BIOTRONIK Inc

BIOTRONIK Inc: 26 FDA recall reports

Every medical device recall the FDA has published for BIOTRONIK Inc. The fault, and how the agency graded it.

26Reports
0Class I
0Ongoing
2021-04-28Most recent

What the classes mean

The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.

ClassWhat the FDA is saying
Class IThere is a reasonable probability that using this will cause serious harm or death.
Class IIIt may cause temporary or reversible harm. Serious harm is unlikely.
Class IIIIt breaks a rule but is unlikely to hurt anyone. Often a labelling fault.

Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.

By class

ClassReports
Class II26

By year reported

YearReports
20151
202125
Checking whether it affects you

The lot or code number is what matters. Not the product name. A recall covers specific lots. Yours may not be one of them.

Check the code on your package against the report. The FDA keeps the current list at fda.gov/safety/recalls.

The reports

The text below is quoted from the FDA record word for word.

BIOTRONIK Ilestro 7 VR-T, DF-1, REF 383578, UDI: 04035479125349 - Product Usage: intended to provide ventricul

Z-1443-2021  ·  Class II  ·  reported 2021-04-28  ·  Terminated

Why it was recalled
There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Product
BIOTRONIK Ilestro 7 VR-T, DF-1, REF 383578, UDI: 04035479125349 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
Where it went
US: Nationwide OUS: Worldwide
How much
1000 UNITS

BIOTRONIK Ilestro 7 VR-T DX, DF-1, REF 383594, UDI: 04035479125509 - Product Usage: intended to provide ventri

Z-1445-2021  ·  Class II  ·  reported 2021-04-28  ·  Terminated

Why it was recalled
There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Product
BIOTRONIK Ilestro 7 VR-T DX, DF-1, REF 383594, UDI: 04035479125509 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
Where it went
US: Nationwide OUS: Worldwide
How much
3664 UNITS

BIOTRONIK INVENTRA 7, HF-T QP DF4 IS4, REF 393012, UDI: 04035479129286 - Product Usage: intended to provide v

Z-1454-2021  ·  Class II  ·  reported 2021-04-28  ·  Terminated

Why it was recalled
There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Product
BIOTRONIK INVENTRA 7, HF-T QP DF4 IS4, REF 393012, UDI: 04035479129286 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
Where it went
US: Nationwide OUS: Worldwide
How much
702 UNITS

BIOTRONIK ILIVIA NEO 7 DR-T DF-1 ProMRI, REF 429529, UDI: 04035479156749 - Product Usage: intended to provide

Z-1464-2021  ·  Class II  ·  reported 2021-04-28  ·  Terminated

Why it was recalled
There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Product
BIOTRONIK ILIVIA NEO 7 DR-T DF-1 ProMRI, REF 429529, UDI: 04035479156749 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
Where it went
US: Nationwide OUS: Worldwide
How much
98 UNITS

BIOTRONIK ITREVIA 7, VR-T DF-4, REF 393041, UDI: 04035479129569 - Product Usage: intended to provide ventricul

Z-1452-2021  ·  Class II  ·  reported 2021-04-28  ·  Terminated

Why it was recalled
There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Product
BIOTRONIK ITREVIA 7, VR-T DF-4, REF 393041, UDI: 04035479129569 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
Where it went
US: Nationwide OUS: Worldwide
How much
670 units

BIOTRONIK Ilestro 7 DR-T, DF-1, REF383562, UDI: 04035479125189 - Product Usage: intended to provide ventricula

Z-1441-2021  ·  Class II  ·  reported 2021-04-28  ·  Terminated

Why it was recalled
There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Product
BIOTRONIK Ilestro 7 DR-T, DF-1, REF383562, UDI: 04035479125189 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
Where it went
US: Nationwide OUS: Worldwide
How much
N/A

BIOTRONIK ILIVIA 7, VR-T DF4 ProMRI, REF 404626, UDI: 04035479142131 - Product Usage: intended to provide ven

Z-1458-2021  ·  Class II  ·  reported 2021-04-28  ·  Terminated

Why it was recalled
There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Product
BIOTRONIK ILIVIA 7, VR-T DF4 ProMRI, REF 404626, UDI: 04035479142131 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
Where it went
US: Nationwide OUS: Worldwide
How much
779 units

BIOTRONIK ITREVIA 7, VR-T DF-1, REF 393040, UDI: 04035479129552 - Product Usage: intended to provide ventricul

Z-1451-2021  ·  Class II  ·  reported 2021-04-28  ·  Terminated

Why it was recalled
There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Product
BIOTRONIK ITREVIA 7, VR-T DF-1, REF 393040, UDI: 04035479129552 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
Where it went
US: Nationwide OUS: Worldwide
How much
499 units

BIOTRONIK IPERIA 7,VR-T, DX, DF-1, ProMRI, REF 393032, UDI: 04035479129477 - Product Usage: intended to provid

Z-1449-2021  ·  Class II  ·  reported 2021-04-28  ·  Terminated

Why it was recalled
There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Product
BIOTRONIK IPERIA 7,VR-T, DX, DF-1, ProMRI, REF 393032, UDI: 04035479129477 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
Where it went
US: Nationwide OUS: Worldwide
How much
3,832 UNITS

BIOTRONIK Ilestro 7 VR-T, DF-4, REF 383580, UDI: 04035479125363 - Product Usage: intended to provide ventricul

Z-1444-2021  ·  Class II  ·  reported 2021-04-28  ·  Terminated

Why it was recalled
There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Product
BIOTRONIK Ilestro 7 VR-T, DF-4, REF 383580, UDI: 04035479125363 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
Where it went
US: Nationwide OUS: Worldwide
How much
323 UNITS

BIOTRONIK Ilestro 7 DR-T, DF-4, REF383564, UDI: 04035479125202 - Product Usage: intended to provide ventricula

Z-1442-2021  ·  Class II  ·  reported 2021-04-28  ·  Terminated

Why it was recalled
There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Product
BIOTRONIK Ilestro 7 DR-T, DF-4, REF383564, UDI: 04035479125202 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
Where it went
US: Nationwide OUS: Worldwide
How much
N/A

BIOTRONIK ILIVIA 7, DR-T DF-1 PROMRI, REF 404622, UDI: 04035479142094 - Product Usage: intended to provide ven

Z-1461-2021  ·  Class II  ·  reported 2021-04-28  ·  Terminated

Why it was recalled
There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Product
BIOTRONIK ILIVIA 7, DR-T DF-1 PROMRI, REF 404622, UDI: 04035479142094 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
Where it went
US: Nationwide OUS: Worldwide
How much
1547 UNITS

BIOTRONIK ITREVIA 7, DR-T QP DF-1, REF 392412, UDI: 04035479128807 - Product Usage: intended to provide ventr

Z-1455-2021  ·  Class II  ·  reported 2021-04-28  ·  Terminated

Why it was recalled
There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Product
BIOTRONIK ITREVIA 7, DR-T QP DF-1, REF 392412, UDI: 04035479128807 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
Where it went
US: Nationwide OUS: Worldwide
How much
658 units

BIOTRONIK Iforia 7,DR-T, DF-1,REF 390064, UDI: 04035479126926 - Product Usage: intended to provide ventricular

Z-1448-2021  ·  Class II  ·  reported 2021-04-28  ·  Terminated

Why it was recalled
There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Product
BIOTRONIK Iforia 7,DR-T, DF-1,REF 390064, UDI: 04035479126926 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
Where it went
US: Nationwide OUS: Worldwide
How much
490 units

BIOTRONIK ILIVIA 7, DR-T DF4 ProMRI, REF 404623, UDI: 04035479142100 - Product Usage: intended to provide ven

Z-1456-2021  ·  Class II  ·  reported 2021-04-28  ·  Terminated

Why it was recalled
There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Product
BIOTRONIK ILIVIA 7, DR-T DF4 ProMRI, REF 404623, UDI: 04035479142100 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
Where it went
US: Nationwide OUS: Worldwide
How much
3497 units

BIOTRONIK INTICA NEO 7 HF-T DF-1 IS-1 ProMRI, REF 429553, UDI: 04035479156831 - Product Usage: intended to pro

Z-1465-2021  ·  Class II  ·  reported 2021-04-28  ·  Terminated

Why it was recalled
There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Product
BIOTRONIK INTICA NEO 7 HF-T DF-1 IS-1 ProMRI, REF 429553, UDI: 04035479156831 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
Where it went
US: Nationwide OUS: Worldwide
How much
8 UNITS

BIOTRONIK Iforia 7, vr-T, DX, DF-1,REF 390090, UDI: 04035479127176 - Product Usage: intended to provide ventri

Z-1447-2021  ·  Class II  ·  reported 2021-04-28  ·  Terminated

Why it was recalled
There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Product
BIOTRONIK Iforia 7, vr-T, DX, DF-1,REF 390090, UDI: 04035479127176 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
Where it went
US: Nationwide OUS: Worldwide
How much
1176 UNITS

BIOTRONIK Iperia 7, DR-T, DF-1, ProMRI REF 392409, UDI: 04035479128777 - Product Usage: intended to provide ve

Z-1446-2021  ·  Class II  ·  reported 2021-04-28  ·  Terminated

Why it was recalled
There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Product
BIOTRONIK Iperia 7, DR-T, DF-1, ProMRI REF 392409, UDI: 04035479128777 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
Where it went
US: Nationwide OUS: Worldwide
How much
1652 UNITS

BIOTRONIK ILIVIA NEO 7 VR-T DF-1 ProMRI, REF 429531, UDI: 04035479156763 - Product Usage: intended to provide

Z-1462-2021  ·  Class II  ·  reported 2021-04-28  ·  Terminated

Why it was recalled
There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Product
BIOTRONIK ILIVIA NEO 7 VR-T DF-1 ProMRI, REF 429531, UDI: 04035479156763 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
Where it went
US: Nationwide OUS: Worldwide
How much
74 UNITS

BIOTRONIK INVENTRA 7, VR-T DX DF-1, REF 399437, UDI: 04035479136321 - Product Usage: intended to provide vent

Z-1460-2021  ·  Class II  ·  reported 2021-04-28  ·  Terminated

Why it was recalled
There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Product
BIOTRONIK INVENTRA 7, VR-T DX DF-1, REF 399437, UDI: 04035479136321 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
Where it went
US: Nationwide OUS: Worldwide
How much
285 Unit

BIOTRONIK INTICA NEO 7 VR-T DX DF-1 ProMRI, REF 429559, UDI: 04035479156855 - Product Usage: intended to provi

Z-1463-2021  ·  Class II  ·  reported 2021-04-28  ·  Terminated

Why it was recalled
There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Product
BIOTRONIK INTICA NEO 7 VR-T DX DF-1 ProMRI, REF 429559, UDI: 04035479156855 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
Where it went
US: Nationwide OUS: Worldwide
How much
17 UNITS

BIOTRONIK ILIVIA 7, VR-T DF-1 ProMRI, REF 404625, UDI: 04035479142124 - Product Usage: intended to provide ve

Z-1457-2021  ·  Class II  ·  reported 2021-04-28  ·  Terminated

Why it was recalled
There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Product
BIOTRONIK ILIVIA 7, VR-T DF-1 ProMRI, REF 404625, UDI: 04035479142124 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
Where it went
US: Nationwide OUS: Worldwide
How much
914 units

BIOTRONIK INVENTRA 7, VR-T DX DF-1 ProMRI, REF 399436, UDI: 04035479136314 - Product Usage: intended to provi

Z-1459-2021  ·  Class II  ·  reported 2021-04-28  ·  Terminated

Why it was recalled
There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Product
BIOTRONIK INVENTRA 7, VR-T DX DF-1 ProMRI, REF 399436, UDI: 04035479136314 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
Where it went
US: Nationwide OUS: Worldwide
How much
N/A

BIOTRONIK IPERIA 7, DR-T DF4 ProMRI, REF 392423, UDI: 04035479128913 - Product Usage: intended to provide vent

Z-1453-2021  ·  Class II  ·  reported 2021-04-28  ·  Terminated

Why it was recalled
There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Product
BIOTRONIK IPERIA 7, DR-T DF4 ProMRI, REF 392423, UDI: 04035479128913 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
Where it went
US: Nationwide OUS: Worldwide
How much
2881 UNITS

BIOTRONIK ITREVIA 7, VR-T DX DF-1, REF 393037, UDI: 04035479129521 - Product Usage: intended to provide ventri

Z-1450-2021  ·  Class II  ·  reported 2021-04-28  ·  Terminated

Why it was recalled
There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Product
BIOTRONIK ITREVIA 7, VR-T DX DF-1, REF 393037, UDI: 04035479129521 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
Where it went
US: Nationwide OUS: Worldwide
How much
386 UNITS

PSW 1203.U/1, PSW 1301.U, PSW 1307.U, PSW 1403.U, and PSW 1501.U Pacemaker Programmer software. Software appr

Z-2376-2015  ·  Class II  ·  reported 2015-08-19  ·  Terminated

Why it was recalled
Ventricular packing: LV software programming versions for BIOTRONIK CRT-P and CRT-D devices are recalled because they are not approved for use in the US.
Product
PSW 1203.U/1, PSW 1301.U, PSW 1307.U, PSW 1403.U, and PSW 1501.U Pacemaker Programmer software. Software approved for use with BIOTRONIK device programmers (trade names: Renamic and ICS 3000). Allows physicians to program devices to pace exclusively in the left ventricle.
Where it went
Nationwide Distribution.
How much
768 units (software)
Read these counts with three things in mind

A firm with more recalls usually makes more products. A contract packer runs hundreds of lines. It will out-recall a company with one product whatever either does. This is not a safety ranking.

Most recalls here are voluntary. A firm found its own fault and pulled the stock. That is the system working. It lands in this database looking the same as a recall the FDA had to push for.

Terminated means the FDA considers the recall finished. It is not a verdict. It does not mean anyone was harmed.