The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.

ClassWhat the FDA is saying
Class IThere is a reasonable probability that using this will cause serious harm or death.
Class IIIt may cause temporary or reversible harm. Serious harm is unlikely.
Class IIIIt breaks a rule but is unlikely to hurt anyone. Often a labelling fault.

Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.

By class

ClassReports
Class III63
Class II46

By year reported

YearReports
20143
201579
20171
20182
20191
202023
Checking whether it affects you

The lot or code number is what matters. Not the product name. A recall covers specific lots. Yours may not be one of them.

Check the code on your package against the report. The FDA keeps the current list at fda.gov/safety/recalls.

The reports

The text below is quoted from the FDA record word for word.

First Check 7 Panel Drug Cup: (AMP/COC/MDMA/MET/OPI/PCP/THC); AMT Item Number 990100B

Z-3050-2020  ·  Class II  ·  reported 2020-10-07  ·  Terminated

Why it was recalled
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Product
First Check 7 Panel Drug Cup: (AMP/COC/MDMA/MET/OPI/PCP/THC); AMT Item Number 990100B
Where it went
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
How much
99,900 units

Noble 1 Step+ Cup (AMT Item Number/ AMT Description): NBCA-13A-B /"Noble 1 Step + Cup" THC50/COC150; OPI30

Z-3039-2020  ·  Class II  ·  reported 2020-10-07  ·  Terminated

Why it was recalled
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Product
Noble 1 Step+ Cup (AMT Item Number/ AMT Description): NBCA-13A-B /"Noble 1 Step + Cup" THC50/COC150; OPI300/MET500; AMP500/BZO300; BAR300/MTD300; BUPG10/TCA1000; MDMA500/OXY100; w/ Cr, Ni, pH, Bl, S.G. Finished Devices", NBCA-14A-PP/"Noble 1 Step + Cup" (THC50/COC150)+(OPI300/MET500)+ (AMP500/BZO300)+(BAR300/MTD300)+ (BOPG10/TCA1000)+(MDMA500/OXY100)+ (PCP25/PPX300) w/ Cr, Ni, pH, Bl, S.G."
Where it went
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
How much
13,700 units

ImmuTest Drugs of Abuse Cup (AMT Item Number/ AMT Description): IMCA-5AB-W / "ImmuTest Drugs of Abuse Cup

Z-3036-2020  ·  Class II  ·  reported 2020-10-07  ·  Terminated

Why it was recalled
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Product
ImmuTest Drugs of Abuse Cup (AMT Item Number/ AMT Description): IMCA-5AB-W / "ImmuTest Drugs of Abuse Cup AMP1000+BZO300+COC300+OPI2000+THC50 With Cr, Ni, pH, Bl, S.G.", IMC-8BBCE/ "ImmuTest Drugs of Abuse Cup 8 Drug Detection: (COC300/THC50)+AMP1000+BZO300+BUP10+MET500+MTD300+OPI300 NO ADULTERATION", IMCA-8CM5 /"ImmuTest Drugs of Abuse Cup (OXY100/MDMA500)+(PCP25/AMP1000)+COC150+MET500+OPI2000+THC50 with Cr, Ni, pH, Bl, S.G."
Where it went
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
How much
27,375 units

Drug Screening Component: Uncut Sheets Part Number/ Description/ LOT Number 100163/ EDI OPI (300) Uncut Shee

Z-3051-2020  ·  Class II  ·  reported 2020-10-07  ·  Terminated

Why it was recalled
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Product
Drug Screening Component: Uncut Sheets Part Number/ Description/ LOT Number 100163/ EDI OPI (300) Uncut Sheet-Card /187876, 100195/ EDI PPX Uncut Sheet-Card/ 187877, 160130/ EDI COC Uncut Sheet-Cup/ 188168, 160180/ EDI PCP Uncut Sheet-Cup/ 189027, 160140/ EDI THC Uncut Sheet Cup /189028, 160170/ EDI BZO Uncut Sheet-Cup/ 189081, 160120/ EDI MET1000 Uncut Sheet-Cup/ 189082, 160160/ EDI OPI2000 Uncut Sheet-Cup / 189123, 160130/ EDI COC Uncut Sheet-Cup/ 189124, 160160/ EDI OPI2000 Uncut Sheet-Cup/ 189166, 160163/ EDI OPI (300) Uncut Sheet-Cup/ 189191, 160110/ EDI AMP Uncut Sheet-Cup/ 189371, 160140/ EDI THC Uncut Sheet Cup/ 189372, 160123/ EDI MET(300) Uncut Sheet-Cup/ 189425, 160170/ EDI BZO Uncut Sheet-Cup/ 189426, 100163/ EDI OPI (300) Uncut Sheet-Card/ 189456, 160170/ EDI BZO Uncut Sheet-Cup/ 189565, 160180/ EDI PCP Uncut Sheet-Cup/ 189601, 160125/ EDI MET(500) Uncut Sheet-Cup/ 189748, 160130/ EDI COC Uncut Sheet-Cup/ 189786, 160160/ EDI OPI2000 Uncut Sheet-Cup / 189844, 160180/ EDI PCP Uncut Sheet-Cup/ 189937, 160172/ EDI BZO(200) Uncut Sheet-CUP/189956 Updated 09/03/2020: EDI Affected products -Catalog Numbers/ Lot Numbers 31105-3/ PI-01081 ...
Where it went
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
How much
3600 units

TDDA Drugs of Abuse Cup(AMT Item/ AMT Description): TDDA-6MBAU-CUP /"TDDA Drugs of Abuse Cup AMP300/BZO200

Z-3044-2020  ·  Class II  ·  reported 2020-10-07  ·  Terminated

Why it was recalled
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Product
TDDA Drugs of Abuse Cup(AMT Item/ AMT Description): TDDA-6MBAU-CUP /"TDDA Drugs of Abuse Cup AMP300/BZO200/COC300/MET300/OPI300/ THC50 with Cr, Ni, pH, Bl, S.G."
Where it went
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
How much
26,775 units

Henry Schein One Step + DOA Cup (AMT Item Number/ AMT Description): 570-1066 / "Henry Schein OneStep+ DOA C

Z-3035-2020  ·  Class II  ·  reported 2020-10-07  ·  Terminated

Why it was recalled
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Product
Henry Schein One Step + DOA Cup (AMT Item Number/ AMT Description): 570-1066 / "Henry Schein OneStep+ DOA Cup 14 Panel AMP500/BAR/BUPG/BZO/COC150/MDMA/MET500/MOP/MTD/OXY/PCP/PPX/TCA/THC (Bl, Cr, Ni, pH, SG)",
Where it went
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
How much
31,800 units

First Check 14 Panel Drug Cup: (BMC style); AMT Item Number 990700B

Z-3047-2020  ·  Class II  ·  reported 2020-10-07  ·  Terminated

Why it was recalled
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Product
First Check 14 Panel Drug Cup: (BMC style); AMT Item Number 990700B
Where it went
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
How much
47,005 units

Consult Diagnostics Drugs Abuse Test Cup (AMT Item Number/ AMT Description): 164-DXA14 /"Consult Diagnostic

Z-3033-2020  ·  Class II  ·  reported 2020-10-07  ·  Terminated

Why it was recalled
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Product
Consult Diagnostics Drugs Abuse Test Cup (AMT Item Number/ AMT Description): 164-DXA14 /"Consult Diagnostics DRUGS OF ABUSE TEST CUP 14-DRUG PANEL WITH ADULTERANTS AMP500/BAR/BUPG/BZO/COC150/MDMA/MET500/MOP/MTD/OXY/PCP/PPX/TCA/THC (Bl, Cr, Ni, pH, SG)",
Where it went
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
How much
31,600 units

ToxCup Drug Screen Cup (AMT Item Number/ AMT Description): HT29A/"ToxCup¿ Drug Screen Cup COC300/MET500/

Z-3045-2020  ·  Class II  ·  reported 2020-10-07  ·  Terminated

Why it was recalled
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Product
ToxCup Drug Screen Cup (AMT Item Number/ AMT Description): HT29A/"ToxCup¿ Drug Screen Cup COC300/MET500/OPI300/THC50/AMP1000/PCP25/OXY100 Adulteration Screen for CR - NI - PH - BL - S.G.", DT14/ "ToxCup¿ Drug Screen Cup THC50/COC150; OPI300/MET500; AMP500/BZO300; BAR300/MTD300; BUPG10/TCA1000; MDMA500/OXY100; PCP25/PPX300", HT15/ "ToxCup¿ Drug Screen Cup AMP1000/COC300/MET500/PCP25/ THC50/OPI2000", HT15A/ ToxCup¿ Drug Screen Cup AMP1000/COC300/MET500/PCP25/ THC50/OPI2000 Adulteration Screen for CR - NI - PH - BL - S.G.", PT15/ "ToxCup¿ Drug Screen Cup AMP1000/COC300/MET500/OPI300/THC50" DT13A/ "ToxCup¿ Drug Screen Cup THC50/COC150; OPI300/MET500; AMP500/BZO300; BAR300/MTD300; BUPG10/TCA1000; MDMA500/OXY100 w/ Cr, Ni, pH, BI, S.G.", DT14A "ToxCup¿ Drug Screen Cup THC50/COC150; OPI300/MET500; AMP500/BZO300; BAR300/MTD300; BUPG10/TCA1000; MDMA500/OXY100; PCP25/PPX300 w/ Cr, Ni, pH, Bl, S.G. Finished Devices", DT12A/"ToxCup¿ Drug Screen Cup THC50/COC150; OPI300/MET500; AMP500/BZO300; BAR300/MTD300; MDMA500/OXY100; PCP25/PPX300 w/ Cr, Ni, pH, BI, S.G.", PT12A/ "ToxCup¿ Drug Screen Cup COC300/MET500/OPI2000/PCP25/THC50 Adulteration...
Where it went
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
How much
189,175 units

iCup Dx14 Drug Screen Cup(AMT Item Number/ AMT Description): X11-I-DXA-2147-011/ "iCup" Dx 14 Drug Screen C

Z-3040-2020  ·  Class II  ·  reported 2020-10-07  ·  Terminated

Why it was recalled
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Product
iCup Dx14 Drug Screen Cup(AMT Item Number/ AMT Description): X11-I-DXA-2147-011/ "iCup" Dx 14 Drug Screen Cup 14 Drug Detection: (THC50/COC150)+(MOP300/MET500)+(AMP500/BZO300)+ (BAR300/MTD300)+(BUPG10/TCA1000)+(MDMA500/OXY100)+(PCP25/PPX300) 5 Adulteration Tests: Cr, Ni, pH, Bl, S.G.", X11-I-DX-1147-022 /"iCup" Dx 14 Drug Screen Cup 14 Drug Detection: (THC50/COC150)+(MOP300/MET500)+(AMP500/BZO300)+ (BAR300/MTD300)+(BUPG10/TCA1000)+(MDMA500/OXY100)+(PCP25/PPX300)"
Where it went
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
How much
136,925 units

First Check 2 Panel Drug Cup: (THC/COC); AMT Item Number 990600B

Z-3048-2020  ·  Class II  ·  reported 2020-10-07  ·  Terminated

Why it was recalled
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Product
First Check 2 Panel Drug Cup: (THC/COC); AMT Item Number 990600B
Where it went
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
How much
41,340 units

QuickTox Drug Screen DipCard(AMT Item/ AMT Description): QT70/ "QuickTox¿ Drug Screen Dipcard 11 Drug Detec

Z-3043-2020  ·  Class II  ·  reported 2020-10-07  ·  Terminated

Why it was recalled
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Product
QuickTox Drug Screen DipCard(AMT Item/ AMT Description): QT70/ "QuickTox¿ Drug Screen Dipcard 11 Drug Detection: COC300+OPI300+MET500+THC50+AMP1000+PCP25+BZO300+BAR300+MTD300+OXY100+MDMA500", QT88/ "QuickTox¿ Drug Screen Dipcard 12 Drug Detection: COC300+OPI300+MET500+THC50+AMP1000+PCP25+BZO300+BAR300+MTD300+OXY100+ MDMA500+BUP10", QT11 / "QuickTox¿ Drug Screen Dipcard 5 Drug Detection: COC300+OPI2000+THC50+AMP1000+PCP25", QT52/ "QuickTox¿ Drug Screen Dipcard 9 Drug Detection: COC300+OPI300+MET500+THC50+AMP1000+BZO300+BAR300+MTD300+OXY100", QT44/ "QuickTox¿ Drug Screen Dipcard 8 Drug Detection: COC300+OPI300+MET500+THC50+AMP1000+BZO300+BAR300+OXY100", QT23/ "QuickTox¿ Drug Screen Dipcard 6 Drug Detection: COC300+OPI2000+MET500+THC50+AMP1000+PCP25", QT80F/ "QuickTox¿ Drug Screen Dipcard 12 Drug Detection: COC300+OPI300+MET500+THC50+AMP1000+PCP25+BZO300+BAR300+MTD300+TCA1000+OXY100+MDMA500 with Positive Control Line", QT52A/ "QuickTox¿ Drug Screen Dipcard 9 Drug Detection: COC300+OPI300+MET500+THC50+AMP1000+BZO300+BAR300+MTD300+OXY100 4 Adulteration Test: Cr, Ni, pH, Ox ", QT20/ "QuickTox" Drug Screen Dipcard 5 Drug Detection:...
Where it went
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
How much
170,425 units

ProScreen" Drugs of Abuse Test (AMT Item/ AMT Description): PSD-6MB/ "ProScreen" Drugs of Abuse Test (MET1

Z-3042-2020  ·  Class II  ·  reported 2020-10-07  ·  Terminated

Why it was recalled
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Product
ProScreen" Drugs of Abuse Test (AMT Item/ AMT Description): PSD-6MB/ "ProScreen" Drugs of Abuse Test (MET1000/OPI2000)+AMP+BZO+COC+THC", PSD-6MTDBO-300 /"ProScreen" Drugs of Abuse Test (COC/THC)+BZO+MTD+OPI300+OXY", PSCUP-5M-W/ ProScreen" Drugs of Abuse Cup AMP/COC/MET1000/OPI2000/THC), PSD-10MOB "ProScreen" Drugs of Abuse Test (COC/THC)+(MET1000/OPI2000)+(OXY/ MDMA)+AMP+BZO+BUP+MTD", PSD-5M / "ProScreen" Drugs of Abuse Test AMP/COC/MET1000/OPI2000/THC", PSCUPA-10OM / ProScreen" Drugs of Abuse Cup (COC/THC)+(MET1000/OPI2000)+(OXY/ MDMA) +(PCP/AMP)+BAR+BZO
Where it went
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
How much
51,075 units

First Check 4 Panel Drug Cup: COC/MET1000/OPI2000/THC): (THC/COC); AMT Item Number 990100B

Z-3049-2020  ·  Class II  ·  reported 2020-10-07  ·  Terminated

Why it was recalled
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Product
First Check 4 Panel Drug Cup: COC/MET1000/OPI2000/THC): (THC/COC); AMT Item Number 990100B
Where it went
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
How much
537,120 units

Alere iScreen Single Drug Detection (AMT Item Number/ AMT Description): X11-IS1 BUP-DX/ "Alere iScreen" Dx S

Z-3029-2020  ·  Class II  ·  reported 2020-10-07  ·  Terminated

Why it was recalled
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Product
Alere iScreen Single Drug Detection (AMT Item Number/ AMT Description): X11-IS1 BUP-DX/ "Alere iScreen" Dx Single Dip Card Single Drug Detection: Buprenorphine (BUP)", X11-IS1 BAR-DX /"Alere iScreen" Dx Single Dip Card Single Drug Detection: Barbiturate (BAR)", X11-IS1 OXY-DX / "Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Oxycodone (OXY)", X11-IS1 BZO-DX / "Alere iScreen" Dx Single Dip Card Single Drug Detection: Benzodiazepine (BZO300)", X11-IS1 MET-DX / "Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Methamphetamine (MET)", X11-IS1 AMP-DX /"Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Amphetamine (AMP)", X11-IS1 MDM-DX / "Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Ecstasy (MDMA)", X11-IS1 MTD-DX /"Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Methadone (MTD)", X11-IS1 MOP-DX/"Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Morphine (MOP)", X11-IS1 THC-DX/"Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Marijuana (THC)", X11-IS1 TCA-DX/ "Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Tricyclic Antidrepessants (TCA)", X11-IS1 PCP-DX /"Alere iScreen¿ Dx Single Dip Card ...
Where it went
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
How much
303,100 units

Alere iScreen Multi Drug Detection (AMT Item Number/ AMT Description): X11-IS12-DX / "Alere iScreen" Dx Mul

Z-3030-2020  ·  Class II  ·  reported 2020-10-07  ·  Terminated

Why it was recalled
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Product
Alere iScreen Multi Drug Detection (AMT Item Number/ AMT Description): X11-IS12-DX / "Alere iScreen" Dx Multi-Drugs of Abuse Dip Test 12 Drug Detection:AMP1000+BAR300+BZO300+COC300+THC50+MTD300+MET500+MOP300+OXY100+PCP25+MDMA500+TCA1000", X11-IS12-DX / "Alere iScreen" Dx Multi-Drugs of Abuse Dip Test 12 Drug Detection: AMP1000+BAR300+BZO300+COC300+THC50+MTD300+MET500+MOP300+OXY100+PCP25+MDMA500+TCA1000", X11-IS10BUP-DX / "Alere iScreen" Dx Multi-Drugs of Abuse Dip Test AMP1000+BAR300+BUP10+BZO300+COC300+THC50+MET500+MTD300+MOP300+OXY100", X11-IS12-DX / "Alere iScreen" Dx Multi-Drugs of Abuse Dip Test 12 Drug Detection: AMP1000+BAR300+BZO300+COC300+THC50+MTD300+MET500+MOP300+OXY100+PCP25+MDMA500+TCA1000", X11-IS10BUP-DX / "Alere iScreen" Dx Multi-Drugs of Abuse Dip Test AMP1000+BAR300+BUP10+BZO300+COC300+THC50+MET500+MTD300+MOP300+OXY100", X11-IS10BUP-DX/ "Alere iScreen" Dx Multi-Drugs of Abuse Dip Test AMP1000+BAR300+BUP10+BZO300+COC300+THC50+MET500+MTD300+MOP300+OXY100", X11-IS10-DX /"Alere iScreen" Dx Multi-Drugs of Abuse Dip Test 10 Drug Detection: Amphetamine, Barbiturate, Benzodiazepine, Cocaine, Marijuana, Methadone, Methamphetamine, Morphine,...
Where it went
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
How much
67,300 units

MicroTox (AMT Item Number/ AMT Description): MTPA-6MBAU / "MicroTox (AMP300/BZO200)+(COC300/THC50)+ (MET30

Z-3038-2020  ·  Class II  ·  reported 2020-10-07  ·  Terminated

Why it was recalled
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Product
MicroTox (AMT Item Number/ AMT Description): MTPA-6MBAU / "MicroTox (AMP300/BZO200)+(COC300/THC50)+ (MET300/MOR300) With Cr, Ni, pH, Bl, S.G.", MTPA-6MBAU/ "MicroTox (AMP300/BZO200)+(COC300/THC50)+ (MET300/MOR300) With Cr, Ni, pH, Bl, S.G."
Where it went
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
How much
12, 100 units

Oratect III Oral Fluid Drug Screen Device (AMT Item Number/ AMT Description): HM12-CE "Oratect¿ III Oral Fl

Z-3041-2020  ·  Class II  ·  reported 2020-10-07  ·  Terminated

Why it was recalled
XXX
Product
Oratect III Oral Fluid Drug Screen Device (AMT Item Number/ AMT Description): HM12-CE "Oratect¿ III Oral Fluid Drug Screen Device MET25+THC40+COC20+AMP25+OPI10+BZO5 "
Where it went
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
How much
1,325 units

MicroScreen (AMT Item Number/ AMT Description): MSCA-6MBAU/ "MicroScreen AMP300/BZO200/COC300/MET300/OPI300/

Z-3037-2020  ·  Class II  ·  reported 2020-10-07  ·  Terminated

Why it was recalled
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Product
MicroScreen (AMT Item Number/ AMT Description): MSCA-6MBAU/ "MicroScreen AMP300/BZO200/COC300/MET300/OPI300/ THC50 with Cr, Ni, pH, Bl, S.G. "
Where it went
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
How much
23,000 units

Alere iCup Drug Detection (AMT Item Number/ AMT Description): I-DXA-10/ "Alere" iCup¿ Dx 10 Drug Detection

Z-3031-2020  ·  Class II  ·  reported 2020-10-07  ·  Terminated

Why it was recalled
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Product
Alere iCup Drug Detection (AMT Item Number/ AMT Description): I-DXA-10/ "Alere" iCup¿ Dx 10 Drug Detection + Adulteration AMP/BAR/BZO/COC/MET/MDMA/MTD/MOP/OXY/THC Bl, Cr, Ni, pH, SG", I-DXA-13B/"Alere" iCup¿ DX Drug Screen Cup 13 Drug Detection + Adulteration AMP/BAR/BZO/BUPG/COC/MET/MDMA/MTD/MOP/OXY/PCP/THC/TCA Bl, Cr, Ni, pH, SG", I-DXA-1127-023-N "Alere" iCup¿ Dx Drugs of Abuse Test (COC/THC)+(MET1000/OPI2000)+(MTD/TCA)+ (OXY/MDMA)+(PCP/AMP)+BAR+BZO With Cr, Ni, pH, Bl, S.G."
Where it went
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
How much
28,525

Fastect¿ II Drug Screen Dipstick Test Single Drug Detection (AMT Item Number/ AMT Description): MF25/ "Fast

Z-3034-2020  ·  Class II  ·  reported 2020-10-07  ·  Terminated

Why it was recalled
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Product
Fastect¿ II Drug Screen Dipstick Test Single Drug Detection (AMT Item Number/ AMT Description): MF25/ "Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: BUP10", QF09/"Fastect¿ II Drug Screen Dipstick Test, 4 Drug Detection: THC50+COC300+OPI300+BZO300", MF20/ "Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: MTD300", MF21/ "Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: TCA1000, MF23/ "Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: OXY100", MF23/ "Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: OXY100", MF18/ "Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: BZO300, MF18/ "Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: BZO300", QF12/ "Fastect¿ II Drug Screen Dipstick Test 4 Drug Detection: THC50+COC300+OPI300+MET500", MF10/ "Fastect¿ II Drug Screen Dipstick Test THC50", MF25/ "Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: BUP10", MF16 /"Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: MET500", MF23 /"Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: OXY100", MF14/ "Fastect¿ II Drug Screen Dipstick Test AMP1000", MF11/ "Fastect¿ II...
Where it went
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
How much
108,000 units

First Check 1 Panel Drug Cup: (THC); AMT Item Number 990400B

Z-3046-2020  ·  Class II  ·  reported 2020-10-07  ·  Terminated

Why it was recalled
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Product
First Check 1 Panel Drug Cup: (THC); AMT Item Number 990400B
Where it went
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
How much
2,555,630 units

CLIA Waived Rapid Drug Test (AMT Item Number/ AMT Description): X07-CLIA-SDDT-25 / CLIA Waived" Inc. Singl

Z-3032-2020  ·  Class II  ·  reported 2020-10-07  ·  Terminated

Why it was recalled
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Product
CLIA Waived Rapid Drug Test (AMT Item Number/ AMT Description): X07-CLIA-SDDT-25 / CLIA Waived" Inc. Single Drug Dipstick Test BUP10, X07-CLIA-SDDT-25 / "CLIA Waived" Inc. Single Drug Dipstick Test BUP10, X07-CLIA-SDDT-13 /"CLIA Waived" Inc. Single Drug Dipstick Test OPI300, X07-CLIA-SDDT-25/"CLIA Waived" Inc. Single Drug Dipstick Test BUP10, X07-CLIA-14-RDTC-A/ "CLIA Waived" Inc. Rapid Drug Test Cup (THC50/COC150)(OPI300/MET500)(AMP500/ BZO300)(BAR300/MTD300)(BUPG10/TCA1000) (MDMA500/OXY100)(PCP25/PPX300) + CR, NI, PH, BL, SG", X07-CLIA-14-RDTC-A/ "CLIA Waived" Inc. Rapid Drug Test Cup (THC50/COC150)(OPI300/MET500)(AMP500/ BZO300)(BAR300/MTD300)(BUPG10/TCA1000) (MDMA500/OXY100)(PCP25/PPX300) + CR, NI, PH, BL, SG", X07-CLIA-12-RDTC-BUP-A/ "CLIA Waived" Inc. Rapid Drug Test Cup (THC50/COC150)(OPI300/MET500)(AMP500/BZO300)(BAR300/MTD300)(BUPG10/TCA1000) (MDMA500/OXY100) + CR, NI, PH, BL, S.G.", X07-CLIA-12-RDTC-BUP-A /"CLIA Waived" Inc. Rapid Drug Test Cup (THC50/COC150)(OPI300/MET500)(AMP500/BZO300)(BAR300/MTD300)(BUPG10/TCA1000) (MDMA500/OXY100) + CR, NI, PH, BL, S.G.", X07-CLIA-14-RDTC-A /"CLIA Waived" Inc. Rapid Drug Test Cup ...
Where it went
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
How much
120, 150

QuickTox 5 Panel Drug Screen DipCard

Z-0692-2020  ·  Class III  ·  reported 2019-12-25  ·  Terminated

Why it was recalled
Mixed Products containing QuickTox 5 Panel Drugs Screen Dipcard (P/N QT11, LOT #188787) and Alere iScreen Drugs of Abuse Dip Test (P/N X11-IS10-22015-DX LOT #188744)
Product
QuickTox 5 Panel Drug Screen DipCard
Where it went
US - AZ, CA, FL, GA, KS, LA, MI, MO, MS, NC, OK, TN, and TX. OUS - None
How much
3400 units (136 kits)

Accutest Drug Test Cup 5+3 Product Usage: The Accutest Drug Test Cup is an in vitro diagnostic test for th

Z-0270-2019  ·  Class III  ·  reported 2018-10-31  ·  Terminated

Why it was recalled
Product was released to market without the adulteration strip included in the product.
Product
Accutest Drug Test Cup 5+3 Product Usage: The Accutest Drug Test Cup is an in vitro diagnostic test for the rapid detection of multiple drugs in human urine. The test is intended as the first step in a two-step process to provide consumers with information concerning the presence or absence of drugs or their metabolites in a urine sample. Information regarding confirmatory testing- the second step in the process-is provided in the package labeling.
Where it went
US in the state of California
How much
40 kit boxes (1000 units)

Alere iScreen Dx Urine Drug Screen Card

Z-1542-2018  ·  Class III  ·  reported 2018-05-09  ·  Terminated

Why it was recalled
The kit box label incorrectly identifies Phencyclidine (PCP), where it should list Tricyclic Antidepressants (TCA) as one of the 10 drug analytes included in the kit box.
Product
Alere iScreen Dx Urine Drug Screen Card
Where it went
Distributed domestically to AZ, FL, IL, ND, NV, PA, TN, TX, VA.
How much
828 kits (20,700 devices)

Alere iScreen Dx Urine Drug Screen Card, 20 Drug Detection, Kit Contents include 26 individually pouched devic

Z-3084-2017  ·  Class III  ·  reported 2017-09-06  ·  Terminated

Why it was recalled
The kit box label incorrectly identifies PCP (instead of TCA) as one of the 10 drug analytes included in the kit. The pouch label, product insert, and the device itself are labeled correctly.
Product
Alere iScreen Dx Urine Drug Screen Card, 20 Drug Detection, Kit Contents include 26 individually pouched devices and 1 IFU insert. The Alere iScreen Dx Drug Screen Card is a rapid urine screening test that can be performed without the use of an instrument. The test utilizes monoclonal antibodies to selectively detect elevated levels of specific drugs in urine
Where it went
Nationwide Distribution to AZ, IL, IN, KY, NV, PA, TN, and TX
How much
263Kits

DrugSmart Cup 10, Item No. 61093D-3C, 61085D, 61015D

Z-0497-2016  ·  Class III  ·  reported 2015-12-30  ·  Terminated

Why it was recalled
Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
Product
DrugSmart Cup 10, Item No. 61093D-3C, 61085D, 61015D
Where it went
Nationwide Distribution.
How much
272 units

11 Panel Card (OPI 300) w/Adult, Item No. 71125AA

Z-0487-2016  ·  Class III  ·  reported 2015-12-30  ·  Terminated

Why it was recalled
Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
Product
11 Panel Card (OPI 300) w/Adult, Item No. 71125AA
Where it went
Nationwide Distribution.
How much
50 kits

ImmuTest Cup 7-panel w/ Adulteration, Item No. IMCA-7P

Z-0512-2016  ·  Class III  ·  reported 2015-12-30  ·  Terminated

Why it was recalled
Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
Product
ImmuTest Cup 7-panel w/ Adulteration, Item No. IMCA-7P
Where it went
Nationwide Distribution.
How much
60 kits

Showing the 30 most recent of 109. The rest are searchable at the FDA enforcement report database.

Read these counts with three things in mind

A firm with more recalls usually makes more products. A contract packer runs hundreds of lines. It will out-recall a company with one product whatever either does. This is not a safety ranking.

Most recalls here are voluntary. A firm found its own fault and pulled the stock. That is the system working. It lands in this database looking the same as a recall the FDA had to push for.

Terminated means the FDA considers the recall finished. It is not a verdict. It does not mean anyone was harmed.

Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.

← All device recalls