DePuy Mitek, Inc., a Johnson & Johnson Co.: 51 FDA recall reports
Every medical device recall the FDA has published for DePuy Mitek, Inc., a Johnson & Johnson Co.. The fault, and how the agency graded it.
51Reports
0Class I
7Ongoing
2025-03-19Most recent
What the classes mean
The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.
Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.
By class
By year reported
Checking whether it affects you
The lot or code number is what matters. Not the product name. A recall covers
specific lots. Yours may not be one of them.
Check the code on your package against the report. The FDA keeps the current
list at fda.gov/safety/recalls.
The reports
The text below is quoted from the FDA record word for word.
COR Disposable Kit, 8 mm. Cartilage Transplant System.
Z-1314-2025 · Class II · reported 2025-03-19 · Ongoing
- Why it was recalled
- Device is missing the pin in the graft loader component.
- Product
- COR Disposable Kit, 8 mm. Cartilage Transplant System.
- Where it went
- Domestic: Nationwide Distribution; Foreign: Austria, Belgium, Brazil, Switzerland, Chile, China, Colombia, Czech Republic, Germany, Spain, Finland, France, French Guyana, Greece, Israel, India, Italy, Japan, South Korea, Luxemburg, Mexico, Netherlands, Poland, Portugal & Singapore.
- How much
- US: 131 units; OUS: 312 units
COR Disposable Kit, 8 mm w/ Perpendicularity. Cartilage Transplant System.
Z-1315-2025 · Class II · reported 2025-03-19 · Ongoing
- Why it was recalled
- Device is missing the pin in the graft loader component.
- Product
- COR Disposable Kit, 8 mm w/ Perpendicularity. Cartilage Transplant System.
- Where it went
- Domestic: Nationwide Distribution; Foreign: Austria, Belgium, Brazil, Switzerland, Chile, China, Colombia, Czech Republic, Germany, Spain, Finland, France, French Guyana, Greece, Israel, India, Italy, Japan, South Korea, Luxemburg, Mexico, Netherlands, Poland, Portugal & Singapore.
- How much
- US: 131 units; OUS: 312 units
DePuy Mitek BIOKNOTLESS Plus w/ Panacryl-Indicated for use in soft tissue to bone fixation in association with
Z-1678-2023 · Class II · reported 2023-06-14 · Ongoing
- Why it was recalled
- IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS¿ Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS¿ Anchor inserter tip detected during surgery may require removal, potentially causing bone damage and longer surgery time
- Product
- DePuy Mitek BIOKNOTLESS Plus w/ Panacryl-Indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212725
- Where it went
- Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Canada, Switzerland, China, Czech Republic, Germany, Egypt, Spain, United Kingdom, Hong Kong, Hungary, Ireland, India, Italy, Korea, Netherlands, New Zealand, Slovakia, Thailand.
- How much
- 540 units
DePuy Mitek BIOKNOTLESS Plus w/ OC-Indicated for use in soft tissue to bone fixation in association with adequ
Z-1679-2023 · Class II · reported 2023-06-14 · Ongoing
- Why it was recalled
- IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS¿ Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS¿ Anchor inserter tip detected during surgery may require removal, potentially causing bone damage and longer surgery time
- Product
- DePuy Mitek BIOKNOTLESS Plus w/ OC-Indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212726
- Where it went
- Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Canada, Switzerland, China, Czech Republic, Germany, Egypt, Spain, United Kingdom, Hong Kong, Hungary, Ireland, India, Italy, Korea, Netherlands, New Zealand, Slovakia, Thailand.
- How much
- 2372 units
DePuy Mitek BIOKNOTLESS Rapid W/ OC-Indicated for use in soft tissue to bone fixation in association with adeq
Z-1680-2023 · Class II · reported 2023-06-14 · Ongoing
- Why it was recalled
- IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS¿ Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS¿ Anchor inserter tip detected during surgery may require removal, potentially causing bone damage and longer surgery time
- Product
- DePuy Mitek BIOKNOTLESS Rapid W/ OC-Indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212722
- Where it went
- Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Canada, Switzerland, China, Czech Republic, Germany, Egypt, Spain, United Kingdom, Hong Kong, Hungary, Ireland, India, Italy, Korea, Netherlands, New Zealand, Slovakia, Thailand.
- How much
- 3328 units
DePuy Mitek BIOKNOTLESS Plus w/ Ethibond-Iindicated for use in soft tissue to bone fixation in association wit
Z-1677-2023 · Class II · reported 2023-06-14 · Ongoing
- Why it was recalled
- IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS¿ Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS¿ Anchor inserter tip detected during surgery may require removal, potentially causing bone damage and longer surgery time
- Product
- DePuy Mitek BIOKNOTLESS Plus w/ Ethibond-Iindicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212724
- Where it went
- Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Canada, Switzerland, China, Czech Republic, Germany, Egypt, Spain, United Kingdom, Hong Kong, Hungary, Ireland, India, Italy, Korea, Netherlands, New Zealand, Slovakia, Thailand.
- How much
- 805 units
DePuy Mitek BIOKNOTLESS Rapid w/ Pan-Indicated for use in soft tissue to bone fixation in association with ade
Z-1681-2023 · Class II · reported 2023-06-14 · Ongoing
- Why it was recalled
- IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS¿ Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS¿ Anchor inserter tip detected during surgery may require removal, potentially causing bone damage and longer surgery time
- Product
- DePuy Mitek BIOKNOTLESS Rapid w/ Pan-Indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212723
- Where it went
- Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Canada, Switzerland, China, Czech Republic, Germany, Egypt, Spain, United Kingdom, Hong Kong, Hungary, Ireland, India, Italy, Korea, Netherlands, New Zealand, Slovakia, Thailand.
- How much
- 1125 units
Healix Knotless ADV BR 4.75 Suture Anchor
Z-0665-2020 · Class II · reported 2019-12-18 · Terminated
- Why it was recalled
- Specific product codes and lots of the Mitek HEALIX ADVANCE" Knotless Anchors were manufactured with BIOCRYL" Biocomposite Material instead of BIOCRYL RAPIDE¿ Biocomposite Material. BIOCRYL and BIOCRYL RAPIDE are biocompatible and effective material choices for absorbable, injection-molded sports medicine implant fixation.
- Product
- Healix Knotless ADV BR 4.75 Suture Anchor
- Where it went
- US: AL AZ CA CT FL IA IL IN LA MA MD ME MI MN MT NC NE NH NJ NM OH OK OR PA TN TX VA WA WI International: Belgium, Chile, China, Czech Republic, France, India, Japan, Korea, Latvia, New Zealand, Slovakia Sweden
- How much
- 299
Healix Knotless ADV BR 5.5 Suture Anchor
Z-0666-2020 · Class II · reported 2019-12-18 · Terminated
- Why it was recalled
- Specific product codes and lots of the Mitek HEALIX ADVANCE" Knotless Anchors were manufactured with BIOCRYL" Biocomposite Material instead of BIOCRYL RAPIDE¿ Biocomposite Material. BIOCRYL and BIOCRYL RAPIDE are biocompatible and effective material choices for absorbable, injection-molded sports medicine implant fixation.
- Product
- Healix Knotless ADV BR 5.5 Suture Anchor
- Where it went
- US: AL AZ CA CT FL IA IL IN LA MA MD ME MI MN MT NC NE NH NJ NM OH OK OR PA TN TX VA WA WI International: Belgium, Chile, China, Czech Republic, France, India, Japan, Korea, Latvia, New Zealand, Slovakia Sweden
- How much
- 599
Mitek Screw and Washer Depth Gauge, Product Code 219024
Z-0208-2019 · Class II · reported 2018-10-31 · Terminated
- Why it was recalled
- The device has an offset in the design that results in a reading approximately 3.2mm shorter than the actual length of the bone hole.
- Product
- Mitek Screw and Washer Depth Gauge, Product Code 219024
- Where it went
- US nationwide in the states of AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, KS, KY, LA, MA, MI, MO, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, TN, TX, UT, VA, VT, WA, WI. The products were distributed to the following foreign countries: Australia, Canada, Chile, China, Chechnya, Denmark, Italy, Japan, Jordan, Netherlands, Poland, Slovenia, Sweden, Turkey, UK.
- How much
- 401
LATARJET EXPERIENCE Sharp Curved Ostetome Product Usage: The Latarjet EXPERIENCE Instability Shoulder Set
Z-1749-2018 · Class II · reported 2018-05-16 · Terminated
- Why it was recalled
- Reports of product loosening or coming unscrewed from the handle.
- Product
- LATARJET EXPERIENCE Sharp Curved Ostetome Product Usage: The Latarjet EXPERIENCE Instability Shoulder Set is intended to provide to the orthopedic surgeon with a means of bone fixation and to assist in the management of reconstructive surgeries.
- Where it went
- US Nationwide Distribution in the states of OH, WA, CO
- How much
- 84
1113 Gravity Tube Set Pk/24 -SH Product Usage: The FMS Gravity Tube Set is intended to deliver irrigation
Z-0695-2018 · Class II · reported 2018-03-07 · Terminated
- Why it was recalled
- Leakage of FMS gravity tub sets (281113) at the bulb and tube connection
- Product
- 1113 Gravity Tube Set Pk/24 -SH Product Usage: The FMS Gravity Tube Set is intended to deliver irrigation fluid from irrigation containers to the operative site during arthroscopic procedures.
- Where it went
- US Nationwide Distribution
- How much
- 15,678
DePuy Mitek LATARJET EXPERIENCE Extraction Driver Product Code: 288216 Product Usage: The Latarjet Cortica
Z-2619-2017 · Class II · reported 2017-06-28 · Terminated
- Why it was recalled
- Combo Screw Driver (Product Code 288211) tip has the increased potential to break intra-operatively when being used at an angle off-axis to screw
- Product
- DePuy Mitek LATARJET EXPERIENCE Extraction Driver Product Code: 288216 Product Usage: The Latarjet Cortical Screw set is intended to treat recurrent shoulder instability by supporting the anteroinferior glenoid with a bony graft
- Where it went
- US: AZ, CA, CO, GA, MA, MI, OH, WA Foreign: Austria, Australia, Belgium, France, Germany, Spain, Switzerland, Netherlands, Poland, UK
- How much
- 115 units
DePuy Mitek LATARJET EXPERIENCE Coracoid Drill Product Code: 288201 Product Usage: The Latarjet Cortical Sc
Z-2616-2017 · Class II · reported 2017-06-28 · Terminated
- Why it was recalled
- Combo Screw Driver (Product Code 288211) tip has the increased potential to break intra-operatively when being used at an angle off-axis to screw
- Product
- DePuy Mitek LATARJET EXPERIENCE Coracoid Drill Product Code: 288201 Product Usage: The Latarjet Cortical Screw set is intended to treat recurrent shoulder instability by supporting the anteroinferior glenoid with a bony graft
- Where it went
- US: AZ, CA, CO, GA, MA, MI, OH, WA Foreign: Austria, Australia, Belgium, France, Germany, Spain, Switzerland, Netherlands, Poland, UK
- How much
- 184 units
DePuy Mitek LATARJET EXPERIENCE Top Hat Tap Product Code: 288203 Product Usage: The Latarjet Cortical Scre
Z-2618-2017 · Class II · reported 2017-06-28 · Terminated
- Why it was recalled
- Combo Screw Driver (Product Code 288211) tip has the increased potential to break intra-operatively when being used at an angle off-axis to screw
- Product
- DePuy Mitek LATARJET EXPERIENCE Top Hat Tap Product Code: 288203 Product Usage: The Latarjet Cortical Screw set is intended to treat recurrent shoulder instability by supporting the anteroinferior glenoid with a bony graft
- Where it went
- US: AZ, CA, CO, GA, MA, MI, OH, WA Foreign: Austria, Australia, Belgium, France, Germany, Spain, Switzerland, Netherlands, Poland, UK
- How much
- 153 units
DePuy Mitek LATARJET EXPERIENCE-Combo Screw Driver Product Code: 288211 Product Usage: The Latarjet Corti
Z-2615-2017 · Class II · reported 2017-06-28 · Terminated
- Why it was recalled
- Combo Screw Driver (Product Code 288211) tip has the increased potential to break intra-operatively when being used at an angle off-axis to screw
- Product
- DePuy Mitek LATARJET EXPERIENCE-Combo Screw Driver Product Code: 288211 Product Usage: The Latarjet Cortical Screw set is intended to treat recurrent shoulder instability by supporting the anteroinferior glenoid with a bony graft
- Where it went
- US: AZ, CA, CO, GA, MA, MI, OH, WA Foreign: Austria, Australia, Belgium, France, Germany, Spain, Switzerland, Netherlands, Poland, UK
- How much
- 173 units
DePuy Mitek LATARJET EXPERIENCE Coracoid Top Hat Drill Product Code: 288202 Product Usage: The Latarjet Cor
Z-2617-2017 · Class II · reported 2017-06-28 · Terminated
- Why it was recalled
- Combo Screw Driver (Product Code 288211) tip has the increased potential to break intra-operatively when being used at an angle off-axis to screw
- Product
- DePuy Mitek LATARJET EXPERIENCE Coracoid Top Hat Drill Product Code: 288202 Product Usage: The Latarjet Cortical Screw set is intended to treat recurrent shoulder instability by supporting the anteroinferior glenoid with a bony graft
- Where it went
- US: AZ, CA, CO, GA, MA, MI, OH, WA Foreign: Austria, Australia, Belgium, France, Germany, Spain, Switzerland, Netherlands, Poland, UK
- How much
- 184 units
FMS neXtra¿ 3.5mm Aggressive Cutter Shaver Blades; Product Code: 283315 Intended to provide controlled cutt
Z-0657-2016 · Class II · reported 2016-01-27 · Terminated
- Why it was recalled
- Product is incorrectly labeled. The blade configuration in the affected lots does not match the description on the label.
- Product
- FMS neXtra¿ 3.5mm Aggressive Cutter Shaver Blades; Product Code: 283315 Intended to provide controlled cutting, shaving, and abrading of tissue during orthopedic procedures.
- Where it went
- Worldwide Distribution. US states of TX, OH, MI, and ME; and the countries of FR, DE, SK, IT, NL, CH, and CZ.
- How much
- 45 units
FMS 3.5mm Ultra-Aggressive Cutter Shaver Blades; Product Code: 287325 Intended to provide controlled cuttin
Z-0658-2016 · Class II · reported 2016-01-27 · Terminated
- Why it was recalled
- Product is incorrectly labeled. The blade configuration in the affected lots does not match the description on the label.
- Product
- FMS 3.5mm Ultra-Aggressive Cutter Shaver Blades; Product Code: 287325 Intended to provide controlled cutting, shaving, and abrading of tissue during orthopedic procedures.
- Where it went
- Worldwide Distribution. US states of TX, OH, MI, and ME; and the countries of FR, DE, SK, IT, NL, CH, and CZ.
- How much
- 115 units
FMS neXtra¿ 3.5mm Full Radius Cutter Shaver Blades; Product Code: 283305 Intended to provide controlled cu
Z-0656-2016 · Class II · reported 2016-01-27 · Terminated
- Why it was recalled
- Product is incorrectly labeled. The blade configuration in the affected lots does not match the description on the label.
- Product
- FMS neXtra¿ 3.5mm Full Radius Cutter Shaver Blades; Product Code: 283305 Intended to provide controlled cutting, shaving, and abrading of tissue during orthopedic procedures.
- Where it went
- Worldwide Distribution. US states of TX, OH, MI, and ME; and the countries of FR, DE, SK, IT, NL, CH, and CZ.
- How much
- 50 units
VAPR¿ TRIPOLAR 90" Degree Suction Electrodes (Model/Catalog No: 225028)
Z-0479-2016 · Class II · reported 2015-12-30 · Terminated
- Why it was recalled
- Mitek identified that on VAPR Tripolar 90 Degree Suction Electrode the ablation and coagulation buttons are colored incorrectly. The wiring for each button function was confirmed as correct and the non-conformance impacted button color only
- Product
- VAPR¿ TRIPOLAR 90" Degree Suction Electrodes (Model/Catalog No: 225028)
- Where it went
- Worldwide Distribution - US Distribution to the states of : AL, CA, CT, FL, KY, MN, MS, ND, NJ, NM and TX., and to the countries of: France, Netherlands and Norway.
- How much
- 127
DePuy Mitek FMS Intermediary Tubing with One-Way valve Product Code: 281142 Product Usage: The Intermedi
Z-0191-2014 · Class II · reported 2013-11-20 · Terminated
- Why it was recalled
- Pillow valve included in the FMS Duo outflow tubing set and the FMS Solo intermediary tubing set may not perform as intended. may result in backflow of irrigation fluid into the One Day Set which may potentially lead to patient cross-contamination
- Product
- DePuy Mitek FMS Intermediary Tubing with One-Way valve Product Code: 281142 Product Usage: The Intermediary Tubing for FMS¿ Fluid Management Systems connects the Inflow Tubing to the arthroscopic sheath during arthroscopic procedures. The Polyvinyl Chloride (PVC) tubing integrates several functional components including: luer locks for attaching the tube to the Inflow tubing and sheath, clamps, and a one-way valve. The FMS Intermediary Tubing is sterile and intended for single use with the FMS Fluid Management Systems in a surgical setting by personnel trained in arthroscopy .
- Where it went
- Worldwide Distribution - USA (Nationwide) and in the countries of: Australia , Austria , Belgium , Canada , Czech Republic , Denmark , Finland , France, Germany , India, Ireland , Israel, Italy , Latvia , Netherlands , Norway , Nepal, Poland , Portugal , Spain , Sweden , Switzerland , United Kingdom.
- How much
- 368 pk
DePuy Mitek FMS Outflow Tubing with One-Way valve Product Code:284649 Product Usage: The Intermediary Tub
Z-0192-2014 · Class II · reported 2013-11-20 · Terminated
- Why it was recalled
- Pillow valve included in the FMS Duo outflow tubing set and the FMS Solo intermediary tubing set may not perform as intended. may result in backflow of irrigation fluid into the One Day Set which may potentially lead to patient cross-contamination
- Product
- DePuy Mitek FMS Outflow Tubing with One-Way valve Product Code:284649 Product Usage: The Intermediary Tubing for FMS¿ Fluid Management Systems connects the Inflow Tubing to the arthroscopic sheath during arthroscopic procedures. The Polyvinyl Chloride (PVC) tubing integrates several functional components including: luer locks for attaching the tube to the Inflow tubing and sheath, clamps, and a one-way valve. The FMS Outflow Tubing with One-Way valve is sterile and intended for single use with the FMS Fluid Management Systems in a surgical setting by personnel trained in arthroscopy.
- Where it went
- Worldwide Distribution - USA (Nationwide) and in the countries of: Australia , Austria , Belgium , Canada , Czech Republic , Denmark , Finland , France, Germany , India, Ireland , Israel, Italy , Latvia , Netherlands , Norway , Nepal, Poland , Portugal , Spain , Sweden , Switzerland , United Kingdom.
- How much
- 29280 units
DePuy Mitek Tissue Liberator -Blade Up Catalog Number: 214623 The Reusable Arthroscopy Instruments are re
Z-1278-2013 · Class II · reported 2013-05-15 · Terminated
- Why it was recalled
- Labeled incorrectly as a Blade Up configuration instead of a Blade Down configuration.
- Product
- DePuy Mitek Tissue Liberator -Blade Up Catalog Number: 214623 The Reusable Arthroscopy Instruments are reusable, non-powered surgical instruments used during arthroscopic surgery of any joint except the spine.
- Where it went
- Worldwide Distribution - USA including KY, NJ and Internationally to Argnetina, Australia, Belgium,Columbia, India, France, Japan, Jordan, Latvija, Israel, Korea, and Great Britain. For questions regarding this recall call 508-977-3813.
- How much
- 27 units
ORTHOCORD One Violet and One Blue Braided Composite Suture with MO-6 Taper Needles, Product Code: 223115.
Z-1175-2013 · Class II · reported 2013-05-01 · Terminated
- Why it was recalled
- The IFU provided with the free strand ORTHOCORD sutures currently indicates that approximately 100% suture strength remains at six weeks after implantation. While this statement is true for Violet ORTHOCORD sutures, it is an inaccurate statement for Blue ORTHOCORD free strand sutures. Based on in-vivo strength retention study, a Blue ORTHOCORD sutures maintain approximately 80% of strength at six
- Product
- ORTHOCORD One Violet and One Blue Braided Composite Suture with MO-6 Taper Needles, Product Code: 223115. For use in general soft tissue approximation, and/or ligation, including orthopedic procedures.
- Where it went
- Nationwide Distribution.
- How much
- 46732
ORTHOCORD BLUE BRAIDED COMPOSITE SUTURE 2 (5 METRIC), Catalog Number: 223111. For use in general soft tissu
Z-1172-2013 · Class II · reported 2013-05-01 · Terminated
- Why it was recalled
- The IFU provided with the free strand ORTHOCORD sutures currently indicates that approximately 100% suture strength remains at six weeks after implantation. While this statement is true for Violet ORTHOCORD sutures, it is an inaccurate statement for Blue ORTHOCORD free strand sutures. Based on in-vivo strength retention study, a Blue ORTHOCORD sutures maintain approximately 80% of strength at six
- Product
- ORTHOCORD BLUE BRAIDED COMPOSITE SUTURE 2 (5 METRIC), Catalog Number: 223111. For use in general soft tissue approximation, and/or ligation, including orthopedic procedures.
- Where it went
- Nationwide Distribution.
- How much
- 65062
ORTHOCORD One Violet and One Blue Braided Composite Suture with OS-6 Reverse Cutting Needles, Product Code: 2
Z-1176-2013 · Class II · reported 2013-05-01 · Terminated
- Why it was recalled
- The IFU provided with the free strand ORTHOCORD sutures currently indicates that approximately 100% suture strength remains at six weeks after implantation. While this statement is true for Violet ORTHOCORD sutures, it is an inaccurate statement for Blue ORTHOCORD free strand sutures. Based on in-vivo strength retention study, a Blue ORTHOCORD sutures maintain approximately 80% of strength at six
- Product
- ORTHOCORD One Violet and One Blue Braided Composite Suture with OS-6 Reverse Cutting Needles, Product Code: 223116. For use in general soft tissue approximation, and/or ligation, including orthopedic procedures.
- Where it went
- Nationwide Distribution.
- How much
- 46501
ORTHOCORD ONE VIOLET AND ONE BLUE BRAIDED COMPOSITE SUTURE 2 (5 METRIC), Catalog Number: 223113. For use i
Z-1173-2013 · Class II · reported 2013-05-01 · Terminated
- Why it was recalled
- The IFU provided with the free strand ORTHOCORD sutures currently indicates that approximately 100% suture strength remains at six weeks after implantation. While this statement is true for Violet ORTHOCORD sutures, it is an inaccurate statement for Blue ORTHOCORD free strand sutures. Based on in-vivo strength retention study, a Blue ORTHOCORD sutures maintain approximately 80% of strength at six
- Product
- ORTHOCORD ONE VIOLET AND ONE BLUE BRAIDED COMPOSITE SUTURE 2 (5 METRIC), Catalog Number: 223113. For use in general soft tissue approximation, and/or ligation, including orthopedic procedures.
- Where it went
- Nationwide Distribution.
- How much
- 84174
ORTHOCORD One Violet and One Blue Braided Composite Suture with MO-7 Taper Needles, Catalog Number: 223114.
Z-1174-2013 · Class II · reported 2013-05-01 · Terminated
- Why it was recalled
- The IFU provided with the free strand ORTHOCORD sutures currently indicates that approximately 100% suture strength remains at six weeks after implantation. While this statement is true for Violet ORTHOCORD sutures, it is an inaccurate statement for Blue ORTHOCORD free strand sutures. Based on in-vivo strength retention study, a Blue ORTHOCORD sutures maintain approximately 80% of strength at six
- Product
- ORTHOCORD One Violet and One Blue Braided Composite Suture with MO-7 Taper Needles, Catalog Number: 223114. For use in general soft tissue approximation, and/or ligation, including orthopedic procedures.
- Where it went
- Nationwide Distribution.
- How much
- 206540
DePuy Mitek 11 MM Fully Fluted Reamer, Sterile Product Number: 232423 DePuy Mitek Sterile Reamers are inte
Z-0529-2013 · Class II · reported 2012-12-26 · Terminated
- Why it was recalled
- Specific units of the DePuy Mitek Sterile Reamers are not labeled correctly, 7 units were labeled as 11MM while the box actually contained a 10 MM reamer
- Product
- DePuy Mitek 11 MM Fully Fluted Reamer, Sterile Product Number: 232423 DePuy Mitek Sterile Reamers are intended to cut through cortical and cancellous bone during Cruciate Ligament Repair/Reconstruction of the knee.
- Where it went
- Nationwide Distribution including IL and NJ
- How much
- 20
Showing the 30 most recent of 51. The rest are searchable at the FDA enforcement report database.
Read these counts with three things in mind
A firm with more recalls usually makes more products. A contract packer runs
hundreds of lines. It will out-recall a company with one product whatever either
does. This is not a safety ranking.
Most recalls here are voluntary. A firm found its own fault and pulled the stock.
That is the system working. It lands in this database looking the same as a recall
the FDA had to push for.
Terminated means the FDA considers the recall finished. It is not a verdict.
It does not mean anyone was harmed.
Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.