Argon Medical Devices, Inc

Argon Medical Devices, Inc: 48 FDA recall reports

Every medical device recall the FDA has published for Argon Medical Devices, Inc. The fault, and how the agency graded it.

48Reports
8Class I
13Ongoing
2026-07-22Most recent
8 Class I on this page

Class I means the FDA judged there was a reasonable probability of serious harm or death. Check the lot numbers against anything you hold.

What the classes mean

The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.

ClassWhat the FDA is saying
Class IThere is a reasonable probability that using this will cause serious harm or death.
Class IIIt may cause temporary or reversible harm. Serious harm is unlikely.
Class IIIIt breaks a rule but is unlikely to hurt anyone. Often a labelling fault.

Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.

By class

ClassReports
Class II40
Class I8

By year reported

YearReports
20121
20143
20152
20162
20174
20199
20208
20218
20221
20231
20242
20251
20266
Checking whether it affects you

The lot or code number is what matters. Not the product name. A recall covers specific lots. Yours may not be one of them.

Check the code on your package against the report. The FDA keeps the current list at fda.gov/safety/recalls.

The reports

The text below is quoted from the FDA record word for word.

Class I

General Biospy Tray, REF: GUTS1000

Z-2706-2026  ·  Class I  ·  reported 2026-07-22  ·  Ongoing

Why it was recalled
Kits containing Lidocaine HCL Injection USP 1% 10mg/mL 5mL ampules subject to another recall recalled due to not meeting release criteria for sterility.
Product
General Biospy Tray, REF: GUTS1000
Where it went
Distribution US nationwide including Puerto Rico; OUS to Trinidad and Tobago.
How much
1800
Class I

Arterial Line Kit 20ga x 6, REF: 400115A

Z-2708-2026  ·  Class I  ·  reported 2026-07-22  ·  Ongoing

Why it was recalled
Kits containing Lidocaine HCL Injection USP 1% 10mg/mL 5mL ampules subject to another recall recalled due to not meeting release criteria for sterility.
Product
Arterial Line Kit 20ga x 6, REF: 400115A
Where it went
Distribution US nationwide including Puerto Rico; OUS to Trinidad and Tobago.
How much
440
Class I

Transvenous Pacing Electrode Tray w/ Bipolar Balloon Pacing Electrode 5F, REF: 008416A, 008566A

Z-2704-2026  ·  Class I  ·  reported 2026-07-22  ·  Ongoing

Why it was recalled
Kits containing Lidocaine HCL Injection USP 1% 10mg/mL 5mL ampules subject to another recall recalled due to not meeting release criteria for sterility.
Product
Transvenous Pacing Electrode Tray w/ Bipolar Balloon Pacing Electrode 5F, REF: 008416A, 008566A
Where it went
Distribution US nationwide including Puerto Rico; OUS to Trinidad and Tobago.
How much
950
Class I

One-Piece Introducer Tray, REF: 497455, 497434 cardiovascular

Z-2705-2026  ·  Class I  ·  reported 2026-07-22  ·  Ongoing

Why it was recalled
Kits containing Lidocaine HCL Injection USP 1% 10mg/mL 5mL ampules subject to another recall recalled due to not meeting release criteria for sterility.
Product
One-Piece Introducer Tray, REF: 497455, 497434 cardiovascular
Where it went
Distribution US nationwide including Puerto Rico; OUS to Trinidad and Tobago.
How much
200
Class I

T-Lok Bone Marrow Tray, REF: BMT1104STL; BMET1104TL; BMT1104CATL

Z-2707-2026  ·  Class I  ·  reported 2026-07-22  ·  Ongoing

Why it was recalled
Kits containing Lidocaine HCL Injection USP 1% 10mg/mL 5mL ampules subject to another recall recalled due to not meeting release criteria for sterility.
Product
T-Lok Bone Marrow Tray, REF: BMT1104STL; BMET1104TL; BMT1104CATL
Where it went
Distribution US nationwide including Puerto Rico; OUS to Trinidad and Tobago.
How much
850
Class I

Bone Marrow Tray, REF: CB0033TL; CB0355TLA

Z-2709-2026  ·  Class I  ·  reported 2026-07-22  ·  Ongoing

Why it was recalled
Kits containing Lidocaine HCL Injection USP 1% 10mg/mL 5mL ampules subject to another recall recalled due to not meeting release criteria for sterility.
Product
Bone Marrow Tray, REF: CB0033TL; CB0355TLA
Where it went
Distribution US nationwide including Puerto Rico; OUS to Trinidad and Tobago.
How much
680

Option"ELITE Vena Cava Filter System REF 352506070E UDI-DI code: 00886333217151 The Option"ELITE Filter i

Z-0589-2026  ·  Class II  ·  reported 2025-12-03  ·  Ongoing

Why it was recalled
Due to complaints of increased resistance when advancing the dilator within the introducer sheath. This issue could potential result in procedural delays and/or risk of venous intimal injuries if excessive force is applied
Product
Option"ELITE Vena Cava Filter System REF 352506070E UDI-DI code: 00886333217151 The Option"ELITE Filter is indicated for the prevention of recurrent pulmonary embolism (PE) via percutaneous placement in the inferior vena cava (IVC) in the following conditions: " Pulmonary thromboembolism when anticoagulants are contraindicated " Failure of anticoagulant therapy in thromboembolic disease " Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced " Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated
Where it went
U.S. (nationwide) distribution to states of: AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, and WV. O.U.S. (international) to countries of: China, Panama and Trinidad and Tobago
How much
Total devices 1821 (US=1794 and O.U.S.=27) units

BioPince Ultra Full Core Biopsy Instrument: REF: REF: 360-1080-01, 360-1080-02, 360-1080-03, 360-1580-01, 360-

Z-3192-2024  ·  Class II  ·  reported 2024-10-02  ·  Ongoing

Why it was recalled
Holes have been identified in biopsy instrument packaging trays compromising the sterile barrier, which may lead to cross contamination, or the introduction of micro-organisms into the access site, which could lead to an infectious process, bacteremia, or sepsis.
Product
BioPince Ultra Full Core Biopsy Instrument: REF: REF: 360-1080-01, 360-1080-02, 360-1080-03, 360-1580-01, 360-1580-02, 360-1580-03, 360-2080-01, 360-2080-02, 360-2080-03, 370-1080-01, 370-1080-02, 370-1080-03, 370-1580-01, 370-1580-02, 370-1580-03; Tru-Core II Biopsy Instrument, REF: 763114100x, 763116100X, 763116160X, 763118100X, 763118200X, 763120100X, 763120160X, 763418200X, 763418250X, 763114200X
Where it went
Worldwide - US Nationwide distribution including in the states of AZ, FL, OH, NY, MO, CO, IL, NC, CA, AK, PA, MN, ND, TX, SC, AR, WI, OK, TN, KS, UT, NJ, VA, MA, KY, IN, AL, DE, MT, NE, WV, NV, HI, OR, WA, NH, LA, CT, MI, GA, MD, ID, MS, IA, RI, NM, VT and the countries of ITA, TUN, ROU, LTU, BEL, GBR, CAN, TUR, POL, PAN, THA, PRT, TWN, AUT, NZL, AUS, BGR, MEX, CHE, NLD, PHL, PER, FRA, SGP, ZAF, CZE, EST, ESP, OMN, NPL, COL, PRI, ISR, JPN, KEN, BHR, ARE, CHN.
How much
95,430

L-CATH PICC S/L Peripherally Inserted Central Catheter, 26Ga (1.9F) 0.60mm x 30cm, REF: 384539

Z-2638-2024  ·  Class II  ·  reported 2024-08-21  ·  Ongoing

Why it was recalled
Peripherally Inserted Central Catheter (PICC) is labeled 26G/1.9Fr x 30cm but package contains incorrect 26G/1.9Fr x 50cm PICC, which could lead to procedural delay.
Product
L-CATH PICC S/L Peripherally Inserted Central Catheter, 26Ga (1.9F) 0.60mm x 30cm, REF: 384539
Where it went
US distribution to the states of: TX, WA, CA, IL
How much
155

Argon Medical Devices, REF 352506070E, Option Elite Retrievable Vena Cava Filter, Suitable of Jugular or Femor

Z-2239-2023  ·  Class II  ·  reported 2023-08-02  ·  Ongoing

Why it was recalled
There is the potential that the vena Cava Filter may not perform as intended which may result in pulmonary embolism.
Product
Argon Medical Devices, REF 352506070E, Option Elite Retrievable Vena Cava Filter, Suitable of Jugular or Femoral Delivery, STERILEO, RxOnly, CE 2797
Where it went
US Nationwide distribution in the state of TX.
How much
3.0

1. Arterial Line Kit 20ga x 6" 2. Arterial Line Kit /1 20ga x 6"

Z-0563-2022  ·  Class II  ·  reported 2022-02-16  ·  Ongoing

Why it was recalled
Guidewire included with two lots of Arterial Line Kits would not fit through the needle of the same kit. This could potentially lead to a slight delay in the procedure and minor blood loss.
Product
1. Arterial Line Kit 20ga x 6" 2. Arterial Line Kit /1 20ga x 6"
Where it went
US Nationwide distribution in the states of AL, CA, DE, IL, IN, LA, MT, NV, OH, TX, VA, TN.
How much
600 ea

SuperCore Biopsy Instrument: REF/Product Description: 701114090/SUPERCORE BIOPSY INSTR 14GA X 9CM; 701114

Z-0354-2022  ·  Class II  ·  reported 2021-12-15  ·  Ongoing

Why it was recalled
Due to component (housing and plunger) detaching from the semi-automatic biopsy instruments.
Product
SuperCore Biopsy Instrument: REF/Product Description: 701114090/SUPERCORE BIOPSY INSTR 14GA X 9CM; 701114150/SUPERCORE BIOPSY INSTR 14GA X 15CM; 701116090/SUPERCORE BIOPSY INSTR 16GA X 9CM; 701118060/SUPERCORE BIOPSY INSTR 18GA X 6CM; 701118090/SUPERCORE BIOPSY INSTR 18GA X 9CM; 701118150/SUPERCORE BIOPSY INSTR 18GA X 15CM; 701118200/SUPERCORE BIOPSY INSTR 18GA X 20CM; 701120090/SUPERCORE BIOPSY INSTR 20GA X 9CM; 701120150/SUPERCORE BIOPSY INSTR 20GA X 15CM; 701120200/SUPERCORE BIOPSY INSTR 20GA X 20CM; 701214090/SUPERCORE BIOPSY INSTR W/COAXIAL - 14G x 9CM; 701214150/SUPERCORE BIOPSY INSTR W/COAXIAL - 14G x 15CM; 701216090/SUPERCORE BIOPSY INSTR W/COAXIAL - 16G x 9CM; 701216150/SUPERCORE BIOPSY INSTR W/COAXIAL - 16G x 15CM; 701218090/SUPERCORE BIOPSY INSTR W/COAXIAL - 18G x 9CM; 701218150/SUPERCORE BIOPSY INSTR W/COAXIAL - 18G x 15CM; 701218200/SUPERCORE BIOPSY INSTR W/COAXIAL - 18G x 20CM; 701220090/SUPERCORE BIOPSY INSTR W/COAXIAL - 20G x 9CM; 701220150/SUPERCORE BIOPSY INSTR W/COAXIAL - 20G x 15CM; 701220200/SUPERCORE BIOPSY INSTR W/COAXIAL - 20G x 20CM; 701114090M/SUPERCORE BIOPSY INSTR 14GA X 9CM-MAMMOTOME; 701114150M/SUPERCORE BIOPSY INSTR 14GA X...
Where it went
Worldwide distribution - U.S. Nationwide distribution in the states of AK, AL, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, and WV. The countries of ALB, ARE, ARG, AUS, AUT, BEL, BGD, BRA, CAN, CHE, CHL, CHN, COL, CZE, DEU, ECU, ESP, EST, FRA, GBR, GRC, HKG, HUN, IND, IRN, ISR, ITA, JOR, JPN, MAR, MEX, MYS, NLD, NPL, NZL, PAK, PER, PHL, POL, PRT, SAU, SGP, SRB, THA, TUR, and TWN.
How much
79,910 devices

Scorpion Portal Vein Access Set - Product Usage: used to gain access to the hepatic vein and guide a sharp pun

Z-2501-2021  ·  Class II  ·  reported 2021-09-29  ·  Ongoing

Why it was recalled
As a result of design changes, sheaths have exhibited cracking/breaking at the tips.
Product
Scorpion Portal Vein Access Set - Product Usage: used to gain access to the hepatic vein and guide a sharp puncture tool (0.040 Stylet or 17ga Needle) through the parenchyma.
Where it went
U.S. Nationwide distribution in the state of WA. O.U.S.: N/A
How much
20 units

Argon Medical Devices TLAB, Transjugular Liver Biopsy System, REF: TL - 18C, 1/18 GA Flexcore Biopsy Needle,

Z-1076-2021  ·  Class II  ·  reported 2021-02-17  ·  Terminated

Why it was recalled
Due to a manufacturing error their is a potential that the sheath within the biopsy kit can separate during use if the tip is flexed or experiences bending stresses. This could result in need for surgical intervention to preclude or reverse permanent damage to a body structure or permanent impairment of a body function.
Product
Argon Medical Devices TLAB, Transjugular Liver Biopsy System, REF: TL - 18C, 1/18 GA Flexcore Biopsy Needle, 1/7F Introducer Sheath W/Curved Metal Stiffener, 4/Tissue Removal Swabs, 1/5f straight catheter, 1/5F Curved Catheter Sterile EO,
Where it went
US: AL AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA,, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TN, TX,, UT, VA, WA, WI, WV, WY OUS: Switzerland, Saudi Arabia, Hong Kong, Thailand, Portugal, Taiwan, Austria, Brazil, Netherlands, Italy, Germany, Colombia, Singapore, Canada, View Nam, Slovakia, Spain
How much
1759 units

Argon Medical Devices TLAB, Transjugular Liver Biopsy System, REF: TL - 19S, 1/19 GA Flexcore Biopsy Needle,

Z-1075-2021  ·  Class II  ·  reported 2021-02-17  ·  Terminated

Why it was recalled
Due to a manufacturing error their is a potential that the sheath within the biopsy kit can separate during use if the tip is flexed or experiences bending stresses. This could result in need for surgical intervention to preclude or reverse permanent damage to a body structure or permanent impairment of a body function.
Product
Argon Medical Devices TLAB, Transjugular Liver Biopsy System, REF: TL - 19S, 1/19 GA Flexcore Biopsy Needle, 1/7F Introducer Sheath W/Curved Metal Stiffener, 4/Tissue Removal Swabs, 1/5f straight catheter, Sterile EO,
Where it went
US: AL AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA,, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TN, TX,, UT, VA, WA, WI, WV, WY OUS: Switzerland, Saudi Arabia, Hong Kong, Thailand, Portugal, Taiwan, Austria, Brazil, Netherlands, Italy, Germany, Colombia, Singapore, Canada, View Nam, Slovakia, Spain
How much
247 units

Argon Medical Devices TLAB, Transjugular Liver Biopsy System, REF: TL - 19, 1/19 GA Flexcore Biopsy Needle,

Z-1073-2021  ·  Class II  ·  reported 2021-02-17  ·  Terminated

Why it was recalled
Due to a manufacturing error their is a potential that the sheath within the biopsy kit can separate during use if the tip is flexed or experiences bending stresses. This could result in need for surgical intervention to preclude or reverse permanent damage to a body structure or permanent impairment of a body function.
Product
Argon Medical Devices TLAB, Transjugular Liver Biopsy System, REF: TL - 19, 1/19 GA Flexcore Biopsy Needle, 7F Introducer Sheath W/Curved Metal Stiffener, Tissue Removal Swabs, 5F Straight Catheter, 5F Curved Catheter, Sterile EO,
Where it went
US: AL AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA,, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TN, TX,, UT, VA, WA, WI, WV, WY OUS: Switzerland, Saudi Arabia, Hong Kong, Thailand, Portugal, Taiwan, Austria, Brazil, Netherlands, Italy, Germany, Colombia, Singapore, Canada, View Nam, Slovakia, Spain
How much
863 unit

Argon Medical Devices TLAB, Transjugular Liver Biopsy System, REF: TL - 18N, 1/18 GA Flexcore Biopsy Needle,

Z-1077-2021  ·  Class II  ·  reported 2021-02-17  ·  Terminated

Why it was recalled
Due to a manufacturing error their is a potential that the sheath within the biopsy kit can separate during use if the tip is flexed or experiences bending stresses. This could result in need for surgical intervention to preclude or reverse permanent damage to a body structure or permanent impairment of a body function.
Product
Argon Medical Devices TLAB, Transjugular Liver Biopsy System, REF: TL - 18N, 1/18 GA Flexcore Biopsy Needle, 1/7F Introducer Sheath W/Curved Metal Stiffener, 4/Tissue Removal Swabs, Sterile EO,
Where it went
US: AL AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA,, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TN, TX,, UT, VA, WA, WI, WV, WY OUS: Switzerland, Saudi Arabia, Hong Kong, Thailand, Portugal, Taiwan, Austria, Brazil, Netherlands, Italy, Germany, Colombia, Singapore, Canada, View Nam, Slovakia, Spain
How much
583 units

Argon Medical Devices TLAB, Transjugular Liver Biopsy System, REF: TL - 19N, 1/19 GA Flexcore Biopsy Needle,

Z-1074-2021  ·  Class II  ·  reported 2021-02-17  ·  Terminated

Why it was recalled
Due to a manufacturing error their is a potential that the sheath within the biopsy kit can separate during use if the tip is flexed or experiences bending stresses. This could result in need for surgical intervention to preclude or reverse permanent damage to a body structure or permanent impairment of a body function.
Product
Argon Medical Devices TLAB, Transjugular Liver Biopsy System, REF: TL - 19N, 1/19 GA Flexcore Biopsy Needle, 1.7F Introducer Sheath W/Curved Metal Stiffener, 4/Tissue Removal Swabs, Sterile EO,
Where it went
US: AL AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA,, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TN, TX,, UT, VA, WA, WI, WV, WY OUS: Switzerland, Saudi Arabia, Hong Kong, Thailand, Portugal, Taiwan, Austria, Brazil, Netherlands, Italy, Germany, Colombia, Singapore, Canada, View Nam, Slovakia, Spain
How much
187 units

Argon Medical Devices TLAB, Transjugular Liver Biopsy System, REF: TL - 18S, 1/18 GA Flexcore Biopsy Needle,

Z-1072-2021  ·  Class II  ·  reported 2021-02-17  ·  Terminated

Why it was recalled
Due to a manufacturing error their is a potential that the sheath within the biopsy kit can separate during use if the tip is flexed or experiences bending stresses. This could result in need for surgical intervention to preclude or reverse permanent damage to a body structure or permanent impairment of a body function.
Product
Argon Medical Devices TLAB, Transjugular Liver Biopsy System, REF: TL - 18S, 1/18 GA Flexcore Biopsy Needle, 1/7F Introducer Sheath W/Curved Metal Stiffener, 4/Tissue Removal Swabs, 1/6f Straight Catheter, Sterile EO,
Where it went
US: AL AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA,, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TN, TX,, UT, VA, WA, WI, WV, WY OUS: Switzerland, Saudi Arabia, Hong Kong, Thailand, Portugal, Taiwan, Austria, Brazil, Netherlands, Italy, Germany, Colombia, Singapore, Canada, View Nam, Slovakia, Spain
How much
483 units

ClosureFast Micro Introducer Sheath Set, REF MIS-7F11; 7Fr/Ch (2.3mm)x11 cm. STERILE|EO; Rx ONLY - Product Usa

Z-2806-2020  ·  Class II  ·  reported 2020-08-26  ·  Terminated

Why it was recalled
Sterile introducer sheath set manufactured under one lot with different expiration dates.
Product
ClosureFast Micro Introducer Sheath Set, REF MIS-7F11; 7Fr/Ch (2.3mm)x11 cm. STERILE|EO; Rx ONLY - Product Usage: Introducer Products are stand-alone components, standard kits, or convenience kits built to user specifications. These products facilitate access to a vein or artery, allowing various catheters to be passed through for local or central diagnostic or therapeutic interventions. Indicated for use in arterial and venous procedures requiring percutaneous introduction of intravascular devices.
Where it went
US Nationwide distribution including in the state of MS.
How much
1000 units

BioPince Full Core Biopsy Instrument 16ga x 15cm.,Use with Optional CO-Axial Needle MCXS1615BP. QTY.: 5/CS -

Z-2427-2020  ·  Class II  ·  reported 2020-07-01  ·  Terminated

Why it was recalled
Incorrect product labeling.
Product
BioPince Full Core Biopsy Instrument 16ga x 15cm.,Use with Optional CO-Axial Needle MCXS1615BP. QTY.: 5/CS - Product Usage: The disposable, automatic BioPince Full Core Biopsy Instrument is used to obtain multiple core samples from soft tissue such as liver, kidney, abdominal masses, etc.
Where it went
Worldwide distribution - US Nationwide including in the states of TX, IL, FL, KS, MD, CA, WA, NY, WI, PA and the countries of United Arab Emirates (Dubai), Spain, Canada.
How much
165 units in U.S. and 330 units Internationally

Skater Introducer Set 6F x 18cm. REF/UDI: 651506300/(01)00886333005000; 655506300/(01)00886333005079 - Product

Z-2109-2020  ·  Class II  ·  reported 2020-06-03  ·  Terminated

Why it was recalled
Tungsten coils of the guidewire included in the introducer kit were detaching.
Product
Skater Introducer Set 6F x 18cm. REF/UDI: 651506300/(01)00886333005000; 655506300/(01)00886333005079 - Product Usage: The Skater Introducer System with radiopaque marker is intended for fine needle puncture and tissue dilation with subsequent positioning of a guidewire for catheter placement in non-vascular procedures.
Where it went
International distribution to the countries of China, Taiwan.
How much
15,795 units

PAC Tray/5 Catheter Introducer Tray 8F, Model 497424, Rx Only, STERILE EO, UDI: 00886333214440 - Product Usage

Z-1938-2020  ·  Class II  ·  reported 2020-05-20  ·  Terminated

Why it was recalled
Arterial Line Kits include thicker curved reverse-cutting needle with a 0-silk suture, instead of the usual straight-cutting needle with a 3-0 silk suture. The thicker suture may be more difficult to tie. The curved needle requires the use of a needle driver, there is a risk of needle stick to the surgeon tries to do it with fingers instead.
Product
PAC Tray/5 Catheter Introducer Tray 8F, Model 497424, Rx Only, STERILE EO, UDI: 00886333214440 - Product Usage: The introducers are indicated for use in arterial and venous procedures requiring percutaneous introduction of intravascular devices. Curved tip Introducer allows for contralateral access to the iliac artery and selected vasculature. Case Label: PAC Tray/5 Catheter Introducer Tray 8F, QTY 10/Bx, Rx Only, STERILE EO
Where it went
US Nationwide distribution including the states of CA, IL, ND, SC, TX, VA. OUS: None
How much
80

Arterial Line Kit 20ga X 3", Model 498106, Rx Only, Sterile EO, UDI: 00886333215577 - Product Usage: Arterial

Z-1937-2020  ·  Class II  ·  reported 2020-05-20  ·  Terminated

Why it was recalled
Arterial Line Kits include thicker curved reverse-cutting needle with a 0-silk suture, instead of the usual straight-cutting needle with a 3-0 silk suture. The thicker suture may be more difficult to tie. The curved needle requires the use of a needle driver, there is a risk of needle stick to the surgeon tries to do it with fingers instead.
Product
Arterial Line Kit 20ga X 3", Model 498106, Rx Only, Sterile EO, UDI: 00886333215577 - Product Usage: Arterial Catheters are indicated for use when in instances where continuous invasive blood pressure readings and/or frequent blood gas measurements are required. Case Label: Arterial Line Kit/1 20ga x 3", Qty: 10/CS, CE 2797
Where it went
US Nationwide distribution including the states of CA, IL, ND, SC, TX, VA. OUS: None
How much
20

Arterial Line Kit 20ga X 3", Model 498411B, Rx Only, Sterile EO, UDI: 00886333215652 - Product Usage: Arteria

Z-1936-2020  ·  Class II  ·  reported 2020-05-20  ·  Terminated

Why it was recalled
Arterial Line Kits include thicker curved reverse-cutting needle with a 0-silk suture, instead of the usual straight-cutting needle with a 3-0 silk suture. The thicker suture may be more difficult to tie. The curved needle requires the use of a needle driver, there is a risk of needle stick to the surgeon tries to do it with fingers instead.
Product
Arterial Line Kit 20ga X 3", Model 498411B, Rx Only, Sterile EO, UDI: 00886333215652 - Product Usage: Arterial Catheters are indicated for use when in instances where continuous invasive blood pressure readings and/or frequent blood gas measurements are required. Case Label: Arterial Line Kit 20ga x 3", Ref 498411B, Qty: 10/CS
Where it went
US Nationwide distribution including the states of CA, IL, ND, SC, TX, VA. OUS: None
How much
150
Class I

0.035"(.89mm)x150 cm guidewire. Model 114135150. Soft Tip 3.5cm. REF/UDI: 114135150/(01)20886333008074. RxOnly

Z-1792-2020  ·  Class I  ·  reported 2020-05-13  ·  Terminated

Why it was recalled
The stiffness of the soft end of the guidewire caused tissue perforation.
Product
0.035"(.89mm)x150 cm guidewire. Model 114135150. Soft Tip 3.5cm. REF/UDI: 114135150/(01)20886333008074. RxOnly.STERILE EO
Where it went
No US consignees. OUS - Sweden, Norway, Denmark, Greece, Slovenia, Turkey, Hong Kong, England
How much
1190 units
Class I

0.035"(.89mm)x80 cm guidewire. Model 114135080. Soft Tip 3.5cm. REF/UDI: 114135080/(01)208863333008067. RxOnly

Z-1791-2020  ·  Class I  ·  reported 2020-05-13  ·  Terminated

Why it was recalled
The stiffness of the soft end of the guidewire caused tissue perforation.
Product
0.035"(.89mm)x80 cm guidewire. Model 114135080. Soft Tip 3.5cm. REF/UDI: 114135080/(01)208863333008067. RxOnly STERILE EO
Where it went
No US consignees. OUS - Sweden, Norway, Denmark, Greece, Slovenia, Turkey, Hong Kong, England
How much
1690 units

(1) Argon Medical Devices Vascuclamp, Radiopaque Vascular Clamp, Small Straight, REF 046-1-200, 2 clamps/packa

Z-0293-2020  ·  Class II  ·  reported 2019-11-13  ·  Terminated

Why it was recalled
The seal formed on the edge of the pouch may be incomplete, which could result in a loss of sterility to the product.
Product
(1) Argon Medical Devices Vascuclamp, Radiopaque Vascular Clamp, Small Straight, REF 046-1-200, 2 clamps/package, 5 packages/box, 10 boxes/case, Sterile, Rx; (2) Argon Medical Devices Vascuclamp, Radiopaque Vascular Clamp, Medium Straight, REF 046-1-300, 2 clamps per package, 5 packages/ shelf box, 10 shelf boxes/case (for a total of 50/2-clamp packages/case), Sterile, Rx; (3) Argon Medical Devices Vascuclamp, Radiopaque Vascular Clamp, Medium Angled, REF 046-1-301, 2 clamps per package, 6 packages/box, 50 boxes/case, Sterile, Rx; and (4) Argon Medical Devices Vascuclamp, Radiopaque Vascular Clamp, Large Angled, REF 046-1-401, 2 clamps per package, 5 packages/box, 50 boxes/case. The firm name on the label is Argon Medical Devices, Inc., Athens, TX.
Where it went
Distribution was nationwide. There was military/government distribution. Foreign distribution was made to Canada, Australia, the Netherlands, and Spain.
How much
Total number in recall for all products in dist: 49,130 units

Argon Medical Devices TRI-JECT, 3 Way IV Administration Organizer without Stopcock, REF 041230001A, 25/box, St

Z-0291-2020  ·  Class II  ·  reported 2019-11-13  ·  Terminated

Why it was recalled
The seal formed on the edge of the pouch may be incomplete, which could result in a loss of sterility to the product.
Product
Argon Medical Devices TRI-JECT, 3 Way IV Administration Organizer without Stopcock, REF 041230001A, 25/box, Sterile, Rx. The firm name on the label is Argon Medical Devices, Inc., Athens, TX.
Where it went
Distribution was nationwide. There was military/government distribution. Foreign distribution was made to Canada, Australia, the Netherlands, and Spain.
How much
Total number in recall for all products in dist: 49,130 units

(1) Argon Medical Devices Double Female LL Adapter, Special Adapter to Convert a Male Fitting to Female LL, RE

Z-0288-2020  ·  Class II  ·  reported 2019-11-13  ·  Terminated

Why it was recalled
The seal formed on the edge of the pouch may be incomplete, which could result in a loss of sterility to the product.
Product
(1) Argon Medical Devices Double Female LL Adapter, Special Adapter to Convert a Male Fitting to Female LL, REF 040183000A, 25/box, Sterile, Rx; and (2) Argon Medical Devices Double Male LL Adapter, Special Adapter to Convert a Female Fitting to Male LL, REF 040184000A, 25/box, Sterile, Rx. The firm name on the labels is Argon Medical Devices, Inc., Athens, TX.
Where it went
Distribution was nationwide. There was military/government distribution. Foreign distribution was made to Canada, Australia, the Netherlands, and Spain.
How much
Total number in recall for all products in dist: 49,130 units

Showing the 30 most recent of 48. The rest are searchable at the FDA enforcement report database.

Read these counts with three things in mind

A firm with more recalls usually makes more products. A contract packer runs hundreds of lines. It will out-recall a company with one product whatever either does. This is not a safety ranking.

Most recalls here are voluntary. A firm found its own fault and pulled the stock. That is the system working. It lands in this database looking the same as a recall the FDA had to push for.

Terminated means the FDA considers the recall finished. It is not a verdict. It does not mean anyone was harmed.

Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.