thejoe
Bard Access Systems Inc.
Bard Access Systems Inc.: 38 FDA recall reports
Every medical device recall the FDA has published for Bard Access Systems Inc.. The fault, and how the agency graded it.
38Reports
2Class I
16Ongoing
2026-07-08Most recent
2 Class I on this pageClass I means the FDA judged there was a reasonable probability of serious harm or death. Check the lot numbers against anything you hold.
What the classes mean
The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.
Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.
By class
By year reported
Checking whether it affects you
The lot or code number is what matters. Not the product name. A recall covers
specific lots. Yours may not be one of them.
Check the code on your package against the report. The FDA keeps the current
list at fda.gov/safety/recalls.
The reports
The text below is quoted from the FDA record word for word.
CK000495A Poly Midline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741121609 CK000566 Poly Midline Cathet
Z-2656-2026 · Class II · reported 2026-07-08 · Ongoing
- Why it was recalled
- Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
- Product
- CK000495A Poly Midline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741121609 CK000566 Poly Midline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741108051 CK000566 Poly Midline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741108051 The Poly Per-Q-Cath Midline catheters are indicated for short or long term peripheral access to the peripheral venous system for selected intravenous therapies and blood sarnpling.
- Where it went
- Worldwide U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The country of Belgium.
- How much
- 3,240
5855240 Power-Trialysis Slim-Cath Straight Short-Term Dialysis Catheter 12 Fr Triple-Lumen UDI-DI Code: 00801
Z-2661-2026 · Class II · reported 2026-07-08 · Ongoing
- Why it was recalled
- Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
- Product
- 5855240 Power-Trialysis Slim-Cath Straight Short-Term Dialysis Catheter 12 Fr Triple-Lumen UDI-DI Code: 00801741107177 5855300 Power-Trialysis Slim-Cath Straight Short-Term Dialysis Catheter 12 Fr Triple-Lumen UDI-DI Code: 00801741107184 5865120 Power-Trialysis Slim-Cath Precurved Extension Short-Term Dialysis Catheter 12 Fr Triple-Lumen UDI-DI Code: 00801741107191 5865200 Power-Trialysis Slim-Cath Precurved Extension Short-Term Dialysis Catheter 12 Fr Triple-Lumen UDI_DI Code: 00801741107214 5865240 Power-Trialysis Slim-Cath Precurved Extension Short-Term Dialysis Catheter 12 Fr Triple-Lumen UDI-DI Code: 00801741107221 The Power-Trialysis Slim-Cath Short-Term Dialysis Catheter, with a third internal lumen for intravenous therapy, power injection of contrast media, and central venous pressure monitoring, is indicated for use in attaining short-term (less than 30 days) vascular access for hemodialysis, hemoperfusion, and apheresis treatments. The catheter is intended to be inserted in the jugular, femoral, or subclavian vein as required. The maximum recommended infusion rate is 5 mL/sec for power injection of contrast media.
- Where it went
- Worldwide U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The country of Belgium.
- How much
- 915
CK000727 PowerMidline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741138478 CK000729 PowerMidline Cathete
Z-2662-2026 · Class II · reported 2026-07-08 · Ongoing
- Why it was recalled
- Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
- Product
- CK000727 PowerMidline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741138478 CK000729 PowerMidline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741138713 P4153108D PowerMidline Catheter 3 Fr Single-lumen UDI-DI Code: 00801741108747 P4154108 PowerMidline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741108754 P4154108D PowerMidline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741108761 The PowerMidline" Catheter is intended for short term peripheral access for selected intravenous therapies, blood sampling, and power injection of contrast media.
- Where it went
- Worldwide U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The country of Belgium.
- How much
- 16,887
CK000880 Provena Midline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741188770 S4153108BDP Provena Midline
Z-2667-2026 · Class II · reported 2026-07-08 · Ongoing
- Why it was recalled
- Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
- Product
- CK000880 Provena Midline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741188770 S4153108BDP Provena Midline Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741154003 S4153108BP Provena Midline Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741153990 S4153108D Provena Midline Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741153952 S4153108DCP Provena Midline Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741233258 S4153108DGP Provena Midline Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741154027 S4254108BD Provena Midline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741154140 S4254108BDP Provena Midline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741154164 S4254108D Provena Midline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741154119 S4254108DCP Provena Midline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741233234 S4254108DGP Provena Midline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741154188 SP4153108D Provena Midline Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741185052 SP4153108DG Provena Midline Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741185069 The Provena" Midline Catheters are intended for short term peripheral access for...
- Where it went
- Worldwide U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The country of Belgium.
- How much
- 24,348
5625240 DuoGlide Short-Term Straight Dialysis Catheter 13 Fr Dual-Lumen UDI-DI Code: 00801741042119 The
Z-2657-2026 · Class II · reported 2026-07-08 · Ongoing
- Why it was recalled
- Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
- Product
- 5625240 DuoGlide Short-Term Straight Dialysis Catheter 13 Fr Dual-Lumen UDI-DI Code: 00801741042119 The intended use of the Short-Term Dialysis Catheter is attaining temporary/short-term (less than 30 days) vascular access for hemodialysis, hemoperfusion, or apheresis therapy via the jugular, subclavian, or femoral vein.
- Where it went
- Worldwide U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The country of Belgium.
- How much
- 310
CK000525B PowerPICC Provena SOLO Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741236129 CK000883 PowerPICC P
Z-2664-2026 · Class II · reported 2026-07-08 · Ongoing
- Why it was recalled
- Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
- Product
- CK000525B PowerPICC Provena SOLO Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741236129 CK000883 PowerPICC Provena Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741188800 CK000884A PowerPICC Provena SOLO Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741219177 CK001000 PowerPICC Provena SOLO Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741223020 CK001066 PowerPICC Provena Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741228544 CK001068 PowerPICC Provena Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741228728 S1173108D PowerPICC Provena Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741130342 S1173108D1 PowerPICC Provena Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741154720 S1173108D4 PowerPICC Provena Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741155031 S1173108D5 PowerPICC Provena Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741155130 S1173108D6 PowerPICC Provena Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741232190 S1193108D3 PowerPICC Provena SOLO Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741154980 S1193108D4 PowerPICC Provena SOLO Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741155086 S1193108D5 PowerPICC Provena SOLO Catheter 3 Fr...
- Where it went
- Worldwide U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The country of Belgium.
- How much
- 29,490
1274108D PowerPICC SV Catheter 4 Fr Dual-Lumen REKW2449 00801741087110 3173108D PowerPICC SV Catheter 3 Fr Si
Z-2659-2026 · Class II · reported 2026-07-08 · Ongoing
- Why it was recalled
- Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
- Product
- 1274108D PowerPICC SV Catheter 4 Fr Dual-Lumen REKW2449 00801741087110 3173108D PowerPICC SV Catheter 3 Fr Single-Lumen REKV2227 00801741027536 The PowerPlCC SV catheters are intended for short or long term peripheral access to the central venous system for intravenous therapy and blood sampling.
- Where it went
- Worldwide U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The country of Belgium.
- How much
- 624
CK000662 PowerPICC SOLO HF Catheter 5 Fr Triple-Lumen UDI-DI Code: 00801741129674 Catheter stylets prov
Z-2658-2026 · Class II · reported 2026-07-08 · Ongoing
- Why it was recalled
- Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
- Product
- CK000662 PowerPICC SOLO HF Catheter 5 Fr Triple-Lumen UDI-DI Code: 00801741129674 Catheter stylets provide internal reinforcement to aid in catheter placement.
- Where it went
- Worldwide U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The country of Belgium.
- How much
- 66
OTP5000 THORA/PARA TRAY, 5FR, NON-VALVED UDI-DI Code: 0110885403284748 PIG1260T SAFE-T-CENTESIS TRAY 6FR X
Z-2655-2026 · Class II · reported 2026-07-08 · Ongoing
- Why it was recalled
- Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
- Product
- OTP5000 THORA/PARA TRAY, 5FR, NON-VALVED UDI-DI Code: 0110885403284748 PIG1260T SAFE-T-CENTESIS TRAY 6FR X 16CM UDI-DI Code: 0110885403108365 TPT1000 THORACENTESIS/PARACENTESIS TRAY UDI-DI Code: 0110885403108488 The Acute Drainage devices are indicated for use in removing fluid from either the pleural or peritoneal spaces during an acute drainage / aspiration procedure.
- Where it went
- Worldwide U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The country of Belgium.
- How much
- 9,500
7617408 Groshong NXT ClearVue Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741035258 The Groshong¿ NXT P
Z-2665-2026 · Class II · reported 2026-07-08 · Ongoing
- Why it was recalled
- Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
- Product
- 7617408 Groshong NXT ClearVue Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741035258 The Groshong¿ NXT PICC provides short (less than 30 days) or long (greater than 30 days) term peripheral access to the central venous system for intravenous therapy or blood sampling.
- Where it went
- Worldwide U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The country of Belgium.
- How much
- 120
CK000980A PowerMidline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741236150 P4254108 PowerMidline Catheter
Z-2663-2026 · Class II · reported 2026-07-08 · Ongoing
- Why it was recalled
- Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
- Product
- CK000980A PowerMidline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741236150 P4254108 PowerMidline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741129278 P4254108D PowerMidline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741129353 The PowerMidline" Catheter is intended for short term peripheral access for selected intravenous therapies, blood sampling, and power injection of contrast media.
- Where it went
- Worldwide U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The country of Belgium.
- How much
- 2,844
4380 ECONOMY SOFT TISSUE BIOPSY TRAY UDI-DI Code: 0110885403108563 4382A SOFT TISSUE BIOPSY TRAY UDI-DI C
Z-2668-2026 · Class II · reported 2026-07-08 · Ongoing
- Why it was recalled
- Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
- Product
- 4380 ECONOMY SOFT TISSUE BIOPSY TRAY UDI-DI Code: 0110885403108563 4382A SOFT TISSUE BIOPSY TRAY UDI-DI Code: 0110885403108501 The Soft Tissue Biopsy devices are indicated for use in the collection of aspirate tissue (e.g. liver, kidney, etc.).
- Where it went
- Worldwide U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The country of Belgium.
- How much
- 11,990
1174108 PowerPICC Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741154935 1295108 PowerPICC SOLO Catheter 5
Z-2666-2026 · Class II · reported 2026-07-08 · Ongoing
- Why it was recalled
- Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
- Product
- 1174108 PowerPICC Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741154935 1295108 PowerPICC SOLO Catheter 5 Fr Dual-Lumen UDI-DI Code: 00801741034602 1174108D3 PowerPICC Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741154935 1174108D4 PowerPICC Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741155048 1174108D5 PowerPICC Catheter 4 Fr Singel-Lumen UDI-DI Code: 00801741155147 1194108D PowerPICC SOLO Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741034510 1194108D1 PowerPICC SOLO Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741154782 1194108D4 PowerPICC SOLO Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741155093 1194108D5 PowerPICC SOLO Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741155192 1275108D PowerPICC Catheter 5 Fr Dual-Lumen UDI-DI Code: 00801741034558 1295108FD PowerPICC SOLO FT Catheter 5 Fr Dual-Lumen UDI-DI Code: 00801741034626 3194108D PowerPICC Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741027727 3275108D PowerPICC Catheter 5 Fr Dual-Lumen UDI-DI Code: 00801741027871 3295108D PowerPICC SOLO FT Catheter 5 Fr Dual-Lumen UDI-DI Code: 00801741028052 9194108D PowerPICC SOLO Catheter 4 Fr Single-Lumen UDI-DI Code:...
- Where it went
- Worldwide U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The country of Belgium.
- How much
- 39,245
5605150 Power-Trialysis Short-Term Straight Dialysis Catheter 13 Fr Triple-Lumen UDI-DI Code: 00801741086953
Z-2660-2026 · Class II · reported 2026-07-08 · Ongoing
- Why it was recalled
- Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
- Product
- 5605150 Power-Trialysis Short-Term Straight Dialysis Catheter 13 Fr Triple-Lumen UDI-DI Code: 00801741086953 5605240 Power-Trialysis Short-Term Straight Dialysis Catheter 13 Fr Triple-Lumen UDI-DI Code: 00801741086977 5605300 Power-Trialysis Short-Term Straight Dialysis Catheter 13 Fr Triple-Lumen UDI-DI Code: 00801741086984 5615120 Power-Trialysis Short-Term Curved Extension Dialysis Catheter 13 Fr Triple-Lumen UDI-DI Code: 00801741075087 5615200 Power-Trialysis Short-Term Curved Extension Dialysis Catheter 13 Fr Triple-Lumen UDI-DI Code: 00801741075100 5615200 Power-Trialysis Short-Term Curved Extension Dialysis Catheter 13 Fr Triple-Lumen UDI-DI Code: 00801741075100 5615200 Power-Trialysis Short-Term Curved Extension Dialysis Catheter 13 Fr Triple-Lumen UDI-DI Code: 00801741075100 5655120 Power-Trialysis Short-Term Alphacurve Dialysis Catheter 13 Fr Triple-Lumen UDI-DI Code: 00801741066054 The Power-Trialysis" Short-Term Dialysis Catheter, with a third internal lumen for intravenous therapy, power injection of contrast media, and central venous pressure monitoring, is indicated for use in attaining short-term (less than 30 days) vascular access...
- Where it went
- Worldwide U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The country of Belgium.
- How much
- 8,795
Class I
BD 4 Fr Single-Lumen PowerPICC Catheters, REF: 3174155, CK000540, 3174118, CK000483, 3174335, 3174355, CK00019
Z-1507-2025 · Class I · reported 2025-04-23 · Ongoing
- Why it was recalled
- Catheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher), which could lead to increased material fatigue leaks, which could lead to infiltration, extravasation, discomfort, phlebitis, bleeding, air embolism, foreign body embolism, infection and interruption to therapy.
- Product
- BD 4 Fr Single-Lumen PowerPICC Catheters, REF: 3174155, CK000540, 3174118, CK000483, 3174335, 3174355, CK000196, 9174118, CK000955, 3174135, CK000617, CK000839, 3194355, 6174118, 3194135, 3194335, 2174108, 9194118, 3194118, 1174118, CK000285C, CK000335A, CK000796, 1194108D, CK000757, CK001042, 9174108, 1174108D1, CK000631B, 3194155, 3174108, CK000567, 1194108D4, 1174108D2, CK000677, CK000864, CK000937A, CK000930, CK001027, 1174108D, CK000867, CK000900, CK000315A, CK000589A, CK000854B, CK000897, CK000953A, CK000586B, CK000549, CK000844, 1194108D1, 1174108D4, 1194108D3, CK000934, CK000986, CK000411, CK001018, 1174108, CK000060, CK000513A, CK001021, CK001030, CK000501A, CK000619A, CK000903, 3194108, 1194108D2, 9194108D, CK000959, CK000613, CK000991, CK000983, CK000881, CK000721, CK000642, CK000289C, CK000894A, CK000914, CK000445, CK000687, CK000524A, CK000819, CK000266A, 1174108D3, 3194108D, CK000545, 1174108D5, 3174108D, CK000775A, 1194108D5, CK000736, CK000739, 9194108, 1194108, CK000660, CK000609A, 9174108D, CK001067, CK000077D, CK000882, CK000874, CK000994, CK000870A, CK000443C, CK000668B, CK000784
- Where it went
- Worldwide - US Nationwide distribution including in the states of WI, NC, TX, NY, CA, MN, MT, RI, MA, OH, CT, SC, IL, CO, VA, NJ, GA, UT, FL, PA, IN, MD, KY, AZ, MI, WA, HI, MO, VT, AL, NM, TN, DE, LA, IA, NH, ID, DC, WV, KS, NE, SD, ND, WY, AK, OK, OR, AR and the countries of Andorra, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Finland, France, French Polynesia, French Guayana, Germany, Greece, Guadeloupe, Guatemala, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Luxembourg, Martinique, Mexico, Netherlands, New Caledonia, New Zealand, Norway, Pakistan, Poland, Portugal, Qatar, Republic of Korea, Reunion, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, Viet Nam.
- How much
- 718,461
Class I
Intraosseous needle and driver. Catalog/Device Name: D015151NK/BD Needle Kit for Powered Driver 15mm x 15G
Z-1397-2022 · Class I · reported 2022-08-03 · Ongoing
- Why it was recalled
- BD Intraosseous Needle Kits may exhibit: 1) Increased force required to remove the stylet from intraosseous needle during placement may cause removal of the entire needle assembly and loss of intraosseous access; 2) Stylet safety mechanism may not engage as the stylet is removed; 3) Metal discs in the powered driver may stick rendering the driver unusable.
- Product
- Intraosseous needle and driver. Catalog/Device Name: D015151NK/BD Needle Kit for Powered Driver 15mm x 15Ga, D015151MK/ BD Manual Driver Needle Kit 15mm x 15Ga, D015251NK/ BD Needle Kit for Powered Driver 25mm x 15Ga, D015251MK/ BD Manual Driver Needle Kit 25mm x 15Ga, D015351NK/ BD Needle Kit for Powered Driver 35mm x 15Ga, D015351MK/ BD Manual Driver Needle Kit 35mm x 15Ga, D015451NK/ BD Needle Kit for Powered Driver 45mm x 15Ga, D015451MK/ BD Manual Driver Needle Kit 45mm x 15Ga, D015551NK/ BD Needle Kit for Powered Driver 55mm x 15Ga, D015551MK/ BD Manual Driver Needle Kit 55mm x 15G, D001001/ BD Intraosseous Powered Driver (drill)
- Where it went
- Distribution US nationwide and Canada.
- How much
- 40,734
5 Fr IR PowerPICC Convenience Kit Catheter Dual Lumen with 70 cm Guidewire Unit Label: PowerPICC REF CK000
Z-1994-2021 · Class II · reported 2021-07-07 · Terminated
- Why it was recalled
- Catheter convenience kits contain the incorrect guidewire. Kits were labeled with 70cm guidewires, but contained a 50cm guidewire.
- Product
- 5 Fr IR PowerPICC Convenience Kit Catheter Dual Lumen with 70 cm Guidewire Unit Label: PowerPICC REF CK000340B BARD MYPICC KIT 5F French Size Dual-Lumen 70cm Guidewire
- Where it went
- U.S. Nationwide distribution in the state of RI.
- How much
- 30 kits
5Fr PowerPICC Catheter Single-Lumen with 70cm Guidewire Instructions For Use: PowerPicc Polyurethand Radio
Z-1991-2021 · Class II · reported 2021-07-07 · Terminated
- Why it was recalled
- Catheter kits contained the incorrect micro-introducers.
- Product
- 5Fr PowerPICC Catheter Single-Lumen with 70cm Guidewire Instructions For Use: PowerPicc Polyurethand Radiology PiCC with Microintroducer
- Where it went
- U.S.: CT, GA, MA, MI, MT, NJ, NV, NY, VA, and WI. O.U.S.: None
- How much
- 180 units
REF 7617400, Groshong NXT ClearVue Catheter, intermediate kit, 4F, Single-Lumen, 60 cm, UDI: (01)1080174104509
Z-1652-2021 · Class II · reported 2021-05-26 · Terminated
- Why it was recalled
- Catheters contain two issues: 1. the inner diameter of the distal connector may be reduced resulting in inability to pass the catheter through the connector and 2. the catch slot on the distal connecter may be deformed resulting in the inability to properly lock
- Product
- REF 7617400, Groshong NXT ClearVue Catheter, intermediate kit, 4F, Single-Lumen, 60 cm, UDI: (01)10801741045094 - Product Usage: is designed for use when central venous catheterization is prescribed.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, CA, CO, FL, GA, IA, IL, IN, KS, KY, LA, MA, MN, MO, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI and the countries of Australia, Belgium, Canada, China, Japan, Taiwan, Hong Kong, Malaysia, Chile, Bangladesh, Korea, Brazil, India, Singapore, Thailand.
- How much
- N/A
REF 7617405J, Groshong NXT ClearVue Catheter, Basic kit, 4F, Single-Lumen, 60 cm, UDI: (01)10801741200592 - Pr
Z-1653-2021 · Class II · reported 2021-05-26 · Terminated
- Why it was recalled
- Catheters contain two issues: 1. the inner diameter of the distal connector may be reduced resulting in inability to pass the catheter through the connector and 2. the catch slot on the distal connecter may be deformed resulting in the inability to properly lock
- Product
- REF 7617405J, Groshong NXT ClearVue Catheter, Basic kit, 4F, Single-Lumen, 60 cm, UDI: (01)10801741200592 - Product Usage: is designed for use when central venous catheterization is prescribed.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, CA, CO, FL, GA, IA, IL, IN, KS, KY, LA, MA, MN, MO, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI and the countries of Australia, Belgium, Canada, China, Japan, Taiwan, Hong Kong, Malaysia, Chile, Bangladesh, Korea, Brazil, India, Singapore, Thailand.
- How much
- N/A
REF 9617408, Groshong NXT ClearVue Catheter with Sherlock Tip Location System (TLS) Stylet, Full Kit, 4F, Sing
Z-1661-2021 · Class II · reported 2021-05-26 · Terminated
- Why it was recalled
- Catheters contain two issues: 1. the inner diameter of the distal connector may be reduced resulting in inability to pass the catheter through the connector and 2. the catch slot on the distal connecter may be deformed resulting in the inability to properly lock
- Product
- REF 9617408, Groshong NXT ClearVue Catheter with Sherlock Tip Location System (TLS) Stylet, Full Kit, 4F, Single-Lumen, TLS, UDI: (01)10801741036009 - Product Usage: is designed for use when central venous catheterization is prescribed.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, CA, CO, FL, GA, IA, IL, IN, KS, KY, LA, MA, MN, MO, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI and the countries of Australia, Belgium, Canada, China, Japan, Taiwan, Hong Kong, Malaysia, Chile, Bangladesh, Korea, Brazil, India, Singapore, Thailand.
- How much
- N/A
REF CK000379, Groshong NXT ClearVue Catheter with Sherlock Tip Location System (TLS) Stylet, Full Kit, 4F, Sin
Z-1666-2021 · Class II · reported 2021-05-26 · Terminated
- Why it was recalled
- Catheters contain two issues: 1. the inner diameter of the distal connector may be reduced resulting in inability to pass the catheter through the connector and 2. the catch slot on the distal connecter may be deformed resulting in the inability to properly lock
- Product
- REF CK000379, Groshong NXT ClearVue Catheter with Sherlock Tip Location System (TLS) Stylet, Full Kit, 4F, Single-Lumen, TLS, CE 2797, UDI: (01)10801741102704 - Product Usage: is designed for use when central venous catheterization is prescribed.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, CA, CO, FL, GA, IA, IL, IN, KS, KY, LA, MA, MN, MO, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI and the countries of Australia, Belgium, Canada, China, Japan, Taiwan, Hong Kong, Malaysia, Chile, Bangladesh, Korea, Brazil, India, Singapore, Thailand.
- How much
- N/A
REF 7655405, Groshong NXT ClearVue Catheter, Basic Kit, 4F, Single-Lumen, 60 cm, UDI: (01)10801741035279 - Pro
Z-1656-2021 · Class II · reported 2021-05-26 · Terminated
- Why it was recalled
- Catheters contain two issues: 1. the inner diameter of the distal connector may be reduced resulting in inability to pass the catheter through the connector and 2. the catch slot on the distal connecter may be deformed resulting in the inability to properly lock
- Product
- REF 7655405, Groshong NXT ClearVue Catheter, Basic Kit, 4F, Single-Lumen, 60 cm, UDI: (01)10801741035279 - Product Usage: is designed for use when central venous catheterization is prescribed.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, CA, CO, FL, GA, IA, IL, IN, KS, KY, LA, MA, MN, MO, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI and the countries of Australia, Belgium, Canada, China, Japan, Taiwan, Hong Kong, Malaysia, Chile, Bangladesh, Korea, Brazil, India, Singapore, Thailand.
- How much
- N/A
REF 7655405J, Groshong NXT ClearVue Catheter, Basic Kit, 4F, Single-Lumen, 60 cm, UDI: (01)10801741200615 - Pr
Z-1657-2021 · Class II · reported 2021-05-26 · Terminated
- Why it was recalled
- Catheters contain two issues: 1. the inner diameter of the distal connector may be reduced resulting in inability to pass the catheter through the connector and 2. the catch slot on the distal connecter may be deformed resulting in the inability to properly lock
- Product
- REF 7655405J, Groshong NXT ClearVue Catheter, Basic Kit, 4F, Single-Lumen, 60 cm, UDI: (01)10801741200615 - Product Usage: is designed for use when central venous catheterization is prescribed.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, CA, CO, FL, GA, IA, IL, IN, KS, KY, LA, MA, MN, MO, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI and the countries of Australia, Belgium, Canada, China, Japan, Taiwan, Hong Kong, Malaysia, Chile, Bangladesh, Korea, Brazil, India, Singapore, Thailand.
- How much
- N/A
REF 9617408D, Groshong NXT ClearVue Catheter with Sherlock Tip Location System (TLS) Stylet, Maximal Barrier K
Z-1662-2021 · Class II · reported 2021-05-26 · Terminated
- Why it was recalled
- Catheters contain two issues: 1. the inner diameter of the distal connector may be reduced resulting in inability to pass the catheter through the connector and 2. the catch slot on the distal connecter may be deformed resulting in the inability to properly lock
- Product
- REF 9617408D, Groshong NXT ClearVue Catheter with Sherlock Tip Location System (TLS) Stylet, Maximal Barrier Kit, 4F, Single-Lumen, TLS, UDI: (01)10801741036016 - Product Usage: is designed for use when central venous catheterization is prescribed.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, CA, CO, FL, GA, IA, IL, IN, KS, KY, LA, MA, MN, MO, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI and the countries of Australia, Belgium, Canada, China, Japan, Taiwan, Hong Kong, Malaysia, Chile, Bangladesh, Korea, Brazil, India, Singapore, Thailand.
- How much
- N/A
REF 9617405, Groshong NXT ClearVue Catheter with Sherlock Tip Location System (TLS) Stylet, Basic Kit, 4F, Sin
Z-1659-2021 · Class II · reported 2021-05-26 · Terminated
- Why it was recalled
- Catheters contain two issues: 1. the inner diameter of the distal connector may be reduced resulting in inability to pass the catheter through the connector and 2. the catch slot on the distal connecter may be deformed resulting in the inability to properly lock
- Product
- REF 9617405, Groshong NXT ClearVue Catheter with Sherlock Tip Location System (TLS) Stylet, Basic Kit, 4F, Single-Lumen, TLS, UDI: (01)10801741035989 - Product Usage: is designed for use when central venous catheterization is prescribed.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, CA, CO, FL, GA, IA, IL, IN, KS, KY, LA, MA, MN, MO, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI and the countries of Australia, Belgium, Canada, China, Japan, Taiwan, Hong Kong, Malaysia, Chile, Bangladesh, Korea, Brazil, India, Singapore, Thailand.
- How much
- N/A
REF 69655405, Groshong NXT ClearVue Catheter with Sherlock Tip Location System (TLS) Stylet, Basic Kit, 4F, Si
Z-1664-2021 · Class II · reported 2021-05-26 · Terminated
- Why it was recalled
- Catheters contain two issues: 1. the inner diameter of the distal connector may be reduced resulting in inability to pass the catheter through the connector and 2. the catch slot on the distal connecter may be deformed resulting in the inability to properly lock
- Product
- REF 69655405, Groshong NXT ClearVue Catheter with Sherlock Tip Location System (TLS) Stylet, Basic Kit, 4F, Single-Lumen, TLS, UDI: (01)10801741147521 - Product Usage: is designed for use when central venous catheterization is prescribed.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, CA, CO, FL, GA, IA, IL, IN, KS, KY, LA, MA, MN, MO, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI and the countries of Australia, Belgium, Canada, China, Japan, Taiwan, Hong Kong, Malaysia, Chile, Bangladesh, Korea, Brazil, India, Singapore, Thailand.
- How much
- N/A
REF 7617408D, Groshong NXT ClearVue Catheter, Maximal Barrier Kit, 4F, Single-Lumen, 60 cm, UDI: (01)108017410
Z-1655-2021 · Class II · reported 2021-05-26 · Terminated
- Why it was recalled
- Catheters contain two issues: 1. the inner diameter of the distal connector may be reduced resulting in inability to pass the catheter through the connector and 2. the catch slot on the distal connecter may be deformed resulting in the inability to properly lock
- Product
- REF 7617408D, Groshong NXT ClearVue Catheter, Maximal Barrier Kit, 4F, Single-Lumen, 60 cm, UDI: (01)10801741035262 - Product Usage: is designed for use when central venous catheterization is prescribed.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, CA, CO, FL, GA, IA, IL, IN, KS, KY, LA, MA, MN, MO, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI and the countries of Australia, Belgium, Canada, China, Japan, Taiwan, Hong Kong, Malaysia, Chile, Bangladesh, Korea, Brazil, India, Singapore, Thailand.
- How much
- N/A
REF 9617407, Groshong NXT ClearVue Catheter with Sherlock Tip Location System (TLS) Stylet, Full Kit, 4F, Sing
Z-1660-2021 · Class II · reported 2021-05-26 · Terminated
- Why it was recalled
- Catheters contain two issues: 1. the inner diameter of the distal connector may be reduced resulting in inability to pass the catheter through the connector and 2. the catch slot on the distal connecter may be deformed resulting in the inability to properly lock
- Product
- REF 9617407, Groshong NXT ClearVue Catheter with Sherlock Tip Location System (TLS) Stylet, Full Kit, 4F, Single-Lumen, TLS, UDI: (01)10801741035996 - Product Usage: is designed for use when central venous catheterization is prescribed.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, CA, CO, FL, GA, IA, IL, IN, KS, KY, LA, MA, MN, MO, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI and the countries of Australia, Belgium, Canada, China, Japan, Taiwan, Hong Kong, Malaysia, Chile, Bangladesh, Korea, Brazil, India, Singapore, Thailand.
- How much
- N/A
REF 9655405, Groshong NXT ClearVue Catheter with Sherlock Tip Location System (TLS) Stylet, Basic Kit, 4F, Sin
Z-1663-2021 · Class II · reported 2021-05-26 · Terminated
- Why it was recalled
- Catheters contain two issues: 1. the inner diameter of the distal connector may be reduced resulting in inability to pass the catheter through the connector and 2. the catch slot on the distal connecter may be deformed resulting in the inability to properly lock
- Product
- REF 9655405, Groshong NXT ClearVue Catheter with Sherlock Tip Location System (TLS) Stylet, Basic Kit, 4F, Single-Lumen, TLS, UDI: (01)10801741036023 - Product Usage: is designed for use when central venous catheterization is prescribed.
- Where it went
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, CA, CO, FL, GA, IA, IL, IN, KS, KY, LA, MA, MN, MO, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI and the countries of Australia, Belgium, Canada, China, Japan, Taiwan, Hong Kong, Malaysia, Chile, Bangladesh, Korea, Brazil, India, Singapore, Thailand.
- How much
- N/A
Showing the 30 most recent of 38. The rest are searchable at the FDA enforcement report database.
Read these counts with three things in mind
A firm with more recalls usually makes more products. A contract packer runs
hundreds of lines. It will out-recall a company with one product whatever either
does. This is not a safety ranking.
Most recalls here are voluntary. A firm found its own fault and pulled the stock.
That is the system working. It lands in this database looking the same as a recall
the FDA had to push for.
Terminated means the FDA considers the recall finished. It is not a verdict.
It does not mean anyone was harmed.
Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.
← All device recalls