Covidien, LP: 126 FDA recall reports
Every medical device recall the FDA has published for Covidien, LP. The fault, and how the agency graded it.
126Reports
26Class I
102Ongoing
2026-06-10Most recent
26 Class I on this pageClass I means the FDA judged there was a reasonable probability of serious harm or death. Check the lot numbers against anything you hold.
What the classes mean
The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.
Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.
By class
By year reported
Checking whether it affects you
The lot or code number is what matters. Not the product name. A recall covers
specific lots. Yours may not be one of them.
Check the code on your package against the report. The FDA keeps the current
list at fda.gov/safety/recalls.
The reports
The text below is quoted from the FDA record word for word.
Covidien EndoStitch" Polysorb Single Use Loading Unit; Product numbers: 170053 and 170071;
Z-2281-2026 · Class II · reported 2026-06-10 · Ongoing
- Why it was recalled
- The presence of holes in the pouch may compromise the package s ability to maintain the sterile barrier and protect the product from environmental exposure.
- Product
- Covidien EndoStitch" Polysorb Single Use Loading Unit; Product numbers: 170053 and 170071;
- Where it went
- Domestic: Nationwide distribution; International: Austria, Canada, Denmark, France, Italy;
- How much
- 2448 units
Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); BOX LOBECTOMIA VATS 3D;
Z-0486-2026 · Class II · reported 2025-11-19 · Ongoing
- Why it was recalled
- A manufacturing issue could lead to a malformed cover tube press which can lead to internal components' movement during firing - leading to the potential for uncontrolled motion.
- Product
- Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); BOX LOBECTOMIA VATS 3D; CFN: BOX06170V4;
- Where it went
- Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Chile, Estonia, Finland, France, Germany, Greece, Hungary, Israel, Italy, Japan, Kuwait, Malaysia, Netherlands, Poland, Portugal, Romania, Saudi Arabia, Slovakia, Spain, United Kingdom.
- How much
- 2 units (OUS only)
Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); CFN: SIGSDL45CTVT;
Z-0485-2026 · Class II · reported 2025-11-19 · Ongoing
- Why it was recalled
- A manufacturing issue could lead to a malformed cover tube press which can lead to internal components' movement during firing - leading to the potential for uncontrolled motion.
- Product
- Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); CFN: SIGSDL45CTVT;
- Where it went
- Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Chile, Estonia, Finland, France, Germany, Greece, Hungary, Israel, Italy, Japan, Kuwait, Malaysia, Netherlands, Poland, Portugal, Romania, Saudi Arabia, Slovakia, Spain, United Kingdom.
- How much
- 3263 units
Chameleon PTA Balloon Catheter with Injection Port 5mm x 40mm x 75cm (CFN CH05-40-75US), 6mm x 40mm x 75cm (CF
Z-0257-2025 · Class II · reported 2024-11-06 · Ongoing
- Why it was recalled
- Specific lots of the Chameleon PTA Balloon Catheter failed to meet in-house balloon pressure testing.
- Product
- Chameleon PTA Balloon Catheter with Injection Port 5mm x 40mm x 75cm (CFN CH05-40-75US), 6mm x 40mm x 75cm (CFN CH06-40-75US), 7mm x 40mm x 75cm (CFN CH07-40-75US), 8mm x 40mm x 75cm (CFN CH08-40-75US), 9mm x 40mm x 75cm (CFN CH09-40-75US), 10mm x 40mm x 75cm (CFN CH10-40-75US), 12mm x 40mm x 75cm (CFN CH12-40-75US)
- Where it went
- Worldwide Distribution: US (Nationwide) and OUS (foreign) to countries of: Austria, Canada, Canary Islands, China, France, French Guiana, Germany, Greece, Guadeloupe, India, Ireland, Israel, Italy, Japan, Netherlands, Portugal, Singapore, Spain and Turkey.
- How much
- 16,336 units
Tri-Staple 2.0 Black Reinforced Intelligent Reload- Applications in abdominal, gynecologic, pediatric and tho
Z-1891-2024 · Class II · reported 2024-06-05 · Ongoing
- Why it was recalled
- Potential damage to the cartridge can result in poor staple formation and/or incomplete staple line, may cause a delay in surgery
- Product
- Tri-Staple 2.0 Black Reinforced Intelligent Reload- Applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection of tissue and creation of anastomosis Model: SIGTRSB60AXT
- Where it went
- Worldwide - US Nationwide distribution in the state of NJ and the countries of Australia, Japan, New Zealand, Saudi Arabia, Singapore, South Africa, Taiwan, Thailand.
- How much
- 1681 units
Sofsilk Braided Silk sutures, Product Number S-2782K. Nonabsorbable, sterile, non-mutagenic surgical suture co
Z-1697-2024 · Class II · reported 2024-05-08 · Ongoing
- Why it was recalled
- One lot of product was exposed to more than the approved number of Ethylene Oxide (EO) sterilization cycles. This may decrease the tensile strength of the sutures over time, which could result in harms such as wound dehiscence, hemorrhage, blood loss, bleeding, tissue breakdown, peritonitis, unspecified infection, vision loss (when used in ophthalmic application), and/or prolonged surgery.
- Product
- Sofsilk Braided Silk sutures, Product Number S-2782K. Nonabsorbable, sterile, non-mutagenic surgical suture composed of natural silk fibers, for use in soft tissue approximation and/or ligation.
- Where it went
- International distribution in the countries of Portugal and Italy.
- How much
- 36 total, OUS distribution only
Palindrome Precision HSI Chronic Catheter Sport Pack 14.5 Fr/Ch (4.8mm) x 19 cm (Symmetrical Tip, Heparin Co
Z-1634-2024 · Class II · reported 2024-05-01 · Ongoing
- Why it was recalled
- Products may not be heparin coated as indicated on the label-leads to a potential loss of benefit due to the absence of the heparin coating
- Product
- Palindrome Precision HSI Chronic Catheter Sport Pack 14.5 Fr/Ch (4.8mm) x 19 cm (Symmetrical Tip, Heparin Coating, Silver Ion Sleeve and Tal VenaTrac Stylet- Intended for acute and chronic hemodialysis, apheresis, and infusion. Model Number (CFN): 8888119369P
- Where it went
- Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canary Islands, Denmark, Faroe Islands, France, Germany, Greece, Iceland, Ireland, Italy, Mayotte, Norway, Poland, Reunion, South Africa, Spain, Sweden, United Kingdom.
- How much
- 305 units
Palindrome Precision H Chronic Catheter Kit 14.5 Fr/Ch (4.8mm) x 23 cm (Symmetrical Tip, Heparin Coating and
Z-1635-2024 · Class II · reported 2024-05-01 · Ongoing
- Why it was recalled
- Products may not be heparin coated as indicated on the label-leads to a potential loss of benefit due to the absence of the heparin coating
- Product
- Palindrome Precision H Chronic Catheter Kit 14.5 Fr/Ch (4.8mm) x 23 cm (Symmetrical Tip, Heparin Coating and Tal VenaTrac Stylet- Intended for acute and chronic hemodialysis, apheresis, and infusion. Model Number (CFN): 8888145044CP
- Where it went
- Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canary Islands, Denmark, Faroe Islands, France, Germany, Greece, Iceland, Ireland, Italy, Mayotte, Norway, Poland, Reunion, South Africa, Spain, Sweden, United Kingdom.
- How much
- 878 units
Covidien Auto Suture" Blunt Tip Trocar, Product Number OMS-T10BT
Z-1470-2024 · Class II · reported 2024-04-10 · Ongoing
- Why it was recalled
- The potential for trocar seal disengagement when using mesh products incorrectly with the device.
- Product
- Covidien Auto Suture" Blunt Tip Trocar, Product Number OMS-T10BT
- Where it went
- Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Aruba, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Colombia, Croatia, Czech Republic, Denmark, Estonia, Finland, France, French Guiana, French Polynesia, Germany, Gibraltar, Greece, Guadeloupe, Hong Kong, Hungary, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea, Republic Of, Luxembourg, Macao, Malaysia, Malta, Martinique, Mayotte, Mexico, Netherlands, New Caledonia, New Zealand, Nigeria, Northern Ireland, Norway, Oman, Poland, Portugal, Puerto Rico, Reunion, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Turkey, United Arab Emirates, United Kingdom, United States, Uzbekistan, Viet Nam, Virgin Islands, U.S.;
- How much
- 165,356 units
Signia" Small Diameter Curved Tip Intelligent Reload, Model SIGSDS30CTVT
Z-1457-2024 · Class II · reported 2024-04-10 · Ongoing
- Why it was recalled
- Under certain firing conditions, reloads were found to articulate in an uncontrolled manner potentially causing disruption to the staple line. The issue is related to components in impacted reloads that are not fully secure, which leads to uncontrolled articulation. The issue can occur using either the Endo GIA" Ultra Universal Stapler or Signia" Stapling System.
- Product
- Signia" Small Diameter Curved Tip Intelligent Reload, Model SIGSDS30CTVT
- Where it went
- Worldwide distribution - US Nationwide and the countries of France, Hong Kong, Israel and Japan.
- How much
- 30 Units
Covidien Signia" Small Diameter Curved Tip Intelligent Reload, Model SIGSDS30CTVT
Z-1458-2024 · Class II · reported 2024-04-10 · Ongoing
- Why it was recalled
- Under certain firing conditions, reloads were found to articulate in an uncontrolled manner potentially causing disruption to the staple line. The issue is related to components in impacted reloads that are not fully secure, which leads to uncontrolled articulation. The issue can occur using either the Endo GIA" Ultra Universal Stapler or Signia" Stapling System.
- Product
- Covidien Signia" Small Diameter Curved Tip Intelligent Reload, Model SIGSDS30CTVT
- Where it went
- Worldwide distribution - US Nationwide and the countries of France, Hong Kong, Israel and Japan.
- How much
- 3,023 units
Covidien Auto Suture" Blunt Tip Trocar, Product Number OMS-T12BT
Z-1468-2024 · Class II · reported 2024-04-10 · Ongoing
- Why it was recalled
- The potential for trocar seal disengagement when using mesh products incorrectly with the device.
- Product
- Covidien Auto Suture" Blunt Tip Trocar, Product Number OMS-T12BT
- Where it went
- Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Aruba, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Colombia, Croatia, Czech Republic, Denmark, Estonia, Finland, France, French Guiana, French Polynesia, Germany, Gibraltar, Greece, Guadeloupe, Hong Kong, Hungary, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea, Republic Of, Luxembourg, Macao, Malaysia, Malta, Martinique, Mayotte, Mexico, Netherlands, New Caledonia, New Zealand, Nigeria, Northern Ireland, Norway, Oman, Poland, Portugal, Puerto Rico, Reunion, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Turkey, United Arab Emirates, United Kingdom, United States, Uzbekistan, Viet Nam, Virgin Islands, U.S.;
- How much
- 93,673 units
Covidien Auto Suture" Blunt Tip Trocar, Product Number OMS-T10BTSNL
Z-1467-2024 · Class II · reported 2024-04-10 · Ongoing
- Why it was recalled
- The potential for trocar seal disengagement when using mesh products incorrectly with the device.
- Product
- Covidien Auto Suture" Blunt Tip Trocar, Product Number OMS-T10BTSNL
- Where it went
- Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Aruba, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Colombia, Croatia, Czech Republic, Denmark, Estonia, Finland, France, French Guiana, French Polynesia, Germany, Gibraltar, Greece, Guadeloupe, Hong Kong, Hungary, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea, Republic Of, Luxembourg, Macao, Malaysia, Malta, Martinique, Mayotte, Mexico, Netherlands, New Caledonia, New Zealand, Nigeria, Northern Ireland, Norway, Oman, Poland, Portugal, Puerto Rico, Reunion, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Turkey, United Arab Emirates, United Kingdom, United States, Uzbekistan, Viet Nam, Virgin Islands, U.S.;
- How much
- 11,796 units
Covidien Auto Suture" Blunt Tip Trocar, Product Number OMS-T12BTNL
Z-1471-2024 · Class II · reported 2024-04-10 · Ongoing
- Why it was recalled
- The potential for trocar seal disengagement when using mesh products incorrectly with the device.
- Product
- Covidien Auto Suture" Blunt Tip Trocar, Product Number OMS-T12BTNL
- Where it went
- Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Aruba, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Colombia, Croatia, Czech Republic, Denmark, Estonia, Finland, France, French Guiana, French Polynesia, Germany, Gibraltar, Greece, Guadeloupe, Hong Kong, Hungary, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea, Republic Of, Luxembourg, Macao, Malaysia, Malta, Martinique, Mayotte, Mexico, Netherlands, New Caledonia, New Zealand, Nigeria, Northern Ireland, Norway, Oman, Poland, Portugal, Puerto Rico, Reunion, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Turkey, United Arab Emirates, United Kingdom, United States, Uzbekistan, Viet Nam, Virgin Islands, U.S.;
- How much
- 44,777 units
Covidien Auto Suture" Blunt Tip Trocar, Product Number OMS-T10BTNL
Z-1466-2024 · Class II · reported 2024-04-10 · Ongoing
- Why it was recalled
- The potential for trocar seal disengagement when using mesh products incorrectly with the device.
- Product
- Covidien Auto Suture" Blunt Tip Trocar, Product Number OMS-T10BTNL
- Where it went
- Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Aruba, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Colombia, Croatia, Czech Republic, Denmark, Estonia, Finland, France, French Guiana, French Polynesia, Germany, Gibraltar, Greece, Guadeloupe, Hong Kong, Hungary, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea, Republic Of, Luxembourg, Macao, Malaysia, Malta, Martinique, Mayotte, Mexico, Netherlands, New Caledonia, New Zealand, Nigeria, Northern Ireland, Norway, Oman, Poland, Portugal, Puerto Rico, Reunion, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Turkey, United Arab Emirates, United Kingdom, United States, Uzbekistan, Viet Nam, Virgin Islands, U.S.;
- How much
- 315,716 units
Covidien Auto Suture" Structural Balloon Trocar, Product Number OMS-T10SB
Z-1465-2024 · Class II · reported 2024-04-10 · Ongoing
- Why it was recalled
- The potential for trocar seal disengagement when using mesh products incorrectly with the device.
- Product
- Covidien Auto Suture" Structural Balloon Trocar, Product Number OMS-T10SB
- Where it went
- Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Aruba, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Colombia, Croatia, Czech Republic, Denmark, Estonia, Finland, France, French Guiana, French Polynesia, Germany, Gibraltar, Greece, Guadeloupe, Hong Kong, Hungary, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea, Republic Of, Luxembourg, Macao, Malaysia, Malta, Martinique, Mayotte, Mexico, Netherlands, New Caledonia, New Zealand, Nigeria, Northern Ireland, Norway, Oman, Poland, Portugal, Puerto Rico, Reunion, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Turkey, United Arab Emirates, United Kingdom, United States, Uzbekistan, Viet Nam, Virgin Islands, U.S.;
- How much
- 124,863 units
Covidien Auto Suture" Blunt Tip Trocar, Product Number OMS-T10BTS
Z-1469-2024 · Class II · reported 2024-04-10 · Ongoing
- Why it was recalled
- The potential for trocar seal disengagement when using mesh products incorrectly with the device.
- Product
- Covidien Auto Suture" Blunt Tip Trocar, Product Number OMS-T10BTS
- Where it went
- Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Aruba, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Colombia, Croatia, Czech Republic, Denmark, Estonia, Finland, France, French Guiana, French Polynesia, Germany, Gibraltar, Greece, Guadeloupe, Hong Kong, Hungary, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea, Republic Of, Luxembourg, Macao, Malaysia, Malta, Martinique, Mayotte, Mexico, Netherlands, New Caledonia, New Zealand, Nigeria, Northern Ireland, Norway, Oman, Poland, Portugal, Puerto Rico, Reunion, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Turkey, United Arab Emirates, United Kingdom, United States, Uzbekistan, Viet Nam, Virgin Islands, U.S.;
- How much
- 18,048 units
AbsorbaTack Absorbable Fixation Device 30 Violet Absorbable Tacks 5 mm- -intended for fixation of prosthetic
Z-0986-2024 · Class II · reported 2024-02-14 · Ongoing
- Why it was recalled
- Packaging of the device may cause a low seal strength in the side seal of the foil pouch potentially resulting in a sterile/moisture barrier breach, compromising the product sterility
- Product
- AbsorbaTack Absorbable Fixation Device 30 Violet Absorbable Tacks 5 mm- -intended for fixation of prosthetic material to soft tissue in minimally invasive and open surgical hernia repair procedures Model ABSTACK30
- Where it went
- US Nationwide distribution in the states of CA, CO, FL, IL, TX.
- How much
- 54 units
Surgilon Braided Nylon sutures: 88861919-31 SURGILON* 4-0 BLK 7X75CMPCT 88861919-41 SURGILON* 3-0 BLK 7X75
Z-0844-2024 · Class II · reported 2024-02-07 · Ongoing
- Why it was recalled
- Specific lots of sutures were sterilized with gamma doses that exceeded the range approved. The extra gamma levels may decrease the strength of the sutures over time, which could result in harms such as wound dehiscence and hemorrhage/blood loss/ bleeding at a critical level.
- Product
- Surgilon Braided Nylon sutures: 88861919-31 SURGILON* 4-0 BLK 7X75CMPCT 88861919-41 SURGILON* 3-0 BLK 7X75CMPCT 88861919-51 SURGILON* 2-0 BLK 7X75CMPCT 88861919-71 SURGILON* 1 BLK 7X75CM PCT
- Where it went
- Worldwide US Nationwide distribution including in the states of AL, AZ, CA, FL, GA, HI, IA, IL, IN, MD, MI, MN, MO, MS, NC, NE, NJ, NV, NY, OH, OK, OR, PA, PR, TN, TX, VA, WA, WV and the countries of Australia, Bahrain, Belgium, Bulgaria, Canary Islands, China, Cyprus, Czech Republic, Finland, France, Germany, Greece, Hong Kong, Hungary, Italy, Japan, Korea, Republic Of Kosovo, Mongolia, Netherlands, New Zealand, Panama, Poland, Portugal, Puerto Rico, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, United Kingdom.
- How much
- 46,812 US; 68,976 OUS
Sofsilk Braided Silk sutures: CS-211 SOFSILK* 4-0 BLK 45CM CVF21 X36 GS-831 SOFSILK* 4-0 BLK 75CM V20 X36 S-1
Z-0845-2024 · Class II · reported 2024-02-07 · Ongoing
- Why it was recalled
- Specific lots of sutures were sterilized with gamma doses that exceeded the range approved. The extra gamma levels may decrease the strength of the sutures over time, which could result in harms such as wound dehiscence and hemorrhage/blood loss/ bleeding at a critical level.
- Product
- Sofsilk Braided Silk sutures: CS-211 SOFSILK* 4-0 BLK 45CM CVF21 X36 GS-831 SOFSILK* 4-0 BLK 75CM V20 X36 S-176 SOFSILK* 0 BLK 12X45CM PCT X24 S-182 SOFSILK* 5-0 BLK 12X45CM PCT X36 S-184 SOFSILK* 3-0 BLK 12X45CM PCT X36 S-185 SOFSILK* 2-0 BLK 12X45CM PCT X36 S-245 SOFSILK* 2-0 BLK 12X60CM PCT X36 S-246 SOFSILK* 0 BLK 6X60CM PCT X36 S-2782K SOFSILK* 4-0 WHI 45CM SS24 DA S-303 SOFSILK* 4-0 BLK 12X75CM PCT X36 S-304 SOFSILK* 3-0 BLK 12X75CM PCT X36 S-305 SOFSILK* 2-0 BLK 12X75CM PCT X36 S-316 SOFSILK* 0 BLK 12X75CM PCT X24 S-317 SOFSILK* 1 BLK 12X75CM PCT X24 S-605 SOFSILK* 2-0 BLK 2X150CM PCT X36 S-606 SOFSILK* 0 BLK 2X150CM PCT X36 S606-12 SOFSILK* 0 BLK 2X150CM PCT X12 S-610 SOFSILK* 5 BLK 2X150CM PCT X12 SOFSILK 2-0 30 BK CV-24 GS453-2 SOFSLK 4-0 24 BK CV-15 DA VS581-2 ***Added 4/1/24*** CS-792 SOFSILK* 0 BLK 100CM GCC90 X36 S-1172 SOFSILK* 6-0 BLK 45CM C1 X36 S-1735 SOFSILK* 4-0 BLK 30CM C1 DA X12 S-1750K SOFSILK* 6-0 BLK 45CM HE1 DAX12 S-1768K SOFSILK* 7-0 BLK 45CM HE6 DAX12 S-1789K SOFSILK* 4-0 BLK 45CM HE3 X12 S-187 SOFSILK* 1 BLK 6X45CM PCT X36 S-243 SOFSILK* 4-0 BLK 12X60CM PCT X36 S-2780K SOFSILK* 6-0 BLK 45CM SS-24DA S-282 SOFSILK* 5 BLK 3X60CM PCT X36 S-318 SOFSILK* 2...
- Where it went
- Worldwide US Nationwide distribution including in the states of AL, AZ, CA, FL, GA, HI, IA, IL, IN, MD, MI, MN, MO, MS, NC, NE, NJ, NV, NY, OH, OK, OR, PA, PR, TN, TX, VA, WA, WV and the countries of Australia, Bahrain, Belgium, Bulgaria, Canary Islands, China, Cyprus, Czech Republic, Finland, France, Germany, Greece, Hong Kong, Hungary, Italy, Japan, Korea, Republic Of Kosovo, Mongolia, Netherlands, New Zealand, Panama, Poland, Portugal, Puerto Rico, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, United Kingdom.
- How much
- 46,812 US; 68,976 OUS
Palindrome Chronic Catheter Kit Symmetrical Tip14.5 Fr/Ch(4.8mm)x23cm
Z-2632-2023 · Class II · reported 2023-10-04 · Ongoing
- Why it was recalled
- Some catheter kits labeled as 23 cm implant length incorrectly included catheters of 28 cm implant length; the actual catheter implant length is identifiable based upon the correct labeling on the catheter body.
- Product
- Palindrome Chronic Catheter Kit Symmetrical Tip14.5 Fr/Ch(4.8mm)x23cm
- Where it went
- Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Brazil, China, Czech Republic, Finland, France, Germany, Italy, Netherlands, Norway, Poland, Portugal, Puerto Rico, Spain, Thailand, United Kingdom.
- How much
- 3050 devices
Class I
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, IC Tray REF:
Z-2372-2023 · Class I · reported 2023-08-23 · Ongoing
- Why it was recalled
- Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
- Product
- MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, IC Tray REF: 8888102004HP
- Where it went
- US Nationwide - Worldwide Distribution Foreign: Argentina Austria Bahrain Bangladesh Belgium Brazil Brunei Darussalam Canada Cayman Islands China Costa Rica Czech Republic Denmark Dominican Republic Finland France French Polynesia Germany Greece Hungary India Iran, Islamic Republic Of Ireland Italy Kuwait Malaysia Maldives Mauritius Mexico Nepal Netherlands New Caledonia Nicaragua Norway Philippines Poland Portugal Puerto Rico Saudi Arabia Singapore Slovakia South Africa Spain Sri Lanka Sweden Switzerland Thailand Turkey United Arab Emirates United Kingdom
- How much
- 685 units
Class I
MAHURKAR 12 Fr Triple Lumen Acute Dialysis Catheter, 20 cm, Curved Extensions, Kit" REF: 8888345
Z-2388-2023 · Class I · reported 2023-08-23 · Ongoing
- Why it was recalled
- Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
- Product
- MAHURKAR 12 Fr Triple Lumen Acute Dialysis Catheter, 20 cm, Curved Extensions, Kit" REF: 8888345629
- Where it went
- US Nationwide - Worldwide Distribution Foreign: Argentina Austria Bahrain Bangladesh Belgium Brazil Brunei Darussalam Canada Cayman Islands China Costa Rica Czech Republic Denmark Dominican Republic Finland France French Polynesia Germany Greece Hungary India Iran, Islamic Republic Of Ireland Italy Kuwait Malaysia Maldives Mauritius Mexico Nepal Netherlands New Caledonia Nicaragua Norway Philippines Poland Portugal Puerto Rico Saudi Arabia Singapore Slovakia South Africa Spain Sri Lanka Sweden Switzerland Thailand Turkey United Arab Emirates United Kingdom
- How much
- 6949 units
Class I
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Curved Extensions, Kit
Z-2391-2023 · Class I · reported 2023-08-23 · Ongoing
- Why it was recalled
- Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
- Product
- MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Curved Extensions, Kit REF: 8888345637HP
- Where it went
- US Nationwide - Worldwide Distribution Foreign: Argentina Austria Bahrain Bangladesh Belgium Brazil Brunei Darussalam Canada Cayman Islands China Costa Rica Czech Republic Denmark Dominican Republic Finland France French Polynesia Germany Greece Hungary India Iran, Islamic Republic Of Ireland Italy Kuwait Malaysia Maldives Mauritius Mexico Nepal Netherlands New Caledonia Nicaragua Norway Philippines Poland Portugal Puerto Rico Saudi Arabia Singapore Slovakia South Africa Spain Sri Lanka Sweden Switzerland Thailand Turkey United Arab Emirates United Kingdom
- How much
- 3965 units
Class I
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 20 cm, Curved Extensions REF: 888
Z-2383-2023 · Class I · reported 2023-08-23 · Ongoing
- Why it was recalled
- Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
- Product
- MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 20 cm, Curved Extensions REF: 8888345520HP
- Where it went
- US Nationwide - Worldwide Distribution Foreign: Argentina Austria Bahrain Bangladesh Belgium Brazil Brunei Darussalam Canada Cayman Islands China Costa Rica Czech Republic Denmark Dominican Republic Finland France French Polynesia Germany Greece Hungary India Iran, Islamic Republic Of Ireland Italy Kuwait Malaysia Maldives Mauritius Mexico Nepal Netherlands New Caledonia Nicaragua Norway Philippines Poland Portugal Puerto Rico Saudi Arabia Singapore Slovakia South Africa Spain Sri Lanka Sweden Switzerland Thailand Turkey United Arab Emirates United Kingdom
- How much
- 295 units
Class I
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 20 cm, Straight Extensions, IC Tray REF:
Z-2371-2023 · Class I · reported 2023-08-23 · Ongoing
- Why it was recalled
- Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
- Product
- MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 20 cm, Straight Extensions, IC Tray REF: 8888102003HP
- Where it went
- US Nationwide - Worldwide Distribution Foreign: Argentina Austria Bahrain Bangladesh Belgium Brazil Brunei Darussalam Canada Cayman Islands China Costa Rica Czech Republic Denmark Dominican Republic Finland France French Polynesia Germany Greece Hungary India Iran, Islamic Republic Of Ireland Italy Kuwait Malaysia Maldives Mauritius Mexico Nepal Netherlands New Caledonia Nicaragua Norway Philippines Poland Portugal Puerto Rico Saudi Arabia Singapore Slovakia South Africa Spain Sri Lanka Sweden Switzerland Thailand Turkey United Arab Emirates United Kingdom
- How much
- 1875 units
Class I
MAHURKAR 12 Fr Triple Lumen Acute Dialysis Catheter, 13 cm, Curved Extensions, Kit" REF: 88883
Z-2384-2023 · Class I · reported 2023-08-23 · Ongoing
- Why it was recalled
- Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
- Product
- MAHURKAR 12 Fr Triple Lumen Acute Dialysis Catheter, 13 cm, Curved Extensions, Kit" REF: 8888345603
- Where it went
- US Nationwide - Worldwide Distribution Foreign: Argentina Austria Bahrain Bangladesh Belgium Brazil Brunei Darussalam Canada Cayman Islands China Costa Rica Czech Republic Denmark Dominican Republic Finland France French Polynesia Germany Greece Hungary India Iran, Islamic Republic Of Ireland Italy Kuwait Malaysia Maldives Mauritius Mexico Nepal Netherlands New Caledonia Nicaragua Norway Philippines Poland Portugal Puerto Rico Saudi Arabia Singapore Slovakia South Africa Spain Sri Lanka Sweden Switzerland Thailand Turkey United Arab Emirates United Kingdom
- How much
- 45880 units
Class I
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 20 cm, Straight Extensions, Kit REF: 88883
Z-2379-2023 · Class I · reported 2023-08-23 · Ongoing
- Why it was recalled
- Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
- Product
- MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 20 cm, Straight Extensions, Kit REF: 8888340629HP
- Where it went
- US Nationwide - Worldwide Distribution Foreign: Argentina Austria Bahrain Bangladesh Belgium Brazil Brunei Darussalam Canada Cayman Islands China Costa Rica Czech Republic Denmark Dominican Republic Finland France French Polynesia Germany Greece Hungary India Iran, Islamic Republic Of Ireland Italy Kuwait Malaysia Maldives Mauritius Mexico Nepal Netherlands New Caledonia Nicaragua Norway Philippines Poland Portugal Puerto Rico Saudi Arabia Singapore Slovakia South Africa Spain Sri Lanka Sweden Switzerland Thailand Turkey United Arab Emirates United Kingdom
- How much
- 7293 units
Class I
MAHURKAR 12 Fr Triple Lumen Acute Dialysis Catheter, 24 cm, Curved Extensions, Kit REF: 8888345637
Z-2390-2023 · Class I · reported 2023-08-23 · Ongoing
- Why it was recalled
- Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
- Product
- MAHURKAR 12 Fr Triple Lumen Acute Dialysis Catheter, 24 cm, Curved Extensions, Kit REF: 8888345637
- Where it went
- US Nationwide - Worldwide Distribution Foreign: Argentina Austria Bahrain Bangladesh Belgium Brazil Brunei Darussalam Canada Cayman Islands China Costa Rica Czech Republic Denmark Dominican Republic Finland France French Polynesia Germany Greece Hungary India Iran, Islamic Republic Of Ireland Italy Kuwait Malaysia Maldives Mauritius Mexico Nepal Netherlands New Caledonia Nicaragua Norway Philippines Poland Portugal Puerto Rico Saudi Arabia Singapore Slovakia South Africa Spain Sri Lanka Sweden Switzerland Thailand Turkey United Arab Emirates United Kingdom
- How much
- 647 units
Class I
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 20 cm, Straight Extensions, PASS Tray RE
Z-2376-2023 · Class I · reported 2023-08-23 · Ongoing
- Why it was recalled
- Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
- Product
- MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 20 cm, Straight Extensions, PASS Tray REF: 8888104003HP
- Where it went
- US Nationwide - Worldwide Distribution Foreign: Argentina Austria Bahrain Bangladesh Belgium Brazil Brunei Darussalam Canada Cayman Islands China Costa Rica Czech Republic Denmark Dominican Republic Finland France French Polynesia Germany Greece Hungary India Iran, Islamic Republic Of Ireland Italy Kuwait Malaysia Maldives Mauritius Mexico Nepal Netherlands New Caledonia Nicaragua Norway Philippines Poland Portugal Puerto Rico Saudi Arabia Singapore Slovakia South Africa Spain Sri Lanka Sweden Switzerland Thailand Turkey United Arab Emirates United Kingdom
- How much
- 1099 units
Showing the 30 most recent of 126. The rest are searchable at the FDA enforcement report database.
Read these counts with three things in mind
A firm with more recalls usually makes more products. A contract packer runs
hundreds of lines. It will out-recall a company with one product whatever either
does. This is not a safety ranking.
Most recalls here are voluntary. A firm found its own fault and pulled the stock.
That is the system working. It lands in this database looking the same as a recall
the FDA had to push for.
Terminated means the FDA considers the recall finished. It is not a verdict.
It does not mean anyone was harmed.
Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.