CME America, LLC

CME America, LLC: 48 FDA recall reports

Every medical device recall the FDA has published for CME America, LLC. The fault, and how the agency graded it.

48Reports
39Class I
0Ongoing
2020-12-09Most recent
39 Class I on this page

Class I means the FDA judged there was a reasonable probability of serious harm or death. Check the lot numbers against anything you hold.

What the classes mean

The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.

ClassWhat the FDA is saying
Class IThere is a reasonable probability that using this will cause serious harm or death.
Class IIIt may cause temporary or reversible harm. Serious harm is unlikely.
Class IIIIt breaks a rule but is unlikely to hurt anyone. Often a labelling fault.

Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.

By class

ClassReports
Class I39
Class II9

By year reported

YearReports
20151
202047
Checking whether it affects you

The lot or code number is what matters. Not the product name. A recall covers specific lots. Yours may not be one of them.

Check the code on your package against the report. The FDA keeps the current list at fda.gov/safety/recalls.

The reports

The text below is quoted from the FDA record word for word.

CME America T-Syringe Pump - T34 - Product Usage: infusion of medications or fluids requiring continuous or in

Z-0530-2021  ·  Class II  ·  reported 2020-12-09  ·  Terminated

Why it was recalled
Multiple reasons: 1) Due to wear and tear of pump motor block mechanism - may cause under infusion situations without the pump alarm being generated. 2) Due to cracks, pump hardware may be damaged by fluid used during the cleaning process.
Product
CME America T-Syringe Pump - T34 - Product Usage: infusion of medications or fluids requiring continuous or intermittent delivery at precisely-controlled infusion rates through clinically acceptable routes of administration, including intravenous, subcutaneous, percutaneous, intra-arterial, epidural, in close proximity to nerves, and into an intra-operative site (soft tissue/body cavity/surgical wound site).
Where it went
Worldwide distribution - U.S. Nationwide distribution including in the states of AZ, AK, CA, CO, CT, FL, ID, IL, KS, LA, MA, MI, MO, MT, NC, NE, NV, Oh, OK, OR, PA, TN, TX, WA, and WI. The country of Canada.
How much
Total pumps = 4,182 (all models)

CME America T-Syringe Pump - T34L PCA - Product Usage: infusion of medications or fluids requiring continuous

Z-0527-2021  ·  Class II  ·  reported 2020-12-09  ·  Terminated

Why it was recalled
Multiple reasons: 1) Due to wear and tear of pump motor block mechanism - may cause under infusion situations without the pump alarm being generated. 2) Due to cracks, pump hardware may be damaged by fluid used during the cleaning process.
Product
CME America T-Syringe Pump - T34L PCA - Product Usage: infusion of medications or fluids requiring continuous or intermittent delivery at precisely-controlled infusion rates through clinically acceptable routes of administration, including intravenous, subcutaneous, percutaneous, intra-arterial, epidural, in close proximity to nerves, and into an intra-operative site (soft tissue/body cavity/surgical wound site).
Where it went
Worldwide distribution - U.S. Nationwide distribution including in the states of AZ, AK, CA, CO, CT, FL, ID, IL, KS, LA, MA, MI, MO, MT, NC, NE, NV, Oh, OK, OR, PA, TN, TX, WA, and WI. The country of Canada.
How much
Total pumps = 4,182 (all models)

CME America T-Syringe Pump - Neofeed - Product Usage: infusion of medications or fluids requiring continuous o

Z-0528-2021  ·  Class II  ·  reported 2020-12-09  ·  Terminated

Why it was recalled
Multiple reasons: 1) Due to wear and tear of pump motor block mechanism - may cause under infusion situations without the pump alarm being generated. 2) Due to cracks, pump hardware may be damaged by fluid used during the cleaning process.
Product
CME America T-Syringe Pump - Neofeed - Product Usage: infusion of medications or fluids requiring continuous or intermittent delivery at precisely-controlled infusion rates through clinically acceptable routes of administration, including intravenous, subcutaneous, percutaneous, intra-arterial, epidural, in close proximity to nerves, and into an intra-operative site (soft tissue/body cavity/surgical wound site).
Where it went
Worldwide distribution - U.S. Nationwide distribution including in the states of AZ, AK, CA, CO, CT, FL, ID, IL, KS, LA, MA, MI, MO, MT, NC, NE, NV, Oh, OK, OR, PA, TN, TX, WA, and WI. The country of Canada.
How much
Total pumps = 4,182 (all models)

CME America T-Syringe Pump - Ameritus 1st Enteral Pump - Product Usage: infusion of medications or fluids requ

Z-0529-2021  ·  Class II  ·  reported 2020-12-09  ·  Terminated

Why it was recalled
Multiple reasons: 1) Due to wear and tear of pump motor block mechanism - may cause under infusion situations without the pump alarm being generated. 2) Due to cracks, pump hardware may be damaged by fluid used during the cleaning process.
Product
CME America T-Syringe Pump - Ameritus 1st Enteral Pump - Product Usage: infusion of medications or fluids requiring continuous or intermittent delivery at precisely-controlled infusion rates through clinically acceptable routes of administration, including intravenous, subcutaneous, percutaneous, intra-arterial, epidural, in close proximity to nerves, and into an intra-operative site (soft tissue/body cavity/surgical wound site).
Where it went
Worldwide distribution - U.S. Nationwide distribution including in the states of AZ, AK, CA, CO, CT, FL, ID, IL, KS, LA, MA, MI, MO, MT, NC, NE, NV, Oh, OK, OR, PA, TN, TX, WA, and WI. The country of Canada.
How much
Total pumps = 4,182 (all models)

CME America T-Syringe Pump - NeoThrive - Product Usage: infusion of medications or fluids requiring continuous

Z-0526-2021  ·  Class II  ·  reported 2020-12-09  ·  Terminated

Why it was recalled
Multiple reasons: 1) Due to wear and tear of pump motor block mechanism - may cause under infusion situations without the pump alarm being generated. 2) Due to cracks, pump hardware may be damaged by fluid used during the cleaning process.
Product
CME America T-Syringe Pump - NeoThrive - Product Usage: infusion of medications or fluids requiring continuous or intermittent delivery at precisely-controlled infusion rates through clinically acceptable routes of administration, including intravenous, subcutaneous, percutaneous, intra-arterial, epidural, in close proximity to nerves, and into an intra-operative site (soft tissue/body cavity/surgical wound site).
Where it went
Worldwide distribution - U.S. Nationwide distribution including in the states of AZ, AK, CA, CO, CT, FL, ID, IL, KS, LA, MA, MI, MO, MT, NC, NE, NV, Oh, OK, OR, PA, TN, TX, WA, and WI. The country of Canada.
How much
Total pumps = 4,182 (all models)

CME America T-Syringe Pump -T34L - Product Usage: infusion of medications or fluids requiring continuous or in

Z-0525-2021  ·  Class II  ·  reported 2020-12-09  ·  Terminated

Why it was recalled
Multiple reasons: 1) Due to wear and tear of pump motor block mechanism - may cause under infusion situations without the pump alarm being generated. 2) Due to cracks, pump hardware may be damaged by fluid used during the cleaning process.
Product
CME America T-Syringe Pump -T34L - Product Usage: infusion of medications or fluids requiring continuous or intermittent delivery at precisely-controlled infusion rates through clinically acceptable routes of administration, including intravenous, subcutaneous, percutaneous, intra-arterial, epidural, in close proximity to nerves, and into an intra-operative site (soft tissue/body cavity/surgical wound site).
Where it went
Worldwide distribution - U.S. Nationwide distribution including in the states of AZ, AK, CA, CO, CT, FL, ID, IL, KS, LA, MA, MI, MO, MT, NC, NE, NV, Oh, OK, OR, PA, TN, TX, WA, and WI. The country of Canada.
How much
Total Pumps = 4,182 (all models)

CMEAmerica BodyGuard Wall Charger: Lot number 2019-0364, Mode/Catalogue number 151-143XL, *+B 01151143X L0/$$7

Z-0305-2021  ·  Class II  ·  reported 2020-11-04  ·  Terminated

Why it was recalled
The wall charger may not properly charge the infusion pump battery.
Product
CMEAmerica BodyGuard Wall Charger: Lot number 2019-0364, Mode/Catalogue number 151-143XL, *+B 01151143X L0/$$72019-0364/ 16D20191101/* Lot Number Label: BODYGUARD LOT 2019-0364 Manufacturing Date: 2019 - 11-01 , LABEL I.D. SM-0642 REV. 01, CMEAmerica Wall Charger Kit, REF 151-143XLP, Contents; Wall Charger, Instructions, +B101150143XLP0/$$7A23456/16D000112010, Manufactured By: CME America LLC, 14988 W. 6th Ave., Suite 830, Golden. CO 80401, USA, SM-0706 Rev 00 External Charger INPUT ; AC 100-240V, 50-60Hz; 10W, 0.3A max., OUTPUT: DC 8.4V, 300mA/DC 9V, 800mA, CME America, LLC , Label ID: SM-0718, Rev. 00 Label, Wall Mount Charger, Pump Service: Use with CME Pumps & Accessories ONLY. Refer Servicing to Qualified Personnel., LABEL I.D. : SM-0756 REV 00
Where it went
U.S. Consignees: 3, IL, MO, KS OUS : 0 "No foreign consignees".
How much
27 chargers
Class I

BodyGuard Microset, REF: A120-161XPS

Z-2557-2020  ·  Class I  ·  reported 2020-07-22  ·  Terminated

Why it was recalled
Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
Product
BodyGuard Microset, REF: A120-161XPS
Where it went
U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
How much
345244
Class I

BodyGuard Microset With Female Luer, REF: A120-160XPSN

Z-2588-2020  ·  Class I  ·  reported 2020-07-22  ·  Terminated

Why it was recalled
Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
Product
BodyGuard Microset With Female Luer, REF: A120-160XPSN
Where it went
U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
How much
50
Class I

BodyGuard Microset, REF: A120-160X2YB

Z-2566-2020  ·  Class I  ·  reported 2020-07-22  ·  Terminated

Why it was recalled
Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
Product
BodyGuard Microset, REF: A120-160X2YB
Where it went
U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
How much
5000
Class I

BodyGuard BodySet, REF: A120-160SAS

Z-2553-2020  ·  Class I  ·  reported 2020-07-22  ·  Terminated

Why it was recalled
Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
Product
BodyGuard BodySet, REF: A120-160SAS
Where it went
U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
How much
3250
Class I

BodyGuard Microset, REF: A120-160XSFK

Z-2558-2020  ·  Class I  ·  reported 2020-07-22  ·  Terminated

Why it was recalled
Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
Product
BodyGuard Microset, REF: A120-160XSFK
Where it went
U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
How much
273700
Class I

CMExpress Microbore Sets, REF: A120-161C

Z-2580-2020  ·  Class I  ·  reported 2020-07-22  ·  Terminated

Why it was recalled
Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
Product
CMExpress Microbore Sets, REF: A120-161C
Where it went
U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
How much
50122
Class I

BodyGuard Microset w/ 0.2 mic filter and lower y-site with female luer, REF: A120-160CYFRV

Z-2589-2020  ·  Class I  ·  reported 2020-07-22  ·  Terminated

Why it was recalled
Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
Product
BodyGuard Microset w/ 0.2 mic filter and lower y-site with female luer, REF: A120-160CYFRV
Where it went
U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
How much
152570
Class I

BodyGuard set with Female Luer, REF: A120-003XSN

Z-2582-2020  ·  Class I  ·  reported 2020-07-22  ·  Terminated

Why it was recalled
Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
Product
BodyGuard set with Female Luer, REF: A120-003XSN
Where it went
U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
How much
7200
Class I

BodyGuard Microset with Filter, REF: A120-160XSFE

Z-2571-2020  ·  Class I  ·  reported 2020-07-22  ·  Terminated

Why it was recalled
Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
Product
BodyGuard Microset with Filter, REF: A120-160XSFE
Where it went
U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
How much
16850
Class I

BodyGuard Microset with Filter and Manual Priming Valve, REF: A120-112XSFV

Z-2583-2020  ·  Class I  ·  reported 2020-07-22  ·  Terminated

Why it was recalled
Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
Product
BodyGuard Microset with Filter and Manual Priming Valve, REF: A120-112XSFV
Where it went
U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
How much
6320
Class I

BodyGuard Microset, REF: A120-161XPSV

Z-2565-2020  ·  Class I  ·  reported 2020-07-22  ·  Terminated

Why it was recalled
Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
Product
BodyGuard Microset, REF: A120-161XPSV
Where it went
U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
How much
291386
Class I

BodyGuard Microset, REF: A120-161XYBS

Z-2563-2020  ·  Class I  ·  reported 2020-07-22  ·  Terminated

Why it was recalled
Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
Product
BodyGuard Microset, REF: A120-161XYBS
Where it went
U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
How much
18520
Class I

BodyGuard Microset, REF: A100-163XSL

Z-2574-2020  ·  Class I  ·  reported 2020-07-22  ·  Terminated

Why it was recalled
Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
Product
BodyGuard Microset, REF: A100-163XSL
Where it went
U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
How much
28600
Class I

Standard BodySet with Needleless Connectors, REF: A120-125XS

Z-2585-2020  ·  Class I  ·  reported 2020-07-22  ·  Terminated

Why it was recalled
Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
Product
Standard BodySet with Needleless Connectors, REF: A120-125XS
Where it went
U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
How much
36286
Class I

BodyGuard Microset with Male Luer Connectors, REF: A100-163XEBF

Z-2581-2020  ·  Class I  ·  reported 2020-07-22  ·  Terminated

Why it was recalled
Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
Product
BodyGuard Microset with Male Luer Connectors, REF: A100-163XEBF
Where it went
U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
How much
8016
Class I

BodyGuard Microset with Filter, REF: A100-164XESF

Z-2572-2020  ·  Class I  ·  reported 2020-07-22  ·  Terminated

Why it was recalled
Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
Product
BodyGuard Microset with Filter, REF: A100-164XESF
Where it went
U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
How much
55455
Class I

BodyGuard Microset, REF: A100-163XE90S

Z-2573-2020  ·  Class I  ·  reported 2020-07-22  ·  Terminated

Why it was recalled
Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
Product
BodyGuard Microset, REF: A100-163XE90S
Where it went
U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
How much
390230
Class I

BodyGuard Microset, REF: A120-003XS2YV

Z-2575-2020  ·  Class I  ·  reported 2020-07-22  ·  Terminated

Why it was recalled
Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
Product
BodyGuard Microset, REF: A120-003XS2YV
Where it went
U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
How much
169856
Class I

BodyGuard Microset w/ 0.2 mic filter and lower y-site with female luer, REF: A120-160XYBF

Z-2584-2020  ·  Class I  ·  reported 2020-07-22  ·  Terminated

Why it was recalled
Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
Product
BodyGuard Microset w/ 0.2 mic filter and lower y-site with female luer, REF: A120-160XYBF
Where it went
U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
How much
6016
Class I

BodyGuard Microset, REF: A120-112XSFK

Z-2559-2020  ·  Class I  ·  reported 2020-07-22  ·  Terminated

Why it was recalled
Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
Product
BodyGuard Microset, REF: A120-112XSFK
Where it went
U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
How much
1273892
Class I

BodyGuard Microset, REF: A120-160XYSF

Z-2561-2020  ·  Class I  ·  reported 2020-07-22  ·  Terminated

Why it was recalled
Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
Product
BodyGuard Microset, REF: A120-160XYSF
Where it went
U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
How much
50
Class I

BodyGuard Microset, REF: A100-163XESV

Z-2555-2020  ·  Class I  ·  reported 2020-07-22  ·  Terminated

Why it was recalled
Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
Product
BodyGuard Microset, REF: A100-163XESV
Where it went
U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
How much
178057
Class I

BodyGuard Microset with Filter, REF: A100-163XESF

Z-2564-2020  ·  Class I  ·  reported 2020-07-22  ·  Terminated

Why it was recalled
Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
Product
BodyGuard Microset with Filter, REF: A100-163XESF
Where it went
U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
How much
95110

Showing the 30 most recent of 48. The rest are searchable at the FDA enforcement report database.

Read these counts with three things in mind

A firm with more recalls usually makes more products. A contract packer runs hundreds of lines. It will out-recall a company with one product whatever either does. This is not a safety ranking.

Most recalls here are voluntary. A firm found its own fault and pulled the stock. That is the system working. It lands in this database looking the same as a recall the FDA had to push for.

Terminated means the FDA considers the recall finished. It is not a verdict. It does not mean anyone was harmed.

Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.