DePuy Mitek, Inc., a Johnson & Johnson

DePuy Mitek, Inc., a Johnson & Johnson Co.: 51 FDA recall reports

Every medical device recall the FDA has published for DePuy Mitek, Inc., a Johnson & Johnson Co.. The fault, and how the agency graded it.

51Reports
0Class I
7Ongoing
2025-03-19Most recent

What the classes mean

The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.

ClassWhat the FDA is saying
Class IThere is a reasonable probability that using this will cause serious harm or death.
Class IIIt may cause temporary or reversible harm. Serious harm is unlikely.
Class IIIIt breaks a rule but is unlikely to hurt anyone. Often a labelling fault.

Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.

By class

ClassReports
Class II51

By year reported

YearReports
201222
20138
20151
20163
20175
20183
20192
20235
20252
Checking whether it affects you

The lot or code number is what matters. Not the product name. A recall covers specific lots. Yours may not be one of them.

Check the code on your package against the report. The FDA keeps the current list at fda.gov/safety/recalls.

The reports

The text below is quoted from the FDA record word for word.

COR Disposable Kit, 8 mm. Cartilage Transplant System.

Z-1314-2025  ·  Class II  ·  reported 2025-03-19  ·  Ongoing

Why it was recalled
Device is missing the pin in the graft loader component.
Product
COR Disposable Kit, 8 mm. Cartilage Transplant System.
Where it went
Domestic: Nationwide Distribution; Foreign: Austria, Belgium, Brazil, Switzerland, Chile, China, Colombia, Czech Republic, Germany, Spain, Finland, France, French Guyana, Greece, Israel, India, Italy, Japan, South Korea, Luxemburg, Mexico, Netherlands, Poland, Portugal & Singapore.
How much
US: 131 units; OUS: 312 units

COR Disposable Kit, 8 mm w/ Perpendicularity. Cartilage Transplant System.

Z-1315-2025  ·  Class II  ·  reported 2025-03-19  ·  Ongoing

Why it was recalled
Device is missing the pin in the graft loader component.
Product
COR Disposable Kit, 8 mm w/ Perpendicularity. Cartilage Transplant System.
Where it went
Domestic: Nationwide Distribution; Foreign: Austria, Belgium, Brazil, Switzerland, Chile, China, Colombia, Czech Republic, Germany, Spain, Finland, France, French Guyana, Greece, Israel, India, Italy, Japan, South Korea, Luxemburg, Mexico, Netherlands, Poland, Portugal & Singapore.
How much
US: 131 units; OUS: 312 units

DePuy Mitek BIOKNOTLESS Plus w/ Panacryl-Indicated for use in soft tissue to bone fixation in association with

Z-1678-2023  ·  Class II  ·  reported 2023-06-14  ·  Ongoing

Why it was recalled
IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS¿ Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS¿ Anchor inserter tip detected during surgery may require removal, potentially causing bone damage and longer surgery time
Product
DePuy Mitek BIOKNOTLESS Plus w/ Panacryl-Indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212725
Where it went
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Canada, Switzerland, China, Czech Republic, Germany, Egypt, Spain, United Kingdom, Hong Kong, Hungary, Ireland, India, Italy, Korea, Netherlands, New Zealand, Slovakia, Thailand.
How much
540 units

DePuy Mitek BIOKNOTLESS Plus w/ OC-Indicated for use in soft tissue to bone fixation in association with adequ

Z-1679-2023  ·  Class II  ·  reported 2023-06-14  ·  Ongoing

Why it was recalled
IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS¿ Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS¿ Anchor inserter tip detected during surgery may require removal, potentially causing bone damage and longer surgery time
Product
DePuy Mitek BIOKNOTLESS Plus w/ OC-Indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212726
Where it went
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Canada, Switzerland, China, Czech Republic, Germany, Egypt, Spain, United Kingdom, Hong Kong, Hungary, Ireland, India, Italy, Korea, Netherlands, New Zealand, Slovakia, Thailand.
How much
2372 units

DePuy Mitek BIOKNOTLESS Rapid W/ OC-Indicated for use in soft tissue to bone fixation in association with adeq

Z-1680-2023  ·  Class II  ·  reported 2023-06-14  ·  Ongoing

Why it was recalled
IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS¿ Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS¿ Anchor inserter tip detected during surgery may require removal, potentially causing bone damage and longer surgery time
Product
DePuy Mitek BIOKNOTLESS Rapid W/ OC-Indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212722
Where it went
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Canada, Switzerland, China, Czech Republic, Germany, Egypt, Spain, United Kingdom, Hong Kong, Hungary, Ireland, India, Italy, Korea, Netherlands, New Zealand, Slovakia, Thailand.
How much
3328 units

DePuy Mitek BIOKNOTLESS Plus w/ Ethibond-Iindicated for use in soft tissue to bone fixation in association wit

Z-1677-2023  ·  Class II  ·  reported 2023-06-14  ·  Ongoing

Why it was recalled
IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS¿ Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS¿ Anchor inserter tip detected during surgery may require removal, potentially causing bone damage and longer surgery time
Product
DePuy Mitek BIOKNOTLESS Plus w/ Ethibond-Iindicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212724
Where it went
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Canada, Switzerland, China, Czech Republic, Germany, Egypt, Spain, United Kingdom, Hong Kong, Hungary, Ireland, India, Italy, Korea, Netherlands, New Zealand, Slovakia, Thailand.
How much
805 units

DePuy Mitek BIOKNOTLESS Rapid w/ Pan-Indicated for use in soft tissue to bone fixation in association with ade

Z-1681-2023  ·  Class II  ·  reported 2023-06-14  ·  Ongoing

Why it was recalled
IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS¿ Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS¿ Anchor inserter tip detected during surgery may require removal, potentially causing bone damage and longer surgery time
Product
DePuy Mitek BIOKNOTLESS Rapid w/ Pan-Indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212723
Where it went
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Canada, Switzerland, China, Czech Republic, Germany, Egypt, Spain, United Kingdom, Hong Kong, Hungary, Ireland, India, Italy, Korea, Netherlands, New Zealand, Slovakia, Thailand.
How much
1125 units

Healix Knotless ADV BR 4.75 Suture Anchor

Z-0665-2020  ·  Class II  ·  reported 2019-12-18  ·  Terminated

Why it was recalled
Specific product codes and lots of the Mitek HEALIX ADVANCE" Knotless Anchors were manufactured with BIOCRYL" Biocomposite Material instead of BIOCRYL RAPIDE¿ Biocomposite Material. BIOCRYL and BIOCRYL RAPIDE are biocompatible and effective material choices for absorbable, injection-molded sports medicine implant fixation.
Product
Healix Knotless ADV BR 4.75 Suture Anchor
Where it went
US: AL AZ CA CT FL IA IL IN LA MA MD ME MI MN MT NC NE NH NJ NM OH OK OR PA TN TX VA WA WI International: Belgium, Chile, China, Czech Republic, France, India, Japan, Korea, Latvia, New Zealand, Slovakia Sweden
How much
299

Healix Knotless ADV BR 5.5 Suture Anchor

Z-0666-2020  ·  Class II  ·  reported 2019-12-18  ·  Terminated

Why it was recalled
Specific product codes and lots of the Mitek HEALIX ADVANCE" Knotless Anchors were manufactured with BIOCRYL" Biocomposite Material instead of BIOCRYL RAPIDE¿ Biocomposite Material. BIOCRYL and BIOCRYL RAPIDE are biocompatible and effective material choices for absorbable, injection-molded sports medicine implant fixation.
Product
Healix Knotless ADV BR 5.5 Suture Anchor
Where it went
US: AL AZ CA CT FL IA IL IN LA MA MD ME MI MN MT NC NE NH NJ NM OH OK OR PA TN TX VA WA WI International: Belgium, Chile, China, Czech Republic, France, India, Japan, Korea, Latvia, New Zealand, Slovakia Sweden
How much
599

Mitek Screw and Washer Depth Gauge, Product Code 219024

Z-0208-2019  ·  Class II  ·  reported 2018-10-31  ·  Terminated

Why it was recalled
The device has an offset in the design that results in a reading approximately 3.2mm shorter than the actual length of the bone hole.
Product
Mitek Screw and Washer Depth Gauge, Product Code 219024
Where it went
US nationwide in the states of AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, KS, KY, LA, MA, MI, MO, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, TN, TX, UT, VA, VT, WA, WI. The products were distributed to the following foreign countries: Australia, Canada, Chile, China, Chechnya, Denmark, Italy, Japan, Jordan, Netherlands, Poland, Slovenia, Sweden, Turkey, UK.
How much
401

LATARJET EXPERIENCE Sharp Curved Ostetome Product Usage: The Latarjet EXPERIENCE Instability Shoulder Set

Z-1749-2018  ·  Class II  ·  reported 2018-05-16  ·  Terminated

Why it was recalled
Reports of product loosening or coming unscrewed from the handle.
Product
LATARJET EXPERIENCE Sharp Curved Ostetome Product Usage: The Latarjet EXPERIENCE Instability Shoulder Set is intended to provide to the orthopedic surgeon with a means of bone fixation and to assist in the management of reconstructive surgeries.
Where it went
US Nationwide Distribution in the states of OH, WA, CO
How much
84

1113 Gravity Tube Set Pk/24 -SH Product Usage: The FMS Gravity Tube Set is intended to deliver irrigation

Z-0695-2018  ·  Class II  ·  reported 2018-03-07  ·  Terminated

Why it was recalled
Leakage of FMS gravity tub sets (281113) at the bulb and tube connection
Product
1113 Gravity Tube Set Pk/24 -SH Product Usage: The FMS Gravity Tube Set is intended to deliver irrigation fluid from irrigation containers to the operative site during arthroscopic procedures.
Where it went
US Nationwide Distribution
How much
15,678

DePuy Mitek LATARJET EXPERIENCE Extraction Driver Product Code: 288216 Product Usage: The Latarjet Cortica

Z-2619-2017  ·  Class II  ·  reported 2017-06-28  ·  Terminated

Why it was recalled
Combo Screw Driver (Product Code 288211) tip has the increased potential to break intra-operatively when being used at an angle off-axis to screw
Product
DePuy Mitek LATARJET EXPERIENCE Extraction Driver Product Code: 288216 Product Usage: The Latarjet Cortical Screw set is intended to treat recurrent shoulder instability by supporting the anteroinferior glenoid with a bony graft
Where it went
US: AZ, CA, CO, GA, MA, MI, OH, WA Foreign: Austria, Australia, Belgium, France, Germany, Spain, Switzerland, Netherlands, Poland, UK
How much
115 units

DePuy Mitek LATARJET EXPERIENCE Coracoid Drill Product Code: 288201 Product Usage: The Latarjet Cortical Sc

Z-2616-2017  ·  Class II  ·  reported 2017-06-28  ·  Terminated

Why it was recalled
Combo Screw Driver (Product Code 288211) tip has the increased potential to break intra-operatively when being used at an angle off-axis to screw
Product
DePuy Mitek LATARJET EXPERIENCE Coracoid Drill Product Code: 288201 Product Usage: The Latarjet Cortical Screw set is intended to treat recurrent shoulder instability by supporting the anteroinferior glenoid with a bony graft
Where it went
US: AZ, CA, CO, GA, MA, MI, OH, WA Foreign: Austria, Australia, Belgium, France, Germany, Spain, Switzerland, Netherlands, Poland, UK
How much
184 units

DePuy Mitek LATARJET EXPERIENCE Top Hat Tap Product Code: 288203 Product Usage: The Latarjet Cortical Scre

Z-2618-2017  ·  Class II  ·  reported 2017-06-28  ·  Terminated

Why it was recalled
Combo Screw Driver (Product Code 288211) tip has the increased potential to break intra-operatively when being used at an angle off-axis to screw
Product
DePuy Mitek LATARJET EXPERIENCE Top Hat Tap Product Code: 288203 Product Usage: The Latarjet Cortical Screw set is intended to treat recurrent shoulder instability by supporting the anteroinferior glenoid with a bony graft
Where it went
US: AZ, CA, CO, GA, MA, MI, OH, WA Foreign: Austria, Australia, Belgium, France, Germany, Spain, Switzerland, Netherlands, Poland, UK
How much
153 units

DePuy Mitek LATARJET EXPERIENCE-Combo Screw Driver Product Code: 288211 Product Usage: The Latarjet Corti

Z-2615-2017  ·  Class II  ·  reported 2017-06-28  ·  Terminated

Why it was recalled
Combo Screw Driver (Product Code 288211) tip has the increased potential to break intra-operatively when being used at an angle off-axis to screw
Product
DePuy Mitek LATARJET EXPERIENCE-Combo Screw Driver Product Code: 288211 Product Usage: The Latarjet Cortical Screw set is intended to treat recurrent shoulder instability by supporting the anteroinferior glenoid with a bony graft
Where it went
US: AZ, CA, CO, GA, MA, MI, OH, WA Foreign: Austria, Australia, Belgium, France, Germany, Spain, Switzerland, Netherlands, Poland, UK
How much
173 units

DePuy Mitek LATARJET EXPERIENCE Coracoid Top Hat Drill Product Code: 288202 Product Usage: The Latarjet Cor

Z-2617-2017  ·  Class II  ·  reported 2017-06-28  ·  Terminated

Why it was recalled
Combo Screw Driver (Product Code 288211) tip has the increased potential to break intra-operatively when being used at an angle off-axis to screw
Product
DePuy Mitek LATARJET EXPERIENCE Coracoid Top Hat Drill Product Code: 288202 Product Usage: The Latarjet Cortical Screw set is intended to treat recurrent shoulder instability by supporting the anteroinferior glenoid with a bony graft
Where it went
US: AZ, CA, CO, GA, MA, MI, OH, WA Foreign: Austria, Australia, Belgium, France, Germany, Spain, Switzerland, Netherlands, Poland, UK
How much
184 units

FMS neXtra¿ 3.5mm Aggressive Cutter Shaver Blades; Product Code: 283315 Intended to provide controlled cutt

Z-0657-2016  ·  Class II  ·  reported 2016-01-27  ·  Terminated

Why it was recalled
Product is incorrectly labeled. The blade configuration in the affected lots does not match the description on the label.
Product
FMS neXtra¿ 3.5mm Aggressive Cutter Shaver Blades; Product Code: 283315 Intended to provide controlled cutting, shaving, and abrading of tissue during orthopedic procedures.
Where it went
Worldwide Distribution. US states of TX, OH, MI, and ME; and the countries of FR, DE, SK, IT, NL, CH, and CZ.
How much
45 units

FMS 3.5mm Ultra-Aggressive Cutter Shaver Blades; Product Code: 287325 Intended to provide controlled cuttin

Z-0658-2016  ·  Class II  ·  reported 2016-01-27  ·  Terminated

Why it was recalled
Product is incorrectly labeled. The blade configuration in the affected lots does not match the description on the label.
Product
FMS 3.5mm Ultra-Aggressive Cutter Shaver Blades; Product Code: 287325 Intended to provide controlled cutting, shaving, and abrading of tissue during orthopedic procedures.
Where it went
Worldwide Distribution. US states of TX, OH, MI, and ME; and the countries of FR, DE, SK, IT, NL, CH, and CZ.
How much
115 units

FMS neXtra¿ 3.5mm Full Radius Cutter Shaver Blades; Product Code: 283305 Intended to provide controlled cu

Z-0656-2016  ·  Class II  ·  reported 2016-01-27  ·  Terminated

Why it was recalled
Product is incorrectly labeled. The blade configuration in the affected lots does not match the description on the label.
Product
FMS neXtra¿ 3.5mm Full Radius Cutter Shaver Blades; Product Code: 283305 Intended to provide controlled cutting, shaving, and abrading of tissue during orthopedic procedures.
Where it went
Worldwide Distribution. US states of TX, OH, MI, and ME; and the countries of FR, DE, SK, IT, NL, CH, and CZ.
How much
50 units

VAPR¿ TRIPOLAR 90" Degree Suction Electrodes (Model/Catalog No: 225028)

Z-0479-2016  ·  Class II  ·  reported 2015-12-30  ·  Terminated

Why it was recalled
Mitek identified that on VAPR Tripolar 90 Degree Suction Electrode the ablation and coagulation buttons are colored incorrectly. The wiring for each button function was confirmed as correct and the non-conformance impacted button color only
Product
VAPR¿ TRIPOLAR 90" Degree Suction Electrodes (Model/Catalog No: 225028)
Where it went
Worldwide Distribution - US Distribution to the states of : AL, CA, CT, FL, KY, MN, MS, ND, NJ, NM and TX., and to the countries of: France, Netherlands and Norway.
How much
127

DePuy Mitek FMS Intermediary Tubing with One-Way valve Product Code: 281142 Product Usage: The Intermedi

Z-0191-2014  ·  Class II  ·  reported 2013-11-20  ·  Terminated

Why it was recalled
Pillow valve included in the FMS Duo outflow tubing set and the FMS Solo intermediary tubing set may not perform as intended. may result in backflow of irrigation fluid into the One Day Set which may potentially lead to patient cross-contamination
Product
DePuy Mitek FMS Intermediary Tubing with One-Way valve Product Code: 281142 Product Usage: The Intermediary Tubing for FMS¿ Fluid Management Systems connects the Inflow Tubing to the arthroscopic sheath during arthroscopic procedures. The Polyvinyl Chloride (PVC) tubing integrates several functional components including: luer locks for attaching the tube to the Inflow tubing and sheath, clamps, and a one-way valve. The FMS Intermediary Tubing is sterile and intended for single use with the FMS Fluid Management Systems in a surgical setting by personnel trained in arthroscopy .
Where it went
Worldwide Distribution - USA (Nationwide) and in the countries of: Australia , Austria , Belgium , Canada , Czech Republic , Denmark , Finland , France, Germany , India, Ireland , Israel, Italy , Latvia , Netherlands , Norway , Nepal, Poland , Portugal , Spain , Sweden , Switzerland , United Kingdom.
How much
368 pk

DePuy Mitek FMS Outflow Tubing with One-Way valve Product Code:284649 Product Usage: The Intermediary Tub

Z-0192-2014  ·  Class II  ·  reported 2013-11-20  ·  Terminated

Why it was recalled
Pillow valve included in the FMS Duo outflow tubing set and the FMS Solo intermediary tubing set may not perform as intended. may result in backflow of irrigation fluid into the One Day Set which may potentially lead to patient cross-contamination
Product
DePuy Mitek FMS Outflow Tubing with One-Way valve Product Code:284649 Product Usage: The Intermediary Tubing for FMS¿ Fluid Management Systems connects the Inflow Tubing to the arthroscopic sheath during arthroscopic procedures. The Polyvinyl Chloride (PVC) tubing integrates several functional components including: luer locks for attaching the tube to the Inflow tubing and sheath, clamps, and a one-way valve. The FMS Outflow Tubing with One-Way valve is sterile and intended for single use with the FMS Fluid Management Systems in a surgical setting by personnel trained in arthroscopy.
Where it went
Worldwide Distribution - USA (Nationwide) and in the countries of: Australia , Austria , Belgium , Canada , Czech Republic , Denmark , Finland , France, Germany , India, Ireland , Israel, Italy , Latvia , Netherlands , Norway , Nepal, Poland , Portugal , Spain , Sweden , Switzerland , United Kingdom.
How much
29280 units

DePuy Mitek Tissue Liberator -Blade Up Catalog Number: 214623 The Reusable Arthroscopy Instruments are re

Z-1278-2013  ·  Class II  ·  reported 2013-05-15  ·  Terminated

Why it was recalled
Labeled incorrectly as a Blade Up configuration instead of a Blade Down configuration.
Product
DePuy Mitek Tissue Liberator -Blade Up Catalog Number: 214623 The Reusable Arthroscopy Instruments are reusable, non-powered surgical instruments used during arthroscopic surgery of any joint except the spine.
Where it went
Worldwide Distribution - USA including KY, NJ and Internationally to Argnetina, Australia, Belgium,Columbia, India, France, Japan, Jordan, Latvija, Israel, Korea, and Great Britain. For questions regarding this recall call 508-977-3813.
How much
27 units

ORTHOCORD One Violet and One Blue Braided Composite Suture with MO-6 Taper Needles, Product Code: 223115.

Z-1175-2013  ·  Class II  ·  reported 2013-05-01  ·  Terminated

Why it was recalled
The IFU provided with the free strand ORTHOCORD sutures currently indicates that approximately 100% suture strength remains at six weeks after implantation. While this statement is true for Violet ORTHOCORD sutures, it is an inaccurate statement for Blue ORTHOCORD free strand sutures. Based on in-vivo strength retention study, a Blue ORTHOCORD sutures maintain approximately 80% of strength at six
Product
ORTHOCORD One Violet and One Blue Braided Composite Suture with MO-6 Taper Needles, Product Code: 223115. For use in general soft tissue approximation, and/or ligation, including orthopedic procedures.
Where it went
Nationwide Distribution.
How much
46732

ORTHOCORD BLUE BRAIDED COMPOSITE SUTURE 2 (5 METRIC), Catalog Number: 223111. For use in general soft tissu

Z-1172-2013  ·  Class II  ·  reported 2013-05-01  ·  Terminated

Why it was recalled
The IFU provided with the free strand ORTHOCORD sutures currently indicates that approximately 100% suture strength remains at six weeks after implantation. While this statement is true for Violet ORTHOCORD sutures, it is an inaccurate statement for Blue ORTHOCORD free strand sutures. Based on in-vivo strength retention study, a Blue ORTHOCORD sutures maintain approximately 80% of strength at six
Product
ORTHOCORD BLUE BRAIDED COMPOSITE SUTURE 2 (5 METRIC), Catalog Number: 223111. For use in general soft tissue approximation, and/or ligation, including orthopedic procedures.
Where it went
Nationwide Distribution.
How much
65062

ORTHOCORD One Violet and One Blue Braided Composite Suture with OS-6 Reverse Cutting Needles, Product Code: 2

Z-1176-2013  ·  Class II  ·  reported 2013-05-01  ·  Terminated

Why it was recalled
The IFU provided with the free strand ORTHOCORD sutures currently indicates that approximately 100% suture strength remains at six weeks after implantation. While this statement is true for Violet ORTHOCORD sutures, it is an inaccurate statement for Blue ORTHOCORD free strand sutures. Based on in-vivo strength retention study, a Blue ORTHOCORD sutures maintain approximately 80% of strength at six
Product
ORTHOCORD One Violet and One Blue Braided Composite Suture with OS-6 Reverse Cutting Needles, Product Code: 223116. For use in general soft tissue approximation, and/or ligation, including orthopedic procedures.
Where it went
Nationwide Distribution.
How much
46501

ORTHOCORD ONE VIOLET AND ONE BLUE BRAIDED COMPOSITE SUTURE 2 (5 METRIC), Catalog Number: 223113. For use i

Z-1173-2013  ·  Class II  ·  reported 2013-05-01  ·  Terminated

Why it was recalled
The IFU provided with the free strand ORTHOCORD sutures currently indicates that approximately 100% suture strength remains at six weeks after implantation. While this statement is true for Violet ORTHOCORD sutures, it is an inaccurate statement for Blue ORTHOCORD free strand sutures. Based on in-vivo strength retention study, a Blue ORTHOCORD sutures maintain approximately 80% of strength at six
Product
ORTHOCORD ONE VIOLET AND ONE BLUE BRAIDED COMPOSITE SUTURE 2 (5 METRIC), Catalog Number: 223113. For use in general soft tissue approximation, and/or ligation, including orthopedic procedures.
Where it went
Nationwide Distribution.
How much
84174

ORTHOCORD One Violet and One Blue Braided Composite Suture with MO-7 Taper Needles, Catalog Number: 223114.

Z-1174-2013  ·  Class II  ·  reported 2013-05-01  ·  Terminated

Why it was recalled
The IFU provided with the free strand ORTHOCORD sutures currently indicates that approximately 100% suture strength remains at six weeks after implantation. While this statement is true for Violet ORTHOCORD sutures, it is an inaccurate statement for Blue ORTHOCORD free strand sutures. Based on in-vivo strength retention study, a Blue ORTHOCORD sutures maintain approximately 80% of strength at six
Product
ORTHOCORD One Violet and One Blue Braided Composite Suture with MO-7 Taper Needles, Catalog Number: 223114. For use in general soft tissue approximation, and/or ligation, including orthopedic procedures.
Where it went
Nationwide Distribution.
How much
206540

DePuy Mitek 11 MM Fully Fluted Reamer, Sterile Product Number: 232423 DePuy Mitek Sterile Reamers are inte

Z-0529-2013  ·  Class II  ·  reported 2012-12-26  ·  Terminated

Why it was recalled
Specific units of the DePuy Mitek Sterile Reamers are not labeled correctly, 7 units were labeled as 11MM while the box actually contained a 10 MM reamer
Product
DePuy Mitek 11 MM Fully Fluted Reamer, Sterile Product Number: 232423 DePuy Mitek Sterile Reamers are intended to cut through cortical and cancellous bone during Cruciate Ligament Repair/Reconstruction of the knee.
Where it went
Nationwide Distribution including IL and NJ
How much
20

Showing the 30 most recent of 51. The rest are searchable at the FDA enforcement report database.

Read these counts with three things in mind

A firm with more recalls usually makes more products. A contract packer runs hundreds of lines. It will out-recall a company with one product whatever either does. This is not a safety ranking.

Most recalls here are voluntary. A firm found its own fault and pulled the stock. That is the system working. It lands in this database looking the same as a recall the FDA had to push for.

Terminated means the FDA considers the recall finished. It is not a verdict. It does not mean anyone was harmed.

Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.

← All device recalls