · Ongoing
- Why it was recalled
- Due to a manufacturing issue and electrical grounding of systems.
- Product
- Brand Name: Elekta Medical Linear Accelerator Product Name: Elekta Synergy, Elekta Harmony, Elekta Infinity, Versa HD Software Version: N/A Product Description: Digital Linear Accelerator Component: No
- Where it went
- Worldwide
- How much
- 5391
MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)
Z-1078-2026 · Class II · reported 2026-01-21 · Ongoing
- Why it was recalled
- Using oncology information system that manages workflow may result in overtreatment if:Plan delivered with Particle Therapy IHE-RO TDW-II interface,With one field per fraction or last field partial treatment,Partial treatment occurs followed by treatment field interruptions with manually recorded partial treatment,Then treatment unit doesn't check meterset value so partial treatment delivered twice
- Product
- MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)
- Where it went
- US Nationwide distribution in the state of TX.
- How much
- 1
Brand Name: Leksell Stereotactic System Product Name: Leksell Vantage Stereotactic System Software Version:
Z-0433-2026 · Class II · reported 2025-11-12 · Ongoing
- Why it was recalled
- Due to increased risk that paint flakes from the engravings may separate from the device and there is the possibility that the flake may inadvertently enter the wound
- Product
- Brand Name: Leksell Stereotactic System Product Name: Leksell Vantage Stereotactic System Software Version: n/a Product Description: The Intended Purpose of Leksell¿ Vantage" Stereotactic System is target localization and fixation of the patient head in a coordinate system in order to perform stereotactic neurosurgical procedures, for example; deep brain stimulation, lesioning, biopsies, targeted injections, aspirations and minimal invasive tumor treatments.
- Where it went
- Worldwide - U.S. Nationwide distribution in the states of AR, CA, CO, FL, ID, IL, LA, MA, MI, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, PA, SC, TN, TX, UT, WA, WI, and WV. The countries of Algeria, Australia, Austria, Belgium, China, Colombia, Costa Rica, Croatia, Czech Republic, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Israel, Italy, Japan, Republic of Korea, Kuwait, Lithuania, Mexico, Netherlands, New Zealand, Norway, Pakistan, Portugal, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, and United Kingdom.
- How much
- 338
MOSAIQ Oncology Information System
Z-1162-2025 · Class II · reported 2025-02-19 · Ongoing
- Why it was recalled
- Image offsets calculated by oncology Information System software may be incorrect for Linear Accelerator (LINACs) not characterized at IEC61217 scale.
- Product
- MOSAIQ Oncology Information System
- Where it went
- US: PA, AL, FL, IA UT OUS: Austria Belgium Bolivia Bosnia and Herzegowina Brazil Canada France Germany Hong Kong Italy Japan Korea, Republic of Malta Netherlands Panama South Africa Spain Taiwan Thailand Turkey Venezuela
- How much
- 64 units
ELEKTA LIMITED Elekta Infinity, REF XRT 0511, XRT 0521, XRT 0531, XRT 0541; Medical Charged-Particle Radiation
Z-0446-2024 · Class II · reported 2023-12-13 · Ongoing
- Why it was recalled
- Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.
- Product
- ELEKTA LIMITED Elekta Infinity, REF XRT 0511, XRT 0521, XRT 0531, XRT 0541; Medical Charged-Particle Radiation Therapy System Accelerator, Linear
- Where it went
- worldwide
- How much
- 291 units
ELEKTA LIMITED Elekta Synergy Platform, REF MRT 8501, MRT 8511; Medical Charged-Particle Radiation Therapy Sys
Z-0444-2024 · Class II · reported 2023-12-13 · Ongoing
- Why it was recalled
- Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.
- Product
- ELEKTA LIMITED Elekta Synergy Platform, REF MRT 8501, MRT 8511; Medical Charged-Particle Radiation Therapy System Accelerator, Linear
- Where it went
- worldwide
- How much
- 264 units
ELEKTA LIMITED Versa HD, REF XRT 2121, 2131; Medical Charged-Particle Radiation Therapy System Accelerator, Li
Z-0445-2024 · Class II · reported 2023-12-13 · Ongoing
- Why it was recalled
- Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.
- Product
- ELEKTA LIMITED Versa HD, REF XRT 2121, 2131; Medical Charged-Particle Radiation Therapy System Accelerator, Linear
- Where it went
- worldwide
- How much
- 451 units
ELEKTA AXESSE, REF XRT 2181, XRT 27891; Medical Charged-Particle Radiation Therapy System Accelerator, Linear
Z-0447-2024 · Class II · reported 2023-12-13 · Ongoing
- Why it was recalled
- Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.
- Product
- ELEKTA AXESSE, REF XRT 2181, XRT 27891; Medical Charged-Particle Radiation Therapy System Accelerator, Linear
- Where it went
- worldwide
- How much
- 47 units
ELEKTA LIMITED Elekta Synergy, REF MRT 8501, MRT 8511; Medical Charged-Particle Radiation Therapy System Accel
Z-0443-2024 · Class II · reported 2023-12-13 · Ongoing
- Why it was recalled
- Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.
- Product
- ELEKTA LIMITED Elekta Synergy, REF MRT 8501, MRT 8511; Medical Charged-Particle Radiation Therapy System Accelerator, Linear
- Where it went
- worldwide
- How much
- 1019 units
ELEKTA LIMITED Precise Digital Accelerator, REF MRT 6001, MRT 6011; Medical Charged-Particle Radiation Therapy
Z-0442-2024 · Class II · reported 2023-12-13 · Ongoing
- Why it was recalled
- Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.
- Product
- ELEKTA LIMITED Precise Digital Accelerator, REF MRT 6001, MRT 6011; Medical Charged-Particle Radiation Therapy System Accelerator, Linear
- Where it went
- worldwide
- How much
- 315 units
Elekta Unity Image-Guided Radiation Therapy System containing an MR gradient coil.
Z-1675-2023 · Class II · reported 2023-06-14 · Ongoing
- Why it was recalled
- There is a low possibility that an electrical connector in the MR gradient coil will overheat on Elekta Unity systems.
- Product
- Elekta Unity Image-Guided Radiation Therapy System containing an MR gradient coil.
- Where it went
- US Nationwide distribution in the states of FL, IA, NJ, NY, OK, PA, TN, WI, MI, LA, KS, NM and TX.
- How much
- 18 devices
Elekta Medical Linear Accelerator (ELMA) included in the below platforms: a. Elekta Infinity; b. Versa H
Z-1665-2023 · Class II · reported 2023-06-07 · Ongoing
- Why it was recalled
- A component within a limited number of Elekta Linacs has been realized to be non-compliant to IEC 60601
- Product
- Elekta Medical Linear Accelerator (ELMA) included in the below platforms: a. Elekta Infinity; b. Versa HD; c. Elekta Synergy
- Where it went
- Worldwide distribution. US distribution to AR, IN, LA, OK, TN, VA, WA. International distribution to Belgium, Canada, Chile, China, France, Germany, India, Iraq, Italy, Jordan, Libyan Arab Jamahiriya, Mauritius, Mexico, Morocco, Poland, Russian Federation, Spain.
- How much
- 46 units
Monaco RTP System. Used to make treatment plans for patients with prescriptions for external beam radiati
Z-1233-2023 · Class II · reported 2023-03-15 · Ongoing
- Why it was recalled
- Re-optimization, after adding contours without forced density outside the external structure, may result in inaccurate dose presentation.
- Product
- Monaco RTP System. Used to make treatment plans for patients with prescriptions for external beam radiation therapy.
- Where it went
- Nationwide distribution in the US. International distribution to Albania, Algeria, Antigua and Barbuda, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegowina, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Georgia, Germany, Ghana, Greece, Hong Kong, Hungary, India, Indonesia, Iran (Islamic Republic of), Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Republic of, Lao People's Democratic Republic, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Macedonia, Malaysia, Malta, Mexico, Morocco, Mozambique, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Pakistan, Palestinian Territory, Occupied, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Rwanda, Saudi Arabia, Senegal, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Suriname, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam, Zimbabwe.
- How much
- 2,020 units
Oncentra Brachy radiation therapy planning system software.
Z-0719-2023 · Class II · reported 2022-12-21 · Ongoing
- Why it was recalled
- A reconstruction error may occur when using the Catheter Bending functionality in Applicator Modeling or Implant Modeling
- Product
- Oncentra Brachy radiation therapy planning system software.
- Where it went
- Distribution US nationwide, including Puerto Rico. There was no military/government distribution.
- How much
- 336 systems
Oncentra Brachy users using version 4.0 and above with Applicator Modeling license 170283-00
Z-0182-2022 · Class II · reported 2021-11-03 · Ongoing
- Why it was recalled
- Under certain circumstances, there can be a difference between the planned and delivered dose of the two ovoid channels.
- Product
- Oncentra Brachy users using version 4.0 and above with Applicator Modeling license 170283-00
- Where it went
- Distribution throughout USA OUS distribution including:Algeria, Argentina, Australia, Austria Bahamas, Bangladesh, Belarus, Belgium, Bolivia, Botswana, Brazil Brunei, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Ghana, Greece, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran (Islamic Republic of), Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kosovo, Libyan Arab Jamahiriya, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Pakistan, Peru, Philippines, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Saudi Arabia, Senegal, Singapore, Slovakia, South Africa, Spain, Sudan, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Vietnam
- How much
- 1072 devices
Elekta Medical Linear Accelerator (EMLA), configured as follows: a. Elekta Precise Digital Accelerator;
Z-0116-2022 · Class II · reported 2021-10-20 · Terminated
- Why it was recalled
- There is a risk that the Diode D1 in the Modulator can fail in a destructive manner when removed from the machine, and that the protective tube surrounding the diode may fail to contain the ejected matter.
- Product
- Elekta Medical Linear Accelerator (EMLA), configured as follows: a. Elekta Precise Digital Accelerator; b. Elekta Harmony; c. Elekta Synergy; d. Elekta Infinity; e. Versa HD
- Where it went
- Distributed nationwide to FL, NC, OH, CA, IL, MD, OR, AL, NY, PA, MA, AK, GA, OR, KY, VA, TX, SD, AZ, NJ, MI, WI, ND, MO, WA, OK, PR, NM, NC, ID, CT, DE, LA, MT, TN, CO, UT, MN, IN, RI, IA, DC, ND, AZ, HI, KS, NE, NH, AR, WY, WV, MS, VT and internationally to Albania, Antigua/Barbuda, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegowina, Botswana, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Finland, France, Gabon, Georgia, Germany, Ghana, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Kosovo, Kyrgystan, Lao People's Democratic Republic, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Mozambique, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Pakistan, Palestinian Territory, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Rwanda, Saudi Arabia, Senegal, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri...
- How much
- 4,405 units
MOSAIQ Oncology Information System
Z-2591-2021 · Class II · reported 2021-10-06 · Ongoing
- Why it was recalled
- A drug strength in MOSAIQ can be changed during the ordering process and approved, but the saved drug strength reverts to the original value.
- Product
- MOSAIQ Oncology Information System
- Where it went
- Distribution was made to CA, CO, FL, ID, LA, MO, MS, NC, NH, NM, NV, NY, OH, OK, PA, SD, and WA. There was no military/government distribution. Foreign distribution was made to Australia, Canada, and the United Kingdom.
- How much
- 31
Elekta Monaco - Product Usage: used to make treatment plans for patients with prescriptions for external beam
Z-2563-2021 · Class II · reported 2021-10-06 · Ongoing
- Why it was recalled
- Contour changes can be saved on an unintended image set. In addition, these contour edits do not cause the dose to be frozen on the plans associated with that image set.
- Product
- Elekta Monaco - Product Usage: used to make treatment plans for patients with prescriptions for external beam radiation therapy.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of PA, MI, WI, NY, TN, NJ, IA, TX and the countries of United Kingdom, Turkey, Switzerland, Sweden, Spain, Netherlands, Korea, Thailand, Japan, Italy, Hong Kong, Germany, France, Denmark, China, Canada, Belgium, Bahrain, Australia.
- How much
- 214 units
Monaco RTP System, Software Builds 5.40.00, 5.40.01, 5.40.02, and 5.51.10
Z-2406-2021 · Class II · reported 2021-09-08 · Ongoing
- Why it was recalled
- If a couch is removed while a plan is loaded, and then added back to the Studyset when the plan is unloaded, it is possible to calculate dose with the couch unassigned to the beams.
- Product
- Monaco RTP System, Software Builds 5.40.00, 5.40.01, 5.40.02, and 5.51.10
- Where it went
- Distribution in US - IA, NJ, NY, TN, PA, TX, and WI OUS distribution in Australia, Bahrain, Belgium, Canada, China, Denmark, France, Germany, Hong Kong, Italy, Japan, Korea, Netherlands, Spain, Sweden, Switzerland, Thailand, Turkey and United Kingdom
- How much
- N/A
Elekta MOSAIC Oncology Information System versions: 2.62, 2.64, 2.70, 2.81, 2.82, and 2.83 hosted on Classic
Z-1783-2021 · Class II · reported 2021-06-09 · Terminated
- Why it was recalled
- Malware attack
- Product
- Elekta MOSAIC Oncology Information System versions: 2.62, 2.64, 2.70, 2.81, 2.82, and 2.83 hosted on Classic Cloud
- Where it went
- Nationwide
- How much
- 40 units
Elekta Unity, Image-Guided Radiation Therapy System
Z-2473-2020 · Class II · reported 2020-07-08 · Terminated
- Why it was recalled
- There have been reports of excessive heating in the MRL Gradient Connection Module.
- Product
- Elekta Unity, Image-Guided Radiation Therapy System
- Where it went
- Distributed in US - PA, NY, and IA. Distributed ROW - Australia, Canada, China, Switzerland, Denmark, Hong Kong, Italy, Netherlands, Sweden, Thailand, Turkey, UK. There was no government/military distribution.
- How much
- 21 devices
Elekta Unity, UDI 05060191071321, Image-guided radiation therapy system - Product Usage: is indicated for radi
Z-2289-2020 · Class II · reported 2020-06-17 · Terminated
- Why it was recalled
- potential risk that Legionella may be present in the Unity machine room, specifically the heat exchanger condensation collection tray.
- Product
- Elekta Unity, UDI 05060191071321, Image-guided radiation therapy system - Product Usage: is indicated for radiation therapy treatments and stereotactic radiation treatments of malignant and benign diseases anywhere in the body as determined by a licensed medical practitioner in accordance with a defined treatment plan.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of PA, WI, NY, TN, NJ, IA, TX, and the countries of Australia, Canada, China, Denmark, Germany, Hong Kong, Italy, Netherlands, Sweden, Switzerland, Thailand, Turkey, UK.
- How much
- 32 units
Showing the 30 most recent of 116. The rest are searchable at the FDA enforcement report database.
Read these counts with three things in mind
A firm with more recalls usually makes more products. A contract packer runs
hundreds of lines. It will out-recall a company with one product whatever either
does. This is not a safety ranking.
Most recalls here are voluntary. A firm found its own fault and pulled the stock.
That is the system working. It lands in this database looking the same as a recall
the FDA had to push for.
Terminated means the FDA considers the recall finished. It is not a verdict.
It does not mean anyone was harmed.
Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.
← All device recalls