Draeger Medical, Inc.

Draeger Medical, Inc.: 78 FDA recall reports

Every medical device recall the FDA has published for Draeger Medical, Inc.. The fault, and how the agency graded it.

78Reports
53Class I
12Ongoing
2024-12-25Most recent
53 Class I on this page

Class I means the FDA judged there was a reasonable probability of serious harm or death. Check the lot numbers against anything you hold.

What the classes mean

The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.

ClassWhat the FDA is saying
Class IThere is a reasonable probability that using this will cause serious harm or death.
Class IIIt may cause temporary or reversible harm. Serious harm is unlikely.
Class IIIIt breaks a rule but is unlikely to hurt anyone. Often a labelling fault.

Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.

By class

ClassReports
Class I53
Class II25

By year reported

YearReports
20133
20141
20154
20166
201810
20199
202110
20224
202326
20245
Checking whether it affects you

The lot or code number is what matters. Not the product name. A recall covers specific lots. Yours may not be one of them.

Check the code on your package against the report. The FDA keeps the current list at fda.gov/safety/recalls.

The reports

The text below is quoted from the FDA record word for word.

Class I

Brand Name: Atlan Product Name: Atlan A300XL Model/Catalog Number: 8621400 The Atlan is an anesthesia

Z-0668-2025  ·  Class I  ·  reported 2024-12-25  ·  Ongoing

Why it was recalled
The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
Product
Brand Name: Atlan Product Name: Atlan A300XL Model/Catalog Number: 8621400 The Atlan is an anesthesia workstation intended for use in anesthetizing adults, pediatrics, and neonates.
Where it went
No Atlan A300 or A300XL devices were imported into the US. Devices were direct shipped from Draegerwerk Ag and Co. KGaA to the following countries: Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Bulgaria, Cambodia, Canada, Chili, China, Colombia, Croatia, Czech Republic, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Hungary, India, Indonesia, Ireland, Italy, Japan, Jordan, Kazakhstan, Kenya, Latvia, Malaysia, Mauritius, Mexico, Morocco, Namibia, Nepal, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Kingdom, Uruguay, United Arab Emirates, and Vietnam.
How much
25 units (OUS Only)
Class I

Brand Name: Atlan Product Name: Atlan A300 Model/Catalog Number: 8211300 The Atlan is an anesthesia wo

Z-0667-2025  ·  Class I  ·  reported 2024-12-25  ·  Ongoing

Why it was recalled
The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
Product
Brand Name: Atlan Product Name: Atlan A300 Model/Catalog Number: 8211300 The Atlan is an anesthesia workstation intended for use in anesthetizing adults, pediatrics, and neonates.
Where it went
No Atlan A300 or A300XL devices were imported into the US. Devices were direct shipped from Draegerwerk Ag and Co. KGaA to the following countries: Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Bulgaria, Cambodia, Canada, Chili, China, Colombia, Croatia, Czech Republic, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Hungary, India, Indonesia, Ireland, Italy, Japan, Jordan, Kazakhstan, Kenya, Latvia, Malaysia, Mauritius, Mexico, Morocco, Namibia, Nepal, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Kingdom, Uruguay, United Arab Emirates, and Vietnam.
How much
280 units (OUS only)
Class I

Brand Name: Atlan Product Name: Atlan A350XL Model/Catalog Number: 8621600 Software Version: Not Applicable

Z-0305-2025  ·  Class I  ·  reported 2024-11-20  ·  Ongoing

Why it was recalled
The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
Product
Brand Name: Atlan Product Name: Atlan A350XL Model/Catalog Number: 8621600 Software Version: Not Applicable Component: Not applicable
Where it went
US Nationwide distribution in the states of AL, AZ, CA, FL, GA, IA, IL, KY, MA, ME, ND, NE, NH, NJ, NY, OH, PA, TN, TX, WI, WY. International distribution: Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Bulgaria, Cambodia, Canada, Chili, China, Columbia, Croatia, Czech Republic, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Hungary, India, Indonesia, Ireland, Italy, Japan, Jordan, Kazakhstan, Kenya, Latvia, Malaysia, Mauritius, Mexico, Morocco, Namibia, Nepal, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Kingdom, Uruguay, United Arab Emirates, Vietnam.
How much
455 units (310 units US, 145 OUS)
Class I

Brand Name: Atlan Product Name: Atlan A350 Model/Catalog Number: 8211500 Software Version: Not applicable

Z-0304-2025  ·  Class I  ·  reported 2024-11-20  ·  Ongoing

Why it was recalled
The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
Product
Brand Name: Atlan Product Name: Atlan A350 Model/Catalog Number: 8211500 Software Version: Not applicable Component: Not applicable
Where it went
US Nationwide distribution in the states of AL, AZ, CA, FL, GA, IA, IL, KY, MA, ME, ND, NE, NH, NJ, NY, OH, PA, TN, TX, WI, WY. International distribution: Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Bulgaria, Cambodia, Canada, Chili, China, Columbia, Croatia, Czech Republic, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Hungary, India, Indonesia, Ireland, Italy, Japan, Jordan, Kazakhstan, Kenya, Latvia, Malaysia, Mauritius, Mexico, Morocco, Namibia, Nepal, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Kingdom, Uruguay, United Arab Emirates, Vietnam.
How much
1,492 units (139 units US, 1,353 OUS)
Class I

Dr¿ger Perseus A500 Anesthesia Workstation

Z-1512-2024  ·  Class I  ·  reported 2024-04-24  ·  Ongoing

Why it was recalled
Unexpected shutdown while operating on battery power.
Product
Dr¿ger Perseus A500 Anesthesia Workstation
Where it went
Domestic: Nationwide Distribution; Foreign: Angola, Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Bolivia, Botswana, Brazil, Brunei Dar-se- S, Bulgaria, Canada, Chili, China, Columbia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, French Polynesia, French Guiana, Georgia, Germany, Gibraltar, Greece, Guadeloupe, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Liechtenstein, Lithuania, Malaysia, Maldives, Mauritius, Mayotte, Mexico, Morocco, Netherlands, New Zealand, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovenia, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Turkmenistan, Uganda, Ukraine, United Kingdom, Uruguay, United Arab Emirates, Vietnam.
How much
16,841 units
Class I

Carina Sub-Acute Care Ventilator

Z-2248-2023  ·  Class I  ·  reported 2023-08-30  ·  Ongoing

Why it was recalled
Presence of polyether polyurethane (PE-PUR) and emission of 1,3-Dichloropropan-2-ol that exceed the acceptable uptake level during continuous use (>30 days) in pediatric patients.
Product
Carina Sub-Acute Care Ventilator
Where it went
US Nationwide Distribution.
How much
11,621 units
Class I

Oxylog 3000 Plus emergency and transport ventilator

Z-2018-2023  ·  Class I  ·  reported 2023-07-19  ·  Ongoing

Why it was recalled
Draeger has become aware of instances where the device stopped working when in use when transitioning from battery to AC operation.
Product
Oxylog 3000 Plus emergency and transport ventilator
Where it went
Distributed Nationwide. OUS Distribution: Afghanistan, Algeria, Andorra, Angola, Antigua/Barbados, Argentina, Australia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Bolivia, Brunei Dar-es-S, Bulgaria, Burkina-Faso, Cambodia, Canada, Cayman Islands, Chile, China, Christmas Island, Columbia, Congo, Costa Rica, Croatia, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Equatorial Guinea, Estonia, Ethiopia, Faeroe, Finland, France, French Polynesia, French Guiana, Gabon, Georgia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Guinea, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Ivory Coast, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Laos, Latvia, Lebanon, Libya, Liechtenstein, Lithuania, Macau, Malawi, Malaysia, Mali, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldavia, Mongolia, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, North Korea, Norway, Oman, Pakistan, Panama, Pap. New Guinea, Paraguay, Peru, Poland, Portugal, Qatar, Reunion, Romania, Russian Fed., Rwanda, San Marino, Saudi Arabia, Senegal, Serbia, Singapore, Sint Maarten, Slovenia, Slovakia, South...
How much
300 units
Class I

VentStar Basic (P)180, MP00351

Z-1440-2023  ·  Class I  ·  reported 2023-05-24  ·  Terminated

Why it was recalled
Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
Product
VentStar Basic (P)180, MP00351
Where it went
Domestic: US Nationwide Distribution.
How much
325 units
Class I

Set2Go Ventilation 12 (A), MP07968

Z-1459-2023  ·  Class I  ·  reported 2023-05-24  ·  Terminated

Why it was recalled
Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
Product
Set2Go Ventilation 12 (A), MP07968
Where it went
Domestic: US Nationwide Distribution.
How much
500 units
Class I

Seattle PAP Plus, MP02260

Z-1439-2023  ·  Class I  ·  reported 2023-05-24  ·  Terminated

Why it was recalled
Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
Product
Seattle PAP Plus, MP02260
Where it went
Domestic: US Nationwide Distribution.
How much
32,050 untis
Class I

VentStar Anesthesia WT (P)180, MP00374

Z-1444-2023  ·  Class I  ·  reported 2023-05-24  ·  Terminated

Why it was recalled
Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
Product
VentStar Anesthesia WT (P)180, MP00374
Where it went
Domestic: US Nationwide Distribution.
How much
130 units
Class I

Anesthesia Set Coax 180, MP03384

Z-1458-2023  ·  Class I  ·  reported 2023-05-24  ·  Terminated

Why it was recalled
Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
Product
Anesthesia Set Coax 180, MP03384
Where it went
Domestic: US Nationwide Distribution.
How much
20 units
Class I

VentStar Watertrap (N) 180, MP00363

Z-1449-2023  ·  Class I  ·  reported 2023-05-24  ·  Terminated

Why it was recalled
Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
Product
VentStar Watertrap (N) 180, MP00363
Where it went
Domestic: US Nationwide Distribution.
How much
125 units
Class I

VentStar Watertrap (P) 180, MP00361

Z-1442-2023  ·  Class I  ·  reported 2023-05-24  ·  Terminated

Why it was recalled
Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
Product
VentStar Watertrap (P) 180, MP00361
Where it went
Domestic: US Nationwide Distribution.
How much
50 units
Class I

VentStar Anesthesia (N) 180, MP00333

Z-1447-2023  ·  Class I  ·  reported 2023-05-24  ·  Terminated

Why it was recalled
Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
Product
VentStar Anesthesia (N) 180, MP00333
Where it went
Domestic: US Nationwide Distribution.
How much
N/A
Class I

Ventstar Coax, MP03373

Z-1454-2023  ·  Class I  ·  reported 2023-05-24  ·  Terminated

Why it was recalled
Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
Product
Ventstar Coax, MP03373
Where it went
Domestic: US Nationwide Distribution.
How much
2,380 units
Class I

ID Circuit Basic (P) 180, MP01340

Z-1446-2023  ·  Class I  ·  reported 2023-05-24  ·  Terminated

Why it was recalled
Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
Product
ID Circuit Basic (P) 180, MP01340
Where it went
Domestic: US Nationwide Distribution.
How much
25 units
Class I

VentStar Basic (P)250, MP00352

Z-1441-2023  ·  Class I  ·  reported 2023-05-24  ·  Terminated

Why it was recalled
Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
Product
VentStar Basic (P)250, MP00352
Where it went
Domestic: US Nationwide Distribution.
How much
200 units
Class I

Anesthesia Circuit Kit Coax 3, MP02732

Z-1453-2023  ·  Class I  ·  reported 2023-05-24  ·  Terminated

Why it was recalled
Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
Product
Anesthesia Circuit Kit Coax 3, MP02732
Where it went
Domestic: US Nationwide Distribution.
How much
329,294 units
Class I

Ventstar Coax 180, MP03374

Z-1455-2023  ·  Class I  ·  reported 2023-05-24  ·  Terminated

Why it was recalled
Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
Product
Ventstar Coax 180, MP03374
Where it went
Domestic: US Nationwide Distribution.
How much
18,160 units
Class I

Anesthesia Circuit Kit Coax 1, MP02730

Z-1452-2023  ·  Class I  ·  reported 2023-05-24  ·  Terminated

Why it was recalled
Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
Product
Anesthesia Circuit Kit Coax 1, MP02730
Where it went
Domestic: US Nationwide Distribution.
How much
200,900 units
Class I

VentStar Basic (N) 180, MP00353

Z-1448-2023  ·  Class I  ·  reported 2023-05-24  ·  Terminated

Why it was recalled
Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
Product
VentStar Basic (N) 180, MP00353
Where it went
Domestic: US Nationwide Distribution.
How much
1,725 units
Class I

VentStar breathing bag Set (P) 110, MP00384

Z-1445-2023  ·  Class I  ·  reported 2023-05-24  ·  Terminated

Why it was recalled
Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
Product
VentStar breathing bag Set (P) 110, MP00384
Where it went
Domestic: US Nationwide Distribution.
How much
25 units
Class I

Anesthesia Circuit Kit Coax HEPA, MP17102

Z-1460-2023  ·  Class I  ·  reported 2023-05-24  ·  Terminated

Why it was recalled
Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
Product
Anesthesia Circuit Kit Coax HEPA, MP17102
Where it went
Domestic: US Nationwide Distribution.
How much
720 units
Class I

Ventstar Coax 230, MP03376

Z-1457-2023  ·  Class I  ·  reported 2023-05-24  ·  Terminated

Why it was recalled
Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
Product
Ventstar Coax 230, MP03376
Where it went
Domestic: US Nationwide Distribution.
How much
40 units
Class I

VentStar Bag Set (N) 110, MP00383

Z-1450-2023  ·  Class I  ·  reported 2023-05-24  ·  Terminated

Why it was recalled
Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
Product
VentStar Bag Set (N) 110, MP00383
Where it went
Domestic: US Nationwide Distribution.
How much
375 units
Class I

VentStar Watertrap (P) 180 w/oLL, MP00362

Z-1443-2023  ·  Class I  ·  reported 2023-05-24  ·  Terminated

Why it was recalled
Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
Product
VentStar Watertrap (P) 180 w/oLL, MP00362
Where it went
Domestic: US Nationwide Distribution.
How much
25 units
Class I

VentStar Coax (P) 150, MP00379

Z-1451-2023  ·  Class I  ·  reported 2023-05-24  ·  Terminated

Why it was recalled
Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
Product
VentStar Coax (P) 150, MP00379
Where it went
Domestic: US Nationwide Distribution.
How much
280 units
Class I

ID Coax 180, MP03375

Z-1456-2023  ·  Class I  ·  reported 2023-05-24  ·  Terminated

Why it was recalled
Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
Product
ID Coax 180, MP03375
Where it went
Domestic: US Nationwide Distribution.
How much
20 units
Class I

NovaStar TS, Standard elbow, Reusable, Material Numbers MP01579 (Size S), MP01580 (Size M), and MP01581 (Size

Z-1138-2023  ·  Class I  ·  reported 2023-03-01  ·  Ongoing

Why it was recalled
The magnets used in the swivel frame/tube set and the headgear clips might impact medical devices (implanted as well as not implanted) and metallic implants.
Product
NovaStar TS, Standard elbow, Reusable, Material Numbers MP01579 (Size S), MP01580 (Size M), and MP01581 (Size L)
Where it went
US Nationwide Distribution
How much
893 units

Showing the 30 most recent of 78. The rest are searchable at the FDA enforcement report database.

Read these counts with three things in mind

A firm with more recalls usually makes more products. A contract packer runs hundreds of lines. It will out-recall a company with one product whatever either does. This is not a safety ranking.

Most recalls here are voluntary. A firm found its own fault and pulled the stock. That is the system working. It lands in this database looking the same as a recall the FDA had to push for.

Terminated means the FDA considers the recall finished. It is not a verdict. It does not mean anyone was harmed.

Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.

← All device recalls