Carestream Health Inc.

Carestream Health Inc.: 86 FDA recall reports

Every medical device recall the FDA has published for Carestream Health Inc.. The fault, and how the agency graded it.

86Reports
0Class I
3Ongoing
2023-12-13Most recent

What the classes mean

The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.

ClassWhat the FDA is saying
Class IThere is a reasonable probability that using this will cause serious harm or death.
Class IIIt may cause temporary or reversible harm. Serious harm is unlikely.
Class IIIIt breaks a rule but is unlikely to hurt anyone. Often a labelling fault.

Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.

By class

ClassReports
Class II83
Class III3

By year reported

YearReports
20121
201357
201414
20153
20163
20173
20191
20201
20233
Checking whether it affects you

The lot or code number is what matters. Not the product name. A recall covers specific lots. Yours may not be one of them.

Check the code on your package against the report. The FDA keeps the current list at fda.gov/safety/recalls.

The reports

The text below is quoted from the FDA record word for word.

DRX Revolution Mobile X-Ray System

Z-0437-2024  ·  Class II  ·  reported 2023-12-13  ·  Ongoing

Why it was recalled
Unexpected failure of electrical components within the CPI generator.
Product
DRX Revolution Mobile X-Ray System
Where it went
Worldwide Distribution: US (Nationwide Distribution); and OUS (Foreign): Algeria, Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, Dominican Rep., Finland, France, Germany, Ghana, Great Britain, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Libya, Macau, Malaysia, Maldives, Mexico, Monaco, Morocco, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Republic of Korea, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sri Lanka, State of Palestine, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, & Vietnam.
How much
3,961 units

The DRX-Revolution Mobile X-Ray System is a diagnostic mobile x-ray system utilizing digital radiography (DR)

Z-0289-2024  ·  Class II  ·  reported 2023-11-15  ·  Ongoing

Why it was recalled
There is a potential for unexpected failure of the electrical components within the Carestream Health Inc. generator installed in your DRX-Revolution Mobile X-ray System(s). This can lead to temporary and self contained thermal overload within the generator. Should an event like this occur, the system will become inoperable and loud noise, burnt smell and smoke may be detectable.
Product
The DRX-Revolution Mobile X-Ray System is a diagnostic mobile x-ray system utilizing digital radiography (DR) technology. the system consists of a self-contained x-ray generator, image receptor(s), imaging display and software for acquiring medical diagnostic images outside of standard stationary x-ray room. The DRX-Revolution system incorporates a flat-panel detector that can be used wirelessly for exams such as in-bed chest projections. The system can also be used to expose CR phosphor screens or films.
Where it went
US Nationwide. Global Distribution.
How much
2402 systems

DRX-Compass/DR-FIT X-ray Systems with Firmware version 2.3.2.0/2.3.2.4. For use in obtaining diagnostic quali

Z-1206-2023  ·  Class II  ·  reported 2023-03-08  ·  Ongoing

Why it was recalled
After pressing and releasing the Z-Axis Motorized buttons on the Tube Head Display, the Overhead Tube Crane (OTC) could potentially continue the movement unexpectedly to a certain distance which could potentially result in a injury
Product
DRX-Compass/DR-FIT X-ray Systems with Firmware version 2.3.2.0/2.3.2.4. For use in obtaining diagnostic quality radiographic images to aid the physician with diagnosis. CATALOG #(s): 8624181 X-RAY SYSTEM CONSOLE PDU 3P 380-480 C/F 8624199 X-RAY SYSTEM CONSOLE PDU 3P 208-240 C/F 8621617 X-RAY SYSTEM CONSOLE PDU 3P 380-480 C/F 8622359 X-RAY SYSTEM CONSOLE PDU 3P 208-240 C/F Note: PDU is a standalone Power Distribution Unit (PDU), plugged into the hospital outlet, which supplies Alternate Current (AC) to the Overhead Tube Crane (OTC).
Where it went
Worldwide distribution - US Nationwide distribution in the states of California, Idaho, Illinois, Massachusetts, Minnesota, Nevada, North Carolina, Texas Wisconsin and the countries of Argentina, Australia, Brazil, China, Finland, France, INDIA, ITALY, Korea, Malaysia, Poland, Republic of Korea.
How much
72 units

Kodak DirectView DR3000/3500 - Product Usage: permanently installed as a diagnostic system intended to generat

Z-0806-2020  ·  Class II  ·  reported 2020-01-22  ·  Terminated

Why it was recalled
Carestream Health has discovered a potential safety problem that can lead to unintended movement of the U-Arm.
Product
Kodak DirectView DR3000/3500 - Product Usage: permanently installed as a diagnostic system intended to generate and control x-rays for examination of various anatomical regions. The DR Operator Console is the user interface. All operator functions except positioning the Bucky assembly and controls for the collimator shutter are accomplished at the operator console.
Where it went
Worldwide distribution in the sate of Iowa and countries of Italy, China, Spain, and Austria.
How much
79

OnSight 3D Extremity System- X-Ray, Tomography Computed System Catalog Numbers: 1743566 (w/ Attached Console

Z-1207-2019  ·  Class II  ·  reported 2019-05-01  ·  Terminated

Why it was recalled
When the user performs the re-assignment of a parent / companion pair, the parent volume is transferred to the new patient but the companion volume will remain in the original patient s exam.
Product
OnSight 3D Extremity System- X-Ray, Tomography Computed System Catalog Numbers: 1743566 (w/ Attached Console) 1743574 (w/ Remote Console) The device is intended to be used for x-ray computed tomography and projection x-ray imaging of upper and lower extremities of adult patients and pediatric patients aged 12 and over.
Where it went
Worldwide Distribution: US (nationwide) to states of: Arizona, California, New York, Alabama, Georgia, Georgia, Illinois, Minnesota, New York, Oregon, Pennsylvania; and countries of: IRAN, ITALY, JERUSALEM, KOREA, PHILIPPINES, PORTUGAL,SAUDI ARABIA, SINGAPORE, SWITZERLAND THAILAND, and UNITED KINGDOM.
How much
37 units

DRX-Revolution Mobile X-Ray System; MODEL DRXR-1; Catalog # 1019397 (all other countries) and Catalog 1060177

Z-2651-2017  ·  Class II  ·  reported 2017-07-05  ·  Terminated

Why it was recalled
Carestream found a single unit where four screws in the assembly were not applied in the manufacturing process. This could result in separation of the tube and yoke assembly.
Product
DRX-Revolution Mobile X-Ray System; MODEL DRXR-1; Catalog # 1019397 (all other countries) and Catalog 1060177 (China only) -- COMMON/USUAL NAME: DRX-Revolution --- The DRX-Revolution is a mobile system used to generate and control X-Rays for diagnostic procedures
Where it went
Worldwide Distribution - US (nationwide) Internationally to Australia, Brazil, Canada, Chile, China, Colombia, Dominican Republic, India, Japan, Israel, Korea, Malaysia, Maldives, Mexico, New Zealand, Peru, Puerto Rico, Singapore, Switzerland, Taipei, Thailand, Turkey, and UAE.
How much
Domestic: 1449 units; Foreign: 1416 units

Carestream DRX-Revolution Mobile X-Ray System, Model DRXR-1, Catalog No. 1019397 (all other countries), Catalo

Z-1138-2017  ·  Class II  ·  reported 2017-02-15  ·  Terminated

Why it was recalled
The firm received a complaint alleging that a DRX Revolution System could not be stopped when applying the brake. The Revolution crashed into an elevator.
Product
Carestream DRX-Revolution Mobile X-Ray System, Model DRXR-1, Catalog No. 1019397 (all other countries), Catalog No. 1060177 (China only) -- Carestream Health, Inc. 150 Verona Street, Rochester, NY 14608
Where it went
Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico, and to the countries of : Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Denmark, Dominican Republic, Dubai, Finland, France, Germany, Ghana, India, Iran, Ireland, Israel, Italy, Japan, Korea, Kuwait, Lebanon, Libya, Malaysia, Maldives, Mexico, Morocco, Netherlands, New Zealand, Norway, Oman, Peru, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates and United Kingdom.
How much
US: 1403 units; Foreign: 1370 units

Carestream Touch Prime, Catalog # 1738830, and Touch Prime XE, Catalog # 1738822, Ultrasound System Diagnosti

Z-1052-2017  ·  Class II  ·  reported 2017-01-25  ·  Terminated

Why it was recalled
Software error; Carestream Health Inc, received a complaint stating that when a user accidentally obtains a measurement value of 0 and corrects the value in the report, the resulting measurement unit is not displayed, i.e., centimeters or millimeters. As such, the user expects that the measurement is taken calculated in centimeters, consistent with other values in the report. In actuality, the measurement is taken in millimeters. When this updated measurement is used in an average calculation, the result appears incorrect as two measurements are interpreted as centimeters while the user corrected value is interpreted as millimeters. If the user selects a Calc Result display as Min or Max, the values are also interpreted as millimeters when centimeters were expected.
Product
Carestream Touch Prime, Catalog # 1738830, and Touch Prime XE, Catalog # 1738822, Ultrasound System Diagnostic ultrasound imaging or fluid flow analysis of the human body
Where it went
Worldwide Distribution - US including SC, GA, TX, OH, IA, NY and foreign distribution to Switzerland and Malaysia
How much
US: 14 units; Foreign: 16 units

Carestream Touch Prime, Catalog Number 1738830, and Carestream Touch Prime XE, Catalog Number 1738822 --- Com

Z-2217-2016  ·  Class II  ·  reported 2016-07-27  ·  Terminated

Why it was recalled
A software issue related to the generic volume measurement functionality could result in the volume of the subject anatomy being overestimated. For example, measurement results of the anatomy may appear enlarged or distended when it actually is not.
Product
Carestream Touch Prime, Catalog Number 1738830, and Carestream Touch Prime XE, Catalog Number 1738822 --- Common Name: Touch Ultrasound Diagnostic ultrasound imaging or fluid flow analysis of the human body
Where it went
Worldwide Distribution: US Distribution to states of: GA, IA, and TX; and country of: Italy.
How much
7 Units (US: 5 units, OUS: 2 units)

CARESTREAM Image Suite V4; Image Suite V4: DICOM STORE SCP: REF/Catalog # 1056191, DICOM STORE SCP/FOR IMAGE

Z-1594-2016  ·  Class II  ·  reported 2016-05-11  ·  Terminated

Why it was recalled
Carestream Health received a complaint related to CARESTREAM Image Suite 4 from a foreign hospital stating that the annotation on the overlay is displayed as "L (Left)", when it should be "R(Right)".
Product
CARESTREAM Image Suite V4; Image Suite V4: DICOM STORE SCP: REF/Catalog # 1056191, DICOM STORE SCP/FOR IMAGE SUITE V4: REF/Catalog # 6566988; Image Suite V4 Bundles: IMAGESUITE STANDALONE PACS: REF/Catalog # 1741289, IMAGESUITE STANDALONE PACS FOR INDIA: REF/Catalog # 1741297; Carestream PRO (Image Suite V4): PRO Medical Wireless GOS System-Desktop: REF/Catalog # 1741891, PRO Wireless System Laptop: REF/Catalog # 1741925, PRO Wireless System - w/o Computer: REF/Catalog # 1741933, PRO Tethered System Desktop: REF/Catalog # 1741941, PRO Tethered System Laptop: REF/Catalog # 1741958, PRO Tether System - w/o Computer: REF/Catalog # 1741966, PRO Fixed System - w/o Computer: REF/Catalog # 1741974, PRO Medical Wireless CsI System-Desktop: REF/Catalog # 1741982, PRO Wireless System Laptop: REF/Catalog # 1742006, PRO Wireless System - w/o Computer: REF/Catalog # 1742014, PRO Tethered System Desktop: REF/Catalog # 1742022, PRO Tethered System Laptop: REF/Catalog # 1742055, PRO Tether System - w/o Computer: Catalog # 1742063; PRO Fixed System - w/o Computer: REF/Catalog # 1742089 -- Made in U.S.A. by: Carestream Health, Inc., 150 Verona Street, Rochester, NY 14608 --- CLASSIFICATION...
Where it went
NY and OH for further distribution nationwide; Foreign distribution to the following countries: Afghanistan, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia & Herzgovena, Brazil, Bulgaria, Cambodia, Canada, Chad, Chile, China, Colombia, Croatia, Czech Rep., Djibouti, France, Germany, Greece, Hungary, India, Indonesia, Iran, Iraq, Ireland, Italy, Jamaica, Japan, Jordan, Kuwait, Kyrgyzstan, Laos, Latvia, Lebanon, Liechtenstein, Lithuania, Macao, Malaysia, Mexico, Montenegro, Myanmar (Burma), Nepal, Netherlands, New Zealand, Niger, Norway, Oman, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Scotland, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Taiwan, thailand, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Vietnam.
How much
US: 37 units, Foreign: 269 units

CARESTREAM Image Suite V3: MINI-PACS/F IMG ST/CLASSIC: REF/Catalog # 1036490; MINI-PACS/F IMG ST/POC&VITA: RE

Z-1593-2016  ·  Class II  ·  reported 2016-05-11  ·  Terminated

Why it was recalled
Carestream Health received a complaint related to CARESTREAM Image Suite 4 from a foreign hospital stating that the annotation on the overlay is displayed as "L (Left)", when it should be "R(Right)".
Product
CARESTREAM Image Suite V3: MINI-PACS/F IMG ST/CLASSIC: REF/Catalog # 1036490; MINI-PACS/F IMG ST/POC&VITA: REF/Catalog # 1036508; MINI-PACS/F IMG ST/CLASSIC/INDIA: REF/Catalog # 1036417; MINI-PACS/F IMG ST/ POC&VITA /INDIA: REF/Catalog # 1036425 -- Made in USA by: Carestream Health, Inc. 150 Verona Street, Rochester, NY 14608 --- CLASSIFICATION NAME: System, Image Processing, Radiological The Carestream Image Suite System is an image management system whose intended use is to receive, process, review, display, print and archive images and data from CR and DR modalities. This excludes mammography applications in the United States.
Where it went
NY and OH for further distribution nationwide; Foreign distribution to the following countries: Afghanistan, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia & Herzgovena, Brazil, Bulgaria, Cambodia, Canada, Chad, Chile, China, Colombia, Croatia, Czech Rep., Djibouti, France, Germany, Greece, Hungary, India, Indonesia, Iran, Iraq, Ireland, Italy, Jamaica, Japan, Jordan, Kuwait, Kyrgyzstan, Laos, Latvia, Lebanon, Liechtenstein, Lithuania, Macao, Malaysia, Mexico, Montenegro, Myanmar (Burma), Nepal, Netherlands, New Zealand, Niger, Norway, Oman, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Scotland, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Taiwan, thailand, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Vietnam.
How much
US: 50 units; Foreign: 2499 units

CARESTREAM DRX-EVOLUTION X-Ray System Product Usage: The device is a permanently installed diagnostic X-r

Z-2449-2015  ·  Class II  ·  reported 2015-09-02  ·  Terminated

Why it was recalled
Software defect that impacts image alignment when using the automatic stitching option for long length images. The defect could cause stitching inconsistencies which may go undetected on a radiograph.
Product
CARESTREAM DRX-EVOLUTION X-Ray System Product Usage: The device is a permanently installed diagnostic X-ray system for general radiographic x-ray imaging including tomography. The tomography feature is not to be used for imaging pediatric patients.
Where it went
Worldwide Distribution - US Nationwide and in the countries: Australia, Canada, Chile, China, India, Korea, Malaysia, New Zealand, Singapore, Switzerland, Taiwan and Thailand.
How much
Domestic: 76 units, Foreign: 160 units

CS 8100, CATALOG # (s): 5303045, 5311129, 5303060, 5303110 Intended to produce complete or segmented tomogr

Z-1377-2015  ·  Class II  ·  reported 2015-04-15  ·  Terminated

Why it was recalled
Units device head descended unexpectedly
Product
CS 8100, CATALOG # (s): 5303045, 5311129, 5303060, 5303110 Intended to produce complete or segmented tomographic digital panoramic X-ray images to be used at the direction of healthcare professionals of the dento-maxillo-facial region of the human anatomy as diagnostic support for pediatric and adult patient.
Where it went
Worldwide Distribution -- US, including the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NV, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, and WV; and the countries of India and Switzerland.
How much
3,110 units (863 domestically & 2,247 internationally)

CS 8100 3D, CATALOG # (s): 5311162, 5311188. Intended to produce complete or segmented tomographic digital

Z-1378-2015  ·  Class II  ·  reported 2015-04-15  ·  Terminated

Why it was recalled
Units device head descended unexpectedly
Product
CS 8100 3D, CATALOG # (s): 5311162, 5311188. Intended to produce complete or segmented tomographic digital panoramic X-ray images and three-dimensional digital X-ray images of the dentomaxillo-facial area to be used at the direction of healthcare professionals as diagnostic support for pediatric and adult patient.
Where it went
Worldwide Distribution -- US, including the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NV, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, and WV; and the countries of India and Switzerland.
How much
470 units (127 domestically & 343 internationally)

DRX-Revolution Mobile X-Ray System, MODEL DRXR-1, Service Code 1243, Catalog Number 1060177 (China) and 101939

Z-0800-2015  ·  Class II  ·  reported 2014-12-24  ·  Terminated

Why it was recalled
While servicing a unit at a customer site, a field engineer discovered a broken bolt from the carriage assembly. Carestream has identified an issue related to the DRX-Revolution Mobile X-Ray System in which bolts that secure the x-ray tube assembly/boom to the device mounting block may break. It has been determined that this issue is a result of an improperly assembled part which holds the x-ray
Product
DRX-Revolution Mobile X-Ray System, MODEL DRXR-1, Service Code 1243, Catalog Number 1060177 (China) and 1019397 (all other countries) --- Carestream Health, Inc. The DRX-Revolution Mobile X-Ray System is a mobile imaging system that incorporates a self contained x-ray generator, image receptor, imaging display and software for acquiring medical diagnostic images outside of a standard x-ray room. The DRX-Revolution Mobile X-Ray System has a motor for driving the system so that the operator does not have to push the system, but can walk behind the system while it is moving from one location to another.
Where it went
Worldwide Distribution: US (nationwide) and countries of: AUSTRALIA, BELGIUM, CANADA, CHILE, CHINA, DENMARK, FINLAND, FRANCE, GERMANY, ITALY, JAPAN, KOREA, KUWAIT, MALAYSIA, NETHERLANDS, NEW ZEALAND, NORWAY, PALESTINE, PHILIPPINES, POLAND, PORTUGAL, SAUDI ARABIA, SINGAPORE, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, and UNITED KINGDOM.
How much
629 units (Domestic: 358 units; Foreign: 271 units)

CARESTREAM DIRECTVIEW CR Software Generates digital mammographic images that can be used for screening and

Z-0820-2015  ·  Class II  ·  reported 2014-12-24  ·  Terminated

Why it was recalled
Reduced mammographic image quality when attempting to print true size multi-format images
Product
CARESTREAM DIRECTVIEW CR Software Generates digital mammographic images that can be used for screening and diagnosis of breast cancer.
Where it went
Worldwide Distribution: US distribution to states of : NY, OH, and NC; and countries of: Argentina, Brazil, China, Chile, Colombia, France, Greece, Hong Kong, India, Japan, Korea, Mexico, Peru, Spain, Taiwan, and Thailand.
How much
59 units

Kodak DirectView DR 7500 Diagnostic X-Ray System Product Usage: The DR 7500 system is a permanently insta

Z-2177-2014  ·  Class II  ·  reported 2014-08-20  ·  Terminated

Why it was recalled
An on-site evaluation of the equipment parts that attach the Beta Assembly (x-ray tube and collimator) to the Over Head Tube Crane (OTC) Telescope was completed by Carestream Health (CSH) Service Engineering. Inspection of these parts revealed that the tab on the lock washer was not bent into the lock nut recess and because the Beta Assembly detached from the OTC telescope, the lock nut was not sufficiently tightened.
Product
Kodak DirectView DR 7500 Diagnostic X-Ray System Product Usage: The DR 7500 system is a permanently installed diagnostic x-ray system composed of 4 main components: an operator console, over head tube crane, x-ray tube assembly including a collimator and stationary generator. In addition the device can use an x-ray table and /or wallstand/bucky to complete x-ray exposures. The DR 7500 System is a permanently installed diagnostic x-ray system for generation of x-rays for examination of various anatomical regions.
Where it went
Worldwide Distribution - US Nationwide and the countries of (United Kingdom, United Arab Emirates, Turkey, Taiwan, Switzerland, Sweden, Spain, South Africa, Portugal, Poland, Oman, New Zealand, Netherlands, Kuwait, Italy, Israel, Hong Kong, Germany, France, Finland, Denmark, Dubai, Czech Republic, China, Korea, Belgium, Australia, and Amsterdam.
How much
589 systems

INSIGHT Dental Film, Catalog Number/REF 811 0785, KODAK and Carestream DENTAL labels --- Carestream Health Inc

Z-2142-2014  ·  Class III  ·  reported 2014-08-13  ·  Terminated

Why it was recalled
The Indicator Dot on the film packet's outer printed paper labeling was incorrectly positioned relative to the location of the raised dot on the dental intraoral film. This may cause confusion for the dental practitioner. If the processed film is placed on the view box by referencing the film's raised dot, the image orientation will be rotated 180 degrees.
Product
INSIGHT Dental Film, Catalog Number/REF 811 0785, KODAK and Carestream DENTAL labels --- Carestream Health Inc.
Where it went
Worldwide Distribution.
How much
1633 boxes

Ultra-speed Dental Film, Catalog Number/REF 834 8658, KODAK and Carestream DENTAL labels --- Carestream Health

Z-2143-2014  ·  Class III  ·  reported 2014-08-13  ·  Terminated

Why it was recalled
The Indicator Dot on the film packet's outer printed paper labeling was incorrectly positioned relative to the location of the raised dot on the dental intraoral film. This may cause confusion for the dental practitioner. If the processed film is placed on the view box by referencing the film's raised dot, the image orientation will be rotated 180 degrees.
Product
Ultra-speed Dental Film, Catalog Number/REF 834 8658, KODAK and Carestream DENTAL labels --- Carestream Health Inc.
Where it went
Worldwide Distribution.
How much
1633 boxes

KODAK DirectView DR 9000 System, Catalog Numbers 1966688 (US and Canada) and 8161937 (outside the US and Cana

Z-2137-2014  ·  Class II  ·  reported 2014-08-13  ·  Terminated

Why it was recalled
A control system failure may cause the U-Arm to move downward unexpectedly when the Radiology Technologist is positioning the equipment using the "Up" or "Down" control buttons. Downward movement occurs at three times the normal rate of speed and will continue until the device control button is released, the Emergency Stop button is activated or the equipment Bump Sensor contacts an object.
Product
KODAK DirectView DR 9000 System, Catalog Numbers 1966688 (US and Canada) and 8161937 (outside the US and Canada) --- A permanently installed diagnostic x-ray system composed of 3 main components: an operator console, U-Arm assembly with x-ray tube including a collimator, and a stationary generator.
Where it went
Worldwide Distribution-USA (nationwide) and the countries of Canada, China, Italy, Germany, Greece, Kuwait, Spain, South Africa, and United Arab Emirates.
How much
Domestic: 100 units; Foreign: 46 units

Model Kodak 2100 Intraoral X-Ray System, Catalog Numbers 5303144, 8011579 -- Carestream Health, Inc. 150 Veron

Z-1127-2014  ·  Class II  ·  reported 2014-03-12  ·  Terminated

Why it was recalled
Carestream received problem reports relating to apparent early failure of Kodak 2X00 Intraoral X-ray device arms leading to arm breakage.
Product
Model Kodak 2100 Intraoral X-Ray System, Catalog Numbers 5303144, 8011579 -- Carestream Health, Inc. 150 Verona Street Rochester, NY 14608 -- Made in France -- TROPHY 77435 Marne La Vallee --- COMMON/USUAL NAME: Dental X-Ray Systems -- CLASSIFICATION NAME: Unit, X-Ray, Extraoral with timer --- There are four major components: x-ray tube, x-ray generator, collimator and image receptor. These systems are permanently installed diagnostic dental x-ray systems for generation of x-rays for examination of the teeth, jaw and oral structures.
Where it went
Worldwide distribution. USA (nationwide) and the country of Canada.
How much
Domestic: 347 units (total for both)

Model Kodak 2200 Intraoral X-Ray System, Catalog Numbers 5303177, 5303169, 5303151, XR-5154323 -- Carestream H

Z-1128-2014  ·  Class II  ·  reported 2014-03-12  ·  Terminated

Why it was recalled
Carestream received problem reports relating to apparent early failure of Kodak 2X00 Intraoral X-ray device arms leading to arm breakage.
Product
Model Kodak 2200 Intraoral X-Ray System, Catalog Numbers 5303177, 5303169, 5303151, XR-5154323 -- Carestream Health, Inc. Made in France -- --- COMMON/USUAL NAME: Dental X-Ray Systems -- CLASSIFICATION NAME: Unit, X-Ray, Extraoral with timer --- There are four major components: x-ray tube, x-ray generator, collimator and image receptor. These systems are permanently installed diagnostic dental x-ray systems for generation of x-rays for examination of the teeth, jaw and oral structures.
Where it went
Worldwide distribution. USA (nationwide) and the country of Canada.
How much
Domestic: 347 units (total for both)

Carestream DRX Evolution System (Standard Q) The DRX-Evolution System (Standard Q) is a permanently install

Z-0774-2014  ·  Class II  ·  reported 2014-02-05  ·  Terminated

Why it was recalled
Carestream received a report of an exposure being initiated without a patient image captured. The patient was imaged 2 times before an image was captured. The field engineer who visited the site reported the system allowed an exposure without the assignment of a bucky or image receptor.
Product
Carestream DRX Evolution System (Standard Q) The DRX-Evolution System (Standard Q) is a permanently installed diagnostic x-ray system for generation of x-rays for examination of various anatomical regions. These products are permanently installed diagnostic x-ray systems composed of 4 main components: an operator console, over head tube crane, x-ray tube assembly including a collimator and stationary generator. In addition the device can use an x-ray table and /or wallstandbucky to complete x-ray exposures.
Where it went
Nationwide Distribution including TX, IL, and GA.
How much
4 units

KODAK DirectView DR 7500 Dual Detector System, MODEL DR 7500, Manufactured by Carestream Health, Inc. Made in

Z-0844-2014  ·  Class II  ·  reported 2014-02-05  ·  Terminated

Why it was recalled
Carestream Health, Inc. has recalled DR 7500 Dual Detector System utilizing Version 5.X Software due to a possible patient image display error.
Product
KODAK DirectView DR 7500 Dual Detector System, MODEL DR 7500, Manufactured by Carestream Health, Inc. Made in U.S.A.
Where it went
worldwide Distribution - US (Nationwide) including states of: IN, PA, MD, MI, NY, GA, MN, OK, VA, TX, WI, CA, IL, TN, WA, NC, SC, MA, NJ, and KY; and Internationally to: Canada, New Zealand, South Korea, Australia, Hong Kong, Saudia Arabia, Guinea, Kuwait, Israel, Oman, South Africa, China, UK, Sweden, Poland, Denmark, Germany, Finland, Czech Republic, Italy, France, Portugal and Spain.
How much
164 units Total (42 domestically & 122 internationally)

DR 7500 with FF WS These products are permanently installed diagnostic x-ray systems composed of 4 main com

Z-0798-2014  ·  Class II  ·  reported 2014-01-29  ·  Terminated

Why it was recalled
Carestream Health Inc. has issued a recall for the Carestream DRX-Evolution (FF WS and CSJ WS) and the Kodak DirectView DR 7500 System due to the potential for unexpected device movement.
Product
DR 7500 with FF WS These products are permanently installed diagnostic x-ray systems composed of 4 main components: an operator console, over head tube crane, x-ray tube assembly including a collimator and stationary generator. In addition the device can use an x-ray table and /or wallstand/bucky to complete x-ray exposures. The DRX-Evolution system is a permanently installed diagnostic x-ray system for generation of x-rays for examination of various anatomical regions.
Where it went
Worldwide Distribution: US (nationwide) including states of: CA, CO, FL, GA, IL,IN, MD, MI, MN, NC, ND, NJ, NY, OH, OK, PA, SC, SD, TN, TX, VA, WA, and WI; and internationally to: Australia Belgium, Canada, China, Finland, France, Germany, Hong Kong, Italy, Netherlands, New Zealand, Poland, Portugal, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, and United Kingdom.
How much
493 units Total (107 units domestically & 386 units internationally)

DRX-Evolution with CSH WS These products are permanently installed diagnostic x-ray systems composed of 4 m

Z-0797-2014  ·  Class II  ·  reported 2014-01-29  ·  Terminated

Why it was recalled
Carestream Health Inc. has issued a recall for the Carestream DRX-Evolution (FF WS and CSJ WS) and the Kodak DirectView DR 7500 System due to the potential for unexpected device movement.
Product
DRX-Evolution with CSH WS These products are permanently installed diagnostic x-ray systems composed of 4 main components: an operator console, over head tube crane, x-ray tube assembly including a collimator and stationary generator. In addition the device can use an x-ray table and /or wallstand/bucky to complete x-ray exposures. The DRX-Evolution system is a permanently installed diagnostic x-ray system for generation of x-rays for examination of various anatomical regions.
Where it went
Worldwide Distribution: US (nationwide) including states of: CA, CO, FL, GA, IL,IN, MD, MI, MN, NC, ND, NJ, NY, OH, OK, PA, SC, SD, TN, TX, VA, WA, and WI; and internationally to: Australia Belgium, Canada, China, Finland, France, Germany, Hong Kong, Italy, Netherlands, New Zealand, Poland, Portugal, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, and United Kingdom.
How much
840 units Total (181 units domestically & 659 units internationally)

DRX-Evolution with FF WS These products are permanently installed diagnostic x-ray systems composed of 4 ma

Z-0796-2014  ·  Class II  ·  reported 2014-01-29  ·  Terminated

Why it was recalled
Carestream Health Inc. has issued a recall for the Carestream DRX-Evolution (FF WS and CSJ WS) and the Kodak DirectView DR 7500 System due to the potential for unexpected device movement.
Product
DRX-Evolution with FF WS These products are permanently installed diagnostic x-ray systems composed of 4 main components: an operator console, over head tube crane, x-ray tube assembly including a collimator and stationary generator. In addition the device can use an x-ray table and /or wallstand/bucky to complete x-ray exposures. The DRX-Evolution system is a permanently installed diagnostic x-ray system for generation of x-rays for examination of various anatomical regions.
Where it went
Worldwide Distribution: US (nationwide) including states of: CA, CO, FL, GA, IL,IN, MD, MI, MN, NC, ND, NJ, NY, OH, OK, PA, SC, SD, TN, TX, VA, WA, and WI; and internationally to: Australia Belgium, Canada, China, Finland, France, Germany, Hong Kong, Italy, Netherlands, New Zealand, Poland, Portugal, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, and United Kingdom.
How much
675 units Total (145 units domestically & 530 units internationally)

Carestream DRX-Revolution Mobile X-Ray System -- Made in U.S.A. by Carestream Health Inc. 150 Verona Street, R

Z-0491-2014  ·  Class II  ·  reported 2014-01-15  ·  Terminated

Why it was recalled
It was discovered that Carestream received a report from a customer that 2 different patients were re-exposed due to the fact that the processed images were not able to be found in the system.
Product
Carestream DRX-Revolution Mobile X-Ray System -- Made in U.S.A. by Carestream Health Inc. 150 Verona Street, Rochester, NY . 14608 --- The DRX Revolution Mobile system is a mobile diagnostic x-ray system for generation of x-rays for examination of various anatomical regions. These products are mobile diagnostic x-ray systems designed to perform radiographic x-ray examinations on all pediatric and adult patients, in all patient treatment areas.
Where it went
Nationwide
How much
310 systerms domestically

Carestream Vue PACS; MX WEB SVR W/3 MX-RAD LIC<15K EXAMS The CARESTREAM PACS is an image management system wh

Z-0225-2014  ·  Class II  ·  reported 2013-11-20  ·  Terminated

Why it was recalled
Carestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a software problem.
Product
Carestream Vue PACS; MX WEB SVR W/3 MX-RAD LIC<15K EXAMS The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute, retrieve and display images and data from all hospital modalities and information systems.
Where it went
Worldwide Distribution - USA including AR, AZ, CA, CO, CT, FL, IA, IL, MA, MD, ME, MN, MO, NC, NH, NV, NY, OH, OK, PA, RI, TN, UT, VA WA, and WI and Internationally to Canada, Japan, and Kuwait.

Carestream Vue PACS; UNLIMITED READING LIC ADD 25K EX/YR The CARESTREAM PACS is an image management system wh

Z-0229-2014  ·  Class II  ·  reported 2013-11-20  ·  Terminated

Why it was recalled
Carestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a software problem.
Product
Carestream Vue PACS; UNLIMITED READING LIC ADD 25K EX/YR The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute, retrieve and display images and data from all hospital modalities and information systems.
Where it went
Worldwide Distribution - USA including AR, AZ, CA, CO, CT, FL, IA, IL, MA, MD, ME, MN, MO, NC, NH, NV, NY, OH, OK, PA, RI, TN, UT, VA WA, and WI and Internationally to Canada, Japan, and Kuwait.

Showing the 30 most recent of 86. The rest are searchable at the FDA enforcement report database.

Read these counts with three things in mind

A firm with more recalls usually makes more products. A contract packer runs hundreds of lines. It will out-recall a company with one product whatever either does. This is not a safety ranking.

Most recalls here are voluntary. A firm found its own fault and pulled the stock. That is the system working. It lands in this database looking the same as a recall the FDA had to push for.

Terminated means the FDA considers the recall finished. It is not a verdict. It does not mean anyone was harmed.