CME America, LLC: 48 FDA recall reports
Every medical device recall the FDA has published for CME America, LLC. The fault, and how the agency graded it.
48Reports
39Class I
0Ongoing
2020-12-09Most recent
39 Class I on this pageClass I means the FDA judged there was a reasonable probability of serious harm or death. Check the lot numbers against anything you hold.
What the classes mean
The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.
Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.
By class
By year reported
Checking whether it affects you
The lot or code number is what matters. Not the product name. A recall covers
specific lots. Yours may not be one of them.
Check the code on your package against the report. The FDA keeps the current
list at fda.gov/safety/recalls.
The reports
The text below is quoted from the FDA record word for word.
CME America T-Syringe Pump - T34 - Product Usage: infusion of medications or fluids requiring continuous or in
Z-0530-2021 · Class II · reported 2020-12-09 · Terminated
- Why it was recalled
- Multiple reasons: 1) Due to wear and tear of pump motor block mechanism - may cause under infusion situations without the pump alarm being generated. 2) Due to cracks, pump hardware may be damaged by fluid used during the cleaning process.
- Product
- CME America T-Syringe Pump - T34 - Product Usage: infusion of medications or fluids requiring continuous or intermittent delivery at precisely-controlled infusion rates through clinically acceptable routes of administration, including intravenous, subcutaneous, percutaneous, intra-arterial, epidural, in close proximity to nerves, and into an intra-operative site (soft tissue/body cavity/surgical wound site).
- Where it went
- Worldwide distribution - U.S. Nationwide distribution including in the states of AZ, AK, CA, CO, CT, FL, ID, IL, KS, LA, MA, MI, MO, MT, NC, NE, NV, Oh, OK, OR, PA, TN, TX, WA, and WI. The country of Canada.
- How much
- Total pumps = 4,182 (all models)
CME America T-Syringe Pump - T34L PCA - Product Usage: infusion of medications or fluids requiring continuous
Z-0527-2021 · Class II · reported 2020-12-09 · Terminated
- Why it was recalled
- Multiple reasons: 1) Due to wear and tear of pump motor block mechanism - may cause under infusion situations without the pump alarm being generated. 2) Due to cracks, pump hardware may be damaged by fluid used during the cleaning process.
- Product
- CME America T-Syringe Pump - T34L PCA - Product Usage: infusion of medications or fluids requiring continuous or intermittent delivery at precisely-controlled infusion rates through clinically acceptable routes of administration, including intravenous, subcutaneous, percutaneous, intra-arterial, epidural, in close proximity to nerves, and into an intra-operative site (soft tissue/body cavity/surgical wound site).
- Where it went
- Worldwide distribution - U.S. Nationwide distribution including in the states of AZ, AK, CA, CO, CT, FL, ID, IL, KS, LA, MA, MI, MO, MT, NC, NE, NV, Oh, OK, OR, PA, TN, TX, WA, and WI. The country of Canada.
- How much
- Total pumps = 4,182 (all models)
CME America T-Syringe Pump - Neofeed - Product Usage: infusion of medications or fluids requiring continuous o
Z-0528-2021 · Class II · reported 2020-12-09 · Terminated
- Why it was recalled
- Multiple reasons: 1) Due to wear and tear of pump motor block mechanism - may cause under infusion situations without the pump alarm being generated. 2) Due to cracks, pump hardware may be damaged by fluid used during the cleaning process.
- Product
- CME America T-Syringe Pump - Neofeed - Product Usage: infusion of medications or fluids requiring continuous or intermittent delivery at precisely-controlled infusion rates through clinically acceptable routes of administration, including intravenous, subcutaneous, percutaneous, intra-arterial, epidural, in close proximity to nerves, and into an intra-operative site (soft tissue/body cavity/surgical wound site).
- Where it went
- Worldwide distribution - U.S. Nationwide distribution including in the states of AZ, AK, CA, CO, CT, FL, ID, IL, KS, LA, MA, MI, MO, MT, NC, NE, NV, Oh, OK, OR, PA, TN, TX, WA, and WI. The country of Canada.
- How much
- Total pumps = 4,182 (all models)
CME America T-Syringe Pump - Ameritus 1st Enteral Pump - Product Usage: infusion of medications or fluids requ
Z-0529-2021 · Class II · reported 2020-12-09 · Terminated
- Why it was recalled
- Multiple reasons: 1) Due to wear and tear of pump motor block mechanism - may cause under infusion situations without the pump alarm being generated. 2) Due to cracks, pump hardware may be damaged by fluid used during the cleaning process.
- Product
- CME America T-Syringe Pump - Ameritus 1st Enteral Pump - Product Usage: infusion of medications or fluids requiring continuous or intermittent delivery at precisely-controlled infusion rates through clinically acceptable routes of administration, including intravenous, subcutaneous, percutaneous, intra-arterial, epidural, in close proximity to nerves, and into an intra-operative site (soft tissue/body cavity/surgical wound site).
- Where it went
- Worldwide distribution - U.S. Nationwide distribution including in the states of AZ, AK, CA, CO, CT, FL, ID, IL, KS, LA, MA, MI, MO, MT, NC, NE, NV, Oh, OK, OR, PA, TN, TX, WA, and WI. The country of Canada.
- How much
- Total pumps = 4,182 (all models)
CME America T-Syringe Pump - NeoThrive - Product Usage: infusion of medications or fluids requiring continuous
Z-0526-2021 · Class II · reported 2020-12-09 · Terminated
- Why it was recalled
- Multiple reasons: 1) Due to wear and tear of pump motor block mechanism - may cause under infusion situations without the pump alarm being generated. 2) Due to cracks, pump hardware may be damaged by fluid used during the cleaning process.
- Product
- CME America T-Syringe Pump - NeoThrive - Product Usage: infusion of medications or fluids requiring continuous or intermittent delivery at precisely-controlled infusion rates through clinically acceptable routes of administration, including intravenous, subcutaneous, percutaneous, intra-arterial, epidural, in close proximity to nerves, and into an intra-operative site (soft tissue/body cavity/surgical wound site).
- Where it went
- Worldwide distribution - U.S. Nationwide distribution including in the states of AZ, AK, CA, CO, CT, FL, ID, IL, KS, LA, MA, MI, MO, MT, NC, NE, NV, Oh, OK, OR, PA, TN, TX, WA, and WI. The country of Canada.
- How much
- Total pumps = 4,182 (all models)
CME America T-Syringe Pump -T34L - Product Usage: infusion of medications or fluids requiring continuous or in
Z-0525-2021 · Class II · reported 2020-12-09 · Terminated
- Why it was recalled
- Multiple reasons: 1) Due to wear and tear of pump motor block mechanism - may cause under infusion situations without the pump alarm being generated. 2) Due to cracks, pump hardware may be damaged by fluid used during the cleaning process.
- Product
- CME America T-Syringe Pump -T34L - Product Usage: infusion of medications or fluids requiring continuous or intermittent delivery at precisely-controlled infusion rates through clinically acceptable routes of administration, including intravenous, subcutaneous, percutaneous, intra-arterial, epidural, in close proximity to nerves, and into an intra-operative site (soft tissue/body cavity/surgical wound site).
- Where it went
- Worldwide distribution - U.S. Nationwide distribution including in the states of AZ, AK, CA, CO, CT, FL, ID, IL, KS, LA, MA, MI, MO, MT, NC, NE, NV, Oh, OK, OR, PA, TN, TX, WA, and WI. The country of Canada.
- How much
- Total Pumps = 4,182 (all models)
CMEAmerica BodyGuard Wall Charger: Lot number 2019-0364, Mode/Catalogue number 151-143XL, *+B 01151143X L0/$$7
Z-0305-2021 · Class II · reported 2020-11-04 · Terminated
- Why it was recalled
- The wall charger may not properly charge the infusion pump battery.
- Product
- CMEAmerica BodyGuard Wall Charger: Lot number 2019-0364, Mode/Catalogue number 151-143XL, *+B 01151143X L0/$$72019-0364/ 16D20191101/* Lot Number Label: BODYGUARD LOT 2019-0364 Manufacturing Date: 2019 - 11-01 , LABEL I.D. SM-0642 REV. 01, CMEAmerica Wall Charger Kit, REF 151-143XLP, Contents; Wall Charger, Instructions, +B101150143XLP0/$$7A23456/16D000112010, Manufactured By: CME America LLC, 14988 W. 6th Ave., Suite 830, Golden. CO 80401, USA, SM-0706 Rev 00 External Charger INPUT ; AC 100-240V, 50-60Hz; 10W, 0.3A max., OUTPUT: DC 8.4V, 300mA/DC 9V, 800mA, CME America, LLC , Label ID: SM-0718, Rev. 00 Label, Wall Mount Charger, Pump Service: Use with CME Pumps & Accessories ONLY. Refer Servicing to Qualified Personnel., LABEL I.D. : SM-0756 REV 00
- Where it went
- U.S. Consignees: 3, IL, MO, KS OUS : 0 "No foreign consignees".
- How much
- 27 chargers
Class I
BodyGuard Microset, REF: A120-161XPS
Z-2557-2020 · Class I · reported 2020-07-22 · Terminated
- Why it was recalled
- Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
- Product
- BodyGuard Microset, REF: A120-161XPS
- Where it went
- U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
- How much
- 345244
Class I
BodyGuard Microset With Female Luer, REF: A120-160XPSN
Z-2588-2020 · Class I · reported 2020-07-22 · Terminated
- Why it was recalled
- Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
- Product
- BodyGuard Microset With Female Luer, REF: A120-160XPSN
- Where it went
- U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
- How much
- 50
Class I
BodyGuard Microset, REF: A120-160X2YB
Z-2566-2020 · Class I · reported 2020-07-22 · Terminated
- Why it was recalled
- Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
- Product
- BodyGuard Microset, REF: A120-160X2YB
- Where it went
- U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
- How much
- 5000
Class I
BodyGuard BodySet, REF: A120-160SAS
Z-2553-2020 · Class I · reported 2020-07-22 · Terminated
- Why it was recalled
- Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
- Product
- BodyGuard BodySet, REF: A120-160SAS
- Where it went
- U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
- How much
- 3250
Class I
BodyGuard Microset, REF: A120-160XSFK
Z-2558-2020 · Class I · reported 2020-07-22 · Terminated
- Why it was recalled
- Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
- Product
- BodyGuard Microset, REF: A120-160XSFK
- Where it went
- U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
- How much
- 273700
Class I
CMExpress Microbore Sets, REF: A120-161C
Z-2580-2020 · Class I · reported 2020-07-22 · Terminated
- Why it was recalled
- Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
- Product
- CMExpress Microbore Sets, REF: A120-161C
- Where it went
- U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
- How much
- 50122
Class I
BodyGuard Microset w/ 0.2 mic filter and lower y-site with female luer, REF: A120-160CYFRV
Z-2589-2020 · Class I · reported 2020-07-22 · Terminated
- Why it was recalled
- Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
- Product
- BodyGuard Microset w/ 0.2 mic filter and lower y-site with female luer, REF: A120-160CYFRV
- Where it went
- U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
- How much
- 152570
Class I
BodyGuard set with Female Luer, REF: A120-003XSN
Z-2582-2020 · Class I · reported 2020-07-22 · Terminated
- Why it was recalled
- Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
- Product
- BodyGuard set with Female Luer, REF: A120-003XSN
- Where it went
- U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
- How much
- 7200
Class I
BodyGuard Microset with Filter, REF: A120-160XSFE
Z-2571-2020 · Class I · reported 2020-07-22 · Terminated
- Why it was recalled
- Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
- Product
- BodyGuard Microset with Filter, REF: A120-160XSFE
- Where it went
- U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
- How much
- 16850
Class I
BodyGuard Microset with Filter and Manual Priming Valve, REF: A120-112XSFV
Z-2583-2020 · Class I · reported 2020-07-22 · Terminated
- Why it was recalled
- Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
- Product
- BodyGuard Microset with Filter and Manual Priming Valve, REF: A120-112XSFV
- Where it went
- U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
- How much
- 6320
Class I
BodyGuard Microset, REF: A120-161XPSV
Z-2565-2020 · Class I · reported 2020-07-22 · Terminated
- Why it was recalled
- Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
- Product
- BodyGuard Microset, REF: A120-161XPSV
- Where it went
- U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
- How much
- 291386
Class I
BodyGuard Microset, REF: A120-161XYBS
Z-2563-2020 · Class I · reported 2020-07-22 · Terminated
- Why it was recalled
- Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
- Product
- BodyGuard Microset, REF: A120-161XYBS
- Where it went
- U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
- How much
- 18520
Class I
BodyGuard Microset, REF: A100-163XSL
Z-2574-2020 · Class I · reported 2020-07-22 · Terminated
- Why it was recalled
- Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
- Product
- BodyGuard Microset, REF: A100-163XSL
- Where it went
- U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
- How much
- 28600
Class I
Standard BodySet with Needleless Connectors, REF: A120-125XS
Z-2585-2020 · Class I · reported 2020-07-22 · Terminated
- Why it was recalled
- Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
- Product
- Standard BodySet with Needleless Connectors, REF: A120-125XS
- Where it went
- U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
- How much
- 36286
Class I
BodyGuard Microset with Male Luer Connectors, REF: A100-163XEBF
Z-2581-2020 · Class I · reported 2020-07-22 · Terminated
- Why it was recalled
- Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
- Product
- BodyGuard Microset with Male Luer Connectors, REF: A100-163XEBF
- Where it went
- U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
- How much
- 8016
Class I
BodyGuard Microset with Filter, REF: A100-164XESF
Z-2572-2020 · Class I · reported 2020-07-22 · Terminated
- Why it was recalled
- Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
- Product
- BodyGuard Microset with Filter, REF: A100-164XESF
- Where it went
- U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
- How much
- 55455
Class I
BodyGuard Microset, REF: A100-163XE90S
Z-2573-2020 · Class I · reported 2020-07-22 · Terminated
- Why it was recalled
- Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
- Product
- BodyGuard Microset, REF: A100-163XE90S
- Where it went
- U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
- How much
- 390230
Class I
BodyGuard Microset, REF: A120-003XS2YV
Z-2575-2020 · Class I · reported 2020-07-22 · Terminated
- Why it was recalled
- Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
- Product
- BodyGuard Microset, REF: A120-003XS2YV
- Where it went
- U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
- How much
- 169856
Class I
BodyGuard Microset w/ 0.2 mic filter and lower y-site with female luer, REF: A120-160XYBF
Z-2584-2020 · Class I · reported 2020-07-22 · Terminated
- Why it was recalled
- Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
- Product
- BodyGuard Microset w/ 0.2 mic filter and lower y-site with female luer, REF: A120-160XYBF
- Where it went
- U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
- How much
- 6016
Class I
BodyGuard Microset, REF: A120-112XSFK
Z-2559-2020 · Class I · reported 2020-07-22 · Terminated
- Why it was recalled
- Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
- Product
- BodyGuard Microset, REF: A120-112XSFK
- Where it went
- U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
- How much
- 1273892
Class I
BodyGuard Microset, REF: A120-160XYSF
Z-2561-2020 · Class I · reported 2020-07-22 · Terminated
- Why it was recalled
- Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
- Product
- BodyGuard Microset, REF: A120-160XYSF
- Where it went
- U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
- How much
- 50
Class I
BodyGuard Microset, REF: A100-163XESV
Z-2555-2020 · Class I · reported 2020-07-22 · Terminated
- Why it was recalled
- Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
- Product
- BodyGuard Microset, REF: A100-163XESV
- Where it went
- U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
- How much
- 178057
Class I
BodyGuard Microset with Filter, REF: A100-163XESF
Z-2564-2020 · Class I · reported 2020-07-22 · Terminated
- Why it was recalled
- Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
- Product
- BodyGuard Microset with Filter, REF: A100-163XESF
- Where it went
- U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
- How much
- 95110
Showing the 30 most recent of 48. The rest are searchable at the FDA enforcement report database.
Read these counts with three things in mind
A firm with more recalls usually makes more products. A contract packer runs
hundreds of lines. It will out-recall a company with one product whatever either
does. This is not a safety ranking.
Most recalls here are voluntary. A firm found its own fault and pulled the stock.
That is the system working. It lands in this database looking the same as a recall
the FDA had to push for.
Terminated means the FDA considers the recall finished. It is not a verdict.
It does not mean anyone was harmed.
Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.