BioFire Diagnostics, LLC

BioFire Diagnostics, LLC: 42 FDA recall reports

Every medical device recall the FDA has published for BioFire Diagnostics, LLC. The fault, and how the agency graded it.

42Reports
0Class I
26Ongoing
2026-05-13Most recent

What the classes mean

The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.

ClassWhat the FDA is saying
Class IThere is a reasonable probability that using this will cause serious harm or death.
Class IIIt may cause temporary or reversible harm. Serious harm is unlikely.
Class IIIIt breaks a rule but is unlikely to hurt anyone. Often a labelling fault.

Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.

By class

ClassReports
Class II42

By year reported

YearReports
20183
20192
20203
20218
20239
202412
20253
20262
Checking whether it affects you

The lot or code number is what matters. Not the product name. A recall covers specific lots. Yours may not be one of them.

Check the code on your package against the report. The FDA keeps the current list at fda.gov/safety/recalls.

The reports

The text below is quoted from the FDA record word for word.

BioFire Joint Infection (JI) Panel REF: RFIT-ASY-0138, 30 test Kit.

Z-2067-2026  ·  Class II  ·  reported 2026-05-13  ·  Ongoing

Why it was recalled
Contamination to in-vitro diagnostic test may result in false positives.
Product
BioFire Joint Infection (JI) Panel REF: RFIT-ASY-0138, 30 test Kit.
Where it went
Worldwide - US Nationwide distribution in the states of TX, NC, UT, MD and the countries of France, Italy, Germany, Spain, Belgium, Switzerland, United Kingdom, Netherlands, Poland, Austria, India, Finland, Czech Republic, Egypt, Lebanon, Morocco, Slovenia, Romania, Dominican Republic, Serbia.
How much
106 kits (3,180 test)

Biofire Spotfire Respiratory/Sore Throat Panel REF: 423485

Z-2081-2026  ·  Class II  ·  reported 2026-05-13  ·  Ongoing

Why it was recalled
Respiratory/sore throat panel test may result in false negative results and control failures.
Product
Biofire Spotfire Respiratory/Sore Throat Panel REF: 423485
Where it went
US Nationwide distribution in the states of MS, TX MN, VA.
How much
2400 pouches

BioFire Respiratory Panel 2.1, IVD, REF: 423742 (30 test kit)

Z-0602-2026  ·  Class II  ·  reported 2025-12-03  ·  Ongoing

Why it was recalled
Due to unintended movement of pouch chemistry, nucleic acid test may result in elevated risk of false negative results and risk of control failures.
Product
BioFire Respiratory Panel 2.1, IVD, REF: 423742 (30 test kit)
Where it went
US:AZ, FL,GA,ID NC, NH, NJ, NY, OR, PA, SC OUS: None
How much
97 Kits (2,910 pouches)

BioFire Respiratory Panel 2.1 plus Kit, REF: 423740, For Filmarray systems

Z-1877-2025  ·  Class II  ·  reported 2025-06-04  ·  Ongoing

Why it was recalled
Due to manufacturing error, respiratory panel may result in elevated control failures and/or false negative results.
Product
BioFire Respiratory Panel 2.1 plus Kit, REF: 423740, For Filmarray systems
Where it went
International distribution in the country of Singapore.
How much
30 kits

FilmArra Pneumonia Panel plus (Pneumoplus), REF: RFIT-LBL-0374

Z-1518-2025  ·  Class II  ·  reported 2025-04-16  ·  Ongoing

Why it was recalled
Increased risk of control failures and false negative test results with multiplexed nucleic acid test.
Product
FilmArra Pneumonia Panel plus (Pneumoplus), REF: RFIT-LBL-0374
Where it went
OUS International Distribution to countries of: France, Italia, Spain, Switzerland, Israel
How much
29 kits

FILMARRAY 2.0 INSTRUMENT, REF: FLM2-ASY-0001, FLM2-ASY-0001-W

Z-3122-2024  ·  Class II  ·  reported 2024-09-18  ·  Ongoing

Why it was recalled
Some automated in vitro diagnostic modules and instruments manufactured or serviced between May 2022 and May 2024 were calibrated with internal equipment that may lead modules and instruments to report higher melting temperatures, which may lead to erroneous results including false positive Salmonella, Giardia lamblia, false positive/negative Norovirus, and false negative MRSA.
Product
FILMARRAY 2.0 INSTRUMENT, REF: FLM2-ASY-0001, FLM2-ASY-0001-W
Where it went
Worldwide - US Nationwide distribution including in the states of NC, TX, OH, CA, KS, AL, WA, NY, PA, IL, WI, KY, LA, NE, FL, UT, IA, SC, VA, AZ, CO, WV, MO, TN, MN, DC, DE, OK, WY, IN, NJ, NM, ID, OR, MS, SD, MI, MT, GA, HI, NV, MD, MA, AR, NH, PR, ND, VT, CT, AK, ME, RI and the countries of Korea, Canada, Singapore distributors, Canada, Canada, China, China, Singapore, Canada, China, Mexico, Japan, Chile, Chile, Singapore, Mexico, Hong King, China, Chile, Singapore, Philippines, Korea, Chile, Korea, Australia, Greece.
How much
1001

FILMARRAY TORCH MODULE BOX, HTFA-ASY-0003, HTFA-ASY-0003-W, HTFA-ASY-0003U

Z-3123-2024  ·  Class II  ·  reported 2024-09-18  ·  Ongoing

Why it was recalled
Some automated in vitro diagnostic modules and instruments manufactured or serviced between May 2022 and May 2024 were calibrated with internal equipment that may lead modules and instruments to report higher melting temperatures, which may lead to erroneous results including false positive Salmonella, Giardia lamblia, false positive/negative Norovirus, and false negative MRSA.
Product
FILMARRAY TORCH MODULE BOX, HTFA-ASY-0003, HTFA-ASY-0003-W, HTFA-ASY-0003U
Where it went
Worldwide - US Nationwide distribution including in the states of NC, TX, OH, CA, KS, AL, WA, NY, PA, IL, WI, KY, LA, NE, FL, UT, IA, SC, VA, AZ, CO, WV, MO, TN, MN, DC, DE, OK, WY, IN, NJ, NM, ID, OR, MS, SD, MI, MT, GA, HI, NV, MD, MA, AR, NH, PR, ND, VT, CT, AK, ME, RI and the countries of Korea, Canada, Singapore distributors, Canada, Canada, China, China, Singapore, Canada, China, Mexico, Japan, Chile, Chile, Singapore, Mexico, Hong King, China, Chile, Singapore, Philippines, Korea, Chile, Korea, Australia, Greece.
How much
3,439

BIOFIRE FILMARRAY Pneumonia Panel plus, REF: RFIT-ASY-0143

Z-2683-2024  ·  Class II  ·  reported 2024-08-28  ·  Ongoing

Why it was recalled
Pneumonia panels have an increased risk of control failures due to an issue in the manufacturing process, which could lead to delayed results for the customer.
Product
BIOFIRE FILMARRAY Pneumonia Panel plus, REF: RFIT-ASY-0143
Where it went
Worldwide - US Nationwide distribution in the states of WV, IN, AL and the countries of France, Italia, Spain, Greece, Japan Ltd, Singapore, Chile, Taiwan, Israel, Romania, Ecuador, Serbia.
How much
24 Kits

BIOFIRE FILMARRAY Pneumonia Panel, REF: RFIT-ASY-0144

Z-2682-2024  ·  Class II  ·  reported 2024-08-28  ·  Ongoing

Why it was recalled
Pneumonia panels have an increased risk of control failures due to an issue in the manufacturing process, which could lead to delayed results for the customer.
Product
BIOFIRE FILMARRAY Pneumonia Panel, REF: RFIT-ASY-0144
Where it went
Worldwide - US Nationwide distribution in the states of WV, IN, AL and the countries of France, Italia, Spain, Greece, Japan Ltd, Singapore, Chile, Taiwan, Israel, Romania, Ecuador, Serbia.
How much
58

BIOFIRE FILMARRAY Gastrointestinal (GI) Panel REF RFIT-ASY-0116 (30 test kit) BIOFIRE FILMARRAY Gastrointe

Z-2514-2024  ·  Class II  ·  reported 2024-08-14  ·  Ongoing

Why it was recalled
Due to increase of false positive Vibrio/Vibrio Cholerae results when using a certain transport medium.
Product
BIOFIRE FILMARRAY Gastrointestinal (GI) Panel REF RFIT-ASY-0116 (30 test kit) BIOFIRE FILMARRAY Gastrointestinal (GI) Panel REF RFIT-ASY-0104 (6 kit test) The BIOFIRE FILMARRAY Gastrointestinal (GI) Panel is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with BIOFIRE¿ FILMARRAY¿ Systems. The BIOFIRE GI Panel is capable of the simultaneous detection and identification of nucleic acids from multiple bacteria, viruses, and parasites directly from stool samples in Cary Blair transport media obtained from individuals with signs and/or symptoms of gastrointestinal infection.
Where it went
U.S.: Nationwide - AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, Virgin Island, VT, WA, WI, WV, and WY. O.U.S.: Worldwide - AR, AU, AT, BE, BR, CA, CL, HK, CN, CO, CZ, DE, FI, FR, GR, HU, IN, IT, JP, MX, NL, NO, PH, PL, PT, SG, ZA, KR, ES, SE, CH, TW, TH, TR, GB, AL, DZ, AO, AM, BH, BY, BO, BA, BW, BG, CR, HR, CY, CD, DO, AN, EC, SV, EE, FJ, GE, GU, GT, GN, HT, HN, IS, ID, IQ, IE, IL, CI, CI, JO, KE, KW, LV, LB, LT, MO, MY, ML, MR, MD, MN, MA, MM, NP, NZ, NZ, NI, OM, PK, PA, PG, PY, PE, QA, RO, SA, RS, SI, TN, TR, UY, AE, VE, VN, and ZW.
How much
67,613 kits (or 2,028,309 pouches). Each pouch is used for each individual test.

BIOFIRE Blood Culture Identification 2 (BCID2) Panel, REF: RFIT-ASY-0147, for FILMARRAY Systems, when used in

Z-2456-2024  ·  Class II  ·  reported 2024-08-07  ·  Ongoing

Why it was recalled
If blood culture identification panel is used in conjunction with specific lots of culture media bottles, then false positive Serratia marcescens results may occur, due to an increased level of non-viable organism from serratia marcescens targets in culture media bottles, false positive result may lead to an inappropriate change in patient therapy.
Product
BIOFIRE Blood Culture Identification 2 (BCID2) Panel, REF: RFIT-ASY-0147, for FILMARRAY Systems, when used in conjunction with Culture Media Bottles: BACT/ALERT FA PLUS, REF: 410851, and BACT/ALERT PF PLUS, REF: 410853
Where it went
Worldwide - US Nationwide distribution including in the states of FL, IL, KY, CO, KS, NC, AK, NY, ND, CA, TX, MD, GA, OH, WI, SD, LA, MT, VA, IA, MO, MA, WV, NE, OK, AZ, OR, SC, MI, IN, ID, TN, WA, NM, NJ, AR, AL, MN, NH, NV, PA, WY, MS, CT and the countries of AE, AM, AR, AT, BA, BG, BH, BN, BR, CA, CH, CI, CK, CN, CO, CZ, DE, DZ, EC, EE, EG, ES, FI, FR, GA, GB, GE, GR, GT, HR, HU, IL, IN, IQ, IS, IT, JO, JP, KK, KR, KW, LB, LT, LV, MK, MM, MX, MY, NG, NO, NW, OM, PA, PH, PK, PL, PT, QA, RO, RS, SA, SI, TN, TR, TW, UG, UY, VN, WF.
How much
99,822

FilmArray Gastrointestinal (GI) Panel RFIT-ASY-0116 (30 test kit) RFIT-ASY-0104 (6 test kit) Gastrointest

Z-1140-2024  ·  Class II  ·  reported 2024-02-21  ·  Ongoing

Why it was recalled
Due to potential signals of increased false positive Norovirus results when using the gastrointestinal (GI) panel.
Product
FilmArray Gastrointestinal (GI) Panel RFIT-ASY-0116 (30 test kit) RFIT-ASY-0104 (6 test kit) Gastrointestinal microorganism multiplex nucleic acid-based assay. In vitro diagnostic.
Where it went
Worldwide - U.S. Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, US Virgin Island, VT, WA, WI, WV, and WY. The countries of Andorra, United Arab Emirates, Albania, Armenia, Argentina, Austria, Australia, Bosnia, Belgium, Bulgaria, Bahrain, Bermuda, Bolivia, Brazil, Botswana, Canada, Congo, Switzerland, Cote D'Ivoire, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Germany, Denmark, Dominican Republic, Algeria, Ecuador, Estonia, Egypt, Spain, Finland, Fiji, France, United Kingdom, Georgia, French Guiana, Gibraltar, Guadeloupe, Greece, Guatemala, Hong Kong, Honduras, Croatia, Hungary, Iceland, Indonesia, Ireland, Israel, India, Italy, Jordan, Japan, Korea, Kuwait, Lebanon, Lithuania, Latvia, Morocco, Macedonia, Mali, Myanmar, Mongolia, Macau, Mauritania, Mexico, Malaysia, Namibia, New Caledonia, Nigeria, Nicaragua, Netherlands, Norway, Nepal, New Zealand, Oman, Panama, Peru, French Polynesia, Papua New Guinea, Philippines, Poland, St. Pierre/Miquelon, Portugal, Paraguay, Qatar,...
How much
64,594 kits (equals 2,303,760 tests)

BIOFIRE Blood Culture Identification 2 (BCID2) Panel when used with: BD BACTEC Lytic Anaerobic medium BD

Z-1139-2024  ·  Class II  ·  reported 2024-02-21  ·  Ongoing

Why it was recalled
Due to a trend in false positive Candida tropicalis results on blood culture panel when using certain BD BACTEC vial mediums
Product
BIOFIRE Blood Culture Identification 2 (BCID2) Panel when used with: BD BACTEC Lytic Anaerobic medium BD BACTEC Peds Plus medium BD BACTEC Plus Aerobic medium BD BACTEC Plus Anaerobic medium BD BACTEC Standard Aerobic medium BD BACTEC Standard Anaerobic medium The BIOFIRE Blood Culture Identification 2 (BCID2) Panel is a multiplex nucleic acid assay for identification of microorganisms and resistance markers from positive blood cultures.
Where it went
Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Andorra, United Arab Emirates, Armenia, Argentina, Austria, Australia, Bosnia, Belgium, Bulgaria, Bahrain, Brunei, Bolivia, Brazil, Canada, Switzerland, Cote D'Ivoire, Chile, China, Colombia, Costa Rica, Cyprus, Cech Republic, Germany, Denmark, Dominican Republic, Algeria, Ecuador, Estonia, Egypt, Spain, Finland, France, Gabon, United Kingdom, Georgia, Guadeloupe, Greece, Guatemala, Hong Kong, Honduras, Croatia, Hungary, Iceland, Indonesia, Ireland, Israel, India, Italy, Jordan, Japan, Korea, Kuwait, Lebanon, Lithuania, Latvia, Morocco, Macedonia, Mali, Mongolia, Macau, Mexico, Malaysia, New Caledonia, Nigeria, Nicaragua, Netherlands, Norway, New Zealand, Oman, Panama, Peru, French Polynesia, Philippines, Pakistan, Poland, St. Pierre/Miquelon, Portugal, Paraguay, Qatar, Reunion, Romania, Serbia, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, El Salvador, Thailand,...
How much
32,910 panels/pouches

FilmArray Pneumonia Panel Plus (Pneumoplus), REF: RFIT-ASY-0143, CED 2797, IVD, Rx Only (30 test kit)

Z-0993-2024  ·  Class II  ·  reported 2024-02-14  ·  Ongoing

Why it was recalled
Their is a potential for false positive results when using Pneumonia panel.
Product
FilmArray Pneumonia Panel Plus (Pneumoplus), REF: RFIT-ASY-0143, CED 2797, IVD, Rx Only (30 test kit)
Where it went
Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Argentina, Bolivia, Brazil, Chili, China, Colombia, Costa Rica, Dominican Republic, Ecuador, Guatemala, Hong Kong, Honduras, Japan, South Korea, Mexico, Nicaragua, Panama, Peru, Philippines, Paraguay, Singapore, El Salvador, Thailand, Taiwan, Uruguay, Venezuela.
How much
214,080 pouches

FilmArray Pneumonia Panel (Pneumo), REF: FIT-ASY-0144, CED 2797, IVD, Rx Only (30 test kit)

Z-0994-2024  ·  Class II  ·  reported 2024-02-14  ·  Ongoing

Why it was recalled
Their is a potential for false positive results when using Pneumonia panel.
Product
FilmArray Pneumonia Panel (Pneumo), REF: FIT-ASY-0144, CED 2797, IVD, Rx Only (30 test kit)
Where it went
Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Argentina, Bolivia, Brazil, Chili, China, Colombia, Costa Rica, Dominican Republic, Ecuador, Guatemala, Hong Kong, Honduras, Japan, South Korea, Mexico, Nicaragua, Panama, Peru, Philippines, Paraguay, Singapore, El Salvador, Thailand, Taiwan, Uruguay, Venezuela.
How much
274,290 units

The BIOFIRE Respiratory Panel 2.1 (RP2.1) is a PCR-based multiplexed nucleic acid test intended for use with t

Z-0704-2024  ·  Class II  ·  reported 2024-01-17  ·  Ongoing

Why it was recalled
A manufacturing event may contribute to elevated rates of false negative results while using gastrointestinal panel and respiratory panel, which could lead to inappropriate antibiotic treatment, no treatment, delay of treatment, and may provide false assurance to someone with a true infection, delayed identification of an outbreak, and may result in additional transmission.
Product
The BIOFIRE Respiratory Panel 2.1 (RP2.1) is a PCR-based multiplexed nucleic acid test intended for use with the BIOFIRE FILMARRAY 2.0 or BIOFIRE FILMARRAY Torch Systems.
Where it went
US: NC, MO, NE, CT, IL, TN, NJ, OH, MA, CA, FL, UT, IA, KS, TX, MD
How much
57 Kits

BIOFIRE FILMARRAY Gastrointestinal (GI) Panel is a qualitative multiplexed nucleic acid-based in vitro diagnos

Z-0703-2024  ·  Class II  ·  reported 2024-01-17  ·  Ongoing

Why it was recalled
A manufacturing event may contribute to elevated rates of false negative results while using gastrointestinal panel and respiratory panel, which could lead to inappropriate antibiotic treatment, no treatment, delay of treatment, and may provide false assurance to someone with a true infection, delayed identification of an outbreak, and may result in additional transmission.
Product
BIOFIRE FILMARRAY Gastrointestinal (GI) Panel is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with BIOFIRE FILMARRAY Systems.
Where it went
US: NC, MO, NE, CT, IL, TN, NJ, OH, MA, CA, FL, UT, IA, KS, TX, MD
How much
53 Kits

BIOFIRE Blood Culture Identification 2 (BCID2) Panel, REF: RFIT-ASY-0147, used with Maine Molecular Quality Co

Z-0451-2024  ·  Class II  ·  reported 2023-12-13  ·  Ongoing

Why it was recalled
Routine quality control testing of affected blood culture identification panel with affected control panel may need to be amended, due to control panel/instrument characteristic changes, that lead C. tropicalis to be amplified in some portion of test runs, but then it's reported as "Not Detected" because the assay amplicon Tm is outside the acceptable limit assigned for calling positive results.
Product
BIOFIRE Blood Culture Identification 2 (BCID2) Panel, REF: RFIT-ASY-0147, used with Maine Molecular Quality Controls, Inc (MMQCI) FilmArray BCID2 Control Panel M416, which is referenced in Quality Control Materials for use with BIOFIRE FILMARRAY and BIOFIRE SPOTFIRE Panels - FLM1-PRT-0251
Where it went
Worldwide Distribution: US (nationwide): DE, OH, NY, IL, KY, FL, GA, KS, CA, WI, NC,, AK, PA, MN, ND, IN, AR, CO, UT, TX, AL, NJ, MD, SD, LA, OR, MA, ID, NE, MT, VA, IA, MO, WV, TN, MI, WY, SC, AZ, WA, NH, OK, HI, MS, DC, VT, CT, NM, RI, ME, NV; and OUS (foreign) to countries of: Algeria, Angola, Armenia, Bahrain, Bolivia, Bosnia- Herzegovina, Brunei, Bulgaria, Costa Rica, Croatia, Cyprus, Dominican Republic, Dutch Antilles, Ecuador, El Salvador, Estonia, Gabon, Georgia, Guam, Guatemala, Honduras, Indonesia, Ireland, Iceland, Israel, Ivory Coast, Jordan, Kosovo, Kuwait, Latvia, Lebanon, Lithuania, Macao, Macedonia, Malaysia, Mali, Mongolia, Morocco, New Zealand, Nicaragua, Oman, Pakistan, Panama, Paraguay, Peru, Qatar, Romania, Saudi Arabia, Serbia , Slovenia, Tunisia, Uganda, United Arab Emirate, Uruguay, Venezuela, Vietnam, Zimbabwe, Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Czech Republic, Spain, France, United Kingdom, Greece, India, Italia, Japan, Korea, Netherlands, Poland, Portugal, Sweden, Singapore, Thailand, Turkey, Taiwan, Switzerland, Austria, Denmark, Finland, Germany, Belgium, Hungary, South Africa, Philippines, Norway, Mexico
How much
32,910

FilmArray TORCH REF HTFA-ASY-0104. HTFA-ASY-0104 is the BIOFIRE TORCH Base which includes barcode scanner, to

Z-0353-2024  ·  Class II  ·  reported 2023-11-29  ·  Ongoing

Why it was recalled
Due to an increased risk of degradation of the power entry switch resulting from arcing inside of the power switch that may result in carbon build-up on the switch contacts. The carbon build-up may lead to excess heating inside of the power switch, subsequently leading to deformation of the power switch case. The deformation can create an opportunity for a power switch electrical short. In addition to an electrical short, deformation may result in an open circuit causing the power switch to fail. This event would only occur after the product is in use.
Product
FilmArray TORCH REF HTFA-ASY-0104. HTFA-ASY-0104 is the BIOFIRE TORCH Base which includes barcode scanner, touch screen computer and USB ports. The BioFire FilmArray Torch (BioFire Torch) is an automated in vitro diagnostic (IVD) device intended for use with FDA-cleared or approved IVD BioFire FilmArray Panels. Additional Part/Reference Numbers as part of the system: - HTFA-ASY-0102: BIOFIRE TORCH Duplex. This is a plastic component that covers the BIOFIRE TORCH Modules. The BIOFIRE TORCH Duplex is only necessary if the system includes more than two modules. - HFTA-SUB-0103: BIOFIRE TORCH Module. The BIOFIRE reagent pouches are inserted into the modules.
Where it went
Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, N, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, Virgin Island, VT, WA, WI, WV, and WY. The countries of Algeria, Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Egypt, El Salvador, France, Germany, Guatemala, India, Indonesia, Italy, Japan, Korea, Kuwait, Malaysia, Mexico, New Zealand, Paraguay, Philippines, Poland, Qatar, Romania, Saudi Arabia, Singapore, Spain, Taiwan, Turkey, United Arab Emirates.
How much
5,406 systems

FILMARRAY Gastrointestinal (GI) Panel (Part No.: RFIT-ASY-0104 & RFIT-ASY-0116) Pouch Module Software

Z-2588-2023  ·  Class II  ·  reported 2023-09-20  ·  Ongoing

Why it was recalled
The firm has identified an increased risk of obtaining false positive Cryptosporidium results from customers using the BIOFIRE¿ FILMARRAY¿ Gastrointestinal (GI) Panel
Product
FILMARRAY Gastrointestinal (GI) Panel (Part No.: RFIT-ASY-0104 & RFIT-ASY-0116) Pouch Module Software
Where it went
US:DE, AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY. OUS: AD, AE, AI, AR, AT, AU, BD, BE, BF, BG, BH, BR, CA, CH, CL, CN, CO, CY, CZ, DE, DK, DZ, EE, ES, FI, FR, GA, GB, GE, GH, GM, GR, GU, HK, HU, IC, ID, IE, IL, IN, IQ, IT, JP, KE, KR, KW, LV, MO, MX, MY, NL, NO, NP, NZ, PG, PH, PL, PT, QA, RS, SA, SE, SG, SI, TH, TN, TR, TW, VN, ZA, ZW
How much
61,069

BioFire Respiratory Panel 2.1 (RP2.1), REF: 423742, For FILMARRAY systems, IVD, Rx Only

Z-2439-2023  ·  Class II  ·  reported 2023-08-30  ·  Ongoing

Why it was recalled
Due to manufacturing issue, panels may result in false negative results.
Product
BioFire Respiratory Panel 2.1 (RP2.1), REF: 423742, For FILMARRAY systems, IVD, Rx Only
Where it went
US: CA CO CT FL GA IA IL IN LA MA MI MO MT NC NE NJ NM NY OH OK PA TN UT WV OUS: Canada, India, Japan, Singapore, Colombia
How much
705 kits (30 pouches/kit)

FilmArray Pneumonia Panel (Pneumo), REF: RFIT-ASY-0145, For FILMARRAY systems, CE 2797, IVD, Rx Only

Z-2440-2023  ·  Class II  ·  reported 2023-08-30  ·  Ongoing

Why it was recalled
Due to manufacturing issue, panels may result in false negative results.
Product
FilmArray Pneumonia Panel (Pneumo), REF: RFIT-ASY-0145, For FILMARRAY systems, CE 2797, IVD, Rx Only
Where it went
US: CA CO CT FL GA IA IL IN LA MA MI MO MT NC NE NJ NM NY OH OK PA TN UT WV OUS: Canada, India, Japan, Singapore, Colombia
How much
24 Kits (30 pouches/kit)

FilmArray Gastrointestinal (GI) Panel for FILMARRAY Systems, REF: RFIT-ASY-0116, (30 pouches/kit)

Z-1392-2023  ·  Class II  ·  reported 2023-04-26  ·  Ongoing

Why it was recalled
Due to manufacturing issue, Gastrointestinal (GI) Panel may result in false negative results.
Product
FilmArray Gastrointestinal (GI) Panel for FILMARRAY Systems, REF: RFIT-ASY-0116, (30 pouches/kit)
Where it went
US Nationwide distribution in the states of AR, CA, DC, IL, KS, LA, MN, NC, NY, OK, WI.
How much
900 pouches (30 kits)

In Vitro Diagnostic (IVD) Panel BioFire BCID2 Panel - REF RFIT-ASY-0147

Z-1251-2023  ·  Class II  ·  reported 2023-03-22  ·  Ongoing

Why it was recalled
Due to temperature excursions with products (products being stored outside of the required temperature specifications) that may cause a reduction in performance. The temperature excursion was as a result of shipping delays.
Product
In Vitro Diagnostic (IVD) Panel BioFire BCID2 Panel - REF RFIT-ASY-0147
Where it went
U.S. Nationwide distribution in the states of FL and GA.
How much
2 kits

In Vitro Diagnostic (IVD) Panel BioFire Respiratory Panel 2.1 (RP2.1) - REF 423742

Z-1252-2023  ·  Class II  ·  reported 2023-03-22  ·  Ongoing

Why it was recalled
Due to temperature excursions with products (products being stored outside of the required temperature specifications) that may cause a reduction in performance. The temperature excursion was as a result of shipping delays.
Product
In Vitro Diagnostic (IVD) Panel BioFire Respiratory Panel 2.1 (RP2.1) - REF 423742
Where it went
U.S. Nationwide distribution in the states of FL and GA.
How much
8 kits

In Vitro Diagnostic (IVD) Panel FilmArray GI Panel - REF RFIT-ASY-0116

Z-1253-2023  ·  Class II  ·  reported 2023-03-22  ·  Ongoing

Why it was recalled
Due to temperature excursions with products (products being stored outside of the required temperature specifications) that may cause a reduction in performance. The temperature excursion was as a result of shipping delays.
Product
In Vitro Diagnostic (IVD) Panel FilmArray GI Panel - REF RFIT-ASY-0116
Where it went
U.S. Nationwide distribution in the states of FL and GA.
How much
2 kits

BioFire Blood Culture Identification 2 (BCID2) Panel Part No: RFIT-ASY-0147 UDI: 00815381020338 Blood Culture

Z-0312-2022  ·  Class II  ·  reported 2021-12-08  ·  Terminated

Why it was recalled
Due to an increased risk of false positive Acinetobacter calcoaceticus- baumannii complex results. The cause may be as a due to non-viable contamination in the blood culture vials.
Product
BioFire Blood Culture Identification 2 (BCID2) Panel Part No: RFIT-ASY-0147 UDI: 00815381020338 Blood Culture Catalog No./Description: 442023/BD BACTEC Plus Aerobic medium in plastic vials 442021/BD BACTEC Lytic Anaerobic medium in plastic vials 442192//BD BACTEC Plus Aerobic medium in glass vials 442022/BD BACTEC Plus Anaerobic medium in plastic vials 442020/BD BACTEC Peds Plus mediums in plastic vials All lots with an expiry date prior to
Where it went
Worldwide distribution - U.S. Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY. The Countries of Algeria, Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Bulgaria, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Dominican Rep/, Ecuador, Egypt, France, Georgia, Germany, Greece, Guam, Guatemala, Hong Kong, Hungary, India, Israel Italy, Kuwait, Latvia, Macedonia, Malaysia, Mexico, Netherlands, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Shanghai, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, United Kingdom, Uruguay, Utd/Arab Emir/.
How much
Total 14,140 Kits [11,155 kits (U.S.) and 2,985 kits (O.U.S.)]

FilmArray Blood Culture Identification (BCID) Panel Product Pouch Label: FilmArray BCID Panel Part No:

Z-2245-2021  ·  Class II  ·  reported 2021-08-18  ·  Terminated

Why it was recalled
Due to product complaints received for false negative results (dropouts) with blood culture identification panel potential caused by a pouch manufacturing error.
Product
FilmArray Blood Culture Identification (BCID) Panel Product Pouch Label: FilmArray BCID Panel Part No: RFIT-ASY-0126 Part No: RFIT-ASY-0127
Where it went
U.S. Nationwide Distribution in the states of : FL, MI, NJ, OH, VA and WI. O.U.S.: None
How much
29 kits

FilmArray Pneumonia Panel (Pneumo) Part Numbers: Part No: RFIT-ASY-0144 / UDI: 00815381020178 (30 test kit

Z-2039-2021  ·  Class II  ·  reported 2021-07-14  ·  Terminated

Why it was recalled
Due to real-time stability study failures resulting in increase rate of false negative results between the 6 to 12 month expiry time-frame.
Product
FilmArray Pneumonia Panel (Pneumo) Part Numbers: Part No: RFIT-ASY-0144 / UDI: 00815381020178 (30 test kit) Part No: RFIT-ASY-0145 / UDI: 00815381020185 (6 test kit)
Where it went
U.S.: AK, AL, AZ, CA, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, IN, KS, KY, KY, LA, MA, MD, ME, MI, MI, MN, MO, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, TN, TX, UT, VA, VT, WA, WI, WV and WY. O.U.S.: Algeria, Angola, Anguilla, Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Bermuda, Bolivia, Bosnia-Herz., Brazil, Bulgaria, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, France, Georgia, Germany, Greece , Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Lebanon, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Myanmar, Nepal, Netherlands, Nicaragua, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, United Kingdom, Uruguay, United Arab Emirates, and Vietnam.
How much
4,462 kits (Globally)

FilmArray Pneumonia Panel plus (Pneumoplus) Part Numbers: Part No: RFIT-ASY-0143 / UDI: 00815381020314 (30

Z-2040-2021  ·  Class II  ·  reported 2021-07-14  ·  Terminated

Why it was recalled
Due to real-time stability study failures resulting in increase rate of false negative results between the 6 to 12 month expiry time-frame.
Product
FilmArray Pneumonia Panel plus (Pneumoplus) Part Numbers: Part No: RFIT-ASY-0143 / UDI: 00815381020314 (30 test kit) Part No: RFIT-ASY-0142 / UDI: 00815381020321 (6 test kit)
Where it went
U.S.: AK, AL, AZ, CA, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, IN, KS, KY, KY, LA, MA, MD, ME, MI, MI, MN, MO, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, TN, TX, UT, VA, VT, WA, WI, WV and WY. O.U.S.: Algeria, Angola, Anguilla, Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Bermuda, Bolivia, Bosnia-Herz., Brazil, Bulgaria, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, France, Georgia, Germany, Greece , Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Lebanon, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Myanmar, Nepal, Netherlands, Nicaragua, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, United Kingdom, Uruguay, United Arab Emirates, and Vietnam.
How much
3,559 kits (Globally)

Showing the 30 most recent of 42. The rest are searchable at the FDA enforcement report database.

Read these counts with three things in mind

A firm with more recalls usually makes more products. A contract packer runs hundreds of lines. It will out-recall a company with one product whatever either does. This is not a safety ranking.

Most recalls here are voluntary. A firm found its own fault and pulled the stock. That is the system working. It lands in this database looking the same as a recall the FDA had to push for.

Terminated means the FDA considers the recall finished. It is not a verdict. It does not mean anyone was harmed.

Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.

← All device recalls