Becton, Dickinson and Company, BD Biosciences: 35 FDA recall reports
Every medical device recall the FDA has published for Becton, Dickinson and Company, BD Biosciences. The fault, and how the agency graded it.
35Reports
0Class I
7Ongoing
2024-07-24Most recent
What the classes mean
The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.
Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.
By class
By year reported
Checking whether it affects you
The lot or code number is what matters. Not the product name. A recall covers
specific lots. Yours may not be one of them.
Check the code on your package against the report. The FDA keeps the current
list at fda.gov/safety/recalls.
The reports
The text below is quoted from the FDA record word for word.
BD Multitest 6-Color TBNK CE-IVD, REF: 644611; BD Multitest 6-Color TBNK Kit with BD Trucount Tubes CE-IVD, R
Z-2351-2024 · Class II · reported 2024-07-24 · Ongoing
- Why it was recalled
- Multitest with compromised raw material used to assess immune system status may result in flow cytometry plot with unusual staining pattern, nonspecific aggregates, high background, double positive event increase on CD3+CD4+vsCD3+CD8+ and/or CD3-CD16+CD56+vsCD3-CD19+, contaminated antibodies, may lead to erroneous or delayed results, additional patient sample, might impact healthcare decisions
- Product
- BD Multitest 6-Color TBNK CE-IVD, REF: 644611; BD Multitest 6-Color TBNK Kit with BD Trucount Tubes CE-IVD, REF: 337166; BD Multitest 6-Color TBNK, IVD, REF: 662967; BD Multitest 6-Color TBNK Kit with BD Trucount Tubes IVD, REF: 662995. BD Multitest" 6-Color TBNK reagent with optional BD Trucount" Tubes is a six-color direct immunofluorescence reagent.
- Where it went
- US: MN, NJ, NY, MD, UT, FL, NC, MA, KY, NE, NV, CA, LA, DC, OH, MI, PA, WA, CT, AL, DE, TN, IN, OR, IL, WI, NH, TX, GA, RI, SD, ID, NM, AR. OUS: CA, SG, NZ, AU, KR, IN, PE, GT, AR, MX, TW, CN, NI, CR, JP, BR, TH, BE, VN
- How much
- 10,027
CD11b APC: ASR, REF: 340936, and CE, REF: 333143
Z-0641-2024 · Class II · reported 2024-01-17 · Ongoing
- Why it was recalled
- In vitro diagnostics product that impacts the quality of cell staining can cause low or dim fluorescence signal, which can lead to potential false negative results and misdiagnosis of hematological malignancy, delayed treatment, and the need for additional patient samples.
- Product
- CD11b APC: ASR, REF: 340936, and CE, REF: 333143
- Where it went
- Worldwide - US Nationwide distribution including in the states of IL, MN, FL, MI, MD, NC, NY, CA, GA, AR, KS, PA, NH, TX, OK, MA, MO, DC, AZ, IN, NJ, TN, WA, VA, CT and the countries of CA, BE, SG, HK, TW, BR, MY, IN, BO, PE, JP, MX, CL, KR, UY, AU.
- How much
- 464 Vials
BD Multitest 6-Color TBNK with optional BD Trucount Tubes (Cat. No. 662995), for use with BD FACSLyric, BD FAC
Z-0173-2023 · Class II · reported 2022-11-09 · Ongoing
- Why it was recalled
- Due to partial label detachment on tubes. This can cause tubes to become stuck together in the pouch or cause automation errors when used with the automated loader for the wash assistant.
- Product
- BD Multitest 6-Color TBNK with optional BD Trucount Tubes (Cat. No. 662995), for use with BD FACSLyric, BD FACSCanto II, and BD FACSCanto flow cytometers to determine the percentages and absolute counts of human lymphocyte subsets in peripheral whole blood for immunophenotyping.
- Where it went
- U.S.: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SD, TN, TX, UT, VA, VT, WA, and WI O.U.S.: Argentina, Australia, Belgium, Bolivia, Brazil, Canada, Chile, China, Columbia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Guatemala, Hong Kong, India, Japan, Korea, Mexico, Malaysia, New Zealand, Panama, Peru, Philippines, Paraguay, Singapore, Thailand, Taiwan, and Uruguay
- How much
- 1,457 tubes (corrected on 11/04/2022)
BD Trucount Tubes (Cat. No. 663028), used for determining absolute counts of leukocytes in blood. BD Trucount
Z-0172-2023 · Class II · reported 2022-11-09 · Ongoing
- Why it was recalled
- Due to partial label detachment on tubes. This can cause tubes to become stuck together in the pouch or cause automation errors when used with the automated loader for the wash assistant. Firm root cause investigation confirmed that label detachment is due to an adhesive change by a tier 2 supplier.
- Product
- BD Trucount Tubes (Cat. No. 663028), used for determining absolute counts of leukocytes in blood. BD Trucount Tubes are designed for use with in vitro diagnostic products such as BD Tritest reagents, and a suitably equipped flow cytometer. BD Trucount Tubes can be used with the BD FACS Loader.
- Where it went
- U.S.: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SD, TN, TX, UT, VA, VT, WA, and WI O.U.S.: Argentina, Australia, Belgium, Bolivia, Brazil, Canada, Chile, China, Columbia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Guatemala, Hong Kong, India, Japan, Korea, Mexico, Malaysia, New Zealand, Panama, Peru, Philippines, Paraguay, Singapore, Thailand, Taiwan, and Uruguay
- How much
- 11,709 tubes (corrected 11/04/2022)
BD Trucount Tubes (Cat. No. 340334), used for determining absolute counts of leucocytes in blood. BD Trucount
Z-0171-2023 · Class II · reported 2022-11-09 · Ongoing
- Why it was recalled
- Due to partial label detachment on tubes. This can cause tubes to become stuck together in the pouch or cause automation errors when used with the automated loader for the wash assistant. Firm root cause investigation confirmed that label detachment is due to an adhesive change by a tier 2 supplier.
- Product
- BD Trucount Tubes (Cat. No. 340334), used for determining absolute counts of leucocytes in blood. BD Trucount tubes are designed for use with in vitro diagnostic products such as BD Tritest reagents, and a suitably equipped flow cytometer. BD Trucount tubes can be used with the BD FACS Loader.
- Where it went
- U.S.: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SD, TN, TX, UT, VA, VT, WA, and WI O.U.S.: Argentina, Australia, Belgium, Bolivia, Brazil, Canada, Chile, China, Columbia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Guatemala, Hong Kong, India, Japan, Korea, Mexico, Malaysia, New Zealand, Panama, Peru, Philippines, Paraguay, Singapore, Thailand, Taiwan, and Uruguay
- How much
- 52,570 tubes (corrected on 11/04/2022)
BD Anti-Kappa PE (Cat. No. 346587) is an analyte specific reagent for the Anti-Kappa antibody, which specifica
Z-1215-2022 · Class II · reported 2022-06-08 · Ongoing
- Why it was recalled
- Anti-Kappa APC & Anti-Kappa PE analyte reagents may exhibit uncharacteristic flow profiles.
- Product
- BD Anti-Kappa PE (Cat. No. 346587) is an analyte specific reagent for the Anti-Kappa antibody, which specifically recognizes the kappa light chains of human immunoglobulins. Anti-Kappa is used for in vitro identification of cells expressing the kappa light chains using flow cytometry.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AL, CA, CO, DC, DE, FL, IA, IL, IN, LA, MA, MD, ME, MI, MN, MO, NC, ND, NY, OH, OK, PA, TN, TX, UT, VA, WA, WI, WV and the country of Taiwan.
- How much
- 199 units
BD Anti-Kappa APC (Cat. No. 341098) is an analyte specific reagent for the Anti-Kappa antibody, which specific
Z-1214-2022 · Class II · reported 2022-06-08 · Ongoing
- Why it was recalled
- Anti-Kappa APC & Anti-Kappa PE analyte reagents may exhibit uncharacteristic flow profiles.
- Product
- BD Anti-Kappa APC (Cat. No. 341098) is an analyte specific reagent for the Anti-Kappa antibody, which specifically recognizes the kappa light chains of human immunoglobulins. Anti-Kappa is used for in vitro identification of cells expressing the kappa light chains using flow cytometry.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AL, CA, CO, DC, DE, FL, IA, IL, IN, LA, MA, MD, ME, MI, MN, MO, NC, ND, NY, OH, OK, PA, TN, TX, UT, VA, WA, WI, WV and the country of Taiwan.
- How much
- 249 units
BD" Mouse IgG2a Isotype Control APC X39 ASR
Z-2333-2021 · Class II · reported 2021-09-01 · Terminated
- Why it was recalled
- Complaints have been received for performance issues related to high background and staining. this issues is related to a manufacturing process error that can potentially cause a false positive staining due to contamination with another antibody.
- Product
- BD" Mouse IgG2a Isotype Control APC X39 ASR
- Where it went
- Worldwide distribution - U.S. Nationwide distribution in the states of CT, FL, KY, NY, and PA. The country of Taiwan.
- How much
- 23 reagents
BD" Multi-Check CD4 Low Control 2.5 mL IVD - Product Usage: intended as a complete process control for immunop
Z-0562-2021 · Class II · reported 2020-12-16 · Terminated
- Why it was recalled
- Recent product complaints showing an unexpected cell population exhibiting dim CD3 expression that may result in inaccurate results gating.
- Product
- BD" Multi-Check CD4 Low Control 2.5 mL IVD - Product Usage: intended as a complete process control for immunophenotyping by flow cytometry.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of AL, CA, CO, CT, DE, FL, GA, IL, IN, KY, MA, MD, MI, NC, NJ, NV, NY, OH, OK, PA, PR, RI, SD, TN, TX, and VA. The countries of Canada, Belgium, Korea, India, Ecuador, Japan, Costa Rica, Australia, Vietnam, Pakistan, Singapore, Thailand, and Peru.
- How much
- 619 vials
BD FACS Sample Prep Assistant (SPA) II - Product Usage: K050191: The BD FACSCanto system with BD FACSCanto cli
Z-2356-2020 · Class II · reported 2020-06-17 · Terminated
- Why it was recalled
- The recalling firm confirmed that BD FACS Sample Prep Assistant (SPA) II and III users may encounter probes with unacceptable wear prior to reaching the 2500 piercings guideline indicated in the Instructions for Use (IFU). When the probe is worn, particles from the rubber cap are more likely to clog the probe, resulting in blocked or inadequate sample dispense.
- Product
- BD FACS Sample Prep Assistant (SPA) II - Product Usage: K050191: The BD FACSCanto system with BD FACSCanto clinical software when used with the BD FACS Sample Prep Assistant II /III is intended for use as an in vitro diagnostic device for identification and enumeration of lymphocyte subsets in human cells in suspension using a lyse / no-wash sample preparation method for flow cytometry.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of AL, CO,CT, GA, IL, MA, MI, MD, NC, NH,NJ, NY, OH, PA, TN, TX, VA, and VT. The countries of Australia, Botswana, Brazil, Canada, Chile, China, Colombia, Costa Rica, Europe, Hong Kong, India, Italy, Jamaica, Japan, Kazakhstan, Malaysia, Martinique, Mexico, New Zealand, Nicaragua, Peru, Puerto Rico (US), Romania, Russia, Singapore, South Africa, South Korea, Swaziland, Switzerland, Turkey, US, Vietnam.
- How much
- 95 units
BD FACS Sample Prep Assistant (SPA) III - Product Usage: Prep Assistant III is intended to prepare human whole
Z-2358-2020 · Class II · reported 2020-06-17 · Terminated
- Why it was recalled
- The recalling firm confirmed that SPA II and III users may encounter probes with unacceptable wear prior to reaching the 2500 piercings guideline indicated in the Instructions for Use (IFU). When the probe is worn, particles from the rubber cap are more likely to clog the probe, resulting in blocked or inadequate sample dispense.
- Product
- BD FACS Sample Prep Assistant (SPA) III - Product Usage: Prep Assistant III is intended to prepare human whole blood for flow cytometric analysis on BD FACSCanto II and BD FACSCalibur flow cytometry systems. Pipetting blood, reagents, and lysing solution using the following previously cleared assays for flow cytometric analysis on BD FACSCanto II flow cytometry systems
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of AL, CO,CT, GA, IL, MA, MI, MD, NC, NH,NJ, NY, OH, PA, TN, TX, VA, and VT. The countries of Australia, Botswana, Brazil, Canada, Chile, China, Colombia, Costa Rica, Europe, Hong Kong, India, Italy, Jamaica, Japan, Kazakhstan, Malaysia, Martinique, Mexico, New Zealand, Nicaragua, Peru, Puerto Rico (US), Romania, Russia, Singapore, South Africa, South Korea, Swaziland, Switzerland, Turkey, US, Vietnam.
- How much
- 467 units
BD FACS Sample Prep Assistant (SPA) II Upgrade - Product Usage: K050191: The BD FACSCanto system with BD FACSC
Z-2357-2020 · Class II · reported 2020-06-17 · Terminated
- Why it was recalled
- The recalling firm confirmed that SPA II and III users may encounter probes with unacceptable wear prior to reaching the 2500 piercings guideline indicated in the Instructions for Use (IFU). When the probe is worn, particles from the rubber cap are more likely to clog the probe, resulting in blocked or inadequate sample dispense.
- Product
- BD FACS Sample Prep Assistant (SPA) II Upgrade - Product Usage: K050191: The BD FACSCanto system with BD FACSCanto clinical software when used with the BD FACS Sample Prep Assistant II /III is intended for use as an in vitro diagnostic device for identification and enumeration of lymphocyte subsets in human cells in suspension using a lyse / no-wash sample preparation method for flow cytometry.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of AL, CO,CT, GA, IL, MA, MI, MD, NC, NH,NJ, NY, OH, PA, TN, TX, VA, and VT. The countries of Australia, Botswana, Brazil, Canada, Chile, China, Colombia, Costa Rica, Europe, Hong Kong, India, Italy, Jamaica, Japan, Kazakhstan, Malaysia, Martinique, Mexico, New Zealand, Nicaragua, Peru, Puerto Rico (US), Romania, Russia, Singapore, South Africa, South Korea, Swaziland, Switzerland, Turkey, US, Vietnam.
- How much
- 51 units
Analyte specific reagent. REF/UDI: 641408/641408-4012-1-1 (US); 641417/641417-1157-1-01 (OUS); 641399/641399-4
Z-1971-2020 · Class II · reported 2020-05-20 · Terminated
- Why it was recalled
- False positive reaction on the reagent.
- Product
- Analyte specific reagent. REF/UDI: 641408/641408-4012-1-1 (US); 641417/641417-1157-1-01 (OUS); 641399/641399-4012-1-01 (US Research Use Only) - Product Usage: CD45 is intended for in vitro diagnostic use in the identification of cells expressing the CD45 antigen, using a BD FAC brand flow cytometer. The flow cytometer must be equipped to detect light scatter and the appropriate fluorescence, and be equipped with appropriate software (such as BD CellQuest, BD CellQuest Pro, BD FACSDiva, or BD FACSCanto clinical software) for data acquisition and analysis.
- Where it went
- Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AZ, CA, CT, FL, GA, IL, IN, LA, MA, MD, MI, MN, MO, NC, NH, NJ, NM, NV, NY, OH, OK, PA, TN, TX, VA, WA, WI and Puerto Rico. The countries of Argentina, Australia, Belgium, Brazil, Canada, China, Hong Kong, India, Japan, Korea, Malaysia, New Zealand, Singapore, Taiwan and Vietnam.
- How much
- 1473 vials.
BD FACSLyric Flow Cytometer, 3-Laser 12 Color Instrument, REF: 663029, when using BD Trucount Tubes
Z-0074-2020 · Class II · reported 2019-10-16 · Terminated
- Why it was recalled
- Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. This may affect the following: initiation of prophylactic therapy for Opportunistic Infections (HIV patients), decisions regarding care/management (PID/immune suppressed patients), laboratory developed tests/experiments, specimen engraftment potency, and use of leucoreduced blood products for transfusion.
- Product
- BD FACSLyric Flow Cytometer, 3-Laser 12 Color Instrument, REF: 663029, when using BD Trucount Tubes
- Where it went
- U.S.: CA, KS, NJ, MI, NY, PA, IL, IN, VA, WA, MD, IA, OK, RI, FL, CO, OH, TX, TN, GA, MO, MN, HI, VT, MA, NE. O.U.S. (Foreign): Germany, Italy, South Korea, Japan, United Kingdom, Poland, France, Belgium, Canada, Latvia, India, Taiwan, Ireland, Algeria, Israel, Iraq, Netherlands, Malaysia, Australia, Sweden, Colombia, Switzerland, Slovakia, Spain, Brazil, Austria, Peru, Iran, Croatia, Lithuania, Czech Republic, Saudi Arabia, Finland, Qatar, South Africa, Denmark, Turkey, Mexico, Indonesia, Ecuador, Vietnam, Singapore
- How much
- 22
BD FACSLyric Flow Cytometer, 2-Laser 4 Color Instrument, REF: 662875, when using BD Trucount Tubes
Z-0066-2020 · Class II · reported 2019-10-16 · Terminated
- Why it was recalled
- Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. This may affect the following: initiation of prophylactic therapy for Opportunistic Infections (HIV patients), decisions regarding care/management (PID/immune suppressed patients), laboratory developed tests/experiments, specimen engraftment potency, and use of leucoreduced blood products for transfusion.
- Product
- BD FACSLyric Flow Cytometer, 2-Laser 4 Color Instrument, REF: 662875, when using BD Trucount Tubes
- Where it went
- U.S.: CA, KS, NJ, MI, NY, PA, IL, IN, VA, WA, MD, IA, OK, RI, FL, CO, OH, TX, TN, GA, MO, MN, HI, VT, MA, NE. O.U.S. (Foreign): Germany, Italy, South Korea, Japan, United Kingdom, Poland, France, Belgium, Canada, Latvia, India, Taiwan, Ireland, Algeria, Israel, Iraq, Netherlands, Malaysia, Australia, Sweden, Colombia, Switzerland, Slovakia, Spain, Brazil, Austria, Peru, Iran, Croatia, Lithuania, Czech Republic, Saudi Arabia, Finland, Qatar, South Africa, Denmark, Turkey, Mexico, Indonesia, Ecuador, Vietnam, Singapore
- How much
- 12
BD FACSLyric Flow Cytometer, 3-Laser 8 Color Instrument, REF: 662877, when using BD Trucount Tubes
Z-0068-2020 · Class II · reported 2019-10-16 · Terminated
- Why it was recalled
- Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. This may affect the following: initiation of prophylactic therapy for Opportunistic Infections (HIV patients), decisions regarding care/management (PID/immune suppressed patients), laboratory developed tests/experiments, specimen engraftment potency, and use of leucoreduced blood products for transfusion.
- Product
- BD FACSLyric Flow Cytometer, 3-Laser 8 Color Instrument, REF: 662877, when using BD Trucount Tubes
- Where it went
- U.S.: CA, KS, NJ, MI, NY, PA, IL, IN, VA, WA, MD, IA, OK, RI, FL, CO, OH, TX, TN, GA, MO, MN, HI, VT, MA, NE. O.U.S. (Foreign): Germany, Italy, South Korea, Japan, United Kingdom, Poland, France, Belgium, Canada, Latvia, India, Taiwan, Ireland, Algeria, Israel, Iraq, Netherlands, Malaysia, Australia, Sweden, Colombia, Switzerland, Slovakia, Spain, Brazil, Austria, Peru, Iran, Croatia, Lithuania, Czech Republic, Saudi Arabia, Finland, Qatar, South Africa, Denmark, Turkey, Mexico, Indonesia, Ecuador, Vietnam, Singapore
- How much
- 8
BD FACSLyric Flow Cytometer, 2-Laser 4 Color Instrument, REF: 651164, when using BD Trucount Tubes
Z-0070-2020 · Class II · reported 2019-10-16 · Terminated
- Why it was recalled
- Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. This may affect the following: initiation of prophylactic therapy for Opportunistic Infections (HIV patients), decisions regarding care/management (PID/immune suppressed patients), laboratory developed tests/experiments, specimen engraftment potency, and use of leucoreduced blood products for transfusion.
- Product
- BD FACSLyric Flow Cytometer, 2-Laser 4 Color Instrument, REF: 651164, when using BD Trucount Tubes
- Where it went
- U.S.: CA, KS, NJ, MI, NY, PA, IL, IN, VA, WA, MD, IA, OK, RI, FL, CO, OH, TX, TN, GA, MO, MN, HI, VT, MA, NE. O.U.S. (Foreign): Germany, Italy, South Korea, Japan, United Kingdom, Poland, France, Belgium, Canada, Latvia, India, Taiwan, Ireland, Algeria, Israel, Iraq, Netherlands, Malaysia, Australia, Sweden, Colombia, Switzerland, Slovakia, Spain, Brazil, Austria, Peru, Iran, Croatia, Lithuania, Czech Republic, Saudi Arabia, Finland, Qatar, South Africa, Denmark, Turkey, Mexico, Indonesia, Ecuador, Vietnam, Singapore
- How much
- 21
BD FACSLyric Flow Cytometer, 3-Laser 10 Color Instrument, REF: 662878, and 66287808 when using BD Trucount Tub
Z-0069-2020 · Class II · reported 2019-10-16 · Terminated
- Why it was recalled
- Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. This may affect the following: initiation of prophylactic therapy for Opportunistic Infections (HIV patients), decisions regarding care/management (PID/immune suppressed patients), laboratory developed tests/experiments, specimen engraftment potency, and use of leucoreduced blood products for transfusion.
- Product
- BD FACSLyric Flow Cytometer, 3-Laser 10 Color Instrument, REF: 662878, and 66287808 when using BD Trucount Tubes
- Where it went
- U.S.: CA, KS, NJ, MI, NY, PA, IL, IN, VA, WA, MD, IA, OK, RI, FL, CO, OH, TX, TN, GA, MO, MN, HI, VT, MA, NE. O.U.S. (Foreign): Germany, Italy, South Korea, Japan, United Kingdom, Poland, France, Belgium, Canada, Latvia, India, Taiwan, Ireland, Algeria, Israel, Iraq, Netherlands, Malaysia, Australia, Sweden, Colombia, Switzerland, Slovakia, Spain, Brazil, Austria, Peru, Iran, Croatia, Lithuania, Czech Republic, Saudi Arabia, Finland, Qatar, South Africa, Denmark, Turkey, Mexico, Indonesia, Ecuador, Vietnam, Singapore
- How much
- 78
BD FACSLyric Flow Cytometer, 2-Laser 6 Color Instrument, REF: 662876, when using BD Trucount Tubes
Z-0067-2020 · Class II · reported 2019-10-16 · Terminated
- Why it was recalled
- Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. This may affect the following: initiation of prophylactic therapy for Opportunistic Infections (HIV patients), decisions regarding care/management (PID/immune suppressed patients), laboratory developed tests/experiments, specimen engraftment potency, and use of leucoreduced blood products for transfusion.
- Product
- BD FACSLyric Flow Cytometer, 2-Laser 6 Color Instrument, REF: 662876, when using BD Trucount Tubes
- Where it went
- U.S.: CA, KS, NJ, MI, NY, PA, IL, IN, VA, WA, MD, IA, OK, RI, FL, CO, OH, TX, TN, GA, MO, MN, HI, VT, MA, NE. O.U.S. (Foreign): Germany, Italy, South Korea, Japan, United Kingdom, Poland, France, Belgium, Canada, Latvia, India, Taiwan, Ireland, Algeria, Israel, Iraq, Netherlands, Malaysia, Australia, Sweden, Colombia, Switzerland, Slovakia, Spain, Brazil, Austria, Peru, Iran, Croatia, Lithuania, Czech Republic, Saudi Arabia, Finland, Qatar, South Africa, Denmark, Turkey, Mexico, Indonesia, Ecuador, Vietnam, Singapore
- How much
- 1
BD FACSLyric Flow Cytometer,3-Laser 10 Color Instrument, REF: 659180 and 65918001, when using BD Trucount Tube
Z-0073-2020 · Class II · reported 2019-10-16 · Terminated
- Why it was recalled
- Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. This may affect the following: initiation of prophylactic therapy for Opportunistic Infections (HIV patients), decisions regarding care/management (PID/immune suppressed patients), laboratory developed tests/experiments, specimen engraftment potency, and use of leucoreduced blood products for transfusion.
- Product
- BD FACSLyric Flow Cytometer,3-Laser 10 Color Instrument, REF: 659180 and 65918001, when using BD Trucount Tubes
- Where it went
- U.S.: CA, KS, NJ, MI, NY, PA, IL, IN, VA, WA, MD, IA, OK, RI, FL, CO, OH, TX, TN, GA, MO, MN, HI, VT, MA, NE. O.U.S. (Foreign): Germany, Italy, South Korea, Japan, United Kingdom, Poland, France, Belgium, Canada, Latvia, India, Taiwan, Ireland, Algeria, Israel, Iraq, Netherlands, Malaysia, Australia, Sweden, Colombia, Switzerland, Slovakia, Spain, Brazil, Austria, Peru, Iran, Croatia, Lithuania, Czech Republic, Saudi Arabia, Finland, Qatar, South Africa, Denmark, Turkey, Mexico, Indonesia, Ecuador, Vietnam, Singapore
- How much
- 236
BD FACSLyric Flow Cytometer, 2-Laser 6 Color Instrument, REF: 651165, when using BD Trucount Tubes
Z-0071-2020 · Class II · reported 2019-10-16 · Terminated
- Why it was recalled
- Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. This may affect the following: initiation of prophylactic therapy for Opportunistic Infections (HIV patients), decisions regarding care/management (PID/immune suppressed patients), laboratory developed tests/experiments, specimen engraftment potency, and use of leucoreduced blood products for transfusion.
- Product
- BD FACSLyric Flow Cytometer, 2-Laser 6 Color Instrument, REF: 651165, when using BD Trucount Tubes
- Where it went
- U.S.: CA, KS, NJ, MI, NY, PA, IL, IN, VA, WA, MD, IA, OK, RI, FL, CO, OH, TX, TN, GA, MO, MN, HI, VT, MA, NE. O.U.S. (Foreign): Germany, Italy, South Korea, Japan, United Kingdom, Poland, France, Belgium, Canada, Latvia, India, Taiwan, Ireland, Algeria, Israel, Iraq, Netherlands, Malaysia, Australia, Sweden, Colombia, Switzerland, Slovakia, Spain, Brazil, Austria, Peru, Iran, Croatia, Lithuania, Czech Republic, Saudi Arabia, Finland, Qatar, South Africa, Denmark, Turkey, Mexico, Indonesia, Ecuador, Vietnam, Singapore
- How much
- 36
BD FACSLyric Flow Cytometer, 3-Laser 8 Color Instrument, REF: 654587, when using BD Trucount Tubes
Z-0072-2020 · Class II · reported 2019-10-16 · Terminated
- Why it was recalled
- Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. This may affect the following: initiation of prophylactic therapy for Opportunistic Infections (HIV patients), decisions regarding care/management (PID/immune suppressed patients), laboratory developed tests/experiments, specimen engraftment potency, and use of leucoreduced blood products for transfusion.
- Product
- BD FACSLyric Flow Cytometer, 3-Laser 8 Color Instrument, REF: 654587, when using BD Trucount Tubes
- Where it went
- U.S.: CA, KS, NJ, MI, NY, PA, IL, IN, VA, WA, MD, IA, OK, RI, FL, CO, OH, TX, TN, GA, MO, MN, HI, VT, MA, NE. O.U.S. (Foreign): Germany, Italy, South Korea, Japan, United Kingdom, Poland, France, Belgium, Canada, Latvia, India, Taiwan, Ireland, Algeria, Israel, Iraq, Netherlands, Malaysia, Australia, Sweden, Colombia, Switzerland, Slovakia, Spain, Brazil, Austria, Peru, Iran, Croatia, Lithuania, Czech Republic, Saudi Arabia, Finland, Qatar, South Africa, Denmark, Turkey, Mexico, Indonesia, Ecuador, Vietnam, Singapore
- How much
- 30
BD FACSLyric 3L 10C with FACSuite Clinical Software v1.1.1, catalog number 662878 BD FACSLyric is a high-pe
Z-0792-2019 · Class III · reported 2019-02-13 · Terminated
- Why it was recalled
- A customer site was identified to have three FACSLyric IVD systems with FACSuite clinical v1.1.1 installed on a computer workstation with windows 10 operating system (OS). Windows 10 OS is not a released configuration for clinical use. It is released for research use only.
- Product
- BD FACSLyric 3L 10C with FACSuite Clinical Software v1.1.1, catalog number 662878 BD FACSLyric is a high-performance flow cytometer designed to support both routine clinical analysis and research for the identification, quantification, and characterization of cells in support of cell analysis. BD FACSuite software is loaded onto an adjacent PC workstation and is used to operate the instrument, acquire samples, and analyze data. Quality control performance, tracking, and reporting are streamlined and automated. Routine tasks such as daily cleaning and shutdown can be programmed to occur automatically. The BD FACSuite clinical software allows users to run BD IVD assays.
- Where it went
- US Distribution in MD.
- How much
- 3 devices
BD FACS Universal Loader (Optional) to the BD FACSLyric Clinical System, Model 662879 (Universal Loader), Rx O
Z-2598-2018 · Class III · reported 2018-08-08 · Terminated
- Why it was recalled
- The solenoid plunger tip length is too short resulting in insufficient engagement to the Universal Loader enclosure to lock during use.
- Product
- BD FACS Universal Loader (Optional) to the BD FACSLyric Clinical System, Model 662879 (Universal Loader), Rx Only. The firm name on the label is Becton, Dickinson and Company, BD Biosciences, San Jose, CA. The Loader offers various settings to resuspend and mix samples. It can draw from 12 x 75-mm tubes in 30- and 40-tube racks. A barcode reader verifies the ID on tube racks and individual tubes in 30- tube racks. A built-in imaging system provides safety checks, such as verifying the correct rack type and tube layout, and ensures the tubes were loaded correctly.
- Where it went
- US Distribution was made to FL, KS, KY, MA, NJ, and SD.
- How much
- 8 devices
CD79b PE (ASR) , Catalog number 335816 ; Hematology: The CD79b antibody recognizes an epitope on the extra
Z-0128-2017 · Class II · reported 2016-10-26 · Terminated
- Why it was recalled
- CD79b (SN8) PE is contaminated with a FITC conjugate of unknown antigenic specificity
- Product
- CD79b PE (ASR) , Catalog number 335816 ; Hematology: The CD79b antibody recognizes an epitope on the extracellular domain of a 3640 kilodalton (kDa) type I membrane glycoprotein. Immunoglobulin (Ig) antigen receptors are composed of a non-covalently associated complex of Ig and two other proteins, Iga and Igb, which have been designated as CD79a and CD79b, respectively. antigen
- Where it went
- Nationwide Distribution to CA, MN, OH and TX.
- How much
- 14 units
BD Tritest CD3 FITC/CD4 PE/CD45 PerCP (IVD); Catalog number 340383. Label states: Contains CD3 FITC/CD4 PE/C
Z-2378-2016 · Class III · reported 2016-08-17 · Terminated
- Why it was recalled
- Internal review found that some vials of CD3/CD4/CD45 state that the product contains gelatin. The product does not contain gelatin and should state in buffer.
- Product
- BD Tritest CD3 FITC/CD4 PE/CD45 PerCP (IVD); Catalog number 340383. Label states: Contains CD3 FITC/CD4 PE/CD45 per CP with Gelatin and 0.1% sodium azide. Hematology: BD Tritest CD3 fluorescein isothiocyanate (FITC)/CD4 phycoerythrin (PE)/CD45 peridinin chlorophyll protein (PerCP) is a three-color direct immunofluorescence reagent for use with a suitably equipped flow cytometer to identify and determine the percentages and absolute counts of mature human T lymphocytes (CD3+) and helper/inducer (CD3+CD4+) Tlymphocyte subsets in erythrocyte-lysed whole blood.
- Where it went
- Nationwide distribution to AR, CA, CO, IL, IN, LA, MA, MD, Ml, MN, NC, NJ, NM, NY, OR, PA, TN, TX, WA
- How much
- 14,089
BD Trucount Tubes: Catalog number: 340334 Hematology: BD Trucount tubes are used for determining absol
Z-2252-2016 · Class II · reported 2016-07-27 · Terminated
- Why it was recalled
- One lot of BD Trucount Absolute Counting Tubes (IVD) may not be sealed properly and exposure to ambient environment can rapidly degrade the performance of the product.
- Product
- BD Trucount Tubes: Catalog number: 340334 Hematology: BD Trucount tubes are used for determining absolute counts of leucocytes in blood.
- Where it went
- USA Distribution to the states of : NC, WA, WI, NJ, PR, MD, NY, MA, CA, NY , MD, KY, IN, VA, SD, PA, OH, AR, AZ and Foreign to : Canada, Belgium and Chile
- How much
- 328 US, 516 International
BD Trucount Absolute Counting Tubes: Model number: 340334 IVD; BD Trucount tubes are used for determining
Z-1570-2016 · Class II · reported 2016-05-04 · Terminated
- Why it was recalled
- One lot of lot of BD Trucount tubes may contain tubes from another lot. There is a 3.3% difference in bead counts between the two lots found within some of the pouches and presents a low risk to patient safety or impact to data if this product is used.
- Product
- BD Trucount Absolute Counting Tubes: Model number: 340334 IVD; BD Trucount tubes are used for determining absolute counts of leucocytes in blood. BD Trucount tubes are designed for use with in vitro diagnostic products such as BD Tritest reagents, and a suitably equipped flow cytometer. BD Trucount tubes can be used with the BD FACS Loader.
- Where it went
- Worldwide Distribution - US including MD, CA GA. WA. PA. CO. TN. FL. NC. DE. VA, CT, LA. NC, NY, OH, MN, NH and Internationally to China, Japan, and India.
- How much
- 1743; 467 US and 1276 International
BD-Retic-Count Reticulocyte Reagent System; Hematology: Used for in vitro diagnostics
Z-1390-2016 · Class III · reported 2016-04-20 · Terminated
- Why it was recalled
- The carton label for one lot of BD Retic-Count Reticulocyte Reagent System had the incorrect storage conditions. The labels on the reagent as well as the Instructions for Use are correct.
- Product
- BD-Retic-Count Reticulocyte Reagent System; Hematology: Used for in vitro diagnostics
- Where it went
- Worldwide Distribution - US Distribution to the states of : CT, MN, FL, NY, CA, MA, DC, UT, VA and IN., and to the countries of : Belgium, Brazil, Canada, China, India, Japan, Korea, Singapore and Taiwan.
- How much
- 205
BD FMC7 FITC: Model 430918. FMC7 is intended for in vitro diagnostic use in the identification of cells expre
Z-0851-2016 · Class III · reported 2016-03-02 · Terminated
- Why it was recalled
- Becton, Dickinson and Company (BD) has determined that the FMC7 FITC (ASR) has an incorrect expiration date listed on the vial label.
- Product
- BD FMC7 FITC: Model 430918. FMC7 is intended for in vitro diagnostic use in the identification of cells expressing FMC7 antigen. Product Usage: Analyte Specific Reagent. FMC7 is intended for in vitro diagnostic use in the identification of cells expressing FMC7 antigen.
- Where it went
- Worldwide Distribution - US Nationwide in the staates of TX, NY, GA, NJ, AL, MA, IL, PA, DC, FL, CA, MD, OK. and coutry of Taiwan.
- How much
- 78
Showing the 30 most recent of 35. The rest are searchable at the FDA enforcement report database.
Read these counts with three things in mind
A firm with more recalls usually makes more products. A contract packer runs
hundreds of lines. It will out-recall a company with one product whatever either
does. This is not a safety ranking.
Most recalls here are voluntary. A firm found its own fault and pulled the stock.
That is the system working. It lands in this database looking the same as a recall
the FDA had to push for.
Terminated means the FDA considers the recall finished. It is not a verdict.
It does not mean anyone was harmed.
Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.