Boston Scientific Corporation

Boston Scientific Corporation: 629 FDA recall reports

Every medical device recall the FDA has published for Boston Scientific Corporation. The fault, and how the agency graded it.

629Reports
138Class I
409Ongoing
2026-07-29Most recent
138 Class I on this page

Class I means the FDA judged there was a reasonable probability of serious harm or death. Check the lot numbers against anything you hold.

What the classes mean

The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.

ClassWhat the FDA is saying
Class IThere is a reasonable probability that using this will cause serious harm or death.
Class IIIt may cause temporary or reversible harm. Serious harm is unlikely.
Class IIIIt breaks a rule but is unlikely to hurt anyone. Often a labelling fault.

Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.

By class

ClassReports
Class II485
Class I138
Class III6

By year reported

YearReports
20126
20139
201423
201517
20167
201718
201855
201938
202067
202155
202252
202331
2024130
202585
202636
Checking whether it affects you

The lot or code number is what matters. Not the product name. A recall covers specific lots. Yours may not be one of them.

Check the code on your package against the report. The FDA keeps the current list at fda.gov/safety/recalls.

The reports

The text below is quoted from the FDA record word for word.

FARADRIVE STEERABLE SHEATH CLEAR, used to facilitate access to the vasculature and into the chambers of the he

Z-2755-2026  ·  Class II  ·  reported 2026-07-29  ·  Ongoing

Why it was recalled
Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.
Product
FARADRIVE STEERABLE SHEATH CLEAR, used to facilitate access to the vasculature and into the chambers of the heart, Material Number 21M402, IFU Reference Number LBL1062
Where it went
Global Distribution. US Nationwide.
How much
74,760 units

FARASTAR Ablation Generator EU, Pulsed Field Ablation Generator, Product ID 61M401; Percutaneous Cardiac Ablat

Z-2753-2026  ·  Class II  ·  reported 2026-07-29  ·  Ongoing

Why it was recalled
Boston Scientific is replacing specific FARASTAR PFA generators which have a configuration that is not compatible for use with the FARAPOINT PFA catheter.
Product
FARASTAR Ablation Generator EU, Pulsed Field Ablation Generator, Product ID 61M401; Percutaneous Cardiac Ablation Catheter
Where it went
International distribution in the countries of Italy, Austria, France, Germany, Belgium.
How much
6 units

FARASTAR Ablation Generator EU, Pulsed Field Ablation Generator, Product ID 61M401; Percutaneous Cardiac Ablat

Z-2750-2026  ·  Class II  ·  reported 2026-07-29  ·  Ongoing

Why it was recalled
Labeling discrepancy: Devices distributed with Instructions For Use (IFU) did not include FARAPOINT PFA catheter information.
Product
FARASTAR Ablation Generator EU, Pulsed Field Ablation Generator, Product ID 61M401; Percutaneous Cardiac Ablation Catheter
Where it went
International distribution in the states of Austria, Spain, France, Germany, Denmark, Italy, Portugal, Great Britain, Ireland.
How much
18 units

FARASTAR Pulsed Field Ablation Generator, Product ID M004PFCE61M401; Percutaneous Cardiac Ablation Catheter

Z-2751-2026  ·  Class II  ·  reported 2026-07-29  ·  Ongoing

Why it was recalled
Labeling discrepancy: Devices distributed with Instructions For Use (IFU) did not include FARAPOINT PFA catheter information.
Product
FARASTAR Pulsed Field Ablation Generator, Product ID M004PFCE61M401; Percutaneous Cardiac Ablation Catheter
Where it went
International distribution in the states of Austria, Spain, France, Germany, Denmark, Italy, Portugal, Great Britain, Ireland.
How much
1 units

FARASTAR PFA Generator, Product ID M004PFCE61M460; Percutaneous Cardiac Ablation Catheter

Z-2752-2026  ·  Class II  ·  reported 2026-07-29  ·  Ongoing

Why it was recalled
Labeling discrepancy: Devices distributed with Instructions For Use (IFU) did not include FARAPOINT PFA catheter information.
Product
FARASTAR PFA Generator, Product ID M004PFCE61M460; Percutaneous Cardiac Ablation Catheter
Where it went
International distribution in the states of Austria, Spain, France, Germany, Denmark, Italy, Portugal, Great Britain, Ireland.
How much
11 units

Boston Scientific TheraSphere 360 Y-90 Management Platform, Model Number M00139509001; Radiological Image Proc

Z-2771-2026  ·  Class II  ·  reported 2026-07-29  ·  Ongoing

Why it was recalled
Software anomaly occurs when a user adds one or more vials to the cart and subsequently returns to the Activity Calculator to modify one or more input. When a user modifies an input and selects "Recalculate", the dose calculation is updated correctly; however, any vials that were previously added to the cart are not automatically updated to reflect the revised calculation.
Product
Boston Scientific TheraSphere 360 Y-90 Management Platform, Model Number M00139509001; Radiological Image Processing System.
Where it went
US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, IL, IN, LA, MA, MN, NE, NY, OH, OK, OR, SC, SD, TN, TX, UT, WA, WI, WV, WY.
How much
68 units

FARADRIVE STEERABLE SHEATH CLEAR OUS, used to facilitate access to the vasculature and into the chambers of th

Z-2757-2026  ·  Class II  ·  reported 2026-07-29  ·  Ongoing

Why it was recalled
Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.
Product
FARADRIVE STEERABLE SHEATH CLEAR OUS, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PFCE21M402, IFU Reference Number 51659727-01
Where it went
Global Distribution. US Nationwide.
How much
193,000 units

FARADRIVE Steerable Sheath Clear IDE, used to facilitate access to the vasculature and into the chambers of th

Z-2754-2026  ·  Class II  ·  reported 2026-07-29  ·  Ongoing

Why it was recalled
Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.
Product
FARADRIVE Steerable Sheath Clear IDE, used to facilitate access to the vasculature and into the chambers of the heart, Material Number 21C602, IFU Reference Number 51636751-01
Where it went
Global Distribution. US Nationwide.
How much
400 units

FARADRIVE STEERABLE SHEATH (CLEAR) - US, used to facilitate access to the vasculature and into the chambers of

Z-2756-2026  ·  Class II  ·  reported 2026-07-29  ·  Ongoing

Why it was recalled
Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.
Product
FARADRIVE STEERABLE SHEATH (CLEAR) - US, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PF21M402, IFU Reference Number 51654495-01
Where it went
Global Distribution. US Nationwide.
How much
508,151 units

Rapid Refill Continuous Injection System, UPN M00566001

Z-2725-2026  ·  Class II  ·  reported 2026-07-22  ·  Ongoing

Why it was recalled
Devices contain recalled Medline Namic Angiographic Control Syringe with Rotating Adaptor. There is a risk of syringe loosening or disconnecting from the manifold, resulting in loose connection or complete disconnection. May lead to difficulty connecting syringe, inability to aspirate or inject properly, fluid leakage, or separation of the syringe from the system, potentially requiring device exchange and prolonging the procedure.
Product
Rapid Refill Continuous Injection System, UPN M00566001
Where it went
US Nationwide distribution. OUS consignee list pending.
How much
153,921 (US); 119,274 (OUS)

CRE Pro Wireguided 12-15mm 240cm

Z-2629-2026  ·  Class II  ·  reported 2026-07-08  ·  Ongoing

Why it was recalled
Potential sterile breach of the pouches in which devices are packaged.
Product
CRE Pro Wireguided 12-15mm 240cm
Where it went
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Bangladesh, Belgium, Bulgaria, Canada, Chile, China, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Estonia, Finland, France, Germany, Great Britain, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Mongolia, Netherlands, New Zealand, Norway, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Tobago, Trinidad, Tunisia, T¿rkiye, United Arab Emirates, Uruguay.
How much
81,543 units

CRE Pro Wireguided 10-12mm 240cm

Z-2628-2026  ·  Class II  ·  reported 2026-07-08  ·  Ongoing

Why it was recalled
Potential sterile breach of the pouches in which devices are packaged.
Product
CRE Pro Wireguided 10-12mm 240cm
Where it went
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Bangladesh, Belgium, Bulgaria, Canada, Chile, China, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Estonia, Finland, France, Germany, Great Britain, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Mongolia, Netherlands, New Zealand, Norway, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Tobago, Trinidad, Tunisia, T¿rkiye, United Arab Emirates, Uruguay.
How much
81,543 units

CRE Pro Wireguided 15-18mm 240cm

Z-2630-2026  ·  Class II  ·  reported 2026-07-08  ·  Ongoing

Why it was recalled
Potential sterile breach of the pouches in which devices are packaged.
Product
CRE Pro Wireguided 15-18mm 240cm
Where it went
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Bangladesh, Belgium, Bulgaria, Canada, Chile, China, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Estonia, Finland, France, Germany, Great Britain, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Mongolia, Netherlands, New Zealand, Norway, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Tobago, Trinidad, Tunisia, T¿rkiye, United Arab Emirates, Uruguay.
How much
81,543 units

CRE Pro Wireguided 6-8mm 240cm

Z-2626-2026  ·  Class II  ·  reported 2026-07-08  ·  Ongoing

Why it was recalled
Potential sterile breach of the pouches in which devices are packaged.
Product
CRE Pro Wireguided 6-8mm 240cm
Where it went
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Bangladesh, Belgium, Bulgaria, Canada, Chile, China, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Estonia, Finland, France, Germany, Great Britain, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Mongolia, Netherlands, New Zealand, Norway, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Tobago, Trinidad, Tunisia, T¿rkiye, United Arab Emirates, Uruguay.
How much
81,543 units

CRE Wireguided 8-10mm 240cm

Z-2633-2026  ·  Class II  ·  reported 2026-07-08  ·  Ongoing

Why it was recalled
Potential sterile breach of the pouches in which devices are packaged.
Product
CRE Wireguided 8-10mm 240cm
Where it went
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Bangladesh, Belgium, Bulgaria, Canada, Chile, China, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Estonia, Finland, France, Germany, Great Britain, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Mongolia, Netherlands, New Zealand, Norway, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Tobago, Trinidad, Tunisia, T¿rkiye, United Arab Emirates, Uruguay.
How much
81,543 units

CRE Wireguided 12-15mm 240cm

Z-2635-2026  ·  Class II  ·  reported 2026-07-08  ·  Ongoing

Why it was recalled
Potential sterile breach of the pouches in which devices are packaged.
Product
CRE Wireguided 12-15mm 240cm
Where it went
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Bangladesh, Belgium, Bulgaria, Canada, Chile, China, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Estonia, Finland, France, Germany, Great Britain, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Mongolia, Netherlands, New Zealand, Norway, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Tobago, Trinidad, Tunisia, T¿rkiye, United Arab Emirates, Uruguay.
How much
81,543 units

CRE Wireguided 15-18mm 240cm

Z-2636-2026  ·  Class II  ·  reported 2026-07-08  ·  Ongoing

Why it was recalled
Potential sterile breach of the pouches in which devices are packaged.
Product
CRE Wireguided 15-18mm 240cm
Where it went
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Bangladesh, Belgium, Bulgaria, Canada, Chile, China, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Estonia, Finland, France, Germany, Great Britain, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Mongolia, Netherlands, New Zealand, Norway, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Tobago, Trinidad, Tunisia, T¿rkiye, United Arab Emirates, Uruguay.
How much
81,543 units

CRE Wireguided 6-8mm 240cm

Z-2632-2026  ·  Class II  ·  reported 2026-07-08  ·  Ongoing

Why it was recalled
Potential sterile breach of the pouches in which devices are packaged.
Product
CRE Wireguided 6-8mm 240cm
Where it went
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Bangladesh, Belgium, Bulgaria, Canada, Chile, China, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Estonia, Finland, France, Germany, Great Britain, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Mongolia, Netherlands, New Zealand, Norway, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Tobago, Trinidad, Tunisia, T¿rkiye, United Arab Emirates, Uruguay.
How much
81,543 units

CRE Wireguided 18-20mm 240cm

Z-2637-2026  ·  Class II  ·  reported 2026-07-08  ·  Ongoing

Why it was recalled
Potential sterile breach of the pouches in which devices are packaged.
Product
CRE Wireguided 18-20mm 240cm
Where it went
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Bangladesh, Belgium, Bulgaria, Canada, Chile, China, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Estonia, Finland, France, Germany, Great Britain, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Mongolia, Netherlands, New Zealand, Norway, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Tobago, Trinidad, Tunisia, T¿rkiye, United Arab Emirates, Uruguay.
How much
81,543 units

CRE Pro Wireguided 8-10mm 240cm

Z-2627-2026  ·  Class II  ·  reported 2026-07-08  ·  Ongoing

Why it was recalled
Potential sterile breach of the pouches in which devices are packaged.
Product
CRE Pro Wireguided 8-10mm 240cm
Where it went
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Bangladesh, Belgium, Bulgaria, Canada, Chile, China, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Estonia, Finland, France, Germany, Great Britain, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Mongolia, Netherlands, New Zealand, Norway, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Tobago, Trinidad, Tunisia, T¿rkiye, United Arab Emirates, Uruguay.
How much
81,543 units

CRE Wireguided 10-12mm 240cm

Z-2634-2026  ·  Class II  ·  reported 2026-07-08  ·  Ongoing

Why it was recalled
Potential sterile breach of the pouches in which devices are packaged.
Product
CRE Wireguided 10-12mm 240cm
Where it went
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Bangladesh, Belgium, Bulgaria, Canada, Chile, China, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Estonia, Finland, France, Germany, Great Britain, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Mongolia, Netherlands, New Zealand, Norway, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Tobago, Trinidad, Tunisia, T¿rkiye, United Arab Emirates, Uruguay.
How much
81,543 units

CRE Pro Wireguided 18-20mm 240cm

Z-2631-2026  ·  Class II  ·  reported 2026-07-08  ·  Ongoing

Why it was recalled
Potential sterile breach of the pouches in which devices are packaged.
Product
CRE Pro Wireguided 18-20mm 240cm
Where it went
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Bangladesh, Belgium, Bulgaria, Canada, Chile, China, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Estonia, Finland, France, Germany, Great Britain, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Mongolia, Netherlands, New Zealand, Norway, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Tobago, Trinidad, Tunisia, T¿rkiye, United Arab Emirates, Uruguay.
How much
81,543 units

Intera Oncology INTERA 3000 Hepatic Artery Infusion Pump, AP-0300H, implantable drug delivery device

Z-2580-2026  ·  Class II  ·  reported 2026-07-01  ·  Completed

Why it was recalled
There is the potential for leakage along the drug pathway from the pump through the end of the catheter.
Product
Intera Oncology INTERA 3000 Hepatic Artery Infusion Pump, AP-0300H, implantable drug delivery device
Where it went
US Nationwide distribution in the state of OH.
How much
1 device

Orca Single Use Air/Water and Suction Valves, Material Number (UPN) SUV-617-50; The Orca Air/Water Valve is in

Z-2413-2026  ·  Class II  ·  reported 2026-06-17  ·  Ongoing

Why it was recalled
Boston Scientific is initiating a medical device removal of Orca Sterile, Single Use Air/Water and Suction Valves due to a higher than anticipated occurrence of the suction button sticking.
Product
Orca Single Use Air/Water and Suction Valves, Material Number (UPN) SUV-617-50; The Orca Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI (gastrointestinal) endoscopic procedure.
Where it went
US Nationwide distribution.
How much
9750 units

LUX-DX II, Arrythmia detector and alarm, Model M302, with SERVER SW LATITUDE DRAGON US, Model 6460, Version 7.

Z-2048-2026  ·  Class II  ·  reported 2026-05-13  ·  Ongoing

Why it was recalled
For some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not collecting PVC Burden data (both M302/M312) or monitoring for Bradyarrhythmia and Pause (M312 only). In addition, for these devices that are not being monitored, the LATITUDE Clarity programming screens have discrepant information, indicating that the monitoring is enabled for these features in one location and not enabled in another location.
Product
LUX-DX II, Arrythmia detector and alarm, Model M302, with SERVER SW LATITUDE DRAGON US, Model 6460, Version 7.5
Where it went
Worldwide distribution - US Nationwide and the country of Belgium.
How much
2 units

LUX-Dx II Plus, Arrythmia detector and alarm, Model M312, with SERVER SW LATITUDE DRAGON EU (Europe), Model 64

Z-2049-2026  ·  Class II  ·  reported 2026-05-13  ·  Ongoing

Why it was recalled
For some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not collecting PVC Burden data (both M302/M312) or monitoring for Bradyarrhythmia and Pause (M312 only). In addition, for these devices that are not being monitored, the LATITUDE Clarity programming screens have discrepant information, indicating that the monitoring is enabled for these features in one location and not enabled in another location.
Product
LUX-Dx II Plus, Arrythmia detector and alarm, Model M312, with SERVER SW LATITUDE DRAGON EU (Europe), Model 6446, Version 7.5, or SERVER SW LATITUDE DRAGON AU (Australia/New Zealand), Model 6448, Version 7.5
Where it went
Worldwide distribution - US Nationwide and the country of Belgium.
How much
148 units
Class I

Boston Scientific Cardiac Resynchronization Therapy (CRT)labeled as: 1. VALITUDE CRT-P EL, Model Number U12

Z-1771-2026  ·  Class I  ·  reported 2026-05-06  ·  Ongoing

Why it was recalled
Updated software, Brady software maintenance release 6 (SMR6), is now available for the ACCOLADE" family of pacemakers and cardiac resynchronization therapy pacemakers (CRT-Ps)1 and the advisory population is expanding to include all CRT-P and dual-chamber extended life (DR-EL) devices. Dual chamber modes are specifically indicated for treatment of the following: " Conduction disorders that require restoration of AV synchrony, including varying degrees of AV block " VVI intolerance (i.e., pacemaker syndrome) in the presence of persistent sinus rhythm " Low cardiac output or congestive heart failure secondary to bradycardia
Product
Boston Scientific Cardiac Resynchronization Therapy (CRT)labeled as: 1. VALITUDE CRT-P EL, Model Number U125; 2. VALITUDE CRT-P EL MRI, Model Number U128; 3. VISIONIST CRT-P EL, Model Number U225; 4. VISIONIST CRT-P EL, Model Number U226; 5. VISIONIST CRT-P EL MRI, Model Number U228.
Where it went
worldwide
How much
685,776 units
Class I

Boston Scientific Pacemaker, Implantable Pulse Generator, Pacemaker (non- Cardiac Re-Synchronization Therapy)

Z-1770-2026  ·  Class I  ·  reported 2026-05-06  ·  Ongoing

Why it was recalled
Updated software, Brady software maintenance release 6 (SMR6), is now available for the ACCOLADE" family of pacemakers and cardiac resynchronization therapy pacemakers (CRT-Ps)1 and the advisory population is expanding to include all CRT-P and dual-chamber extended life (DR-EL) devices. Dual chamber modes are specifically indicated for treatment of the following: " Conduction disorders that require restoration of AV synchrony, including varying degrees of AV block " VVI intolerance (i.e., pacemaker syndrome) in the presence of persistent sinus rhythm " Low cardiac output or congestive heart failure secondary to bradycardia
Product
Boston Scientific Pacemaker, Implantable Pulse Generator, Pacemaker (non- Cardiac Re-Synchronization Therapy) labeled as: 1. ESSENTIO SR SL, Model Number L100; 2. ESSENTIO DR SL Pacemaker, Model Number L101; 3. ESSENTIO SR SL MRI Pacemaker, Model Number L110; 4. ESSENTIO DR SL MRI Pacemaker, Model Number L111; 5. ESSENTIO DR EL Pacemaker, Model Number L121; 6. ESSENTIO DR EL MRI Pacemaker, Model Number L131; 7. PROPONENT SR SL Pacemaker, Model Number L200; 8. PROPONENT DR SL Pacemaker, Model Number L201; 9. PROPONENT DR (VDD) SL Pacemaker, Model Number L209; 10. PROPONENT SR SL Pacemaker, Model Number L210; 11. PROPONENT DR SL MRI Pacemaker, Model Number L211; 12. PROPONENT DR EL Pacemaker, Model Number L221; 13. PROPONENT DR EL MRI Pacemaker, Model Number L231*; 14. ACCOLADE SR SL Pacemaker, Model Number L300; 15. ACCOLADE DR SL Pacemaker, Model Number L301; 16. ACCOLADE SR SL MRI Pacemaker, Model Number L310; 17. ACCOLADE DR SL MRI Pacemaker, Model Number L311; 18. ACCOLADE DR EL Pacemaker, Model Number L321; 19. ACCOLADE DR EL MRI Pacemaker, Model Number L331*; 20. ALTRUA 2 SR SL Pacemaker, Model Number S701; 21. ALTRUA 2 DR SL...
Where it went
worldwide
How much
718,456 units

EndoVive Safety PEG Kit; Outer Box Material Number (UPN): M00509001; Inner Box Material Number (UPN): M00509

Z-1551-2026  ·  Class II  ·  reported 2026-03-18  ·  Ongoing

Why it was recalled
Recall of kits which utilized ChloraPrep Triple Swabsticks from BD. Open or compromised seals on the sterile swabstick packaging, which could result in loss of sterility or drying of the antiseptic solution. This component is placed outside of the sterile barrier of the tray manufactured by Boston Scientific, which contains the majority of the kit components. As such, inclusion of the affected ChloraPrep Triple Swabsticks does not degrade the integrity of the other components within the PEG kit.
Product
EndoVive Safety PEG Kit; Outer Box Material Number (UPN): M00509001; Inner Box Material Number (UPN): M00509000;
Where it went
US Nationwide distribution.
How much
806 units
Class I

HOT AXIOS Stent and Electrocautery-Enhanced Delivery System 6mm x 8mm; Material Number (UPN): M00553520;

Z-1161-2026  ·  Class I  ·  reported 2026-02-11  ·  Ongoing

Why it was recalled
Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most serious reasonably foreseeable outcome in cases where the first flange is unable to deploy or expand would be a cascade of events resulting in additional intervention, either endoscopic or surgical, to remove the stent and close the puncture site.
Product
HOT AXIOS Stent and Electrocautery-Enhanced Delivery System 6mm x 8mm; Material Number (UPN): M00553520;
Where it went
Worldwide distribution. US Nationwide including Puerto Rico; Algeria, Argentina, Austria, Belgium, Brazil, Canada, CHILE, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, JAPAN, Jordan, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Morocco, Netherlands, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, PERU, Poland, Portugal, Romania, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, UAE, and United Kingdom.
How much
266 units

Showing the 30 most recent of 629. The rest are searchable at the FDA enforcement report database.

Read these counts with three things in mind

A firm with more recalls usually makes more products. A contract packer runs hundreds of lines. It will out-recall a company with one product whatever either does. This is not a safety ranking.

Most recalls here are voluntary. A firm found its own fault and pulled the stock. That is the system working. It lands in this database looking the same as a recall the FDA had to push for.

Terminated means the FDA considers the recall finished. It is not a verdict. It does not mean anyone was harmed.

Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.

← All device recalls