BIOTRONIK Inc: 26 FDA recall reports
Every medical device recall the FDA has published for BIOTRONIK Inc. The fault, and how the agency graded it.
26Reports
0Class I
0Ongoing
2021-04-28Most recent
What the classes mean
The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.
Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.
By class
By year reported
Checking whether it affects you
The lot or code number is what matters. Not the product name. A recall covers
specific lots. Yours may not be one of them.
Check the code on your package against the report. The FDA keeps the current
list at fda.gov/safety/recalls.
The reports
The text below is quoted from the FDA record word for word.
BIOTRONIK Ilestro 7 VR-T, DF-1, REF 383578, UDI: 04035479125349 - Product Usage: intended to provide ventricul
Z-1443-2021 · Class II · reported 2021-04-28 · Terminated
- Why it was recalled
- There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
- Product
- BIOTRONIK Ilestro 7 VR-T, DF-1, REF 383578, UDI: 04035479125349 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
- Where it went
- US: Nationwide OUS: Worldwide
- How much
- 1000 UNITS
BIOTRONIK Ilestro 7 VR-T DX, DF-1, REF 383594, UDI: 04035479125509 - Product Usage: intended to provide ventri
Z-1445-2021 · Class II · reported 2021-04-28 · Terminated
- Why it was recalled
- There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
- Product
- BIOTRONIK Ilestro 7 VR-T DX, DF-1, REF 383594, UDI: 04035479125509 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
- Where it went
- US: Nationwide OUS: Worldwide
- How much
- 3664 UNITS
BIOTRONIK INVENTRA 7, HF-T QP DF4 IS4, REF 393012, UDI: 04035479129286 - Product Usage: intended to provide v
Z-1454-2021 · Class II · reported 2021-04-28 · Terminated
- Why it was recalled
- There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
- Product
- BIOTRONIK INVENTRA 7, HF-T QP DF4 IS4, REF 393012, UDI: 04035479129286 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
- Where it went
- US: Nationwide OUS: Worldwide
- How much
- 702 UNITS
BIOTRONIK ILIVIA NEO 7 DR-T DF-1 ProMRI, REF 429529, UDI: 04035479156749 - Product Usage: intended to provide
Z-1464-2021 · Class II · reported 2021-04-28 · Terminated
- Why it was recalled
- There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
- Product
- BIOTRONIK ILIVIA NEO 7 DR-T DF-1 ProMRI, REF 429529, UDI: 04035479156749 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
- Where it went
- US: Nationwide OUS: Worldwide
- How much
- 98 UNITS
BIOTRONIK ITREVIA 7, VR-T DF-4, REF 393041, UDI: 04035479129569 - Product Usage: intended to provide ventricul
Z-1452-2021 · Class II · reported 2021-04-28 · Terminated
- Why it was recalled
- There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
- Product
- BIOTRONIK ITREVIA 7, VR-T DF-4, REF 393041, UDI: 04035479129569 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
- Where it went
- US: Nationwide OUS: Worldwide
- How much
- 670 units
BIOTRONIK Ilestro 7 DR-T, DF-1, REF383562, UDI: 04035479125189 - Product Usage: intended to provide ventricula
Z-1441-2021 · Class II · reported 2021-04-28 · Terminated
- Why it was recalled
- There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
- Product
- BIOTRONIK Ilestro 7 DR-T, DF-1, REF383562, UDI: 04035479125189 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
- Where it went
- US: Nationwide OUS: Worldwide
- How much
- N/A
BIOTRONIK ILIVIA 7, VR-T DF4 ProMRI, REF 404626, UDI: 04035479142131 - Product Usage: intended to provide ven
Z-1458-2021 · Class II · reported 2021-04-28 · Terminated
- Why it was recalled
- There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
- Product
- BIOTRONIK ILIVIA 7, VR-T DF4 ProMRI, REF 404626, UDI: 04035479142131 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
- Where it went
- US: Nationwide OUS: Worldwide
- How much
- 779 units
BIOTRONIK ITREVIA 7, VR-T DF-1, REF 393040, UDI: 04035479129552 - Product Usage: intended to provide ventricul
Z-1451-2021 · Class II · reported 2021-04-28 · Terminated
- Why it was recalled
- There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
- Product
- BIOTRONIK ITREVIA 7, VR-T DF-1, REF 393040, UDI: 04035479129552 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
- Where it went
- US: Nationwide OUS: Worldwide
- How much
- 499 units
BIOTRONIK IPERIA 7,VR-T, DX, DF-1, ProMRI, REF 393032, UDI: 04035479129477 - Product Usage: intended to provid
Z-1449-2021 · Class II · reported 2021-04-28 · Terminated
- Why it was recalled
- There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
- Product
- BIOTRONIK IPERIA 7,VR-T, DX, DF-1, ProMRI, REF 393032, UDI: 04035479129477 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
- Where it went
- US: Nationwide OUS: Worldwide
- How much
- 3,832 UNITS
BIOTRONIK Ilestro 7 VR-T, DF-4, REF 383580, UDI: 04035479125363 - Product Usage: intended to provide ventricul
Z-1444-2021 · Class II · reported 2021-04-28 · Terminated
- Why it was recalled
- There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
- Product
- BIOTRONIK Ilestro 7 VR-T, DF-4, REF 383580, UDI: 04035479125363 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
- Where it went
- US: Nationwide OUS: Worldwide
- How much
- 323 UNITS
BIOTRONIK Ilestro 7 DR-T, DF-4, REF383564, UDI: 04035479125202 - Product Usage: intended to provide ventricula
Z-1442-2021 · Class II · reported 2021-04-28 · Terminated
- Why it was recalled
- There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
- Product
- BIOTRONIK Ilestro 7 DR-T, DF-4, REF383564, UDI: 04035479125202 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
- Where it went
- US: Nationwide OUS: Worldwide
- How much
- N/A
BIOTRONIK ILIVIA 7, DR-T DF-1 PROMRI, REF 404622, UDI: 04035479142094 - Product Usage: intended to provide ven
Z-1461-2021 · Class II · reported 2021-04-28 · Terminated
- Why it was recalled
- There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
- Product
- BIOTRONIK ILIVIA 7, DR-T DF-1 PROMRI, REF 404622, UDI: 04035479142094 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
- Where it went
- US: Nationwide OUS: Worldwide
- How much
- 1547 UNITS
BIOTRONIK ITREVIA 7, DR-T QP DF-1, REF 392412, UDI: 04035479128807 - Product Usage: intended to provide ventr
Z-1455-2021 · Class II · reported 2021-04-28 · Terminated
- Why it was recalled
- There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
- Product
- BIOTRONIK ITREVIA 7, DR-T QP DF-1, REF 392412, UDI: 04035479128807 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
- Where it went
- US: Nationwide OUS: Worldwide
- How much
- 658 units
BIOTRONIK Iforia 7,DR-T, DF-1,REF 390064, UDI: 04035479126926 - Product Usage: intended to provide ventricular
Z-1448-2021 · Class II · reported 2021-04-28 · Terminated
- Why it was recalled
- There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
- Product
- BIOTRONIK Iforia 7,DR-T, DF-1,REF 390064, UDI: 04035479126926 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
- Where it went
- US: Nationwide OUS: Worldwide
- How much
- 490 units
BIOTRONIK ILIVIA 7, DR-T DF4 ProMRI, REF 404623, UDI: 04035479142100 - Product Usage: intended to provide ven
Z-1456-2021 · Class II · reported 2021-04-28 · Terminated
- Why it was recalled
- There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
- Product
- BIOTRONIK ILIVIA 7, DR-T DF4 ProMRI, REF 404623, UDI: 04035479142100 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
- Where it went
- US: Nationwide OUS: Worldwide
- How much
- 3497 units
BIOTRONIK INTICA NEO 7 HF-T DF-1 IS-1 ProMRI, REF 429553, UDI: 04035479156831 - Product Usage: intended to pro
Z-1465-2021 · Class II · reported 2021-04-28 · Terminated
- Why it was recalled
- There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
- Product
- BIOTRONIK INTICA NEO 7 HF-T DF-1 IS-1 ProMRI, REF 429553, UDI: 04035479156831 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
- Where it went
- US: Nationwide OUS: Worldwide
- How much
- 8 UNITS
BIOTRONIK Iforia 7, vr-T, DX, DF-1,REF 390090, UDI: 04035479127176 - Product Usage: intended to provide ventri
Z-1447-2021 · Class II · reported 2021-04-28 · Terminated
- Why it was recalled
- There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
- Product
- BIOTRONIK Iforia 7, vr-T, DX, DF-1,REF 390090, UDI: 04035479127176 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
- Where it went
- US: Nationwide OUS: Worldwide
- How much
- 1176 UNITS
BIOTRONIK Iperia 7, DR-T, DF-1, ProMRI REF 392409, UDI: 04035479128777 - Product Usage: intended to provide ve
Z-1446-2021 · Class II · reported 2021-04-28 · Terminated
- Why it was recalled
- There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
- Product
- BIOTRONIK Iperia 7, DR-T, DF-1, ProMRI REF 392409, UDI: 04035479128777 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
- Where it went
- US: Nationwide OUS: Worldwide
- How much
- 1652 UNITS
BIOTRONIK ILIVIA NEO 7 VR-T DF-1 ProMRI, REF 429531, UDI: 04035479156763 - Product Usage: intended to provide
Z-1462-2021 · Class II · reported 2021-04-28 · Terminated
- Why it was recalled
- There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
- Product
- BIOTRONIK ILIVIA NEO 7 VR-T DF-1 ProMRI, REF 429531, UDI: 04035479156763 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
- Where it went
- US: Nationwide OUS: Worldwide
- How much
- 74 UNITS
BIOTRONIK INVENTRA 7, VR-T DX DF-1, REF 399437, UDI: 04035479136321 - Product Usage: intended to provide vent
Z-1460-2021 · Class II · reported 2021-04-28 · Terminated
- Why it was recalled
- There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
- Product
- BIOTRONIK INVENTRA 7, VR-T DX DF-1, REF 399437, UDI: 04035479136321 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
- Where it went
- US: Nationwide OUS: Worldwide
- How much
- 285 Unit
BIOTRONIK INTICA NEO 7 VR-T DX DF-1 ProMRI, REF 429559, UDI: 04035479156855 - Product Usage: intended to provi
Z-1463-2021 · Class II · reported 2021-04-28 · Terminated
- Why it was recalled
- There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
- Product
- BIOTRONIK INTICA NEO 7 VR-T DX DF-1 ProMRI, REF 429559, UDI: 04035479156855 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
- Where it went
- US: Nationwide OUS: Worldwide
- How much
- 17 UNITS
BIOTRONIK ILIVIA 7, VR-T DF-1 ProMRI, REF 404625, UDI: 04035479142124 - Product Usage: intended to provide ve
Z-1457-2021 · Class II · reported 2021-04-28 · Terminated
- Why it was recalled
- There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
- Product
- BIOTRONIK ILIVIA 7, VR-T DF-1 ProMRI, REF 404625, UDI: 04035479142124 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
- Where it went
- US: Nationwide OUS: Worldwide
- How much
- 914 units
BIOTRONIK INVENTRA 7, VR-T DX DF-1 ProMRI, REF 399436, UDI: 04035479136314 - Product Usage: intended to provi
Z-1459-2021 · Class II · reported 2021-04-28 · Terminated
- Why it was recalled
- There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
- Product
- BIOTRONIK INVENTRA 7, VR-T DX DF-1 ProMRI, REF 399436, UDI: 04035479136314 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
- Where it went
- US: Nationwide OUS: Worldwide
- How much
- N/A
BIOTRONIK IPERIA 7, DR-T DF4 ProMRI, REF 392423, UDI: 04035479128913 - Product Usage: intended to provide vent
Z-1453-2021 · Class II · reported 2021-04-28 · Terminated
- Why it was recalled
- There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
- Product
- BIOTRONIK IPERIA 7, DR-T DF4 ProMRI, REF 392423, UDI: 04035479128913 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
- Where it went
- US: Nationwide OUS: Worldwide
- How much
- 2881 UNITS
BIOTRONIK ITREVIA 7, VR-T DX DF-1, REF 393037, UDI: 04035479129521 - Product Usage: intended to provide ventri
Z-1450-2021 · Class II · reported 2021-04-28 · Terminated
- Why it was recalled
- There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
- Product
- BIOTRONIK ITREVIA 7, VR-T DX DF-1, REF 393037, UDI: 04035479129521 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
- Where it went
- US: Nationwide OUS: Worldwide
- How much
- 386 UNITS
PSW 1203.U/1, PSW 1301.U, PSW 1307.U, PSW 1403.U, and PSW 1501.U Pacemaker Programmer software. Software appr
Z-2376-2015 · Class II · reported 2015-08-19 · Terminated
- Why it was recalled
- Ventricular packing: LV software programming versions for BIOTRONIK CRT-P and CRT-D devices are recalled because they are not approved for use in the US.
- Product
- PSW 1203.U/1, PSW 1301.U, PSW 1307.U, PSW 1403.U, and PSW 1501.U Pacemaker Programmer software. Software approved for use with BIOTRONIK device programmers (trade names: Renamic and ICS 3000). Allows physicians to program devices to pace exclusively in the left ventricle.
- Where it went
- Nationwide Distribution.
- How much
- 768 units (software)
Read these counts with three things in mind
A firm with more recalls usually makes more products. A contract packer runs
hundreds of lines. It will out-recall a company with one product whatever either
does. This is not a safety ranking.
Most recalls here are voluntary. A firm found its own fault and pulled the stock.
That is the system working. It lands in this database looking the same as a recall
the FDA had to push for.
Terminated means the FDA considers the recall finished. It is not a verdict.
It does not mean anyone was harmed.
Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.