Cook Incorporated: 85 FDA recall reports
Every medical device recall the FDA has published for Cook Incorporated. The fault, and how the agency graded it.
85Reports
6Class I
84Ongoing
2026-05-13Most recent
6 Class I on this pageClass I means the FDA judged there was a reasonable probability of serious harm or death. Check the lot numbers against anything you hold.
What the classes mean
The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.
Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.
By class
By year reported
Checking whether it affects you
The lot or code number is what matters. Not the product name. A recall covers
specific lots. Yours may not be one of them.
Check the code on your package against the report. The FDA keeps the current
list at fda.gov/safety/recalls.
The reports
The text below is quoted from the FDA record word for word.
Class I
Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20 GPN G11916 Catheters for use in angiographic
Z-1934-2026 · Class I · reported 2026-05-13 · Ongoing
- Why it was recalled
- Marker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms associated with device fragmentation/separation, and vessel injury. In a worst-case scenario, device fragmentation and separation may cause life-threatening harm or death.
- Product
- Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20 GPN G11916 Catheters for use in angiographic procedures are available in a variety of French sizes, endhole sizes, lengths, materials and designs (e.g., polyethylene or nylon, non-braided or braided with 1:1 torque). The choice of catheter material for different angiographic procedures should be based on the physician s experience.
- Where it went
- US distribution nationwide. OUS distribution to Brunei, Brazil, Canda, China, Colombia, Costa Rica, Hong Kong, Israel, India, Japan, South Korea, Macau, New Zealand, Singapore, El Salvador, Taiwan, Uruguay
- How much
- 11,253
Class I
Aurous Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20-01 GPN G47308 Sizing catheters are sin
Z-1935-2026 · Class I · reported 2026-05-13 · Ongoing
- Why it was recalled
- Marker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms associated with device fragmentation/separation, and vessel injury. In a worst-case scenario, device fragmentation and separation may cause life-threatening harm or death.
- Product
- Aurous Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20-01 GPN G47308 Sizing catheters are single-use, sterile catheters with gold marker bands. These catheters are for use in angiographic procedures and available in a variety of lengths (70 and 100 cm) and tip configurations (straight, PIG, and VCF).
- Where it went
- US distribution nationwide. OUS distribution to Brunei, Brazil, Canda, China, Colombia, Costa Rica, Hong Kong, Israel, India, Japan, South Korea, Macau, New Zealand, Singapore, El Salvador, Taiwan, Uruguay
- How much
- 271
Class I
Beacon Tip Centimeter Sizing Catheter RPN/GPN NR5.0-35-70-P-10S-0-CSC-20 G31213 NR5.0-35-100-P-10S-0-CSC-2
Z-1936-2026 · Class I · reported 2026-05-13 · Ongoing
- Why it was recalled
- Marker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms associated with device fragmentation/separation, and vessel injury. In a worst-case scenario, device fragmentation and separation may cause life-threatening harm or death.
- Product
- Beacon Tip Centimeter Sizing Catheter RPN/GPN NR5.0-35-70-P-10S-0-CSC-20 G31213 NR5.0-35-100-P-10S-0-CSC-20 G31214 NR5.0-35-100-P-10S-PIG-CSC-20 G31216 NR5.0-35-70-P-10S-VCF-CSC-20 G31220 NR5.0-35-100-P-10S-VCF-CSC-20 G31221 Sizing catheters are single-use, sterile catheters with gold marker bands. These catheters are for use in angiographic procedures and available in a variety of lengths (70 and 100 cm) and tip configurations (straight, PIG, and VCF).
- Where it went
- US distribution nationwide. OUS distribution to Brunei, Brazil, Canda, China, Colombia, Costa Rica, Hong Kong, Israel, India, Japan, South Korea, Macau, New Zealand, Singapore, El Salvador, Taiwan, Uruguay
- How much
- 4,245
COOK MEDICAL Arterial Pressure Monitoring Tray, Reference Part Number APMY-2.5-15-0-2.5 , Order Number G58412;
Z-1778-2026 · Class II · reported 2026-04-15 · Ongoing
- Why it was recalled
- Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
- Product
- COOK MEDICAL Arterial Pressure Monitoring Tray, Reference Part Number APMY-2.5-15-0-2.5 , Order Number G58412; Reference Part Number APMY-4.0-21J-1-12, Order Number G58419
- Where it went
- Worldwide distribution.
- How much
- 20 units
Melker Universal Cuffed Emergency Cricothyrotomy Catheter Set: Reference Part Number C-TCCSB-500-UNI, Order N
Z-1785-2026 · Class II · reported 2026-04-15 · Ongoing
- Why it was recalled
- Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
- Product
- Melker Universal Cuffed Emergency Cricothyrotomy Catheter Set: Reference Part Number C-TCCSB-500-UNI, Order Number G32193; Reference Part Number C-TCCSB-500-UNI-01, Order Number G47665.
- Where it went
- Worldwide distribution.
- How much
- 476 units
COOK MEDICAL Cook¿ Strange Bile Duct Stone Exploration Set: Reference Part Number C-CSRC-7.5D-40-STRANGE, Ord
Z-1784-2026 · Class II · reported 2026-04-15 · Ongoing
- Why it was recalled
- Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
- Product
- COOK MEDICAL Cook¿ Strange Bile Duct Stone Exploration Set: Reference Part Number C-CSRC-7.5D-40-STRANGE, Order Number G09108
- Where it went
- Worldwide distribution.
- How much
- 49 units
COOK MEDICAL Wayne Pneumothorax Set: Reference Part Number C-UTPT-1020-WAYNE-IMH, Order Number G56532; Refe
Z-1791-2026 · Class II · reported 2026-04-15 · Ongoing
- Why it was recalled
- Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
- Product
- COOK MEDICAL Wayne Pneumothorax Set: Reference Part Number C-UTPT-1020-WAYNE-IMH, Order Number G56532; Reference Part Number C-UTPT-1400-WAYNE-112497-IMH, Order Number G56535.
- Where it went
- Worldwide distribution.
- How much
- 104 units
COOK MEDICAL Blue Rhino¿ G2-Multi Percutaneous Tracheostomy Introducer Tray: Reference Part Number C-PTISY-10
Z-1780-2026 · Class II · reported 2026-04-15 · Ongoing
- Why it was recalled
- Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
- Product
- COOK MEDICAL Blue Rhino¿ G2-Multi Percutaneous Tracheostomy Introducer Tray: Reference Part Number C-PTISY-100-HC-G-NA, Order Number G57703; Reference Part Number C-PTISY-100-UNL-HC-G-NA, Order Number G57706; Reference Part Number C-PTISY-100-UNS-HC-G-NA, Order Number G57704.
- Where it went
- Worldwide distribution.
- How much
- 441 units
COOK MEDICAL Approach CTO-18 Micro Wire Guide, Reference Part Number CMW-14-190-18G, Order Number G50788; Refe
Z-1777-2026 · Class II · reported 2026-04-15 · Ongoing
- Why it was recalled
- Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
- Product
- COOK MEDICAL Approach CTO-18 Micro Wire Guide, Reference Part Number CMW-14-190-18G, Order Number G50788; Reference Part Number CMW-14-300-18G, Order Number G50792
- Where it went
- Worldwide distribution.
- How much
- 75 units
COOK MEDICAL Spectrum¿ Central Venous Tray: Reference Part Number C-UQLMYJ-1001J-RSC-ABRM-HC-FST-A-RD, Order
Z-1789-2026 · Class II · reported 2026-04-15 · Ongoing
- Why it was recalled
- Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
- Product
- COOK MEDICAL Spectrum¿ Central Venous Tray: Reference Part Number C-UQLMYJ-1001J-RSC-ABRM-HC-FST-A-RD, Order Number G44800; Reference Part Number C-UTLMYJ-701J-ABRM-HC-IHI-FST-A-RD, Order Number G44809; Reference Part Number C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, Order Number G44806; Reference Part Number C-UTLMYJ-701J-RSC-ABRM-HC-IHI-FST-A-RD, Order Number G44807.
- Where it went
- Worldwide distribution.
- How much
- 2348 units
COOK MEDICAL Spectrum¿ Central Venous Catheter Set: Reference Part Number C-UDLM-401J-ABRM-HC, Order Number G
Z-1788-2026 · Class II · reported 2026-04-15 · Ongoing
- Why it was recalled
- Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
- Product
- COOK MEDICAL Spectrum¿ Central Venous Catheter Set: Reference Part Number C-UDLM-401J-ABRM-HC, Order Number G43983; Reference Part Number C-UDLM-401J-ABRM-HC-CAH, Order Number G24754; Reference Part Number C-UDLM-401J-LSC-ABRM-HC, Order Number G43986; Reference Part Number C-UDLM-401J-PED-ABRM-HC, Order Number G43984; Reference Part Number C-UDLM-401J-PED-ABRM-HC-CAH, Order Number G24752; Reference Part Number C-UDLM-401J-RSC-ABRM-HC, Order Number G43985; Reference Part Number C-UDLM-501J-ABRM-HC, Order Number G43995; Reference Part Number C-UDLM-501J-PED-ABRM-HC, Order Number G43996; Reference Part Number C-UDLM-501J-RSC-ABRM-HC, Order Number G43997; Reference Part Number C-UDLM-801J-ABRM-HC, Order Number G50817; Reference Part Number C-UQLM-1001J-ABRM-HC-RD, Order Number G51537; Reference Part Number C-UQLM-1001J-RSC-ABRM-HC-RD, Order Number G51539; Reference Part Number C-UTLM-501J-ABRM-HC, Order Number G44005; Reference Part Number C-UTLM-501J-ABRM-HC-CAH, Order Number G24803; Reference Part Number C-UTLM-501J-LSC-ABRM-HC, Order Number G44008; Reference Part Number C-UTLM-501J-PED-ABRM-HC, Order Number G44006; Reference Part Number...
- Where it went
- Worldwide distribution.
- How much
- 5080 units
COOK MEDICAL Approach CTO-6 Micro Wire Guide, Reference Part Number CMW-14-190-6G, Order Number G50786
Z-1775-2026 · Class II · reported 2026-04-15 · Ongoing
- Why it was recalled
- Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
- Product
- COOK MEDICAL Approach CTO-6 Micro Wire Guide, Reference Part Number CMW-14-190-6G, Order Number G50786
- Where it went
- Worldwide distribution.
- How much
- 15 units
COOK MEDICAL Thal-Quick Chest Tube Tray: Reference Part Number C-TQTSY-3200, Order Number G07242
Z-1790-2026 · Class II · reported 2026-04-15 · Ongoing
- Why it was recalled
- Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
- Product
- COOK MEDICAL Thal-Quick Chest Tube Tray: Reference Part Number C-TQTSY-3200, Order Number G07242
- Where it went
- Worldwide distribution.
- How much
- 2 units
COOK MEDICAL Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Set: Reference Part Number C-PTIS-100-H
Z-1779-2026 · Class II · reported 2026-04-15 · Ongoing
- Why it was recalled
- Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
- Product
- COOK MEDICAL Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Set: Reference Part Number C-PTIS-100-HC-G-EU, Order Number G57685; Reference Part Number C-PTIS-100-HC-G-NA, Order Number G57682; Reference Part Number C-PTIS-100-UNL-HC-G-EU, Order Number G57687; Reference Part Number C-PTIS-100-UNS-HC-G-EU, Order Number G57686.
- Where it went
- Worldwide distribution.
- How much
- 916 units
COOK MEDICAL Multipurpose Drainage Catheter: Reference Part Number ULT8.5-38-25-P-CLMS-8-RH-FS, Order Number
Z-1786-2026 · Class II · reported 2026-04-15 · Ongoing
- Why it was recalled
- Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
- Product
- COOK MEDICAL Multipurpose Drainage Catheter: Reference Part Number ULT8.5-38-25-P-CLMS-8-RH-FS, Order Number G29775
- Where it went
- Worldwide distribution.
- How much
- 20 units
COOK MEDICAL Ciaglia Blue Rhino Percutaneous Tracheostomy Introducer Set: Reference Part Number C-PTIS-100-FL
Z-1781-2026 · Class II · reported 2026-04-15 · Ongoing
- Why it was recalled
- Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
- Product
- COOK MEDICAL Ciaglia Blue Rhino Percutaneous Tracheostomy Introducer Set: Reference Part Number C-PTIS-100-FLEX-HC, Order Number G32092; Reference Part Number C-PTIS-100-HC, Order Number G12115.
- Where it went
- Worldwide distribution.
- How much
- 484 units
COOK MEDICAL NCompass¿ Nitinol Stone Extractor: Reference Part Number C-NTSE-2.4-115-NC3, Order Number G36250
Z-1787-2026 · Class II · reported 2026-04-15 · Ongoing
- Why it was recalled
- Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
- Product
- COOK MEDICAL NCompass¿ Nitinol Stone Extractor: Reference Part Number C-NTSE-2.4-115-NC3, Order Number G36250; Reference Part Number C-NTSE-2.4-115-NCT4, Order Number G36251.
- Where it went
- Worldwide distribution.
- How much
- 70 units
COOK MEDICAL Approach CTO-12 Micro Wire Guide: Reference Part Number CMW-14-190-12G, Order Number G50787; Re
Z-1776-2026 · Class II · reported 2026-04-15 · Ongoing
- Why it was recalled
- Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
- Product
- COOK MEDICAL Approach CTO-12 Micro Wire Guide: Reference Part Number CMW-14-190-12G, Order Number G50787; Reference Part Number CMW-14-300-12G, Order Number G50791
- Where it went
- Worldwide distribution.
- How much
- 49 units
COOK MEDICAL Wayne Pneumothorax Tray: Reference Part Number C-UTPTY-1400-WAYNE-112497-IMH, Order Number G5653
Z-1792-2026 · Class II · reported 2026-04-15 · Ongoing
- Why it was recalled
- Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
- Product
- COOK MEDICAL Wayne Pneumothorax Tray: Reference Part Number C-UTPTY-1400-WAYNE-112497-IMH, Order Number G56537; Reference Part Number C-UTPTY-1400-WAYNE-112497-IMH, Order Number G56537; Reference Part Number C-UTPTY-1400-WAYNE-112497-IMH, Order Number G56537; Reference Part Number C-UTPTY-1400-WAYNE-112497-IMH, Order Number G56537.
- Where it went
- Worldwide distribution.
- How much
- 318 units
COOK MEDICAL Cook¿ Spectrum¿ Central Venous Catheter Tray: Reference Part Number C-UTLMY-701J-ABRM-CUSTOM-000
Z-1783-2026 · Class II · reported 2026-04-15 · Ongoing
- Why it was recalled
- Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
- Product
- COOK MEDICAL Cook¿ Spectrum¿ Central Venous Catheter Tray: Reference Part Number C-UTLMY-701J-ABRM-CUSTOM-0003, Order Number G56424; Reference Part Number C-UTLMY-701J-ABRM-CUSTOM-0018, Order Number G21053; Reference Part Number C-UTLMY-701J-ABRM-CUSTOM-0042, Order Number G34914; Reference Part Number C-UTLMY-701J-ABRM-CUSTOM-0047, Order Number G44431; Reference Part Number C-UTLMY-701J-ABRM-CUSTOM-0048, Order Number G44432.
- Where it went
- Worldwide distribution.
- How much
- 1872 units
COOK MEDICAL Cook Staged Extubation Set: Reference Part Number C-CAE-14.0-83-EXTU, Order Number G24073
Z-1782-2026 · Class II · reported 2026-04-15 · Ongoing
- Why it was recalled
- Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
- Product
- COOK MEDICAL Cook Staged Extubation Set: Reference Part Number C-CAE-14.0-83-EXTU, Order Number G24073
- Where it went
- Worldwide distribution.
- How much
- 95 units
Flexor Check-Flo Introducers and Sets, Reference Part Numbers: 1) KCFW-9.0-18/38-45-RB-ANL0-HC; 2) KCFW-9
Z-0993-2026 · Class II · reported 2025-12-31 · Ongoing
- Why it was recalled
- Cook Medical identified that products supplied in the affected device lots may have been manufactured out of specification. Impacted products may have been cut to the incorrect length, not correctly trimmed, and not inspected properly.
- Product
- Flexor Check-Flo Introducers and Sets, Reference Part Numbers: 1) KCFW-9.0-18/38-45-RB-ANL0-HC; 2) KCFW-9.0-18/38-45-RB-ANL1-HC; 3) KCFW-10.0-35-45-RB-HFANL0-HC; 4) KCFW-10.0-35-45-RB-HFANL1-HC; 5) KCFW-12.0-35-45-RB-HFANL0-HC; 6) KCFW-12.0-35-45-RB-HFANL1-HC; 7) KCFW-9.0-35-38.5-RB-RTPS-100; 8) KCFW-10.0-35-RB-RTPS-100; 9) KCFW-9.0-38-55-RB-RAABE; 10) KCFW-9.0-38-70-RB-RAABE; 11) KCFW-9.0-38-80-RB; 12) KCFW-10.0-38-30-RB; 13) KCFW-10.0-38-40-RB; 14) KCFW-10.0-38-80-RB; 15) KCFW-10.0-38-RB; 16) KCFW-12.0-38-40-RB; 17) KCFW-12.0-38-80-RB; Flexor Introducers and Guiding Sheaths are designed to perform as a guiding sheath and/or introducer sheath. Flexor Introducers and Guiding Sheaths incorporate a Flexor shaft with a hemostasis valve and are provided with one or more dilators. These devices are available in various sizes, lengths and configurations. Configurations include differences in shaft construction, such as varying shaft stiffness and distal tip material and shape, dilator material, hydrophilic coating and/or distal radiopaque markers.
- Where it went
- Worldwide distribution.
- How much
- 7952 units
R¿sch-Uchida Transjugular Liver Access Sets, Reference Part Numbers: RUPS-100; The R¿sch-Uchida Transjugula
Z-0995-2026 · Class II · reported 2025-12-31 · Ongoing
- Why it was recalled
- Cook Medical identified that products supplied in the affected device lots may have been manufactured out of specification. Impacted products may have been cut to the incorrect length, not correctly trimmed, and not inspected properly.
- Product
- R¿sch-Uchida Transjugular Liver Access Sets, Reference Part Numbers: RUPS-100; The R¿sch-Uchida Transjugular Liver Access Set is intended for transjugular liver access in diagnostic and interventional procedures.
- Where it went
- Worldwide distribution.
- How much
- 23 units
Ring Transjugular Intrahepatic Access Sets, Reference Part Numbers: 1) RTPS-100; 2) RTPS-100-10.0; The
Z-0994-2026 · Class II · reported 2025-12-31 · Ongoing
- Why it was recalled
- Cook Medical identified that products supplied in the affected device lots may have been manufactured out of specification. Impacted products may have been cut to the incorrect length, not correctly trimmed, and not inspected properly.
- Product
- Ring Transjugular Intrahepatic Access Sets, Reference Part Numbers: 1) RTPS-100; 2) RTPS-100-10.0; The Ring Transjugular Intrahepatic Access Set is intended for transjugular liver access in diagnostic and interventional procedures.
- Where it went
- Worldwide distribution.
- How much
- 15 units
Class I
Beacon Tip 5.0 Fr Angiographic catheters are designed to facilitate diagnostic and therapeutic procedures. Cat
Z-1902-2025 · Class I · reported 2025-06-25 · Ongoing
- Why it was recalled
- Cook Medical identified that catheters supplied in the affected device lots may experience tip separation.
- Product
- Beacon Tip 5.0 Fr Angiographic catheters are designed to facilitate diagnostic and therapeutic procedures. Catheters are available in various configurations. Configurations include different shaft lengths and tip curves and some catheters are manufactured with a hydrophilic coating. Refer to the product label for product specifications (e.g., catheter length, distal curve configuration).
- Where it went
- Worldwide distribution.
- How much
- 26986 units
Check=Flo Performer Introducer, 10.0Fr, Reference Part Number(RPN) REF RCFW-10.0-0-38-30-RB, Order Number (GP
Z-1786-2025 · Class II · reported 2025-05-28 · Ongoing
- Why it was recalled
- Cook Medical identified that devices from the affected device lots may contain the incorrectly sized introducer sheath.
- Product
- Check=Flo Performer Introducer, 10.0Fr, Reference Part Number(RPN) REF RCFW-10.0-0-38-30-RB, Order Number (GPN) REF G08686,
- Where it went
- Worldwide distribution - US Nationwide and the countries of Belgium, Germany, Spain, France, Ireland, Italy Netherlands, Switzerland, South Africa, Slovenia.
- How much
- 99 units
Check=Flo Performer Introducer, 14.0Fr, Reference Part Number(RPN) REF RCFW-14.0-0-38-30-RB, Order Number (GP
Z-1787-2025 · Class II · reported 2025-05-28 · Ongoing
- Why it was recalled
- Cook Medical identified that devices from the affected device lots may contain the incorrectly sized introducer sheath.
- Product
- Check=Flo Performer Introducer, 14.0Fr, Reference Part Number(RPN) REF RCFW-14.0-0-38-30-RB, Order Number (GPN) REF G08957,
- Where it went
- Worldwide distribution - US Nationwide and the countries of Belgium, Germany, Spain, France, Ireland, Italy Netherlands, Switzerland, South Africa, Slovenia.
- How much
- 89 units
Tornado Embolization Microcoil, intended for embolization of selective vessel supply to arteriovenous malforma
Z-1762-2025 · Class II · reported 2025-05-21 · Ongoing
- Why it was recalled
- Cook Medical identified that devices from the affected device lots may contain the incorrectly sized embolization coil.
- Product
- Tornado Embolization Microcoil, intended for embolization of selective vessel supply to arteriovenous malformations and other vascular lesions, RPN MWCE-18S-3/2-TORNADO (GPN G08261), RPN MWCE-18S-3/2-TORNADO-081800 (GPN G13102), RPN MWCE-18S-4/2-TORNADO (GPN G08357)
- Where it went
- worldwide - No US distribution.
- How much
- 2952 units
Approach CTO Micro Wire Guide, RPN CMW-14-190-25G, GPN G50789
Z-0499-2025 · Class II · reported 2024-11-27 · Ongoing
- Why it was recalled
- Devices from the affected device lots have labels that state the incorrect expiration dates.
- Product
- Approach CTO Micro Wire Guide, RPN CMW-14-190-25G, GPN G50789
- Where it went
- International Distribution to: Korea only
- How much
- 10 units
Approach Hydro ST Micro Wire Guide; Catalog prefix HMW Reference Part Numbers: HMW-14-135-ST HMW-14-190-S
Z-0347-2025 · Class II · reported 2024-11-20 · Ongoing
- Why it was recalled
- Affected device lots have labels that state the incorrect expiration dates.
- Product
- Approach Hydro ST Micro Wire Guide; Catalog prefix HMW Reference Part Numbers: HMW-14-135-ST HMW-14-190-ST HMW-14-300-ST
- Where it went
- Worldwide - US Nationwide distribution.
- How much
- 5,953 US; 2,528 OUS
Showing the 30 most recent of 85. The rest are searchable at the FDA enforcement report database.
Read these counts with three things in mind
A firm with more recalls usually makes more products. A contract packer runs
hundreds of lines. It will out-recall a company with one product whatever either
does. This is not a safety ranking.
Most recalls here are voluntary. A firm found its own fault and pulled the stock.
That is the system working. It lands in this database looking the same as a recall
the FDA had to push for.
Terminated means the FDA considers the recall finished. It is not a verdict.
It does not mean anyone was harmed.
Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.