Degania Silicone, Ltd.: 43 FDA recall reports
Every medical device recall the FDA has published for Degania Silicone, Ltd.. The fault, and how the agency graded it.
43Reports
0Class I
0Ongoing
2020-01-22Most recent
What the classes mean
The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.
Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.
By class
By year reported
Checking whether it affects you
The lot or code number is what matters. Not the product name. A recall covers
specific lots. Yours may not be one of them.
Check the code on your package against the report. The FDA keeps the current
list at fda.gov/safety/recalls.
The reports
The text below is quoted from the FDA record word for word.
Medline Thermistor Foley Catheter 14 FR 5-10 ml, non sterile Catalog Number: 102201101463MD {Medline code 553
Z-0809-2020 · Class II · reported 2020-01-22 · Terminated
- Why it was recalled
- Temperature deviation did not meet release specifications
- Product
- Medline Thermistor Foley Catheter 14 FR 5-10 ml, non sterile Catalog Number: 102201101463MD {Medline code 55346) Monitor urinary output and bladder temperature in addition to facilitating urine drainage.
- Where it went
- US Distribution to IL
- How much
- 2080 units
Cardinal Health Jackson-Pratt Channel Drain 19FR, Full flutes, with trocar Cardinal Health: PN: JP-2191 An
Z-2264-2019 · Class II · reported 2019-08-21 · Terminated
- Why it was recalled
- 15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body
- Product
- Cardinal Health Jackson-Pratt Channel Drain 19FR, Full flutes, with trocar Cardinal Health: PN: JP-2191 An introduction/drainage catheter is a device that is a flexible single or multilumen tube intended to be used to introduce nondrug fluids into body cavities other than blood vessels, drain fluids from body cavities, or evaluate certain physiologic conditions.
- Where it went
- Nationwide Distribution
- How much
- 6960
Cardinal Health Jackson-Pratt Channel Drain 19FR, 3/4 flutes, with trocar Cardinal Health: PN JP-2225 An
Z-2265-2019 · Class II · reported 2019-08-21 · Terminated
- Why it was recalled
- 15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body
- Product
- Cardinal Health Jackson-Pratt Channel Drain 19FR, 3/4 flutes, with trocar Cardinal Health: PN JP-2225 An introduction/drainage catheter is a device that is a flexible single or multilumen tube intended to be used to introduce nondrug fluids into body cavities other than blood vessels, drain fluids from body cavities, or evaluate certain physiologic conditions.
- Where it went
- Nationwide Distribution
- How much
- 2160
Cardinal Health Jackson-Pratt Channel Drain 15FR Cardinal PN:JP-2223 An introduction/drainage catheter
Z-2260-2019 · Class II · reported 2019-08-21 · Terminated
- Why it was recalled
- 15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body
- Product
- Cardinal Health Jackson-Pratt Channel Drain 15FR Cardinal PN:JP-2223 An introduction/drainage catheter is a device that is a flexible single or multilumen tube intended to be used to introduce nondrug fluids into body cavities other than blood vessels, drain fluids from body cavities, or evaluate certain physiologic conditions.
- Where it went
- Nationwide Distribution
- How much
- 640 units
Smith & Nephew Negative Pressure Wound Therapy 15FR Channel Wound Drain PN: 02872 An introduction/drainag
Z-2267-2019 · Class II · reported 2019-08-21 · Terminated
- Why it was recalled
- 15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body
- Product
- Smith & Nephew Negative Pressure Wound Therapy 15FR Channel Wound Drain PN: 02872 An introduction/drainage catheter is a device that is a flexible single or multilumen tube intended to be used to introduce nondrug fluids into body cavities other than blood vessels, drain fluids from body cavities, or evaluate certain physiologic conditions.
- Where it went
- Nationwide Distribution
- How much
- 2700
Cardinal Health Jackson-Pratt Channel Drain 15FR, Full flutes, with trocar Product Number: JP-2189 An
Z-2262-2019 · Class II · reported 2019-08-21 · Terminated
- Why it was recalled
- 15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body
- Product
- Cardinal Health Jackson-Pratt Channel Drain 15FR, Full flutes, with trocar Product Number: JP-2189 An introduction/drainage catheter is a device that is a flexible single or multilumen tube intended to be used to introduce nondrug fluids into body cavities other than blood vessels, drain fluids from body cavities, or evaluate certain physiologic conditions.
- Where it went
- Nationwide Distribution
- How much
- 6080
Cardinal Health Jackson-Pratt Channel Drain 19FR, Full flutes Cardinal Health: PN JP-2190 An introduction/
Z-2263-2019 · Class II · reported 2019-08-21 · Terminated
- Why it was recalled
- 15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body
- Product
- Cardinal Health Jackson-Pratt Channel Drain 19FR, Full flutes Cardinal Health: PN JP-2190 An introduction/drainage catheter is a device that is a flexible single or multilumen tube intended to be used to introduce nondrug fluids into body cavities other than blood vessels, drain fluids from body cavities, or evaluate certain physiologic conditions.
- Where it went
- Nationwide Distribution
- How much
- 4080
Cardinal Health Jackson-Pratt Channel Drain 15FR, Full flutes Cardinal Health PN: JP-2188 An introductio
Z-2261-2019 · Class II · reported 2019-08-21 · Terminated
- Why it was recalled
- 15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body
- Product
- Cardinal Health Jackson-Pratt Channel Drain 15FR, Full flutes Cardinal Health PN: JP-2188 An introduction/drainage catheter is a device that is a flexible single or multilumen tube intended to be used to introduce nondrug fluids into body cavities other than blood vessels, drain fluids from body cavities, or evaluate certain physiologic conditions.
- Where it went
- Nationwide Distribution
- How much
- 3040
Medline Silicone Drain 15Fr 3/16" Full Fluted PN: DYNJWE2188HB An introduction/drainage catheter is a de
Z-2266-2019 · Class II · reported 2019-08-21 · Terminated
- Why it was recalled
- 15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body
- Product
- Medline Silicone Drain 15Fr 3/16" Full Fluted PN: DYNJWE2188HB An introduction/drainage catheter is a device that is a flexible single or multilumen tube intended to be used to introduce nondrug fluids into body cavities other than blood vessels, drain fluids from body cavities, or evaluate certain physiologic conditions.
- Where it went
- Nationwide Distribution
- How much
- 200
Foley Catheter with temperature sensor 400TM 16FR, Catalogue Number 102201101680TY Product Usage: This is
Z-2766-2018 · Class II · reported 2018-09-26 · Terminated
- Why it was recalled
- The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
- Product
- Foley Catheter with temperature sensor 400TM 16FR, Catalogue Number 102201101680TY Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.
- Where it went
- Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
- How much
- 92150
Temperature Sensing 100% Silicone Foley Catheter, sterile, Catalogue Number 102201100880MD Product Usage:
Z-2772-2018 · Class II · reported 2018-09-26 · Terminated
- Why it was recalled
- The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
- Product
- Temperature Sensing 100% Silicone Foley Catheter, sterile, Catalogue Number 102201100880MD Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.
- Where it went
- Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
- How much
- 1000
Temperature sensor Foley catheter 10FR, Catalogue Number 102205101080AB Product Usage: This is sterile, si
Z-2756-2018 · Class II · reported 2018-09-26 · Terminated
- Why it was recalled
- The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
- Product
- Temperature sensor Foley catheter 10FR, Catalogue Number 102205101080AB Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.
- Where it went
- Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
- How much
- 500
Temperature Sensor Catheter 16FR, Catalogue Number 102201101663BI Product Usage: This is sterile, single u
Z-2761-2018 · Class II · reported 2018-09-26 · Terminated
- Why it was recalled
- The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
- Product
- Temperature Sensor Catheter 16FR, Catalogue Number 102201101663BI Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.
- Where it went
- Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
- How much
- 16133
Temperature sensing catheter 18FR, Catalogue Number 102203101863DO Product Usage: This is sterile, single
Z-2783-2018 · Class II · reported 2018-09-26 · Terminated
- Why it was recalled
- The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
- Product
- Temperature sensing catheter 18FR, Catalogue Number 102203101863DO Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.
- Where it went
- Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
- How much
- 2461
Temperature sensing catheter 16FR, Catalogue Number 102203101663DO Product Usage: This is sterile, single
Z-2782-2018 · Class II · reported 2018-09-26 · Terminated
- Why it was recalled
- The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
- Product
- Temperature sensing catheter 16FR, Catalogue Number 102203101663DO Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.
- Where it went
- Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
- How much
- 34627
Foley Catheter with temperature sensor 400TM 8FR, Catalogue Number 102201100880TY Product Usage: This is s
Z-2763-2018 · Class II · reported 2018-09-26 · Terminated
- Why it was recalled
- The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
- Product
- Foley Catheter with temperature sensor 400TM 8FR, Catalogue Number 102201100880TY Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.
- Where it went
- Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
- How much
- 4332
Foley Temperature Sensor Catheter 12FR, Catalogue Number 102201101280JP Product Usage: This is sterile, si
Z-2770-2018 · Class II · reported 2018-09-26 · Terminated
- Why it was recalled
- The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
- Product
- Foley Temperature Sensor Catheter 12FR, Catalogue Number 102201101280JP Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.
- Where it went
- Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
- How much
- 250
Temperature Sensing 100% Silicone Foley Catheter, sterile, Catalogue Number 102201101680MD Product Usage:
Z-2775-2018 · Class II · reported 2018-09-26 · Terminated
- Why it was recalled
- The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
- Product
- Temperature Sensing 100% Silicone Foley Catheter, sterile, Catalogue Number 102201101680MD Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.
- Where it went
- Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
- How much
- 500
Foley Temperature Sensor Catheter 14FR, Catalogue Number 102201101480JP Product Usage: This is sterile, si
Z-2771-2018 · Class II · reported 2018-09-26 · Terminated
- Why it was recalled
- The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
- Product
- Foley Temperature Sensor Catheter 14FR, Catalogue Number 102201101480JP Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.
- Where it went
- Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
- How much
- 5000
Temperature Sensor Catheter 14FR, Catalogue Number 102201101463BI Product Usage: This is sterile, single u
Z-2760-2018 · Class II · reported 2018-09-26 · Terminated
- Why it was recalled
- The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
- Product
- Temperature Sensor Catheter 14FR, Catalogue Number 102201101463BI Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.
- Where it went
- Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
- How much
- 16350
Temperature sensor Foley catheter 8FR, Catalogue Number 102205100880AB Product Usage: This is sterile, sin
Z-2755-2018 · Class II · reported 2018-09-26 · Terminated
- Why it was recalled
- The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
- Product
- Temperature sensor Foley catheter 8FR, Catalogue Number 102205100880AB Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.
- Where it went
- Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
- How much
- 11488
Thermistor Foley catheter, nonsterile, Catalogue Number 102201100863MD Product Usage: This is sterile, singl
Z-2779-2018 · Class II · reported 2018-09-26 · Terminated
- Why it was recalled
- The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
- Product
- Thermistor Foley catheter, nonsterile, Catalogue Number 102201100863MD Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.
- Where it went
- Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
- How much
- 330
Temperature sensing catheter 16FR, Catalogue Number 102201101663NU Product Usage: This is sterile, single
Z-2785-2018 · Class II · reported 2018-09-26 · Terminated
- Why it was recalled
- The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
- Product
- Temperature sensing catheter 16FR, Catalogue Number 102201101663NU Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.
- Where it went
- Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
- How much
- 1000
Temperature Sensor Catheter 8FR, Catalogue Number 102201100863BI Product Usage: This is sterile, single us
Z-2757-2018 · Class II · reported 2018-09-26 · Terminated
- Why it was recalled
- The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
- Product
- Temperature Sensor Catheter 8FR, Catalogue Number 102201100863BI Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.
- Where it went
- Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
- How much
- 1588
Thermistor Foley catheter, nonsterile, Catalogue Number 102201101663MD Product Usage: This is sterile, sin
Z-2778-2018 · Class II · reported 2018-09-26 · Terminated
- Why it was recalled
- The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
- Product
- Thermistor Foley catheter, nonsterile, Catalogue Number 102201101663MD Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.
- Where it went
- Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
- How much
- 14326
Temperature Sensor Catheter 18FR, Catalogue Number 102201101863BI Product Usage: This is sterile, single u
Z-2762-2018 · Class II · reported 2018-09-26 · Terminated
- Why it was recalled
- The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
- Product
- Temperature Sensor Catheter 18FR, Catalogue Number 102201101863BI Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.
- Where it went
- Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
- How much
- 1025
Temperature Sensor Catheter 12FR, Catalogue Number 102201101263BI Product Usage: This is sterile, single u
Z-2759-2018 · Class II · reported 2018-09-26 · Terminated
- Why it was recalled
- The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
- Product
- Temperature Sensor Catheter 12FR, Catalogue Number 102201101263BI Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.
- Where it went
- Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
- How much
- 1028
Temperature Sensor Catheter 10FR, Catalogue Number 102201101063BI Product Usage: This is sterile, single u
Z-2758-2018 · Class II · reported 2018-09-26 · Terminated
- Why it was recalled
- The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
- Product
- Temperature Sensor Catheter 10FR, Catalogue Number 102201101063BI Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.
- Where it went
- Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
- How much
- 475
Thermistor Foley catheter, nonsterile, Catalogue Number 102201101863MD Product Usage: This is sterile, sin
Z-2777-2018 · Class II · reported 2018-09-26 · Terminated
- Why it was recalled
- The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
- Product
- Thermistor Foley catheter, nonsterile, Catalogue Number 102201101863MD Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.
- Where it went
- Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
- How much
- 2784
Temperature sensor Foley catheter 18FR, Catalogue Number 102205101880AB Product Usage: This is sterile, si
Z-2754-2018 · Class II · reported 2018-09-26 · Terminated
- Why it was recalled
- The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
- Product
- Temperature sensor Foley catheter 18FR, Catalogue Number 102205101880AB Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.
- Where it went
- Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
- How much
- 153
Showing the 30 most recent of 43. The rest are searchable at the FDA enforcement report database.
Read these counts with three things in mind
A firm with more recalls usually makes more products. A contract packer runs
hundreds of lines. It will out-recall a company with one product whatever either
does. This is not a safety ranking.
Most recalls here are voluntary. A firm found its own fault and pulled the stock.
That is the system working. It lands in this database looking the same as a recall
the FDA had to push for.
Terminated means the FDA considers the recall finished. It is not a verdict.
It does not mean anyone was harmed.
Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.
← All device recalls