Draeger Medical, Inc.: 78 FDA recall reports
Every medical device recall the FDA has published for Draeger Medical, Inc.. The fault, and how the agency graded it.
78Reports
53Class I
12Ongoing
2024-12-25Most recent
53 Class I on this pageClass I means the FDA judged there was a reasonable probability of serious harm or death. Check the lot numbers against anything you hold.
What the classes mean
The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.
Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.
By class
By year reported
Checking whether it affects you
The lot or code number is what matters. Not the product name. A recall covers
specific lots. Yours may not be one of them.
Check the code on your package against the report. The FDA keeps the current
list at fda.gov/safety/recalls.
The reports
The text below is quoted from the FDA record word for word.
Class I
Brand Name: Atlan Product Name: Atlan A300XL Model/Catalog Number: 8621400 The Atlan is an anesthesia
Z-0668-2025 · Class I · reported 2024-12-25 · Ongoing
- Why it was recalled
- The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
- Product
- Brand Name: Atlan Product Name: Atlan A300XL Model/Catalog Number: 8621400 The Atlan is an anesthesia workstation intended for use in anesthetizing adults, pediatrics, and neonates.
- Where it went
- No Atlan A300 or A300XL devices were imported into the US. Devices were direct shipped from Draegerwerk Ag and Co. KGaA to the following countries: Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Bulgaria, Cambodia, Canada, Chili, China, Colombia, Croatia, Czech Republic, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Hungary, India, Indonesia, Ireland, Italy, Japan, Jordan, Kazakhstan, Kenya, Latvia, Malaysia, Mauritius, Mexico, Morocco, Namibia, Nepal, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Kingdom, Uruguay, United Arab Emirates, and Vietnam.
- How much
- 25 units (OUS Only)
Class I
Brand Name: Atlan Product Name: Atlan A300 Model/Catalog Number: 8211300 The Atlan is an anesthesia wo
Z-0667-2025 · Class I · reported 2024-12-25 · Ongoing
- Why it was recalled
- The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
- Product
- Brand Name: Atlan Product Name: Atlan A300 Model/Catalog Number: 8211300 The Atlan is an anesthesia workstation intended for use in anesthetizing adults, pediatrics, and neonates.
- Where it went
- No Atlan A300 or A300XL devices were imported into the US. Devices were direct shipped from Draegerwerk Ag and Co. KGaA to the following countries: Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Bulgaria, Cambodia, Canada, Chili, China, Colombia, Croatia, Czech Republic, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Hungary, India, Indonesia, Ireland, Italy, Japan, Jordan, Kazakhstan, Kenya, Latvia, Malaysia, Mauritius, Mexico, Morocco, Namibia, Nepal, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Kingdom, Uruguay, United Arab Emirates, and Vietnam.
- How much
- 280 units (OUS only)
Class I
Brand Name: Atlan Product Name: Atlan A350XL Model/Catalog Number: 8621600 Software Version: Not Applicable
Z-0305-2025 · Class I · reported 2024-11-20 · Ongoing
- Why it was recalled
- The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
- Product
- Brand Name: Atlan Product Name: Atlan A350XL Model/Catalog Number: 8621600 Software Version: Not Applicable Component: Not applicable
- Where it went
- US Nationwide distribution in the states of AL, AZ, CA, FL, GA, IA, IL, KY, MA, ME, ND, NE, NH, NJ, NY, OH, PA, TN, TX, WI, WY. International distribution: Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Bulgaria, Cambodia, Canada, Chili, China, Columbia, Croatia, Czech Republic, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Hungary, India, Indonesia, Ireland, Italy, Japan, Jordan, Kazakhstan, Kenya, Latvia, Malaysia, Mauritius, Mexico, Morocco, Namibia, Nepal, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Kingdom, Uruguay, United Arab Emirates, Vietnam.
- How much
- 455 units (310 units US, 145 OUS)
Class I
Brand Name: Atlan Product Name: Atlan A350 Model/Catalog Number: 8211500 Software Version: Not applicable
Z-0304-2025 · Class I · reported 2024-11-20 · Ongoing
- Why it was recalled
- The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
- Product
- Brand Name: Atlan Product Name: Atlan A350 Model/Catalog Number: 8211500 Software Version: Not applicable Component: Not applicable
- Where it went
- US Nationwide distribution in the states of AL, AZ, CA, FL, GA, IA, IL, KY, MA, ME, ND, NE, NH, NJ, NY, OH, PA, TN, TX, WI, WY. International distribution: Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Bulgaria, Cambodia, Canada, Chili, China, Columbia, Croatia, Czech Republic, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Hungary, India, Indonesia, Ireland, Italy, Japan, Jordan, Kazakhstan, Kenya, Latvia, Malaysia, Mauritius, Mexico, Morocco, Namibia, Nepal, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Kingdom, Uruguay, United Arab Emirates, Vietnam.
- How much
- 1,492 units (139 units US, 1,353 OUS)
Class I
Dr¿ger Perseus A500 Anesthesia Workstation
Z-1512-2024 · Class I · reported 2024-04-24 · Ongoing
- Why it was recalled
- Unexpected shutdown while operating on battery power.
- Product
- Dr¿ger Perseus A500 Anesthesia Workstation
- Where it went
- Domestic: Nationwide Distribution; Foreign: Angola, Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Bolivia, Botswana, Brazil, Brunei Dar-se- S, Bulgaria, Canada, Chili, China, Columbia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, French Polynesia, French Guiana, Georgia, Germany, Gibraltar, Greece, Guadeloupe, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Liechtenstein, Lithuania, Malaysia, Maldives, Mauritius, Mayotte, Mexico, Morocco, Netherlands, New Zealand, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovenia, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Turkmenistan, Uganda, Ukraine, United Kingdom, Uruguay, United Arab Emirates, Vietnam.
- How much
- 16,841 units
Class I
Carina Sub-Acute Care Ventilator
Z-2248-2023 · Class I · reported 2023-08-30 · Ongoing
- Why it was recalled
- Presence of polyether polyurethane (PE-PUR) and emission of 1,3-Dichloropropan-2-ol that exceed the acceptable uptake level during continuous use (>30 days) in pediatric patients.
- Product
- Carina Sub-Acute Care Ventilator
- Where it went
- US Nationwide Distribution.
- How much
- 11,621 units
Class I
Oxylog 3000 Plus emergency and transport ventilator
Z-2018-2023 · Class I · reported 2023-07-19 · Ongoing
- Why it was recalled
- Draeger has become aware of instances where the device stopped working when in use when transitioning from battery to AC operation.
- Product
- Oxylog 3000 Plus emergency and transport ventilator
- Where it went
- Distributed Nationwide. OUS Distribution: Afghanistan, Algeria, Andorra, Angola, Antigua/Barbados, Argentina, Australia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Bolivia, Brunei Dar-es-S, Bulgaria, Burkina-Faso, Cambodia, Canada, Cayman Islands, Chile, China, Christmas Island, Columbia, Congo, Costa Rica, Croatia, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Equatorial Guinea, Estonia, Ethiopia, Faeroe, Finland, France, French Polynesia, French Guiana, Gabon, Georgia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Guinea, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Ivory Coast, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Laos, Latvia, Lebanon, Libya, Liechtenstein, Lithuania, Macau, Malawi, Malaysia, Mali, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldavia, Mongolia, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, North Korea, Norway, Oman, Pakistan, Panama, Pap. New Guinea, Paraguay, Peru, Poland, Portugal, Qatar, Reunion, Romania, Russian Fed., Rwanda, San Marino, Saudi Arabia, Senegal, Serbia, Singapore, Sint Maarten, Slovenia, Slovakia, South...
- How much
- 300 units
Class I
VentStar Basic (P)180, MP00351
Z-1440-2023 · Class I · reported 2023-05-24 · Terminated
- Why it was recalled
- Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
- Product
- VentStar Basic (P)180, MP00351
- Where it went
- Domestic: US Nationwide Distribution.
- How much
- 325 units
Class I
Set2Go Ventilation 12 (A), MP07968
Z-1459-2023 · Class I · reported 2023-05-24 · Terminated
- Why it was recalled
- Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
- Product
- Set2Go Ventilation 12 (A), MP07968
- Where it went
- Domestic: US Nationwide Distribution.
- How much
- 500 units
Class I
Seattle PAP Plus, MP02260
Z-1439-2023 · Class I · reported 2023-05-24 · Terminated
- Why it was recalled
- Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
- Product
- Seattle PAP Plus, MP02260
- Where it went
- Domestic: US Nationwide Distribution.
- How much
- 32,050 untis
Class I
VentStar Anesthesia WT (P)180, MP00374
Z-1444-2023 · Class I · reported 2023-05-24 · Terminated
- Why it was recalled
- Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
- Product
- VentStar Anesthesia WT (P)180, MP00374
- Where it went
- Domestic: US Nationwide Distribution.
- How much
- 130 units
Class I
Anesthesia Set Coax 180, MP03384
Z-1458-2023 · Class I · reported 2023-05-24 · Terminated
- Why it was recalled
- Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
- Product
- Anesthesia Set Coax 180, MP03384
- Where it went
- Domestic: US Nationwide Distribution.
- How much
- 20 units
Class I
VentStar Watertrap (N) 180, MP00363
Z-1449-2023 · Class I · reported 2023-05-24 · Terminated
- Why it was recalled
- Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
- Product
- VentStar Watertrap (N) 180, MP00363
- Where it went
- Domestic: US Nationwide Distribution.
- How much
- 125 units
Class I
VentStar Watertrap (P) 180, MP00361
Z-1442-2023 · Class I · reported 2023-05-24 · Terminated
- Why it was recalled
- Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
- Product
- VentStar Watertrap (P) 180, MP00361
- Where it went
- Domestic: US Nationwide Distribution.
- How much
- 50 units
Class I
VentStar Anesthesia (N) 180, MP00333
Z-1447-2023 · Class I · reported 2023-05-24 · Terminated
- Why it was recalled
- Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
- Product
- VentStar Anesthesia (N) 180, MP00333
- Where it went
- Domestic: US Nationwide Distribution.
- How much
- N/A
Class I
Ventstar Coax, MP03373
Z-1454-2023 · Class I · reported 2023-05-24 · Terminated
- Why it was recalled
- Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
- Product
- Ventstar Coax, MP03373
- Where it went
- Domestic: US Nationwide Distribution.
- How much
- 2,380 units
Class I
ID Circuit Basic (P) 180, MP01340
Z-1446-2023 · Class I · reported 2023-05-24 · Terminated
- Why it was recalled
- Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
- Product
- ID Circuit Basic (P) 180, MP01340
- Where it went
- Domestic: US Nationwide Distribution.
- How much
- 25 units
Class I
VentStar Basic (P)250, MP00352
Z-1441-2023 · Class I · reported 2023-05-24 · Terminated
- Why it was recalled
- Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
- Product
- VentStar Basic (P)250, MP00352
- Where it went
- Domestic: US Nationwide Distribution.
- How much
- 200 units
Class I
Anesthesia Circuit Kit Coax 3, MP02732
Z-1453-2023 · Class I · reported 2023-05-24 · Terminated
- Why it was recalled
- Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
- Product
- Anesthesia Circuit Kit Coax 3, MP02732
- Where it went
- Domestic: US Nationwide Distribution.
- How much
- 329,294 units
Class I
Ventstar Coax 180, MP03374
Z-1455-2023 · Class I · reported 2023-05-24 · Terminated
- Why it was recalled
- Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
- Product
- Ventstar Coax 180, MP03374
- Where it went
- Domestic: US Nationwide Distribution.
- How much
- 18,160 units
Class I
Anesthesia Circuit Kit Coax 1, MP02730
Z-1452-2023 · Class I · reported 2023-05-24 · Terminated
- Why it was recalled
- Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
- Product
- Anesthesia Circuit Kit Coax 1, MP02730
- Where it went
- Domestic: US Nationwide Distribution.
- How much
- 200,900 units
Class I
VentStar Basic (N) 180, MP00353
Z-1448-2023 · Class I · reported 2023-05-24 · Terminated
- Why it was recalled
- Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
- Product
- VentStar Basic (N) 180, MP00353
- Where it went
- Domestic: US Nationwide Distribution.
- How much
- 1,725 units
Class I
VentStar breathing bag Set (P) 110, MP00384
Z-1445-2023 · Class I · reported 2023-05-24 · Terminated
- Why it was recalled
- Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
- Product
- VentStar breathing bag Set (P) 110, MP00384
- Where it went
- Domestic: US Nationwide Distribution.
- How much
- 25 units
Class I
Anesthesia Circuit Kit Coax HEPA, MP17102
Z-1460-2023 · Class I · reported 2023-05-24 · Terminated
- Why it was recalled
- Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
- Product
- Anesthesia Circuit Kit Coax HEPA, MP17102
- Where it went
- Domestic: US Nationwide Distribution.
- How much
- 720 units
Class I
Ventstar Coax 230, MP03376
Z-1457-2023 · Class I · reported 2023-05-24 · Terminated
- Why it was recalled
- Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
- Product
- Ventstar Coax 230, MP03376
- Where it went
- Domestic: US Nationwide Distribution.
- How much
- 40 units
Class I
VentStar Bag Set (N) 110, MP00383
Z-1450-2023 · Class I · reported 2023-05-24 · Terminated
- Why it was recalled
- Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
- Product
- VentStar Bag Set (N) 110, MP00383
- Where it went
- Domestic: US Nationwide Distribution.
- How much
- 375 units
Class I
VentStar Watertrap (P) 180 w/oLL, MP00362
Z-1443-2023 · Class I · reported 2023-05-24 · Terminated
- Why it was recalled
- Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
- Product
- VentStar Watertrap (P) 180 w/oLL, MP00362
- Where it went
- Domestic: US Nationwide Distribution.
- How much
- 25 units
Class I
VentStar Coax (P) 150, MP00379
Z-1451-2023 · Class I · reported 2023-05-24 · Terminated
- Why it was recalled
- Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
- Product
- VentStar Coax (P) 150, MP00379
- Where it went
- Domestic: US Nationwide Distribution.
- How much
- 280 units
Class I
ID Coax 180, MP03375
Z-1456-2023 · Class I · reported 2023-05-24 · Terminated
- Why it was recalled
- Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
- Product
- ID Coax 180, MP03375
- Where it went
- Domestic: US Nationwide Distribution.
- How much
- 20 units
Class I
NovaStar TS, Standard elbow, Reusable, Material Numbers MP01579 (Size S), MP01580 (Size M), and MP01581 (Size
Z-1138-2023 · Class I · reported 2023-03-01 · Ongoing
- Why it was recalled
- The magnets used in the swivel frame/tube set and the headgear clips might impact medical devices (implanted as well as not implanted) and metallic implants.
- Product
- NovaStar TS, Standard elbow, Reusable, Material Numbers MP01579 (Size S), MP01580 (Size M), and MP01581 (Size L)
- Where it went
- US Nationwide Distribution
- How much
- 893 units
Showing the 30 most recent of 78. The rest are searchable at the FDA enforcement report database.
Read these counts with three things in mind
A firm with more recalls usually makes more products. A contract packer runs
hundreds of lines. It will out-recall a company with one product whatever either
does. This is not a safety ranking.
Most recalls here are voluntary. A firm found its own fault and pulled the stock.
That is the system working. It lands in this database looking the same as a recall
the FDA had to push for.
Terminated means the FDA considers the recall finished. It is not a verdict.
It does not mean anyone was harmed.
Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.
← All device recalls