Aomori Olympus Co., Ltd.

Aomori Olympus Co., Ltd.: 113 FDA recall reports

Every medical device recall the FDA has published for Aomori Olympus Co., Ltd.. The fault, and how the agency graded it.

113Reports
0Class I
2Ongoing
2023-09-13Most recent

What the classes mean

The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.

ClassWhat the FDA is saying
Class IThere is a reasonable probability that using this will cause serious harm or death.
Class IIIt may cause temporary or reversible harm. Serious harm is unlikely.
Class IIIIt breaks a rule but is unlikely to hurt anyone. Often a labelling fault.

Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.

By class

ClassReports
Class II113

By year reported

YearReports
2021111
20232
Checking whether it affects you

The lot or code number is what matters. Not the product name. A recall covers specific lots. Yours may not be one of them.

Check the code on your package against the report. The FDA keeps the current list at fda.gov/safety/recalls.

The reports

The text below is quoted from the FDA record word for word.

Olympus Distal Attachment (in Olympus Disposable EMR Kit)- Designed to be attached to the distal end of the en

Z-2522-2023  ·  Class II  ·  reported 2023-09-13  ·  Ongoing

Why it was recalled
The Distal Attachment contain di-2-ethylhexyl phthalate (DEHP), which exceeds the permissible exposure level calculated according to ISO 10993-17: 2002
Product
Olympus Distal Attachment (in Olympus Disposable EMR Kit)- Designed to be attached to the distal end of the endoscope. Disposable EMR Kit: K-001, K-002, K-003, K-004, K-011, K-012
Where it went
US Nationwide distribution.
How much
1,048 units

Reusable distal attachments Models: MH-462, MH-463, MH-464, MH-465, MH-466, MH-483, MH-587, MH-588(US), MH-5

Z-2523-2023  ·  Class II  ·  reported 2023-09-13  ·  Ongoing

Why it was recalled
The Distal Attachment contain di-2-ethylhexyl phthalate (DEHP), which exceeds the permissible exposure level calculated according to ISO 10993-17: 2002
Product
Reusable distal attachments Models: MH-462, MH-463, MH-464, MH-465, MH-466, MH-483, MH-587, MH-588(US), MH-589, MH-590, MH-591, MH-592, MH-593, MH-594, MH-595, MH-596, MH-597, MH-598, MAJ-289, MAJ-290, MAJ-291, MAJ-292, MAJ-293, MAJ-294, MAJ-295, MAJ-296, MAJ-297, MAJ-663, MAJ-664, MAJ-665, MAJ-666
Where it went
US Nationwide distribution.
How much
1,252 units

INJECTOR FORCEMAX 25G 2.5MM, model no. NM-400U-0425 - Product Usage: Used with endoscopes to perform endoscopi

Z-1018-2021  ·  Class II  ·  reported 2021-02-17  ·  Terminated

Why it was recalled
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Product
INJECTOR FORCEMAX 25G 2.5MM, model no. NM-400U-0425 - Product Usage: Used with endoscopes to perform endoscopic injection for the treatment of esophageal and gastric varices and for submucosal injection in the digestive tract.
Where it went
US Nationwide distribution.
How much
7,400,000 (globally); 260,395 (US)

KD-611L ELECTROSURGICAL KNIFE, model no. KD-611L - Product Usage: Used with endoscopes and electrosurgical uni

Z-0974-2021  ·  Class II  ·  reported 2021-02-17  ·  Terminated

Why it was recalled
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Product
KD-611L ELECTROSURGICAL KNIFE, model no. KD-611L - Product Usage: Used with endoscopes and electrosurgical units to cut tissue using high-frequency current within the upper digestive tract.
Where it went
US Nationwide distribution.
How much
7,400,000 (global); 260,395 (US)

K-202 GUIDE SHEATH KIT 2.0MM, model no. K-202 - Product Usage: Used with endoscopes to collect cells or tissue

Z-1051-2021  ·  Class II  ·  reported 2021-02-17  ·  Terminated

Why it was recalled
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Product
K-202 GUIDE SHEATH KIT 2.0MM, model no. K-202 - Product Usage: Used with endoscopes to collect cells or tissue specimens in the respiratory organs.
Where it went
US Nationwide distribution.
How much
7,400,000 (globally); 260,395 (US)

NEEDLEMASTER 3MMx 23G LOWER, model no. NM-610U-0323 - Product Usage: Used with endoscopes to perform endoscopi

Z-1040-2021  ·  Class II  ·  reported 2021-02-17  ·  Terminated

Why it was recalled
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Product
NEEDLEMASTER 3MMx 23G LOWER, model no. NM-610U-0323 - Product Usage: Used with endoscopes to perform endoscopic vascular or submucosal injection in the GI tract.
Where it went
US Nationwide distribution.
How much
7,400,000 (globally); 260,395 (US)

SD-230U-20 ELECTROSURGICAL SNARE, model no. SD-230U-20 - Product Usage: Used with endoscopes to electrosurgica

Z-0987-2021  ·  Class II  ·  reported 2021-02-17  ·  Terminated

Why it was recalled
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Product
SD-230U-20 ELECTROSURGICAL SNARE, model no. SD-230U-20 - Product Usage: Used with endoscopes to electrosurgically resect tissue within the digestive tract.
Where it went
US Nationwide distribution.
How much
7,400,000 (globally); 260,395 (US)

B7-2LA BALLOON CATHETER 13MM DIA 3.2, model no. B7-2LA - Product Usage: Used with endoscopes for injection of

Z-0964-2021  ·  Class II  ·  reported 2021-02-17  ·  Terminated

Why it was recalled
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Product
B7-2LA BALLOON CATHETER 13MM DIA 3.2, model no. B7-2LA - Product Usage: Used with endoscopes for injection of contrast medium or other medical fluid into the biliary or pancreatic tract.
Where it went
US Nationwide distribution.
How much
7,400,000 (global); 260,395 (US)

INJECTOR FORCEMAX UP 2.8MMCH 23G 6MM, model no. NM-400L-0623 - Product Usage: Used with endoscopes to perform

Z-1012-2021  ·  Class II  ·  reported 2021-02-17  ·  Terminated

Why it was recalled
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Product
INJECTOR FORCEMAX UP 2.8MMCH 23G 6MM, model no. NM-400L-0623 - Product Usage: Used with endoscopes to perform endoscopic injection for the treatment of esophageal and gastric varices and for submucosal injection in the digestive tract.
Where it went
US Nationwide distribution.
How much
7,400,000 (globally); 260,395 (US)

INJECTOR FORCEMAX UP 2.8MMCH 25G 6MM, model no. NM-400L-0625 - Product Usage: Used with endoscopes to perform

Z-1000-2021  ·  Class II  ·  reported 2021-02-17  ·  Terminated

Why it was recalled
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Product
INJECTOR FORCEMAX UP 2.8MMCH 25G 6MM, model no. NM-400L-0625 - Product Usage: Used with endoscopes to perform endoscopic injection for the treatment of esophageal and gastric varices and for submucosal injection in the digestive tract.
Where it went
US Nationwide distribution.
How much
7,400,000 (globally); 260,395 (US)

INJECTOR FORCEMAX UP 2.8MMCH 21G 6MM, model no. NM-400L-0621 - Product Usage: Used with endoscopes to perform

Z-1010-2021  ·  Class II  ·  reported 2021-02-17  ·  Terminated

Why it was recalled
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Product
INJECTOR FORCEMAX UP 2.8MMCH 21G 6MM, model no. NM-400L-0621 - Product Usage: Used with endoscopes to perform endoscopic injection for the treatment of esophageal and gastric varices and for submucosal injection in the digestive tract.
Where it went
US Nationwide distribution.
How much
7,400,000 (globally); 260,395 (US)

PR-V614M 2-LUMEN CANNULA, model no. PR-V614M - Product Usage: Used with endoscopes to inject contrast medium i

Z-0969-2021  ·  Class II  ·  reported 2021-02-17  ·  Terminated

Why it was recalled
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Product
PR-V614M 2-LUMEN CANNULA, model no. PR-V614M - Product Usage: Used with endoscopes to inject contrast medium in the biliary or pancreatic duct.
Where it went
US Nationwide distribution.
How much
7,400,000 (global); 260,395 (US)

KD-620UR ELECTROSURGICAL SNARE, model no. KD-620UR - Product Usage: Used with endoscopes and electrosurgical u

Z-0978-2021  ·  Class II  ·  reported 2021-02-17  ·  Terminated

Why it was recalled
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Product
KD-620UR ELECTROSURGICAL SNARE, model no. KD-620UR - Product Usage: Used with endoscopes and electrosurgical units to cut tissue using high-frequency current within the upper digestive tract.
Where it went
US Nationwide distribution.
How much
7,400,000 (global); 260,395 (US)

INJECTOR FORCEMAX LOWER 25G 6MM, model no. NM-400U-0625 - Product Usage: Used with endoscopes to perform endos

Z-1022-2021  ·  Class II  ·  reported 2021-02-17  ·  Terminated

Why it was recalled
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Product
INJECTOR FORCEMAX LOWER 25G 6MM, model no. NM-400U-0625 - Product Usage: Used with endoscopes to perform endoscopic injection for the treatment of esophageal and gastric varices and for submucosal injection in the digestive tract.
Where it went
US Nationwide distribution.
How much
7,400,000 (globally); 260,395 (US)

INJECTOR FORCEMAX LOWER 23G 3MM, model no. NM-400U-0323 - Product Usage: Used with endoscopes to perform endos

Z-1016-2021  ·  Class II  ·  reported 2021-02-17  ·  Terminated

Why it was recalled
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Product
INJECTOR FORCEMAX LOWER 23G 3MM, model no. NM-400U-0323 - Product Usage: Used with endoscopes to perform endoscopic injection for the treatment of esophageal and gastric varices and for submucosal injection in the digestive tract.
Where it went
US Nationwide distribution.
How much
7,400,000 (globally); 260,395 (US)

B-V233P-A 3-LUMEN EXTRACTION BALLOON V, model no. B-V233P-A - Product Usage: Used with endoscopes to endoscopi

Z-0966-2021  ·  Class II  ·  reported 2021-02-17  ·  Terminated

Why it was recalled
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Product
B-V233P-A 3-LUMEN EXTRACTION BALLOON V, model no. B-V233P-A - Product Usage: Used with endoscopes to endoscopically remove stones such as calculi, pancreatic and common bile duct stones.
Where it went
US Nationwide distribution.
How much
7,400,000 (global); 260,395 (US)

INJECTOR FORCEMAX UP 2.8MMCH 25G 4MM, model no. NM-400L-0425 - Product Usage: Used with endoscopes to perform

Z-0995-2021  ·  Class II  ·  reported 2021-02-17  ·  Terminated

Why it was recalled
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Product
INJECTOR FORCEMAX UP 2.8MMCH 25G 4MM, model no. NM-400L-0425 - Product Usage: Used with endoscopes to perform endoscopic injection for the treatment of esophageal and gastric varices and for submucosal injection in the digestive tract.
Where it went
US Nationwide distribution.
How much
7,400,000 (globally); 260,395 (US)

INJECTOR FORCEMAX SIF 270CM 23G 4MM, model no. NM-401L-0423 - Product Usage: Used with endoscopes to perform e

Z-1024-2021  ·  Class II  ·  reported 2021-02-17  ·  Terminated

Why it was recalled
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Product
INJECTOR FORCEMAX SIF 270CM 23G 4MM, model no. NM-401L-0423 - Product Usage: Used with endoscopes to perform endoscopic injection for the treatment of esophageal and gastric varices and for submucosal injection in the digestive tract.
Where it went
US Nationwide distribution.
How much
7,400,000 (globally); 260,395 (US)

INJECTOR FORCEMAX UP 2.8MMCH 23G 6MM, model no. NM-400L-0623 - Product Usage: Used with endoscopes to perform

Z-0999-2021  ·  Class II  ·  reported 2021-02-17  ·  Terminated

Why it was recalled
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Product
INJECTOR FORCEMAX UP 2.8MMCH 23G 6MM, model no. NM-400L-0623 - Product Usage: Used with endoscopes to perform endoscopic injection for the treatment of esophageal and gastric varices and for submucosal injection in the digestive tract.
Where it went
US Nationwide distribution.
How much
7,400,000 (globally); 260,395 (US)

NA-401D-1321 ASPIRATION NEEDLE, model no. NA-401D-1321 - Product Usage: Used with endoscopes to collect cells

Z-1054-2021  ·  Class II  ·  reported 2021-02-17  ·  Terminated

Why it was recalled
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Product
NA-401D-1321 ASPIRATION NEEDLE, model no. NA-401D-1321 - Product Usage: Used with endoscopes to collect cells or tissue specimens in the respiratory organs.
Where it went
US Nationwide distribution.
How much
7,400,000 (globally); 260,395 (US)

KD-655U ELECTROSURGICAL SNARE, model no. KD-655U - Product Usage: Used with endoscopes and electrosurgical uni

Z-0983-2021  ·  Class II  ·  reported 2021-02-17  ·  Terminated

Why it was recalled
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Product
KD-655U ELECTROSURGICAL SNARE, model no. KD-655U - Product Usage: Used with endoscopes and electrosurgical units to cut tissue using high-frequency current within the upper digestive tract.
Where it went
US Nationwide distribution.
How much
7,400,000 (global); 260,395 (US)

NA-201SX-4022-B ASPIRATION NEEDLE ViziShot 22G, model no. NA-201SX-4022-B - Product Usage: Used with ultrasoun

Z-1061-2021  ·  Class II  ·  reported 2021-02-17  ·  Terminated

Why it was recalled
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Product
NA-201SX-4022-B ASPIRATION NEEDLE ViziShot 22G, model no. NA-201SX-4022-B - Product Usage: Used with ultrasound endoscopes for ultrasound guided fine needle aspiration of submucosal and extramural lesions of the tracheobronchial tree and esophagus.
Where it went
US Nationwide distribution.
How much
7,400,000 (globally); 260,395 (US)

INJECTOR FORCEMAX UP 2.8MMCH 21G 6MM, model no. NM-400L-0621 - Product Usage: Used with endoscopes to perform

Z-0998-2021  ·  Class II  ·  reported 2021-02-17  ·  Terminated

Why it was recalled
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Product
INJECTOR FORCEMAX UP 2.8MMCH 21G 6MM, model no. NM-400L-0621 - Product Usage: Used with endoscopes to perform endoscopic injection for the treatment of esophageal and gastric varices and for submucosal injection in the digestive tract.
Where it went
US Nationwide distribution.
How much
7,400,000 (globally); 260,395 (US)

NEEDLEMASTER 3MMx 23G LOWER, model no. NM-610L-0523 - Product Usage: Used with endoscopes to perform endoscopi

Z-1035-2021  ·  Class II  ·  reported 2021-02-17  ·  Terminated

Why it was recalled
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Product
NEEDLEMASTER 3MMx 23G LOWER, model no. NM-610L-0523 - Product Usage: Used with endoscopes to perform endoscopic vascular or submucosal injection in the GI tract.
Where it went
US Nationwide distribution.
How much
7,400,000 (globally); 260,395 (US)

INJECTOR FORCEMAX SIF 270CM 23G 4MM, model no. NM-400Y-0423 - Product Usage: Used with endoscopes to perform e

Z-1008-2021  ·  Class II  ·  reported 2021-02-17  ·  Terminated

Why it was recalled
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Product
INJECTOR FORCEMAX SIF 270CM 23G 4MM, model no. NM-400Y-0423 - Product Usage: Used with endoscopes to perform endoscopic injection for the treatment of esophageal and gastric varices and for submucosal injection in the digestive tract.
Where it went
US Nationwide distribution.
How much
7,400,000 (globally); 260,395 (US)

NEEDLEMASTER 6MMx 23G LOWER, model no. NM-610U-0623 - Product Usage: Used with endoscopes to perform endoscopi

Z-1047-2021  ·  Class II  ·  reported 2021-02-17  ·  Terminated

Why it was recalled
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Product
NEEDLEMASTER 6MMx 23G LOWER, model no. NM-610U-0623 - Product Usage: Used with endoscopes to perform endoscopic vascular or submucosal injection in the GI tract.
Where it went
US Nationwide distribution.
How much
7,400,000 (globally); 260,395 (US)

NEEDLEMASTER 5MMx 21G UPPER, model no. NM-610L-0521 - Product Usage: Used with endoscopes to perform endoscopi

Z-1034-2021  ·  Class II  ·  reported 2021-02-17  ·  Terminated

Why it was recalled
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Product
NEEDLEMASTER 5MMx 21G UPPER, model no. NM-610L-0521 - Product Usage: Used with endoscopes to perform endoscopic vascular or submucosal injection in the GI tract.
Where it went
US Nationwide distribution.
How much
7,400,000 (globally); 260,395 (US)

K-203 GUIDE SHEATH Kit 2.6MM, model no. K-203 - Product Usage: Used with endoscopes to collect cells or tissue

Z-0972-2021  ·  Class II  ·  reported 2021-02-17  ·  Terminated

Why it was recalled
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Product
K-203 GUIDE SHEATH Kit 2.6MM, model no. K-203 - Product Usage: Used with endoscopes to collect cells or tissue specimens in the respiratory organs.
Where it went
US Nationwide distribution.
How much
7,400,000 (global); 260,395 (US)

INJECTOR FORCEMAX LOWER 23G 6MM, model no. NM-400U-0623 - Product Usage: Used with endoscopes to perform endos

Z-1006-2021  ·  Class II  ·  reported 2021-02-17  ·  Terminated

Why it was recalled
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Product
INJECTOR FORCEMAX LOWER 23G 6MM, model no. NM-400U-0623 - Product Usage: Used with endoscopes to perform endoscopic injection for the treatment of esophageal and gastric varices and for submucosal injection in the digestive tract.
Where it went
US Nationwide distribution.
How much
7,400,000 (globally); 260,395 (US)

INJECTOR FORCEMAX LOWER 23G 3MM, model no. NM-400U-0323 - Product Usage: Used with endoscopes to perform endos

Z-1001-2021  ·  Class II  ·  reported 2021-02-17  ·  Terminated

Why it was recalled
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Product
INJECTOR FORCEMAX LOWER 23G 3MM, model no. NM-400U-0323 - Product Usage: Used with endoscopes to perform endoscopic injection for the treatment of esophageal and gastric varices and for submucosal injection in the digestive tract.
Where it went
US Nationwide distribution.
How much
7,400,000 (globally); 260,395 (US)

Showing the 30 most recent of 113. The rest are searchable at the FDA enforcement report database.

Read these counts with three things in mind

A firm with more recalls usually makes more products. A contract packer runs hundreds of lines. It will out-recall a company with one product whatever either does. This is not a safety ranking.

Most recalls here are voluntary. A firm found its own fault and pulled the stock. That is the system working. It lands in this database looking the same as a recall the FDA had to push for.

Terminated means the FDA considers the recall finished. It is not a verdict. It does not mean anyone was harmed.

Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.