Abbott Vascular: 37 FDA recall reports
Every medical device recall the FDA has published for Abbott Vascular. The fault, and how the agency graded it.
37Reports
23Class I
9Ongoing
2022-10-26Most recent
23 Class I on this pageClass I means the FDA judged there was a reasonable probability of serious harm or death. Check the lot numbers against anything you hold.
What the classes mean
The FDA sorts every recall into one of three classes. The class is the whole story. It is also the part people skip.
Class II is the bulk of every list here. That matters. A long list of Class II says something different from one Class I.
By class
By year reported
Checking whether it affects you
The lot or code number is what matters. Not the product name. A recall covers
specific lots. Yours may not be one of them.
Check the code on your package against the report. The FDA keeps the current
list at fda.gov/safety/recalls.
The reports
The text below is quoted from the FDA record word for word.
Product/Part: MitraClip XTR Clip Delivery System US/CDS0601-XTR, MitraClip G4 Clip Delivery System NT US/CDS0
Z-0115-2023 · Class II · reported 2022-10-26 · Ongoing
- Why it was recalled
- An increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked, have been observed. Contributing causes may be clip locking component material properties and certain use conditions. Unsuccessful clip procedure can lead to recurrence of severe mitral regurgitation, device embolization, or tissue injury.
- Product
- Product/Part: MitraClip XTR Clip Delivery System US/CDS0601-XTR, MitraClip G4 Clip Delivery System NT US/CDS0701-NT, MitraClip G4 Clip Delivery System NTW US/CDS0701-NTW, MitraClip G4 Clip Delivery System XT US/CDS0701-XT, MitraClip G4 Clip Delivery System XTW US/CDS0701-XTW. For cardiac procedures.
- Where it went
- Worldwide distribution - US Nationwide including in the states of PR, WA, NE, MO, IA, KS, TN, MN, ND, MD, VA, MA, PA, DC, LA, KY, CA, NV, IL, WI, TX, CT, VT, AL, AR, GA, MS, MI, OH, FL, NC, NJ, CO, NY, DE, OR, SC, IN, HI, AZ, NH, ME, WV, OK, ID, UT, MT, NM, RI, AK, SD and the countries of AE, AR, AT, AU, BE, BG, BR, BY, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, EC, EE, EG, ES, FI, FR, GB, GR, HK, HU, ID, IE, IL, IN, IR, IT, JO, JP, KR, KW, LB, LU, MT, MX, MY, NL, NO, NZ, OM, PE, PH, PK, PL, PT, QA, RE, RO, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW.
- How much
- 18,042
Product/Part: MitraClip NTR Clip Delivery System US/CDS0601-NTR, MitraClip NTR Clip Delivery System CE/CDS060
Z-0116-2023 · Class II · reported 2022-10-26 · Ongoing
- Why it was recalled
- An increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked, have been observed. Contributing causes may be clip locking component material properties and certain use conditions. Unsuccessful clip procedure can lead to recurrence of severe mitral regurgitation, device embolization, or tissue injury.
- Product
- Product/Part: MitraClip NTR Clip Delivery System US/CDS0601-NTR, MitraClip NTR Clip Delivery System CE/CDS0602-NTR, MitraClip XTR Clip Delivery System CE/CDS0602-XTR, MitraClip G4 Clip Delivery System NT CE/CDS0702-NT, MitraClip G4 Clip Delivery System NTW CE/CDS0702-NTW, MitraClip G4 Clip Delivery System XT CE/CDS0702-XT, MitraClip G4 Clip Delivery System XTW CE/CDS0702-XTW, MitraClip G4 Clip Delivery Sys NT ROW/CDS0705-NT, MitraClip G4 Clip Delivery Sys NTW ROW/CDS0705-NTW, MitraClip G4 Clip Delivery Sys XT ROW/CDS0705-XT, MitraClip G4 Clip Delivery Sys XTW ROW/CDS0705-XTW, TriClip XT Clip Delivery Sys CE/TCDS0202-XT, TriClip NT Clip Delivery Sys CE/TCDS0202-NT, TriClip XT Clip Delivery Sys ROW/TCDS0205-XT, TriClip NT Clip Delivery Sys ROW/TCDS0205-NT, TriClip G4 NT Delivery System CE/TCDS0302-NT, TriClip G4 NTW Delivery System CE/TCDS0302-NTW, TriClip G4 XT Delivery System CE/TCDS0302-XT, TriClip G4 XTW Delivery System CE/TCDS0302-XTW, TriClip G4 NT Delivery System ROW/TCDS0305-NT, TriClip G4 NTW Delivery System ROW/TCDS0305-NTW, TriClip G4 XT Delivery System ROW/TCDS0305-XT, TriClip G4 XTW Delivery System ROW/TCDS0305-XTW, TriClip NT Clip Delivery Sys...
- Where it went
- Worldwide distribution - US Nationwide including in the states of PR, WA, NE, MO, IA, KS, TN, MN, ND, MD, VA, MA, PA, DC, LA, KY, CA, NV, IL, WI, TX, CT, VT, AL, AR, GA, MS, MI, OH, FL, NC, NJ, CO, NY, DE, OR, SC, IN, HI, AZ, NH, ME, WV, OK, ID, UT, MT, NM, RI, AK, SD and the countries of AE, AR, AT, AU, BE, BG, BR, BY, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, EC, EE, EG, ES, FI, FR, GB, GR, HK, HU, ID, IE, IL, IN, IR, IT, JO, JP, KR, KW, LB, LU, MT, MX, MY, NL, NO, NZ, OM, PE, PH, PK, PL, PT, QA, RE, RO, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW.
- How much
- 20,045
Product/Part: MitraClip G4 CDS NT REPAIR-MR Clinical/CDS0704-NT, MitraClip G4 CDS NTW REPAIR-MR Clinical/CDS0
Z-0117-2023 · Class II · reported 2022-10-26 · Ongoing
- Why it was recalled
- An increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked, have been observed. Contributing causes may be clip locking component material properties and certain use conditions. Unsuccessful clip procedure can lead to recurrence of severe mitral regurgitation, device embolization, or tissue injury.
- Product
- Product/Part: MitraClip G4 CDS NT REPAIR-MR Clinical/CDS0704-NT, MitraClip G4 CDS NTW REPAIR-MR Clinical/CDS0704-NTW, MitraClip G4 CDS XT REPAIR-MR Clinical/CDS0704-XT, MitraClip G4 CDS XTW REPAIR-MR Clinical/CDS0704-XTW, TriClip NT Clip Delivery Sys CLIN/TCDS0203-NT, TriClip XT Clip Delivery Sys CLIN/TCDS0203-XT, TriClip G4 NT Delivery System CLIN/TCDS0303-NT, TriClip G4 NTW Delivery System CLIN/TCDS0303-NTW, TriClip G4 XT Delivery System CLIN/TCDS0303-XT, TriClip G4 XTW Delivery System CLIN/TCDS0303-XTW, TriClip NT Clip Delivery Sys CLIN/TCDS0203-NT, TriClip XT Clip Delivery Sys CLIN/TCDS0203-XT. For cardiac procedures.
- Where it went
- Worldwide distribution - US Nationwide including in the states of PR, WA, NE, MO, IA, KS, TN, MN, ND, MD, VA, MA, PA, DC, LA, KY, CA, NV, IL, WI, TX, CT, VT, AL, AR, GA, MS, MI, OH, FL, NC, NJ, CO, NY, DE, OR, SC, IN, HI, AZ, NH, ME, WV, OK, ID, UT, MT, NM, RI, AK, SD and the countries of AE, AR, AT, AU, BE, BG, BR, BY, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, EC, EE, EG, ES, FI, FR, GB, GR, HK, HU, ID, IE, IL, IN, IR, IT, JO, JP, KR, KW, LB, LU, MT, MX, MY, NL, NO, NZ, OM, PE, PH, PK, PL, PT, QA, RE, RO, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW.
- How much
- 767
Class I
Dragonfly OpStar Imaging Catheter, REF 1014651/1014652, 0.068 in, 0.014 in, 135 cm, STERILE EO
Z-1095-2022 · Class I · reported 2022-05-25 · Ongoing
- Why it was recalled
- There is a potential that the proximal marker may separate from the imaging catheter
- Product
- Dragonfly OpStar Imaging Catheter, REF 1014651/1014652, 0.068 in, 0.014 in, 135 cm, STERILE EO
- Where it went
- US:AL AR AZ CA CO CT DC FL GA HI IA IL IN KS KY MA MD ME MI MN MO MT NC NE NH NJ NM NV NY OH OK OR PA PR RI SC TN TX UT VA VT WA WI, PR OUS: AE, AT, AW, BE, CA, CH, DE, DK, IT, JP, LU, NL, NO, NZ, PL, EE, ES, FI, FR, GB, HK, HU, IE, PT, RO, SE, SG, SI, TH
- How much
- 14,544 UNITS
Plus 30 Priority Pack w/.096 RHV - REF 1000183 WPL2122277-01 (20-03-01) Plus 30 Priority Pack w/.115 RHV -
Z-0925-2022 · Class II · reported 2022-04-20 · Ongoing
- Why it was recalled
- Due to an increase in complaint trend for leaks and intermittent/loose connections.
- Product
- Plus 30 Priority Pack w/.096 RHV - REF 1000183 WPL2122277-01 (20-03-01) Plus 30 Priority Pack w/.115 RHV - REF 1000185-115 WPL2122279-01 (2019-03-01)
- Where it went
- U.S Nationwide Distribution.: AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, TN, TX, VA, WA, and WI. O.U.S.: United Arab Emirates, France, Pakistan, Argentina, United Kingdom, Portugal, Australia, Greece, R¿union, Belgium, Guam, Romania, Brazil, Indonesia, Russian Federation, Bahamas, Israel, Saudi Arabia, Canada, India, Singapore, Switzerland, Italy, Slovakia, Chile, Jamaica, Thailand, China, Japan, Taiwan Republic of China, Colombia, Korea (South), Costa Rica, Luxembourg, Uruguay, Czech Republic, Mexico, Vietnam, Germany, Netherlands, South Africa, Dominican Republic, New Zealand, Spain, and Panama
- How much
- 7 lots
INDEFLATOR Plus 30 - REF 1000183 WPL2122277-01 (2019-03-01)
Z-0924-2022 · Class II · reported 2022-04-20 · Ongoing
- Why it was recalled
- Due to an increase in complaint trend for leaks and intermittent/loose connections.
- Product
- INDEFLATOR Plus 30 - REF 1000183 WPL2122277-01 (2019-03-01)
- Where it went
- U.S Nationwide Distribution.: AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, TN, TX, VA, WA, and WI. O.U.S.: United Arab Emirates, France, Pakistan, Argentina, United Kingdom, Portugal, Australia, Greece, R¿union, Belgium, Guam, Romania, Brazil, Indonesia, Russian Federation, Bahamas, Israel, Saudi Arabia, Canada, India, Singapore, Switzerland, Italy, Slovakia, Chile, Jamaica, Thailand, China, Japan, Taiwan Republic of China, Colombia, Korea (South), Costa Rica, Luxembourg, Uruguay, Czech Republic, Mexico, Vietnam, Germany, Netherlands, South Africa, Dominican Republic, New Zealand, Spain, and Panama
- How much
- 6 lots
20/30 INDEFLATOR -REF 1000184 WPL2122268-01 (2019-03-01)
Z-0921-2022 · Class II · reported 2022-04-20 · Ongoing
- Why it was recalled
- Due to an increase in complaint trend for leaks and intermittent/loose connections.
- Product
- 20/30 INDEFLATOR -REF 1000184 WPL2122268-01 (2019-03-01)
- Where it went
- U.S Nationwide Distribution.: AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, TN, TX, VA, WA, and WI. O.U.S.: United Arab Emirates, France, Pakistan, Argentina, United Kingdom, Portugal, Australia, Greece, R¿union, Belgium, Guam, Romania, Brazil, Indonesia, Russian Federation, Bahamas, Israel, Saudi Arabia, Canada, India, Singapore, Switzerland, Italy, Slovakia, Chile, Jamaica, Thailand, China, Japan, Taiwan Republic of China, Colombia, Korea (South), Costa Rica, Luxembourg, Uruguay, Czech Republic, Mexico, Vietnam, Germany, Netherlands, South Africa, Dominican Republic, New Zealand, Spain, and Panama
- How much
- 17 lots
20/30 Priority Pack Accessory Kit/.096 RHV - REF 1000186 WPL2122270-01 (2019-03-01) 20/30 Priority Pack w/
Z-0922-2022 · Class II · reported 2022-04-20 · Ongoing
- Why it was recalled
- Due to an increase in complaint trend for leaks and intermittent/loose connections.
- Product
- 20/30 Priority Pack Accessory Kit/.096 RHV - REF 1000186 WPL2122270-01 (2019-03-01) 20/30 Priority Pack w/.115 RHV - REF 1000186-115 WPL2122270-02 (2019-03-01)
- Where it went
- U.S Nationwide Distribution.: AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, TN, TX, VA, WA, and WI. O.U.S.: United Arab Emirates, France, Pakistan, Argentina, United Kingdom, Portugal, Australia, Greece, R¿union, Belgium, Guam, Romania, Brazil, Indonesia, Russian Federation, Bahamas, Israel, Saudi Arabia, Canada, India, Singapore, Switzerland, Italy, Slovakia, Chile, Jamaica, Thailand, China, Japan, Taiwan Republic of China, Colombia, Korea (South), Costa Rica, Luxembourg, Uruguay, Czech Republic, Mexico, Vietnam, Germany, Netherlands, South Africa, Dominican Republic, New Zealand, Spain, and Panama
- How much
- 23 lots
20/30 Priority Pack w/COPILOT - REF 1003327 WPL2122272-01 (2019-03-01)
Z-0923-2022 · Class II · reported 2022-04-20 · Ongoing
- Why it was recalled
- Due to an increase in complaint trend for leaks and intermittent/loose connections.
- Product
- 20/30 Priority Pack w/COPILOT - REF 1003327 WPL2122272-01 (2019-03-01)
- Where it went
- U.S Nationwide Distribution.: AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, TN, TX, VA, WA, and WI. O.U.S.: United Arab Emirates, France, Pakistan, Argentina, United Kingdom, Portugal, Australia, Greece, R¿union, Belgium, Guam, Romania, Brazil, Indonesia, Russian Federation, Bahamas, Israel, Saudi Arabia, Canada, India, Singapore, Switzerland, Italy, Slovakia, Chile, Jamaica, Thailand, China, Japan, Taiwan Republic of China, Colombia, Korea (South), Costa Rica, Luxembourg, Uruguay, Czech Republic, Mexico, Vietnam, Germany, Netherlands, South Africa, Dominican Republic, New Zealand, Spain, and Panama
- How much
- 131 lots
Xience Sierra TM Everolimus Eluting Coronary Stent System, RX 3.5mm x 33mmStrength: 100 ¿g/cm¿, Nominal Everol
Z-2370-2020 · Class II · reported 2020-06-24 · Terminated
- Why it was recalled
- Incorrect expiration date
- Product
- Xience Sierra TM Everolimus Eluting Coronary Stent System, RX 3.5mm x 33mmStrength: 100 ¿g/cm¿, Nominal Everolimus Content: 209 ¿g; Lot number 903224A,
- Where it went
- US Nationwide distribution including in the states of NJ, HI, MD, FL, NV, CA, KS and PR. No foreign consignee
- How much
- 13 units
Class I
NC TRAVELER Coronary Dilatation Catheter, REF 1013157-15, Rx Only,NC Traveler RX 4.0 X 15MM The NC Travele
Z-1142-2020 · Class I · reported 2020-02-26 · Terminated
- Why it was recalled
- Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
- Product
- NC TRAVELER Coronary Dilatation Catheter, REF 1013157-15, Rx Only,NC Traveler RX 4.0 X 15MM The NC Traveler RX Coronary Dilatation Catheter is not cleared for commercial distribution in the United States.
- Where it went
- Worldwide Distribution: US (Nation-wide) and OUS (countries of): Algeria Jordan Argentina Korea Armenia Kuwait Aruba Lebanon Australia Malaysia Austria, Mauritius Bahrain Mexico Bangladesh Morocco Belgium Netherlands Brazil New Zealand Brunei Norway Bulgaria Oman Canada Pakistan Chile Panama China Peru Colombia Poland Costa Rica Portugal Croatia R¿union Cyprus Romania Czech Republic Russian Federation Denmark Saudi Arabia Dominican Republic Singapore Egypt Slovakia Estonia Slovenia Finland South Africa France Spain Germany Sri Lanka Greece State of Palestine Hong Kong ...
- How much
- Total 791 Traveler (0 US and 791 OUS)
Class I
NC TREK Coronary Dilatation Catheter, REF 1012453-20 Rx Only, NC TREK RX 4.50 X 8MM BDC The NC TREK RX Cor
Z-1132-2020 · Class I · reported 2020-02-26 · Terminated
- Why it was recalled
- Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
- Product
- NC TREK Coronary Dilatation Catheter, REF 1012453-20 Rx Only, NC TREK RX 4.50 X 8MM BDC The NC TREK RX Coronary Dilatation Catheter is indicated for: a) balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis, for the purpose of improving myocardial perfusion b) balloon dilatation of a coronary artery occlusion, for the purpose of restoring coronary flow in patients with ST-segment elevation myocardial infarction c) balloon dilatation of a stent after implantation (balloon models 2.00 mm 5.00 mm only)
- Where it went
- Worldwide Distribution: US (Nation-wide) and OUS (countries of): Algeria Jordan Argentina Korea Armenia Kuwait Aruba Lebanon Australia Malaysia Austria, Mauritius Bahrain Mexico Bangladesh Morocco Belgium Netherlands Brazil New Zealand Brunei Norway Bulgaria Oman Canada Pakistan Chile Panama China Peru Colombia Poland Costa Rica Portugal Croatia R¿union Cyprus Romania Czech Republic Russian Federation Denmark Saudi Arabia Dominican Republic Singapore Egypt Slovakia Estonia Slovenia Finland South Africa France Spain Germany Sri Lanka Greece State of Palestine Hong Kong ...
- How much
- Total 39687 NC TREK units (13891 US and 25796 OUS)
Class I
NC TREK Coronary Dilatation Catheter, REF 1012453-20 Rx Only, NC TREK RX 4.50 X 8MM BDC The NC TREK RX Coro
Z-1133-2020 · Class I · reported 2020-02-26 · Terminated
- Why it was recalled
- Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
- Product
- NC TREK Coronary Dilatation Catheter, REF 1012453-20 Rx Only, NC TREK RX 4.50 X 8MM BDC The NC TREK RX Coronary Dilatation Catheter is indicated for: a) balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis, for the purpose of improving myocardial perfusion b) balloon dilatation of a coronary artery occlusion, for the purpose of restoring coronary flow in patients with ST-segment elevation myocardial infarction c) balloon dilatation of a stent after implantation (balloon models 2.00 mm 5.00 mm only)
- Where it went
- Worldwide Distribution: US (Nation-wide) and OUS (countries of): Algeria Jordan Argentina Korea Armenia Kuwait Aruba Lebanon Australia Malaysia Austria, Mauritius Bahrain Mexico Bangladesh Morocco Belgium Netherlands Brazil New Zealand Brunei Norway Bulgaria Oman Canada Pakistan Chile Panama China Peru Colombia Poland Costa Rica Portugal Croatia R¿union Cyprus Romania Czech Republic Russian Federation Denmark Saudi Arabia Dominican Republic Singapore Egypt Slovakia Estonia Slovenia Finland South Africa France Spain Germany Sri Lanka Greece State of Palestine Hong Kong ...
- How much
- Total 39687 NC TREK units (13891 US and 25796 OUS)
Class I
NC TRAVELER Coronary Dilatation Catheter, REF 1013158-08, Rx Only,NC Traveler RX 4.5 X 8MM The NC Traveler
Z-1143-2020 · Class I · reported 2020-02-26 · Terminated
- Why it was recalled
- Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
- Product
- NC TRAVELER Coronary Dilatation Catheter, REF 1013158-08, Rx Only,NC Traveler RX 4.5 X 8MM The NC Traveler RX Coronary Dilatation Catheter is not cleared for commercial distribution in the United States.
- Where it went
- Worldwide Distribution: US (Nation-wide) and OUS (countries of): Algeria Jordan Argentina Korea Armenia Kuwait Aruba Lebanon Australia Malaysia Austria, Mauritius Bahrain Mexico Bangladesh Morocco Belgium Netherlands Brazil New Zealand Brunei Norway Bulgaria Oman Canada Pakistan Chile Panama China Peru Colombia Poland Costa Rica Portugal Croatia R¿union Cyprus Romania Czech Republic Russian Federation Denmark Saudi Arabia Dominican Republic Singapore Egypt Slovakia Estonia Slovenia Finland South Africa France Spain Germany Sri Lanka Greece State of Palestine Hong Kong ...
- How much
- Total 791 Traveler (0 US and 791 OUS)
Class I
NC TRAVELER Coronary Dilatation Catheter, REF 1013157-12, Rx Only,NC Traveler RX 4.0 X 12MM The NC Travele
Z-1141-2020 · Class I · reported 2020-02-26 · Terminated
- Why it was recalled
- Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
- Product
- NC TRAVELER Coronary Dilatation Catheter, REF 1013157-12, Rx Only,NC Traveler RX 4.0 X 12MM The NC Traveler RX Coronary Dilatation Catheter is not cleared for commercial distribution in the United States.
- Where it went
- Worldwide Distribution: US (Nation-wide) and OUS (countries of): Algeria Jordan Argentina Korea Armenia Kuwait Aruba Lebanon Australia Malaysia Austria, Mauritius Bahrain Mexico Bangladesh Morocco Belgium Netherlands Brazil New Zealand Brunei Norway Bulgaria Oman Canada Pakistan Chile Panama China Peru Colombia Poland Costa Rica Portugal Croatia R¿union Cyprus Romania Czech Republic Russian Federation Denmark Saudi Arabia Dominican Republic Singapore Egypt Slovakia Estonia Slovenia Finland South Africa France Spain Germany Sri Lanka Greece State of Palestine Hong Kong ...
- How much
- Total 791 Traveler (0 US and 791 OUS)
Class I
NC TREK Coronary Dilatation Catheter, REF 1012455-08 Rx Only, NC TREK RX 5.00 X 8MM BDC The NC TREK RX Cor
Z-1136-2020 · Class I · reported 2020-02-26 · Terminated
- Why it was recalled
- Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
- Product
- NC TREK Coronary Dilatation Catheter, REF 1012455-08 Rx Only, NC TREK RX 5.00 X 8MM BDC The NC TREK RX Coronary Dilatation Catheter is indicated for: a) balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis, for the purpose of improving myocardial perfusion b) balloon dilatation of a coronary artery occlusion, for the purpose of restoring coronary flow in patients with ST-segment elevation myocardial infarction c) balloon dilatation of a stent after implantation (balloon models 2.00 mm 5.00 mm only)
- Where it went
- Worldwide Distribution: US (Nation-wide) and OUS (countries of): Algeria Jordan Argentina Korea Armenia Kuwait Aruba Lebanon Australia Malaysia Austria, Mauritius Bahrain Mexico Bangladesh Morocco Belgium Netherlands Brazil New Zealand Brunei Norway Bulgaria Oman Canada Pakistan Chile Panama China Peru Colombia Poland Costa Rica Portugal Croatia R¿union Cyprus Romania Czech Republic Russian Federation Denmark Saudi Arabia Dominican Republic Singapore Egypt Slovakia Estonia Slovenia Finland South Africa France Spain Germany Sri Lanka Greece State of Palestine Hong Kong ...
- How much
- Total 39687 NC TREK units (13891 US and 25796 OUS)
Class I
NC TREK Coronary Dilatation Catheter, REF 1012453-20 Rx Only, NC TREK RX 4.00 X 20MM BDC The NC TREK RX Co
Z-1131-2020 · Class I · reported 2020-02-26 · Terminated
- Why it was recalled
- Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
- Product
- NC TREK Coronary Dilatation Catheter, REF 1012453-20 Rx Only, NC TREK RX 4.00 X 20MM BDC The NC TREK RX Coronary Dilatation Catheter is indicated for: a) balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis, for the purpose of improving myocardial perfusion b) balloon dilatation of a coronary artery occlusion, for the purpose of restoring coronary flow in patients with ST-segment elevation myocardial infarction c) balloon dilatation of a stent after implantation (balloon models 2.00 mm 5.00 mm only)
- Where it went
- Worldwide Distribution: US (Nation-wide) and OUS (countries of): Algeria Jordan Argentina Korea Armenia Kuwait Aruba Lebanon Australia Malaysia Austria, Mauritius Bahrain Mexico Bangladesh Morocco Belgium Netherlands Brazil New Zealand Brunei Norway Bulgaria Oman Canada Pakistan Chile Panama China Peru Colombia Poland Costa Rica Portugal Croatia R¿union Cyprus Romania Czech Republic Russian Federation Denmark Saudi Arabia Dominican Republic Singapore Egypt Slovakia Estonia Slovenia Finland South Africa France Spain Germany Sri Lanka Greece State of Palestine Hong Kong ...
- How much
- Total 39687 NC TREK units (13891 US and 25796 OUS)
Class I
NC TREK Coronary Dilatation Catheter, REF 1012454-15 Rx Only, NC TREK RX 4.50 X 15MM BDC The NC TREK RX Co
Z-1134-2020 · Class I · reported 2020-02-26 · Terminated
- Why it was recalled
- Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
- Product
- NC TREK Coronary Dilatation Catheter, REF 1012454-15 Rx Only, NC TREK RX 4.50 X 15MM BDC The NC TREK RX Coronary Dilatation Catheter is indicated for: a) balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis, for the purpose of improving myocardial perfusion b) balloon dilatation of a coronary artery occlusion, for the purpose of restoring coronary flow in patients with ST-segment elevation myocardial infarction c) balloon dilatation of a stent after implantation (balloon models 2.00 mm 5.00 mm only)
- Where it went
- Worldwide Distribution: US (Nation-wide) and OUS (countries of): Algeria Jordan Argentina Korea Armenia Kuwait Aruba Lebanon Australia Malaysia Austria, Mauritius Bahrain Mexico Bangladesh Morocco Belgium Netherlands Brazil New Zealand Brunei Norway Bulgaria Oman Canada Pakistan Chile Panama China Peru Colombia Poland Costa Rica Portugal Croatia R¿union Cyprus Romania Czech Republic Russian Federation Denmark Saudi Arabia Dominican Republic Singapore Egypt Slovakia Estonia Slovenia Finland South Africa France Spain Germany Sri Lanka Greece State of Palestine Hong Kong ...
- How much
- Total 39687 NC TREK units (13891 US and 25796 OUS)
Class I
NC TREK Coronary Dilatation Catheter, REF 1012453-12 Rx Only, NC TREK RX 4.00 X 12MM BDC The NC TREK RX Co
Z-1129-2020 · Class I · reported 2020-02-26 · Terminated
- Why it was recalled
- Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
- Product
- NC TREK Coronary Dilatation Catheter, REF 1012453-12 Rx Only, NC TREK RX 4.00 X 12MM BDC The NC TREK RX Coronary Dilatation Catheter is indicated for: a) balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis, for the purpose of improving myocardial perfusion b) balloon dilatation of a coronary artery occlusion, for the purpose of restoring coronary flow in patients with ST-segment elevation myocardial infarction c) balloon dilatation of a stent after implantation (balloon models 2.00 mm 5.00 mm only)
- Where it went
- Worldwide Distribution: US (Nation-wide) and OUS (countries of): Algeria Jordan Argentina Korea Armenia Kuwait Aruba Lebanon Australia Malaysia Austria, Mauritius Bahrain Mexico Bangladesh Morocco Belgium Netherlands Brazil New Zealand Brunei Norway Bulgaria Oman Canada Pakistan Chile Panama China Peru Colombia Poland Costa Rica Portugal Croatia R¿union Cyprus Romania Czech Republic Russian Federation Denmark Saudi Arabia Dominican Republic Singapore Egypt Slovakia Estonia Slovenia Finland South Africa France Spain Germany Sri Lanka Greece State of Palestine Hong Kong ...
- How much
- Total 39687 NC TREK units (13891 US and 25796 OUS)
Class I
NC TREK Coronary Dilatation Catheter, REF 1012454-20 Rx Only, NC TREK RX 4.50 X 20MM BDC The NC TREK RX Co
Z-1135-2020 · Class I · reported 2020-02-26 · Terminated
- Why it was recalled
- Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
- Product
- NC TREK Coronary Dilatation Catheter, REF 1012454-20 Rx Only, NC TREK RX 4.50 X 20MM BDC The NC TREK RX Coronary Dilatation Catheter is indicated for: a) balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis, for the purpose of improving myocardial perfusion b) balloon dilatation of a coronary artery occlusion, for the purpose of restoring coronary flow in patients with ST-segment elevation myocardial infarction c) balloon dilatation of a stent after implantation (balloon models 2.00 mm 5.00 mm only)
- Where it went
- Worldwide Distribution: US (Nation-wide) and OUS (countries of): Algeria Jordan Argentina Korea Armenia Kuwait Aruba Lebanon Australia Malaysia Austria, Mauritius Bahrain Mexico Bangladesh Morocco Belgium Netherlands Brazil New Zealand Brunei Norway Bulgaria Oman Canada Pakistan Chile Panama China Peru Colombia Poland Costa Rica Portugal Croatia R¿union Cyprus Romania Czech Republic Russian Federation Denmark Saudi Arabia Dominican Republic Singapore Egypt Slovakia Estonia Slovenia Finland South Africa France Spain Germany Sri Lanka Greece State of Palestine Hong Kong ...
- How much
- Total 39687 NC TREK units (13891 US and 25796 OUS)
Class I
NC TRAVELER Coronary Dilatation Catheter, REF 1013157-08, Rx Only,NC Traveler RX 4.0 X 8MM The NC Traveler
Z-1140-2020 · Class I · reported 2020-02-26 · Terminated
- Why it was recalled
- Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
- Product
- NC TRAVELER Coronary Dilatation Catheter, REF 1013157-08, Rx Only,NC Traveler RX 4.0 X 8MM The NC Traveler RX Coronary Dilatation Catheter is not cleared for commercial distribution in the United States.
- Where it went
- Worldwide Distribution: US (Nation-wide) and OUS (countries of): Algeria Jordan Argentina Korea Armenia Kuwait Aruba Lebanon Australia Malaysia Austria, Mauritius Bahrain Mexico Bangladesh Morocco Belgium Netherlands Brazil New Zealand Brunei Norway Bulgaria Oman Canada Pakistan Chile Panama China Peru Colombia Poland Costa Rica Portugal Croatia R¿union Cyprus Romania Czech Republic Russian Federation Denmark Saudi Arabia Dominican Republic Singapore Egypt Slovakia Estonia Slovenia Finland South Africa France Spain Germany Sri Lanka Greece State of Palestine Hong Kong ...
- How much
- Total 791 Traveler (0 US and 791 OUS)
Class I
NC TREK Coronary Dilatation Catheter, REF 1012455-120 Rx Only,NC TREK RX 5.00 X 20MM BDC The NC TREK RX Co
Z-1139-2020 · Class I · reported 2020-02-26 · Terminated
- Why it was recalled
- Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
- Product
- NC TREK Coronary Dilatation Catheter, REF 1012455-120 Rx Only,NC TREK RX 5.00 X 20MM BDC The NC TREK RX Coronary Dilatation Catheter is indicated for: a) balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis, for the purpose of improving myocardial perfusion b) balloon dilatation of a coronary artery occlusion, for the purpose of restoring coronary flow in patients with ST-segment elevation myocardial infarction c) balloon dilatation of a stent after implantation (balloon models 2.00 mm 5.00 mm only)
- Where it went
- Worldwide Distribution: US (Nation-wide) and OUS (countries of): Algeria Jordan Argentina Korea Armenia Kuwait Aruba Lebanon Australia Malaysia Austria, Mauritius Bahrain Mexico Bangladesh Morocco Belgium Netherlands Brazil New Zealand Brunei Norway Bulgaria Oman Canada Pakistan Chile Panama China Peru Colombia Poland Costa Rica Portugal Croatia R¿union Cyprus Romania Czech Republic Russian Federation Denmark Saudi Arabia Dominican Republic Singapore Egypt Slovakia Estonia Slovenia Finland South Africa France Spain Germany Sri Lanka Greece State of Palestine Hong Kong ...
- How much
- Total 39687 NC TREK units (13891 US and 25796 OUS)
Class I
NC TREK Coronary Dilatation Catheter, REF 1012453-08 Rx Only, NC TREK RX 4.00 X 8MM BDC The NC TREK RX Cor
Z-1128-2020 · Class I · reported 2020-02-26 · Terminated
- Why it was recalled
- Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
- Product
- NC TREK Coronary Dilatation Catheter, REF 1012453-08 Rx Only, NC TREK RX 4.00 X 8MM BDC The NC TREK RX Coronary Dilatation Catheter is indicated for: a) balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis, for the purpose of improving myocardial perfusion b) balloon dilatation of a coronary artery occlusion, for the purpose of restoring coronary flow in patients with ST-segment elevation myocardial infarction c) balloon dilatation of a stent after implantation (balloon models 2.00 mm 5.00 mm only)
- Where it went
- Worldwide Distribution: US (Nation-wide) and OUS (countries of): Algeria Jordan Argentina Korea Armenia Kuwait Aruba Lebanon Australia Malaysia Austria, Mauritius Bahrain Mexico Bangladesh Morocco Belgium Netherlands Brazil New Zealand Brunei Norway Bulgaria Oman Canada Pakistan Chile Panama China Peru Colombia Poland Costa Rica Portugal Croatia R¿union Cyprus Romania Czech Republic Russian Federation Denmark Saudi Arabia Dominican Republic Singapore Egypt Slovakia Estonia Slovenia Finland South Africa France Spain Germany Sri Lanka Greece State of Palestine Hong Kong ...
- How much
- Total 39687 NC TREK units (13891 US and 25796 OUS)
Class I
NC TRAVELER Coronary Dilatation Catheter, REF 1013158-12, Rx Only,NC Traveler RX 4.5 X 12MM The NC Traveler
Z-1144-2020 · Class I · reported 2020-02-26 · Terminated
- Why it was recalled
- Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
- Product
- NC TRAVELER Coronary Dilatation Catheter, REF 1013158-12, Rx Only,NC Traveler RX 4.5 X 12MM The NC Traveler RX Coronary Dilatation Catheter is not cleared for commercial distribution in the United States.
- Where it went
- Worldwide Distribution: US (Nation-wide) and OUS (countries of): Algeria Jordan Argentina Korea Armenia Kuwait Aruba Lebanon Australia Malaysia Austria, Mauritius Bahrain Mexico Bangladesh Morocco Belgium Netherlands Brazil New Zealand Brunei Norway Bulgaria Oman Canada Pakistan Chile Panama China Peru Colombia Poland Costa Rica Portugal Croatia R¿union Cyprus Romania Czech Republic Russian Federation Denmark Saudi Arabia Dominican Republic Singapore Egypt Slovakia Estonia Slovenia Finland South Africa France Spain Germany Sri Lanka Greece State of Palestine Hong Kong ...
- How much
- Total 791 Traveler (0 US and 791 OUS)
Class I
NC TREK Coronary Dilatation Catheter, REF 1012455-12 Rx Only, NC TREK RX 5.00 X 12MM BDC The NC TREK RX Co
Z-1137-2020 · Class I · reported 2020-02-26 · Terminated
- Why it was recalled
- Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
- Product
- NC TREK Coronary Dilatation Catheter, REF 1012455-12 Rx Only, NC TREK RX 5.00 X 12MM BDC The NC TREK RX Coronary Dilatation Catheter is indicated for: a) balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis, for the purpose of improving myocardial perfusion b) balloon dilatation of a coronary artery occlusion, for the purpose of restoring coronary flow in patients with ST-segment elevation myocardial infarction c) balloon dilatation of a stent after implantation (balloon models 2.00 mm 5.00 mm only)
- Where it went
- Worldwide Distribution: US (Nation-wide) and OUS (countries of): Algeria Jordan Argentina Korea Armenia Kuwait Aruba Lebanon Australia Malaysia Austria, Mauritius Bahrain Mexico Bangladesh Morocco Belgium Netherlands Brazil New Zealand Brunei Norway Bulgaria Oman Canada Pakistan Chile Panama China Peru Colombia Poland Costa Rica Portugal Croatia R¿union Cyprus Romania Czech Republic Russian Federation Denmark Saudi Arabia Dominican Republic Singapore Egypt Slovakia Estonia Slovenia Finland South Africa France Spain Germany Sri Lanka Greece State of Palestine Hong Kong ...
- How much
- Total 39687 NC TREK units (13891 US and 25796 OUS)
Class I
NC TREK Coronary Dilatation Catheter, REF 1012455-15 Rx Only,NC TREK RX 5.00 X 15MM BDC The NC TREK RX Cor
Z-1138-2020 · Class I · reported 2020-02-26 · Terminated
- Why it was recalled
- Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
- Product
- NC TREK Coronary Dilatation Catheter, REF 1012455-15 Rx Only,NC TREK RX 5.00 X 15MM BDC The NC TREK RX Coronary Dilatation Catheter is indicated for: a) balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis, for the purpose of improving myocardial perfusion b) balloon dilatation of a coronary artery occlusion, for the purpose of restoring coronary flow in patients with ST-segment elevation myocardial infarction c) balloon dilatation of a stent after implantation (balloon models 2.00 mm 5.00 mm only)
- Where it went
- Worldwide Distribution: US (Nation-wide) and OUS (countries of): Algeria Jordan Argentina Korea Armenia Kuwait Aruba Lebanon Australia Malaysia Austria, Mauritius Bahrain Mexico Bangladesh Morocco Belgium Netherlands Brazil New Zealand Brunei Norway Bulgaria Oman Canada Pakistan Chile Panama China Peru Colombia Poland Costa Rica Portugal Croatia R¿union Cyprus Romania Czech Republic Russian Federation Denmark Saudi Arabia Dominican Republic Singapore Egypt Slovakia Estonia Slovenia Finland South Africa France Spain Germany Sri Lanka Greece State of Palestine Hong Kong ...
- How much
- Total 39687 NC TREK units (13891 US and 25796 OUS)
Class I
NC TREK Coronary Dilatation Catheter, REF 1012453-15 Rx Only, NC TREK RX 4.00 X 15MM BDC The NC TREK RX Co
Z-1130-2020 · Class I · reported 2020-02-26 · Terminated
- Why it was recalled
- Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
- Product
- NC TREK Coronary Dilatation Catheter, REF 1012453-15 Rx Only, NC TREK RX 4.00 X 15MM BDC The NC TREK RX Coronary Dilatation Catheter is indicated for: a) balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis, for the purpose of improving myocardial perfusion b) balloon dilatation of a coronary artery occlusion, for the purpose of restoring coronary flow in patients with ST-segment elevation myocardial infarction c) balloon dilatation of a stent after implantation (balloon models 2.00 mm 5.00 mm only)
- Where it went
- Worldwide Distribution: US (Nation-wide) and OUS (countries of): Algeria Jordan Argentina Korea Armenia Kuwait Aruba Lebanon Australia Malaysia Austria, Mauritius Bahrain Mexico Bangladesh Morocco Belgium Netherlands Brazil New Zealand Brunei Norway Bulgaria Oman Canada Pakistan Chile Panama China Peru Colombia Poland Costa Rica Portugal Croatia R¿union Cyprus Romania Czech Republic Russian Federation Denmark Saudi Arabia Dominican Republic Singapore Egypt Slovakia Estonia Slovenia Finland South Africa France Spain Germany Sri Lanka Greece State of Palestine Hong Kong ...
- How much
- Total 39687 NC TREK (13891 US and 25796 OUS)
Abbott MitraClip XTR Clip Delivery System, UDI: 08717648226366, Part: CDS0601-XTR, Sterile EO, Rx Only
Z-1687-2019 · Class II · reported 2019-06-05 · Terminated
- Why it was recalled
- Reports of implantable mitral valve repair system clips unexpectedly opening and becoming nonfunctional, resulting from unintended excessive force applied during the Clip implantation. Excessive force can also result in unexpected movement of clip arms. The inability to close and remove the device has lead to surgery and additional intervention.
- Product
- Abbott MitraClip XTR Clip Delivery System, UDI: 08717648226366, Part: CDS0601-XTR, Sterile EO, Rx Only
- Where it went
- U.S.: FL, UT, IN, OK, TX, CO, LA, AZ, MO, AL, GA, CA, AR, KS, IL, HI, IA, KY, MA, DC, NY, NV, NM, NJ, MT, MD, NC, CT, MI, MN, OH, ND, NH, OR, PA, WI, WA, VA, WV, TN, SC, RI, DE, ME, NE, SD, MS, ID, VT. Foreign (OUS): AUSTRALIA, AUSTRIA, BELGIUM, BULGARIA, Canada, CROATIA, CZECH REPUBLIC, DENMARK, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, HUNGARY, INDONESIA, IRELAND, ISRAEL, ITALY, KUWAIT, LEBANON, MACAU, NETHERLANDS, NEW ZEALAND, NORWAY, POLAND, PORTUGAL, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SPAIN, SWEDEN, SWITZERLAND, THAILAND, TURKEY, UNITED KINGDOM
- How much
- N/A
PLUS 30 PRIORITY PACK Accessory Kit Product Usage: Is recommended for use during vascular procedures in co
Z-0960-2019 · Class II · reported 2019-03-06 · Terminated
- Why it was recalled
- Incorrect expiration being entered for one lot.
- Product
- PLUS 30 PRIORITY PACK Accessory Kit Product Usage: Is recommended for use during vascular procedures in conjunction with interventional and / or diagnostic devices (e.g., balloon dilatation catheters, atherectomy devices, stent delivery systems, intravascular ultrasound devices).
- Where it went
- US Nationwide Distribution - NC and NY
- How much
- 26 units
Class I
Absorb Bioresorbable Vascular Scaffold (BVS) system (Australia). The Absorb BVS is a temporary scaffold indic
Z-2151-2017 · Class I · reported 2017-06-07 · Terminated
- Why it was recalled
- Abbott Vascular is initiating a recall of the Absorb Bioresorbable Vascular Scaffold (BVS) System due to studies showing elevated rates of major adverse events, specifically, myocardial infraction and scaffold thrombosis when compared to patients treated with the Xience metallic drug eluting stent.
- Product
- Absorb Bioresorbable Vascular Scaffold (BVS) system (Australia). The Absorb BVS is a temporary scaffold indicated for improving coronary luminal diameter that will eventually resorb and potentially facilitate normalization of vessel function in patients with ischemic heart disease due to de novo native coronary artery lesions. The treated lesion length should be less than the nominal scaffolding length (8 mm, 12 mm, 18 mm, 23 mm, 28 mm) with reference vessel diameters e 2.0 mm and d 3.8 mm.
- Where it went
- Australia
- How much
- 590 units
Showing the 30 most recent of 37. The rest are searchable at the FDA enforcement report database.
Read these counts with three things in mind
A firm with more recalls usually makes more products. A contract packer runs
hundreds of lines. It will out-recall a company with one product whatever either
does. This is not a safety ranking.
Most recalls here are voluntary. A firm found its own fault and pulled the stock.
That is the system working. It lands in this database looking the same as a recall
the FDA had to push for.
Terminated means the FDA considers the recall finished. It is not a verdict.
It does not mean anyone was harmed.
Source: openFDA enforcement reports, retrieved 2026-08-05. FDA dataset last updated 2026-07-29.
← All device recalls